CHEM 334 Quantitative Analysis Laboratory

Similar documents
Volumetric Measurement Techniques. Technique #1 Use of a Burette. Technique #2 Use of a Pipette. Technique #3 Use of a Volumetric Flask

Experimental Procedure. Lab 406

Experiment 8 Introduction to Volumetric Techniques I. Objectives

Density of an Unknown

Chemistry Calibration of a Pipet and Acid Titration

Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE

H 3 O + (aq) + P 2- (aq)

Name: Date: AP Chemistry. Titrations - Volumetric Analysis. Steps for Solving Titration Problems

NaOH (aq) + HCl (aq) NaCl (aq) + H 2 O (l)

CHEM 334 Quantitative Analysis Laboratory

TRATION: ANALYSIS OF VINE

Acid-Base Titration. M M V a

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Titration of Acetic Acid in Vinegar Minneapolis Community and Technical College v.9.17

Introduction to Spectroscopy: Analysis of Copper Ore

Experiment 20: Analysis of Vinegar. Materials:

Titration with an Acid and a Base

Experiment 20-Acid-Base Titration: Standardization of KOH and Determination of the Molarity and/or Percent Composition of an Acid Solution

ANALYTICAL CHEMISTRY. Basic Skills in Chemical Laboratory

Practice Lab. Balances and calibration of volumetric tools

Accuracy and Precision of Laboratory Glassware: Determining the Density of Water

Experiment 7: ACID-BASE TITRATION: STANDARDIZATION OF A SOLUTION

The Synthesis and Analysis of Aspirin

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

Basic Equipment. 2D.1 Equipment for Measuring Mass

POGIL LAB EXERCISE 15 HOW DO YOU STANDARDIZE AN ACID AND BASE?

Calibration of Volumetric Glassware. Prepared by Allan Fraser May 2016 APPLICATION Note 1

Australian National Titration Competition. Basics of Titration

#09 Investigating the Relationship between the Mass of a Liquid and its Volume Ken Lyle, St. John s School, Houston, TX

Experiment: Titration

Titration of an Unknown Acid

CHEM Practice to be done before the lab. Experiment 9 Introduction to Volumetric Techniques II. Objectives

TITRATION OF AN ACID WITH A BASE

Density of Aqueous Sodium Chloride Solutions

Introduction to Spectroscopy: Analysis of Copper Ore

Chemical Reactions: Titrations

Learn to do quantitative titration reactions. Observe the mole ratios of several simple chemical reactions.

Determination of the K a Value and Molar Mass of an Unknown Weak Acid

EXPERIMENT #8 Acid-Base I: Titration Techniques

PURPOSE: 1. To illustrate an oxidation-reduction titration with potassium permanganate 2. To determine the percent mass of iron in an unknown.

Experiment 1. Determination of the Density of Water and an Unknown Solid Sample

Introduction to Small Scale Chemistry

Ascorbic Acid Titration of Vitamin C Tablets

Basic Equipments and Instruments used in Chemistry laboratory: Balance: It is an instrument for measuring mass.

DETERMINATION OF THE SOLUBILITY PRODUCT OF GROUPII HYDROXIDES

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15

Density of Aqueous Sodium Chloride Solutions

In this laboratory exercise we will determine the percentage Acetic Acid (CH 3 CO 2 H) in Vinegar.

GENERAL INSTRUCTIONS GENERAL PREPARATION

Checklist for Titration

SOLVOLYSIS OF tert-butyl CHLORIDE: TESTING A MECHANISM

Safety Note: Safety glasses and laboratory coats are required when performing this experiment

NOTE: YOU WILL BE USING THIS SOLUTION IN BOTH, THIS EXPERIMENT AND EXP 12B. IF YOU WASTE THE SOLUTION YOU MAY RUN OUT BEFORE YOU HAVE FINISHED EXP 12B

Assessment of Accuracy and Precision

Chemistry 143 Acid Base Titration Dr. Caddell. Titrating Acid

EXPERIMENT 22 SOLUBILITY OF A SLIGHTLY SOLUBLE ELECTROLYTE

Introduction to Chemistry Techniques Prelab (Week 1) 2. Determine the number of significant figures in each of the following numbers.

+ H 2 O Equation 1. + NaOH CO 2 Na

Chemistry Lab: Introduction to Measurement

2 burets (50 ml) Standard solution of NaOH (0.600 M) Phenolphthalein indicator

Titrations Worksheet and Lab

Identification Of The Common Laboratory Glassware, Pipettes And Equipment. BCH 312 [Practical]

Experiment 7: Titration of an Antacid

Introduction to Spectroscopy: Analysis of Copper Ore

6. Common Laboratory Techniques

Chemistry 143 Experiment #11 Acid Base Titration Dr. Caddell. Titrating Acid

Acid-Base Titrations

Chemistry 101 Experiment 12B-ANALYSIS OF COMMERCIAL VINEGARS

Potentiometric measurement of ph

Introduction to Spectroscopy: Analysis of Copper Ore

Experiment #7. Titration of Vinegar

GENERAL INSTRUCTIONS GENERAL PREPARATION

To see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide:

Experiment #10: Analysis of Antacids

(b) Write the chemical equation for the dissolution of NaOH in water.

Citric Acid Analysis L-6-1. MOISTURE (Karl Fischer)

Rate law Determination of the Crystal Violet Reaction Using the Isolation Method

Molarity of Acetic Acid in Vinegar A Titration Experiment

Corn Sugar (crude & refined) Analysis F MOISTURE (Karl Fischer)

Experiment 2: Analysis of Commercial Bleach Solutions

POS Checklist. Preparing solutions: Dilution. vs. Dissolving vs. Dilution. Ice Tea. 5 Preparing Solutions Dilution.notebook.

CHM111 Lab Titration of Vinegar Grading Rubric

6 Acid Base Titration

MEASUREMENT: PART II

Conductometric Titration & Gravimetric Determination of a Precipitate

Standardizing a Solution of Sodium Hydroxide. Evaluation copy

The Thermodynamics of the Solubility of Borax

Experimental Procedure Lab 402

The floating egg problem

Experiment 10. Acid Base Titration

2. MASS AND VOLUME MEASUREMENTS Mass measurement Analytical and standard laboratory balances Pre-lab Exercises

Final Concentration 0 excess 0.1 M 0.1 M

PURPOSE: To determine the Rate Law for the following chemical reaction:

Experimental Procedure

Determination of Orthophosphate Ion

Titrimetric Determination of EASTMAN Color Developing Agent, CD-3, in Process ECN-2 Developer with Sulfato Cerate

Spectrophotometric Determination of Ferrocyanide in Effluents

Analysis of Hypochlorite in Bleach

This chapter contains detailed directions for performing a variety of chemical

Practice Examination #1

Transcription:

Calibration of Volumetric Glassware Introduction Volumetric glassware is a class of glass vessels that are calibrated to contain or deliver certain volumes of substances. Graduated cylinders, pipettes (also pipets), burettes (also burets) and volumetric flasks are examples as are beakers and Erlenmeyer flasks. It can be seen from this list that the term "calibrated" can mean "to a varying degree". The manufacturer's stated accuracy of beakers and Erlenmeyer flasks is typically 5% while that for volumetric flasks is generally much higher. In this laboratory experiment, the three types of glassware typically used by an analytical chemist: a volumetric flask, a volumetric pipet and a buret will be calibrated. These tools are used extensively when performing gravimetric and titrimetric Analyses. In order to avoid introducing systematic errors into measurements, each of these instruments must be properly calibrated. And, to reduce the random errors inherent when using these instruments, their proper use must be thoroughly understood. The quality of the measurements obtained from these tools depends heavily on the care taken in calibrating and in using each instrument. Procedures Figure 2. Volumetric and Graduated Pipettes. Figure 1. Volumetric flask. Figure 3. Burette. In the most precise work it is never safe to assume that the volume delivered by or contained in any volumetric instrument is exactly the amount indicated by the calibration mark. Instead, the amount of water delivered by or contained in the volumetric apparatus is quantified by mass to effect the calibration. This mass is then converted to the desired volume using the tabulated density of water and the following standard relation: d = m / v (1) where v is measured volume, m is mass of water used to fill the volume and d is density of the water. Colorado State University 2017.01.25 Page 1 of 6

Calibration of Volumetric Glassware All volumetric apparatus is either purchased with a Calibration Certificate or calibrated by a chemist in this fashion. Systematic Errors Affecting Volumetric Measurements The volume occupied by a given mass of liquid varies with temperature, as does the volume of the device that holds the liquid. 20 C has been chosen as the normal temperature for calibration of much volumetric glassware. Glass is a fortunate choice for volumetric ware as it has a relatively small coefficient of thermal expansion; a glass vessel that holds 1.00000 L at 15 C holds 1.00025 L at 25 C. If desired, the volume (V) obtained at a temperature (T) can be corrected to 20 C by use of: V20 C = VT [1 + 0.000025 (20 T)] (2) In most work, this correction is small enough it may be ignored. However, the thermal expansion of the contained liquid is frequently of importance. Dilute aqueous -1 solutions have a coefficient of volume thermal expansion of about 0.00025 C. Thus, 1.000 L of water at 15 C will occupy 1.002 L at 25 C. A correction for this expansion is often applied in rigorous calibration procedures. Parallax is another source of error when using volumetric ware. A correction for this expansion must frequently be applied during calibration procedures. Frequently, graduation marks encircle the girth of the apparatus to aid in preventing this artifact by permitting the chemist to align the line of sight. Figure 4. Parallax error from incorrect Lines of sight. Figure 5. The meniscus of water. Readings that are taken from a line of sight that is either too high or too low will result otherwise. Tips for Correct Use of Volumetric Glassware Pipettes. The pipette is used to transfer a volume of solution from one container to another. Most volumetric pipettes are calibrated To-Deliver (TD) with a certain amount of the liquid remaining in the tip and as a film along the inner barrel after delivery of the liquid. The liquid in the tip should not be blownout. Pipets of the "blow-out" variety will usually have a ground glass ring at the top. And, drainage rates Colorado State University 2017.01.25 Page 2 of 6

from the pipet must be carefully controlled so as to leave a uniform and reproducible film along the inner glass surface. Measuring Pipets will be gradated in appropriate units. Once the pipet is cleaned and ready to use, make sure the outside of the tip is dry. Then rinse the pipet with the solution to be transferred. Insert the tip into the liquid to be used and draw enough of the liquid into the pipet to fill a small portion of the bulb. Hold the liquid in the bulb by placing your fore finger over the end of the stem. Withdraw the pipet from the liquid and gently rotate it at an angle so as to wet all portions of the bulb. Drain out and discard the rinsing liquid. Repeat this once more. To fill the pipet, insert it vertically in the liquid, with the tip near the bottom of the container. Apply suction to draw the liquid above the graduation mark. Quickly place a forefinger over the end of the stem. Withdraw the pipet from the liquid and use a dry paper to wipe off the stem. Now place the tip of the pipet against the container from which the liquid has been withdrawn and drain the excess liquid such that the meniscus is at the graduation mark. Figure 5. Pipet usage Move the pipet to the receiving container and allow the liquid to flow out (avoiding splashing) of the pipet freely. When most of the liquid has drained from the pipet, touch the tip to the wall of the container until the flow stops and for an additional count of ten. Volumetric Flasks. The volumetric flask is used to prepare Standard Solutions or in diluting a sample. Most of these flasks are calibrated To-Contain (TC) a given volume of liquid. When using a flask, the solution or solid to be diluted is added and solvent is added until the flask is about two-thirds full. It is important to rinse down any solid or liquid that has adhered to the neck. Swirl the solution until it is thoroughly mixed. Now add solvent until the meniscus is at the calibration mark. If any droplets of solvent adhere to the neck, use a piece of tissue to blot these out. Stopper the flask securely and invert the flask at least ten times. Burets. The Buret is used to accurately deliver a variable amount of liquid. Fill the buret to above the zero mark and open the stopcock to fill the tip. Work air bubbles out of the tip by rapidly squirting the liquid through the tip or tapping the tip while solution is draining. The initial buret reading is taken a few seconds, ten to twenty, after the drainage of liquid has ceased. Holding a white piece of paper with a heavy black mark on behind the buret can highlight the meniscus. Place the flask into which the liquid is to be drained on a white piece of paper. (This is done during a titration to help visualize color changes that occur during the titration.) The flask is swirled with the righthand while the stopcock is manipulated with the left- hand. The buret is opened and allowed to drain freely until near the point where liquid will no longer be added to the flask. Smaller additions are made as the end-point of the addition is neared. Allow a few seconds after closing the stopcock before making any readings. At the end-point, read the buret in a manner similar to that above. Colorado State University 2017.01.25 Page 3 of 6

As with pipets, drainage rates must be controlled so as to provide a reproducible liquid film along the inner barrel of the buret. Cleaning of Volumetric Glassware. Cleaning of volumetric glassware is necessary to not only remove any contaminants, but to ensure its accurate use. The film of water that adheres to the inner glass wall of a container as it is emptied must be uniform. Two or three rinses with tap water, a moderate amount of agitation with a dilute detergent solution, several rinses with tap water, and two or three rinses with distilled water are generally sufficient if the glassware is emptied and cleaned immediately after use. If needed, use a warm detergent solution (60-70 C). A buret or test tube brush can be used in the cleaning of burets and the neck of volumetric flasks. Volumetric flasks can be filled with cleaning solution directly. Pipets and burets are filled by inverting them and drawing the cleaning solution into the device using suction. Avoid getting cleaning solution in the stopcock. Allow the warm cleaning solution to stand in the device for about fifteen minutes; never longer than 20 minutes. Drain the cleaning solution and rinse thoroughly with tap water and finally two or three times with distilled water. Pipets and burets should be rinsed at least once with the solution they are to filled with before use. A General Calibration Procedure As was noted above, volumetric glassware is calibrated by measuring the mass of water that is containedin or delivered-by the device. This mass data is then converted to volume data using the tabulated density of water (See Appendix) at the temperature of calibration. (In very accurate work, the thermometer is calibrated also as an incorrect temperature reading will lead to the use of an incorrect density for water. This, in turn, will give an inaccurate volume calibration.) Finally, this volume data is corrected to the standard temperature of 20 C. This can be accomplished using (2) to describe the thermal volume expansion coefficient of water. Further details concerning calibration of laboratory glassware can be found in the NIST publication identified in the references. In this experiment a volumetric flask, a measuring pipette and a volumetric pipette will be using gravimetrically determined water. In each case, the measured mass of the calibrating water will be standardized to 20 C. Procedure Record the laboratory temperature. For this experiment use distilled water contained in one of the carboys - not directly from the distilled water tap. (Why?) Clean and dry (in a 105 C oven for a few minutes, followed by cooling to room temperature) a weighing bottle and its cap Use detergent if necessary followed by tap water then distilled water (also from its own tap, why?) rinses. Use a 105 C oven for a few minutes for drying and cool to room temperature. Do not handle the clean glassware with bare hands. (Why?) Colorado State University 2017.01.25 Page 4 of 6

Measuring Pipette. Fill a 10 ml measuring pipette with deionized water, then drain it to ensure that no droplets remain on the inner walls. If droplets remain, clean the pipette thoroughly and repeat. Fill the pipette and deliver approximately two milliliters into the weighed bottle. Estimate all volume readings to the nearest 0.01 ml. Cap the bottle and reweigh. Repeat this procedure for samples of 4, 6 and 8 ml. Perform three replicate measurements for the pipette. Volumetric Pipette. Fill a 10 ml volumetric pipette with deionized water and drain to ensure that no droplets remain on the inner walls. If droplets remain, clean the pipette thoroughly and repeat. Fill the pipette to the fill line and deliver the total volume of liquid into a dry bottle. Cap the weighing bottle and reweigh. Perform three replicate measurements for the pipette. Volumetric Flask. Calibrate a 10 ml volumetric flask by weighing it empty (and dry) and filled to the mark with distilled water. Perform three replicate measurements for the flask. Results Volumetric Pipette. Use the density of water to convert the mass of water to the volume delivered by your pipette. Determine the average and 95% confidence interval for the three replicate measurements for the pipette. If the 95% confidence intervals are greater than the tolerance for the 10 ml pipette and if time permits repeat these measurements. Volumetric Flask. Convert the mass of water contained in your volumetric flask to volume. Determine the average volume and 95% confidence interval. If time permits, repeat the calibration if the 95% confidence intervals greater than the tolerance for the 10 ml flask. In the future, use this average value as the exact volume contained in the flask. Graduated Pipette. Convert the mass of water to the volume delivered by the pipette. Plot the average measured volume obtained from your replicate measurements delivered versus the expected volume delivered as read on the pipette for the 10 ml pipette. Plot y-value error bars corresponding to the standard deviation obtained from your replicate measurements. (Can you see the error bars?) Using a linear regression analysis, determine the best-fit equation for the relationship between measured and expected volumes. Include both slope and intercept in the analysis. Report the slope, its standard deviation, the y-intercept, its standard deviation and the R 2 value of the calibration line. Calculate the 95% confidence interval for each value of expected volume. Discussion Prepare a Laboratory Report as described previously. Present all measured data in the report together with the expected data. Describe and explain the relationships between the two sets of data. Include good-quality figures and tables to substantiate your discussions. Include answers to the several "why?" questions that appear in this description. Include a quantitative error analysis in your discussion. Compare your assessment of the various volumetric glassware accuracy values to the manufacturer's stated values. References Tufte, E.R., "Envisioning Information" (1990), Graphic Press, CT, Chapters 1 and 6. Colorado State University 2017.01.25 Page 5 of 6

Pierce, C., Haenisch, E.L. and Sawyer, D.T., "Quantitative Analysis" (1948) Fourth Edition Wiley & Sons, several locations. Lembeck, J., The Calibration of Small Volumetric Laboratory Glassware, NBSIR 74-461. Available at: http://ts.nist.gov/measurementservices/calibrations/upload/74-461.pdf. Harris, D.C., "Quantitative Chemical Analysis" (2007) 7th edition, Freeman & Co., NY. Appendix 1. The density of water as a function of temperature. Temperature ( C) Density (g/ml) 15 0.999 102 6 16 0.998 946 0 17 0.998 777 9 18 0.998 598 6 19 0.998 408 2 20 0.998 207 1 21 0.997 995 5 22 0.997 773 5 23 0.997 541 5 24 0.997 299 5 25 0.997 047 9 26 0.996 786 7 27 0.996 516 2 28 0.996 236 5 29 0.995 947 8 30 0.995 650 2 Colorado State University 2017.01.25 Page 6 of 6