Biomedical Instrumentation System

Similar documents
Transduction Based on Changes in the Energy Stored in an Electrical Field

ECE Circuit Theory. Final Examination. December 5, 2008

Solved Problems. Electric Circuits & Components. 1-1 Write the KVL equation for the circuit shown.

Measurement Systems. Biomedical Instrumentation. Biomedical Instrumentation. Lecture 3. Biomedical Instrumentation. Biomedical Instrumentation System

Coverage: through entire course material (updated July 18, 2017) Terminology/miscellaneous. for a linear instrument:

Control Engineering BDA30703

4/3/2019. Advanced Measurement Systems and Sensors. Dr. Ibrahim Al-Naimi. Chapter one. Introduction to Measurement Systems

Step Response for the Transfer Function of a Sensor

Chapter 3. 1 st Order Sine Function Input. General Solution. Ce t. Measurement System Behavior Part 2

2.004 Dynamics and Control II Spring 2008

Problem Weight Total 100

BIOSIGNAL PROCESSING. Hee Chan Kim, Ph.D. Department of Biomedical Engineering College of Medicine Seoul National University

Automatic Control Systems. -Lecture Note 15-

Damped Oscillation Solution

2. The following diagram illustrates that voltage represents what physical dimension?

Department of Mechanical and Aerospace Engineering. MAE334 - Introduction to Instrumentation and Computers. Final Examination.

SAMPLE EXAMINATION PAPER

CMPT 889: Lecture 3 Fundamentals of Digital Audio, Discrete-Time Signals

ENGR 2405 Chapter 8. Second Order Circuits

Module I Module I: traditional test instrumentation and acquisition systems. Prof. Ramat, Stefano

Index. Index. More information. in this web service Cambridge University Press

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam

Biological Systems Modeling & Simulation. Konstantinos P. Michmizos, PhD

Silicon Capacitive Accelerometers. Ulf Meriheinä M.Sc. (Eng.) Business Development Manager VTI TECHNOLOGIES

Chapter 10: Sinusoidal Steady-State Analysis

ELG4112. Electromechanical Systems and Mechatronics

BE 3600 BIOMEDICAL INSTRUMENTATION (LAB) - WEEK 2

Chapter 4 Transients. Chapter 4 Transients

BENG 186B Winter 2012 Final

Chapter 10: Sinusoids and Phasors

Instrument types and performance characteristics

Physics 116A Notes Fall 2004

BME 50500: Image and Signal Processing in Biomedicine. Lecture 5: Correlation and Power-Spectrum CCNY

Name. Section. Short Answer Questions. 1. (20 Pts) 2. (10 Pts) 3. (5 Pts) 4. (10 Pts) 5. (10 Pts) Regular Questions. 6. (25 Pts) 7.

To find the step response of an RC circuit

RLC Circuit (3) We can then write the differential equation for charge on the capacitor. The solution of this differential equation is

Math Assignment 5

Test II Michael R. Gustafson II

C(s) R(s) 1 C(s) C(s) C(s) = s - T. Ts + 1 = 1 s - 1. s + (1 T) Taking the inverse Laplace transform of Equation (5 2), we obtain

Refresher course on Electrical fundamentals (Basics of A.C. Circuits) by B.M.Vyas

Source-Free RC Circuit

Basics of Electric Circuits

12. Introduction and Chapter Objectives

BME/EECS Biomedical Instrumentation & Design Matt O Donnell

Electrical Circuits Lab Series RC Circuit Phasor Diagram

QUESTION BANK DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING UNIT I - INTRODUCTION SYLLABUS

Final Exam: 11 May 2001, Friday

Chapter 7 Vibration Measurement and Applications

Process Control & Instrumentation (CH 3040)

Open Loop Tuning Rules

Delay Coordinate Embedding

Linear systems, small signals, and integrators

WEEKS 8-9 Dynamics of Machinery

Transient Response of a Second-Order System

Chapter 5 Frequency Domain Analysis of Systems

Review of Linear Time-Invariant Network Analysis

Section 4.9; Section 5.6. June 30, Free Mechanical Vibrations/Couple Mass-Spring System

Lecture 11 - AC Power

Electrical Engineering 3BA3: Structure of Biological Materials

BENG 186B Winter 2014 Quiz 3. March 5, NAME (Last, First): This quiz is closed book and closed notes. You may use a calculator for algebra.

RLC Circuits. 1 Introduction. 1.1 Undriven Systems. 1.2 Driven Systems

Sinusoids. Amplitude and Magnitude. Phase and Period. CMPT 889: Lecture 2 Sinusoids, Complex Exponentials, Spectrum Representation

Sensing and Sensors: Fundamental Concepts

MATHEMATICS FOR ENGINEERING TRIGONOMETRY TUTORIAL 3 PERIODIC FUNCTIONS

31545 Medical Imaging systems

COLLEGE OF ENGINEERING DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING QUESTION BANK SUBJECT CODE & NAME : EE1354-MODERN CONTROL SYSTEMS

INSTRUMENTAL ENGINEERING

4.9 Free Mechanical Vibrations

Chapter 2: Unit and Measurement

CMPT 889: Lecture 2 Sinusoids, Complex Exponentials, Spectrum Representation

CITY UNIVERSITY LONDON. MSc in Information Engineering DIGITAL SIGNAL PROCESSING EPM746

Dr Ian R. Manchester Dr Ian R. Manchester AMME 3500 : Review

Introduction to Vibration. Professor Mike Brennan

ME 515 Mechatronics. Overview of Computer based Control System

2. Determine whether the following pair of functions are linearly dependent, or linearly independent:

Introduction and Background

FEEDBACK CONTROL SYSTEMS

Chapter 10 Sinusoidal Steady State Analysis Chapter Objectives:

EE 205 Dr. A. Zidouri. Electric Circuits II. Frequency Selective Circuits (Filters) Low Pass Filter. Lecture #36

JC: You want to see self-organization at work? Test each part of the circuit one by one. You ll see that the whole is more than the sum of its parts

Chapter 5 Frequency Domain Analysis of Systems

Modeling and Experimentation: Mass-Spring-Damper System Dynamics

Manufacturing Equipment Control

EECE 2150 Circuits and Signals Final Exam Fall 2016 Dec 16

EE 4BD4 Lecture 26. Plethysmography

Displacement at very low frequencies produces very low accelerations since:

Lectures on Medical Biophysics Department of Biophysics, Medical Faculty, Masaryk University in Brno. Biocybernetics

Theory and Design for Mechanical Measurements

Principles of Bioelectronics Design Lecture #1

Noise - irrelevant data; variability in a quantity that has no meaning or significance. In most cases this is modeled as a random variable.

Video 5.1 Vijay Kumar and Ani Hsieh

Fall 2011 ME 2305 Network Analysis. Sinusoidal Steady State Analysis of RLC Circuits

Sinusoidal Steady-State Analysis

Cast of Characters. Some Symbols, Functions, and Variables Used in the Book

Microsensors. G.K. Ananthasuresh Professor, Mechanical Engineering Indian Institute of Science Bangalore, , India

Designing Information Devices and Systems II Spring 2016 Anant Sahai and Michel Maharbiz Midterm 2

(A) (B) (D) (C) 1.5. Amplitude (volts) 1.5. Amplitude (volts) Time (seconds) Time (seconds)

Final Exam April 30, 2013

A Proposed Warped Choi Williams Time Frequency Distribution Applied to Doppler Blood Flow Measurement

Transcription:

BME/EECS 458 - Biomedical Instrumentation & Design Matt O Donnell I.0 Introduction What is a biomedical instrument? To many it s an EKG machine, to others it s a chemical biosensor, and to some it s a medical imaging system. There are a lot of instruments that qualify as biomedical instruments. Current estimates of the worldwide market in biomedical instruments is over $200 billion. Even though there is a wide variety of instruments, almost all of them can be modeled using the simple diagram below. Biomedical Instrumentation System S E N S E A C T U A T E Electronics Interface Computation Tissue Interface All biomedical instruments must interface with biological materials. That interface can by direct contact or by indirect contact (e.g., induced fields). A sensor must: detect biochemical, bioelectrical, or biophysical parameters reproduce the physiologic time response of these parameters provide a safe interface with biological materials BME/EECS 458 - Chapter 1 1

An actuator must: deliver external agents via direct or indirect contact control biochemical, bioelectrical, or biophysical parameters provide a safe interface with biologic materials The electronics interface must: match electrical characteristics of the sensor/actuator with computation unit preserve signal to noise ratio of sensor preserve efficiency of actuator preserve bandwih (i.e., time response) of sensor/actuator provide a safe interface with the sensor/actuator provide a safe interface with the computation unit provide secondary signal processing functions for the system The computation unit must: provide primary user interface provide primary control for the overall system provide data storage for the system provide primary signal processing functions for the system maintain safe operation of the overall system In this course we will study only sensing systems - this means the system front-end will always be a sensing element. Other than this restriction, we will cover all aspects of typical biomedical instrumentation systems. We will do them in the following order: A. Basic Sensors and Principles - including biopotential electrodes B. Electronic Interfacing - including system noise figure, system bandwih, preamps, post-amps, instrumentation amps, A/D and D/A converters, aliasing, triggering and signal averaging C. Computation - including data capture and signal processing D. Systems - complete system response using specific examples (electromyogram, pressure sensors and blood pressure measurements, flow sensors and blood flow measurements, and chemical biosensors) BME/EECS 458 - Chapter 1 2

I.1 Types of Biomedical Instruments Types of Biomedical Instrumentation Systems Direct / Indirect Invasive / Noninvasive Contact / Remote Sense / Actuate Real-time / Static Direct/Indirect - The sensing system measure a physiologic parameter directly, such as the average volume blood flow in an artery, or measures a parameter related to the physiologic parameter of interest (e.g., EKG recording at the body surface is related to propagation of the action potential in the heart but is not a measurement of the propagation waveform). Invasive/Noninvasive - Direct electrical recording of the action potential in nerve fibers using an implantable electrode system is an example of an invasive sensor. An imaging system measuring blood flow dynamics in an artery (e.g., ultrasound color flow imaging of the carotid artery) is an example of a noninvasive sensor. Contact/Remote - A strain gauge sensor attached to a muscle fiber can record deformations and forces in the muscle. An MRI or ultrasound imaging system can measure internal deformations and forces without contacting the tissue. Sense/Actuate - A sensor detects biochemical, bioelectrical, or biophysical parameters. An actuator delivers external agents via direct or indirect contact and/or controls biochemical, bioelectrical, or biophysical parameters. An automated insulin delivery pump is an example of a direct, contact actuator. Noninvasive surgery with high intensity, focused ultrasound (HIFU) is an example of a remote, noninvasive actuator. Real-time/Static - Static instruments measure temporal averages of physiologic parameters. Real-time instruments have a time response faster than or equal to the physiologic time constants of the sensed parameter. For example a real-time, ultrasound Doppler system can measure changes in arterial blood velocity over a cardiac cycle. SEE CHAPTER 1 FIGURES FOR SYSTEM EXAMPLES BME/EECS 458 - Chapter 1 3

I.2 General Instrument Performance Parameters To compare different instruments, quantitative criteria are needed. Performance parameters are usually subdivided into two classes based on the frequency of the input signal. Static characteristics describe the performance for dc or very low frequency inputs. The properties of the output for a wide range of constant inputs demonstrate the quality of the measurement. Dynamic characteristics require a full differential equation description of system performance - more on that in the next section. Complete characteristics are usually approximated by the sum of static and dynamic characteristics. Some common static performance parameters are listed below. Accuracy The accuracy of a single measured quantity is the difference between the true value and the measured value divided by the true value : True - Measured Accuracy = True It is often quoted as ± percentage. Often the true value is unknown over all operating conditions, so the true value is approximated with some standard. Precision The precision of a measurement expresses the number of distinguishable alternatives from which a given result is selected. On most modern instrumentation systems the precision is ultimately determined by the analog to digital converter (A/D) characteristics. Resolution The smallest quantity that can be measured with certainty is the resolution. Resolution expresses the degree to which nearly equal values of a quantity can be discriminated. Reproducibility The ability of an instrument to give the same output for equal inputs applied over some period of time is called reproducibility. Drift is the primary limit on reproducibility - see below. Sensitivity Sensitivity describes changes in system output for a given change in a single input. It is quantified by holding all inputs constant except one. This one input is varied BME/EECS 458 - Chapter 1 4

incrementally over the normal operating range, producing a range of outputs needed to compute the sensitivity. Zero (Offset) Drift Offset drift is one parameter determining reproducibility. It is measured by monitoring the system output with no change in input. Any changes that occur are simply result of system offset. Sensitivity Drift Sensitivity drift is the second primary contributor to irreproducibility. It causes error proportional to the magnitude of the input. These drift parameters are summarized in a typical sensor sensitivity curve below. BME/EECS 458 - Chapter 1 5

Linearity A linear system satisfies the condition: If x 1 y 1 x 2 y 2 then the system is linear if and only if: ax 1 + bx 2 ay 1 + by 2 This is the simple expression of the superposition principle for a linear system. There are many ways to express deviations from linearity for a practical system. For dynamic systems, multitone tests are often used, where the magnitude of beat frequencies between the individual tone frequencies can quantify the level of nonlinearity. For static systems, independent nonlinearity measures as shown below are often used. BME/EECS 458 - Chapter 1 6

Dynamic Range The dynamic range defines the ratio between the maximum undistorted signal (i.e., maximum input signal satisfying the linearity specification for the sensor) and the minimum detectable signal for a given set of operating conditions. Often the dynamic range is quoted on a logarithmic scale (i.e., db scale). Input Impedance The instantaneous rate at which energy is transferred by a system (i.e., the power) is simply proportional to the product of an effort variable (e.g., voltage, pressure, force) with a flow variable (current, flow, velocity). The generalized impedance, Z, is the ratio of the phasor equivalent of the steady-state sinusoidal effort variable to the phasor equivalent of the steady-state flow variable: ~ ~ Z = V ~ I where the tilde denotes phasor variables (i.e., magnitude and phase). The phase is related to the response lag of the system to a sinusoidal input - more about this for dynamic systems. I.3 General Dynamic Performance Parameters Most biomedical instruments must process signals that change with time. The dynamics of the measurement system, therefore, must be chosen to properly reproduce the dynamics of the physiologic variables the system is sensing. In this course we will only consider linear, time invariant systems unless otherwise explicitly noted. For such systems, the dynamics can be fully described by simple differential equations of the form: a n n d y(t) n +... + a dy(t) + a y(t) = b m d y(t) +... + b dx(t) + b x(t) 1 0 m m 1 0 where x(t) is the input signal (usually the physiologic parameter of interest), y(t) is the output signal (usually the electronic signal), and the a s and b s are constants determined by the physical characteristics of the sensor system. Most practical sensor front-ends are described by differential equations of zero, first or second order (i.e., n=0,1,2), and derivatives of the input are usually absent, so m=0. Linear, time invariant systems are simply characterized by their response to sinusoidal inputs of the form x(t) = A sin(ωt), where the output is a sinusoid at precisely the same frequency of the form y(t) = B(ω) sin(ωt + φ(ω)). This simple characteristics is captured in the system transfer function, defined as function of angular frequency ω=2πf : BME/EECS 458 - Chapter 1 7

~ ~ ω H ( ω ) = Y( ~ ) X( ω) = b ( jω) +... + b ( jω) + b m m a ( jω) +... + a ( jω) + a n n 1 0 1 0 where j = (-1) 1/2 and H(ω) is written in complex notation, where the magnitude of H(ω) equals the ratio B(ω)/A and the phase of H(ω) represents the physical phase lag φ(ω). Using the transfer function notation, the dynamic response of simple zero, first, and second order systems are presented below. Zero Order System A linear potentiometer can be used to measure displacement and represents a simple example of a zero order system. The differential equation describing its operation is: and the transfer function is: a y(t) = b x(t) 0 0 ~ ~ ω H ( ω ) = Y( ~ ) X( ω) b0 = = K a 0 Note there is no phase lag between output and input at ALL frequencies. This means the step response is instantaneous, as illustrated below. BME/EECS 458 - Chapter 1 8

First Order System If the instrument contains a single energy storage element, the a first order differential equation describes system dynamics dy(t) a 1 + a 0 y(t) = b 0 x(t) with associate transfer function: ~ ~ ω K H ( ω ) = Y( ~ ) = X( ω) 1+ jωτ where K=b 0 /a 0 and τ=a 1 /a 0. The RC low pass filter shown below is an example of a first order system. Note the phase lag is a function of frequency and creates a delayed step response. The system does not pass frequencies much greater than ω=1/τ. Consequently, for a first order sensor system there should be no significant frequency components in the physiologic input parameter greater than this cutoff frequency. BME/EECS 458 - Chapter 1 9

Second Order System A second order system has two levels of energy storage with dynamics described by the differential equation 2 d y(t) a 2 + a dy(t) 2 1 0 0 + a y(t) = b x(t) with associate transfer function: ~ ~ ω K H ( ω ) = Y( ~ ) = X( ω) ( j ω/ ω ) + ( 2ζj ω/ ω ) + 1 where the static sensitivity is K=b 0 /a 0, the undamped natural frequency is ω n = (a 0 /a 2 ) 1/2, and the dimensionless damping ratio is ζ=a 1 /(2(a o a 2 ) 1/2 ). A mechanical force-measuring system shown below illustrates the properties of a second order system. Note the step response for underdamped, critically damped, and overdamped cases. Again, significant components in the input variable must be at frequencies less than natural frequency of the second order system. In later sections we will see how the pre-amp, post-amp, digitization system and digital signal processing system must be matched to the transfer characteristics of the sensor element. n 2 n BME/EECS 458 - Chapter 1 10