Experiment 7: Adsorption Spectroscopy I, Determination of Iron with 1,10 Phenanthroline

Similar documents
Spectrophotometric Determination of Iron

Experiment 1 (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with 1,10- Phenanthroline

Preparation of Standard Curves. Principle

Expt 4: Determination of Iron by Absorption Spectrophotometry Calibration solutions: Iron(II)ammonium sulfate = ferrous ammonium sulfate

The ROXI Colorimeter & Fluorimeter. Laboratory Application I. Colorimetric measurements via Beer s Law.

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

2 SPECTROSCOPIC ANALYSIS

Answers to spectroscopy questions. 1. Consider the spectrum below. Questions a f refer to this spectrum.

ANALYSIS OF ZINC IN HAIR USING FLAME ATOMIC ABSORPTION SPECTROSCOPY

Chemistry 141 Laboratory Spectrometric Determination of Iron Concentration Lab Lecture Notes 8/29/2011 Dr. Abrash

Lab #12: Determination of a Chemical Equilibrium Constant

DETERMINATION OF AN EQUILIBRIUM CONSTANT

A Fluorometric Analysis of Quinine in Tonic Water

A Study of Beer s Law Prelab

Lab Investigation 4 - How could you make more of this dye?

CHEM Lab 7: Determination of an Equilibrium Constant using Spectroscopy

Experimental Procedure Overview

Investigating Transition Metal Complexes

An Introduction to Ultraviolet-Visible Molecular Spectrometry (Chapter 13)

EXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy

Chem 321 Lecture 18 - Spectrophotometry 10/31/13

Introduction to Spectroscopy: Analysis of Copper Ore

Chem 310 rd. 3 Homework Set Answers

Basics of UV-Visible Spectroscopy *

EXPERIMENT #3 A Beer's Law Study

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

DETERMINATION OF K c FOR AN EQUILIBRIUM SYSTEM

EXPERIMENT 12 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of a Commercial Aspirin Tablet

Spectrophotometric Determination of an Equilibrium Constant

Skill Building Activity 2 Determining the Concentration of a Species using a Vernier Spectrometer

MORE LIGHTS, COLOR, ABSORPTION!

Concepts, Techniques. Concepts, Techniques 9/11/2012. & Beer s Law. For a simple, transparent, COLORED material, e.g. ROYGBV

MASTERING THE VCE 2014 UNIT 3 CHEMISTRY STUDENT SOLUTIONS

Chemistry 213. A KINETIC STUDY: REACTION OF CRYSTAL VIOLET WITH NaOH LEARNING OBJECTIVES

Experiment 6: Determination of the Equilibrium Constant for Iron Thiocyanate Complex

MOLEBIO LAB #4: Using a Spectrophotometer

Beer's Law and Data Analysis *

Spectroscopy of Atoms and Molecules

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

A Spectrophotometric Analysis of Calcium in Cereal

Beers Law Instructor s Guide David T. Harvey

Determining Riboflavin Content in a Multivitamin: Things to do before the lab

EXPERIMENT 6: Photometric Determination of an Equilibrium Constant

Chemistry 1215 Experiment #11 Spectrophotometric Analysis of an Unknown Brass Sample

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013

Spectroscopy Primer. for ultraviolet and visible absorbance spectroscopy. by Stephanie Myers Summer 2015

Spectrophotometry. Introduction

The Determination of an Equilibrium Constant

Rate law Determination of the Crystal Violet Reaction Using the Isolation Method

Experiment 12H, Parts A and B

Introduction to Spectroscopy: Analysis of Copper Ore

SPECTROPHOTOMETRY/BEER S LAW LECTURE HONORS CHEMISTRY NAME

Concentrations that absorb. Measuring percentage transmittance of solutions at different concentrations

EXPERIMENT 5: PHOTOMETRIC DETERMINATION OF AN EQUILIBRIUM CONSTANT

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

FLAME PHOTOMETRY AIM INTRODUCTION

Spectroscopy. Page 1 of 8 L.Pillay (2012)

Experiment 13I THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Brooklyn College Department of Chemistry

Clinical Chemistry (CHE221) Professor Hicks Week 1. Statistics Made Slightly Less Boring and Introduction to Spectrophotometry. Accuracy vs Precision

The Determination of an Equilibrium Constant

Determination of an Equilibrium Constant

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Practical 1 Procedure Iron by Titrimetry

The Fluorometric Determination of Acetylsalicylic Acid in an Aspirin Tablet

The Determination of an Equilibrium Constant

EXPERIMENT 6 INTRODUCTION TO SPECTROSCOPY

A Determination of DNA-DAPI Binding using Fluorescence Spectroscopy

RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION

Colorimetric analysis of aspirin content in a commercial tablet

Exercise 6: Determination of Hardness of Water

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

Experiment 7A ANALYSIS OF BRASS

Introduction to Spectroscopy: Analysis of Copper Ore

CHEM 334 Quantitative Analysis Laboratory

Introduction. The amount of radiation absorbed may be measured in a number of ways: Transmittance, T = P / P 0 % Transmittance, %T = 100 T

Example: How would we prepare 500. ml of M sodium sulfate in water?

Atomic Emission and Molecular Absorption Spectra

1901 Application of Spectrophotometry

For this lab, you will determine the purity of the aspirin by titration and by spectrophotometric analysis.

CHEMISTRY 135 General Chemistry II. Determination of an Equilibrium Constant

REVIEW OF LAB TECHNIQUES

CHEMICAL CALCULATIONS - RELATING MASS AND ATOMS

Determination of an Equilibrium Constant

EXPERIMENT 14. ACID DISSOCIATION CONSTANT OF METHYL RED 1

Introduction to Spectroscopy: Analysis of Copper Ore

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15

CHEM 1471 Kinetics of Phenolphthalein Decolorization Fall 2010 (Buckley) Lab Under Development

DR/4000 PROCEDURE. IRON, Total

CHEMICAL KINETICS E + 2B 2C + D (1)

REVIEW OF LAB TECHNIQUES

aa + bb cc + dd Equation 1

1iI1E. The Determination of 0 an Equilibrium Constant [LU. Computer

A Fluorometric Analysis of Quinine in Tonic Water

Spectrophotometric Determination of an Equilibrium Constant 1

1. Preliminary qualitative analysis of unknown substances (liquid or solid).

Clinical Chemistry (CHE 221) Experiment # 3 Analysis of rare body fluids by UV-Vis Absorption Spectroscopy

Bio 120 Lab 5: Quantitative Analysis

Laboratory Measurements and Procedures

Transcription:

Procedure Experiment 7: Adsorption Spectroscopy I, Determination of Iron with 1,10 Phenanthroline John Smith CHE226.001 4/10/95 In this experiment, we aim to determine the Na concentration of an unknown sample by measuring an emission intensity of unknowns and comparing the measurements to a calibration curve created from solutions of known concentrations. First, turn in a 100mL volumetric flask to TA for unknown at least one day before doing the experiment. Weigh out about 0.07g of ferrous ammonium sulfate, dissolve in water, and transfer to a 1000mL volumetric flask. Add 2.5mL of concentrated sulfuric acid and dilute to the mark. Now, prepare the standards. Using a pipet place 1, 5, 10, 20, and 35ml of standard iron solution into five 100mL volumetric flasks. Also, place 50mL of H 2 O in a sixth flask (for blank). Treat all samples (including unknown) in the following manner: pipet 1mL of hydroxylamine solution, 10mL of 1,10 phenathraline solution and 8mL of sodium acetate solution into each flask and dilute to the mark. Mix and allow to stand for 10 min. Now, on the Spec 20 set the %T to zero with no cuvette in the compartment, using the disk on the left front. Set the %T to 100 using the right front dial, with the blank in the compartment. Recheck these settings frequently. Set the wavelength at 505nm (instructed to do so by Shawn) and record the %T within 0.2% for each of the standards and the usual unknown. Calculate the absorbance using A= log(%t/100) and the concentration for the standard. Plot a graph of absorbance vs. concentration of the standards using Tablecurve or Excel. For detailed procedure refer to: Handout, Experiment 7, Absorption Spectroscopy I: Determination of Iron with 1,10 phenanthroline. For further information on adsorption spectroscopy refer to: Skoog, Holler, and West, Chapters 20,21, and 22. Pp. 383 439. Theory Diagram of Spec 20 Narrow band Red of λ Light Source Slit Diffraction violet Detector gradient Sample 0 %T 100

In preparing the standards and the unknown the reaction of the ferrous iron with 1,10 phenanthroline yielded a red complex that served as a good sensitive method for determining iron: The molar absorptivity of this complex is 11,100 at 508nm. Prior to the development of the redorange color, hydroxylamine is added as a reducing agent to put the iron the ferrous state: The sodium acetate is added to adjust the ph to between 6 and 9. The color intensity remains relatively constant between ph 2 and 9. During the analysis of the samples, the spec 20 is calibrated by using the blank sample which contained all species except the iron solution. By doing this, we hoped to minimize the effects caused by reflection, absorption, ect. due to the presence of the other elements (ie, the cuvette and solvent). Failure to use a blank would have caused inaccurate % transmittance readings. Now, by placing the subsequent standards in the spec 20, the decreased %T was due only to the presence of the iron solution. In passing through the cuvette the power of the light beam is decreased as a result of the interactions between the photons and absorbing particles (the iron solution). The %T reading on the spec 20 results from the following ratio: Where p is the power of the beam prior to entering the cuvette and P o is the power after passing through. The absorbance was calculated from the %T using the equation: It was not read from the scale on the Spec 20. Beer s law states that absorbance is linearly related to the generation of the absorbing species, c, and the path length, b, of the radiation in the absorbing medium. (Skoog, 405). In my graph of absorbance vs. concentration, I used Tablecurve. In asking it to draw a curve that best fit my data points, it produced an exponential curve as its first choice, which obviously is not a linear graph. I went and showed the graph to Dr. Holler (After Shawn told me to get some assistance in making it a linear curve). He, at first, started to help me redo the graph. Then, he thought since the exponential graph was in fact the best fit, he said solve for c and use it. So, technically my graph is not linear, which would suggest it did not obey Beer s Law. However the curve to the exponential is very slight, so essentially Beer s Law was obeyed. Structure of 1,10 phenanthroline

Data Amount of ferrous ammonium sulfate: 0.0724g %Transmittance at λ=505nm Conc(ppm) %T 0.103 92.8% 0.515 76.6% 1.03 57.1% 2.06 32.3% 3.61 15.1% Unknown 34.0% Absorbance at 505nm Conc(ppm) Abs 0.103 0.0324 0.515 0.116 1.03 0.243 2.06 0.491 3.61 0.821 Unknown 0.468 Calibration Curve

Calculations Concentration of ferrous ammonium sulfate solution (FAS) Calculation of absorbance: Concentration of diluted standards: Concentration of unknown, determined from graph:

Molarity of unknown: Results and Error Analysis Mean absorbance of unknown: 0.468 C Fe2+ of unknown: 3.56x10 5 M Parts per million: 1.99ppm I have no numerical error analysis for this experiment. Error sources will be discussed in discussion. Discussion In this experiment I determined the iron concentration in my unknown solution. I found that the concentration for my unknown was 3.56x10 5 M which is equivalent to 1.00ppm. This value was determined by obtaining data from solutions of known concentration. These solutions were placed in the spec 20 and a %Transmittance reading was produced. The %T was used to determine the absorbance for the standards and the unknown. Then by plotting a graph of absorbance vs. concentration, an equation was produced from which the unknown concentration was determined (the eq. was solved for x concentration, which was the only unknown quantity in the equation). Error may have resulted from the fact that two different cuvettes were used. One was used for the blank solution (because It was used for frequent readings) and the other for the various standards and the unknown. Although the two cuvettes are very similar in all their characteristics, they are not exactly alike. Therefore, the amount of light reflected, scattered, and absorbed, ect. is different and thus, the %T is somewhat affected. In short, setting the 100%T with the cuvette containing the blank solution might have been different that if the cuvette used for the samples was also used for the blank solution.

Additional error may have resulted from pipetting slightly different amounts of each reagent inot the different flasks. Naturally, there will always be some type of deviation when pipetting the same quantity into each flask. This deviation may have caused an increased or decreased amount of absorbance by solution components not of interest. For example, if the contents of the blank solution were slightly different than the 1mL standard the %T would be slightly affected because of the increased or decreased presence of other components. This error can be minimized by simply being careful in your pipetting techniques. The first error mentioned could be minimized by using only one cuvette for all samples, including the blank.