Thermocouples calibration and analysis of the influence of the length of the sensor coating

Similar documents
Base metal thermocouples drift rate dependence from thermoelement diameter

Veerapong Kanchanawongkul*

EXPERIMENT VARIATION OF THERMO-EMF WITH TEMPERATURE. Structure. 7.1 Introduction Objectives

Thermal analysis of superconducting undulator cryomodules

Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE

Linear Control Design for a Plastic Extruder

Using a Mercury itc with thermocouples

Temperature Scales. Temperature, and Temperature Dependent on Physical Properties. Temperature. Temperature Scale

ASEN 2002 Experimental Laboratory 1: Temperature Measurement and an Blow Dryer Test

AC-DC Transfer Difference Measurements using a Frequency Output Thermal Converter

SEN TRONIC AG 1 A 6 6 / "

Sensing, Computing, Actuating

ME 365 EXPERIMENT 5 FIRST ORDER SYSTEM IDENTIFICATION APPLIED TO TEMPERATURE MEASUREMENT SYSTEMS

EM375 MECHANICAL ENGINEERING EXPERIMENTATION THERMOCOUPLE LABORATORY

Experiment 3. Electrical Energy. Calculate the electrical power dissipated in a resistor.

Heating surface thermal stabilization for pool boiling on porous coverings

Study on the natural air cooling design of electronic equipment casings: Effects of the height and size of outlet vent on the flow resistances

Laboratory 12: Three Thermodynamics Experiments

In-process Measuring of Capacitance Per Unit Length for Single-core Electric Wires

C as an Aid in the Teaching (Learning) of Heat Transfer

Never switch on the equipment without the assistants explicit authorization!

Design and Development of Impact Load Sensor for Dynamic Testing Purposes

Critical Current Properties of HTS Twisted Stacked-Tape Cable in Subcooled- and Pressurized-Liquid Nitrogen

Measurement of Temperature in the Plastics Industry

PHYS320 ilab (O) Experiment 2 Instructions Conservation of Energy: The Electrical Equivalent of Heat

Heat transfer coefficient distribution over the inconel plate cooled from high temperature by the array of water jets

15 - THERMAL AND CHEMICAL EFFECTS OF CURRENTS Page 1 ( Answers at the end of all questions )

TEMPERATURE MEASUREMENT. GROUP 3 Noah Beauchamp Kyle Camacho Jack Connolly Curtis Yu

ME 105 Mechanical Engineering Laboratory Spring Quarter Experiment #2: Temperature Measurements and Transient Conduction and Convection

Lab 1f Boiling Heat Transfer Paradox

Stirling engine with Cobra3

Experimental and numerical investigations of heat transfer and thermal efficiency of an infrared gas stove

Ideal Gas Law and Absolute Zero

Experiment 5: Thermocouples (tbc 1/14/2007, revised 3/16/2007, 3/22,2007, 2/23/2009, 3/13/2011)

Experimental and numerical investigation on particle-induced liquid metal flow using Lorentz force velocimetry

Lecture 36: Temperatue Measurements

An industrial radiography exposure device based on measurement of transmitted gamma-ray intensity

Department of Mechanical Engineering ME 96. Free and Forced Convection Experiment. Revised: 25 April Introduction

Inductance and Current Distribution Analysis of a Prototype HTS Cable

Fundamentals of Circuits I: Current Models, Batteries & Bulbs

Designing a Thermostat Worksheet

1. Mark the correct statement(s)

OPPA European Social Fund Prague & EU: We invest in your future.

Specific Heat. Power Supply Beaker Beam Balance Conecting wires ice. Assembly

Harnessing the Power of Arduino for the Advanced Lab

Sensors and Actuators Sensors Physics

Specific Heat. Power Supply Beaker Beam Balance Conecting wires ice. Assembly

Student Experiments HEAT 2. Manual P9160-5C.

Temperature Measurements Using Type K Thermocouples and the Fluke Helios Plus 2287A Datalogger Artmann, Nikolai; Vonbank, R.; Jensen, Rasmus Lund

An autonomous underwater telescope for measuring the scattering of light in the deep sea

Lab E3: The Wheatstone Bridge

Circular Motion and Centripetal Force

Development of high-efficiency Stirling cryocoolers for high temperature superconducting motors

Interferometric determination of thermal expansion coefficient of piston/cylinder unit preliminary investigation

Thermochemistry/Calorimetry. Determination of the enthalpy of vaporization of liquids LEC 02. What you need: What you can learn about

An Introduction to Insulation Resistance Testing

Metrological Characterization of Hardness Indenter Calibration System

Effect of Discontinuities and Penetrations on the Shielding Efficacy of High Temperature Superconducting Magnetic Shields

Capacitor investigations

15. Compare the result with the value you have taken above Compare the calculated pressure value with the actual pressure value that you have

Measurements & Instrumentation

Our Drop Counter sensor now features housing for two electrode sensors, an anti-twist mechanism, an indicator LED and two cable guides.

Home-built Apparatus for Measuring Thermal Conductivity of Glass and Polymer Materials

Earlier Lecture. In the earlier lecture, we have seen non metallic sensors like Silicon diode, Cernox and Ruthenium Oxide.

Heat transfer coefficient of near boiling single phase flow with propane in horizontal circular micro channel

Engineering of centrifugal dust-collectors based on parallel comparing tests applying computer modelling

12. Heat of melting and evaporation of water

1. How much heat was needed to raise the bullet to its final temperature?

LMM-H04 Mass Air Flow Sensor

The Underground Experimental Investigation of Thermocouples

Thermal and electrical conductivity of metals

Correction for PMT temperature dependence of the LHCf calorimeters

PAPER 2 THEORY QUESTIONS

Contribution to the study of neon-nitrogen mixtures at low temperatures

A method for overcoming thermo EMF of shunt connected meters for measuring electric energy

High strain rate fracture behaviour of fused silica

Name Partner. Thermal Physics. Part I: Heat of Vaporization of Nitrogen. Introduction:

About the different types of variables, How to identify them when doing your practical work.

Experiment 14 It s Snow Big Deal

Conductometric Titration & Gravimetric Determination of a Precipitate

Electricity. Measuring the force on current-carrying conductors in a homogeneous magnetic field. LEYBOLD Physics Leaflets P

Research on State-of-Charge (SOC) estimation using current integration based on temperature compensation

Fundamentals of Heat Transfer (Basic Concepts)

Saltwater Conductivity: Trends of the Periodic Table

CHARGED PARTICLES IN FIELDS

c. VH: Heating voltage between the collector and emitter.

1. This question is about modelling the thermal processes involved when a person is running.

Exploring Physics and Math with the CBL System

Motion Control of the Pressing Rollers in the Paper Tape Crimping Machine Based on Servo Control

Pin Fin Lab Report Example. Names. ME331 Lab

T h e rm i s t o r s

Handout 10: Heat and heat transfer. Heat capacity

4.1. Physics Module Form 4 Chapter 4 - Heat GCKL UNDERSTANDING THERMAL EQUILIBRIUM. What is thermal equilibrium?

MECHANICAL ENGINEERING TECHNOLOGY ESSENTIALS FOR LABORATORY REPORTS

Coulomb s Law PHYS 296

Technical Notes. Introduction. PCB (printed circuit board) Design. Issue 1 January 2010

I. Introduction and Objectives

Finite element analysis of combined magnetoelectric- electrodynamic vibration energy converter

Study of the influence of heat transfer of a CLT beam through FEM

PHYSICS PAPER QUESTIONS

Transcription:

Journal of Physics: Conference Series OPEN ACCESS Thermocouples calibration and analysis of the influence of the length of the sensor coating To cite this article: M Noriega et al 2015 J. Phys.: Conf. Ser. 582 012029 View the article online for updates and enhancements. Related content - THE DIMENSIONS OF THERMOCOUPLE RECEIVERS Edison Pettit - Temperature determination of the Si SiC eutectic fixed point using thermocouples Suherlan, Yong-Gyoo Kim, Wukchul Joung et al. - Fabrication of a nickel freezing-point cell and preliminary study of its behaviour Y-G Kim, K S Gam and K H Kang This content was downloaded from IP address 37.44.195.42 on 01/01/2018 at 09:54

Thermocouples calibration and analysis of the influence of the length of the sensor coating M Noriega, R Ramírez, R López, M Vaca, J Morales, H Terres, A Lizardi, S Chávez Universidad Autónoma Metropolitana-Azcapotzalco. Distrito Federal, México. E-mail: rlc@correo.azc.uam.mx Abstract. This paper presents the design and construction of a lab prototype, with a much lower cost compared to the ones commercially sold, enabling the manufacture of thermocouples which are then calibrated to verify their functionality and acceptance. We also analyze the influence of the external insulation over the wires, to determine whether it influences temperature measurement. The tested lengths ranged from 0.00 m up to 0.030 m. The thermocouple was compared against the behavior of a thermocouple of the same type that was purchased with a commercial supplier. The obtained measurement showed less than 1 C difference in some points. This makes the built thermocouple reliable, since the standard allows a difference of up to 2.2 C. 1. Introduction Temperature is an important factor that should be taken into account in the majority of energy exchange processes, thermometers are instruments specifically designed to measure temperature, there is a wide variety of thermometers today and their operating principles are equally extensive. The thermometers are the instruments that are used for temperature measurement, a type of them, contact thermometers, should be placed in physical contact in the middle of the medium from which you want to know the temperature, for example: in production processes or exchange of energy, the medium can consist of an electric oven, in exhaust gases of a combustion engine, water at the entrance of a cooling tower, bioreactors, etc. These contact thermometers base its operation in thermal equilibrium that is established by direct contact of sensor and the medium of interest. Thermocouples are a type of contact temperature gauges; they are widely used in the industrial branch due to their capacity of measuring a wide range of temperatures, their reliability, easy operation and the possibility of registration of the measurement automatically. A thermocouple is composed of two conducting wires made of different material; these cables have a point in common. This union works as measuring point (sensing joint), which comes in contact with the medium in which you want to determine the temperature, at the other end, wires are connected to an electronic circuit that delivers a reference voltage (the voltage is usually associated with a temperature of 0 C). When thermocouple wires are subjected to a difference in temperature between the measuring joint and the end connected to the reference circuit, there is a potential difference based on the change of temperatures (Seebeck effect). One of the most widely used thermocouples is the K-type, because of its low cost compared to other types of thermocouples. The commercial cost of these devices is high, although the material to build them is usually inexpensive; the cost of the commercial product rises considerably, up to eight times. The punctual Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. Published under licence by Ltd 1

measurement of temperature has different key point applications in laboratory work, especially when it comes to evaluating the thermodynamic behavior of some systems as for example in solar energy, heat transfer, thermal machines, etc. There are some prototypes which require up to 50 measurements in different sites at the same time, which makes the cost of this kind of experience rather high, considering the use of one or more thermocouples for each point. In order to reduce costs, to the extent possible, in the area of Thermo-fluids, at UAM-A, it was decided to design and build an apparatus that would manufacture the thermocouples used in the instrumentation of the prototypes. We subsequently performed the calibration of the produced thermocouples. A question that arises when these instruments are evaluated is if the distance of coating that is removed to the thermocouple measuring wires influence on the precision of the measurement or not. The objectives of this work were to present the design and construction of the prototype for the thermocouples manufacturing and to investigate whether the distance that is removed from the wire coating had a significant impact on the temperature measurements. 2. Description of the manufacturing prototype and calibration of thermocouples For the manufacture of thermocouples, type K (Chromel-Alumel) cable for thermocouple was used, the joint between the two metals that is required for the measurement of temperature was made producing a short circuit; this merged both metals forming an amalgam, and therefore the principle of operation of the prototype is to produce this phenomenon. The components of the device are: (A) a mechanical arm, (H) a ceramic rod, (D) a tip to join the thermocouple, (B, C) thermocouple attachment lugs, and (E, F) the power switch box. One end of the lever arm is firmly fixed to a base (D) at the other end is placed a piece of charcoal which helps to close the electrical circuit with the tip of the thermocouple (C). This material was selected because it was the only one that would not leave residues and also it would not stick to the fused union of the temperature sensor; the ceramic bar (H) is fixed and its function is to serve as support and insulation when the short circuit between the carbon and the tip of the thermocouple occurs; the tip of the thermocouple to be joined is the type K-thermocouple cable. The power switch (F) is located in the switch box; it is a 10 A thermomagnetic switch, which must be activated before making any temperature measurement (F). Finally, there is a plug for connection to the Electric power supply. The diagram of the device is presented in Figure 1 and the built prototype, in Figure 2. Fig.1. Prototype diagram. Fig. 2. Image of the finished prototype. LabView software in which the instrumentation (ADAM modules) is integrated, was used in the registration of measurements. The proper programming for reading in real time and data storage for their subsequent management was performed. 2

A virtual instrument (program) which allows the measurement, registration and presenting the graphs for four temperatures on screen at the desired time, was created. The main window of this LabView program is shown in figure 3. Fig. 3. Main window of the program. 3. Manufacture of thermocouple The thermocouple manufacturing was carried out with a section of 0.40 m type K-cable, a length of 0.02 m of the insulating material at one end was withdrawn, this side should be placed in the holddown shoe of the prototype for the manufacture of thermocouples; a length of 0.005 m of the insulation material is removed from the end of wires, then both wires are united with a rat-tail type lashing; this would be the temperature sensor. This end was placed in the ceramics holding of the positive pole of the apparatus and then the soldering thermocouple wires by the action of shortcircuit proceeded. The thermocouple was withdrawn from the device making sure that the molten tip was perfect, i.e. that a small ball with both wires, free of any kind of fracture, had been formed. For details on the design and use of the available equipment, [2] can be consulted. For the first test the length of the insulation removed at the thermocouple was considered null, i.e., the thermocouple had two wires covered completely, then lengths of 0.003 m, 0.005 m, 0. 010 m, 0.015 m, 0.020 m 0.025 m and 0.030 m were withdrawn, and each test was repeated four times to verify that there was no significant variation in the measurements, or in its case, to obtain an average of them. A commercial thermocouple of the same type was employed as reference, and served as a pattern for measurements. 4. Experimental procedure. The experiment was conducted by placing 750 ml of water at 0 C, in a beaker of 1 liter of volume, where two thermocouples were inserted into the center of the bowl and exactly at the same depth, to ensure that both were in the same isothermal region and to avoid, to the extent possible, variations in temperature due to the flow currents. The vessel was warmed on an electrical thermal plaque until the water reached boiling point. At the same time both ends of thermocouples were connected to the ADAM module and the LabView program was started on the computer and the registration of measurements, performed every second, begins. The employed array is shown in Figure 3. 3

Fig. 3. Measurement arrangement: a) pattern thermocouple, b) thermocouple, c) electric heater, d) ADAM unit, e) computer, and f) water. 5. Results All the tests that were conducted began at 0 C and ended at approximately at 92 C, in each experiment the average of the four tests was obtained, most of the results were almost equal, and the duration of each time was approximately 26 minutes. The graph of temperature against time obtained when the two wires of the thermocouple were completely covered with plastic is presented in Figure 4. There are some differences of maximum 1 C in some spots. In all tests no greater value difference was observed. Fig. 4. Graph of temperature against time, with both wires of the thermocouple fully covered with insulating material. In the case of the thermocouple for which 0.030 m of the insulating cover was withdrawn, the variation of temperature for some points did not exceed 1 C, as shown in the graph of Figure 5. According to the ASTM E230/E230M-12 standard [1], in the temperature range of 0 C to 760 C a difference of up to 2.2 C is allowed, therefore the built thermocouple was considered acceptable for measurements in the specified temperature range. Subsequently, the described procedure was repeated several times to this same thermocouple, i.e., a length of 0.003 m of plastic was removed and four tests were performed, until it a final length of 0.030 m of cover was withdrawn. 4

Fig. 5. Graph of temperature against time, for the thermocouple without 0.030 m insulating material. Figure 6, shows the graph of the temperature of the thermocouple to be calibrated against the temperature of the reference thermocouple, in order to find the percentage of error between two measurements and the value of of each of the performed experiences. In the case of both wires completely covered the = 0.9996, the percentage of average error found was 1.2%, which makes it acceptable. Fig 6. Fit graph for the insulated thermocouple, the value of is very close to the unit. The values of the percentage of error and the value of for every experienced thermocouple are presented in Table 1. The maximum value of the percentage of retrieved error was observed when we withdrew a length of 0.030 m of the insulating material and the value of in all cases was very close to the unit. The conclusion from these results is that the insulating material of a thermocouple can be removed up to 0.025 m length and readings obtained will be within the recommendation mentioned in the corresponding ASTM standard. 5

6. Conclusions. An economical lab prototype that enables the production of thermocouples which are subsequently calibrated for its functionality and acceptance has been built. The influence of the insulation covering of the wires in temperature measurement was analyzed. The experienced lengths of the uncovered portion of the wire ranged from 0.00 m up to 0.030 m, and were compared against the behavior of a thermocouple of the same type that was purchased with a commercial distributor. The obtained measurements presented a variation on some points of less than 1 C, which confirms the built thermocouple as reliable, since the standard allows a difference of up to 2.2 C. Table 1. Error percentage and value of p for each evaluated condition of the thermocouple. Non-insulated length [m] % Error 0.000 1.10 0.99967 0.003 0.71 0.99972 0.005 1.27 0.99967 0.010 0.43 0.99969 0.015 0.65 0.99977 0.020 1.90 0.99976 0.025 0.60 0.9997 0.030 2.90 0.99944 References [1] ASTM E230/E230M-12 2012 [2] López C.R., Morales G. J., Vaca M. M., Lizardi R. A., Terres P. H, Lara V. A. 2013. Equipo para Laboratorio: Fabricación y Calibración de Termopares UAM-Azcapotzalco 6