Muliple Gas#3 plus Sulfur GC Configuration

Similar documents
Thermal Soil Desorber Instructions September 2006

NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25)

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian

Application Note S/SL

Medical Cannabis Pesticide Screening using the SRI 8610C GC

Far UV Absorbance Detector

Gas Chromatography. Introduction

CHAPTER 6 GAS CHROMATOGRAPHY

Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary.

*Correspondence to:

Chromatography. Gas Chromatography

Monomer Analysis. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Introduction to Gas Chromatography

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Selection of a Capillary

Petrochemical. Transformer Oil Gas Analysis - TOGA

How To Select the Correct GC Column. Simon Jones Application Engineer

Selection of a Capillary GC Column

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

GC Instruments. GC Instruments - Sample Introduction

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Introduction to Capillary GC

Chapter 27: Gas Chromatography

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Chemistry Gas Chromatography: Separation of Volatile Organics

Gas Chromatography notes

Chapter 11 Conventional Gas Chromatography

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

GAS CHROMATOGRAPHY (GC)

Application Note 032

Method for the determination of 1,3-butadiene

Making Improvements to Today s Natural Gas Analysis Systems

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

2401 Gas (liquid) Chromatography

Gas Chromatography (GC)

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.

Determination of Volatile Substances Proof of Food Adulteration

Analysis of Sulfur-Containing Flavor Compounds By GC/MS With A PFPD

Chapter 31 Gas Chromatography. Carrier Gas System

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Understanding Gas Chromatography

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

Understanding the Capillary GC Column: How to Choose the Correct Type and Dimension

G a s C h r o m a t o g r a p h y Determination of Ethanol in Wine by Head-Space Gas Chromatography

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

Experimental Determination of Pyrolysis Products from Carbon/Resin Ablative Materials

Gas chromatography. Advantages of GC. Disadvantages of GC

Selective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of 2,2 Dinitrobiphenyl with TiO 2 and UV light irradiation

1.0 Reactive Gas Chromatography Design and Setup Procedure for Catalyst Pelletization & Inlet Packing Software and Method.

Abstract. Introduction

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

Continuous Improvement in Petroleum/Petrochemical Analysis HP s Family of Innovative PLOT Columns

637. Thiamethoxam. HPLC method

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Gas Chromatography. Rosa Yu, David Reckhow CEE772 Instrumental Methods in Environmental Analysis CEE 772 #16 2

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

7 INSTRUMENTAL CHROMATOGRAPHY

Analysis for Natural Gas and Similar Gaseous Mixtures by Gas Chromatography

Instrumental Chemical Analysis

SPIKING ANALYSIS OF DRY GAS

Use of the SFC-MS Splitter Kit G

Product Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas

Virtually Particle-Free Rt -Silica BOND Columns

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC

Method for the determination of dimethyl sulfate

Agilent G2350A Atomic Emission Detector (AED)

High-Speed Gas and Headspace Analysis for the Process-Line and Laboratory: SIFT- MS IFPAC 2017

COOKBOOK FOR THE PERKINELMER SWAFER TM TECHNOLOGY FOR GC APPLICATIONS

Thermal Conductivity Detector Analysis of Hydrogen Using Helium Carrier Gas and HayeSep D Columns

Skoog/Holler/Crouch Chapter 26 Principles of Instrumental Analysis, 6th ed. CHAPTER 26

AppNote 7/1994. Friedhelm Rogies, Andreas Hoffmann Gerstel GmbH & Co.KG, Aktienstrasse , D Mülheim an der Ruhr, Germany

Disadvantage: Destructive Technique once analyzed by GC, the sample is lost

Experiment 8: Chlorination of 1-Chlorobutane

Fall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers.

STANDARD OPERATING PROCEDURES

Chemistry Instrumental Analysis Lecture 31. Chem 4631

INVESTIGATING GAS CHROMATOGRAPHY rev 8/12

10/27/10. Chapter 27. Injector typically 50 C hotter than oven

Using a mini-gas Chromatograph (GC): Identifying Unknown Compounds

Gas Chromatography. Chromatography Laboratory Course. Dr. Christian Jungnickel Chromatography Course GC September 2005

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Permanent Gas Analysis by Gas Chromatography Vacuum Ultraviolet Spectroscopy

Gas Chromatographic Analysis of Atmospheric Sulfur Compounds Using Cryogenic Sampling and A Modified Thermal Desorber

DEHYDRATION OF ALCOHOLS-GAS CHROMATOGRAPHY

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Gas chromatography. Flow measurement. Flow control. Injection methods. Flow measurement

LCT. Detailed Crude Oil Analysis: Processes for the future

Chemistry 101 Chapter 14 Liquids & Solids

Methods for the determination of vinyl chloride

a) ion-ion attractions b) London dispersion forces c) hydrogen bonding forces d) dipole-dipole attractions

Mass Spectrometry. General Principles

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Gas Chromatography CHEM Dr. Reem M. Alghanmi st term

Fundamentals of GC: Introduction: Sample introduction:

Transcription:

The SRI Model 8610C Gas Chromatograph ( GC ) configured as a MultipleGas#3 plus Sulfur is designed to measure H2, O2, N2, CO, CO2, H2O, C1 through C5 hydrocarbons and also H2S, COS/SO2, and other sulfur molecules such as mercaptans, CS2, DMS, DMDS, Thiophenes and more in a single analysis. The GC is equipped with three detectors: The Thermal Conductivity Detector ( ) measures all non-sulfur molecules from 500ppm to 100% The FID/Methanizer detector ( FID/meth ) measures hydrocarbons plus CO and CO2 from 1ppm to 50,000ppm The Flame Photometric/FID combo detector ( FPD/FID ) measures all sulfur molecules plus hydrocarbons. FPD/FID FID/meth Inside the column oven are three columns. The Haysep-D ( HD ) and MoleSieve ( MS13X ) columns together separate H2, O2, N2, CO, CO2 and C1-C5 hydrocarbons as well as water. The 60meter MXT1 capillary column separates the sulfur molecules and also hydrocarbons from C1-C10 60MXT1 Capillary Haysep-D MS13X www.srigc.com Page 1

The 60meter MXT1 capillary column is connected to the Haysep-D column using a tee fitting. When the sample is injected, it is split so about half the sample flows into the HD column and the other half flows into the capillary column. Notice how the cap column is inserted into the sample delivery tube in such a way that the sample splits cleanly. If the connection is not made in this way, the peaks will not be as sharp. The other end of the capillary column is connected to the FPD/FID detector using a Swagelok nut and graphite ferrule. The and FID/methanizer detectors are connected in series, so all molecules exit the HD/MS13x columns and flow first through the and then exit the and flow into the FID/methanizer via a 1/16 stainless steel tube. www.srigc.com Page 2

A 10 port Valco gas sampling valve is mounted in the heated valve oven. The valve is plumbed as shown in the diagram. In the Load position the carrier gas flows through the three columns and into the detectors but the sample loop is isolated so sample can be loaded into the loop. MS13X column FID/meth FPD/FID When the analysis is started, the valve rotates to the Inject position so the carrier gas now pushes the sample out of the loop, to the tee fitting where it splits into two paths. FID/meth tee FPD/FID www.srigc.com Page 3

The sample loop is loaded with new sample by flushing the loop with 10ml or more of fresh sample. The loop itself is 1ml, but it takes 10ml to completely flush it. Sulfur compounds like H2S can be lost on active metal surfaces, so we use a special kind of tubing to make the loop called Sulfinert. 1ml Sulfinert loop Sample can be pushed through the loop with a syringe, or it can flow continuously, or it can be pulled through with suction ( vacuum pump ). For sulfur compounds it is important to use Teflon tubing to avoid losses. The GCs temperature program and event table is set up as shown. Users may expect to make modifications depending on the exact molecules being measured. www.srigc.com Page 4

Since there are four detectors, the PeakSimple software will show four chromatograms simultanouosly on the computer monitor. FPD FID part of FPD/FID FID/methanizer Zooming in on the FPD chromatogram you can see Hydrogen Sulfide ( H2S ), Carbonyl Sulfide ( COS ) and Di Methyl sulfide ( DMS ). The FPD is blind to hydrocarbons like methane. The FID part of the FPD/FID combo detector detects methane and DMS ( C2H6S ), but does not detect H2S and COS since the FID only detects molecules with carbon-hydrogen bonds. When the FPD/FID combo detector is optimized for best sulfur detection, the FID sensitivity and range is reduced to less that what a normal FID detector would deliver. FPD FID part of FPD/FID www.srigc.com Page 5

The FID/methanizer detects all hydrocarbons from methane to hexane and also CO and CO2 which are converted to methane by the methanizer. Detection from 1ppm to 50,000ppm are possible. Shown are methane, CO and CO2 at 1000ppm. FID/methanizer The detects Hydrogen, Oxygen, Nitrogen, Methane, CO, CO2, Water and all molcules with boiling points below hexane. Detection of most molecules is possible from 500ppm to 100%. Hydrogen can be detected from 10ppm to 100% if using Nitrogen or Argon carrier, but this increases detection limits for everything else from 500ppm to about 5000ppm. With helium carrier detection limit for hydrogen is about 10,000ppm to 100%. In some cases, the GC can be equipped with a second detector, valve and column specifically to detect Hydrogen while helium carrier is used for the other molecules. www.srigc.com Page 6