Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

Similar documents
TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

Bureau International des Poids et Mesures

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays

M [scale units/s] of the system

Bureau International des Poids et Mesures

Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation

Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays

Abstract An indirect comparison of the standards for reference air kerma rate for

Bureau International des Poids et Mesures

4.1b - Cavity Theory Lecture 2 Peter R Al mond 2011 Overview of Lecture Exposure (W/e)air Exposure Exposure and and and Air Air Kerma

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the NIM, China and the BIPM in 60 Co gamma radiation

Comparison of the standards of air kerma of the BNM-LPRI and the BIPM for 137CS y rays

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck

Comparison of the standards of air kerma of the GUM and the BIPM for 60Co 'Y rays

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the LNE-LNHB, France and the BIPM in 60 Co gamma radiation

Air kerma rate measurements from a miniature x-ray source using free-air ionization chambers

Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays

Comparison of the standards for absorbed dose to water of the ENEA-INMRI (Italy) and the BIPM for 60 Co γ rays

Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network /2017

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden

Comparison of the standards for absorbed dose to water of the VSL and the BIPM for 60 Co gamma rays

Ionizing Radiation Dosimetry and Medical Physics

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Reference Dosimetry for Megavoltage Therapy Beams: Electrons

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS

Determination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1)

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS

Chapter 9: Calibration of Photon and Electron Beams

RECOMBINATION PARAMETERS OF SOME FABRICATED IONIZATION CHAMBERS

Dosimetry: Electron Beams

Chapiter VII: Ionization chamber

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

Efficiencies of Some Spherical Ion Chambers in Continuous and Pulsed Radiation: A Numerical Evaluation

TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers

Comparisons of the standards for air kerma of the GUM and the BIPM for 60 Co and 137 Cs gamma radiation

8/2/2012. Larry DeWerd has a partial interest in Standard Imaging

Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kv to 300 kv

PRELIMINARY CHARACTERIZATION OF THE NIS FREE-AIR CHAMBER STANDARD AT THE BIPM

Optimization of microionization chambers for small-field reference dosimetry

Comparison of air kerma measurements for tungsten anode based mammography x-ray beam qualities (EURAMET.RI(I)-S4.1)

H Bjerke 1, J Plagnard 2, J-M Bordy 2, A Kosunen 3, J Huikari 3, L Persson 4, and P O Hetland 1

Implementation of the IAEA-AAPM Code of Practice for the dosimetry of small static fields used in external beam radiotherapy

Principles of applied dosimetry - illustrated by ionometry. Lesson FYSKJM4710 Eirik Malinen

Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015

Comparisons between IRD and BIPM exposure and air-kerma standards. for cobalt-60 gamma rays. by C.E. de Almeida* and M.-T. Niatel

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields

BIPM Key Comparison BIPM.RI(I)-K6 PROTOCOL 3.0 CCRI(I) 6 November 2017

Calibration of a pencil ionization chamber with and without preamplifier

Metrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT)

Comparison of V and 10 V DC Voltage References

Chapter V: Cavity theories

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe

Determination of contributions of scatter and distance error to the source strength of 192 Ir HDR brachytherapy source

Essentials of expressing measurement uncertainty

N. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required

Small Field Dosimetric Measurements with TLD-100, Alanine, and Ionization Chambers

FINAL REPORT. European co-operation for Accreditation (EA) Interlaboratory Comparison IR3: Calibration of a Radiation Protection Dosimeter

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

BUREAU INTERNATIONAL DES POIDS ET MESURES

Instrumentation for Verification of Dose

Measurement of Tritium in Helium

Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams

Radiation Quantities and Units

Investigation of the standard temperature- pressure correction factor at low x-ray energies

General Overview of Gas Filled Detectors

Work Programme. of the International Bureau of Weights and Measures. for the four years Comité international des poids et mesures

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

Updating reference dosimetry a decade after TG-51

Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration

MONTE CARLO SIMULATIONS USED TO ESTABLISH A FRENCH NATIONAL HDR BRACHYTERAPY REFERENCE

PRIMARY STANDARDS of AIR KERMA for 60 CO and X-RAYS & ABSORBED DOSE in PHOTON and ELECTRON BEAMS. Malcolm McEwen

Instrumentation for Verification of Dose

Shielding of Ionising Radiation with the Dosimetry & Shielding Module

) for Varian TrueBeam high-dose-rate therapy beams

IAEA TRS May 2001 (V.10A) PUBLISHED BY THE IAEA ON BEHALF OF IAEA, WHO, PAHO, AND ESTRO INTERNATIONAL ATOMIC ENERGY AGENCY IAEA

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification.

APMP key comparison for the measurement of air kerma for 60 Co (APMP.RI(I)-K1.1)

Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

First international comparison of primary absorbed dose to water standards in the medium-energy X-ray range

Monte Carlo Simulation concerning Particle Therapy

ERRATA LIST AND UPDATES TO IAEA TRS-398 (2000) 1

BUREAU INTERNATIONAL DES POIDS ET MESURES

Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI

Final Report August 2010

Simulation Modeling in Dosimetry

Bureau International des Poids et Mesures, Sèvres, France. *KEBS, Kenya Bureau of Standards P.O. Box , Nairobi, Kenya

Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin

Jeancarlo Torres, CHP. Naval Surface Warfare Center Carderock Division

Symposium on NIST Centennial July 26, 2001 AAPM Salt Lake City

Transcription:

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti 1 and Ladislav Czap 2 1 National Institute of Standards and Technology (NIST), Gaithersburg, MD USA 2 International Atomic Energy Agency (IAEA), Vienna, Austria. Abstract A comparison of the air kerma standards for 137 Cs and 60 Co gamma ray beams was performed between the NIST and the IAEA. Two reference class ionization chambers owned by the IAEA were used as part of this comparison and were calibrated at each facility. The calibration coefficients, N K, were determined for both chambers and in both gamma-ray beams. The measurements were performed at the IAEA and NIST facilities starting in the fall of 2009 and were completed in 2010. The comparison ratio of the calibration coefficients for each chamber, N K,IAEA / N K,NIST, between the IAEA and NIST were 0.999 and 0.997 for the 137 Cs and 60 Co gamma ray beams respectively. The relative standard uncertainty for each of these ratios is 0.5 %. 1. Introduction The air kerma standards for 137 Cs and 60 Co gamma ray beams of the National Institute of Standards and Technology (NIST) and the International Atomic Energy Agency (IAEA) were compared for the first time under the auspices of the regional metrology organization known as the Inter-America Metrology System (SIM). This supplementary comparison is identified as SIM.RI(I)-S1. For this comparison, two reference class chambers owned by the IAEA were used. The calibration coefficient, N K, was obtained for each chamber in each radiation protection level beam at both facilities. The comparison results are expressed as the ratio of the calibration coefficients obtained at each laboratory. 2. Transfer chambers Two reference class ionization chambers owned by the IAEA were used during this comparison. The physical properties of these two chambers are listed in Table 1. Both chambers are spherical and made of C552 air equivalent plastic. The nominal volumes of the chambers were 100 cm 3 and 1000 cm 3 for the models A5 and PTW TN32002, respectively. They were aligned with the stems perpendicular to the gamma-ray beam and with the white mark on the chamber facing the source of radiation. A negative voltage was applied to the outer wall while the center electrode was kept at ground potential. In this way negative charge was collected. 1/8

Table 1. Description of the chambers. Chamber Name and Model a Serial Number Sensitive volume (nominal) Outside Diameter (nominal) Wall Thickness (nominal) Applied Potential /cm 3 / cm / mm / V Exradin A5 PTW TN32002 XY032721 100 6 3 300 277 1000 14 3 400 3. Reference beams at the NIST and at the IAEA. 3.1. Reference rates at the NIST The gamma-ray measurements were made in the NIST 137 Cs and 60 Co gamma ray beam calibration facilities. The air kerma rates in these facilities were obtained by applying a decay correction to the reference air kerma rate. The reference air kerma rates were determined using a suite of six graphite-wall cavity ionization chambers which constitute the NIST primary standard instruments. Details on the determination of the construction and evaluation of the reference air kerma rates are in [1]. In brief, the reference air kerma rates, K &, at the NIST were determined from the equation ( Wair /e) ( ) graphite / ( / ) air S ρ μ ρ k, I K & = air en graphite i Vρ 1 g (1) air i where I is the measured ionization current, V is the volume of air cavity, ρ air is the density of dry air at the measurement temperature and pressure, W air /e is the mean energy expended in dry air by electrons per ion pair formed (33.97 J/C), g is the mean fraction of the initial kinetic energy of secondary electrons liberated by photons that is lost through radiative processes in air. The value of g for 60 Co has been reevaluated [2] and found to be consistent with the previous value [3] of 0.0032 adopted by the NIST, k i, are correction factors, S / ρ is the ratio of the mean collision mass stopping powers, ( ) graphite air ( μ / ρ) air en is the ratio of photon mass energy-absorption coefficients. graphite a Certain commercial equipment, instruments, and materials are identified in this work in order to specify adequately the experimental procedure. Such identification does not imply recommendation nor endorsement by NIST nor does it imply that the material or equipment identified is the best available for the purposes described in this work. 2/8

The value of the quantities and factors used in (1) can be found in earlier publications [1, 4 and 5]. The correction factors k i include: a humidity correction for the effects of moist air; a correction for the loss of ionization due to ion recombination, k sat ; and corrections for stem scattering, and wall absorption and scattering ( kwall ). The value of the wall correction, kwall, was revised by Seltzer and Bergstrom [2] resulting in a change of 0.90 % for the 137 Cs air kerma rates and 1.05 % for the 60 Co air kerma rates. This change was implemented on July 1, 2003. The decay-corrected daily values of the air kerma rate are checked routinely using a large variety of reference class chambers. A history is maintained for the calibration coefficient for these reference class chambers which can be traced back to the original reference date. The gamma beams used were collimated and are uniform over the area across the chamber. The source to detector distances, the beam size and air kerma rates used for each chamber are listed in Table 2a. Table 2a. Beam parameters at the NIST Chamber Model Serial Number Source Distance Beam Radius Air Kerma Rate A5 XY032721 / cm / cm / (Gy/s) 137 Cs 195 41 2.53 x 10 5 60 Co 150 29 7.90 x 10 7 137 Cs 307 58 1.06 x 10 6 TN32002 277 60 Co 307 59 1.86 x 10 7 3.2. Reference air kerma rates at the IAEA The gamma-ray measurements were made in the IAEA 137 Cs and 60 Co gamma ray beams. The air kerma rates were determined using the secondary standard ionization chamber LS-01. This chamber was calibrated in terms of air kerma for 137 Cs and 60 Co gamma ray beams at BIPM in June 2008. As of 1 November 2009, a correction reflecting the change of the BIPM standards has been introduced [6, 7]. The reference air kerma rates, K, at the IAEA were determined from the equation K & k N k (2) = I kt, P, H sat K BIPM where I is the measured current k T,P,H correction of the ionization chamber response for the environmental conditions k sat correction for the loss of ionization due to ion recombination N K air kerma calibration coefficient k BIPM correction to reflect a change of the primary standard [6]. 3/8

The air kerma rates at the reference point of the IAEA calibration facility are checked routinely with various ionization chambers as a part of the Quality Assurance system. A history of air kerma rates determined is maintained. The gamma beams used were collimated and are uniform over the cross-sectional area of the chamber volume. The source to detector distances, the beam size and kerma rates used for each chamber are listed in Table 2b Table 2b. Beam parameters at IAEA. Chamber Model Serial Number Source Distance Beam Radius Air Kerma Rate A5 XY032721 / cm / cm / (Gy/s) 137 Cs 200 50 1.07 x 10 6 60 Co 200 50 9.48 x 10 8 137 Cs 300 75 2.38 x 10 6 TN32002 277 60 Co 300 75 2.10 x 10 7 4. Calibration coefficients For each chamber, the air kerma calibration coefficient, N K was determined by evaluating the ratio N K = K & / I (3) where K & is the air-kerma rate at each calibration facility (determined by applying a decay correction to the original reference air-kerma rate value) and I is the ionization current measured with one of the chambers and corrected for influence quantities. The air kerma rate, K &, is expressed in units of Gy/s while the ionization current is expressed in units of C/s. The value of the ionization current I includes corrections to account for deviations of the ambient temperature and pressure from the NIST reference conditions of 295.15 K and 101.325 kpa, respectively. No correction for electronic recombination was applied to the measurements since the magnitude of the correction was the same at both facilities (of the order of 0.05 %) and therefore the effect on the comparison result was negligible. 5. Procedure, measurement conditions at the NIST and the IAEA. At both facilities each calibration was made by using the substitution method in which the transfer chambers listed in Table 1 are calibrated following the prior calibration of a check chamber. A check chamber is any of the reference class chambers owned by the IAEA and the NIST for which a significant amount of data has been collected over the years. In this 4/8

way the response and stability of the chamber is well known in the radiation beams in which they are used. A successful calibration of a check chamber ensures that the measurement system and the air kerma rates delivered meet the expected results. For all the measurements performed, the chambers are positioned with the stem perpendicular to the beam axis. The chambers are aligned relative to the source using the center of the detection volume using an alignment telescope. The uncertainty in the position is better than 0.1 mm. Since the measurement distances are of the order of 200 cm, the uncertainty in the position results in a negligible contribution to the overall uncertainty of the measurements (less than 0.01 %). The distances at which the detectors were positioned at both the NIST and the IAEA are listed in Tables 2a and 2b. The exact beam geometry was not prescribed in the protocol so that both participants could use the appropriate geometry depending on the field size and field uniformities available and the conditions that best represents routine calibrations at their facility. The influence of field size was not investigated for this comparison. At both facilities the chamber background current was measured before and after the chamber was exposed to the source of radiation. The average of the background currents was determined and subtracted from the measured ionization current produced in the chamber when exposed to the source of radiation. Typical background currents were of the order of 0.1 % of the ionization current. The integration time used both at the NIST and the IAEA was 60 s. At both facilities the measured ionization currents were normalized to standard values of temperature and pressure. The normalizing value for pressure at the NIST and the IAEA is 101 325 kpa. The normalization values for temperature are usually 295.15 K and 293.15 K at the NIST and the IAEA, respectively, but for the comparison, the normalization of 295.15 K was used. The humidity was monitored and recorded and was typically in a range of 30 %, however the currents were not corrected for humidity. Only the negative polarity measurements were used for the comparison results, resulting in a negative charge collected on the electrometers. No correction was applied to the calibration coefficient to account for polarity effect since at both facilities the same polarity was used. 6. Uncertainties The uncertainties associated with the measurements made at the NIST and the IAEA are listed in Tables 3a and 3b. The various sources of uncertainty are grouped according to Type A and Type B methods of evaluation, following the guidelines described elsewhere [8]. Type A evaluations are based on statistical analysis while Type B evaluations are based on scientific judgment. The uncertainty of the calibration coefficient, N K, is obtained essentially by combining the uncertainties of the standard air kerma rate, K & std, and the ionization current corrected for all influence quantities. As shown in the table, the relative fractional combined uncertainty of the calibration coefficients, N K, at the NIST is 0.0034 while the value at the IAEA is 0.0037. This results in relative combined uncertainties of 0.34 % and 0.35 %, respectively and relative expanded uncertainties of 0.68 % and 0.74 %, respectively. The relative expanded uncertainty is obtained by multiplying the relative combined uncertainties by a coverage factor of k = 2. The relative expanded uncertainty is considered to have approximately the significance of a 95 % confidence interval. Table 4 lists the relative combined uncertainty of the comparison ratios obtained in both gamma-ray beams. 5/8

Table 3a. Relative uncertainties associated with the measurements made at the NIST NIST Relative standard uncertainty u ia u ib Charge 0.0010 0.0010 Time 0.0005 Air density correction (temperature and pressure) 0.0003 Distance 0.0002 k sat, loss of ionization due to recombination 0.0001 0.0005 Probe orientation 0.0001 Humidity 0.0006 Chamber current, u i corrected for all influence 0.0010 0.0014 quantities 0.0017 K & Standard 0.0029 N K,NIST combined standard uncertainty 0.0034 Table 3b. Relative uncertainties associated with the measurements made at IAEA IAEA Relative standard uncertainty u ia u ib Air density correction (temperature) 0.0002 0.0003 Air density correction (pressure) 0.0005 0.0006 Chamber positioning 0.0001 Current measurement 0.0006 Chamber current, u i corrected for all influence 0.0005 0.0004 quantities 0.0006 K & Standard 0.0035 N K,IAEA combined standard uncertainty 0.0036 Table 4. Relative uncertainties associated with the comparison results Relative standard uncertainty N K,NIST 0.0034 N K,IAEA 0.0036 N K,IAEA / N K,NIST 0.0049 K & IAEA/ K & NIST 0.0049 u i 6/8

7. Results and discussion Table 5 shows a summary of the comparison results. It includes the actual value of the calibration coefficients N K,NIST and N K,IAEA obtained at the NIST and the IAEA facilities, respectively for each chamber in a given reference beam. The ratio of the calibration coefficients obtained with each chamber in a given reference beam is also shown in Table 5. The result of the comparison for each reference beam can be determined from the average of the ratios N K,IAEA /N K,NIST obtained with each chamber. The final comparison ratios are 0.999 and 0.997 for 137 Cs and 60 Co, respectively. This represents an agreement of 0.1 % for 137 Cs and 0.3 % for 60 Co between NIST and IAEA. Table 5: Comparison results Beam Chamber Model b b N N K,IAEA N K,IAEA / K,NIST N K,NIST per chamber c / (Gy/C) / (Gy/C) N K,IAEA / N K,NIST 137 Cs 60 Co A5 3.038 0 x 10 5 3.040 4 x 10 5 0.999 (5) TN32002 2.520 3 x 10 4 2.522 9 x 10 4 0.999 (5) A5 2.988 0 x 10 5 3.001 8 x 10 5 0.995 (5) TN32002 2.479 8 x 10 4 2.484 7 x 10 4 0.998 (5) 0.999 (5) 0.997 (5) 8. Conclusions The work presented here constitutes the first bilateral comparison between the IAEA and the NIST for air kerma from 137 Cs and 60 Co gamma ray beams. The realization of the air kerma standard at both metrological institutes is made differently. At the NIST the air kerma is realized from primary standards directly by evaluating (1). In the case of the IAEA the realization of the air kerma is obtained evaluating (2) using reference standards with traceability to the SI via the BIPM. Despite these different approaches the results obtained from the measurements performed in this work, show a good level of agreement of the air kerma standards maintained at both metrological institutes. An agreement of well within one standard uncertainty is obtained for the air kerma from 137 Cs and 60 Co protection-level gamma ray beams in this SIM.RI(I)-S1 supplementary comparison. b Numbers in subscript are included to minimize rounding errors when the air-kerma rates are used in calculations but do not represent the level of accuracy of the measurements. c The number in parenthesis represents the standard uncertainty on the final digit i.e. a value of 0.005. 7/8

8. References [1] Loftus, T.P., Weaver, J.T., 1974. Standardization of 60 Co and 137 Cs gamma-ray beams in terms of exposure. J. Res. Nat. Bur. Stand. 78A, 465-476. [2] Seltzer, S.M., Bergstrom, P.M., 2003. Changes in the U.S. primary standards for the air kerma from gamma-ray beams. J. Res. Natl. Inst. Stand. Technol. 108, 359-381. [3] M. Boutillon, Report CCEMRI (I)/ 85-18 to Section I of CCEMRI (Consultative Committee for Standards for the Measurement of Ionizing Radiation), Bureau International des Poids et Mesures, Sèvres, France (1985). [4] Minniti, R., NIST Quality Manual, Ionizing Radiation Division QM-II, Procedure 4. Calibration of gamma-ray radiation detectors. Gaithersburg, MD: NIST, 2004; [5] Minniti, R., Chen-Mayer H., Seltzer S. M., Huq S. M., Bryson L., Slowey T., Micka J. A., DeWerd L. A., Wells N., Hanson W. F., and Ibbott G. S. (2006). The US radiation dosimetry standards for 60 Co therapy level beams, and the transfer to the AAPM accredited dosimetry calibration laboratories. Med Phys 33:1074. [6] BIPM, private communication, 2009-Oct-14. [7] Allisy-Roberts P.J., Burns D.T., Kessler C., Measuring conditions used for the calibration of national ionometric standards at the BIPM, Rapport BIPM-2009/04, 20 pp [8] ANSI/NCSL Z540-2-1997, 1997. U.S. Guide to the Expression of Uncertainty in Measurement, American National Standard for Expressing Uncertainty. (NCSL, Boulder, CO). 8/8