Carbon and Heteronuclear NMR on the Bruker

Similar documents
HMQC HSQC and HMBC. Gradient HMQC, HMBC on the Bruker400 and 500

NMR: PRACTICAL ASPECTS

The NMR Spectrum - 13 C. NMR Spectroscopy. Spin-Spin Coupling 13 C NMR. A comparison of two 13 C NMR Spectra. H Coupled (undecoupled) H Decoupled

τ 1 > 1/J - if this lifetime is significantly shortened, the coupling (splitting of the signal) will not be observed

X-Nucleus NMR: 31 P and 19 F

BASIC NMR HANDBOOK Written by M. A. Eastman Copyright 1997, 2001, 2013, 2015, 2018

NMR FACILITY NEWSLETTER

Carbon 13 NMR NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY

How to perform 2D NMR Experiments on the Varian/Agilent VNMRS 500 when using the Chempack interface

High-Resolutio n NMR Techniques i n Organic Chemistry TIMOTHY D W CLARIDGE

North Carolina State University Department of Chemistry Varian NMR Training Manual

1H 1D-NOE Difference Spectra and Spin-Saturation Transfer Experiments on the GN500

NMR Data workup using NUTS

NMR at UNC: Tips, Tricks, and Techniques

Introduction to 1D and 2D NMR Spectroscopy (4) Vector Model and Relaxations

Magnetic Nuclei other than 1 H

Nuclear magnetic resonance spectroscopy II. 13 C NMR. Reading: Pavia Chapter , 6.7, 6.11, 6.13

PAPER No. 12: ORGANIC SPECTROSCOPY. Module 19: NMR Spectroscopy of N, P and F-atoms

Pulsar. Delivering NMR to your benchtop

NMR PRAKTIKUM. Data processing Data acquisition... 17

Determining C-H Connectivity: ghmqc and ghmbc (VnmrJ-2.2D Version: For use with the new Software)

Chemistry Department

2D NMR: HMBC Assignments and Publishing NMR Data Using MNova

Relaxation, Multi pulse Experiments and 2D NMR

PROTEIN NMR SPECTROSCOPY

Once again, a good way to keep track of these is to fill out the table below: Nucleus power pw90 spectral width offset

Operation of the Bruker 400 JB Stothers NMR Facility Department of Chemistry Western University

Login -the operator screen should be in view when you first sit down at the spectrometer console:

SSSC Discovery Series NMR2 Multidimensional NMR Spectroscopy

The Use of NMR Spectroscopy

Nuclear Magnetic Resonance

EDGEWOOD CHEMICAL BIOLOGICAL CENTER

Chapter 13: Nuclear Magnetic Resonance (NMR) Spectroscopy direct observation of the H s and C s of a molecules

Laboration 8a. Relaxation, T 1 -measurement with inversion recovery

NMR TRAINING. What to Cover

H-C one-bond correlations: HSQC Preliminaries.

Chapter 13 Structure t Determination: Nuclear Magnetic Resonance Spectroscopy

Getting started with NMR

High-resolution probes for 19 F applications. Aitor Moreno Bruker BioSpin, Switzerland BeNeLux NMR User Meeting 2016, Brussels 25 th of November 2016

Determining Chemical Structures with NMR Spectroscopy the ADEQUATEAD Experiment

ONE AND TWO DIMENSIONAL NMR SPECTROSCOPY

Structure Determination: Nuclear Magnetic Resonance Spectroscopy

NMR NEWS June To find tutorials, links and more, visit our website

Center for Sustainable Environmental Technologies, Iowa State University

Stable isotope analytical methods

CHEM Chapter 13. Nuclear Magnetic Spectroscopy (Homework) W

Protein NMR. Part III. (let s start by reviewing some of the things we have learned already)

Non Uniform Sampling in Routine 2D Correlation Experiments

Inverse Detection in Multinuclear NMR

NMR = Nuclear Magnetic Resonance

Module 20: Applications of PMR in Structural Elucidation of Simple and Complex Compounds and 2-D NMR spectroscopy

Two Dimensional (2D) NMR Spectroscopy

NMR Practical Considerations Gregory R. Cook, NDSU Monday, January 28, 13

H-C one-bond correlations: HSQC Preliminaries.

Double-Resonance Experiments

Nuclear Magnetic Resonance

16.1 Introduction to NMR Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy. Spectroscopy 4/11/2013

NMR Spectroscopy: A Quantum Phenomena

COSY type experiments exploring through-bond homonuclear correlations

NMR course at the FMP: NMR of organic compounds and small biomolecules - II -

[1] [2] 10 CH 3 COOH 9 1D 1 H NMR

Chemistry 416 Spectroscopy Fall Semester 1997 Dr. Rainer Glaser

NMR course at the FMP: Practical aspects

8.2 The Nuclear Overhauser Effect

Physical Background Of Nuclear Magnetic Resonance Spectroscopy

Z 2 (OMe) = Z 3 (Cl) = 1.4. Calculated: Measured: Deviation:

Experience the most powerful benchtop NMR spectrometer

BMB/Bi/Ch 173 Winter 2018

Homonuclear 2DJ spectroscopy - HOMO2DJ

Homonuclear Broadband Decoupling via PSYCHE Element Benjamin Görling, Aitor Moreno, Wolfgang Bermel

Ethyl crotonate. Example of benchtop NMR on small organic molecules

MOLECULAR SPECTROSCOPY AND PHOTOCHEMISTRY

It is possible to choose the temperature for each experiment by setting a temperature under the Temp pane (under the Standard panel).

BMB/Bi/Ch 173 Winter 2018

Chapter 13 Nuclear Magnetic Resonance Spectroscopy

5 Heteronuclear Correlation Spectroscopy

Nuclear Magnetic Resonance H-NMR Part 1 Introduction to NMR, Instrumentation, Sample Prep, Chemical Shift. Dr. Sapna Gupta

Filtered/edited NOESY spectra

William H. Brown & Christopher S. Foote

Ammonium Sulfate as a Standard for 33S-NMR Spectra

Nuclear Magnetic Resonance Spectroscopy Chem 4010/5326: Organic Spectroscopic Analysis Andrew Harned

Experiment 4: Equilibrium Thermodynamics of a Keto-Enol Tautomerism Reaction

Basic One- and Two-Dimensional NMR Spectroscopy

Agilent s new solution for obtaining routinely quantitative results from NMR measurements. Magnetic Resonance Systems

Nuclear Magnetic Resonance Spectroscopy

NMR-spectroscopy of proteins in solution. Peter Schmieder

C NMR Spectroscopy

Adam Wu, Jean Michel Lauzon, Indah Andriani, and Brian R. James*

Tools for Structure Elucidation

Technical Note. Introduction

THE NUCLEAR OVERHAUSER EFFECT IN STRUCTURAL AND CONFORMATIONAL ANALYSIS

UI500NB Spectrometer: Advanced 1D and 2D NMR Experiments With Application to Structure Elucidation of Small Organic Molecules

Supporting Information

With that first concept in mind, it is seen that a spinning nucleus creates a magnetic field, like a bar magnet

Ibuprofen. Example of benchtop NMR on small organic molecules

NMR SPECTROSCOPY DR. M. KANJIA. Copyright reserved NMRS. Application to reproduce to Dr M Kanjia

HSQC spectra for three proteins

Química Orgânica I. Nuclear Magnetic Resonance Spectroscopy (II) Ciências Farmacêuticas Bioquímica Química AFB QO I 2007/08 1 AFB QO I 2007/08 2

CHEM311 FALL 2005 Practice Exam #3

To Do s. Read Chapter 3. Complete the end-of-chapter problems, 3-1, 3-3, 3-4, 3-6 and 3-7. Answer Keys are available in CHB204H

Transcription:

Carbon and Heteronuclear NMR on the Bruker There are several different types of carbon spectra such as a normal qualitative spectrum, DEPT, coupled, and those with and without NOE. This handout discusses these experiments and the parameters important for them. IMPORTANT NOTE: when doing carbon spectra, one should not use the rga command. The appropriate value of rg is already set. Simply start the acquisition with zg. Carbon and heternuclear direct detection experiments can only be performed on the DPX400 machine. Qualitative Carbon Spectra Routine carbon spectra are acquired broad band-proton-decoupled, with NOE (which provides up to a factor of 3 signal enhancement), and under optimum conditions of sensitivity (hopefully). These conditions result in qualitative spectra, i.e., the intensities do not reflect the number of carbons in the molecule. See below for more on this. 1. Read in the carbon parameters: type rpar carbon. Lock and shim. 2. Tune the Probe: The next step is to tune the carbon channel of the probe. This is essential since if the probe is not tuned, no signal may be observed. To tune the probe, type wobb, and under Acquire at the top, select observe fid window (or type in acqu ). In a few seconds a tuning curve with a dip will be displayed. Tuning and matching capacitors are now adjusted to make the dip both centered and as deep as possible. The matching and tuning capacitors on the DPX400 are gold sliders with numbered indicating positions. There is a chart near the probe that lists for each nucleus an approximate set of numbers for the tuning and matching. Adjust the least significant bit until the LEDs of the preamplifier next to the probe show a minimum number of lights. The optimum will be only green LEDs are lit, and all red and yellow LEDs are off. If you have not done this before, ask for assistance. Do NOT tune carbon and nitrogen channels on the DMX500 and 600. Tuning the probe is very important for sensitivity and proper pulse widths and is essential for DEPT. You should always tune the carbon channel. You should also tune the proton channel if you intend to do DEPT. To do so, push the channel select button on the preamp box and wait until the wobble curve changes to proton (this takes a few seconds) and tune using the proton rods. Type stop to end this tuning procedure. 3. Start the acquisition by typing zg (do not type rga ). To look at the spectrum without stopping the acquisition, type tr, and then after the next scan is completed, type ef. To stop scanning, type halt. Yufeng Wei, Ph.D. (212) 327-7220 1

Signal-to-noise The signal-to-noise improves as the square root of the number of scans and thus the square-root of the time of the acquisition. An approximate guide to experiment length (based on strychnine, MW=334. It is the moles that are important so scale according your molecule) is as follows: Sample concentration (mg/ml) 50 15 6 3 Time 5 min 30 min 3 hrs 12 hrs ly sign-up for the time you need. Spectra that show no baseline noise have used too much spectrometer time. Carbon spectra may be essential but they are very time-consuming. With modern NMR spectrometers, a 2D carbon-proton correlation spectrum, such as HMQC or HSQC, requires much less time than a single carbon spectrum. The signal-to-noise of a HSQC is much higher than that for a 1D carbon for a given amount of time. If you are only interested in protonated carbons, the HSQC is to be greatly preferred over the 1D carbon. About sample size and tubes When sample quantity is very limited, it is advantageous to limit the amount of solvent in which it is dissolved. If a normal 5mm tube is used, however, this cannot be less than about 500 ml (400 ml for carbon spectra) without causing serious lineshape problems (shimming problems) and the attendant loss of signal-to-noise. When one reduces the solvent quantity in a normal 5mm tube, it is important that the sample be centered within the coil. To do this, center the sample about the scored line on the plastic depth gauge. There are special tubes (made by Shigemi) and inserts (made by Wilmad), however, that can be used to restrict the active volume and, hence, reduce the amount of solvent without causing lineshape problems. These work reasonably well, with the Shigemi tubes being superior but more costly. Decoupling modes The pulse sequence for the experiments described (except DEPT) here is divided into 2 periods, separated by the pulse. The parameter defining the length of each period is given below: relaxation delay pulse acquisition of FID (d1) (p1) (aq) Yufeng Wei, Ph.D. (212) 327-7220 2

When the decoupler is on during the relaxation delay, the NOE effect is developed. When the decoupler is on during acquisition, a decoupled spectrum is acquired. Since these periods are independent, the two effects are independent, i.e., one can have a coupled spectrum with NOE or a decoupled spectrum without NOE. The normal qualitative carbon spectrum is decoupled with NOE and hence, the decoupler is on all the time. Bruker NMR instruments, the pulse program must be changed to change the decoupler mode. The pulse program is the NMR sequence that is executed upon giving the zg command. In the table below are the pulse programs to collect a simple 1D carbon or hetero-nuclei spectrum: Pulse program Decoupler status during d1 Decoupler status during aq Name of mode zgpg30 zggd30 zgig30 zg30 Continuous Gated Inverse-gated The variable pulprog contains the name of the pulse program to be used for acquisition. To change the pulse program, type pulprog and enter a pulse program. For each of the above, the actual pulse length used in the sequence is p1*0.30. Line Broadening The processing parameter, lb, controls the amount of exponential multiplication applied to the FID when it is transformed with the ef command (or the em command). The ft command does not apply exponential multiplication. Carbon spectra are always processed with some amount of line broadening and, thus, the ef command should always be used to transform carbon and other hetero-nuclear spectra. The normal value for carbon spectra is 3. Larger values, up to about 6, result in broader lines but higher signal-to-noise, while smaller values decrease line width (and increase resolution) but increase the noise. Quantitative Spectra Normal carbon spectra are qualitative in that the intensities do not correspond to the number of carbons present. There are two reasons for this: some carbons may not receive the full NOE and different carbons have different relaxation times (T 1 ); carbons with long T 1 s experience saturation which results in reduced intensity. In general, carbons with no directly bonded protons suffer most from reduced NOE and long T 1 s. Quantitative spectra are taken without NOE, decoupled, and with a very long relaxation delay. (This decoupling mode is referred to as inverse-gated.) Change the following parameters to the following values after setting up for a qualitative spectrum: Yufeng Wei, Ph.D. (212) 327-7220 3

pulprog: zgig d1: 60 Quantitative spectra take an extremely long time to acquire and should only be done overnight on concentrated samples. Emphasizing Quaternary and Carbonyl Carbons Without resorting to quantitative spectra, one may adjust the qualitative parameters to emphasize the quaternary carbons (defined in this case as any carbon with no directly bonded protons). First, follow the setup for qualitative spectra. The pulse width should be reduced to approximately 20 (the normal value is 30 ). To calculate the proper value, multiply p1 by 2/3 and enter the result for p1. (Remember the zgpg30 program actually uses p1*0.30 as the actual pulse.) Also, set d1 = 1-2 s. Measuring Coupling Constants The simplest approach to measuring carbon-proton coupling constants is to take the carbon spectra coupled. For sensitivity reasons, these spectra are usually acquired with NOE. (This decoupling mode is referred to as gated.) Setup for a qualitative spectrum and then change the following parameters: pulprog: zggd30 Since the peak intensity is now divided into multiplet instead of singlet, this experiment must run approximately 10 times longer than the qualitative, decoupled one, in order to get equivalent signal-to-noise spectra. Measuring two and three bond carbon-proton couplings, 2,3 J CH, is usually not possible from 1D spectra. Special 2D methods, such as the EXSIDE technique, are available for 2,3 J CH measurements. Determining Multiplicity DEPT The multiplicity refers to the number of directly attached protons, i.e., whether the carbon is a methine (CH), methylene (CH 2 ), or methyl (CH 3 ). In DEPT spectra, all carbons without directly bonded protons are suppressed. The DEPT technique is about as sensitive as a qualitative carbon spectra. There are two useful versions of DEPT spectra. The version is determined by the pulse program: Yufeng Wei, Ph.D. (212) 327-7220 4

dept90: CH only dept135: CH, CH 3 up, CH 2 down The dept135 experiment is the most useful. The phase of the peak gives the multiplicity the CH and CH 3 peaks will point in one direction and the CH 2 peaks will be pointing in the opposite direction. The dept90 ideally gives only CH peaks but in practice this means that CH peaks are enhanced relative to others. Note that with two DEPT spectra, dept90 and dept135, and the qualitative carbon spectra, complete multiplicity information about all carbons is obtained. To run a dept spectrum, setup for a qualitative spectrum (rpar carbon) and change following parameters: pulprog: dept135 (or dept90) d1: 2 ds: 8 szero: 10 This last parameter raises the baseline of the plot so that inverted peaks are not cut off. It is important to tune both the carbon and proton channels of the probe prior to taking a dept spectrum. Dept spectra depend on one-bond carbon-proton couplings that vary from approximately 120Hz (sp3) to 160 Hz (sp2) to over 200 Hz (sp). The sequence can only be optimized for one particular coupling value. The variable cnst2 contains this value and is usually set to 145 Hz, a compromise value. You may change this if, for instance, you are only interested in aliphatic carbons. Automating the process It is sometimes desired to be able to do a qualitative carbon spectrum as well as both dept spectra in series without being present at the spectrometer to start each separate experiment. The command xau multizg solves this problem. It executes multiple acquisitions on sequential experiment numbers. For example, assume your first experiment is <mycarbondata 2 1 x jimmy> while the second is <mycarbondata 3 1 x jimmy> and the third is <mycarbondata 4 1 x jimmy>. When you type xau multizg it will ask you how many experiments you wish to perform. If your current data set is <mycarbondata 2 1 x jimmy> and you enter 3, it will collect data for <mycarbondata 2 1 x jimmy>, <mycarbondata 3 1 x jimmy>, and <mycarbondata 4 1 x jimmy> in this order. Note that the data sets must already exist and have the correct parameters set. Thus, to use this method, do the following: a. Setup for a carbon experiment in the current data set. b. Type edc and increment the experiment number (expno) by one. This creates a new data set and carries over the current parameters (those for the carbon experiment). Change parameters appropriately for a dept spectrum, if desired. c. Repeat step b if desired Yufeng Wei, Ph.D. (212) 327-7220 5

d. Return to the first experiment number (lowest numerically) in this series of experiments, by typing re expno, where expno is the experimental number of the first experiment. Alternatively, you may use the search command. e. Type xau multizg and enter the number of experiments to be performed. Nuclei other than carbon All direct detection heteronuclear NMR experiments are done on the DPX400. The current heternuclear parameter list includes: boron11 carbon deuterium n15 phosphorus silicon Other nuclei are possible. Ask facility director. Some nuclei ( 15 N as an example) have a negative gyro-magnetic ratio. This has implications for the pulse sequence one should use. When the gyro-magnetic ratio is negative, the NOE from protons is opposite in sign to its natural signal. If it happens that the NOE signal is nearly equal to the natural signal, they will cancel and give no apparent signal. Inverse gated decoupling mode (pulprog = zgig30, see above, or DEPT) is preferred to avoid the buildup of NOE for these nuclei. For some nuclei, the INEPT or DEPT experiment can provide much higher signal-to-noise than the standard 1D experiment. This is true only for peaks that have observable couplings to proton. For example, for 29Si and 15N, the DEPT provides 2.5 times more signal than the standard 1D with NOE. Again, it is highly recommended to run 2D heternuclear correlation experiments (HMQC and HSQC) if you are to observed only protonated carbon or nitrogen nuclei, for their much higher sensitivity than direct carbon or nitrogen detection. In general, direct 15 N detection should not be done unless 2D correlation experiments cannot give desired signals. External Referencing Often in heteronuclear NMR, one uses an external reference. This refers to having a chemical shift reference in a separate NMR tube from your sample. If possible, internal referencing is always superior. To insure correct external referencing, proper procedures must be followed. The NMR lock controls the frequency calibration. The lock must know the identity of the lock Yufeng Wei, Ph.D. (212) 327-7220 6

solvent at all times. For each sample, one must lock and choose the correct solvent. For instance, if the reference is in CDCl 3, and the sample is in acetone, after the reference is placed in the magnet, one must lock and choose CDCl 3, and then take the spectrum. For the sample, one must first lock and choose acetone, and then take the spectrum. Sometimes, the lock will simply come on, meaning that it is locked without being told which solvent to lock on. This is wrong and will lead to incorrect referencing. Always re-lock when changing samples when working with external references. An additional complication arises when the solvent has more than one resonance. Toluene, for example, has two sets of peaks. Normally the methyl group is chosen by the lock. The lock, however, may find the aromatic peaks first and lock on these. If this occurs, all referencing will be off by the difference in chemical shift of the two solvent peaks. For this reason, when working in solvents with multiple resonances, one must manually check that the lock has found the correct peak. Yufeng Wei, Ph.D. (212) 327-7220 7