Chem 434 Instrumental Analysis Test 1

Similar documents
very high temperature for excitation not necessary generally no plasma/arc/spark AAS

Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry

Chem 434 -Instrumental Analysis Hour Exam 1

2101 Atomic Spectroscopy

FLAME PHOTOMETRY AIM INTRODUCTION

3 - Atomic Absorption Spectroscopy

high temp ( K) Chapter 20: Atomic Spectroscopy

Atomization. In Flame Emission

Chemistry Instrumental Analysis Lecture 18. Chem 4631

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

AN INTRODUCTION TO ATOMIC SPECTROSCOPY

9/13/10. Each spectral line is characteristic of an individual energy transition

Spectroscopy Problem Set February 22, 2018

Emission spectrum of H

a. An emission line as close as possible to the analyte resonance line

CH. 21 Atomic Spectroscopy

Optical Atomic Spectroscopy

COMPARISON OF ATOMIZERS

Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT.

Application of IR Raman Spectroscopy

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Review: Properties of a wave

Ch. 8 Introduction to Optical Atomic Spectroscopy

Atomic Absorption & Atomic Fluorescence Spectrometry

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Advanced Pharmaceutical Analysis

Chem 321 Lecture 18 - Spectrophotometry 10/31/13

II. Spectrophotometry (Chapters 17, 19, 20)

Lecture 7: Atomic Spectroscopy

Partial Energy Level Diagrams

Analytical Chemistry II

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

Wavelength λ Velocity v. Electric Field Strength Amplitude A. Time t or Distance x time for 1 λ to pass fixed point. # of λ passing per s ν= 1 p

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

ATOMIC SPECROSCOPY (AS)

Lecture Outline: Spectroscopy (Ch. 4)

Ultraviolet-Visible and Infrared Spectrophotometry

Chem 155 Quiz 3 Review Topics: Quiz 3 outline

VALLIAMMAI ENGINEERING COLLEGE SRM Nagar, Kattankulathur

The Spectral Classification of Stars

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES

EXPERIMENT 3 Flame Tests & Electron Configuration

Chapter 4 Spectroscopy

Example: model a star using a two layer model: Radiation starts from the inner layer as blackbody radiation at temperature T in. T out.

Spectroscopy. Page 1 of 8 L.Pillay (2012)

Instrumental methods of analysis

Chapter 4. Spectroscopy. Dr. Tariq Al-Abdullah

Intro to Galaxies Light and Atoms - I

9/28/10. Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Valence Electronic Structure. n σ* transitions

1 P a g e h t t p s : / / w w w. c i e n o t e s. c o m / Physics (A-level)

Chem 310 rd. 3 Homework Set Answers

FTIR measurement of NH 3, HCN, SO 2, H 2 S and COS in pulverized lignite oxy-fuel flames Daniel Fleig, Stefan Hjärtstam and Daniel Kühnemuth

Chemistry Instrumental Analysis Lecture 3. Chem 4631

Light and Atoms. ASTR 1120 General Astronomy: Stars & Galaxies. ASTR 1120 General Astronomy: Stars & Galaxies !ATH REVIEW: #AST CLASS: "OMEWORK #1

Training Coarse ECOPROBE 5 CHEMISTRY RS DYNAMICS

ATOC 3500/CHEM 3151 Air Pollution Chemistry Lecture 1

CHAPTER 8 Introduction to Optical Atomic Spectrometry

1. Cyclic voltammetry involves the measurement of a diffusion controlled at an electrode in which the is controlled. (4 points)

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

Atomic Structure and Periodicity

R O Y G B V. Spin States. Outer Shell Electrons. Molecular Rotations. Inner Shell Electrons. Molecular Vibrations. Nuclear Transitions

you-try-it-04.xlsx Step-by-Step Guide ver. 9/4/2009

Temperature Measurement

Application of Hydroxyl (OH) Radical Ultraviolet Absorption Spectroscopy to Rocket Plumes

Lecture 6 - spectroscopy

Lecture- 08 Emission and absorption spectra

Chapter 11. Important to distinguish between them. They are not interchangeable. They mean very different things when used in physics Internal Energy

PRINCIPLE OF ICP- AES

Light sources. Excited gas atoms are the primaty source of radiation in combustion and discharge lamps. lamps is not continuous!

Ultraviolet-Visible and Infrared Spectrophotometry

Design strategy for Low e windows with effective insulation

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES

Infrared Spectroscopy: Identification of Unknown Substances

K20: Temperature, Heat, and How Heat Moves

1 WHAT IS SPECTROSCOPY?

UNIT 4 Part 2 Measurements of Environmental Air Pollution Parameters j

Energy. Position, x 0 L. Spectroscopy and the Particle-in-a-Box. Introduction

10/31/2017. Calculating the temperature of earth (The greenhouse effect) IR radiation. The electromagnetic spectrum

Atomic Absorption Spectroscopy

Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006

Chapter 5 Light and Matter

The Nature of Light. We have a dual model

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.

Cork Institute of Technology. Summer 2005 Instrumental Analysis (Time: 3 Hours) Section A

Chapter 11. Energy in Thermal Processes

Signal to noise Source of noise Signal to noise enhancement

Spectroscopy in Astronomy

Chapter 4 Ultraviolet and visible spectroscopy Molecular Spectrophotometry

Chromatography & instrumentation in Organic Chemistry

Chemistry Instrumental Analysis Lecture 17. Chem 4631

Page 1 of 9 Chem 201 Lecture 6a Summer 09. Announcement: Fluoride experiment will replace the Cu-ASV experiment for those who have not yet done Cu-ASV

Name: Partner(s): 1102 or 3311: Desk # Date: Spectroscopy Part I

Spectroscopy and the Particle in a Box

Atomic Emission Spectroscopy

Atomic Theory and Periodic Trends Practice AP Chemistry Questions

Clinical Chemistry (CHE221) Professor Hicks Week 1. Statistics Made Slightly Less Boring and Introduction to Spectrophotometry. Accuracy vs Precision

Properties of Electromagnetic Radiation Chapter 5. What is light? What is a wave? Radiation carries information

2. Why do the discharge tubes get hot after running for a period of time?

The Physics of Light, part 2. Astronomy 111

Transcription:

Chem 434 Instrumental Analysis Test 1 Name: 1. (15 points) In Chapter 5 we discussed four sources of instrumental noise: Thermal Noise, Shot Noise, Flicker Noise, and Environmental noise. Discuss the differences between these kinds of noise, how they are different from each other, and how you can try to minimize each of these types of noise. Thermal Noise (Johnson Noise) - random fluctuations of electrons with the electronic circuits Can t change k or R, but can change T or f cool circuit to reduce T, slow response time of instrument to reduce f. Shot Noise - arises when electrons jump across junctions i RMS = sqrt(2ie f), I and e fixed so only thing you can do is change response time of instrument to reduce f Flicker noise- Cause not well understood. Inversely proportional to frequency, worst at 0 frequency - use chopper to make signal high frequency, can also be reduced by using certain components like wire wound resistors instead of C resistors Environmental noise - Fluctuations in environment that are picked up either directly or indirectly by circuitry of instrument. Twos ways to fix. One is to shield all electronics and shorten all leads to a minimum. The other is to recognize that certain frequency regions are relatively noise free, so change electronics to operate at these noise free frequencies. 2. (10 points) Discuss the differences between line spectra, band spectra and continuum spectra. Focus on where the spectra come from and how and why they are different. Line spectra - Spectra associated with electronic transitions within an individual atom. Tend to be very sharp with lines widths in the fractions of a nanometer Band spectra - Spectra associated with molecules made up of 2 or more atoms. While these may also be electronic transitions, they are not nearly as sharp as atomic spectra because additional vibrations and rotation states that do not exist for individual atoms broaden Band spectra so they can have linewidths of 10's to 100's of nm. Continuum Spectra - Also called Black-Body radiation. A broad continuous background of emissions that is tied to the temperature of a material rather than distinctive electronic, vibrational, or rotational transitions with a given molecules. The peak wavelength depends directly on the temperature of the material and the linewidth is not well defined because emission occurs as a continuum of light emission.

3. (10 points) What are the differences between an interference filter and an absorption filter? An interference filter is built by having two metal films separated by a transparent dielectric (insulator). Interference effect of light hitting the two metal films causes light at certain wavelengths to be removed through destructive interference. Exact wavelength of this interference can be calculated from the thickness of the film. Tend to allow light though with a very narrow bandwidth. Effective from UV to IR relatively expensive Adsorption Filter - glass or gelatin or plastic that contains a dye that absorbs light at certain wavelengths. Bandwidths are much larger than inference filters and tem to be less selective, but they are cheaper 2 4. (10 points) What are the three sources of line broadening in atomic spectroscopy what is the physical effect that causes the broadening, and approximately what is the line width due to each kind of broadening. Uncertainty in transition time tied with uncertainly in frequency makes linewidth.0001a called natural line width Doppler Broadening - molecules in flame are moving around puts Doppler shifts on top of wavelengths take width out to.01a Pressure Broadening - molecules colliding with each other or with other atoms in the flame broadens out further to.1-.01 A

3 5. (10 points) It is observed that in a hot flame the emission intensities of the sodium lines at 589.0 and 589.6 are greater in a sample solution that contains KCl than when this compound is absent. Suggest an explanation. Na, being a Group I metal is a species that you would suspect might be ionized fairly easily. Since K is also a Group I metal, but is even lower on the periodic table than Na, you would expect it to be more easily ionized. Thus I suspect that this is an example of ion suppression. In the presence of K the K gets ionized first and this suppresses the ionization of Na, so you get a stronger signal for Na. In the absence of K, Na itself can + 0 become ionized, to Na, thus lowering the concentration of Na in the flame and lowering its signal. 6. (10 points) Describe the structure of a flame. What are the different zones, what is the chemistry that is going on in each of these regions, Where are the hottest and coldest parts of the flame, why do some atoms give a stronger absorbance at the base of a flame, the middle of a flame, at the tip of a flame? Primary Combustion Zone - light blue inner for of the flame. See band spectra for C 2 CH and radicals, not at equilibrium so not used for analytical purposes. Interzonal region - region between primary combustion and secondary combustion. Size varies with air/fuel mix. Is minimal in flames with 1:1 fuel to oxygen. Very large in fuel rich flames. Has lots of free atoms, contains the hottest part of the flame about 1cm above the primary combustion zone. Secondary Combustion zone - products mixing with atmospheric oxygen so getting more oxides. Also cooler so usually not used. Height of maximum signal for any atom is a hard to predict. Have most excited atoms in the hottest part of the flame, but have various reaction that change atomic atoms into various ions at other parts of flame. Species that can be easily oxidized are best examined low on the flame while other atoms that resist ionization can be sampled much higher on the flame. 7. (15 points) Compare and contrast electrothermal atomization and flame atomization. Start by describing the physical set up of each method, then compare their strengths and weaknesses. Electrothermal - sample is placed in a closed graphite oven heated to evaporate, then ash, then atomize sample. Sample can be examined by passing light directly through the oven. Tends to be simpler with less sample preparation and little interference effect because there is no flame. Also tends to be more sensitive and uses less material (a few l s) because, once atomized, the atoms stay in the light path for about a second. Flame - Simple only if sample is readily soluble in some solvent, otherwise can have extensive sample preparation getting sample into solution. Once sample is in solution the liquid is simply aspirated into the gas that feeds the flame. Not as sensitive and uses much more material that electrothermal, but much more reproducible to the uncertainty in any measurement is much smaller.

4 8 (20 points) TAKE HOME QUESTION The following data was taken form the calibration of an instrument Conc (ppm) # of replication Signal Standard deviation (ppm) 0 5 0.025 0.011 3 5 11.5 0.015 6 5 19.5 0.016 9 5 31.2 0.016 12 5 39.1 0.018 15 5 45 0.02 Calculate the following for this instrument: Sensitivity Detection limit Dynamic range Limit of Quantitation Limit of linearity First plot the data I m suspicious of the last point, so let s plot it again, droping the last point This looks like a tighter fit, so I m going to throw out the last point, assuming it is outside my linear range and work with the above calibration curve.

5 Sensitivity Calibration sensitivity is simply the slope = 3.2617 units/ppm but as the book mentions this is not a great number Analytical sensitivity, =slope/std dev Is a better number. In the above data the std dev varies with the concentration, so I will just used the average std. dev. =.016 = 3.2617/.016 = 204 Detection limit In class I gave the detection limit as the minimum concentration that is 3 std. dev. above the baseline signal. The book says that you should have 20-30 blank measurements to establish the baseline signal, we you only have 5 in the above data table. Oh well, deal with it! S min = 0.025 + 3(.011) S =.058 min This is the minimum signal, now I have to subtract the baseline from that signal and change to concentration by dividing by the slope: (S -.S )/slope = Conc = (.058-.025)/3.2617 =.01 ppm min blank min Dynamic range LOL (Limit of Linearity) - LOQ (limit of quantitation) LOL I will take to be the last concentration that was linear, 12 ppm LOQ is taken as the concentration that give a signal 10x the std. dev. of the baseline =.025 + 10(.011) Signal at 10x std.dev. Of baseline =.025 + 10(.011) =.135 Conc at this signal = (S - S )/slope = (.135-.025)/3.2617 =.034 ppm LOL Blank Dynamic range = 12-.034 = 11.97 ppm The book says that the dynamic range should be a least two orders of magnitude. What does this mean? An order of magnitude is a power of 10. The LOL ia.034, an order of magnitude above this is.34, a second order of magnitude is 3.4 and a third order of magnitude is 34. Our LOQ is at 12 ppm which is somewhere between the second and third order of magnitude, so this is a reasonable method. Limit of Quantitation Done above: LOQ is taken as the concentration that give a signal 10x the std. dev. of the baseline =.025 + 10(.011) Signal at 10x std.dev. Of baseline =.025 + 10(.011) =.135 Conc at this signal = (S - S )/slope = (.135-.025)/3.2617 =.034 ppm LOL Blank Limit of linearity Done above: LOL I will take to be the last concentration that was linear, 12 ppm