Calibration of a pencil ionization chamber with and without preamplifier

Similar documents
CALIBRATION OF IONIZATION CHAMBERS FOR COMPUTED TOMOGRAPHY BEAMS IN BRAZIL: THE PRESENT REALITY*

Intercomparison of Ionization Chambers in Standard X-Ray Beams, Radioprotection Level

INVESTIGATIONS OF THE FRICKE GEL (FXG) DOSIMETER DEVELOPED AT IPEN IRRADIATED WITH 60 Co GAMMA RAYS

COMPARISON OF ABSORBED DOSE TO AIR CALIBRATION FACTORS FOR A PARALLEL PLATE IONIZATION CHAMBER*

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1

Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network /2017

Comparison of air kerma measurements for tungsten anode based mammography x-ray beam qualities (EURAMET.RI(I)-S4.1)

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co

The Radiation Effect of 60 Co Gamma Rays on Polycarbonate detector.

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Use of a radioactive check device for redundancy check of ionization chambers

M [scale units/s] of the system

Method to evaluate the L/D Ratio of Neutron Imaging Beams

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS

Commissioning of the Beta Secondary Standard (BSS2)

City University of Hong Kong

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

CALIBRATION OF DOSEMETERS USED IN MAMMOGRAPHY WITH DIFFERENT X RAY QUALITIES: EUROMET PROJECT NO. 526

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *

Reference Dosimetry for Megavoltage Therapy Beams: Electrons

CONSIDERATION ON THE H p(10) AND H*(10) SECONDARY STANDARD CHAMBER CHARACTERISTICS

N. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer

OXYGEN INHIBITION AND COATING THICKNESS EFFECTS ON UV RADIATION CURING OF WEATHERFAST CLEARCOATS STUDIED BY PHOTO-DSC

Dosimetry: Electron Beams

TECHNIQUE OF MEASURING THE FIRST IONIZATION COEFFICIENT IN GASES

Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays

Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration

Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3)

Research Article Evaluation of the X-Ray Absorption by Gold Nanoparticles Solutions

COMPARISON OF PERSONNEL RADIATION MONITORING DOSIMETERS DESIGNED FOR MEDICAL FIELD

Electron density and effective atomic number images generated by dual energy imaging with a 320-detector CT system: A feasibility study

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden

Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015

THERMOLUMINESCENCE PROPERTIES OF OLIVINE

Ionizing Radiation Dosimetry and Medical Physics

Development of an irradiation system for a small size continuous run multipurpose gamma irradiator*

USE OF DLP FOR ESTABLISHING THE SHIELDING OF MULTI- DETECTOR COMPUTED TOMOGRAPHY ROOMS

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

Monte Carlo Simulation concerning Particle Therapy

Radhakrishnan B*, Kurup P G G**, Ramakrishnan G***, Chandralingam S****

Initial Certification

Efficiencies of Some Spherical Ion Chambers in Continuous and Pulsed Radiation: A Numerical Evaluation

General characteristics of radiation dosimeters

Proposed Room Requirements for CT System

Simulation Modeling in Dosimetry

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

MEASUREMENT OF DIAGNOSTIC X-RAY SPECTRA USING CdZnTe DETECTOR

Trace element quality control analysis of environmental samples at the Neutron Activation Analysis Laboratory, IPEN, São Paulo, Brazil

EVALUATION OF Ra, Th, K AND RADIUM EQUIVALENT ACTIVITY IN SAND SAMPLES FROM CAMBURI BEACH, VITÓRIA, ESPÍRITO SANTO, BRAZIL.

THE ACTIVITY CALIBRATOR

Evaluation of Phantom Equivalent Materials in Polychromatic Diagnostic X-Ray Beam

Neutron Spectrometry in Mixed Fields: Characterisation of the RA-1 Reactor Workplace

Work Programme. of the International Bureau of Weights and Measures. for the four years Comité international des poids et mesures

B E A M Q U A L I T Y S P E C I F I C A T I O N I N K I L O-V O L T A G E R A D I O T H E R A P Y

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required

Evaluation of Correction Factors Applied in Photon Calibration of NIS TE Neutron Ionization Chambers

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields

) for Varian TrueBeam high-dose-rate therapy beams

Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kv to 300 kv

Characterization of Low-Energy (6-30 kev) Response of Polish TLDs (MTS-N, MCP-N) with Synchrotron Radiation and Determination of Some TLD Quantities.

Occupational Radiation Protection at Accelerator Facilities: Challenges

ความร พ นฐานและเทคน คการบ าร งร กษา เคร องโดสคาล เบรเตอร (Dose Calibrator) ส ว ทย ป ณณช ยยะ

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers

Computation of energy imparted in diagnostic radiology

Jeancarlo Torres, CHP. Naval Surface Warfare Center Carderock Division

7. a XV-2 high spatial resolution lm detector (Kodak). Important parameters of these detectors are given in Table1. The ionization chambers and the di

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS

Optimization of microionization chambers for small-field reference dosimetry

Half-life data a critical review of TECDOC-619 update

Small Field Dosimetric Measurements with TLD-100, Alanine, and Ionization Chambers

ISO Radiological protection Sealed radioactive sources General requirements and classification

R function for residual analysis in linear mixed models: lmmresid

Ion-Chamber Survey Meter OD-02

Tritium concentration analysis of groundwater samples from the environmental monitoring program at IPEN, São Paulo, Brazil

IAEA TRS May 2001 (V.10A) PUBLISHED BY THE IAEA ON BEHALF OF IAEA, WHO, PAHO, AND ESTRO INTERNATIONAL ATOMIC ENERGY AGENCY IAEA

Comparison of some human tissues and some commonly used thermoluminescent dosimeters for photon energy absorption

CALCULATION OF THE DETECTOR-CONTRIBUTION TO ZIRCONIUM PEAKS IN EDXRF SPECTRA OBTAINED WITH A SI-DRIFT DETECTOR

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays

STUDY ON TRACE ELEMENT DETERMINATION IN HUMAN HEAD HAIR USING NEUTRON ACTIVATION ANALYSIS

Bureau International des Poids et Mesures

MECHANICAL AND THERMAL PROPERTIES OF COMMERCIAL MULTILAYER PET/PP FILM IRRADIATED WITH ELECTRON-BEAM

Energy response for high-energy neutrons of multi-functional electronic personal dosemeter

Chapter 10 Acceptance Tests and Commissioning Measurements

Bases of radioisotope diagnostic methods

Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry

Radioactive Waste Management

CALIBRATION OF IN VITRO BIOASSAY METHODOLOGY FOR DETERMINATION OF 131 I IN URINE

Real time evaluation of overranging in helical computed tomography

Comparison between TG-51 and TRS-398: Electron Contamination Effect on Photon Beam Quality Specification.

INTERCOMPARISON OF PERFORMANCE CHARACTERISTICS OF OSLDs. AND TLDs USED FOR INDIVIDUAL MONITORING

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe

Transcription:

Calibration of a pencil ionization chamber with and without preamplifier Ana F. Maia and Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares Comissão Nacional de Energia Nuclear Av. Prof. Lineu Prestes, 2242 05508-000, São Paulo, Brazil afmaia@ipen.br / lcaldas@ipen.br Abstract. The pencil ionization chamber is a cylindrical dosimeter developed for computed tomography beams. Many kinds of ionization chambers have a preamplifier connected to the chamber to make it electrically more stable, specially for field instruments. In this study, the performance of a Victoreen pencil ionization chamber with the original preamplifier and after its removal was compared. The objective of the preamplifier removal was to enable connecting the chamber to other kinds of electrometers available in the laboratory. To assure the proper performance of the chamber in its new condition, the main quality control tests were performed, as short and long term stability, leakage current and energy dependence in diagnostic and mammography calibration beams. The behavior of the pencil ionization chamber before and after the removal of the preamplifier was very similar, and the results obtained were always within the limits of international recommendations. The results obtained in both situations allow, if necessary, the preamplifier removal of the system without lack of precision in the measurements. 1. Introduction The pencil ionization chamber was developed for computed tomography (CT) dosimetric purposes, and since its introduction it has been very often used. The pencil ionization chamber allows the determination of the computed tomography dose index (CTDI), the most common dose quantity in CT dosimetry [1-5]. Many kinds of ionization chambers were manufactured with a signal digitizing preamplifier, that guarantees a better electrical stability, and it is useful specially for field instruments. However, in most cases, the preamplifier can only be coupled to a specific electrometer. Nevertheless, for laboratory instruments it may be interesting to remove the preamplifier to earn more versatility. The aim of this study was to evaluate the performance of a Victoreen pencil ionization chamber with the original preamplifier and after its removal. This pencil ionization chamber could have been bought without its preamplifier from the manufacturer, but at that time it did not seem to be necessary. Only recently this need appeared, and a laboratory adaptation was made removing the preamplifier and using a triaxial connector to enable the use of the chamber with other kinds of electrometers available in the laboratory. The main quality control tests, as short and long term stability, leakage current and energy dependence in diagnostic and mammography calibration beams, were performed to verify the performance of the chamber with and without the preamplifier. The study was realized using the available radiation beams at the laboratory. 2. Materials and Methods A Victoreen pencil ionization chamber, model 660-6, was tested coupled to a Victoreen electrometer, model 660, when using its original preamplifier, and coupled to a PTW electrometer, model UNIDOS 10001, after the preamplifier removal. The chamber presents a sensitive volume of 3.2 cm³, a sensitive length of 10 cm, and it is not sealed. The measures of this chamber are proportional to the irradiated length. 1

A 90 Sr + 90 Y check source, Physikalisch-Technische Werkstätten (PTW; 5.77 MBq, 2003), was used to perform the repeatability and reproducibility tests. For most types of ionization chambers, supports for repeatability and reproducibility tests are commercially available. However, this is not the case of supports for pencil ionization chambers. Therefore, a homemade acrylic support was utilized for those two tests [6]. It was not possible to measure the leakage current before the preamplifier removal, because of its manual zero adjustment. After the removal, the leakage current was measured before and after irradiation with the PTW check source. Two X-ray systems were used for the energy dependence test. The first one was a diagnostic radiology level equipment, Medicor Mövek Röntgengyara, model Neo-Diagnomax, that operates from 40 to 125 kv at the radiographic mode, and from 45 to 100 kv at the fluoroscopic mode. Diagnostic qualities defined by the International Electrotechnical Commission, IEC 61267 [7], were used in this system, and their parameters are listed in Table I. The reference system for these qualities was a parallel plate ionization chamber with 1 cm³ of sensitive volume, PTW, model 77334, with a PTW electrometer, model UNIDOS 10001. This chamber was calibrated by PTW, with traceability to the German Primary Standard Dosimetry Laboratory, Physikalisch-Technische Bundesanstalt (PTB). Table I. IEC diagnostic radiology qualities in the Medicor Mövek Röntgengyara X-ray equipment. Radiation Quality Voltage (kv) Total Filtration (mmal) Half-Value Layer (mmal) Effective Energy (kev) Direct Beams RQR3 52 2.5 1.82 32.0 RQR5 70 2.5 2.45 39.2 RQR7 90 2.5 3.10 46.0 Attenuated Beams RQA3 52 12.5 4.0 38.8 RQA4 63 18.5 5.7 45.6 RQA5 70 23.5 7.1 51.8 RQA6 80 29.5 8.4 57.9 RQA7 90 32.5 9.1 62.9 The second X-rays system was a low energy system Rigaku Denki Co. Ltd. Generator, type Geigerflex, and a Phillips tube, model PW 2184/00. This equipment operates from 20 to 60 kv. Mammography qualities similar to those from the National Institute of Standards and Technology (NIST), defined by the International Atomic Energy Agency (IAEA) [8] were used in this system. In Table II the parameters of these qualities are listed. The reference system utilized in this case was a parallel plate ionization chamber Radcal Corporation, model 10x5-6M, with a Radcal Corporation electrometer, model 9015; its sensitive volume was 6 cm³. This chamber has a calibration certificate from the Center for Devices and Radiological Health, Food and Drug Administration, USA, and its calibration is traceable to NIST. 2

Table II. Mammography qualities in the Rigaku Denki X-ray equipment. Radiation Quality Half-Value Layer (mmal) Effective Energy (kev) Voltage (kv) Direct Beams (Total Filtration: 0.06 mmmo) RXM20 20 0.28 13.6 RXM23 22.5 0.32 14.8 RXM25 25 0.33 15.1 RXM28 27.5 0.34 15.3 RXM30 30 0.35 15.6 RXM32 32.5 0.37 16.0 RXM35 35 0.38 16.2 Attenuated Beams (Total Filtration: 0.06 mmmo + 2mmAl) RXM20x 20 0.52 18.5 RXM23x 22.5 0.56 18.7 RXM25x 25 0.58 18.8 RXM28x 27.5 0.61 19.0 RXM30x 30 0.67 19.5 RXM32x 32.5 0.72 19.7 RXM35x 35 0.85 21.6 3. Results and Discussion The repeatability test was performed by taking several measurements with the chamber exposed to the check source under reproducible conditions. According to international recommendations [9], the maximum acceptable coefficient of variation is 1% for CT specific chambers. The highest coefficient of variation obtained was 0.32%, before the preamplifier removal, and 0.22%, after the preamplifier removal, that are values within the limit. The reproducibility test was obtained by plotting the results of the repeatability test, before and after the preamplifier removal, in function of time, so the ionization chamber long-term stability could be observed for both conditions. As stated by the norm IEC 61674 [9], the medium value obtained in each repeatability test must not differ from the reference value more than 3%. FIG. 1 and 2 show the results obtained for the pencil ionization chamber. All values obtained in this work were within the recommended limit. The leakage current was measured during 20 minutes, before and after irradiation, and the maximum leakage current obtained was 0.34% of the measurement obtained in the repeatability test. According to the norm IEC 61674 [9], the leakage current of a dosimeter shall not exceed 5% of the minimum effective air kerma rate of the range in use for at least 1 minute. The maximum leakage current obtained before irradiation represented only 1.4% of the ionization current produced by the minimum air kerma rate used in this study (0.534mGy/min). 3

FIG. 1. Reproducibility test of the Victoreen pencil ionization chamber, with the preamplifier, using a 90 Sr + 90 Y check source. FIG. 2. Reproducibility test of the Victoreen pencil ionization chamber, without the preamplifier, using a 90 Sr + 90 Y check source. 4

For the energy dependence test, the chamber was calibrated in the beam qualities described in Tables I and II. The calibration coefficients [10] obtained are showed in Tables III and IV. The energy dependence (difference between the highest and the lowest calibration coefficients in percentage) of the pencil ionization chamber for those qualities is showed in Table V for both cases (with and without the use of preamplifier). FIG. 3, 4, 5 and 6 show the energy dependence curves of the chamber in those qualities. The values were normalized for the highest energy qualities. The results were similar in most of the cases for the condition of use and without use of the preamplifier, but for the attenuated diagnostic beams the lowest energy dependence was obtained after the preamplifier removal. Probably these results are consequence of the interference of the manual zero adjustment of the preamplifier in the measures that were close to the inferior display limit of the electrometer. Table III. Calibration coefficients for a Victoreen pencil ionization chamber in IEC diagnostic radiology standard beams, with and without the use of its original preamplifier. Calibration Coefficients Radiation Quality With Preamplifier (Dimensionless) Without Preamplifier (x 10 8 Gy/C) RQR3 1.04 ± 0.05 1.03 ± 0.05 RQR5 1.06 ± 0.06 1.02 ± 0.05 RQR7 1.05 ± 0.05 1.04 ± 0.06 RQA3 1.03 ± 0.07 1.08 ± 0.06 RQA4 1.15 ± 0.06 1.11 ± 0.06 RQA5 1.20 ± 0.07 1.10 ± 0.06 RQA6 1.19 ± 0.06 1.11 ± 0.06 RQA7 1.20 ± 0.06 1.12 ± 0.06 Table IV. Calibration coefficients for a Victoreen pencil ionization chamber in mammography standard beams, with and without the use of its original preamplifier. Calibration Coefficients Radiation Quality With Preamplifier (Dimensionless) Without Preamplifier (x 10 8 Gy/C) RXM20 1.054 ± 0.036 1.060 ± 0.035 RXM23 1.039 ± 0.034 1.048 ± 0.035 RXM25 1.027 ± 0.034 1.044 ± 0.035 RXM28 1.025 ± 0.034 1.040 ± 0.034 RXM30 1.022 ± 0.034 1.037 ± 0.034 RXM32 1.017 ± 0.033 1.031 ± 0.034 RXM35 1.009 ± 0.033 1.027 ± 0.034 RXM20x 0.991 ± 0.038 1.013 ± 0.036 RXM23x 0.996 ± 0.040 1.010 ± 0.035 RXM25x 0.992 ± 0.034 1.006 ± 0.035 RXM28x 1.000 ± 0.034 1.008 ± 0.035 RXM30x 1.001 ± 0.035 1.010 ± 0.035 RXM32x 1.004 ± 0.034 1.014 ± 0.035 RXM35x 1.012 ±0.034 1.023 ± 0.038 5

Table V. Energy dependence of the pencil ionization chamber for several radiation qualities. Radiation Quality IEC diagnostic radiology qualities Direct beams IEC diagnostic radiology qualities Attenuated beams Mammography qualities Direct beams Mammography qualities Attenuated beams Energy Dependence With Preamplifier Without Preamplifier (%) (%) 1.9 2.0 16.5 3.7 4.5 3.2 2.1 1.7 FIG. 3. Energy dependence curve for the Victoreen pencil ionization chamber, before and after the preamplifier removal, in IEC diagnostic radiology qualities, direct beams. The calibration coefficients were normalized for the highest energy quality (46.0 kev). 6

FIG. 4. Energy dependence curve for the Victoreen pencil ionization chamber, before and after the preamplifier removal, in IEC diagnostic radiology qualities, attenuated beams. The calibration coefficients were normalized for the highest energy quality (62.9 kev). FIG. 5. Energy dependence curve for the Victoreen pencil ionization chamber, before and after the preamplifier removal, in mammography qualities, direct beams. The calibration coefficients were normalized for the highest energy quality (16.2 kev). 7

FIG. 6. Energy dependence curve for the Victoreen pencil ionization chamber, before and after the preamplifier removal, in mammography qualities, attenuated beams. The calibration coefficients were normalized for the highest energy quality (21.6 kev). The performance of the pencil ionization chamber with and without the preamplifier was similar and satisfactory, except when the air kerma rates were close to the inferior display limit of the electrometer used before the preamplifier removal. 4. Conclusions This study shows that it is possible to easily adapt a pencil ionization chamber, originally with a preamplifier, to be used with other kinds of electrometers than its original Victoreen electrometer. The results obtained before and after the preamplifier removal were similar in behavior, showing that the adaptation did not imply in lose if precision in the measurements. 5. Acknowledgements The authors wish to acknowledge the partial financial support of Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP) and Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq), Brazil. 6. References 1. Jucius, R.A. and Kambic, G.X. Radiation dosimetry in computed tomography (CT), SPIE Proc., 127:286-295, (1977). 2. Poletti, J.L. An ionization chamber based CT dosimetry system, Phys. Med. Biol., 29:725-731, (1984). 3. Pavlicek, W. Evaluation of the MDH Industries, Inc. pencil chamber for direct beam CT measurements, Health Phys., 37:773-774, (1979). 4. Bochud, F.O., Grecescu, M. and Valley, J.F. Calibration of ionization chambers in air kerma length, Phys. Med. Biol., 46:2477-2487, (2001). 8

5. Suzuki, A. and Suzuki, M.N. Use of a pencil-shaped ionization chamber for measurement of exposure resulting from a computed tomography scan, Med. Phys. 5(6):536-539, (1978). 6. Maia, A.F., Caldas, L.V.E. Performance of a pencil ionization chamber in various radiation beams, Appl. Radiat. Isot., 58:595-601 (2003). 7. International Electrotechnical Commission. Medical diagnostic X-ray equipment - Radiation conditions for use in determination of characteristics. IEC 61267. Genève (1994). 8. International Atomic Energy Agency. The use of plane parallel ionization chambers in high energy electron and photon beams. Technical Reports Series No. 381. Vienna (1997). 9. International Electrotechnical Commission. Medical electrical equipment - Dosimeters with ionization chamber and/or semi-conductor detectors as used in X-ray diagnostic imaging. IEC 61674. Genève (1997). 10. Meghzifene, A., Shortt, K.R. Calibration factor or calibration coefficient? SSDL Newsletter, 46:33-33, (2002). 9