Atomic Force Microscopy

Similar documents
Lecture 4 Scanning Probe Microscopy (SPM)

Introduction to Scanning Probe Microscopy

General concept and defining characteristics of AFM. Dina Kudasheva Advisor: Prof. Mary K. Cowman

MRSEC. Refrigerator Magnet Activity Guide. Quick Reference Activity Guide. a) b) c) = north = south. Activity Materials

Basic Laboratory. Materials Science and Engineering. Atomic Force Microscopy (AFM)

Measuring Newton's Constant of Universal Gravitation using a Gravitational Torsion Balance

Chapter 10. Nanometrology. Oxford University Press All rights reserved.

MS482 Materials Characterization ( 재료분석 ) Lecture Note 11: Scanning Probe Microscopy. Byungha Shin Dept. of MSE, KAIST

Nanoscale work function measurements by Scanning Tunneling Spectroscopy

Scanning Tunneling Microscopy

NIS: what can it be used for?

Techniken der Oberflächenphysik (Techniques of Surface Physics)

Module 26: Atomic Force Microscopy. Lecture 40: Atomic Force Microscopy 3: Additional Modes of AFM

Scanning Probe Microscopy. Amanda MacMillan, Emmy Gebremichael, & John Shamblin Chem 243: Instrumental Analysis Dr. Robert Corn March 10, 2010

SOLID STATE PHYSICS PHY F341. Dr. Manjuladevi.V Associate Professor Department of Physics BITS Pilani

Contents. What is AFM? History Basic principles and devices Operating modes Application areas Advantages and disadvantages

Atomic and molecular interactions. Scanning probe microscopy.

And Manipulation by Scanning Probe Microscope

672 Advanced Solid State Physics. Scanning Tunneling Microscopy

Program Operacyjny Kapitał Ludzki SCANNING PROBE TECHNIQUES - INTRODUCTION

Introduction to the Scanning Tunneling Microscope

MSE 321 Structural Characterization

MSE 321 Structural Characterization

Introduction to Scanning Probe Microscopy Zhe Fei

Scanning tunneling microscopy

Nanotechnology is an ever expanding field with improved growth every year.

Imaging Methods: Scanning Force Microscopy (SFM / AFM)

From nanophysics research labs to cell phones. Dr. András Halbritter Department of Physics associate professor

Instrumentation and Operation

Characterization of MEMS Devices

tip of a current tip and the sample. Components: 3. Coarse sample-to-tip isolation system, and

Citation for published version (APA): Mendoza, S. M. (2007). Exploiting molecular machines on surfaces s.n.

Nanoscale characteristics by Scanning Tunneling Spectroscopy

Atomic Force Microscopy (AFM) Part I

nano-ta: Nano Thermal Analysis

A SCIENTIFIC APPROACH TO A STICKY PROBLEM

Experiment C-10 Titration of a Strong Acid and a Strong Base

EXPLORING SCANNING PROBE MICROSCOPY WITH MATHEMATICA

STM: Scanning Tunneling Microscope

Scanning Probe Microscopy. EMSE-515 F. Ernst

ELEC 1908 The Electric Potential (V) March 28, 2013

CHARACTERIZATION of NANOMATERIALS KHP

Bridge between research in modern physics and entrepreneurship in nanotechnology. Quantum Physics

Vacuum Kelvin Force Probe Research Richard Williams August 1st 2008

LAB 6 - GRAVITATIONAL AND PASSIVE FORCES

2: SIMPLE HARMONIC MOTION

Probing Molecular Electronics with Scanning Probe Microscopy

Electrical and Magnetic Properties of High Temperature Superconductors Using Varying forms of Data Acquisition

Nanotechnology Fabrication Methods.

Force and Motion 20 N. Force: Net Force on 2 kg mass = N. Net Force on 3 kg mass = = N. Motion: Mass Accel. of 2 kg mass = = kg m/s 2.

AFM for Measuring Surface Topography and Forces

Scanning Force Microscopy

Supporting Information

Part I. Two Force-ometers : The Spring Scale and The Force Probe

Simple Harmonic Motion and Damping

Nanoscale Surface Physics PHY 5XXX

Scanning Tunneling Microscopy Transmission Electron Microscopy

Physics Lab 202P-4. Understanding Electric Potential NAME: LAB PARTNERS:

Measurement of Bone Strength and Stiffness using 3-Point Bending

Evaluation copy. Ground Temperature Measurements. computer OBJECTIVES MATERIALS PRE-LAB QUESTIONS

What is the velocity profile downstream of the sudden expansion? What is the relationship between the velocity profile and the flow rate?

Static Equilibrium. Torque - also known as moment of force. (Serway Sec. 11.1) Rigid objects in static equilibrium. (Serway Secs. 12.1, 12.

MECH 466. Micro Electromechanical Systems. Laboratory Manual Laboratory #3: Stiction of MEMS and Strength of MEMS Materials

Scanning Probe Microscopy (SPM)

Characterization Tools

MATERIAL SCIENCE AND TECHONOLOGY-1. Scanning Tunneling Microscope, STM Tunneling Electron Microscope, TEM HATİCE DOĞRUOĞLU

Application of electrostatic force microscopy in nanosystem diagnostics

Chapter 2 Correlation Force Spectroscopy

Nitride HFETs applications: Conductance DLTS

Coefficient of Friction Lab

Scanning Probe Microscopy (SPM)

Ecole Franco-Roumaine : Magnétisme des systèmes nanoscopiques et structures hybrides - Brasov, Modern Analytical Microscopic Tools

Software Verification

Hooke s Law. Equipment. Introduction and Theory

PHY222 Lab 2 - Electric Fields Mapping the Potential Curves and Field Lines of an Electric Dipole

Synthesis, Characterization and Tribological Evaluation of SDS stabilized. Antiwear Lubricant Additives

Understanding the properties and behavior of groups of interacting atoms more than simple molecules

Laboratory Exercise X Most Dangerous Places to Live in the United States Based on Natural Disasters

MSE640: Advances in Investigation of Intermolecular & Surface Forces

A NEW APPROACH TOWARDS PROPERTY NANOMEASUREMENTS USING IN-SITU TEM

Simple Harmonic Motion

Nanotechnology. Gavin Lawes Department of Physics and Astronomy

Lecture 26 MNS 102: Techniques for Materials and Nano Sciences

Magnetic Fields. Experiment 1. Magnetic Field of a Straight Current-Carrying Conductor

DETERMINATION OF THE ADHESION PROPERTIES OF MICA VIA ATOMIC FORCE SPECTROSCOPY

bio-molecular studies Physical methods in Semmelweis University Osváth Szabolcs

Observing Specialized Cells

Information on the test material EDS-TM002 and the BAM software package EDX Spectrometer Test for determination of the spectrometer performance

Measuring Properties of Elastomers using AFM Force Curves

Paper Session I-A - Development of a Microscopy Platform for the International Space Station

Lab 1: Electric Interactions I

Atomic Force Microscopy imaging and beyond

Harmonic Motion. Mass on a Spring. Physics 231: General Physics I Lab 6 Mar. 11, Goals:

VEDA - Virtual Environment for Dynamic Atomic Force Microscopy

Math 261 Sampling Distributions Lab Spring 2009

Physics 1050 Experiment 3. Force and Acceleration

Surface Studies by Scanning Tunneling Microscopy

1986 s Nobel Prize in Physics

SCANNING-PROBE TECHNIQUES OR APPARATUS; APPLICATIONS OF SCANNING-PROBE TECHNIQUES, e.g. SCANNING PROBE MICROSCOPY [SPM]

Control of Dynamics of SPM Probes for Non-destructive Defectoscopy

Transcription:

Journal of the Advanced Undergraduate Physics Laboratory Investigation Volume 0 Issue 0 Premiere Issue Article 2 6-6-2012 Atomic Force Microscopy Tyler Lane Minnesota State University - Moorhead, lanety@mnstate.edu Follow this and additional works at: http://opus.ipfw.edu/jaupli Opus Citation Lane, Tyler (2012) "Atomic Force Microscopy," Journal of the Advanced Undergraduate Physics Laboratory Investigation: Vol. 0: Iss. 0, Article 2. Available at: http://opus.ipfw.edu/jaupli/vol0/iss0/2 This Article is brought to you for free and open access by Opus: Research & Creativity at IPFW. It has been accepted for inclusion in Journal of the Advanced Undergraduate Physics Laboratory Investigation by an authorized administrator of Opus: Research & Creativity at IPFW. For more information, please contact admin@lib.ipfw.edu.

Lane: Atomic Force Microscopy Atomic Force Microscopy Tyler Lane, Garrett Marcus I. INTRODUCTION AND BACKGROUND Scanning tunneling microscopy (STM) and atomic force microscopy (AFM) are powerful tools for the examination of surfaces. The research, development, and application of the STM and AFM methods are currently making rapid progress [1]. In 1986, Gerd Binnig and Heinrich Rohrer won the Nobel Prize in Physics for the invention of the scanning tunneling microscope (STM) and the fact that it could achieve atomic resolution [2]. Later the AFM was invented and became very promising because it can be used to image a huge variety of samples which do not need to be electrical conductors. The AFM detects the force interaction between a sample and a very tiny tip mounted on a cantilever. The force interaction between the sample and tip is related to the deflection of the cantilever, so the more the tip presses into the sample the greater the deflection of the cantilever and the greater the force exercised on the sample. II. EXPERIMENTAL METHODS The samples that will be viewed are a chip structure in silicon a.k.a. RAM, a CD stamper gold clusters, and nanotubes. All of these samples are in an extended kit that came with the Nanosurf software. The AFM uses atomic forces to detect the surface of the sample. The AFM detects the force interaction between a sample and a very tiny tip mounted on a cantilever. Atomic forces attract the tip which causes the cantilever to bend, as illustrated in the setup diagram. The force interaction between the sample and tip is related to the deflection of the cantilever. A feedback system keeps the deflection constant as the tip scans over the sample surface. Published by Opus: Research & Creativity at IPFW, 2012 1

Journal of the Advanced Undergraduate Physics Laboratory Investigation, Vol. 0 [2012], Iss. 0, Art. 2 Setup of AFM experiment: Start up the easyscan 2 or a similar program that is used with the AFM. Be sure to load proper parameters otherwise the images produced will not come out clear. Be sure to ground the AFM sample mount if that option is available. Wear gloves when necessary. Also make sure the AFM is level before and during the duration of the experiment. Place the sample on to the sample mount, and carefully slide it under the AFM. Positioning the tip above the sample is the next step. Theree are two different modes the AFM can operate in: static mode and dynamic mode. Static mode is when the probe tip is always in contact with the sample and the deflection of the cantilever is set by the user. Dynamic mode is when the probe tip is Figure 1 oscillated and makes intermittent contact with the surface; in some cases, not touching the surface at all. In this particular experiment, dynamic mode was used Figure 2 http://opus.ipfw.edu/jaupli/vol0/iss0/2 2

Lane: Atomic Force Microscopy so as to not break the tip off the cantilever. Next, use the software to prepare the tip to approach the sample. After that, get as close to the sample as you can manually using the side view on the AFM. See figure 1 and figure 2. In figure 1, looking through the side view lens, you can estimate the position of the sample surface. It is half way between the cantilever and its reflection. Advance the tip until the cantilever is close to, but not touching the surface. See figure 2. If the tip touches the surface, the tip might break, and then it must be replaced. This process may vary depending on what AFM and mode is used. After the tip is in position, click the approach button. This will automatically lower the tip closer to the sample at a safe distance to begin the imaging process. Once the image is complete, stop the AFM and click retract to raise the tip high enough to remove and view a different sample. Again, this may be different for different software programs that work with an AFM. Save the image as a.nid file, otherwise it will not be saved so measurements can be taken. Repeat the above steps for each sample for the imaging process. If any problems arise, read the manuals of the AFM and of the easyscan 2 or the software used in this experiment. III. RESULTS The results are as follows: RAM CD Gold Clusters The measurements were made using the built in features of the Nanosurf software. Published by Opus: Research & Creativity at IPFW, 2012 3

Journal of the Advanced Undergraduate Physics Laboratory Investigation, Vol. 0 [2012], Iss. 0, Art. 2 The RAM had two parts measured: the distance between the dots across from it and the distance to the next set of dots. These are labeled a and b on the previous page. The CD also had two measurements: the first was the length of a short bump and the second was the length of a long bump. Again, these are labeled a and b on the previous page. For the gold clusters, the Root-Mean-Square (RMS) roughness, S q, was measured over different areas of the sample image. There were no nanotubes found on the sample they were supposed to be on. Hence no images of the nanotubes are shown in this report. Table 1 shows the measurements and final values for each sample. The final values are as follows: Sample RAM length a RAM length b CD length a CD length b Gold clusters Nanotubes Final Experimental Value 12.5±0.2µm 7.1±0.2µm 3.5±0.2µm 2.5±0.3µm 5.1±0.8nm N/A Finally these results were compared to the different companies on page 36 of the lab manual to determine which company made these samples. IV. CONCLUSION The results strongly support that the samples were made by Company #4. The following table shows the experimental results vs. the results presented by Company #4: Sample Final Experimental Company #4 Value RAM length a 12.5±0.2µm 10.0±0.3 µm RAM length b 7.1±0.2µm 7.2±0.7µm CD length a 3.5±0.2µm 3.5±0.3µm CD length b 2.5±0.3µm 2.5±0.7µm Gold clusters 5.1±0.8nm 5.2±2.3nm Nanotubes N/A 25-30nm From the table above, the only measurement really in question is the RAM length a. This may be due to the fact that the judgment of where the center of each http://opus.ipfw.edu/jaupli/vol0/iss0/2 4

Lane: Atomic Force Microscopy dot was is different for each person. The measured value was greater than all possible choices listed by the companies, but the closest value of all the companies was Company #4. If the measured value with its uncertainty overlapped and could potentially be one of the values from the lab manual of one of the companies, it was declared in agreement with that company, otherwise it was in disagreement. Again, RAM length a, was the only sample in disagreement with Company #4, but every other measured sample was in agreement. The data overall strongly supports that Company #4 was the creator of the samples. V. APPENDIX. Table 1 shows the measurements and the uncertainty. The uncertainty of each measurement was calculated by taking the average of the lengths, and then taking the standard deviation of the those same lengths. So the final value reads the average ± standard deviation. Table 1: Measurements of AFM Samples and Standard Deviation with final value as Average±Standard Dev. (Uncertainty Measurements) sample 1 RAM (micrometers) Average Standard Dev. Final Value length a 12.6 12.75 12.56 12.37 12.33 12.522 0.1728 12.5±0.2µm length b 6.989 7.182 7.416 7.008 6.782 7.0754 0.2374 7.1±0.2µm CD sample 2 (micrometers) length a 3.339 3.44 3.582 3.438 3.849 3.5296 0.1984 3.5±0.2µm length b 2.262 2.369 2.614 2.342 3.024 2.5222 0.3098 2.5±0.3µm sample 3 Gold clusters Sq 4.438 5.5377 4.7194 4.5351 6.2068 5.0874 0.7612 5.1±0.8nm In calculating these uncertainties, page 100 from Taylor was used as a reference for the standard deviations. Published by Opus: Research & Creativity at IPFW, 2012 5

Journal of the Advanced Undergraduate Physics Laboratory Investigation, Vol. 0 [2012], Iss. 0, Art. 2 REFERENCES [1] Magonov, Dr. Sergei N., and Dr. Myung-Hwan Whangbo. Frontmatter. In Surface Analysis with STM and AFM, Wiley VCH Verlag GmbH, p. I XII. http://onlinelibrary.wiley.com/doi/10.1002/9783527615117.fmatter/summ ary?systemmessage=wiley+online+library+will+be+disrupted+24+mar ch+from+10-14+gmt+%2806-10+edt%29+for+essential+maintenance (Accessed March 21, 2012). [2] Physics Lab Manual: Experimental Physics II. Physics 306, spring 2012. p.32-114. Minnesota State University Moorhead Physics Department, 2012. [3] Taylor, John Robert. An introduction to error analysis: the study of uncertainties in physical measurements. University Science Books, 1997. Print. http://opus.ipfw.edu/jaupli/vol0/iss0/2 6