Activity in the FID Detection Port: A Big Problem if Underestimated

Similar documents
An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

Gas Chromatography. Introduction

Trajan SGE GC Columns

Introduction and Principles of Gas Chromatography

Understanding Gas Chromatography

Get Selective. By Jaap de Zeeuw

A New Web-Based Application for Modeling Gas Chromatographic Separations

Ultra-Inert chemistry for Trace Level Analysis

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

Introduction to Capillary GC

Chapter 27: Gas Chromatography

CHAPTER 6 GAS CHROMATOGRAPHY

Chemistry Gas Chromatography: Separation of Volatile Organics

Chromatography. Gas Chromatography

2401 Gas (liquid) Chromatography

Gas Chromatography (GC)

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Secrets of GC Column Dimensions

LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT. Thomas Wenzel, Bates College. In-class Problem Set Extraction.

Introduction to Capillary GC

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

1.0 Reactive Gas Chromatography Design and Setup Procedure for Catalyst Pelletization & Inlet Packing Software and Method.

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

A New Web-Based Application for Modeling Gas Chromatographic Separations. Dr. Hansjoerg Majer Restek Corporation Market Development Manager Europe

Analysis of Sulfur-Containing Flavor Compounds By GC/MS With A PFPD

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

Selection of a Capillary GC Column

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Capillary GC Column Selection and Method Development A Primer on Column Parameters and Instrument Conditions

High Performance Liquid Chromatography

Column Dimensions. GC Columns and Consumables. Mark Sinnott Application Engineer. March 12, 2010

Uncontrolled Copy. SOP-060 Chromatography, Tune Evaluation and Troubleshooting. Table of Contents. 1. Principle... 3

Selection of a Capillary

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Column Selection. there is more to life than a boiling point column. Jaap de Zeeuw Restek Corporation, The Netherlands. Copyrights: Restek Corporation

Introduction to Gas Chromatography

Water Injections in GC - Does Water Cause Bleed?

Theory and Instrumentation of GC. Sample Introduction

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Instrumental Chemical Analysis

Headspace Technology for GC and GC/MS: Features, benefits & applications

Hydrogen as a Carrier Gas in the Analysis of Polar/Non-Polar Compounds Using the Polyarc System. Application Note. Author. Introduction.

Chapter 1. Chromatography. Abdul Muttaleb Jaber

Gas Chromatography. Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid

GAS CHROMATOGRAPHY (GC)

Chapter 31 Gas Chromatography. Carrier Gas System

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Chapter 11 Conventional Gas Chromatography

Virtually Particle-Free Rt -Silica BOND Columns

environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)

COOKBOOK FOR THE PERKINELMER SWAFER TM TECHNOLOGY FOR GC APPLICATIONS

Rtx-BAC Plus 1 and Rtx-BAC Plus 2 Columns Advanced Technology for Fast, Reliable Measurement of Alcohol in Blood

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

NEW Rtx -BAC Plus 1 and Rtx -BAC Plus 2 Columns

GC Instruments. GC Instruments - Sample Introduction

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Evaluation of Capillary Columns for General Performance Parameters

What is Chromatography?

Determination of Volatile Substances Proof of Food Adulteration

Lab 3 Guide: Gas Chromatography (GC) (Sept 8-14)

GC Instruments. GC Instruments - Columns

GAS CHROMATOGRAPHY. Mobile phase is a gas! Stationary phase could be anything but a gas

METHANOLYSIS OF ACETAL

AppNote 6/2004. Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons SUMMARY

Partitioning. Separation is based on the analyte s relative solubility between two liquid phases or a liquid and solid.

GC and GCMS Fundamentals of Instrument Troubleshooting and Maintenance. Kirk Lokits Ph.D. Applications Chemist

Chromatography and Functional Group Analysis

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

Disadvantage: Destructive Technique once analyzed by GC, the sample is lost

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Volatile organic compounds (VOCs):

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Gas Chromatography (Chapter 2 and 3 in The essence of chromatography)

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

Fundamentals of GC: Introduction: Sample introduction:

Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary.

Agilent G3212 GC-APCI Source

Dynamic headspace (DHS) technique: set-up and parameter control for GC/MS analysis of odorant formulations

Paints and varnishes Determination of volatile organic compound (VOC) content. Part 2: Gas-chromatographic method

Highly Sensitive and Rugged GC/MS/MS Tool

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY

Gas Chromatography notes

Chromatography. Chromatography is a combination of two words; * Chromo Meaning color * Graphy representation of something on paper (writing)

HPLC Background Chem 250 F 2008 Page 1 of 24

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

7 INSTRUMENTAL CHROMATOGRAPHY

Transcription:

Activity in the FID Detection Port: A Big Problem if Underestimated Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands It is commonly known in gas chromatography, that many problems can be traced to the injection system (e.g. sample, syringe, inlet), which is often a primary place to look at possible issues. This is a valid statement, as 80% of trouble is related to injection conditions. One must also be aware that activity may be caused by other contributions. For instance, if we look at non-symmetrical peaks, there are more important area s to look at. Not only the columns used can be overloaded or poorly deactivated, also the contribution of the detector has a huge impact on peak shape and response.here it is shown that flame tips can adsorb up to 90% of component and cause tailing on polar as well as base/acidic components. Though it looks columns are not performing, it s the detector that is the problem. Introduction There is an ongoing need to measure lower levels of components in different kind of matrices. Often, troublesome components are more polar in nature, which challenges the analysis further, especially if lower concentrations are to be reported. In addition, in most labs, the user is given less time to optimize systems. Chromatography knowledge is fading and this is becoming an increasing challenge. It s all about eluting and measuring a correct chromatographic peak. Such a peak should look like a Gaussian curve. Any deformation from this curve is caused by an anomaly in the sampling and separation processes. The peak shape for a component in a typical GC can be affected in three ways: Injection and focusing Separation Detection Injection and focusing As stated before, 80% of problems Figure 1 Figure 1: Impact of column activity on peak height.

Figure 2 Figure 3 Figure 2: Impact of column activity on retention time of a polar analyte on columns with different deactivations. Figure 3: A 30 m x 0.25 mm Rxi-5Sil MS column installed in correct position in the detector. Peaks are symmetrical and show correct response. in GC are caused by injection issues. Users need to understand the processes occurring in the injection port. These processes depend from the technique used. In splitted injection, the sample evaporates quickly and injection, evaporation, homogenization and splitting must be performed in a fraction of a second. Any interaction with liners, glass wool or residual septum particles will slow down the injection and peak shape will be affected. Split liners usually are narrow to get high speed and optimal splitting conditions. Because the injection is performed as a narrow band, a focusing effect is not required. This is different when using the splitless, cold on-column or large volume injection techniques. Concentrations are lower and as the injection takes a longer time to get the components onto the column, there is also more time to interact. Separation The separation pathway is very long, depending on the column used. Within the column, any residual active sites will deform the ideal peak shape. Figure 1 shows what can happen when activity is present. Peaks start to become lower in height (lower sensitivity) and when they look like shark-fins we can expect impact on retention also. The retention time for a strong tailing peak depends on two mechanisms: 1. Gas liquid interaction, which is responsible for the normal retention. 2. Activity interaction, which is a different retention mechanism, based on adsorption. The degree of activity makes a peak elute at the wrong position. An example is shown in Figure 2 where the same amount of alcohol is injected on different PONA columns. On a poorly deactivated surface, the alcohols elute at completely different retention times, due to activity. Only a proper deactivation makes sure alcohols elute at similar expected positions. Detection The purpose of a detector is to detect the components. Usually the detector is set 10-20 ºC above the highest oven-temperature we use to elute the analytes from the column.

Sometimes detector temperatures are set above the maximum temperature of the stationary phase. For instance, polyethylene glycols will not be suitable at a temperature above 270 º C: they will decompose rapidly forming activity. One must be very careful with that especially after reading the next paragraphs. How important is correct installation at the detection port? In the detector, there is always some distance between the end of the column and the actual detection. What can happen here? Looking in the literature there is very little Figure 4 experimentation to find. It seems that activity at the end of the GC system is often neglected. Here we discuss the detector, but the same is also true for transfer lines, valves, connectors, flow-switching devices etc. Columns and transfer lines can develop high activity, just in standby position. This happens with any detector, where the carrier pressure is stopped and the detector gas-flows and temperatures are still high. For a FID, this is a big challenge and reason for doing some testing. What happens inside FID liner if you Figure 5 Figure 5: Column ends 4 cm below top of flame-tip. Response of polar compounds diminishes even furhter. Figure 4: Positioning the column outlet 1 cm lower in the 6890 FID, a big difference in response is observed. Also peak tailing starts to develop. change column or when the pressure is stopped for some reason? Officially the detector should be cooled down when doing maintenance, and gases must be switched off. But is that always done in a routine laboratory? The FID usually is at high temperature, so if the column flow disappears, or the detector connection is opened, we can expect water/oxygen to enter this area. When air is on, there is always a little overpressure on top of the detector that will drive air into the jet. This is an ideal scenario to develop activity. If a GC column is tested for inertness, it is a valid concern to be aware of the discussion above. Columns are tested with nano-grams of highly polar, base/acid analytes to verify performance. Experimental A series of experiments was initiated to demonstrate if there is a detector contribution. A Rxi-5Sil MS column was installed in an Agilent 6890N GC and operated under normal column performance test conditions as listed below: GC: Agilent 6890N Column: Rxi-5Sil MS, 30m x 0.25mm d f = 0.25μm

Figure 6 1-decylamine and undecanol. No column would pass QA if it showed this behaviour. The last 10 mm of the flame tip in the 6890 was highly activated causing huge adsorption and loss of many components. To maximize impact of the detector the column was placed 4 cm lower in the detector. This would not be very realistic, but it provides a good example on what happens in the last 40 mm of transport. The problems increased significantly, (see Figure 5). As detectors are also used at lower temperatures, the temperature was lowered to 250 º C. Now we see almost a loss of 85-90% response Figure 7 (see Figure 6). The next step was to reinstall the column to the correct position. Figure 7 shows the chromatogram. The response was restored completely, demonstrating again the importance of correct position of the column outlet during installation. Other systems We also checked an old HP-5890 system we use in our lab. Here another column was installed at different positions in the detector. Remarkably in this intance the position of the column in the detector did not make any difference Figure 6: FID- detector cooled down from 300 to 250 ºC. Significant loss of response for polar compounds, caused by increased activity at lower temperature. Oven: 135 º C Carrier gas : He, 22.9 psi, constant flow Injection: split, 200 ml/min, 0.6 μl Temp: 250 º C Sample: test mixture for Rxi column performance testing Detector: FID, temp. 300 º C The column was installed according to the manual: the column end positioned 1-2 mm from the top of the flame tip. We analysed the test mixture and obtained the chromatogram shown in Figure 3. The chromatogram looked as expected, with good response and good peak shapes. This became our baseline chromatogram. Now we cooled the oven to 40 º C, slightly opened the detector connection and pulled the column back proximate 10 mm. The connection was sealed again and oven temperature was restored. Figure 4 showed the impact of the 10 mm lower positioning of the column outlet in the FID. The last 10 mm of the FID caused a reduction in response for the polar analytes of almost a factor of 2. Note the response relative to peak 5 (tridecane). Also, significant tailing develope on the polar compounds 2,6-hexane diol, 4-chlorophenol, Figure 7: Column re-positioned near flame tip when detector was at 250 ºC. System performance is completely restored.

Figure 8 Figure 9 Comparison: correct installation and 5 cm below flame tip in old HP-5890 Figure 9: Example of SS and Siltek deactivated flame tips for agilent FID. Figure 8: Standard test in HP-5890 GC using correct (black) and challenging (red) positions for column outlet. System response is perfect. at any or temperature tested. Even when the column was cut 1 cm above the ferrule, the resulting chromatograms were very similar, see (Figure 8). The only difference with the newer systems, is the length of the FID-jet. The newer 6890 FIDs have a much shorter jet. However, this cannot explain the big differences observed. The activity /reactivity of the flame tip in FID always depend on the moment and can occur in any brand of GC. One can replace the flame tip for special deactivated ones (there are Siltek deactivated tips available, see Figure 9), but activity in detection systems can also often be masked by conditioning columns at the upper temperature as some bleed products are formed. These bleed products can deactivate the detector, so it will perform OK. This conditioningdeactivation is a known method for deactivating transfer lines and traps for example, in SCLD and MS systems. Also one can assume when a GC is used with adsorption columns (alumina, silica, porous polymers, charcoal or molecular sieves), it can be possible that some dust particles will be retained in the flame tip. Such particles are highly active. Conclusion For good GC we need a fully inert gas path. Activity can be present not only in the injector and the column, but also in the detector. For many it may seem trivial to install columns the correct way; however, practically the importance of the position of column in a FID detector is underestimated. If detection ports are well deactivated, one will see no measurable impact. In contrast, if there is activity developing, only a few millimeters of contact with the sample gas-path may lead to loss of components. The use of inert liners and inert columns, but for best results, one needs to be aware why columns are to be installed in the certain positions in injection and detection systems. References 1] J. de Zeeuw, R.Lautamo, G.Stidsen, C.English and S,Reese, American Lab, Vol. 40, nr.12, p.11-15. This article was written by Jaap de Zeeuw. Jaap is a GC specialist working for Restek.