Comparison of Polychromatic and Monochromatic X-rays for Imaging

Similar documents
ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

1Meherun Nahar, Sazzad, 3Abdus Sattar Mollah and 4Mir Mohammad Akramuzzaman Correspondence Address INTRODUCTION MATERIALS AND METHODS Mathematical

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

X-ray absorption. 4. Prove that / = f(z 3.12 ) applies.

Atomic and nuclear physics

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

Absorbers for medical X-ray detectors with optimum spatial resolution: a simulation study

DICOM Correction Item

Nuclear Instruments and Methods in Physics Research A 417 (1998) 86 94

X-ray Interaction with Matter

Rad T 290 Worksheet 2

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Electron density and effective atomic number images generated by dual energy imaging with a 320-detector CT system: A feasibility study

1-D Fourier Transform Pairs

X-Ray Emission and Absorption

Study of Monte Carlo Simulator for Estimation of Anti-Scatter Grid Physical Characteristics on IEC 60627:2013-Based

SYNCHROTRON RADIATION: NEW TREND IN X-RAY MAMMOGRAPHY

Atomic and nuclear physics

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

Theoretical analysis of comparative patient skin dose and exposure technique approaches in planar radiography

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Basic principles of x-ray production

Health Physics Services Ltd

Characteristic X-rays of molybdenum

We have seen how the Brems and Characteristic interactions work when electrons are accelerated by kilovolts and the electrons impact on the target

INTERACTIONS OF RADIATION WITH MATTER

EFFECT OF SOURCE X-RAY ENERGY SPECTRA ON THE DETECTION OF FLUORESCENCE PHOTONS FROM GOLD NANOPARTICLES

Basic physics Questions

X-ray Spectroscopy. Danny Bennett and Maeve Madigan. October 12, 2015

Laser-Accelerated protons for radiation therapy

AQA Physics /7408

Metal Artifact Reduction and Dose Efficiency Improvement on Photon Counting Detector CT using an Additional Tin Filter

Modeling of Li-Ion-Batteries to Optimize the Results Gained by Neutron Imaging

X-RAY SPECTRA. Theory:

Data Acquisition and Image Formation Methods for Multi Energy CT

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

PD233: Design of Biomedical Devices and Systems

Atomic and nuclear physics

This is the last of our four introductory lectures. We still have some loose ends, and in today s lecture, we will try to tie up some of these loose

ELEG 479 Lecture #6. Mark Mirotznik, Ph.D. Associate Professor The University of Delaware

For the next several lectures, we will be looking at specific photon interactions with matter. In today s lecture, we begin with the photoelectric

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Absorption of X-rays

Physics of Radiography

1. Which of the following statements is true about Bremsstrahlung and Characteristic Radiation?

size within a local area. Using the image as a matched lter, a nite number of scales are searched within a local window to nd the wavelet coecients th

Cauchois Johansson x-ray spectrograph for kev energy range

Practical 1P4 Energy Levels and Band Gaps

Thermal Characteristics of Rotating Anode X-ray Tube with Emissivity in Aging Process for Digital Radiography

X-ray induced singlet oxygen generation by. nanoparticle-photosensitizer conjugates for. photodynamic therapy: determination of singlet

Andrew D. Kent. 1 Introduction. p 1

Physics of Radiography

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS

LASER-COMPTON SCATTERING AS A POTENTIAL BRIGHT X-RAY SOURCE

MEASUREMENT OF TEMPORAL RESOLUTION AND DETECTION EFFICIENCY OF X-RAY STREAK CAMERA BY SINGLE PHOTON IMAGES

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

Diffractometer. Geometry Optics Detectors

EUV Reflectivity measurements on Acktar Sample Magic Black

Evaluation of Phantom Equivalent Materials in Polychromatic Diagnostic X-Ray Beam

Practical 1P4 Energy Levels and Band Gaps

FXA UNIT G485 Module X-Rays. Candidates should be able to : I = I 0 e -μx

FAFF20: Multispectral Imaging

Computation of energy imparted in diagnostic radiology

11/19/2014. Chapter 3: Interaction of Radiation with Matter in Radiology and Nuclear Medicine. Nuclide Families. Family Nuclides with Same: Example

Shell Atomic Model and Energy Levels

Laboratory Manual 1.0.6

Reconstruction for Proton Computed Tomography: A Monte Carlo Study

LAB 01 X-RAY EMISSION & ABSORPTION

Midterm Review. Yao Wang Polytechnic University, Brooklyn, NY 11201

Computational Toolset for X-ray Spectral Analysis

Radiation Detection and Measurement

High-Resolution Neutron Diffraction Monochromators for Neutron Diffractometry

A. I, II, and III B. I C. I and II D. II and III E. I and III

In today s lecture, we want to see what happens when we hit the target.

X-Ray Photoelectron Spectroscopy (XPS)

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

X-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination

APPLIED RADIATION PHYSICS

ATTENUATION STUDIES ON DRY AND HYDRATED CROSS-LINKED HYDROPHILIC COPOLYMER MATERIALS AT 8.02 TO kev USING X-RAY FLUORESCENT SOURCES

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

11/10/2014. Chapter 1: Introduction to Medical Imaging. Projection (Transmission) vs. Emission Imaging. Emission Imaging

Pixels GaAs Detectors for Digital Radiography. M.E. Fantacci. and. Abstract

Revision Guide. Chapter 7 Quantum Behaviour

Two-Material Decomposition From a Single CT Scan Using Statistical Image Reconstruction

Photoelectron spectroscopy Instrumentation. Nanomaterials characterization 2

Study of the phase contrast for the characterization of the surface of microshell

CASSY Lab. Manual ( )

X-ray imaging with amorphous selenium: Theoretical feasibility of the liquid crystal light valve for radiography

RED. BLUE Light. Light-Matter

SPARCLAB. Source For Plasma Accelerators and Radiation Compton with Laser And Beam

Atomization. In Flame Emission

INTRODUCTION Strained Silicon Monochromator Magnesium Housing Windows for Monochromator Shutter and Collimator Fission Detector HOPG Monochromator

Development of new educational apparatus to visualize scattered X-rays

Training program for x-ray inspection film (RT)

Changing and challenging times for service crystallography. Electronic Supplementary Information

X-ray Spectroscopy. c David-Alexander Robinson & Pádraig Ó Conbhuí. 14th March 2011

X-ray practical: Crystallography

X-ray Absorption Spectroscopy

Spectrophotometry. Introduction

Transcription:

Comparison of Polychromatic and Monochromatic X-rays for Imaging M. Hoheisel 1, P. Bernhardt 1, R. Lawaczeck 2, and H. Pietsch 2 1 Siemens AG Medical Solutions, Forchheim, Germany 2 Schering AG Imaging Diagnostics and Radiopharmaceuticals Research, Berlin, Germany ABSTRACT Monochromatic X-rays have been proposed for medical imaging, especially in the mammographic energy range. Our previous investigations have shown that the contrast of objects such as lesions or contrast media can be enhanced considerably by using monochromatic X-rays instead of the common polychromatic spectra. Admittedly, only one specific polychromatic spectrum and one monochromatic energy have been compared so far. In this work, we investigated the contrast yielded by a series of different X-ray spectra obtained by varying tube voltage and beam filtering. This resulted in spectra of different mean energies and spectral widths. The objects under examination were aqueous solutions containing different chemical elements such as I, Gd, Dy, Yb, and Bi. A monoenergetic spectrum at 17.5 was obtained using a mammographic X-ray tube with a Mo anode and a monochromator equipped with a HOPG crystal. Moreover, we simulated quasi-monoenergetic spectra at different energies and with different widths. As a result, we demonstrated that in many cases spectra with an energetic width of some yield an equivalent contrast to monoenergetic radiation at the same energy. Therefore, the advantage in image contrast of monochromatic X-rays at 17.5 over narrow-band polychromatic X-ray spectra obtained by appropriate filtering is only slight. Thus, the additional expenditure on a mammography system with HOPG monochromator that can deliver only a small X-ray dose and the unfavorable slot-scan geometry can be avoided. Moreover, we carried out simulations of monochromatic versus polychromatic spectra throughout the whole radiographic energy range. We found advantages in using monochromatic X-rays at higher energies and thicker objects that will justify their application for diagnostic imaging in a number of specific cases. Keywords: Medical X-ray imaging, monochromatic source, Highly Oriented Pyrolytic Graphite (HOPG), spectral width, mammography, signal-difference-to-noise ratio, contrast enhancement, contrast media 1. INTRODUCTION In the field of medical X-ray imaging, much effort has been directed towards finding an optimum spectrum for every specific diagnostic situation. This includes the choice of the anode material of the X-ray tube, the tube voltage, as well as material and thickness of additional filters 1. To visualize a certain lesion, the ratio of contrast to noise should be as high as possible, where contrast means the signal difference between lesion and surrounding tissue. Since the absorption coefficient becomes lower with increasing energy, a low energy will deliver high contrast. Unfortunately, with lower energy the patient dose increases while less quanta will reach the detector and be registered. This calls for the use of high energies. It is evident that there is an optimum energy in between. This gives rise to the demand for a monochromatic X-ray source at this very energy. A simulation study on this subject has been conducted for instance by Boone and Seibert 2. Up to now, monochromatic X-rays have been produced by synchrotrons or parametric X-ray sources 3, which are very expensive and complex methods. Therefore, synchrotron radiation has not found its way into everyday clinical life. The common approach is to use as narrow a continuous spectrum as possible, which is accomplished by absorption filters. Thick filters are extremely useful for this purpose, but are limited by the increase in tube load. In our previous work 4 we pursued another approach. We used an ordinary X-ray tube, a highly oriented pyrolytic graphite (HOPG) mirror, and a slit defining the Bragg angle for the desired energy. This arrangement forms a fan beam source. Together with a linear detector we have a slot-scan system. Diekmann et al. 5 demonstrated that this approach Medical Imaging 2006: Physics of Medical Imaging, edited by Michael J. Flynn, Jiang Hsieh, Proceedings of SPIE Vol. 6142, 614209, (2006) 1605-7422/06/$15 doi: 10.1117/12.651037 Proc. of SPIE Vol. 6142 614209-1

might be advantageous in some situations. Nevertheless, such a slot-scan system with a monochromator has several drawbacks, such as a quantum flux that is lower by a factor of 20 thus necessitating multiple scans to reach the dose required for a sufficient noise level, or extremely long scanning time leading to motion artifacts. Therefore, we tried to work out whether a well-adapted polychromatic spectrum can yield the same performance as monochromatic X-ray radiation. 2. METHOD This work is based mainly on simulations. We used a series of different phantoms that span the whole range of medical diagnostic procedures. For the mammographic energy range we used 2 cm to 8 cm thick slabs composed of 25% adipose and 75% glandular tissue. For the radiographic energy range we applied water phantoms of between 10 cm and 44 cm thickness. In each phantom we implanted thin pieces of different materials, i.e. the chemical elements Al, Y, I, Gd, Dy, Yb, and Bi (in order of their atomic numbers 13, 39, 53, 64, 66, 70, and 83, respectively), representing lesions or contrast agents. We investigated in this study the signal-difference-to-noise ratio (SDNR) of these materials. The SDNR is a well-suited quantity to characterize the contrast behavior of an object, since it is invariant to many image-processing operations, e.g. windowing. As a figure of merit we chose Q = SDNR²/K with the air kerma K at the surface of the phantom, because this variable easily allows the comparison of results derived from different measurements. In one case, Q = SDNR²/D with the absorbed dose D was used instead. The simulations included monochromatic X-rays as well as polychromatic spectra. A deterministic simulation yielded the contrast emerging from the different material samples. An additional Monte-Carlo simulation produced the noise figures. Scattered radiation was ignored in the course of this work. Details of the simulations will be published elsewhere 6. The monochromatic energy E x was varied over a sufficiently broad spectral range to yield a Q(E x ) function with a maximum. Other simulations have been carried out using polychromatic spectra which are given by the X-ray tube and filtration parameters (anode material, tube voltage, filter material and thickness). The idea was to substitute the optimal monochromatic radiation with a broader X-ray spectrum at the same mean energy. Then the signal is generated by both the optimum energy and suboptimal spectral contributions. One aim of this study was to evaluate to what extent a broader spectrum can lead to the same or nearly the same SDNR as the monochromatic spectrum. Furthermore, we carried out measurements on real phantoms. The objects under examination were 10 mm high cuvettes with aqueous solutions containing different chemical elements such as I, Gd, Dy, Yb, and Bi in a concentration of 20 mmol/l each. The measured contrast was defined as the quotient between the signal intensity behind these objects and the signal intensity behind a similar cuvette containing pure water. The X-ray system for monochromatic imaging was compiled as follows: The basic system was a Siemens Mammomat 300 mammography device, contributing the X-ray tube, the high voltage generator, the operating console, and the mechanical setup. The tube had a molybdenum anode and was operated with 35 kv voltage and 100 ma current in most cases. A permanent pre-filtration of 30 µm Mo was used. Admittedly, we could use only one single monochromatic energy, 17.5, which is the K α energy of molybdenum. The monochromator could have been tuned to other energies also, but only delivers a sufficiently high quantum flux at the K α line. The detector was a CCD (Thales TH9570), 4092 x 200 pixels, 54 µm in size in the binning mode. For the polychromatic spectra we used the same mammography system with its Mo anode without monochromator and varied the tube voltage in the range 23 kv to 35 kv. Extra beam filtering was accomplished by a series of 1 mm to 4 mm thick aluminum sheet enabling us to obtain spectra of different mean energy (17.4 to 27.4 ) and different spectral width listed in Table 1. The resulting spectra could also be simulated and are plotted in Fig. 1. All spectra show a pronounced peak at the K α energy of 17.5, a second, smaller peak at the K β energy of 19.6, and a more or less large continuum up to the maximum energy corresponding to the tube voltage selected. Therefore, the overall spectra are rather broad; giving just the spectral width as a parameter is certainly inadequate to characterize each spectrum. Proc. of SPIE Vol. 6142 614209-2

Table 1 Tube voltage Al filter thickness Mean energy Width 23 kv 1 mm 17.4 1.8 29 kv 1 mm 18.7 2.9 35 kv 1 mm 20.6 4.9 35 kv 2 mm 23.1 5.6 35 kv 3 mm 25.5 5.4 35 kv 4 mm 27.4 4.7 Normalized spectral density [arbitrary units] 23 kv, 30 µm Mo + 1 mm Al 29 kv, 30 µm Mo + 1 mm Al 35 kv, 30 µm Mo + 1 mm Al 35 kv, 30 µm Mo + 2 mm Al 35 kv, 30 µm Mo + 3 mm Al 35 kv, 30 µm Mo + 4 mm Al 12 14 16 18 20 22 24 26 28 30 32 34 Figure 1: Series of mammographic X-ray spectra from a tube with Mo anode, 30 µm Mo pre-filtration, different tube voltages, and different Al filter thicknesses. 3. SIMULATION RESULTS First we studied the optimum (monochromatic) X-ray energy E x for different situations characterized by the phantoms described above. Fig. 2 shows the figure of merit Q as a function of quantum energy for a specific case (Ca sample in a series of mammography phantoms of different thickness). At each thickness a clear maximum is visible which denotes the optimum energy E x. As a result, the optimum energy E x can be derived as a function of phantom thickness for a given kind of phantom and contrast sample. This is plotted in the insert of Fig. 2. From many of these graphs one can compile diagrams such as Fig. 3. In Fig. 3 Q is again plotted as a function of quantum energy for a 20 cm-thick water phantom and various contrast objects which are characterized by their different K α energies. The maxima of these curves, i.e. the optimum quantum energies E x are 45 for those materials with a K α energy below 45. Contrast objects with K α > 45 lead to pronounced maxima near their respective K α energies. Proc. of SPIE Vol. 6142 614209-3

SDNR²/D [arbitrary units] 45 40 35 30 25 20 15 10 2 cm 3 cm 4 cm 5 cm 6 cm 7 cm 8 cm Optimum quantum energy 26 25 24 23 22 21 20 19 18 1 2 3 4 5 6 7 8 9 Phantom thickness cm 5 0 16 18 20 22 24 26 28 30 32 34 36 38 40 42 Figure 2: Figure of merit Q = SDNR²/D as a function of quantum energy assuming monochromatic X-rays for a calcification in a series of mammography phantoms (25% fatty, 75% glandular tissue) of different thickness. Every curve shows a clear maximum at its optimum energy E x. The insert shows E x as a function of phantom thickness. 18 SDNR²/K [arbitrary units] 16 14 12 10 8 6 Bi 15 µm (Ek = 90.5 ) Yb 15 µm (Ek = 61.3 ) Dy 15 µm (Ek = 53.8 ) Gd 15 µm (Ek = 50.2 ) I 15 µm (Ek = 33.2 ) Y 30 µm (Ek = 17.0 ) Al 500 µm (Ek = 1.6 ) 4 2 0 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 110 115 120 125 Figure 3: Figure of merit Q = SDNR²/K as a function of quantum energy assuming monochromatic X-rays for different contrast objects in a 20 cm-thick water phantom. The curves are guides to the eye. Every curve shows a clear maximum at its optimum energy E x. If the K-edge energy E k of the contrast object is above 45, E x is found in the vicinity of E k. Proc. of SPIE Vol. 6142 614209-4

For the simulations using polychromatic spectra we varied the tube voltage from 40 kv to 150 kv in steps of 5 kv. The additional filtration was 0 mm, 0.1 mm, 0.2 mm, 0.3 mm, 0.6 mm, or 0.9 mm for each voltage step. Again, Q was calculated for the same contrast elements as above and different phantom thicknesses. Similar to the results shown in Fig. 3, a maximum can be found for every phantom and contrast object. The X-ray spectrum which produced this maximum was then calculated from the tube voltage and filtration parameters. Fig. 4 is a compilation of these spectra. 0,12 Spectral density [arbitrary units] 0,10 0,08 0,06 0,04 Y 30 µm / I 15 µm / Bi 15 µm (50 kv) Al 500 µm (55 kv) Gd 15 µm (70 kv) Dy 15 µm (75 kv) Yb 15 µm (85 kv) 0,02 0,00 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 Figure 4: A series of X-ray spectra which were found to yield the optimum figure of merit Q = SDNR²/K for the different contrast objects (given in the caption) in a 20 cm-thick water phantom. Although all spectra have been obtained with maximum filter thickness (0.9 mm) their spectral width is rather high. It is interesting to note that all spectra of Fig. 4 have been obtained with the maximum filter thickness, i.e. 0.9 mm, which is the maximum filter thickness implemented in the Siemens Axiom X-ray systems. Thicker filters would require a much higher tube power which is currently not available. Thicker filters lead to narrower spectra which come closer to ideal monochromatic spectra. As long as the tube voltage is kept below 65 kv, the K α emission line of the W anode at 59.3 does not play a significant role and the resulting spectra in Fig. 4 consist of a single peak. Higher tube voltages result in rather broad spectra with a pronounced superimposed peak at 59.3 and, at even higher voltages, a second peak at 67.2, the K β emission line of W. The consequence of these peaks is that it is impossible to raise the mean energy of the spectrum significantly above 70, even when the tube voltage is set to the highest value we have on-hand, i.e. 150. The optimum Q values found with monochromatic (Q mono ) and polychromatic (Q poly ) spectra for the different contrast objects at a 20 cm and, since the optima depend markedly on the phantom thickness, a 40 cm water phantom are listed in Table 2. All polychromatic spectra applied were those shown in Fig. 4, where a filtration of 0.9 mm Cu and different tube voltages had been used. Furthermore, the quotient Q poly /Q mono is given for each case to show in how far Q poly can compete with Q mono. Similar relations between monochromatic and polychromatic radiation on phantoms of different thickness had already been found by Oosterkamp 7. Proc. of SPIE Vol. 6142 614209-5

Table 2 Contrast 20 cm water phantom 40 cm water phantom object Q monochromatic Q polychromatic Q poly /Q mono Q monochromatic Q polychromatic Q poly /Q mono Al 5.380E+05 4.698E+05 87.3% 7.224E+03 5.252E+03 72.7% Y 2.326E+06 2.001E+06 86.0% 2.452E+04 1.603E+04 65.4% I 3.130E+06 2.647E+06 84.6% 3.442E+04 2.251E+04 65.4% Gd 1.241E+07 6.019E+06 48.5% 1.869E+05 9.937E+04 53.2% Dy 1.627E+07 6.252E+06 38.4% 2.451E+05 1.177E+05 48.0% Yb 7.627E+06 2.563E+06 33.6% 1.543E+05 5.184E+04 33.6% Bi 5.389E+06 4.559E+06 84.6% 1.098E+05 3.932E+04 35.8% The polychromatic spectra resulting from our simulations are unrealistic from a practical point of view. For all spectra a 0.9 mm Cu filtration was used to make the spectra as narrow as possible. Since a sufficiently high dose is required to produce high-quality radiograms, in most cases no additional filter is used. Table 3 shows the Q values under the assumption that no additional filters have been applied. Table 3 Contrast 20 cm water phantom 40 cm water phantom object Q monochromatic Q polychromatic Q poly /Q mono Q monochromatic Q polychromatic Q poly /Q mono Al 5.380E+05 2,324E+05 43.2% 7.224E+03 2,580E+03 35.7% Y 2.326E+06 9,454E+05 40.6% 2.452E+04 6,993E+03 28.5% I 3.130E+06 1,177E+06 37.6% 3.442E+04 9,874E+03 28.7% Gd 1.241E+07 2,610E+06 21.0% 1.869E+05 4,024E+04 21.5% Dy 1.627E+07 2,894E+06 17.8% 2.451E+05 4,667E+04 19.0% Yb 7.627E+06 1,217E+06 16.0% 1.543E+05 2,186E+04 14.2% Bi 5.389E+06 2,164E+06 40.2% 1.098E+05 2,242E+04 20.4% Taking practical boundary conditions into account, even well-adapted polychromatic spectra deliver Q values roughly a factor of two lower than narrow polychromatic spectra, and up to a factor of seven lower than the Q values that would be achieved with monochromatic X-rays, as can be seen from Table 3. 4. EXPERIMENTAL RESULTS For the experimental investigations we followed a slightly different approach since we had to use a fixed monochromatic energy of 17.5. We used the spectra denoted in Table 1 and Fig. 1. Then we determined Q for I, Gd, Dy, Yb, and Bi as a function of the mean energy (Fig. 5). Q increases when the mean energy is lowered and reaches almost the same value as in the monochromatic case. The different contrast objects were 10 mm high cuvettes with aqueous solutions as explained above. Therefore, the thickness of the phantom was held constant at 1 cm and not varied as in the simulations. The contrast for all elements increased according to their attenuation coefficients, i.e. with the atomic number of the contrasting element from I to Bi. Proc. of SPIE Vol. 6142 614209-6

SDNR²/K [arbitrary units] 0,50 0,45 0,40 0,35 0,30 0,25 0,20 0,15 Yb I Gd Dy Bi 0,10 0,05 0,00 16 18 20 22 24 26 28 Figure 5: Figure of merit Q = SDNR²/K as a function of the mean energy of different polychromatic spectra. The objects consisted of aqueous solutions of I, Gd, Dy, Yb, and Bi with 20 µmol/cm² each. The corresponding Q values for the monochromatic case are plotted at 17.5 in larger symbols. With a mean energy of 17.5 nearly the same Q values can be reached. 5. CONCLUSIONS In this study we investigated the optimum energy of monochromatic X-rays for medical diagnostic applications. The achievable SDNR values were compared with those of appropriate polychromatic spectra. Simulations with monochromatic spectra confirm the well-known fact that in most cases there is an optimum quantum energy for a good Q = SDNR²/K. Fortunately, in mammography Q shows a rather broad maximum as can be seen from Fig. 2. Deviations of ± 2 will reduce SDNR by some 10% which often might be tolerable. As a general trend, thicker objects require higher quantum energies for sufficient penetration. However, depending on the details of interest present in these objects, e.g. contrast media, the optimum energy may be different. The optimum energy results primarily from the thickness of the background object (see Al, Y, and I in Fig. 3). If the K-edge energy of the dominating chemical element in the detail is below this energy, the optimum energy is not noticeably changed. With a higher K-edge energy a significantly better Q value can be obtained at a quantum energy slightly above this K-edge energy. Using polychromatic spectra one can yield similar SDNR values as long as the maxima of Q as a function of quantum energy are sufficiently broad and the polychromatic spectra are accordingly narrow. This is generally the case at low energies such as applied in mammography. Table 2 shows that the Q values are only some 10% to 30% lower for a 20 cm and a 40 cm thick phantom, and for monochromatic compared to polychromatic X-rays, respectively. Proc. of SPIE Vol. 6142 614209-7

If contrast media with a higher Z are present, the optimum shifts to a much higher energy, as can be seen in Fig. 3. This calls for a polychromatic spectrum at this higher energy. Unfortunately, in this energy range the X-ray spectra are dominated by the K α and K β emission lines of W. The total spectra become rather broad and mean energies significantly above 70 are not feasible (see Fig. 4). These relations lead to an interesting finding. Bi is a material with a very high Z of 83 and a K-edge energy of 90.5. Therefore, the optimum quantum energy for a phantom with a Bi detail is around 95. One would expect that the optimum polychromatic spectrum for Bi would require a very high tube voltage. Conversely a tube voltage of only 50 kv leads to the best Q value. The explanation is that even the highest tube voltage available cannot push the mean energy of the spectrum high enough. That is why the optimum polychromatic spectrum for Bi is found to have a mean energy of only 45 which coincidences with the second (lower) maximum of the Bi curve in Fig. 3. In general it was found that using thick absorption filters next to the tube is beneficial in all cases. Unfortunately, this is limited by the maximum tube load the system can deliver. The experiments we carried out at a monochromatic energy of 17.5 and with polychromatic spectra produced by a mammography system confirmed that with subtly adapted polychromatic spectra one can yield nearly the same SDNR as in the monochromatic case. Therefore, using a sophisticated monochromatic scanning setup for mammography appears not to be justified. Under radiographic conditions the situation is similar. However, there is some benefit from using monochromatic X-rays instead of polychromatic spectra. With monochromatic X-rays of appropriate energy, Q values of about a factor of two to three higher than for optimized polychromatic spectra might be reached, as can be seen from Table 2. Admittedly, it is much more complicated to produce monochromatic radiation at theses high energies (> 40 ). The Bragg angles become very small which results in very long system setups (several meters) that will not fit in a clinical environment. Moreover, cutting out a single energy out of the bremsstrahlung spectrum is not effective and leads to a low quantum flux and extremely long exposure times. Finally, a scanning setup is necessary to utilize monochromatic radiation. ACKNOWLEDGEMENT We would like to thank M. Murphy for critically reading the manuscript. REFERENCES 1. M. Säbel, H. Aichinger, Recent developments in breast imaging, Physics in Medicine and Biology 41, 315-368, 1996 2. J. M. Boone, J. A. Seibert, A comparison of mono- and poly-energetic X-ray beam performance for radiographic and fluoroscopic imaging, Medical Physics 21, 1853-1863, 1994 3. M. A. Piestrup, X. Wu, V. V. Kaplan, S. R. Uglov, J. T. Cremer, D. W. Rule, R. B. Fiorito, A design of mammography units using a quasimonochromatic X-ray source, Review of Scientific Instruments 72, 2159, 2001 4. M. Hoheisel, R. Lawaczeck, H. Pietsch, V. Arkadiev, Advantages of monochromatic X-rays for imaging, Proceedings of SPIE 5745, 1087-1095, 2005 5. F. Diekmann, S. Diekmann, K. Richter, U. Bick, T. Fischer, R. Lawaczeck, W. R. Press, K. Schon, H. J. Weinmann, V. Arkadiev, A. Bjeoumikhov, N. Langhoff, J. Rabe, P. Roth, J. Tilgner, R. Wedell, M. Krumrey, U. Linke, G. Ulm, B. Hamm, Near monochromatic X-rays for digital slot-scan mammography: initial findings, European Radiology 14, 1641-1646, 2004 6. P. Bernhardt, T. Mertelmeier, M. Hoheisel, X-ray spectrum optimization of full-field digital mammography: simulation and phantom study, submitted to Medical Physics 7. W. J. Oosterkamp, Monochromatic X-rays for medical fluoroscopy and radiography, Medica Mundi 7, 68-77, 1962 Proc. of SPIE Vol. 6142 614209-8