PF-FLO REFERENCE TEST AT THE MARTIN-LUTHER UNIVERSITY HALLE-WITTENBERG

Similar documents
Figure 1 Enlargement of Powdered Activated Carbon by microscope.

Turbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas

Chapter 6 Pneumatic Transport and Standpipes

Numerical Simulation of Elongated Fibres in Horizontal Channel Flow

Vane Type Rotary Actuators Series Variations

Chapter 6 Pneumatic Transport

FIELD TEST OF WATER-STEAM SEPARATORS FOR THE DSG PROCESS

Standard Practices for Air Speed Calibration Testing

Controlling Downstream Particle Segregation with a Twisted Standpipe

Mixing Process of Binary Polymer Particles in Different Type of Mixers

1) Batching Methods (Add Weigh or Loss In Weight) based on batch dispensing time, time between batches, head room, floor space, accuracy and budget.

REPORT issued by an Accredited Testing Laboratory

FLOW MEASUREMENT. INC 102 Fundamental of Instrumentation and Process Control 2/2560

Applied Fluid Mechanics

Gravimetric coal dust firing and dry lignite firing in power plants

Fluidisational velocity, resistance of permeable material layer

Measurement of Dust Cloud Characteristics in Industrial Plants 1

Chapter 3. Experimentation and Data Acquisition

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

Energy transport in metal nanoparticle plasmon waveguides

DEM Simulation on the Mixing of Rice Husk Powder in a Screw Feeder with Various Screws

Applied Fluid Mechanics

ATOMISATION TECHNOLOGY TO MINIMIZE THE EFFECTS OF NOZZLE WEAR ON THE DROPLET SIZE

Technique for the electric and magnetic parameter measurement of powdered materials

An Essential Requirement in CV Based Industrial Appliances.

Lab Section Date. ME4751 Air Flow Rate Measurement

Solids Fractions and Flow Characteristics in a CFB as Measured with a Fiber Optic Probe

ME332 FLUID MECHANICS LABORATORY (PART II)

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

Table of Contents Why? (10 min.) How? (2 min.) Q&A (3 min.)

AEROSPACE ENGINEERING DEPARTMENT. Second Year - Second Term ( ) Fluid Mechanics & Gas Dynamics

Geophysical Applications GPR Ground Penetrating Radar

APPENDIX 1 DESCRIPTION OF HOT WIRE ANEMOMETER

EFFECT OF PARTICLES IN A TURBULENT GAS-PARTICLE FLOW WITHIN A 90 o BEND

ME332 FLUID MECHANICS LABORATORY (PART I)

Control Your Product Quality by Controlling Your Product's Conveying Velocity

Piping Systems and Flow Analysis (Chapter 3)

3D Simulation of the Plunger Cooling during the Hollow Glass Forming Process Model, Validation and Results

1P22/1P92 Exam Review Problems 2013 Friday, January 14, :03 AM. Chapter 20

Safety Engineering. -Static Electricity and Charge Accumulation-

Structural Engineering Research Group (SERG) April 10, 2013 Dielectric Modeling of Hydrated Cement Paste Panels

Anemometer Calibration Requirements for Wind Energy Applications

Application of an ultrasonic velocity profile monitor in a hydraulic laboratory

Introduction and Challenges

UNIT I ELECTROSTATIC FIELDS

Experimental Investigation on Segregation of Binary Mixture of Solids by Continuous Liquid Fluidization

Correlation Between Entry Velocity, Pressure Drop And Collection Efficiency In A Designed Stairmands Cyclone.

Flow Measurement in Pipes and Ducts COURSE CONTENT

Particle size and shape distribution of stable dust analysed with laser diffraction and imaging technique

Experiment No.4: Flow through Venturi meter. Background and Theory

CHAPTER 4 TEST REVIEW

2 Navier-Stokes Equations

Solids Flowmeters Can you go with the Flow? Solids Flowmeters for Industrial Applications

Comparison of discharge measurements - Thermodynamic to US Clamp- On, stationary US and Needle Opening Curve

Experimental and Numerical Investigation of Two- Phase Flow through Enlarging Singularity

= A x (t) + B utt), by d{ _-=-=-

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

Strategy in modelling irregular shaped particle behaviour in confined turbulent flows

Use of Near Infrared Spectroscopy for in- and off-line performance determination of continuous and batch powder mixers: opportunities & challenges

UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS General Certificate of Education Advanced Level

A) n 1 > n 2 > n 3 B) n 1 > n 3 > n 2 C) n 2 > n 1 > n 3 D) n 2 > n 3 > n 1 E) n 3 > n 1 > n 2

VK2 volume flow controller Overview

Lesson 37 Transmission Of Air In Air Conditioning Ducts

Photonic/Plasmonic Structures from Metallic Nanoparticles in a Glass Matrix

NORTH SEA FLOW MEASUREMENT WORKSHOP 2004 In. St Andrews, Scotland

Manufacturing Equipment Control

CFD Simulation of Dilute Gas-Solid Flow in 90 Square Bend

CONTROL OF MICROWAVE HEATING IN RECTANGULAR WAVEGUIDE

STUDY OF DESIGN OF CYCLONE SEPARATOR UNDER COLLECTION EFFICIENCY AND AIR DENSITY EFFECT

Experiment (4): Flow measurement

Flow Measurement in Pipes and Ducts

Sound Rating of Ducted Air Moving and Conditioning Equipment

Liquid Feed Injection in a High Density Riser

SELECCIÓN DE CAUDALIMETROS. Tablas de Fabricantes

AP Physics C Mechanics Objectives

ROLLER BEARING FAILURES IN REDUCTION GEAR CAUSED BY INADEQUATE DAMPING BY ELASTIC COUPLINGS FOR LOW ORDER EXCITATIONS

Wet Collectors. Type 1: Spray Chamber Scrubber 10/30/2013. EVE 402 Air Pollution Generation and Control. Chapter #5 Lectures (Part 5)

Fig.8-1 Scheme of the fluidization column

Relativistic Behavior Detection through Electron Acceleration

Modeling Complex Flows! Direct Numerical Simulations! Computational Fluid Dynamics!

Cyclones. Vane Axial Cyclone 10/30/2013. EVE 402 Air Pollution Generation and Control. Chapter #5 Lectures (Part 4) A mechanical gas cleaning device

ACCELERATOR MASS SPECTROMETRY AT THE LUND PELLETRON ACCELERATOR

DESIGN AND PERFORMANCE OF THE CONVERGING-DIVERGING VORTEX FLOWMETER

6.5 Optical-Coating-Deposition Technologies

HVAC Clinic. Duct Design

Experimental Investigation of the Velocity Distribution near the Swirl Generator of a Uniflow Cyclone for Performance Data Prediction

Chapter 11. Vibrations and Waves

W 18e Heat Capacity Ratio γ

Outline of College Physics OpenStax Book

EQUIPMENT Beta spectrometer, vacuum pump, Cs-137 source, Geiger-Muller (G-M) tube, scalar

CAMSIZER. Catalyst Beads, Powders, Extrudates Presentation. Dynamic Image Analysis ISO conform

Performance of a Venturi Scrubber in the removal of fine powder. from a confined gas stream

: PM CEM Optical Technology and PS-11

1 One-dimensional analysis

Process Control, 3P4 Assignment 6

Microwave Phase Shift Using Ferrite Filled Waveguide Below Cutoff

Chapter Four fluid flow mass, energy, Bernoulli and momentum

SAMPLE PAPER-05 (solved) PHYSICS (Theory) Class XII. Time allowed: 3 hours Marks: 70

Effects of Horizontal Vibration on Hopper Flows of Granular Materials. R.C. weathers', M.L. ~unt', C.E. ~rennen~, A.T. ~ e e ~ and C.R.

J10M.1 - Rod on a Rail (M93M.2)

Transcription:

PF-FLO REFERENCE TEST AT THE MARTIN-LUTHER UNIVERSITY HALLE-WITTENBERG Martin-Luther-Universität Halle-Wittenberg AMC Power PROMECON Fachbereich Ingenieurwissenschaften 15 Hopper Avenue Lehrstuhl für Mechanische Verfahrenstechnik Santa Rosa, California 9543 699 Halle (Saale) U.S.A. Germany

PF-FLO REFERENCE TEST AT THE MARTIN-LUTHER UNIVERSITY HALLE-WITTENBERG CONTENTS Page 1. Introduction... 1 2. Description of the Test Facilities... 3 2.1 The Testing Plant... 3 2.2 The Pf-FLO Mass Flow Measurement... 4 2.2.1 Density measurement... 4 2.2.2 Velocity measurement... 5 2.2.3 Calculation of the Mass Flow... 6 2.3 Pf-FLO Test Configuration... 6 2.4 The Test Medium... 8 2.5 Feeder Calibration... 9 3. Testing Procedure... 11 4. Results... 14 4.1 Pf-FLO Measurement Accuracy... 14 4.1.1 Absolute Deviation... 15 4.1.2 Repeatability... 16 4.2 Influence of the Particle Size... 17 4.2.1 Velocity Measurement... 17 4.2.2 Density Measurement... 19 4.2.3 Mass flow measurement... 2 5. Abstract... 23

Figures Page Fig. 2.1: Schematic drawing of the test plant... 3 Fig. 3.1: Range of pf-concentrations based on feeder mass flow and transport air flow... 11 Fig. 3.2: Density measurement... 12 Fig. 3.3: Velocity measurement... 12 Fig. 3.4: Resulting mass flow and feeder signal... 12 Fig. 3.5: Mass flow of feeder versus Pf-FLO... 12 Fig. 4.1: Evaluation of all test runs with 66 / 225 µm particles... 14 Fig. 4.2: Repeatability of channel for 66-225 µm particles... 15 Fig. 4.3 Averaged particle velocities at channel... 17 Fig. 4.4: Acceleration along the test duct of the 225 µm particles... 18 Fig. 4.5: Influence of the mass flow on the velocity of the particle mix in Test IV-VI.. 18 Fig. 4.6: Density measurement with 66 µm particles Test V... 19 Fig. 4.7: Density measurement with 225 µm particles Test I... 19 Fig. 4.8: Influence of the particle size on the Pf-FLO measurement... 2 Fig. 4.9: Estimated deviation by modeled particle size distribution... 22 Tables Table 2.1:Bulk density and frequency shift for fixed-bed powder of pulverized black Page coal and glass particles... 8 Table 3.1:Test run number for each particle size... 12 Table 4.1:Standard deviation and mean error for individual particle fractions... 15 Table 4.2:Standard deviation of the individual channels with 66 µm particles... 16 Table 4.3:Standard deviation of the individual channels with 225 µm particles... 16 Table 4.4:Standard deviation of the individual channels with 66-225 µm particle mix. 16 Table 4.5:Ratio of arbitrary units to mass and the resulting mass frequency factor k fd for each particle fraction... 21

1. Introduction Measurement of particle concentration or mass flow rate in pipeline systems (i.e. pneumatic conveying) is essential for numerous technical applications, such as conveying of pulverized coal in power plants or conveying systems in cement factories. Of major importance is the detection of the particulate flow in the entire cross-section of a pipe. In the past this has generally only been achieved in pipe elements where the particulate concentration is homogeneously distributed over the entire cross-section. For such a measurement different techniques are available, namely extractive methods utilizing probes and non-extractive methods employing electromagnetic waves or particle charging. From the first inspection an isokinetic sampling probe seems to be the simplest approach, however, in order to measure the particulate flux in the entire pipe section it is necessary to systematically position the extraction probe at defined locations across the entire pipe cross-section. Due to probe erosion damage, extractive sampling is only suitable for periodic measurement. For continuous measurement, non-extractive methods are more favorable, where the sensing instrumentation is mounted in-situ. One approach is the detection of the electrostatic charge of moving particles. Unfortunately, the resulting signal is not only affected by particle concentration, but also by gas temperature and particle velocity. The method used for the investigation documented in this report is based on utilizing microwaves emitted and detected by screw-in sensors. The generated microwave field covers the entire pipe cross-section and hence allows the determination of the averaged particle concentration over that cross-section. The principle of the method will be outlined below. The microwave based experiments were conducted on an air-particle flow loop established at the Lehrstuhl für Mechanische Verfahrenstechnik of the Martin-Luther University Halle-Wittenberg. In order to consider different conveying conditions, probes were installed at multiple locations of the conveying pipe, namely in an upward flow with almost homogeneous dust distribution, behind a vertical-horizontal bend where roping is likely to occur, and in an almost fully developed state of a horizontal pipe. The particle introduction was achieved using a calibrated screw feeder, which also allowed the comparison of the particulate mass flow with the result from the microwave instrument. Since the handling of pulverized coal could not be safely performed at the test facility, spherical glass beads of two different mean diameters were used as coal substitutes. It is acknowledged that the different material density and particle shape of the glass particles results in a slightly different conveying behavior, primarily in the form of lower particle velocities, but otherwise the general flow behavior of coal and glass particle is extremely similar.

This report presents a description of the test facility and the measurement principle. The measurements are presented in comparison with the calibrated screw feeder, and a detailed discussion of the accuracy achieved with the microwave instrument is provided.

2. Description of the Test Facilities The reference test was carried out at the Merseburg test plant. The test facility is designed with a closed loop for the particle flow and an open end for the transport air. This arrangement ensures particle recycling via a cyclone back to the feeder without significant particle mass loss, for re-introduction at a controlled rate/concentration. For safety reasons the test plant was operated with glass beads of two different diameters instead of pulverized coal. Particle load and transport air velocity were varied during the test series in a range simulating that which naturally occurs with pneumatically transported coal (see test matrix, Figure 3.1 and Table 3.1). 2.1 The Testing Plant The test duct layout is drawn in Figure 2.1. Two rotary piston blowers, operating in parallel and controlled by fan speed frequency converters, providing a velocity range of about 46 to 92 ft³/s for the transport air. Air Outlet Ch 3 Ch 2 16.5 ft. Ch 1 Air Inlet Cyclone Rotary Valve 1 ft. Ch Bagfilter Hopper Screw Feeder Fig. 2.1: Schematic drawing of the test plant The particles are introduced to the airflow by a screw feeder, transported through the pipe and separated in a cyclone. Out of the cyclone the separated particles are fed

by a rotary valve back into the hopper of the screw feeder. The transport air is exhausted through a bag filter. The screw feeder is frequency controlled over a range of 35 rpm. The horizontal run downstream of the feeder has a rectangular cross section, whereas the vertical and the upper horizontal pipes where the Pf-FLO measurements are located are round pipes having an inner diameter of 4.86. The airflow velocity is measured by a multi-point Pitot probe positioned upstream of the feeder. In addition, the airflow static pressure and temperature are also monitored. 2.2 The Pf-FLO Mass Flow Measurement The Pf-FLO system independently measures the density and velocity of pulverized fuel in a two-phase flow. After a zero calibration of the empty transport pipe is performed for the density measurement process, absolute mass flow can be calculated using the product of the separate density and coal velocity signals. 2.2.1 Density measurement Using the pipe as a wave-guide, the particulate concentration or density is measured with transmitted microwaves that cover the full cross section of the pipe. Starting with the known microwave transmission characteristic determined during empty pipe zeroing, the varying dielectric load caused by changing pulverized fuel (pf) concentrations produces a measurable frequency shift. The basics of this measurement can be described as follows 1) : The cut-off frequency (λ cut ) of a round wave-guide is in this application the frequency of the H 11 mode. The wavelength of the H 11 mode is a function of the diameter of the pipe. Equation 1. λ cut π D = = 1. 71 D Χ mn wherein D is the diameter of the pipe and X mn is the solution of the Bessel function. 1) Kummer: Grundlagen der Mikrowellentechnik, chapter 3.4 and 4.3; Berlin: 1986

The frequency (ƒ) of the H 11 mode depends on the dielectric ε r and the magnetic µ r properties of the volume in the wave-guide. Equation 2. f = ε r µ c r λ cut where c is the constant for the speed of light An unloaded pipe filled with air has a ε r of 1 and a µ r of 1. Hard coal has an ε r of 4 and µ r of 1. The volumetric ratio of pulverized coal to air at a coal concentration of.312 lb/ft³ is 1 to 25. Since the resulting ε r changes between loaded and empty pipe in terms of 1/25 the series expansion of Equation 2 can be used with its linear term. This gives a linear relation between the frequency and pf load within the concentration range typically found in coal fired power plants. The Pf-FLO system couples microwaves in the range of the cut-off frequency into a pipe section using a pair of sensors, one sensor functioning as the transmitter and a second sensor as the receiver. The exact frequency of the H 11 mode is calculated by scanning the transmitted microwave signal amplitudes. A change in the concentration of pf in a given pipe changes the measured microwave frequency: The higher the concentration, the lower the frequency. The frequency shift caused by the pf is calculated by subtracting the frequency f ε of the loaded pipe from the frequency f of the empty pipe. This frequency shift is transformed into a density signal (ρ) by the frequency density factor k fd. Equation 3. ( f f ) k fd ρ = ε Where f is determined by the empty pipe zeroing process. Changes in pipe diameter caused by temperature do affect the measured frequency. Temperature compensation of the measured frequency uses the pipe surface temperature and the linear expansion factor for that pipe material. 2.2.2 Velocity measurement The velocity measurement uses a cross-correlation method for comparing the stochastic signals of the electrostatic charged particles at two sensor locations of known separation (see Figure 2.2). An evaluation of the velocity sensor signals gives the time shift or time of flight (τ m ). Using the distance between the sensors (L), solid particle velocity (v s ) can be calculated as follows: Equation 4. L vs = τ m

By using this method only particle velocity is measured, which in most instances differs from and is slower than the transport air velocity in a two-phase flow. This difference, or velocity slip, is a function of such factors as pipe configuration, specific weight and size of particles. Signal 1 Signal 2 Velocity = L / τ m Cross-correlation Fig. 2.2: Velocity measurement principle: signals and resulting cross-correlation function 2.2.3 Calculation of the Mass Flow The mass flow dm dt measurement as follows: is calculated from the density and the particulate velocity Equation 5. dm dt = ρ v s The Pf-FLO system is calibrated to a known mass flow of the mill or pipe by adjusting the frequency density factor k fd in Equation 3, which in turn depends on the pipe diameter. The factor k fd is kept constant for all pipes with the same diameter. 2.3 Pf-FLO Test Configuration The 4.86 diameter test duct pipe has a cut-off frequency of approximately 1.4 GHz. The standard microwave generating unit of the Pf-FLO system has been selected to provide frequencies up to only a 1 GHz level required for the range of larger coal pipe

sizes found in power plants. For the test runs conducted it was necessary to replace the standard model generator with a similar model having an extended frequency range of up to 2. GHz. Corresponding to the smaller inner diameter of the test duct, the sensor antenna was also scaled down in length. Distances between sensors and rods in the test runs were the standard distances based on a pipe having a diameter D, as show in Figure 2.3. 5.25 D.87 D 1. D.75 D.75 D 1. D.87 D rod sensor Fig. 2.3: Arrangement of sensors and rods at individual measurement locations The Pf-FLO system uses wear resistant Tungsten Carbide rods to keep the propagation of the microwaves within the certain measurement zone of the pipe. Without the rods, the density measurement would be disturbed by reflected signals caused by pipe bends, orifice plates, isolation valves, etc., located upstream and/or downstream of the measurement zone. The optional fifth rod perpendicular to the sensors and located at their midpoint provides an additional signal short cut for depressing the propagation of 9 polarized H 11 modes. Without knowing the actual mass frequency factor for the test pipe size, all channels were initially set to a.u. k fd = 5 Hz m This factor was kept constant for all measurements in the test. The resulting units for dm measured density (ρ) and mass flow are in arbitrary units [a.u.]. dt

2.4 The Test Medium The test plant could not be used with black coal for safety reasons. Therefore, glass spheres were used, with such properties as particle size, dielectric constant, and electrostatic charging similar to pulverized coal. Typically 85 % 95 % by weight of pulverized coal particles downstream of the mill s classifier are smaller than 9 µm and.3 % or less are bigger than 225 µm. The two glass particle sizes of 66 µm and 225 µm used for this test represent the main fraction and the biggest possible size fraction of particles in coal pipes. The manufacturer of the glass beads specifies a glass density of 158.6 lb/ft³ and an ε r of 2.28 at visible light. The ε r may be slightly different for microwaves due to dispersion. The dielectric properties of milled coal and the glass spheres were tested in a microwave resonator chamber. It was found that the frequency shift in this measurement was dependent upon the dielectric properties on the bulk density of the pulverized medium. By calculating the frequency shift per mass, the influence of the sphere packing were eliminated. The results are displayed in Table 2.1. Medium Bulk density [lbs/ft³] Frequency shift/ mass [MHz/lb] Glass spheres 88.1 124.1 Black coal (Primero) 35.8 2.1 Black coal (Blumenthal) 35.8 193.9 Black coal (Knurrow) 41.8 193.5 Table 2.1: Bulk density and frequency shift for fixed-bed powder of pulverized black coal and glass particles The frequency shift at the same mass flow caused by glass is about 2/3 of the tested coal. Therefore, the expected frequency shift for the mass flow measurement will only be about 1/3 less for glass than for coal with the same mass. This ensures a good comparability between the test data obtained with glass particles used as the test medium versus that which would have been obtained had coal been able to be used for the test medium. The density for raw coal is between 78. and 81.8 lbs/ft³. Taking this density into account, glass particles of the same size are about two times heavier than coal

particles. The weight differential plus the shape of the particles, spherical for glass and polyhedral for coal, give glass aerodynamic properties which result in a greater velocity differential or slip between the airflow and the glass particles. The electrostatic charging depends on particle collisions and particle conductivity. The velocity measurement needs a certain amount of electrostatic charge to correlate the sensor signals into a reliable time of flight measurement. Charging signal strengths for both size glass beads and bead mixtures were sufficiently high to obtain accurate time of flight measurements. Induced by the substantially greater number of particle amount within the airflow, the signal strength of 66 µm particles was about five times higher than for the 225 µm particles. 1.6 Gravimetric Particle Size Distribution Rel. particle distribution [%] 1.4 1.2 1.8.6.4.2 5 1 15 2 25 3 Particle size [µm] Fig. 2.4: Particle distribution as a function of particle size for the 5/5 mix of 66 µm and 225 µm particles Beside the pure 66 µm and 225 µm particles, a 5/5 mix by weight was also tested. Figure 2.4 shows the gravimetric distribution of particle sizes. 2.5 Feeder Calibration To calibrate the feeder, glass beads were fed by the frequency controlled feeder into a container for 3 seconds and their mass was weighed. This procedure was repeated twice for each particle size in steps of 5 rpm from to 35 rpm. The average of both sets of measurements was used for the feeder calibration.

The repeatability of the feeder calibration was then tested by 1 individual measurements with the 66 µm particles at 15 rpm. They were all in the range of ±.9 % by weight. This was acceptable since the aim of the tests was not to examine the characteristics of the screw feeder. And with all four sensor locations measuring physically the same airflow/particle mixture, any scattering of the feeder is eliminated as a common variable..55 Feeder Calibration.44 Mass flow [lbs/s].33.22.11 66 225 µm mix 225 µm 66 µm 5 1 15 2 25 3 35 4 Feeder speed [rpm] Fig. 2.5: Mass flow versus feeder speed for different particle fractions The mass flow of the feeder is shown in Figure 2.5 for the specific particle fractions. The mass flow at a particular feeder speed depends on the particle size distribution. The mix of the two size fractions has the tightest packing and thus shows the highest mass flow. The 66 µm and 225 µm particles have different mass flows since for particles <1 µm, adhesion forces influence the flowability within the screw feeder.

3. Testing Procedure The test runs have been made under the aspect of realistic airflow velocities and particle concentrations. Within the capacity of the fan, three velocity levels were chosen at 72 ft/s, 82 ft/s, and 92 ft/s, representing normal transport velocities in utility plants. With constant air velocities the feeder speed was varied between - 3 rpm in steps of 5 rpm..5 Particle Concentration Range.44 Concentration [lbs/ft³].37.31.25.19.13.6 5 1 15 2 25 3 35 Feeder speed [rpm] Fig. 3.1: Range of pf-concentrations based on feeder mass flow and transport air flow The pf concentrations in utility plants usually range between.12 to.31 lbs/ft³. Figure 3.1 shows the range of the expected pf concentration based on the ratio of feeder mass flow and the airflow during the tests. Table 3.1 gives an overview of the different test runs: From the total number of 15 test runs there were six runs with the 66 µm particles, six runs with the particle mix and three runs with the 225 µm particles.

Particle Size Test Numbers 66 µm I,VI II,V III,IV 225 µm I II III 66-225 µm mix I,IV II,V III, VI 72 ft/s 82 ft/s 92 ft/s Gas Velocity Table 3.1: Test run number for each particle size The following diagrams illustrate the data acquired for all test runs: Diagram Figures 3.2 and 3.3 show density and velocity measurement, and Figure 3.4 shows the resulting mass flow of the 66 225 µm particles of Test Number V. Each feeder step was kept constant for at least 15 minutes to get about 2 individual measurements. From the last 15 measurements of each feeder step the average was taken and plotted against the feeder mass flow in Figure 3.5. Density [a.u./ft] 61 54 48 42 36 3 24 18 12 6 13:4 13:12 Densities 66-225 µm, Test V 13:2 13:29 13:37 13:46 13:54 14:2 14:11 14:19 14:28 14:36 14:44 14:53 15:1 CH CH 1 CH 2 CH 3 feeder 15:1 15:18 Fig. 3.2: Density measurement Fig. 3.3: Velocity measurement Mass flow [a.u./s] 45 4 35 3 25 2 15 1 5 13:4 13:11 Mass Flows 66-225 µm, Test V 13:19 13:26 13:34 13:41 13:49 13:57 14:4 14:12 14:19 14:27 14:34 14:42 14:49 14:57 15:5 CH CH 1 CH 2 CH 3 feeder 15:12 15:2 4 35 3 25 2 15 1 5 35 3 25 2 15 1 5 Feeder speed [rpm] Feeder speed [rpm] Fig. 3.4: Resulting mass flow and feeder signal Fig. 3.5: Mass flow of feeder versus Pf-FLO Velocity [ft/s] Pf-FLO Mass flow [a.u./s] 98 82 66 49 33 16 45 4 35 3 25 2 15 1 5 13:4 13:11 Velocities 66-225 µm, Test V 13:19 13:26 13:34 13:41 13:49 13:57 14:4 14:12 14:19 14:27 14:34 Mass Flow of Feeder vs. Pf-FLO, 66-225 µm, Test I - VI 198 397 595 793 992 119 1389 1587 Feeder mass flow [lbs/hr] CH CH 1 CH 2 CH 3 CH CH 1 CH 2 CH 3 14:42 14:49 14: 57 15:5 15:12 15:2

All test runs have been plotted as displayed in Figure 3.5. As there is only a constant factor between [a.u./s] and [g/s], a unified y-axis scaling was used to help evaluate the influence of different particle sizes (see also Figure 4.8).

4. Results 4.1 Pf-FLO Measurement Accuracy Figures 4.1 and 4.2 illustrate results only for the 5/5 particle mix. Results for other particle fractions are listed in the tabulations in Sections 4.1.1 and 4.1.2. 45 Mass Flow of Feeder versus Pf-FLO; 66-225 µm, Test I - VI 4 Pf-FLO mass flow [a.u/s] 35 3 25 2 15 1 5 198 397 595 793 992 119 1389 1587 Feeder mass flow [lbs/hr] Channel Channel 1 Channel 2 Channel 3 lin average Standard deviation [a.u./s]: 158 Fig. 4.1: Evaluation of all test runs with 66 / 225 µm particles The diagram in Figure 4.1 shows the evaluated results for all measuring channels versus the feeder mass flow. The mass flow signals of all channels were averaged for each feeder step and one particle fraction, and a linear coefficient was determined for it. With this coefficient the linear average was calculated as it can be seen in the diagram, indicated with lin. average. Based on the linear average the standard deviation was determined for each measuring channel as listed in Table 4.1. The repeatability of one channel for all tests and one particle size is exemplarily displayed in Figure 4.2. For other measuring channels and particle fractions, see tabulations in Section 4.1.2.

45 Repeatability of Channel ; 66 225 µm Particles, All Tests (Standard Deviation: 3.3 %) 4 Pf-FLO mass flow [a.u./s] 35 3 25 2 15 1 5 Test No. I 72 ft/s Test No. II 82 ft/s Test No. III 92 ft/s Test No. IV 72 ft/s Test No. V 82 ft/s Test No. VI 92 ft/s 397 793 119 1587 Feeder mass flow [lbs/hr] Fig. 4.2: Repeatability of channel for 66-225 µm particles 4.1.1 Absolute Deviation The standard deviation of one particle fraction from the linear average of all channels is listed in Table 4.1. The errors in % refer to the maximum mass flow determined at 3 rpm feeder speed. Particle Size Channel Test No. Number of Measurements Standard Deviation [a.u./s] Max. Mass Flow [a.u./s] Mean Error % 66 µm CH - CH3 Test I - VI 144 19 2462 4.4% 225 µm CH - CH3 Test I - III 72 132 352 3.8% 66-225 µm CH - CH3 Test I - VI 144 158 4 3.9% Table 4.1: Standard deviation and mean error for individual particle fractions

4.1.2 Repeatability The relative deviation of one channel in all tests shows its repeatability. This includes the scattering of the feeder but excludes systematic deviations from one channel in comparison to the others. Results for each channel are listed in the Tables 4.2 to 4.4. Standard Deviation Number of To Linear Average Mean Error Channel Test No. Measurements [a.u./s] % CH Test I - VI 36 75.1 3.1% CH 1 Test I - VI 36 92.5 3.8% CH 2 Test I - VI 36 93.4 3.8% CH 3 Test I - VI 36 72.9 3.% The error in % refers to the maximum mass flow at 3 rpm: 2462 [a.u./s] Table 4.2: Standard deviation of the individual channels with 66 µm particles Standard Deviation Number of To Linear Average Mean Error Channel Test No. Measurements [a.u./s] % CH Test I - III 18 132.3 3.8% CH 1 Test I - III 18 65. 1.8% CH 2 Test I - III 18 151.9 4.3% CH 3 Test I - III 18 111.7 3.2% The error in % refers to the maximum mass flow at 3 rpm: 352 [a.u./s] Table 4.3: Standard deviation of the individual channels with 225 µm particles Standard Deviation Number of To Linear Average Mean Error Channel Test No. Measurements [a.u./s] % CH Test I - VI 36 131.7 3.3% CH 1 Test I - VI 36 114.1 2.9% CH 2 Test I - VI 36 163.9 4.1% CH 3 Test I - VI 36 141.2 3.5% The error in % refers to the maximum mass flow at 3 rpm: 4 [a.u./s] Table 4.4: Standard deviation of the individual channels with 66-225 µm particle mix

4.2 Influence of the Particle Size Another purpose of the tests was to quantify the influence of particle sizes. As the 225 µm particles can only be found in smaller percentages in pulverized coal, it is a practical fraction to resolve particle size dependent influences on density and velocity measurement. The transferability of the results to the operating condition of coal fired power plants have to be viewed in relation to the real particle size distributions in coal pipes. In Section 4.2.3 the results out of the tests are evaluated. 4.2.1 Velocity Measurement Measured Particle Velocities at Channel 98 Particle velocity [ft/s] 82 66 49 33 16 225µm 66µm 66-225 µm mix 72 ft/s gas velocity 82 ft/s gas velocity 92 ft/s gas velocity Fig. 4.3 Averaged particle velocities at channel The measurements with 225 µm particles showed a difference of about 16.4 ft/s between airflow and particle velocity. This is due to the change in the aerodynamic properties which increase the slip between particles and gas. The weight of particles changes proportional to D³ but the cross section only changes proportional to D². The velocity of the 66 µm particles, as well as the 5/5 mix, was found to be very close to the airflow velocity. The reason can be found by comparing the electrostatic signal strength. The electrostatic signal strength of the 225 µm particles was found to be significantly lower than for 66 µm. But the particle number increases with the relation of particle diameters to the power of three (see above). With a 5/5 particle mixture by weight, the number of 66 µm particles is about 28 times greater than for the 225 µm particles. The cross correlation method resolves the time shift of the sensor signals by comparing their highest identity. If the signal strength of two time

shifts is of the same order, it might be possible to distinguish between the two velocities. In case of the particle mix the signal strength of the 225 µm particles was below the noise signal level of the 66 µm particles. Therefore, it is obvious that only the velocity of the 66 µm particles has been measured. The error in relation to the realistic particle size distribution is estimated in Section 4.2.3. 98 Velocities of the 225 µm Particles 82 velocity [ft/s] 66 49 33 CH CH1 CH2 CH3 16 72 ft/s gas velocity 82 ft/s gas velocity 92 ft/s gas velocity Fig. 4.4: Acceleration along the test duct of the 225 µm particles In the tests which measured 225 µm particles only, channel 3 was found to have higher velocities than the other channels. This can be explained by the position of this sensor pair located at the end of the horizontal test duct with the longest straight run after a bend (see Figure 2.1). This leads to a certain acceleration, especially for the bigger sized particles. Influence of Mass Flow on Velocity of the Particle Mix 98 92 velocity [ft/s] 85 79 72 CH CH 1 CH 2 CH 3 67 397 793 119 1587 Feeder [lbs/hr] Fig. 4.5: Influence of the mass flow on the velocity of the particle mix in Test IV-VI

Figure 4.5 shows the influence of the mass flow on particle velocity. This effect, here illustrated for the particle mix, is obvious when the averaged velocity of each feeder step is plotted over the mass flow as it is done in Figure 4.5. Each bundle of the four channels represents one step of the airflow velocity. The higher the airflow velocity the higher the influence from pf load in the pipe. Channel 2 with the shortest distance from a bend seems to be affected most. It is assumed that this effect is related to particle interaction between 66 µm and 225 µm particles, the latter having significantly lower velocities. 4.2.2 Density Measurement Density [a.u./ft] 42.7 36.6 3.5 24.4 18.3 6.1 13:32 13:41 Densities 66 µm Particles, Test V CH CH 1 CH 2 CH 3 feeder 13:5 13:58 14:7 14:16 14:25 14:34 14:42 14:51 15: 15:9 15:18 15:27 15:36 15:45 15:54 Fig. 4.6: Density measurement with 66 µm particles Test V. Densities 225 µm Particles, Test I 4 35 3 25 2 15 1 5 Feeder speed [rpm] Density [a.u./ft] 76.2 61. 45.7 3.5 15.2 1:14 1:2 Fig. 4.7: CH CH 1 CH 2 CH 3 feeder 1:27 1:33 1:4 1:46 1:53 1:59 11:6 11:12 11:19 11:25 11:32 11:38 11:45 11:51 11:58 12:4 12:11 12:17 Density measurement with 225 µm particles Test I 4 35 3 25 2 15 1 5 Feeder speed [rpm]

The diagrams in Figures 4.6 and 4.7 show the particle size dependent scattering of the densities in two test runs. The more extended scattering of the density signal for 66 µm particles also increases with the load or particle numbers. With the 225 µm particles there was less scattering although the densities in Figure 4.7 were nearly twice as high. The fluctuation quantity for the particle mix is higher than the one for the 225 µm, but less than the one for the 66 µm particles and was also influenced by the load (see Figure 3.4). The scattering of measurement is proved to be realistic and relates to the density fluctuations of the particle flow. The different behavior can be explained with the mean free path between particle collisions. The 66 µm particles have less particlewall collisions but more particle-particle collisions in a smaller volume. Local high and low density concentrations do not average out within the pipe volume that has been measured. 4.2.3 Mass flow measurement 45 Influence of the Particle Size 4 35 Ratio ([a.u.] to [lbs]): 2.734 2.885 Pf-FLO mass flow [a.u./s] 3 25 2 15 1 5 2.356 lin average 66 µm lin average 2 µm lin average 66-2 µm 198 397 595 793 992 119 1389 1587 Feeder mass flow [lb/hr] Fig. 4.8: Influence of the particle size on the Pf-FLO measurement

The diagram in Fig. 4.8 shows the ratio of arbitrary units [a.u./sec] to mass [g] taken from the linear averages of the four measurement channels. This factor was found to be a function of the used particle fraction. The temporarily applied arbitrary units to frequency factor of 5 (see section 2.3) has to be divided by this ratio to get the calibrated mass flow signal for each particle fraction. The resulting k fd is displayed in Table 4.5. Particle Fraction ratio [a.u./lbs] k fd [g/m khz] 66 µm particles 2.3562 26.7 225 µm particles 2.7342 23. 66-225 µm particle mix 2.8854 21.8 Table 4.5: Ratio of arbitrary units to mass and the resulting mass frequency factor k fd for each particle fraction The deviation of 13.8 % between the 66 and 225 µm particles can be regarded as dependent on particle size. The deviation of the 66 225 µm particle mix is due to the deviation of the velocity measurement described in Section 4.2.1. The results have to be compared with the real particle size distribution in a coal pipe after classifier. These deviations have influence only on the absolute accuracy but not on the relative accuracy between several pipes of one mill since a segregation of particle fractions between several pipes is not probable. The following calculations are linear estimations of the error in real particle size distributions within the results of the tests: For an examplary particle distribution of 15% >9 µm and.2 % >225 µm it was assumed to have a discrete mixture out of 84.8 % 66 µm particles, 15 % 145 µm particles and.2 % 225 µm particles. The relation of the diameters was taken to interpolate the velocity of the 145 µm particles linear between the velocities of the 66 µm and 225 µm particles. Also the relation of the diameters was taken to interpolate the deviation of density measurement for 145 µm particles linear between the densities of the 66 and 225 µm particles.

Deviation caused by Particle Size Relative mass flow [%] 12 1 8 6 4 2 1. 11.2 11.8 12.5 225 µm 145 µm 66 µm 1% 66 µm 89.9% 66µm 1.% 145µm.1% 225µm 84.8% 66µm 15.% 145µm.2% 225µm 79.% 66µm 2.% 145µm 1.% 225µm Fig. 4.9: Estimated deviation by modeled particle size distribution The results of the estimation are shown in Figure 4.9. If the distribution changes from column 2 to 3, the estimated error is about.6 % of the mass flow. The distribution of the last column shows a 2.5 % error but this change in particle distribution is meant to be quite unrealistic in an optimized milling process and will also influence the combustion badly.

5. Abstract A reference test at the pneumatic conveying test plant of the Lehrstuhl für Mechanische Verfahrenstechnik at the University of Halle-Wittenberg was established to prove the accuracy of a flow measurement system for air-solid flows. The test facility consists of a calibrated screw feeder, a pipe system with vertical and horizontal elements, and the particle separation equipment. Glass beads were used as a test medium whose physical properties are comparable to coal dust if taking into account the measurement principle. In addition to the single sized test materials with diameters of 66 µm and 225 µm, a 5/5 mixture by weight of both particle sizes was used. The experimental matrix for the tests covered the usual operational range for the throughput and the velocity in coal pipes of power plants. In total, four measurement instruments were located at two locations in the upward run and two locations in the horizontal run of the test pipe. From the measured density and velocity signals of the particles the mass flow was calculated in each case and compared with the calibrated feeder signal. The measuring error was related to a single standard deviation. As a result, the measured deviation from the feeder signal is < 4.5 %; this applies to the entirety of all four measuring points and all particle fractions. For individual sensors the deviation lies in the range between 1.8 % to 4.3 %, and is in this case not significantly dependent on the used particle size. In addition, investigations of the influence of the particle size were carried out. Within the wide range of the used particle fractions, the density and velocity measurement showed some size dependencies. However, the measured differences have only little influence on the accuracy (<.6 %) since in utility plants the grading of coal dust usually changes only in a comparable small range.