Demonstration of the new InterMet radiosondes system installed at the Tanzania Meterological Agency, Dar-es-Salaam

Similar documents
Results from WMO High Quality Radiosonde Comparison, Mauritius As related to planning Reference Upper Air Observations for GCOS

Introduction to upper air measurements with radiosondes and other in situ observing systems. John Nash, C. Gaffard,R. Smout and M.

Overview of Met Office Intercomparison of Vaisala RS92 and RS41 Radiosondes

RA III Radiosonde Training Workshop, Buenos Aires, May 2006

Lessons from Mauritius WMO intercomparison and Results of temperature measurement of the aluminized boom sensor of Modem s radiosonde

Met Office Intercomparison of Vaisala RS92 and RS41 Radiosondes

WORLD METEOROLOGICAL ORGANIZATION. INSTRUMENTS AND OBSERVING METHODS REPORT No. 83

PRESENTATIONS ON RECENT NATIONAL TESTS/COMPARISONS. Recent Tests and Comparisons of Radiosonde Operated by Japan Meteorological Agency

W O R L D M E T E O R O L O G I C A L O R G A N I Z A T I O N COMMISSION FOR INSTRUMENTS AND METHODS OF OBSERVATION JOINT

WMO INTERNATIONAL RADIOSONDE COMPARISON

OPAG on Integrated Observing Systems. Workshop to Improve the Usefulness of Operational Radiosonde Data. (Submitted by the Secretariat)

Instruments and Methods of Observation Programme

RECENT WORLDWIDE GPS RADIOSONDE PERFORMANCE Incorporating the review of WMO GPS questionnaire, 2001

Comparison of Vaisala Radiosondes RS41 and RS92 WHITE PAPER

Recent Application of The Accurate Temperature Measuring (ATM) Radiosonde

* * * Table (1) Table (2)

Observation Development, Met Office, Exeter

VAISALA RADIOSONDE RS41 COVER IMPROVEMENT

China BeiDou radiosonde system

VAISALA RS92 RADIOSONDES OFFER A HIGH LEVEL OF GPS PERFORMANCE WITH A RELIABLE TELEMETRY LINK

Report of CoreTemp2017: Intercomparison of dual thermistor radiosonde (DTR) with RS41, RS92 and DFM09 radiosondes

SOME STEP OF QUALITY CONTROL OF UPPER-AIR NETWORK DATA IN CHINA. Zhiqiang Zhao

Specifications for a Reference Radiosonde for the GCOS Reference. Upper-Air Network (GRUAN)

The Vaisala AUTOSONDE AS41 OPERATIONAL EFFICIENCY AND RELIABILITY TO A TOTALLY NEW LEVEL.

Country Report for Japan (Submitted by Kenji Akaeda, Japan Meteorological Agency)

The Meisei sonde data product

Instruments and Methods of Observation Programme, the Report of the President of CIMO. Report to Cg-XV May 2007 Dr. J. Nash President of CIMO

The GRUAN Observing Station Payerne - Switzerland

Deriving Meteorological Data from free-to-air Mode-S broadcasts in an Australian Context.

of Russian demonstrated Fig 1. Russian allowed to the Vektor-M and more difficult. Performance in winter.

W 0 R L D M E T E 0 R 0 L 0 G I C A L 0 R G A N I Z A T I 0 N INSTRUMENTS AND OBSERVING METHODS REPORT NO. 40. WMO/TD-No

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

The GRUAN Observing Station Payerne - Switzerland

Status of GRUAN certification for French sites

High accuracy. Proven reliability. / SOUNDING SOLUTIONS

Field experiments at Sodankylä

P2.6 EVALUATION OF THE WVSS-II MOISTURE SENSOR USING CO-LOCATED IN-SITU AND REMOTELY SENSED OBSERVATIONS

DYNAMO/CINDY Sounding Network

Operational Applications of Awos Network in Turkey

NIWA s Lauder Site 45.0 S, E 370m a.s.l.

1 Introduction. Station Type No. Synoptic/GTS 17 Principal 172 Ordinary 546 Precipitation

5.1 Use of the Consensus Reference Concept for Testing Radiosondes. Joe Facundo and Jim Fitzgibbon, Office of Operational Systems,

MetConsole AWOS. (Automated Weather Observation System) Make the most of your energy SM

Exploitation of ground based GPS for Climate and Numerical Weather Prediction applications COST action 716

The current status, functions, challenges and needs of South Sudan Meteorological Department (SSMD)

GRUAN Data Flow and RS92 Data Product v2

EMADDC. towards operational collection of Mode-S EHS observations in Europe

W O R L D M E T E O R O L O G I C A L O R G A N I Z A T I O N COMMISSION FOR INSTRUMENTS AND METHODS OF OBSERVATION

Management in Surface, Climate and Upper-air Observations in RAII,Tokyo, Assistant Director Bangladesh Meteorological Department

GCOS Reference Upper- Air Network

*Ryan N. Brown and Daniel Brewer CyberData Technologies, Inc. Sterling, VA

Revitalization of the GCOS Surface and Upper Air Network Stations. Richard K. Thigpen GCOS Secretariat

The Vaisala Reference Radiosonde Program: First Results and Future Plans

Use of lightning data to improve observations for aeronautical activities

GRUAN Station Report for Ny-Ålesund

Meteorological Service

Country Report. India Meteorolog. gical ldepartment

The German Contribution to the WMO GAW Programme:

What Measures Can Be Taken To Improve The Understanding Of Observed Changes?

Performance of Radar Wind Profilers, Radiosondes, and Surface Flux Stations at the Southern Great Plains (SGP) Cloud and Radiation Testbed (CART) Site

Implementation Guidance of Aeronautical Meteorological Observer Competency Standards

Site Report: Lauder, New Zealand

MEASUREMENT OF UPPER-AIR PRESSURE, TEMPERATURE AND HUMIDITY

National Report on Weather Forecasting Service

A Rain Sensitive House Window Closes Automatically When Raining

New Peltier-based chilled-mirror hygrometer SKYDEW

Proper Data Management Responsibilities to Meet the Global Ocean Observing System (GOOS) Requirements

USE OF SATELLITE INFORMATION IN THE HUNGARIAN NOWCASTING SYSTEM

The Payerne Meteolabor Radiosonde

Ground-based temperature and humidity profiling using microwave radiometer retrievals at Sydney Airport.

NWS Sterling Field Support Center (SFSC)

QUALITY MANAGEMENT IN SURFACE, CLIMATE AND UPPER-AIR OBSERVATIONS IN CHINA

EAS 535 Laboratory Exercise Weather Station Setup and Verification

Vaisala AviMet Automated Weather Observing System

Radiosonde O-B statistics in the stratosphere

KING EDWARD POINT OBSERVATORY MAGNETIC DATA

DETERMINATION OF THE POWER LAW EXPONENT FOR SOUTHERN HIGHLANDS OF TANZANIA

2 Calculating Biases from Radiosonde Intercomparison Experiments

GRUAN Site Report for Ny-Ålesund

MEASUREMENT OF UPPER AIR PRESSURE, TEMPERATURE AND HUMIDITY

WMO INTERCOMPARISON OF GPS RADIOSONDES

CHAPTER 12. MEASUREMENT OF UPPER-AIR PRESSURE, TEMPERATURE AND HUMIDITY

A Case Study on Diurnal Boundary Layer Evolution

LUAMI CAMPAIGN FALL 2008 SUMMARY

Precipitation type detection Present Weather Sensor

Monthly Magnetic Bulletin

AWOS. Automated Weather Observing Systems COASTAL

Correction for Dry Bias in Vaisala Radiosonde RH Data

CLIMATE CHANGE ADAPTATION BY MEANS OF PUBLIC PRIVATE PARTNERSHIP TO ESTABLISH EARLY WARNING SYSTEM

Development of a UK National Water Vapour Processing System

Inter-comparison of Raingauges in a Sub-tropical Environment

Monthly Magnetic Bulletin

Change management for the RS92 phase out

KING EDWARD POINT OBSERVATORY MAGNETIC DATA

GRUAN Site Report for La Réunion

Barry M. Lesht. Environmental Research Division Argonne National Laboratory Argonne, Illinois 60439

Guidance on Aeronautical Meteorological Observer Competency Standards

Contribution of Norwegian partners (Aanderaa Data Instruments and NIVA) to Safeport project ( ). Final report

Atmospheric Sounding: Basics, Research and Development Radiosonde, Dropsonde and Driftsonde Junhong (June) Wang NCAR Earth Observing Laboratory

Government of Sultanate of Oman Public Authority of Civil Aviation Directorate General of Meteorology. National Report To

Transcription:

Demonstration of the new InterMet radiosondes system installed at the Tanzania Meterological Agency, Dar-es-Salaam J. Nash, R. Smout, M. Smees Met Office, Exeter,UK C. Bower NOAA-NWS, Silver Spring, Md, USA S. Mbuya, L. Mganga and S. Ndonde Tanzania Meterological Agency 1. Introduction An IMS 1600 upper air sounding system was installed by the manufacturer in Dar-es- Salaam, Tanzania in August 2004. The upper air station is located at Dar-es-Salaam airport. The IMS 1500 radiotheodolite, see Fig.2, was installed on the top of an observatory building, as advised by WMO representatives (see Fig.1), with the ground station computers located in the upper floor of the building. Fig.1 View of observing station at Dar-es-Salaam Airport from the balloon shed. A contract was placed with the UK Met Office to provide a demonstration test of this system, with reports to be generated for the US donors supporting the regeneration of radiosonde stations within the GCOS network. The test took place from 18 to 30 October 2004, with a lot of help from Carl Bower of NOAA and from a wide range of staff from the Tanzania Meteorological Agency.

2. Structure of demonstration test Plans for the test had to be revised daily, once it was realised that the Tanzanian staff needed more instruction in operational radiosonde procedures than had been provided by the Intermet engineers responsible for the installation. Thus, the competed demonstration test consisted of:- 14 individual Vaisala RS92 flights, used to train TMA staff in procedures. (These flights were processed by DigiCora III software using equipment brought to Tanzania by the UK team. Measurements were usually performed by TMA staff. The radiosonde were ground checked as recommended by the manufacturer, so the humidity sensors should have been regenerated from contamination before launch.) individual Sippican MKII training flights. nighttime simultaneous comparisons between Sippican and Vaisala, 5 daytime simultaneous comparisons between Sippican and Vaisala, nighttime simultaneous comparisons between MODEM and Vaisala 2 daytime simultaneous comparisons between MODEM and Vaisala. The limited number of simultaneous flights were sufficient to identify areas where more work was needed on the system. It was found that the IMS system was not really ready for a larger scale quantitative test. Fig.2 shows a Sippican radiosonde prepared for launch, with the chip thermistor deployed facing upwards and the wire support bent to hold the thermistor well above the top of the radiosonde. Earlier radiosonde testing in 2004 at Camborne, UK had shown that this mode of sensor deployment gave reproducible results with the Sippican MkII. Fig.2 Sippican MKII Radiosonde and Intermet IMS 1500 radiotheodolite.

The radiosondes in the twin flights were suspended under each end of a length of plastic water pipe. See Fig.3. Launching a twin flight is shown in Fig.4. The night time measurements were usually launched as the sun was setting, so the end of the flights was in the dark. Fig.3 Preparing to launch a Sippican / Vaisala RS92 twin flight, Dar-es-Salaam. The balloons used for the testing were a mixture of Pawan and Totex 350g balloons with burst heights typically between 19 and 23 km respectively. Six Totex 1200g balloons lifted radiosondes to heights above 30 km. The 1200g balloons were a little large for the balloon shed. However, the hydrogen generator was easily able to supply sufficient gas for this type of balloon. The large balloons performed best in the nighttime flights. Working arrangements for balloon filling were improved during the trial with the installation of an automatic balloon filler and a table on which balloons were prepared, see Fig.5. Results of the test The IMS 1600 system clearly needed more testing than had occurred before it was deployed in Tanzania. A lot of the system outputs were inconsistent within the various modules of the software. Temperature corrections applied to radiosonde measurements were not made with the correct local time. The origin of corrections applied was not well documented and there appeared to be a significant lack of cooperation between the radiosonde supplier and the IMS software engineers. Operational measurements should be made uncorrected until reliable correction schemes can be implemented.

Fig. 4 Launch of twin flight at Dar-es-Salaam Fig.5 Use of automatic balloon filler during the test.

A computational error for the pressure derived from geometric heights by IMS prevented operational use of the MODEM GPS radiosondes. e.g. the 100 hpa geopotential height was in error by more than 2 km. The tracking of the radiotheodolite appeared good and able to meet user requirements for wind in the tropics, but the algorithms used for generating winds were not computing values at the resolution needed [or indicated by the software settings], i.e. 1 minute near the ground and 1 to 4 minutes in the stratosphere, see the results from the twin flight comparison in Fig.6. Fig.6 Simultaneous measurements of wind from Vaisala (SGP) and IMS (MKII4) from flight 32 on 28.10.04 Mounting the radiotheodolite on the roof in Dar-es-Salaam hinders the automatic tracking of the radiosonde at launch. Launch routines would probably be easier if the radiotheodolite were mounted on the ground. Then, the person who launched the balloon could be responsible for checking that it was tracking on the main beam.

General conclusions 1. Further action is required to provide a more reliable method of communicating data from Dar-es-Salaam onto the GTS. Message generations needs to be made more reliable, see Fig.7 where a considerable manual effort was required to generate the TEMP message in a suitable format. 2. Re-establishing radiosonde measurements in a location such as Dar-es-Salaam where radiosonde ascents had not been made for more than 10 years requires a concerted training effort with staff present from both the manufacturer and from a National Meteorological Service with the necessary experience in radiosonde operations. 3. It would be beneficial if a suitable expert was sent to Tanzania to complete the training started during the demonstration test and to implement improved message communications. This should be considered once IMS are ready to upgrade the system software. 4. The hydrogen generator installation worked well and is clearly capable of supporting the use of 800 or 1000g balloons on a regular basis. It is recommended that suitable funding be provided for the use of the larger balloons at Dar-es-Salaam in the long term. 5. Met Office staff were grateful for the help and enthusiasm of the staff from the Tanzania Meteorological Agency, who made the test successful. Fig.7 Working together to generate a satisfactory TEMP message.