Calibration (The Good Curve) Greg Hudson EnviroCompliance Labs, Inc.

Similar documents
And how to do them. Denise L Seman City of Youngstown

Laboratory Techniques 100: Back To Basics. Carol Injasoulian Lab Manager City of Bay City April 29,2015

Measurement Uncertainty: A practical guide to understanding what your results really mean.

Bio 120 Lab 5: Quantitative Analysis

D E T E R M I N A T I O N O F K e q L A B

Linearization of Nonlinear Equations

DETERMINATION OF AN EQUILIBRIUM CONSTANT

9 Equilibrium. Aubrey High School PreAP -Chemistry. Name Period Date / /

A Fluorometric Analysis of Quinine in Tonic Water

Skill Building Activity 2 Determining the Concentration of a Species using a Vernier Spectrometer

regression analysis is a type of inferential statistics which tells us whether relationships between two or more variables exist

of the ferric thiocyanate. This was done by creating the solutions and putting them into a

Introduction to Spectroscopy: Analysis of Copper Ore

Experiment 11 Beer s Law

Experiment 11 Beer s Law

Spectrophotometric Determination of Ferrocyanide in Effluents

-However, this definition can be expanded to include: biology (biometrics), environmental science (environmetrics), economics (econometrics).

Chemical Kinetics: Integrated Rate Laws. ** updated Procedure for Spec 200 use **

Quality Assurance is what we do to get the right answer for our purpose FITNESS FOR PURPOSE

Developers Guide. Development of new instrument applications for D-dimer

Linear Regression. Simple linear regression model determines the relationship between one dependent variable (y) and one independent variable (x).

Experiment 1 (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with 1,10- Phenanthroline

Beers Law Instructor s Guide David T. Harvey

EPAs New MDL Procedure What it Means, Why it Works, and How to Comply

DETERMINATION OF K c FOR AN EQUILIBRIUM SYSTEM

ANALYSIS OF ZINC IN HAIR USING FLAME ATOMIC ABSORPTION SPECTROSCOPY

Water & Wastewater. Laboratory Math

Exp 03 - Reaction Rate

CORRELATION AND REGRESSION

Introduction to Spectroscopy: Analysis of Copper Ore

AP Chemistry Laboratory #15: Reaction Rate of Crystal Violet and Sodium Hydroxide. Lab days: Thursday and Friday, February 1-2, 2018

Proposed Procedures for Determining the Method Detection Limit and Minimum Level

Rate law Determination of the Crystal Violet Reaction Using the Isolation Method

Analytical Performance & Method. Validation

CHEMICAL KINETICS E + 2B 2C + D (1)

4/22/2010. Test 3 Review ANOVA

KINETICS OF THE PERMANGANATE- ISOPROPYL ALCOHOL REACTION

Experiment 8: DETERMINATION OF AN EQUILIBRIUM CONSTANT

Lab Investigation 4 - How Fast Does the Crystal Violet Decolorize?

Method Validation. Role of Validation. Two levels. Flow of method validation. Method selection

Determination of Orthophosphate Ion

Concentrations that absorb. Measuring percentage transmittance of solutions at different concentrations

Determination of Orthophosphate Ion

Statistics: Error (Chpt. 5)

Determining the Concentration of a Solution: Beer s Law

CHEM 3420 /7420G Instrumental Analysis

Chemistry 213. A KINETIC STUDY: REACTION OF CRYSTAL VIOLET WITH NaOH LEARNING OBJECTIVES

CHEM Lab 7: Determination of an Equilibrium Constant using Spectroscopy

Experimental Procedure Overview

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

[C] [D] K eq = [A] [B]

MassLynx 4.1 Peak Integration and Quantitation Algorithm Guide

Basic Statistics. 1. Gross error analyst makes a gross mistake (misread balance or entered wrong value into calculation).

Spectrophotometric Determination of pka of Phenol Red

Chemical Equilibrium: Finding a Constant, Kc

Spectrometric Determination of the Acid Dissociation Constant of an Acid-base Indicator

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 03: EXPERIMENTAL ERROR

Kinetics of Crystal Violet Bleaching

Experimental design. Matti Hotokka Department of Physical Chemistry Åbo Akademi University

MORE LIGHTS, COLOR, ABSORPTION!

Chapter 12 - Part I: Correlation Analysis

Intro to Linear Regression

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

ELISA QUALITY ASSURANCE Analytical Phase

Experiment 13I THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Chapter 5. Errors in Chemical Analysis 熊同銘.

Intro to Linear Regression

Chemical Kinetics. The dependence of reaction rate on concentration is given by the rate law: rate = k[a] x [B] y [C] z (1)

EXPERIMENT 30A1: MEASUREMENTS. Learning Outcomes. Introduction. Experimental Value - True Value. 100 True Value

Chemical Kinetics Prelab. 4. Why do the solutions have to be mixed quickly before measuring the absorbance data?

Lab 1 Uniform Motion - Graphing and Analyzing Motion

Results and Analysis 10/4/2012. EE145L Lab 1, Linear Regression

KINETICS: INITIAL RATES

EXPERIMENT #3 A Beer's Law Study

PREPARATION FOR CHEMISTRY LAB: FLUORIDE IN WATER

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION

STAT 350 Final (new Material) Review Problems Key Spring 2016

Introduction to Spectroscopy: Analysis of Copper Ore

Chapter 4: Verification of compendial methods

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

EXPERIMENT 14. ACID DISSOCIATION CONSTANT OF METHYL RED 1

Stat 101 L: Laboratory 5

Spectrophotometric Determination of an Equilibrium Constant

Business Statistics. Lecture 9: Simple Regression

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

REVIEW OF LAB TECHNIQUES

The New MDL Procedure How To s. Presented by: Marcy Bolek - Alloway

Concepts, Techniques. Concepts, Techniques 9/11/2012. & Beer s Law. For a simple, transparent, COLORED material, e.g. ROYGBV

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry

D. Experiment Summary

THE TEMPERATURE DEPENDENCE OF THE EQUILIBRIUM CONSTANT

EXPERIMENT 23. Determination of the Formula of a Complex Ion INTRODUCTION

Predicted Y Scores. The symbol stands for a predicted Y score

ADVANCED ANALYTICAL LAB TECH (Lecture) CHM

Variance. Standard deviation VAR = = value. Unbiased SD = SD = 10/23/2011. Functional Connectivity Correlation and Regression.

The Determination of an Equilibrium Constant

Quality Control Procedures for Graphite Furnace AAS using Avanta Software

Experiment 6: Determination of the Equilibrium Constant for Iron Thiocyanate Complex

Basics of UV-Visible Spectroscopy *

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 03: EXPERIMENTAL ERROR

Transcription:

Calibration (The Good Curve) Greg Hudson EnviroCompliance Labs, Inc. greghudson@envirocompliance.com www.envirocompliance.com

Abstract It should come as no surprise that the correlation coefficient is only one way to judge the quality of a calibration curve. The fact is that there are numerous tools available to evaluate how good a calibration curve represents a given data set. With a basic understanding of how to use regression analysis, some simple methods can be applied to determine both the quality of necessary to improve calibration technique.

Overview The black box versus thinking outside the box. The black box is the regression algorithm. Thinking outside the box is evaluating analytical processes or biases which may effect the quality of the curve.

What is a Curve? STANDARD CURVE: A plot of concentrations of known analyte standards versus the instrument response to the analyte. Calibration standards are prepared by successively diluting a standard solution to produce working standards which cover the working range of the instrument. Standards should be prepared at the frequency specified in the appropriate section... SW-846

OK, what do we really mean by curve? The ability to predict behavior by establishing known outcome. Simply put, getting a result with reasonable certainty.

What so important about a good Curve? Easier to correct problems if the calibration is bad Higher confidence in analytical results

A Peek inside the Black Box Basic linear regression by least squares LINEAR Eq.: x=(y-b)/m or y=mx+b where m = slope or rise over run b = intercept (curve crosses the y axis) (Σxy - (Σx * Σy y /n)) Slope = --------------------------- (Σx^2 - ((Σ x)^2 / n)) intercept = (Σy/n) - (m *Σx / n)

The Correlation Coefficient An estimate of the remainders derived from the regression. How do we calculate it? (Σxy - (Σx * Σy y / n))^2 r = ---------------------------------------------- ((Σx^2 - ((Σx)^2 / n)))*(σy^2 - ((Σy)^2 / n)) What does it mean? For our purpose it must be >0.995 The correlation coefficient only indicates variance from the averages; therefore it is not the best indicator of curve quality.

r = 0.997 45 40 35 30 25 20 15 10 5 0 0 10 20 30 40 50 Std 1 3 1 Std 2 2 2 Std 3 1 3 Std 4 10 10 Std 5 20 20 Std 6 40 40

r = 0.9999 0.1 0.09 0.08 0.07 0.06 0.05 0.04 0.03 0.02 0.01 0 0 5 10 15 20 25 Std 1 1 0.011 90.90909 Std 2 2 0.015 133.3333 Std 3 5 0.028 178.5714 Std 4 10 0.048 208.3333 Std 5 20 0.091 219.7802

Outside the Box Beer's Law states that the absorbance of a solution is directly proportional to the concentration. (Ideally linear) What effects Beer? Noise is most often the function of the detector Quenching - At high concentration, Beer's law will not continue to behave Limiting factors of method or reagents. Most methods are intended to operate at specific trace ranges. Dilution after reaction or extraction beyond trace levels may bias final results. Capture

Evaluation of Curve Quality Slope vs. RF Response Factor is the ratio of Concentration/Response and represents the instantaneous slope for a given point. Variance of RF The average of the RFs should be close to the slope. The variance of the RFs is an excellent indication of curve quality. RSD <10% Linear Calibration Range Study - LCR LCR is simply plugging the response of each standard into the line function and determining the variance from predicted value (conc.). These should be <20% and ideally <10% Intercept - Should not be greater than the lowest standard. Visual Date should appear to be linear and fall along the curve. Equal Data points on either side Generally, data points should be evenly distributed around the curve. If most of the points fall to one side of the line or are not evenly divided by the line, there may be a problem with the curve.

Problems Effecting Curve Quality Bias Bias is any part of the analytical process which may cause unequal treatment from one part of the procedure to another- such as using that "special" beaker for blanks. Intercept High intercepts imply bias for instrument background or contamination. Tilt Tilt or skewed sloped is often the result of poor instrument performance. Higher order curves Calibration curves which are not linear may imply problems in the analytical process such as saturation or degradation. Bimodal data Bimodal data may appear to have two distinct 'lines' and is most often related to dilution technique like using a eppendorf for the lower part of a curve and a pipet for the upper part of the curve. Scatter Random divergence in a curve can result from numerous factors including extraction/digestion, poor technique, or dirty or malfunctioning instrumentation among others.

Problems Effecting Curve Quality Forcing through zero Linear regression should not be forced through zero for two reasons: 1) Good information can be gained from the natural intercept. 2) The regression formula is skewed by passing through zero. Span The entire range of the analytical curve can be problematic if it goes beyond the linear range of the instrument or system. Spacing of stds Standards should represent the analytical range evenly. Keep in mind that the Least Squares algorithm can be biases by using higher concentrations. Standards should not exceed five (5) fold changes in concentration at most. Dilution and general pipetting technique It is critical to use consistent and precise dilution techniques for standards and all spiking or QC functions. Use of varying means for dilutions can add significant variance for each standard.

Corrective Actions Perform Maintenance Clean glassware Check dilutions and be consistent Remake standard and reagents (INCLUDING STOCKS) Re-run standards Re-zero Avoid: Deleting standards Running extra standards Excessive re-zeroing "Special" Behavior (i.e. using one cuvette for standards and another for samples)

Calibration (The Good Curve) Discussion/Questions Greg Hudson EnviroCompliance Labs, Inc. greghudson@envirocompliance.com www.envirocompliance.com