Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Similar documents
Methanol Extraction of high level soil samples by USEPA Method 8260C

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Optimizing. Abstract: is standardd. procedures. altered to

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Solid Phase Micro Extraction of Flavor Compounds in Beer

Analytical Trap Comparison for USEPA Method 8260C

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Optimizing Standard Preparation

Maximizing Sample Throughput In Purge And Trap Analysis

Maximizing Production While Minimizing Costs

Solid Phase Micro Extraction of Flavor Compounds in Beer

2017 EPA Method Update Rule and EPA Method 624.1

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

EPA TO-17 Volatile Organic Compounds

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Ed George and Anaïs Viven Varian, Inc.

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Target Compound Results Summary

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Determination of Volatile Organic Compounds in Air

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Volatile Organic Compounds in Every Day Food

ANALYTICAL REPORT. Job Number: Job Description: Transform Complete

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

CALA Directory of Laboratories

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

ANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

STANDARD OPERATING PROCEDURES

Meeting NJ Low Level TO-15 Air Testing Method Requirements

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Study of Residual Solvents in Various Matrices by Static Headspace

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Associates. Project No.: October 25, 2017

Thermal Desorption Technical Support

Volatile Organic Compounds in Water PBM

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

Standard Operating Procedure for the Analysis of Purgeable Organic Compounds (VOC s) in Water by Capillary Column Gas Chromatography/Mass Spectrometry

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

ANALYTICAL REPORT. Job Number: SDG Number: Job Description: Joint Base Cape Cod

Copies: Dave Favero, RACER File

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Building 280A and Building 450 Soil Sampling Results Richmond Field Station Site Berkeley Global Campus at Richmond Bay, Richmond, California

McCAMPBELL ANALYTICAL INC Willow Pass Road Pittsburg CA

Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic Compounds

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

In-Tube Extraction Sample Preparation for Gas Chromatography

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

Headspace Technology for GC and GC/MS: Features, benefits & applications

Method 1624 Revision C Volatile Organic Compounds by Isotope Dilution GCMS

Date Issued: April 01, 2013 Expiration Date: June 30, 2013

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Table 1 Soil and Groundwater Sampling Summary Table

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

ORGANIC COMPOUNDS IN WATER BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY USING NITROGEN PURGE GAS

Laboratory Report Number(s) Lone Tree Road (Deep) 10/5/15, 10/19/15, 6/13/16. Address Sample Date(s) Analyses Requested

Detection of Environmental Contaminants Caused by the Oil Spill in the Gulf of Mexico by GC and Hplc

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

1. Introduction. 2. Sampling Activities

January 19, Dear Mr. Nightingale:

Transcription:

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260 Application Note Abstract Due to regional shortages and increasing costs of helium, the preferred carrier gas in gas chromatography, alternative carrier gas options are increasing in demand. This study will evaluate the Teledyne Tekmar Atomx VOC Sample Prep System in conjunction with a Thermo Scientific TRACE 1310 / ISQ GCMS while employing hydrogen GC carrier gas and nitrogen purge gas to perform EPA method 8260C. A working linear calibration curve, Method Detection Limits (MDLs), and a BFB tuning evaluation will be presented to determine the validity of alternative helium gases for VOC analysis by USEPA 8260C. Introduction The EPA developed method 8260C to determine volatile organic compounds (VOCs) in soil and water. 1 Helium is the preferred GC carrier gas, as well as the most widely used purge gas. Unfortunately, helium is a nonrenewable gas which has caused issues with supply and increasing costs. Conversely, alternative carrier gases such as hydrogen and nitrogen are renewable and can be derived from generators. Many laboratories are making the conscious effort to cut costs and shift to helium free instruments and methodologies, but must also be able to produce reliable results. This study will evaluate using these renewable gas sources - hydrogen for the GC/MS and nitrogen for the Atomx - as substitutes for helium in EPA Method 8260C. Calibration standards will we evaluated by linear calibration curves per the method criteria. A calibration curve will be generated from 0.5 ppb to 200 ppb, and must meet the criteria of %RSD below 20% or a linear regression value (r²) of at least 0.99. 1 The minimum average response factor is also defined in EPA Method 8260C. MDLs will be assessed for 100% or 1 ppb variance of the actual concentration. Teledyne Tekmar s Atomx is a VOC sample prep system that integrates a Purge and Trap Concentrator with a Multi-Matrix Autosampler. It employs an 80-position vial autosampler capable of running water, soil and automated methanol extractions, and also has the capability of prepping samples ahead, while the previous sample is being analyzed by GC, allowing for faster analysis time. The Atomx has the ability to automatically add Internal Standards (IS) and Surrogate Sample (SS) from three separate standard addition vessels, with volume increment variables of 1, 2, 5, 10, and 20 µl. Experimental-Instrument Conditions For this study, the Atomx VOC Sample Prep System was coupled to a Thermo Scientific TRACE 1310 GC with ISQ MS Detector. The analytical trap used was a #9 (proprietary). Hydrogen was used as the carrier gas and nitrogen as the purge gas. The GC was equipped with a Rtx -VMS 20m x 0.18mm x 1µm column. The GC/MS parameters are outlined in Tables 1 and 2. Table 3 outlines the Atomx conditions.

GC Parameters MS Parameters GC: Thermo Scientific TRACE 1310 MS: ISQ Column: Rtx -VMS 20 m x 0.18 mm ID x 1.0 µm Ion Source: 280 C Oven Program: 35 C for 3 min; 14 C/min to 100 C; 25 to 210; hold for 2 min Solvent Delay: 0.8 min Inlet: 200 C Scan Range: 35-260 Column Flow: 0.8 ml/min Dwell Time: 0.15 sec Gas: Hydrogen Emission Current: 40 µa Split: 60:1 Detector Gain: 3.00e 5 Pressure: MS Transfer Line: 5 psi 230 C Chrom Filter Width: Tables 1 & 2: GC and MS Parameters Atomx 8260C Water Parameters 2.0 sec Variable Value Variable Value Valve oven Temp 140 C Dry Purge Flow 100 ml/min Transfer Line Temp 140 C Dry Purge Temp 20 C Sample Mount Temp 90 C Methanol Needle Rinse Off Water Heater Temp 90 C Methanol Needle Rinse Volume 3.0 ml Sample Vial Temp 20 C Water Needle Rinse Volume 7.0 ml Sample Equilibrate Time 0.00 min Sweep Needle Time 0.25 min Soil Valve Temp 50 C Desorb Preheat Time 245 C Standby Flow 10 ml/min GC Start Signal Start of Desorb Purge Ready Temp 40 C Desorb Time 2 min Condensate Ready Temp 45 C Drain Flow 300 ml/min Presweep Time 0.25 min Desorb Temp 250 C Prime Sample Fill Volume 3.0 ml Methanol Glass rinse Off Sample Volume 5.0 ml Number of Methanol Glass Rinses 1 Sweep Sample Time 0.25 min Methanol Glass Rinse Volume 3.0 ml Sweep Sample Flow 100 ml/min Number of Water Bake Rinses 1 Sparge Vessel Heater OFF Water Bake Rinse Volume 7.0 ml Sparge Vessel Temp 20 C Bake Rinse Sweep Time 0.25 min Prepurge Time 0.00 min Bake Rinse Sweep Flow 100 ml/min Prepurge Flow 0 ml/min Bake Rinse Drain Time 0.40 min Purge Time 11.00 min Bake Time 2.00 min Purge Flow 40 ml/min Bake Flow 400 ml/min Purge Temp 20 C Bake Temp 280 C Condensate Purge Temp 20 C Condensate Bake Temp 200 C Dry Purge Time 0.5 min Table 3: Atomx 8260C Parameters (items in yellow were not used)

Calibration and Method Detection Limits A 50 ppm stock standard was brought to volume with methanol. The stock standard was diluted in deionized water to generate calibration standards from 0.5 ppb to 200 ppb. These standards were transferred to headspace free 40 ml VOA vials. A chromatogram of a 10 ppb calibration standard can be found in Figure 1. Internal Standard (IS) and Surrogate standards (SS) were made in methanol at 50 ppm. IS/SS were dispensed by the internal standard vessels in 5 µl aliquots for a final concentration of 50 ppb. Thermo TraceFinder software was used to process the calibration data for each compound using extracted ion mode. The primary and secondary characteristic ions are prescribed in method 8260C. The relative response factors (RRF) were used to calculate the percent relative standard deviation (%RSD). Full calibration data can be found in Table 4. MDLs were determined by running a 1 ppb standard in seven repetitions. The MDL must be less than 1. The results can be found in table 4. Compound Avg RRF % RSD MDL Compound Avg RRF % RSD MDL Dichlorodifluoromethane 0.275 8.37 0.217 2-Chloroethylvinyl ether 0.313 8.65 0.152 Chloromethane 0.657 7.2 0.105 cis-1,3-dichloropropene 0.853 11.17 0.155 Bromomethane* 0.998 0.207 Toluene 2.389 5.93 0.134 Chloroethane 0.314 6.08 0.296 Tetrachloroethylene 0.576 7.91 0.373 Trichlorofluoromethane 0.631 11.95 0.239 2-Hexanone 1.207 13.23 0.132 Ethyl ether 0.542 11.42 0.148 4-Methyl-2-pentanone 1.155 11.8 0.151 1,1-Dichloroethene 0.838 8.47 0.143 trans 1,3-Dichloropropene 0.782 12.1 0.105 Vinyl Chloride 0.355 6.57 0.251 1,1,2-Trichloroethane 0.4 10.19 0.131 Carbon Disulfide 2.132 7.59 0.204 Ethyl Methacrylate 0.691 14.5 0.112 1,1,2-Trichlorotrifluoroethane 0.607 11.22 0.139 Dibromochloromethane 0.241 11.9 0.181 Acetonitrile 3.106 12.59 0.433 1,3-Dichloropropane 1.015 7.82 0.158 Allyl chloride 3.14 12.49 0.428 1,2-dibromoethane 0.469 9.25 0.19 Methylene Chloride 2.726 6.99 0.204 Butyl acetate 1.552 13.54 0.155 Acetone 0.642 17.62 0.438 Chlorobenzene 1.304 5.29 0.084 trans-1,2-dichloroethene 1.032 7.22 0.09 Ethylbenzene 2.456 5.94 0.095 Methyl acetate 1.768 8.27 0.222 1,1,1,2-Tetrachloroethane 0.611 11.64 0.12 Methyl-tert-butyl-Ether 2.49 9.33 0.122 m, p-xylene 1.995 5.6 0.171 Propionitrile 0.093 9.58 0.322 Isopropylbenzene 0.383 6.26 0.124 tert-butyl Alcohol 0.246 11.28 0.159 o-xylene 2.054 5.87 0.113 Ethyl-tert-butyl-Ether 0.256 6.45 0.136 Bromoform 0.206 14.72 0.164 Chloroprene 0.766 9.25 0.061 Styrene 1.49 6.83 0.075 1,1-Dichloroethane 1.253 5.55 0.143 Amyl acetate 1.65 13.27 0.075 Acrylonitrile 0.254 9.8 0.225 Bromobenzene 1.165 5.25 0.141 Vinyl acetate 3.249 10.23 0.288 n Propylbenzene 2.823 5.99 0.196 cis-1,2-dichloroethene 0.881 7.16 0.092 1,1,2,2-Tetracholorethane 0.576 9.66 0.13

2,2-Dichloropropane 0.959 7.51 0.1 4-Chlorotoluene 2.865 6.21 0.069 Bromochloromethane 1.483 8.33 0.118 1,2,3-Trichloropropane 0.323 8.87 0.151 Trichloroethene 0.075 8.2 0.152 1,2,3-Trimethylbenzene 3.434 6.16 0.138 Chloroform 1.132 6.58 0.095 trans-1,4-dichloro-2-butene 0.238 14.58 0.105 Carbon Tetrachloride 0.515 12.21 0.149 2-Chlorotoluene 1.777 6.18 0.161 1,1,1-Trichloroethane 0.85 9.68 0.148 Pentachloroethane 2.42 8.02 0.23 Ethyl Acetate 1.558 10.01 0.352 sec-butylbenzene 4.085 7.44 0.277 Tetrahydrofuran 0.575 11.85 0.176 1,2,4-Trimethylbenzene 0.124 8.96 0.438 Isobutanol 0.978 8.28 0.446 tert-butylbenzene 2.799 7.75 0.303 Isopropyl acetate 0.986 7.07 0.378 p-isopropyltoluene 2.835 7.69 0.293 1,1-Dichloropropene 0.962 6.11 0.122 1,3-Dichlorobenzene 1.879 8.18 0.209 2-Butanone 0.43 6.85 0.192 1,4-Dichlorobenzene 1.93 7.24 0.181 Benzene 2.137 4.08 0.143 n-butylbenzene 3.371 7.91 0.318 tert-amyl-methyl Ether 2.176 8.77 0.16 1,2-dichlorobenzene 1.844 7.77 0.129 1,2-Dichloroethane 0.508 5.37 0.206 1,2-Dibromo-3-chloropropane 0.158 11.57 0.212 Dibromomethane 0.277 6.15 0.139 Nitrobenzene 0.054 12.28 0.953 1,2-Dichloropropane 0.455 5.23 0.101 Hexachlorobutadiene 0.194 7.52 0.256 Brodichloromethane 0.42 8.59 0.098 1,2,4-Trichlorobenzene 1.138 9.8 0.48 Methyl Methacrylate 1.17 10.51 0.218 Naphthalene 2.478 8.85 0.14 Propyl acetate 1.256 8.27 0.134 1,2,3 Trichlorobenzene 1.09 9.98 0.399 Table 4: Calibration Data for EPA Method 8260C using Hydrogen Carrier Gas (*denotes linear regression) Figure 1: Total Ion Chromatogram (TIC of 10 ppb Calibration Standard.) 4-bromofluorobenzene (BFB) tuning requirements met all the ion abundance criteria for EPA Method 8260 1 (see Figure 2).

Figure 2: Passing BFB Tune using Hydrogen Carrier Gas. Conclusions The Atomx VOC Sample Prep System, used in conjunction with a Thermo Scientific TRACE 1310 / ISQ GCMS, was able to pass all performance requirements of US EPA method 8260C using hydrogen carrier gas and nitrogen purge gas. This is both a budget and environmentally friendlier way to analyze VOCs. As this study demonstrates, these readily available gases are viable alternatives for this analysis, although the method itself must be further evaluated and updated to allow for their use. Acknowledgement Teledyne Tekmar would like to thank Terry Jeffers of Thermo Scientific for his support with this application. References 1. USEPA Method 8260C, Volatile Organic Compounds by Gas Chromatography/Mass Spectrometry (GC/MS), Revision 3, August 2006.