Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Similar documents
Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Nitrogen, Total Inorganic

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

Oxygen Demand, Chemical

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Oxygen Demand, Chemical

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Chloramine (Mono) and Nitrogen, Free Ammonia

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

Method to 0.50 mg/l NH 3 N Powder Pillows

Chlorine, Free and Total, High Range

Method to 0.50 mg/l NH 3 N Powder Pillows

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 )

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

Chlorine, Free and Total, High Range

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Chlorine, Free and Total, Low Range

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

DR/4000 PROCEDURE NITRATE. 2. The display will show:

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

DR/4000 PROCEDURE. IRON, Total

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Trihalomethane Formation Potential (THMFP)

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Quaternary Ammonium Compounds

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

CHLORINE, TOTAL (0 to 4.00 mg/l)

NITROGEN, AMMONIA, High Range, Test N Tube

DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample.

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

TPH (Total Petroleum Hydrocarbons)

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure

NITROGEN, TOTAL, HR, Test N Tube

Mercaptoacetic Acid Method*

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

DR/4000 PROCEDURE. HARDNESS, Total

NITROGEN, TOTAL, Test N Tube * Method 10071

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

SILICA, Low Range (0 to 1.60 mg/l)

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

Polychlorinated Biphenyls (PCB) in Soil

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

DPD Test N Tube Method *

Manganese III Digestion Method * (with chloride removal)

NITRITE, Low Range (0 to mg/l NO 2- -N)

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced.

DR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys.

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Hach Company TNTplus TM Phosphorus Spectrophotometric Measurement of Phosphorus in Water and Wastewater

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

ALLOWAY METHOD OUTLINE

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION

p-dimethylaminobenzaldehyde Method*

METHOD DETERMINATION OF PHOSPHORUS BY SEMI-AUTOMATED COLORIMETRY

SIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER COLORIMETRIC DETERMINATION OF CYANIDE

Steam-Assisted Gravity Drainage Water Analysis Handbook

SC400 Colorimeter. Silica. Photometer-System. Instruction Manual. SiO 2. Bedienungsanleitung Seite Instruction Manual Page 19 33

Steam-Assisted Gravity Drainage Water Analysis Handook

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

SILICA, Low Range SiLR

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

Standard Operating Procedure for: Total Phosphorus (3010R02 Total P.doc) Missouri State University. and

Phosphate total HR TT mg/l P b) Phosphomolybdenum Blue

O Sodium Bicarbonate, NaHCO 3 Ammonium Molybdate Tetrahydrate, (NH 4. O 4 Sodium Chloride, NaCl Ascorbic Acid, C 6 H 8 O 6

Phosphate total LR TT mg/l P b) Phosphomolybdenum Blue

PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION

C 214 Multiparameter Bench Photometer

MERCURY (0.1 to 2.5 µg/l) Method 10065

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

#18 Analysis of Water (Chlorine, Phosphorus, Fluoride, and Color)

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2

Determination of Orthophosphate Ion

Phosphate HR T 1-80 mg/l P Vanadomolybdate

Determination of Orthophosphate Ion

Transcription:

Phosphorus, Total DOC316.53.01121 USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method 8190 0.06 to 3.50 mg/l PO 3 4 (0.02 to 1.10 mg/l P) Test N Tube Vials Scope and application: For water, wastewater and seawater. 1 USEPA accepted for reporting wastewater analyses (Standard Methods 4500-P E). Test preparation Instrument-specific information Table 1 shows all of the instruments that have the program for this test. The table also shows adapter and light shield requirements for the instruments that use them. To use the table, select an instrument, then read across to find the applicable information for this test. Table 1 Instrument-specific information for test tubes Instrument Adapters Light shield DR 6000, DR 5000 DR 3900 LZV849 DR 3800, DR 2800, DR 2700 LZV646 DR 1900 9609900 (D 1 ) DR 900 4846400 Cover supplied with the instrument 1 The D adapter is not available with all instrument versions. Before starting Install the instrument cap on the DR 900 cell holder before ZERO or READ is pushed. DR 3900, DR 3800, DR 2800 and DR 2700: Install the light shield in Cell Compartment #2 before this test is started. For the best results, measure the reagent blank value for each new lot of reagent. Replace the sample with deionized water in the test procedure to determine the reagent blank value. Subtract the reagent blank value from the sample results automatically with the reagent blank adjust option. The test range for total phosphate is limited to 0.06 to 3.5 mg/l PO 4 3. Test results that are more than 3.5 mg/l can be used to estimate dilution ratios, but should NOT be used for reporting purposes. If the test result is more than 3.5 mg/l, dilute the sample and repeat the digestion and the colorimetric test. Clean all glassware with 6.0 N (1:1) hydrochloric acid, then fully rinse with deionized water to remove contaminants. The reagent that is used in this test is corrosive. Use protection for eyes and skin and be prepared to flush any spills with running water. Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective equipment. Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory agencies for further disposal information. Items to collect Description Quantity Total Phosphorus Test 'N Tube Reagent Set 1 DRB200 Reactor 1 1

Items to collect (continued) Description Quantity Funnel, micro 1 Light shield or adapter (For information about sample cells, adapters or light shields, refer to Instrument-specific information on page 1.) 1 Pipet, TenSette, 1.0- to 10.0-mL, with pipet tips 1 Test tube rack 1 Water, deionized varies Refer to Consumables and replacement items on page 5 for order information. Sample collection and storage Collect samples in clean glass or plastic bottles that have been cleaned with 1:1 hydrochloric acid and rinsed with deionized water. Analyze the samples as soon as possible for best results. Do not use a detergent that contains phosphate to clean the sample bottles. The phosphate in the detergent will contaminate the sample. To preserve samples for later analysis, adjust the sample ph to less than 2 with concentrated sulfuric acid (about 2 ml per liter). No acid addition is necessary if the sample is tested immediately. Keep the preserved samples at or below 6 C (43 F) for a maximum of 28 days. Let the sample temperature increase to room temperature before analysis. Before analysis, adjust the ph to 7 with 5.0 N sodium hydroxide standard solution. Correct the test result for the dilution caused by the volume additions. Acid persulfate digestion for Test 'N Tubes Start 1. Start the DRB200 Reactor. Preheat to 150 C. Refer to the DRB200 manual. 2. Start program 536 P Total/AH PV TNT. For information about sample cells, adapters or light shields, refer to Instrumentspecific information on page 1. 3. Add 5.0 ml of sample to the Total Phosphorus Test Vial. 4. Add the contents of one Potassium Persulfate Powder Pillow for Phosphonate to the vial. Note: Although the program name can be different between instruments, the program number does not change. 2 Phosphorus, Total, PhosVer 3 TNT Method (3.50 mg/l)

5. Put the cap on the vial. Shake to dissolve the powder. 6. Insert the vial into the reactor. Close the reactor. 7. Start the instrument timer. A 30-minute reaction time starts. 8. When the timer expires, carefully remove the vial from the reactor. Set the vial in a test tube rack. Let the vial cool to room temperature. 9. Add 2 ml of 1.54 N Sodium Hydroxide Standard Solution to the vial. 10. Put the cap on the vial. Invert to mix. 11. Clean the vial. 12. Insert the vial into the 16-mm cell holder. Zero 13. Push ZERO. The display shows 0.00 mg/l PO 4 3. 14. Add the contents of one PhosVer 3 Powder Pillow to the vial. 15. Put the cap on the vial. Shake to mix for 20 30 seconds. The powder will not dissolve completely. 16. Start the instrument timer. A 2-minute reaction time starts. Measure the sample within two to eight minutes after the timer expires. Phosphorus, Total, PhosVer 3 TNT Method (3.50 mg/l) 3

Read 17. Clean the vial. 18. Insert the vial into the 16-mm cell holder. 19. Push READ. Results show in mg/l PO 4 3. Interferences Interfering substance Aluminum Arsenate Chromium Copper Sulfide Iron Nickel Highly buffered samples or extreme sample ph Silica Silicate Turbidity or color Zinc Interference level More than 200 mg/l Interferes at any level More than 100 mg/l More than 10 mg/l More than 90 mg/l More than 100 mg/l More than 300 mg/l Can prevent the correct ph adjustment of the sample by the reagents. Sample pretreatment may be necessary. More than 50 mg/l More than 10 mg/l Samples with a high amount of turbidity can give inconsistent results. The acid in the reagents can dissolve some of the suspended particles and variable desorption of orthophosphate from the particles can occur. More than 80 mg/l Accuracy check Standard additions method (sample spike) Use the standard additions method (for applicable instruments) to validate the test procedure, reagents and instrument and to find if there is an interference in the sample. Items to collect: Phosphate 10-mL Ampule Standard, 50-mg/L as PO 4 3 Ampule breaker Pipet, TenSette, 0.1 1.0 ml and tips Mixing cylinders, 25-mL (3) 1. Use the test procedure to measure the concentration of the sample, then keep the (unspiked) sample in the instrument. 2. Go to the Standard Additions option in the instrument menu. 3. Select the values for standard concentration, sample volume and spike volumes. 4. Open the standard solution. 5. Prepare three spiked samples: use the TenSette pipet to add 0.1 ml, 0.2 ml and 0.3 ml of the standard solution, respectively, to three 25-mL portions of fresh sample. Mix well. 4 Phosphorus, Total, PhosVer 3 TNT Method (3.50 mg/l)

6. Use the test procedure to measure the concentration of each of the spiked samples. Start with the smallest sample spike. Measure each of the spiked samples in the instrument. 7. Select Graph to compare the expected results to the actual results. Note: If the actual results are significantly different from the expected results, make sure that the sample volumes and sample spikes are measured accurately. The sample volumes and sample spikes that are used should agree with the selections in the standard additions menu. If the results are not within acceptable limits, the sample may contain an interference. Standard solution method Use the standard solution method to validate the test procedure, the reagents and the instrument. Items to collect: 3.0-mg/L phosphate standard solution 1. Use the test procedure to measure the concentration of the standard solution. 2. Compare the expected result to the actual result. Note: The factory calibration can be adjusted slightly with the standard adjust option so that the instrument shows the expected value of the standard solution. The adjusted calibration is then used for all test results. This adjustment can increase the test accuracy when there are slight variations in the reagents or instruments. Method performance The method performance data that follows was derived from laboratory tests that were measured on a spectrophotometer during ideal test conditions. Users can get different results under different test conditions. Program Standard Precision (95% confidence interval) Sensitivity Concentration change per 0.010 Abs change 536 3.00 mg/l PO 4 3 2.93 3.07 mg/l PO 4 3 0.06 mg/l PO 4 3 Summary of method Phosphates present in organic and condensed inorganic forms (meta-, pyro- or other polyphosphates) must be converted to reactive orthophosphate before analysis. Pretreatment of the sample with acid and heat provides the conditions for hydrolysis of the condensed inorganic forms. Organic phosphates are converted to orthophosphates by heating with acid and persulfate. Orthophosphate reacts with molybdate in an acid medium to produce a mixed phosphate/molybdate complex. Ascorbic acid then reduces the complex, giving an intense molybdenum blue color. The measurement wavelength is 880 nm (DR 1900: 710 nm) for spectrophotometers or 610 nm for colorimeters. Pollution prevention and waste management Reacted samples contain molybdenum and must be disposed of as a hazardous waste. Dispose of reacted solutions according to local, state and federal regulations. Consumables and replacement items Required reagents Description Quantity/test Unit Item no. Water, deionized varies 100 ml 27242 Total Phosphorus Test N Tube Reagent Set 1 50 tests 2742645 Includes: PhosVer 3 Phosphate Reagent Powder Pillow, 10-mL 1 50/pkg 2106046 Potassium Persulfate Powder Pillow 1 pillow 50/pkg 2084766 Phosphorus, Total, PhosVer 3 TNT Method (3.50 mg/l) 5

Consumables and replacement items (continued) Description Quantity/test Unit Item no. Sodium Hydroxide, 1.54 N varies 100 ml 2743042 Total and Acid Hydrolyzable Test Vials (not sold separately) 1 50/pkg Required apparatus Description Quantity/test Unit Item no. DRB 200 Reactor, 110 VAC option, 15 x 16-mm wells 1 each LTV082.53.40001 DRB 200 Reactor, 220 VAC option, 15 x 16-mm wells 1 each LTV082.52.40001 Funnel, micro, poly 1 each 2584335 Light shield, DR 3900 1 each LZV849 Light shield, DR 3800, DR 2800, DR 2700 1 each LZV646 Pipet, TenSette, 1.0 10.0 ml 1 each 1970010 Pipet Tips, for TenSette Pipet, 1.0 10.0 ml 2 250/pkg 2199725 Test tube rack 1 each 1864100 Recommended standards and apparatus Description Unit Item no. Drinking Water Standard, Mixed Parameter, Inorganic for F -, NO 3 N, PO 3 4, SO 2 4 500 ml 2833049 Phosphate Standard Solution, 50-mg/L, 10-mL Voluette Ampules 16/pkg 17110 Phosphate Standard Solution, 1-mg/L as PO 3 4 500 ml 256949 Phosphate Standard Solution, 3-mg/L as PO 3 4 946 ml 2059716 Wastewater Effluent Standard Solution, Mixed Parameter, for NH 3 -N, NO 3 -N, PO 4 3, COD, SO 4 2, TOC 500 ml 2833249 Ampule Breaker, 10-mL Voluette Ampules each 2196800 Optional reagents and apparatus Description Unit Item no. Mixing cylinder, graduated, 25-mL each 189640 Pipet, volumetric, Class A, 2-mL each 1451536 Hydrochloric Acid Solution, 6.0 N (1:1) 500 ml 88449 Sodium Hydroxide Standard Solution, 5.0 N 1 L 245053 Sulfuric Acid, concentrated, ACS 500 ml 97949 Pipet, TenSette, 0.1 1.0 ml each 1970001 Pipet tips for TenSette Pipet, 0.1 1.0 ml 50/pkg 2185696 Pipet tips for TenSette Pipet, 0.1 1.0 ml 1000/pkg 2185628 Bottle, sampling, with cap, low density polyethylene, 250-mL 12/pkg 2087076 Paper, ph, 0 14 ph range 100/pkg 2601300 Water, deionized 4 L 27256 Thermometer, non-mercury, 10 to +225 C each 2635700 Finger cots 2/pkg 1464702 6 Phosphorus, Total, PhosVer 3 TNT Method (3.50 mg/l)

Optional standards Description Unit Item no. Phosphate Standard Solution, 10-mg/L as PO 3 4 946 ml 1420416 Phosphate Standard Solution, 15-mg/L as PO 3 4 100 ml 1424342 Phosphate Standard Solution, 100-mg/L as PO 4 100 ml 1436832 Phosphate Standard Solution, 10-mL ampule, 500 mg/l as PO 4 16/pkg 1424210 Phosphate Standard Solution, 500-mg/L as PO 4 100 ml 1424232 Phosphorus, Total, PhosVer 3 TNT Method (3.50 mg/l) 7

FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company/Hach Lange GmbH, 1989 2014. All rights reserved. 01/2014, Edition 9