Impact of Pore Exclusion on Reversed-Phase HPLC Column Performance

Similar documents
Improve HPLC Separations by Carefully Matching Column Pore-Size to Solute-Size

Superficially Porous Silica Particles with Wide Pores for Biomolecular Separations

Breaking the speed limit: Fast, high-resolution peptide and tryptic digest separations using fused-core particles

Sepax SRT -C, Zenix -C SEC Phases

Bringing uhplc Performance to the Separation of Peptides

New 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns

Changing the way you think about HPLC

Advantages of Agilent AdvanceBio SEC Columns for Biopharmaceutical Analysis

Alternatives to Sub-2 µm UHPLC Columns

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

LC Technical Information

How proteins separate on reverse-phase HPLC

Phenyl-Hexyl. UHPLC Columns. Alternate, complementary selectivity to C18 and C8 bonded phases

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

penta-hilic UHPLC COLUMNS

C18 Column. Care & Use Sheet

penta-hilic UHPLC COLUMNS

P R O D U C T B U L L E T I N. Fused-Core particle technology for hyper-fast and super-rugged HPLC columns

SRT -10/10C (10 µm ) Fast Protein Purification For FPLC, HPLC and Prep LC. Sepax Technologies, Inc. Preparative Size Exclusion Chromatography

Clearing the Confusion: GPC, SEC, GFC What, When, Why, and How?

Trends in Method Development for HPLC and UHPLC... the movement toward using solid-core particles

Chromatography- Separation of mixtures CHEM 212. What is solvent extraction and what is it commonly used for?

Improving HPLC Column Selection and System Performance

Advantages and Applications of Revolutionary Superficially Porous Particle Columns in Liquid Chromatography

HICHROM. Chromatography Columns and Supplies. LC COLUMNS HALO and HALO-5. Catalogue 9. Hichrom Limited

Gel Permeation Chromatography (GPC) or Size Exclusion Chromatography (SEC)

Method Development with ZirChrom's Ion Exchange Phases

NOTES. Zenix SEC-80. Sepax Zenix SEC-80 provides a valuable solution

Zirconia: the Ideal Substrate for Ion-Exchange LC and LC-MS

Reversed Phase Solvents

LC and LC/MS Column Selection Flow Chart

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

ANALYSIS OF BASIC COMPOUNDS

Logical Method Development for Peptides and Proteins Using RP-HPLC and SEC

Effects of Secondary Interactions in Size Exclusion Chromatography

Successfully Scaling and Transferring HPLC and UPLC Methods

Fast Protein and Peptide Separations Using Monolithic Nanocolumns and Capillary Columns

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns

Small Molecule Selectivity Sampler

Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents

Basic Principles for Purification Using Supercritical Fluid Chromatography

The Challenge: Sepax SEC C-line columns!

Packings for HPLC. Packings for HPLC

Liquid Chromatography

Peptide and protein analysis by capillary HPLC Optimization of chromatographic and instrument parameters. Application. Angelika Gratzfeld-Huesgen

High Speed 2D-HPLC Through the Use of Ultra-Fast High Temperature HPLC as the Second Dimension

High Resolution Fast LC

High Performance Liquid Chromatography

BSA pegylation PG Journal of chromatographic A, 1147 (2007)

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

2007 Waters Corporation. Application of UPLC to Peptide Mapping

ACE Excel SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability

HPLC & LC-MS. Techniques to Improve Speed, Resolution and Sensitivity without Investing in Capital Equipment

CHANGING THE WAY YOU THINK ABOUT HPLC

NOTES. Fusion Protein Separation. Highlighted FACTS:

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Effects of Aqueous Sample Content and Aqueous Co-Solvent Composition on UPC 2 Separation Performance

Separation of Structurally Similar Steroids on HALO C18 and PFP

Open Column Chromatography, GC, TLC, and HPLC

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

Superose High-performance Columns

HPLC COLUMNS WILEY-VCH. Theory, Technology, and Practice. Uwe D. Neue with a contribution from M. Zoubair El Fallah

Choosing Columns and Conditions for the Best Peak Shape

1.7 m Fortis UHPLC Columns

Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

The power of multi-detector SEC in the analysis of antibodies

LECTURE 2. Advanced Separation Science Techniques Present and Future Separation Tools

Introduction to Chromatography

QUALITY HPLC COLUMNS PARTICLE PORE PAGE SIZE SIZE. C18 Bidentate C18 Extremely long column life and rugged 4um 100Å 3

ACE SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability

ACE SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability

Applications and Properties of New Polymeric Mixed Mode Cation Exchange Chromatography Media

Electron Transfer Dissociation of N-linked Glycopeptides from a Recombinant mab Using SYNAPT G2-S HDMS

LIQUID CHROMATOGRAPHY

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

High Speed vs. High Resolution Analyses in HPLC: A Critical Performance Comparison of Column Options Using Poppe and Kinetic Plots

APPLICATIONS TN Peptide Purification Method Development: Application for the Purification of Bivalirudin. Introduction

Multi-mode Separations Using Zirconiabased Stationary Phases

Analytical Chemistry

High Performance Liquid Chromatography

High Performance Liquid Chromatography

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system

Introduction to Chromatographic Separations

Chromatographic Separation

Quick guide to selecting columns and standards for Gel Permeation Chromatography and Size Exclusion Chromatography SELECTION GUIDE

Lecture 26: More on Gel Filtration Chromatography and the Trypsin Resurrection Experiment

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Separation Methods Based on Distributions in Discrete Stages (02/04/15)

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Too Polar for Reversed Phase What Do You Do?

Fast HPLC Theory and Practice of Fast HPLC

H P L C C O L U M N S

Part 2. Overview of HPLC Media

UPLC Method Development and Validation

Transcription:

Impact of Pore Exclusion on Reversed-Phase HPLC Column Performance Richard A. Henry, Technical Consultant, 983 Greenbriar Drive, State College, PA 16801 (hyperlc@comcast.net) Stephanie Schuster, AMT, Inc., Silverside Road, Wilmington DE 19810 (SSchuster@advanced-materials-tech.com) 11-15-2016 EAS 2016 RAH 1

Two Main HPLC and UHPLC Particle Types Totally Porous Particle (TPP) Superficially Porous Particle (SPP) 300 Å pores 12-2016 LCGC 2016 RAH 2

Surface-Layer Porosity in 400Å Superficially-Porous Silica 1, 2 Large solutes diffuse 10-100 times more slowly than small ones. Unique performance and separation speed of SPP particles arise from a combination of large pores with short diffusion paths. Proven SPP advantages, including speed, for interactive modes (RP, NP, HILIC) may extend to the size exclusion mode (SEC). 12-2016 LCGC 2016 RAH 3

dv/dlog(w) Pore Volume (cm³/g Å) dv/dlog(w) Pore Volume (cm³/g Å) Common Pore-Size Ranges for Fused-Core Silica (SPP) 0.69 0.59 90Å 160Å 400Å 0.24 0.49 90 Å 0.19 0.39 160 Å 400 Å 0.14 0.29 0.09 0.19 0.09 0.04-0.01 10 100 1000 Pore Width (Å) -0.01 Small drugs Small peptides Large peptides Small proteins Large proteins Water (18 Da) Small drugs (< 5000 Da) Small peptides (< 2000 Da) Large peptides/small proteins (< 15,000 Da) Large proteins (15,000 to 500,000 Da) 2.0 Å 5-10 Å 20-30 Å 40-50 Å 100-200 Å 12-2016 LCGC 2016 RAH 4

How Pore-Size Impacts Column Performance 3 Porous and superficially porous particles are required in HPLC columns to create adequate surface area and retention. Small solutes have essentially full access to mesopores in 90Å- 130Å particles to allow rapid phase-solute interaction; however, even partial exclusion can adversely affect column separation when solute size approaches the pore diameter. When too confined, solutes lose freedom to diffuse and interact with the surface. Both equilibrium constants and (interaction) kinetics may be impacted. Ideally, one should know MW and size for all sample solutes and particle pore structure for HPLC columns (and match them carefully) in order to develop reproducible methods with optimum separation. 12-2016 LCGC 2016 RAH 5

Retention Equation for Pure Size Exclusion HPLC 4,5 V M = V 0 + V P (1) V R = V 0 + K Size V P (2) K Size varies from zero for total pore exclusion to one for total pore occupation. Separation by a pure SEC mechanism correlates well with solute hydrodynamic radius and pore geometry. The subject of this study is the point on the curve where solute size adversely affects column performance. 12-2016 LCGC 2016 RAH 6

Retention Equation for Pure Sorption HPLC Modes Retention by stationary phase interaction: V R = V M + K Sorp V S = V 0 + V P + K Sorp V S (3) K Sorp can vary from 0 to infinity, and V S stationary phase volume which is related to surface area. Equations (2) and (3) are usually employed separately under conditions where only one mode dominates. SEC mode behaves differently compared to retention modes such as RPC and can be very useful for analysis of complex samples by multi-dimensional techniques. 12-2016 LCGC 2016 RAH 7

Retention Equation When SEC and Sorption Modes Overlaps 6-10 When both large and small molecules are present, they will not experience equal access to the pores, therefore retention must be predicted by combining equations: V R = V 0 + K Size (V P + K Sorp V S ) (4) Equation (4) reduces to Equation (3) for small molecules (K Size = 1). When molecules become large enough to become partly excluded from pores (K Size < 1), retention factors will be reduced and efficiency (rate of mass transfer) may also be reduced. Rule of thumb: To eliminate undesirable size exclusion effects, select a column having an average pore diameter that is 10X the diameter of the largest solute in the sample. 12-2016 LCGC 2016 RAH 8

SEC Mode Overlap Impacts Column Performance Klein defined an occupational partition coefficient (K OC ) that is essentially equivalent to K size. The occupational partition coefficient impacts V S and solute retention in a complex fashion. Mode overlap regularly occurs between size exclusion and stationary phase retention modes of HPLC with microporous column particles. A complete understanding of this process is needed to optimize any HPLC method, especially one involving large-molecule separations. Klein used a cylindrical pore model and equation to explain the steric exclusion process based on solute-to-pore radius ratio. K OC = (1- a/r) 2 = K Size (5) where K OC is the occupation partition coefficient,(a) is the solute radius and (r) is the pore radius. Both (a) and (r) are variables for different samples and columns, respectively. 12-2016 LCGC 2016 RAH 9

Impact of Size Exclusion on Solute Retention 6 r a r a Cylindrical pore model Conical pore model Pore radius (r) varies within a range for porous particles; an average is usually compared. 1 0.9 0.8 Occupation Partition Coefficient, K OC Increasing MW or smaller pore Solute radius (a )increases with MW; an optimum ratio (a/r) exists for each sample component. As solutes become larger, they are excluded from pores to create the calibration curve, but they also lose access to surface area. K OC 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 Using this formula, solutes having 1/10 of the pore diameter can access ca. 80% of the pore volume and allow phase interaction to dominate. K OC = (1- a/r) 2 0.05 0.1 0.2 0.5 0.75 1 a/r 12-2016 LCGC 2016 RAH 10

Effect of Pore-Size on Efficiency for High MW Solute 11 Columns: 4.6 x 100 mm; Mobile phase: 31.5% ACN/68.5% Water/0.1% TFA; Flow rate: 1.0 ml/min; Temperature: 60 C; Sample: Bovine Insulin, MW = 5733; Injection volume = 10 µl 160 Å 90 Å Fused-Core Particles Bovine insulin: MW 5733 High efficiency is the key to high peak capacity; the larger 160Å pore particle shows higher efficiency and peak capacity for insulin and other large peptides compared to 90Å. Giddings predicted this but more explanation is needed. Based on SEC-150 column calibration, Bovine Insulin is < 50% excluded from the 160Å column pores, but likely is > 50% excluded from 90Å column pores. 12-2016 LCGC 2016 RAH 11

Absorbance (mau) 90 Å Pore-Size Column - Too Small for Many Peptides 12 90Å Fused-Core C18 0.233 0.058 0.063 0.178 0.243 1. Ribonuclease A (13,700 g/mol) 2. Bovine Insulin (5733 g/mol) 3. Human Insulin (5808 g/mol) 4. Cytochrome c (12,400 g/mol) 5. Lysozyme (14,300 g/mol) 160Å Fused-Core ES-C18 0.078 0.041 0.043 0.075 0.083 0 2 4 6 8 10 Time (min.) Columns: 100 mm x 4.6 mm; mobile phase: A: water/0.1% TFA; B: acetonitrile/0.1% TFA; gradient: 25 42% B in 10 min; flow rate: 1.5 ml/min; temperature: 30 C; detection: 215 nm; Peak widths in minutes above each peak. Retention, efficiency and sensitivity are lost with a 90Å column due to pore-crowding for the larger peptides The 160Å column is preferred for peptides. Surface areas 2.7 um 90 A 135 m 2 /g 2.7 um 160 A 90 m 2 /g 12-2016 LCGC 2016 RAH 12

Impact of Size Exclusion on Zone Broadening 13 Sources of solute zone spreading in packed-bed liquid chromatography have been widely studied. Schure has recently submitted a comprehensive analysis of HPLC zone broadening that includes a description of hindered diffusion inside pores (relative to free mobile phase diffusion) as solute size approaches average pore diameter. Hindered diffusion in cylindrical pore as function of ratio of molecule radius to pore radius. Although phase interaction is still the dominant mode at a radius ratio of 0.1, the pore diffusion coefficient has already dropped by almost 50% according to Schure. 12-2016 LCGC 2016 RAH 13

log MW Calibration Curve for Size Exclusion Chromatography* Chromatographic Conditions columns: Sepax Zenix SEC-150, 30 cm x 4.6 mm I.D., 3 µm mobile phase: 0.2 M potassium phosphate, ph 7.0 flow rate: pressure: 0.25 ml/min 66 bar column temp.: 25 C detector: injection: 0.5 µl samples: UV 215, 280 nm listed below Analyte MW (Da) Size (Å) K size 1 thyroglobulin 667000 200 0.00 2 SigmaMab 150000 0.06 3 IgG 150000 100 0.05 4 BSA 66400 0.16 5 ovalbumin 45000 0.24 6 myoglobin 17000 40 0.33 7 ribonuclease A 13700 0.38 8 bovine insulin 5700 0.60 9 neurotensin 1700 0.66 10 vitamin B12 1350 20 0.85 11 angiotensin II 1000 0.74 12 uracil 112 1.00 7 6 5 4 3 2 Solute/pore radius 0.1-1.0 1 2,3 4 5 6 7 y = -1.4279x + 8.2024 R² = 0.9641 1 V 0 V M 0 1.50 2.00 2.50 3.00 3.50 4.00 4.50 Elution Volume Solute/pore radius < 0.1 8 9 10 * Unpublished data by AMT Applications Lab 11 12 12-2016 LCGC 2016 RAH 14

105 95 85 75 65 55 45 35 25 15 5-5 105 95 85 75 65 55 45 35 25 15 5-5 160 Å Pore-Size Column- Too Small for Large Proteins 15 160 Å ES-C18 2.7 µm Fused-Core 0.0384 0.0328 0.0413 0.0396 0.1019 0.0849 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 400 Å ES-C18 3.4 µm Fused-Core 0.0291 0.0240 Time (min.) 0.0287 0.0247 0.0467 0.0315 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Time (min.) Peak width (min) at 50% shown Peak Identities: 1. Ribonuclease A 13.7 kda 2. Cytochrome c 12.4 kda 3. Lysozyme 14.3 kda 4. α-lactalbumin 14.2 kda 5. Catalase 60 kda (subunit) 6. Enolase 46.7 kda Efficiency and sensitivity are lost with the 160Å column due to pore-crowding for the larger proteins. The 400Å column is preferred for proteins. Columns: 4.6 x 100 mm Gradient: 23-50 %B in 15 minutes A: water/0.1% TFA B: ACN/0.1% TFA Flow rate: 1.5 ml/min Detection: 215 nm Temp: 60 C Injection: 5 µl Peak widths in minutes above each peak. Surface areas 2.7 um 160 A 90 m 2 /g 3.4 um 400 A 15 m 2 /g 12-2016 LCGC 2016 RAH 15

Absorbance (mau) Absorbance (mau) Large Protein Comparison on 160 Å and 400 Å Columns 14 9 8 7 6 5 4 3 2 1 0-1 9 8 7 6 5 4 3 2 1 0-1 Myosin cellular protein >200 kda Heavy chain subunits @ ~220 kda 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5 Time (min.) 2.7 µm Fused-Core 160 Å ES-C18 2.7 µm Fused-Core Prototype 400 Å ES-C18 Efficiency is higher on 400Å pore size particle and retention is equal, even with much lower surface area. The 400Å column is preferred for Myosin. Surface areas 2.7 um 160 Å 90 m 2 /g 2.7 um 400 Å 29 m 2 /g Conditions: columns= 2.1 100 mm; temperature= 80 C; flow rate = 0.45 ml/min. Gradient 35 65% B in 15 min: A = 0.1% trifluoroacetic acid; B = acetonitrile with 0.1% trifluoroacetic acid. Detection = 215 nm; injection volume = 1 μl; sample = myosin, 400 kda initial, but denatured in mobile phase. 12-2016 LCGC 2016 RAH 16

SILU Lite SigmaMAb Comparison on 160Å and 400Å 15 Solute size of MAb is now approaching the average pore diameter of the 160Å column. The 400Å column is preferred. 2.7 µm Fused-Core Prototype 160 Å C4 Surface areas 2.7 um 160 Å 90 m 2 /g 3.4 um 400 Å 15 m 2 /g Width 50% = 0.3055 min Corrected Relative Retention* = 13.30 3.4 µm Fused-Core 400 Å C4 Width 50% = 0.0991 min Corrected Relative Retention* = 14.08 * Retention time relative to solvent front Time (min.) 0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 Columns: 150 mm x 2.1 mm; mobile phase: A: water/0.1% difluoroacetic acid; B: acetonitrile/0.1% difluoroacetic acid; gradient: 27-37% B in 20 min; flow rate: 0.4 ml/min; temperature: 80 C; injection volume: 2 µl; detection: 280 nm 12-2016 LCGC 2016 RAH 17

mau, 254 nm Relative Scale Wide-Pore Column Comparisons for Small Molecules 15 Acquity, 1.7 µm Fused-Core, 3.4 µm BEH 300Å C4 1.7 µm; ~90 m 2 /g 1-chloro-4-nitrobenzene 1.0 0.8 0.6 0.4 0.2 Fused-Core 400Å C4 3.4 µm; 15 m 2 /g 1-chloro-4-nitrobenzene Retention time, min Either column would work for smaller molecules. 0.0 Efficiency (N 1-chloro-4- nitrobenzene) Retention (k 1-chloro-4- nitrobenzene) Chromatographic Conditions columns: Fused-Core C4, 15 cm x 2.1 mm I.D., 3.4 µm Acquity UPLC Protein BEH C4, 15 cm x 2.1 mm I.D., 1.7 µm mobile phase: [A] water; [B] acetonitrile; (75:25, A:B) flow rate: 0.3 ml/min pressure: 552 bar (BEH); 177 bar (Fused-Core) column temp.: 25 C detector: 254 nm injection: 0.2 µl sample: uracil, phenol, propiophenone, 1-chloro-4-nitrobenzene 12-2016 LCGC 2016 RAH 18

mau, 280 nm Relative Scale 18 16 14 12 10 8 Column Comparisons for Monoclonal Antibody 15 BEH 300Å C4 1.7 µm; ~90 m 2 /g Acquity, 1.7 µm Fused-Core, 3.4 µm 6 4 2 0-2 18 16 14 12 10 0 2 4 6 8 10 12 Fused-Core 400Å C4 3.4 µm; 15 m 2 /g 8 6 4 2 0-2 0 2 4 6 8 10 12 Retention time, min Fused-Core 400Å has more retention, at lower surface area, and much higher efficiency than Acquity 300Å. Fused-Core 400Å column would be preferred for this MAb molecule. Chromatographic Conditions columns: Fused-Core C4, 15 cm x 2.1 mm I.D., 3.4 µm Acquity UPLC Protein BEH C4, 15 cm x 2.1 mm I.D., 1.7 µm mobile phase A: 0.1% DFA in water mobile phase B: 0.1% DFA in acetonitrile flow rate: 0.4 ml/min gradient: 29 to 35% B in 12 min initial pressure: 322 bar (BEH); 115 bar (Fused-Core) column temp.: 80 C detector: 280 nm injection: 2 µl sample: SILu Lite SigmaMAb; 0.5 mg/ml 12-2016 LCGC 2016 RAH 19

Recognizing a Column Pore-Exclusion Problem A pore exclusion problem might exist: 1. When high MW solutes elute close to or even before low MW solutes, under either isocratic or gradient conditions. 2. When high MW solutes show much lower efficiency than expected compared to lower MW solutes. 3. When higher MW solutes show improved retention and efficiency after a larger pore column is installed. Larger pore-size analytical columns should be included in column screening, especially when sample composition is unknown. The use of SEC columns for early screening is recommended. Analytical columns should be screened initially in RP or another retention mode with a low or moderate MW test mixes to establish expected retention and efficiency performance. 12-2016 LCGC 2016 RAH 20

Heavy and Light Chain Reduced Fragments of SigmaMAb 15 10 mv 9 8 7 6 5 4 3 BEH 300Å C4 1.7 µm; ~90 m 2 /g 0.081 0.262 Pressure = 342 bar 2 1 0-1 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 min 10 mv bar 350 300 250 200 150 100 50 0 bar Peak Identities: 1. Light Chain (x2) 23 kda 2. Heavy Chain (x2) 50 kda Reduced SigmaMAb 0.5 µg/2 µl (0.1% TFA) 20-40% AcN/0.1% Formic acid in 20 min 0.5 ml/min; 80 C 2.1 mm x 150 mm 9 8 7 6 5 4 3 Fused-Core 400Å C4 3.4 µm; 15 m 2 /g 0.088 0.101 Pressure = 118 bar 2 100 1 50 0-1 0 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 min 350 300 250 200 150 Both columns perform well for the light chain fragment but the lower surface area 400Å has better retention. The 400Å column is preferred because of both narrow peaks and more retention for the fragments. 12-2016 LCGC 2016 RAH 21

Using SEC Columns to Estimate Solute Size Select SEC columns with similar pore size to high resolution analytical columns available. Two or more SEC columns are useful. Select standards with known size for calibrating target solutes; samples are often injected separately since SEC is a low resolution technique. Solute markers near V 0 and V M are very important. Pore exclusion would not be expected for target solutes that elute from SEC columns close to V M (solute/pore radius <0.1). Performance loss would be expected for solutes eluting with K size < 0.5 (solute/pore radius 0.1-1.0) due to poor phase access and hindered pore diffusion within the analytical column. Solutes of interest should never be retained by phase interaction under optimum SEC conditions. Retention by stationary phase makes solutes appear too small. Nothing should elute after V M in an SEC experiment. Under certain conditions, RP and other columns may be useful with strong solvents for estimating solute size 16. 12-2016 LCGC 2016 RAH 22

Conclusions Column retention and efficiency can both be lost for larger sample components when size exclusion mode significantly overlaps with stationary phase retention modes to partially exclude large solutes from stationary phase and interfere with diffusion processes. Guidelines have been proposed that solutes should be less than ca. 10% of the column exclusion limit for optimum performance. For example, if the exclusion limit is 150,000 Da, largest solute size to avoid significant peak retention and efficiency loss would be 15,000 Da. Preliminary sample screening by SEC can save valuable time by identifying solutes that might be subject to pore exclusion and require evaluation of larger pore columns. When data on solute sizes and average column pore size is not available for selecting optimum radius ratios in advance, the use of larger pore columns is recommended for screening unknown samples. 12-2016 LCGC 2016 RAH 23

References 1. J. J. Kirkland, et al., J. Chromatogr. A, 1315 (2012) 118-126. 2. J. J. Kirkland, et al., LCGC N. A., 33 No. 11 Nov 2015 856-865. 3. J. C. Giddings, Unified Separation Science, Chapter 2, Entropy Effects in Porous Media, Wiley (1991); J. C. Giddings, Dynamics of Chromatography: Principles and Theory, CRC Press, Boca Raton, FL (2002). 4. J. J. Kirkland, J. Chromatogr. 125 (1976) 231-250. 5. High Performance Liquid Chromatography, J. H. Knox, ed. Edinburgh University Press (1978). 6. J. Klein and K. Treichel, Chromatographia, 10 (1977) 604-610. 7. F. Eisenbeis and S. Ehlerding, Kontakte 1/78 (1978) 22-29. 8. K. Unger, Porous Silica, Chapter 9 Size-Exclusion, Elsevier (1979). 9. H. Engelhardt and D. Mathes, J. Chromatogr., 185 (1979) 305-319. 10. G. Barka and P. Hoffman, J. Chromatogr., 389 (1987) 273-278. 11. S. Schuster et al., J. Chromatogr. Sci., 48 (2010) 566-571. 12. J.J. Kirkland, et al., J. Pharm. Anal., 3 (2013) 303 312. 13. M. Schure, submitted to J. Chromatogr. A, (2016). 14. B.M. Wagner, et al., J. Chromatogr. A, 1264 (2012) 22 30. 15. Unpublished data by AMT Applications Lab (2016). 16. Private communication from Joe Foley. 12-2016 LCGC 2016 RAH 24

Acknowledgements Jack Kirkland and Barry Boyes of AMT for general column discussions. Mark Schure of Kroungold Analytical, Inc. for conversations about particles, pores and peak broadening. Hillel Brandes of Supelco Division of MilliporeSigma for discussions about preparing SEC calibration plots. Will Miles and Bob Moran of AMT for chromatographic data acquisition. Trademarks: Fused-Core is a registered trademark of Advanced Materials Technology, Inc. Acquity is a registered trademark of Waters Corporation. SigmaMAB is a trademark of Sigma-Aldrich Co. LLC. Zenix is a trademark Sepax Technologies, Inc. 12-2016 LCGC 2016 RAH 25