On the Calibration of Alpha Sciences Proportional Counters

Similar documents
The need for a large-area low emissivity alpha particle standard

Quality Assurance. Purity control. Polycrystalline Ingots

Alpha-energies of different sources with Multi Channel Analyzer (Item No.: P )

Alpha-Energies of different sources with Multi Channel Analyzer

Lab NUC. Determination of Half-Life with a Geiger-Müller Counter

Modern Physics Laboratory (Physics 6180/7180)

Physics 1000 Half Life Lab

Alpha-particle Stopping Powers in Air and Argon

Radiation (Particle) Detection and Measurement

Energy loss of alpha particles - Prelab questions

Gamma Spectroscopy. References: Objectives:

Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University

DUNPL Preliminary Energy Calibration for Proton Detection

ACCURATE MEASUREMENTS OF SURFACE EMISSION RATE FOR LARGE-AREA ALPHA AND BETA REFERENCE SOURCES

Scintillation Detector

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

THE COMPTON EFFECT Last Revised: January 5, 2007

Radioactivity. Lecture 6 Detectors and Instrumentation

General Overview of Gas Filled Detectors

RANGE OF ALPHAS. Advanced Laboratory, Physics 407 University of Wisconsin Madison, Wisconsin 53706

3. Perturbed Angular Correlation Spectroscopy

He-3 Neutron Detectors

Energy calibration of the threshold of Medipix for ATLAS

Gamma and X-Ray Detection

Absorption and Backscattering ofβrays

High Purity Germanium Detector Calibration at ISOLDE

MEASURING THE LIFETIME OF THE MUON

Efficiency and Attenuation in CdTe Detectors

1. Introduction. Etim I.P. 1, William E. S. 2, Ekwe S.O. 3. Licensed Under Creative Commons Attribution CC BY

Determination of the shielding power of different materials against gamma radiation

DETERMINING FUNDAMENTAL PARAMETERS OF A SINGLE-PHOTOTUBE LIQUID SCINTILLATION COUNTER

Statistics of Radioactive Decay

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

CHAPTER 12 TEST REVIEW

MEASUREMENT OF THE D-D NEUTRON GENERATION RATE BY PROTON COUNTING

Physics 248, Spring 2009 Lab 6: Radiation and its Interaction with Matter

EXAMINATION QUESTIONS (6)

MEASUREMENT AND DETECTION OF RADIATION

PHYS 3446 Lecture #12

EEE4106Z Radiation Interactions & Detection

RADIOACTIVITY IN THE AIR

Nuclear Physics Laboratory. Gamma spectroscopy with scintillation detectors. M. Makek Faculty of Science Department of Physics

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

Introduction to Radiation Monitoring

Activities at the Laboratory of the Nuclear Engineering Department of the Polytechnic University of Valencia

Chapter 4 Scintillation Detectors

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440

Development of a Silicon PIN Diode X-Ray Detector

Joint Institute for Nuclear Research. Flerov Laboratory of Nuclear Reactions. University Center LABORATORY WORK

Radioactivity. PC1144 Physics IV. 1 Objectives. 2 Equipment List. 3 Theory

9/27 JUNE 2003 SUMMER STAGE PARTICLES REVELATION THROUGH CERENKOV AND SCINTILLATION COUNTER AND THE CEBAF EXPERIMENT

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

A Study of Radioactivity and Determination of Half-Life

HEROICA: a test facility for the characterization of BEGe detectors for the GERDA experiment

Proportional Counters

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

Absorption and Backscattering of β-rays

Chem 481 Lecture Material 3/20/09

Radiometric Measurements of Environmental Radioactivity

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Diffraction: spreading of waves around obstacles (EM waves, matter, or sound) Interference: the interaction of waves

Radiation Detectors. How do we detect ionizing radiation? What are these effects? Types of Ionizing Radiation Detectors

Detection and measurement of gamma-radiation by gammaspectroscopy

BEGe Detector studies update

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler

Experimental Particle Physics

1 of :32

E. K. A. ADVANCED PHYSICS LABORATORY STATISTICS OF COUNTING WITH A GEIGER COUNTER ARTIFICIAL RADIOACTIVITY

ORTEC Experiment 24. Measurements in Health Physics. Equipment Required

LAAPD Performance Measurements in Liquid Xenon

GAMMA RAY SPECTROSCOPY

Received October 7, 1963

THE GEIGER-MULLER TUBE AND THE STATISTICS OF RADIOACTIVITY

High-Resolution Gamma-Ray and Neutron Detectors For Nuclear Spectroscopy

PARTICLES REVELATION THROUGH SCINTILLATION COUNTER

A proposal to study gas gain fluctuations in Micromegas detectors

Mass of the electron m 0

Chapter 30 Nuclear Physics and Radioactivity

Radioactivity. (b) Fig shows two samples of the same radioactive substance. The substance emits β-particles. Fig. 12.1

Looking at Cosmic Muons to verify Einstein's Special Relativity

Comparison of the Photo-peak Efficiencies between the Experimental Data of 137 Cs Radioactive Source with Monte Carlo (MC) Simulation Data

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures:

Analysis of γ spectrum

CRaTER Science Requirements

PANDA-?? A New Detector for Dark Matter Search

Application Note. Understanding Performance Specifications for Low Background Alpha Beta Counters. FOM What Is It and Is It Useful?

Quenchings in crystals!

Beta Spectroscopy. Glenn F. Knoll Radiation Detection and Measurements, John Wiley & Sons, Inc. 2000

Radioactivity APPARATUS INTRODUCTION PROCEDURE

Absolute activity measurement of 85 Kr with proportional counters

Resolution. σ z. = σ z. If D(z) is gaussian the relation between FWHM and standard deviation is. = Δz 2.36

The Quantum Mechanics of Alpha Decay

Neutron Spectrometer Based on a Proton Telescope with Electronic Collimation of Recoil Protons

Effect of the time characteristics of the Compton camera on its performance

BEAM BEAM. cathode anode cathode

Experiment Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado

Preview from Notesale.co.uk Page 4 of 35

LAB 4: Gamma-ray coincidence spectrometry (2018)

RITU and the GREAT Spectrometer

This paper should be understood as an extended version of a talk given at the

Transcription:

IBM Research On the Calibration of Alpha Sciences Proportional Counters Michael Gordon (gordonm@us.ibm.com) Ken Rodbell IBM Research 1101 Kitchawan Road Yorktown Heights, NY 10598 10/26/2011 2011 IBM Corporation

Outline IBM Research & IBM Systems and Technology Group How is the discriminator set on Alpha Sciences counter and what is the alpha-particle energy associated with the setting? What is the impact of cutting out the low energy alphas from a sample? Introduction Background Overview of the Alpha Sciences proportional counters Our low-energy alpha-particle source Results Conclusion 2 10/26/2011

Introduction IBM Research & IBM Systems and Technology Group We review a technique for determining the alpha-particle energy associated with a discriminator setting. We have a low energy thick alpha-particle source that was developed for this work. The alpha particles from this source stop in the gas within the active volume of the detector (important for association between pulse-height analysis and alpha-particle energy). We compared the count rate of the source vs discriminator setting in the proportional counter, and the count rate of the same source with a silicon surface barrier detector to obtain an energy scale. 3 10/26/2011

Background IBM Research & IBM Systems and Technology Group Many labs have Alpha Sciences proportional counters Material labs Semiconductor labs Recently, several labs participated in a round-robin study of both low and ultralow emissivity samples, Multicenter comparison of alpha particle measurements and methods typical of semiconductor processing * Very few constraints were placed on the measurement No sample preparation or measurement protocol Samples were in many pieces that could be tiled to account for different sizes of the tray/active area of the counter There was a ~ 2X difference between the largest and smallest sample emissivity measured It was proposed that this difference could be due to the discriminator settings on the individual detectors Most people assume that the discriminators are set to 1 MeV * http://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=5784521 4 10/26/2011

Cutaway section of an old Alpha Sciences counter anode wires Cross section of counter cathode plane 6 mm anode 6 mm cathode sample tray 5 10/26/2011

SRIM simulations of alpha-particles in Ar gas The counter gas used is Ar/CH 4 0.25 0.2 120 104.8 mm 4.25 96 de/dx (MeV/mm) 5 72 48 Range (mm) 0.05 36.2 mm 1.43 24 0 0 1 2 3 4 5 6 7 8 9 10 0 6 10/26/2011

SRIM simulations of alpha-particles in Ar gas 0.25 20 de/dx (MeV/mm) 0.2 5 0.05 16 12 8 4 Range (mm) 0 0 0.25 0.5 0.75 1 1.25 1.5 1.75 2 0 ~1.1 MeV α-particles have a 6 mm range higher-energy alpha particle will not stop in the detector volume 7 10/26/2011

The low-energy α-source We need a low-energy α-source (so that the α s stop in the active volume in the counter) We made an alpha-particle source from natural Samarium since it s one of the lowest energy naturally-occurring α-particle emitter Sm has 3 naturally-occurring α-emitting radioactive isotopes, only 147 Sm is important, due to the shorter ½ life compared to 148 Sm, or 149 Sm Sm Isotope α-energy (MeV) ½-life (years) Abundance (%) Specific activity (α/sec-g) 147 Sm 2.25 1.06E11 15 850 148 Sm 1.93 7E15 11.2 0.013 149 Sm 1.07 >1E16 (stable?) 13.8 Ln2 N( atoms / mol) SA = x T1 / 2(sec) A( g / mol) http://www.nndc.bnl.gov/chart/ Chart of the Nuclides, GE, 13 th ed. 8 10/26/2011

Table of natural alpha-particle emitters and their energies Source- unknown, yellowed sheet in my lab 9 10/26/2011

Energy spectrum from Sm source, in vacuum Counts/ hr/ channel 10 1 measured spectrum noise difference 0 1 2 3 4 5 110310-sm-1 10 10/26/2011

Energy spectrum from 232 Th source, in vacuum 00 Counts/ hr/ channel 0 10 difference measured spectrum β s, noise 1 0 2 4 6 8 10 110218-232th-1 11 10/26/2011

Alpha Sciences counter with Ortec discrete electronics Using off-the-shelf electronics allows for easy adjustment and repair +HV amp discrim. CAMAC counter Charge-sensitive preamp MCA a b Anode Cathode Sample/ tray Analytical form for cylindrical geo. V E ( r) = r ln( b / a) For r=a, V=800V, E~10 6 V/m Pulse Amplitude 2 MeV 1 MeV Ion saturation Prop region Applied Voltage Applied Bias From Knoll Geiger region 12 10/26/2011

Pulse-height spectrum from sources in the Alpha Sciences counter Sm source 232 Th source 0 0 noise, β s Counts/ hr/ channel 10 1 Counts/ hr/ channel 10 1 0.01 0 200 400 600 800 0 A2110311-2 Channel 0.01 0 200 400 600 800 0 A2110311-1 Channel Ultra-low activity sample 0 Counts/ hr/ channel 10 1 noise 0.01 0 200 400 600 800 0 GC9-111024 Channel 13 10/26/2011

Energy spectrum of Sm source, energy loss from sample through cathode 2 mm gap Cpunts/ hr/ channel 10 1 gap 2.5 mm gap 0.01 0 0.5 1 1.5 2 102.5 3 ~1 MeV energy loss in air and mylar- before α s reach active area vacuum air air + mylar Cpunts/ hr/ channel 1 0.01 0 0.5 1 1.5 2 102.5 3 Cpunts/ hr/ channel 1 3 mm gap 0.01 0 0.5 1 1.5 2 2.5 3 14 10/26/2011

Energy spectrum of 232 Th source, energy loss from sample through cathode 000 Counts/ hr/ channel 00 0 10 1 2.5 mm gap, air + mylar 2.5 mm gap, in vacuum 0 2 4 6 8 10 Passage of the alphas in the gap and mylar causes reduction in maximum energy and in the flux 15 10/26/2011

Results IBM Research & IBM Systems and Technology Group Integrate the measured alpha particle energy spectrum from the samarium source, using the silicon detector, above a given threshold (eg E>0.25 MeV, E>0.5 MeV, E>0.75 MeV and E>1.0 MeV) Repeat integration for each of the source-to-detector gaps Plot the resulting integrated count rate vs energy Integrate the count rate for the Alpha Sciences counter above a given ADC voltage threshold (eg V>0.5V, V>0.9V, V>1.5V,V >2.0V) using the discrete electronics We can correlate the pulse height with energy- given the MCA spectrum from the Alpha Sciences counter, and the maximum energy in the alpha particle energy spectra 16 10/26/2011

Results 2 mm gap Si detector data gas counter data 450 400 Counts/ hr/ channel 350 300 250 200 150 O.5 0.9 1.5 2.0 50 20 400 2.5 0 350 0 0.5 1 1.5 2300 2.5 3 data from Si detector fits the gas counter data with a ~2.5 mm gap Counts/ hr/ channel 450 250 200 150 50 20 2.5 mm gap 0 350 0 0.5 1 1.5 2300 2.5 3 Counts/ hr/ channel 450 400 250 200 150 50 20 3 mm gap 0 count rate with this source is ~1-1.3 MeV 0 0 0.5 1 1.5 2 2.5 3 17 10/26/2011

Response of Alpha Sciences counter, original electronics Sm source Th source 300 80000 250 75000 counts/ hr 200 150 counts/ hr 70000 65000 50 60000 0 0 0.05 5 0.2 0.25 0.3 0.35 A1111018 Discriminator setting (V) 55000 0 0.05 5 0.2 0.25 0.3 0.35 A1111019 Discriminator setting (V) Zero count rate corresponds to a total energy loss of 2.25 MeV Raising the discriminator level lowers the detection efficiency (lowers the count rate) for a thick source The slope of count rate vs discriminator setting, and the discriminator setting where the count rate goes to zero, will depend on the amplifier gain 18 10/26/2011

Conclusion IBM Research & IBM Systems and Technology Group We have shown the energy spectrum from a low-energy Sm α-particle source. There is an appreciable energy-loss of low-energy α s before they reach the active gas volume in the Alpha Sciences counters. While the discriminator can be set higher to reduce the background count rate, it cuts out the detection of low-energy α s. The efficiency of detection in these counters is energy-dependent. What energy α s are really important for samples that we measure? What is the α-particle energy associated with the discriminator setting on your Alpha Sciences counters? 19 10/26/2011