Bureau International des Poids et Mesures

Similar documents
Bureau International des Poids et Mesures

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays

Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays

Degrees of equivalence for the key comparison BIPM.RI(I)-K3 between national primary standards for medium-energy x-rays

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the NRC, Canada and the BIPM in medium-energy x-rays

Bureau International des Poids et Mesures

PRELIMINARY CHARACTERIZATION OF THE NIS FREE-AIR CHAMBER STANDARD AT THE BIPM

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the LNE-LNHB, France and the BIPM in 60 Co gamma radiation

Comparison of the standards of air kerma of the GUM and the BIPM for 60Co 'Y rays

Comparison of the standards for absorbed dose to water of the ENEA-INMRI (Italy) and the BIPM for 60 Co γ rays

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the NIM, China and the BIPM in 60 Co gamma radiation

Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

Abstract An indirect comparison of the standards for reference air kerma rate for

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira

Comparison of the standards of air kerma of the BNM-LPRI and the BIPM for 137CS y rays

Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI

Air kerma rate measurements from a miniature x-ray source using free-air ionization chambers

A Correction Factor for Effects of Scattered X-rays at Calibration of Ionization Chambers in Low Energy X-ray Standard Fields

Comparisons of the standards for air kerma of the GUM and the BIPM for 60 Co and 137 Cs gamma radiation

First international comparison of primary absorbed dose to water standards in the medium-energy X-ray range

Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kv to 300 kv

Determination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1)

M [scale units/s] of the system

Comparison of air kerma measurements for tungsten anode based mammography x-ray beam qualities (EURAMET.RI(I)-S4.1)

N. E. Ipe*, K. E. Rosser, C. J. Moretti, J. W. Manning and M. J. Palmer

Comparison of the standards for absorbed dose to water of the VSL and the BIPM for 60 Co gamma rays

Investigation of the standard temperature- pressure correction factor at low x-ray energies

Chapiter VII: Ionization chamber

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden

4.1b - Cavity Theory Lecture 2 Peter R Al mond 2011 Overview of Lecture Exposure (W/e)air Exposure Exposure and and and Air Air Kerma

FINAL REPORT. European co-operation for Accreditation (EA) Interlaboratory Comparison IR3: Calibration of a Radiation Protection Dosimeter

BIPM Key Comparison BIPM.RI(I)-K6 PROTOCOL 3.0 CCRI(I) 6 November 2017

Comparisons between IRD and BIPM exposure and air-kerma standards. for cobalt-60 gamma rays. by C.E. de Almeida* and M.-T. Niatel

Summary of the ICRM Life Sciences Working Group Meeting

H Bjerke 1, J Plagnard 2, J-M Bordy 2, A Kosunen 3, J Huikari 3, L Persson 4, and P O Hetland 1

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

Basic physics Questions

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS

RECOMBINATION PARAMETERS OF SOME FABRICATED IONIZATION CHAMBERS

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

PRIMARY STANDARDS of AIR KERMA for 60 CO and X-RAYS & ABSORBED DOSE in PHOTON and ELECTRON BEAMS. Malcolm McEwen

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs

Updating reference dosimetry a decade after TG-51

2012 ANS WINTER MEETING NOVEMBER 11-15, 2012

Evaluation of Phantom Equivalent Materials in Polychromatic Diagnostic X-Ray Beam

Accelerators and radiation spectra

Bureau International des Poids et Mesures Comparison of ozone reference standards of the CHMI and the BIPM

Monte Carlo simulations as a tool for the development of a new reference ionisation chamber

Rad T 290 Worksheet 2

Non-radioactive radiation sources. Lesson FYSKJM4710 Eirik Malinen

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

Total radiation measurements of thermodynamic temperature

8/2/2012. Larry DeWerd has a partial interest in Standard Imaging

General characteristics of radiation dosimeters

Mass determination of 1 kg silicon spheres for Avogadro project

Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015

Ionizing Radiation Dosimetry and Medical Physics

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz

Metrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT)

Work Programme. of the International Bureau of Weights and Measures. for the four years Comité international des poids et mesures

Code of Practice for the Absorbed Dose Determination in High Energy Photon and Electron Beams

BUREAU INTERNATIONAL DES POIDS ET MESURES

Leakage Tests for Am-241 Solid Sources Used for Liquid Xenon Detector Monitoring

CALIBRATION OF DOSEMETERS USED IN MAMMOGRAPHY WITH DIFFERENT X RAY QUALITIES: EUROMET PROJECT NO. 526

Chapter V: Cavity theories

Radhakrishnan B*, Kurup P G G**, Ramakrishnan G***, Chandralingam S****

International comparison of extrapolation chamber measurements of the absorbed dose rate in tissue for beta radiation

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers

INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN

Comments on ICRU Report 64: Dosimetry of High-Energy Photon Beams based on Standards of Absorbed Dose to Water

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

Final Report August 2010

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Bureau International des Poids et Mesures

Reference Dosimetry for Megavoltage Therapy Beams: Electrons

Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3)

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu

Draft protocol of APMP.RI(I)-K8 Reference air kerma rate for HDR Ir-192 brachytherapy sources

In vacuum ID beam line shielding commissioning and direct gasbremsstrahlung measurements at Synchrotron SOLEIL

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Measurement Science and Standards. Ernesto Mainegra-Hing, Hong Shen, Malcolm McEwen Ionizing Radiation Standards Report PIRS-2350 June 2017

BUREAU INTERNATIONAL DES POIDS ET MESURES. Comparison of Quantum Hall Effect resistance standards of the PTB and the BIPM

X-ray Spectroscopy. Danny Bennett and Maeve Madigan. October 12, 2015

WATT BALANCES AND THE FUTURE OF THE KILOGRAM

Derivation of factors for estimating the scatter of diagnostic x-rays from walls and ceiling slabs

Ba (Z = 56) W (Z = 74) preferred target Mo (Z = 42) Pb (Z = 82) Pd (Z = 64)

Final Report on CIPM key comparison of multiples and submultiples of the kilogram (CCM.M-K2)

Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the PTB, Germany and the BIPM in low-energy x-rays

National Physical Laboratory (NPL), UK

The metrology of radioactivity. C. Michotte, BIPM

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required

Calibration of the GNU and HSREM neutron survey instruments

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

ACCURATE MEASUREMENTS OF SURFACE EMISSION RATE FOR LARGE-AREA ALPHA AND BETA REFERENCE SOURCES

Transcription:

Bureau International des Poids et Mesures Comparison of the air-kerma standards of the ENEA-INMRI and the BIPM in the low-energy x-ray range D.T. Burns, M.P. Toni and M. Bovi November 1999 Pavillon de Breteuil, F-92312 Sèvres Cedex

Comparison of the air-kerma standards of the ENEA-INMRI and the BIPM in the low-energy x-ray range by D.T. Burns, M.P. Toni* and M. Bovi* Bureau International des Poids et Mesures, F-92312 Sèvres Cedex *Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, C.R. Casaccia, c.p. 2400 Rome, Italy Abstract A direct comparison has been made between the air-kerma standards of the ENEA-INMRI and the BIPM in the low-energy x-ray range. The results show the standards to be in agreement to around 0.35 % at reference beam qualities up to 50 kv. Measurements were also made at the 80 kv quality in order to estimate the electron-loss correction for the ENEA standard at this quality. 1. Introduction A direct comparison has been made between the air-kerma standards of the Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti (ENEA-INMRI) and the Bureau International des Poids et Mesures (BIPM) in the x-ray range from 10 kv to 50 kv. The comparison took place at the BIPM in September 1998 using the reference conditions recommended by the CCRI [1]. Additional measurements were made at 80 kv to determine the electron-loss correction for the ENEA-INMRI standard at this quality. This required the measurement of aperture and wall transmission correction factors. 2. Determination of the air-kerma rate For a free-air ionization chamber standard with measuring volume V, the air-kerma rate is determined by the relation I W K& air 1 = k i (1) ρ airv e 1 g air i where I is the ionization current, ρ air is the density of air under reference conditions, W air is the mean energy expended by an electron of charge e to produce an ion pair in dry air, g air is the fraction of the initial electron energy lost by bremsstrahlung production in air, and Π k i is the product of the correction factors to be applied to the standard. The values used for the physical constants ρ air and W air /e are given in Table 1. For use with this value for ρ air, the ionization current I must be corrected for the difference between the density of the air of the measuring volume at the time of measurement and the value given in the table 1. 1 For an air temperature T and pressure P in the measuring volume, this involves a temperature correction T / T 0, a pressure correction P 0 / P, a humidity correction k h = 0.9980, and the factor 1.0002 to account for the change in the compressibility of air between T ~ 293 K and T 0 = 273.15 K. 1

3. Details of the standards Both free-air chamber standards used in the present comparison are of the conventional parallelplate design. The measuring volume V is defined by the diameter of the chamber aperture and the length of the collecting region. The BIPM air-kerma standard is described in [2] and [3]. Details of the ENEA-INMRI standard, which has not previously been compared with the BIPM standard, are given in [4]. The main dimensions, the measuring volume and the polarizing voltage for each chamber are shown in Table 2. Table 1. Physical constants used in the determination of the air-kerma rate Constant Value s 10 2 ρ air 1.293 0 kg m 3 0.01 W air / e 33.97 J C 1 0.15 s is the relative standard uncertainty. Density of dry air at T 0 = 273.15 K and P 0 = 101 325 Pa. Table 2. Main characteristics of the standards Chamber BIPM ENEA- INMRI Aperture diameter / mm 9.941 8.014 Air path length / mm 100.0 65.12 Collector length / mm 15.466 40.738 Electrode separation / mm 70 60 Collector width / mm 71 60 Measuring volume / mm 3 1 200.4 2 054.9 Polarizing voltage / V 1 500 1 600 This is the quoted value of 64.30 mm plus 0.82 mm due to three screws supporting the aperture. 3. Comparison procedure 3.1 BIPM irradiation facility and reference beam qualities The comparison was carried out in the BIPM low-energy x-ray laboratory, which houses a constant-potential generator (maximum usable generating potential 80 kv) and a tungsten-anode x-ray tube with an inherent filtration of 2.9 mm beryllium. Both the generating potential and the tube current are stabilized using feedback systems constructed at the BIPM; this results in a very high stability and obviates the need for a transmission current monitor. The variation in the measured ionization current over the duration of a comparison introduces a relative standard uncertainty of typically 4 10 4. The radiation qualities used in the range from 10 kv to 50 kv are those recommended by the CCRI [1] and are given in Table 3 in ascending half-value layer (HVL) from left to right. Measurements were also made at the 80 kv quality indicated in the table. 2

3.2 Correction factors The correction factors applied to the ionization current measured at each radiation quality, together with their associated uncertainties, are given in Table 4 for the BIPM standard and in Table 5 for the ENEA-INMRI standard (the correction factors for the ENEA-INMRI standard at the 80 kv quality are discussed in Section 4.3). Table 3. Characteristics of the BIPM reference radiation qualities Generating potential / kv 10 30 25 50(b) 50(a) 80 Al filtration / mm 0 0.208 2 0.372 3 1.008 2 3.989 3.041 Al HVL / mm 0.036 0.176 0.250 1.020 2.257 3.01 µ air / 10 3 mm 1 1.757 0.415 0.304 0.091 0.046 0.042 K & BIPM / mgy s 1 0.56 3.31 1.13 1.57 0.34 0.61 Air attenuation coefficient at 293.15 K and 100 000 Pa, measured at the BIPM for an air path length of 100 mm. Table 4. Correction factors for the BIPM standard Generating potential / kv 10 30 25 50(b) 50(a) 80 s 10 2 type A type B Air attenuation k a 1.192 1 1.042 4 1.030 9 1.009 1 1.004 6 1.004 2 0.03 0.01 Scattered radiation k sc 0.994 4 0.995 6 0.995 7 0.996 6 0.997 1 0.997 4-0.07 Electron loss k e 1.000 0 1.000 0 1.000 0 1.000 0 1.000 0 1.010 0-0.01 Ion recombination k s 1.000 7 1.001 9 1.001 0 1.001 1 1.000 6 1.000 6 0.02 0.01 Field distortion k d 1.000 0 1.000 0 1.000 0 1.000 0 1.000 0 1.000 0-0.07 Aperture edge transmission k l 1.000 0 1.000 0 1.000 0 1.000 0 1.000 0 0.999 7-0.01 Wall transmission k p 1.000 0 1.000 0 1.000 0 1.000 0 1.000 0 1.000 0 0.01 - Humidity k h 0.998 0 0.998 0 0.998 0 0.998 0 0.998 0 0.998 0-0.03 1 g air 1.000 0 1.000 0 1.000 0 1.000 0 1.000 0 1.000 0-0.01 These are nominal values for 293.15 K and 100 000 Pa; each measurement is corrected using the air temperature and pressure measured at the time. For the 80 kv quality, the value for s 10 2 is 0.10. The largest correction at low energies is that due to the attenuation of the x-ray fluence along the air path between the reference plane and the centre of the collecting volume. The correction factor k a is calculated using the measured air-attenuation coefficients µ air given in Table 3 (in units of inverse length). In practice, the values used deviate slightly from those given in the tables. This is because the attenuation varies with the temperature and pressure of the air in the 3

chamber and the values for k a used are corrected for this effect. The value for k a for the ENEA- INMRI chamber at 10 kv has been increased by the factor 1.000 9 to account for the larger mean air-attenuation coefficient for an air path length of 65.12 mm (the values given in Table 3 were measured at the BIPM for an air path length of 100 mm). This effect is negligible at the other radiation qualities. All ionization measurements are also corrected for changes in air attenuation arising from variations in the temperature and pressure of the ambient air between the radiation source and the reference plane, using the air-attenuation coefficients given in Table 3. All measured ionization currents are corrected for ion recombination. The measured values for the ion recombination correction k s for the BIPM standard are given in Table 4. For the ENEA- INMRI standard, the values for k s given in Table 5 were calculated for the BIPM air-kerma rates given in Table 3 using [5]. These values were checked by performing the same calculation for the air-kerma rates at the ENEA-INMRI; these latter values agree within 0.01 % with the values measured at the ENEA-INMRI 2. Table 5. Correction factors for the ENEA-INMRI standard Generating potential / kv 10 30 25 50(b) 50(a) type A s 10 2 type B Air attenuation k a 1.122 2 1.027 4 1.020 0 1.005 9 1.003 0 0.03 0.01 Scattered radiation k sc 0.994 3 0.995 0 0.994 7 0.996 7 0.997 1-0.1 Electron loss k e 1.000 1.000 1.000 1.000 1.000-0.1 Ion recombination k s 1.000 5 1.000 9 1.000 6 1.000 7 1.000 5-0.05 Field distortion k d 1.000 1.000 1.000 1.000 1.000-0.1 Aperture edge transmission k l 1.000 0 1.000 0 1.000 0 1.000 0 1.000 0-0.05 Wall transmission k p 1.000 0 1.000 0 1.000 0 1.000 0 1.000 0-0.05 Humidity k h 0.998 0 0.998 0 0.998 0 0.998 0 0.998 0-0.03 1 g air 1.000 0 1.000 0 1.000 0 1.000 0 1.000 0-0.01 These are nominal values for 293.15 K and 100 000 Pa; each measurement is corrected using the air temperature and pressure measured at the time. This value is derived from the uncertainty of µ air at the BIPM. The value for s 10 2 used at the ENEA- INMRI is 0.2. 3.3 Chamber positioning and measurement procedure The ENEA-INMRI chamber was positioned close to the BIPM chamber and both remained fixed throughout the comparison; the alternation of measurements between chambers was carried out by displacement of the radiation source. Alignment on the beam axis was measured to around 0.2 mm and this position was reproducible to better than 0.01 mm. An off-axis displacement of 0.1 mm changes the measured current by no more than 0.03 % at 10 kv and at 50 kv. No 2 Except for the 50 kv(a) quality, for which the calculated value is 1.0005 and the measured value 1.0010. The calculated value is used in this report. 4

correction is applied for the radial non-uniformity of the beam. The reference plane for each chamber was positioned at 500 mm from the radiation source for all qualities up to 50 kv and at 750 mm for the 80 kv quality. This distance was measured to 0.03 mm and was reproducible to better than 0.01 mm. The beam diameter in the reference plane is 45 mm for all qualities up to 50 kv and 68 mm for the 80 kv quality. The air temperature for the ENEA-INMRI chamber was measured using a BIPM mercury thermometer calibrated to 0.02 K and positioned in the holder of the ENEA-INMRI chamber. The leakage current was measured before and after each series of ionization current measurements and a correction made based on the mean of these leakage measurements. For the BIPM chamber the leakage current was less than 0.05 % of the ionization current and for the ENEA-INMRI chamber less than 0.01 %. Each series consisted of five measurements. The relative standard uncertainty of the mean of a series of five measurements was typically less than 2 10 4. Taking into account the relative standard uncertainty of 4 10 4 arising from the repeatability over the duration of a comparison of current measurements using the ENEA- INMRI standard, a type A relative standard uncertainty of 5 10 4 is taken for current measurements using this chamber. For both chambers, measurements were made at both polarities to correct for any polarity effect. The measured difference was typically 0.11 % for the BIPM chamber and less than 0.02 % for the ENEA-INMRI chamber. 4. Supporting measurements 4.1 Comparison of apertures The ENEA-INMRI aperture of diameter 8.014 mm was positioned in the BIPM chamber, replacing the BIPM aperture of diameter 9.941 mm. The current measured at 30 kv by the BIPM standard under these conditions was corrected for the decrease in aperture diameter and for the decrease in ion recombination using [5]. The resulting air-kerma rate determination was 0.09 % less than that determined using the BIPM aperture (measured with a relative standard uncertainty of 2 10 4 ). 4.2 Comparison of methods for measuring air attenuation The air-attenuation correction for each standard was determined using the air-attenuation coefficients µ air measured at the BIPM, as given in Table 3. These are measured using a tube of length 270 mm positioned approximately midway between the added filters and the reference plane. By reducing the air pressure in the tube to approximately 64 kpa and measuring the decrease in the ionization current, µ air is determined for an air path length of 100 mm. For the 10 kv radiation quality, additional measurements are made over a range of air pressures to determine µ air for other air path lengths, from which the correction factor 1.0009 for the ENEA- INMRI air path length of 65.12 mm is derived. Note that the thin beryllium windows of the tube are included in the stated inherent filtration (2.9 mm beryllium). The ENEA-INMRI method for determining µ air is by the addition of an aluminium tube of internal diameter 25 mm to the front of the chamber, which moves the aperture towards the radiation source by a known distance. The source must be moved back by this distance. The decrease in the ionization current is a measure of the air attenuation arising from the additional air path between the aperture and the collecting volume. Two sets of measurements were performed at 10 kv. In the first set, a tube of length 50.4 mm was added to the ENEA-INMRI standard and the radiation source moved back by this amount. From the measured decrease in current, the air-attenuation coefficient was evaluated as µ air = 1.706 10 3 mm 1 (at 293.15 K and 100 kpa). For the second set of measurements a tube of length 95.0 mm was used, giving µ air = 1.687 10 3 mm 1. From these results one can 5

6 Rapport BIPM-99/11 interpolate (linearly) the value µ air = 1.700 10 3 mm 1 for an air path length of 65 mm. Using this value rather than the BIPM value yields an air-kerma rate determination which is smaller by 0.37 % for the ENEA-INMRI standard and by 0.57 % for the BIPM standard. 4.3 Measurements at 80 kv Although the ENEA-INMRI standard is not designed for use above 50 kv, the opportunity was taken to compare the two standards at 80 kv. This allows an experimental determination of the electron-loss correction k e for the ENEA-INMRI standard at this quality, on the assumption that all other correction factors are known. Those for the BIPM standard are given in Table 4. The value for k e is derived from measurements at the BIPM using the magnetic field technique and has been verified by comparison with the BIPM medium-energy free-air chamber standard and by recent Monte Carlo calculations [6]. For the ENEA-INMRI standard, k sc was taken to be the same as for the BIPM standard (their values at 50 kv are very similar), k s was calculated using [5], k d was taken to be the same as at lower energies and 1 g air was taken to be unity. This leaves only the corrections for aperture-edge transmission k l and wall transmission k p for the ENEA-INMRI standard, which were measured. Aperture-edge transmission was measured using a dummy aperture of the same material and thickness as the real aperture but without the hole. At 80 kv no transmission was measurable at the 0.02 % level. Wall transmission was measured using a large lead stopper. At 80 kv the walltransmission correction k p = 0.995 2 was determined with an estimated relative standard uncertainty of 0.02 %. 5. Uncertainties The uncertainties associated with the primary standards and with the results of the comparison are listed in Table 6. In general, the quoted uncertainties are representative of those associated with routine air-kerma rate determinations at both institutions. The uncertainty of µ air is that of the BIPM determination and is significantly less than that quoted for determinations at the ENEA-INMRI (see Table 5 and footnote). The uncertainties associated with the measurement of the ionization current and with chamber positioning are those which apply to measurements at the BIPM. The uncertainties of the ratios K & ENEA-INMRI K& BIPM take into account correlations in the type B uncertainties associated with the determination of the ionization current, the humidity correction and the physical constants. Correlations between the values for k sc are not taken into account, although these are derived from the same basic data. 6. Results and discussion The comparison results are given in Table 7. General agreement at the level of 0.3 % to 0.4 % is observed, which is approximately 1.5 times the relative standard uncertainty. The result of the aperture comparison accounts for almost 0.1 % of the difference and so the standards can be considered to be in acceptable agreement. No significant trend with radiation quality is observed The result of the comparison at 80 kv is K & ENEA -INMRI K& BIPM = 0.969 1, with a relative standard uncertainty of 0.27 % (including the 0.10 % uncertainty of k e for the BIPM standard). Comparing this with the mean result obtained in the range from 10 kv to 50 kv, one can deduce the value k e = 1.028 for the ENEA-INMRI standard at this quality. Using this method, many correlated uncertainties cancel and the relative standard uncertainty of this value for k e is estimated to be 0.12 %. This result will be used for comparison with Monte Carlo calculations. Of particular concern is the difference in the BIPM and the ENEA-INMRI determinations of the air-attenuation coefficient µ air for the 10 kv quality at the BIPM. This difference is very much

larger than the estimated uncertainty of the measurement techniques. Using the ENEA-INMRI value yields an air-kerma rate determination which is smaller by 0.37 % for the ENEA-INMRI standard. Around 0.1 % of this difference may be explained by a change in the photon-scatter correction k sc when using the ENEA-INMRI technique, which is not taken into account. The use of the ENEA-INMRI value for µ air for the BIPM standard at 10 kv yields an air-kerma rate determination which is smaller by 0.57 %. The net effect of these changes is an increase in the comparison result at 10 kv by 0.20 % which worsens the consistency of the results as a function of beam quality. From this evidence one might deduce that the BIPM determination of µ air is the closer estimate. A summary of the results of BIPM comparisons of air-kerma standards for low-energy x-rays, including the present comparison, is presented in Annex A. Table 6. Uncertainties associated with the comparison results Standard BIPM ENEA-INMRI s 10 2 type A type B type A type B Ionization current 0.02 0.02 0.05 0.02 Volume 0.03 0.05-0.05 Positioning 0.01 0.01 0.01 0.06 Correction factors (excl. k h ) 0.04 0.10 0.03 0.19 Humidity k h - 0.03-0.03 Physical constants - 0.15-0.15 0.06 0.19 0.06 0.26 K & Standard 0.20 0.27 K & K& 0.25 ENEA-INMRI BIPM Takes account of correlations in Type B uncertainties. Table 7. Comparison results Generating potential / kv 10 30 25 50(b) 50(a) K & K& 0.9972 0.9958 0.9956 0.9966 0.9966 ENEA-INMRI BIPM 7

References [1] BIPM, Qualités de rayonnements, CCEMRI (Section I), 1972, 2, R15. [2] BOUTILLON M., HENRY W.H. and LAMPERTI P.J., Comparison of exposure standards in the 10-50 kv x-ray region, Metrologia, 1969, 5, 1-11. [3] BOUTILLON M., Measurement conditions used for the calibration of ionization chambers at the BIPM, 1996, Rapport BIPM-96/1. [4] MORETTI C.J., HEATON J.A., LAITANO R.F. AND TONI M.P., Intercomparison of low-energy primary standards for x-ray exposure of NPL and ENEA, 1991, NPL Report RSA (EXT) 19. [5] BOUTILLON M., Volume recombination parameter in ionization chambers, Physics in Medicine and Biology, 1998, 43, 2061-2072. [6] BURNS D.T., Consistent set of calculated values for electron-loss and photon-scatter corrections for parallel-plate free-air chambers, 1999, CCRI(I)/99-4, 14 p. 8

Annex A The results of BIPM comparisons of air-kerma standards for low-energy x-rays are presented in Table A.1. For laboratories which have compared more than once at the BIPM, only the results of the most recent comparison are included. The same data are presented in graphical form in Figure A.1. Table A.1 Results of BIPM low-energy x-ray comparisons, expressed as K. & & NMI K BIPM NMI Country Date Generating potential / kv 10 kv 30 kv 25 kv 50 kv(b) 50 kv(a) NRC Canada 1966 1.0007 1.0003 - - 0.9995 ETL Japan 1972 0.9963 0.9963 - - 1.0032 CIEMAT Spain 1979 1.0021 1.0011 1.0013 1.0018 1.0025 OMH Hungary 1988 0.9973-0.9994 1.0010 1.0020 GUM Poland 1994 0.9963 0.9973-0.9968 0.9977 NMi Netherlands 1996 0.9972 0.9984-0.9984 0.9963 NPL UK 1997 0.9983 0.9980 0.9995-0.9977 NIST USA 1998 0.9950 0.9943 0.9949 0.9938 0.9956 OFMET Switzerland 1998 0.9994 0.9993 0.9994 0.9984 0.9985 ENEA Italy 1998 0.9972 0.9958 0.9956 0.9966 0.9966 The results for this laboratory are provisional; BIPM report still in preparation. 9

1.015 1.010 1.005 10 kv 30 kv 25 kv 50 kv(b) 50 kv(a) CIEMAT ENEA ETL GUM NIST NMi NPL NRC OFMET OMH K NMI / K BIPM 1.000 0.995 0.990 0.985 Figure A.1. Results of BIPM low-energy x-ray comparisons, expressed as the ratio of the air-kerma rate determined by the NMI standard to that determined by the BIPM standard. The results for the NPL are provisional. 10