A Quadrupole-Orbitrap Hybrid Mass Spectrometer Offers Highest Benchtop Performance for In-Depth Analysis of Complex Proteomes

Similar documents
HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Relative quantification using TMT11plex on a modified Q Exactive HF mass spectrometer

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

Targeted protein quantification

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

HR/AM Targeted Peptide Quantitation on a Q Exactive MS: A Unique Combination of High Selectivity, Sensitivity and Throughput

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Peptide Targeted Quantification By High Resolution Mass Spectrometry A Paradigm Shift? Zhiqi Hao Thermo Fisher Scientific San Jose, CA

Q Exactive TM : A True Qual-Quan HR/AM Mass Spectrometer for Routine Proteomics Applications. Yi Zhang, Ph.D. ThermoFisher Scientific

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Increasing the Multiplexing of Protein Quantitation from 6- to 10-Plex with Reporter Ion Isotopologues

MetWorks Metabolite Identification Software

Quan/Qual Analyses. Unmatched Confidence for. Thermo Scientific Q Exactive Orbitrap LC-MS/MS System. Identify Quantify Confirm

High-Field Orbitrap Creating new possibilities

Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 *

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Self-assembling covalent organic frameworks functionalized. magnetic graphene hydrophilic biocomposite as an ultrasensitive

of mass spectrometry

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Relative Quantitation of TMT-Labeled Proteomes Focus on Sensitivity and Precision

Quantitation of TMT-Labeled Peptides Using Higher-Energy Collisional Dissociation on the Velos Pro Ion Trap Mass Spectrometer

Thermo Scientific Pesticide Explorer Collection. Start-to-finish. workflows for pesticide analysis

High-Resolution Accurate-Mass (HRAM) Phthalate Screening using Direct Analysis in Real Time (DART) Ambient Ionization

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

Improving Intact Antibody Characterization by Orbitrap Mass Spectrometry

Modeling Mass Spectrometry-Based Protein Analysis

Introduction to Fourier Transform Infrared Spectroscopy

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

for XPS surface analysis

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Key Words Q Exactive Focus, Orbitrap, veterinary drugs, small molecule HRAM quantitation, small molecule HRAM screening, UHPLC, vdia

Rapid Distinction of Leucine and Isoleucine in Monoclonal Antibodies Using Nanoflow. LCMS n. Discovery Attribute Sciences

Mass spectrometry-based proteomics has become

A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzcloud Library

Designed for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS

Dual-Pressure Linear Ion Trap Mass Spectrometer Improving the Analysis of Complex Protein Mixtures

Quantitative and Qualitative Confirmation of Pesticides in Beet Extract Using a Hybrid Quadrupole-Orbitrap Mass Spectrometer

Fast, Effective XPS Point Analysis of Metal Components

An Evaluation of Various High-Resolution, Accurate-Mass Scan Modes for In Vitro Drug Discovery Screening

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Thermo Scientific LTQ Velos Dual-Pressure Linear Ion Trap

Fast Protein and Peptide Separations Using Monolithic Nanocolumns and Capillary Columns

Powerful Scan Modes of QTRAP System Technology

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry

TOMAHAQ Method Construction

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

The Power of LC MALDI: Identification of Proteins by LC MALDI MS/MS Using the Applied Biosystems 4700 Proteomics Analyzer with TOF/TOF Optics

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Tandem mass spectra were extracted from the Xcalibur data system format. (.RAW) and charge state assignment was performed using in house software

Analysis of a Verapamil Microsomal Incubation using Metabolite ID and Mass Frontier TM

Introduction to Fourier Transform Infrared Spectroscopy

Methods for proteome analysis of obesity (Adipose tissue)

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Overview - MS Proteomics in One Slide. MS masses of peptides. MS/MS fragments of a peptide. Results! Match to sequence database

Guide to Peptide Quantitation. Agilent clinical research

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

PosterREPRINT COMPARISON OF PEAK PARKING VERSUS AUTOMATED FRACTION ANALYSIS OF A COMPLEX PROTEIN MIXTURE. Introduction

HOWTO, example workflow and data files. (Version )

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Protein Quantitation II: Multiple Reaction Monitoring. Kelly Ruggles New York University

IN QUALITATIVE ANALYSIS,

Monitoring Protein PEGylation with Ion Exchange Chromatography

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software

Thermo Scientific K-Alpha + XPS Spectrometer. Fast, powerful and accessible chemical analysis for surface and thin film characterization

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Protein Quantitation II: Multiple Reaction Monitoring. Kelly Ruggles New York University

Direct Analysis using Paper-Spray Mass Spectrometry: Method Development for the Rapid Screening of Drugs of Abuse for Forensic Toxicology

Figure S1. Interaction of PcTS with αsyn. (a) 1 H- 15 N HSQC NMR spectra of 100 µm αsyn in the absence (0:1, black) and increasing equivalent

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS

Key Words Nanoparticles, spicp-ms, ICP-MS, Qtegra ISDS Software, AF4-ICP-MS

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

Analyst Software. Peptide and Protein Quantitation Tutorial

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Charles Yang, Dipankar Ghosh, Jonathan Beck Thermo Fisher Scientific, San Jose, CA, USA

MS Amanda, a Universal Identification Algorithm Optimized for High Accuracy Tandem Mass Spectra

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness

T H E J O U R N A L O F C E L L B I O L O G Y

Analysis of Labeled and Non-Labeled Proteomic Data Using Progenesis QI for Proteomics

4000 Q TRAP LC/MS/MS System. Advanced Linear Ion Trap technology at the highest level of sensitivity 4000 QTRAP. LC/MS/MS System

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Proteome-wide label-free quantification with MaxQuant. Jürgen Cox Max Planck Institute of Biochemistry July 2011

Transcription:

A Quadrupole-Orbitrap Hybrid Mass Spectrometer Offers Highest Benchtop Performance for In-Depth Analysis of Complex Proteomes Zhiqi Hao 1, Yi Zhang 1, Shannon Eliuk 1, Justin Blethrow 1, Dave Horn 1, Vlad Zabrouskov 1, Markus Kellmann 2, and Andreas F. Huhmer 1 1 Thermo Fisher Scientific, San Jose, CA, USA; 2 Thermo Fisher Scientific, Bremen, Germany Application Note 552 Key Words Q Exactive, Orbitrap, TripleTOF 5600, Protein ID, Yeast Goal To evaluate the performance of a Q Exactive mass spectrometer for peptide identification, particularly its ability to access low-abundance proteins in a complex proteome. Introduction Mass spectrometry has become a widely used powerful tool for high-throughput analysis of complex proteomes 1,2. Recent advances in mass spectrometry have made it possible to identify several thousand proteins and their modifications in complex proteomes in a single LC-MS experiment 3,4. However, for routine analysis of extremely complex biological mixtures, two major challenges remain: sample complexity and range of protein abundance that often exceeds the capability of many currently available instruments. The Thermo Scientific Orbitrap mass analyzer, which was first described in 2000 5, has now become a mainstream analyzer because of its high sensitivity, high resolution and accurate-mass capabilities. In complex matrices, a single MS survey spectrum can contain hundreds of different, co-eluting molecular species of which only a fraction are often selected for identification by datadependent MS/MS. Adding to the challenge is the need to detect very low-abundance species in the presence of highly abundant ones. Consequently, a wide dynamic range in the survey scan, a fast sampling rate for MS/MS generation with improved sensitivity, and efficient fragmentation that produces high-quality spectra for low-abundance species are required for identification of very low-abundance proteins, which are often of particular biological interest due to their importance in cellular regulatory pathways 6. Here we introduce the newly developed Thermo Scientific Q Exactive hybrid quadrupole-orbitrap mass spectrometer (Figure 1). Coupling a quadrupole as a front-end mass filter to an Orbitrap mass analyzer, the Q Exactive instrument is capable of precursor ion selection for both selected ion monitoring (SIM) and data-dependent MS/MS experiments. The quadrupole mass filter allows transfer of ions of only a specified mass range into the c-trap for accumulation, thus improving sensitivity for both SIM and MS/MS experiments. In addition to the common features of an Orbitrap mass analyzer such as high mass accuracy, large inter- and intra-scan dynamic range, and low detection limits, the Q Exactive instrument s advanced signal processing provides ultra-high resolution of 140,000 FWHM at m/z 200 and higher scan speeds at all resolution settings. In addition, parallel ion filling / fragmentation and Orbitrap detection improves practical acquisition rates, allowing up to 12 high-quality MS/MS events per second. The multiplexing scan function, which performs a single Orbitrap detection for multiple precursors isolated individually, greatly improves throughput and detection limits for targeted analyses 7. Finally, the S-lens increases the transmission of ions into the Q Exactive instrument, increasing the scan rate by reducing the ion injection time or increasing sensitivity by injecting more ions in those instances when the ion injection time has reached its maximum (max ion time). In this study, a Q Exactive MS was evaluated for peptide identification from a yeast total cell lysate. The Q Exactive instrument s ability to access low-abundance proteins in a complex proteome was investigated. The performance of the Q Exactive instrument was benchmarked against a TripleTOF 5600 mass spectrometer (AB Sciex). HCD Cell for Precursor Ion Fragmentation C-Trap for Ion Accumulation in Parallel with Orbitrap Detection Orbitrap with Enhanced Resolution Quadrupole Mass Filter for Precursor Ion Selection S-Lens for Improved Sensitivity Bent Flatapole Ion Transfer Tube Figure 1. Schematic representation of the Q Exactive quadrupole mass spectrometer

2 Experimental Sample Preparation Yeast (Saccharomyces cerevisiae) cell lysate tryptic digest was a courtesy of Professor Steve Gygi at Harvard University. Liquid Chromatography A Thermo Scientific Easy-nLC II HPLC system was used in a single-column configuration for all Q Exactive MS experiments. A 75-μm x 15-cm, 3-μm particle C18 packedtip nano-lc column was used with a 110-minute gradient. A Thermo Scientific Acclaim PepMap C18 column (75-μm x 15-cm, 3-µm particle size) with a 75-minute gradient was used with both the Q Exactive MS and TripleTOF 5600 MS for the benchmark study. Mass Spectrometry The Q Exactive instrument was operated using a datadependent top-10 experiment with 70K resolution for the full MS scans, 17.5K resolution for high energy collisional dissociation (HCD) MS/MS scans and a dynamic exclusion of 30 seconds. Detailed instrument parameters are listed in Table 1. The TripleTOF 5600 was run by a skilled operator in a core lab according to the manufacturer s recommendations for this type of analysis. The instrument was operated using a data-dependent top-20 experiment with 30K resolution and 250-ms beam time for the full MS, and 15K resolution and 50-ms beam time for the MS/MS. The instrument was externally calibrated every 4 hours using a dedicated LC/MS run. Table 1. Q Exactive method parameters Parameters Settings Full MS scan range 400-2000 MS/MS fixed first mass 100 AGC Max injection time (ms) 1e6, full MS 5e4, MS/MS 2, full MS 60, MS/MS Isolation width (m/z) 2.0 NCE 27 Under fill ratio 1% Dynamic Exclusion 30 sec Data Analysis Thermo Scientific Proteome Discoverer software version 1.3 with the MASCOT and/or the SEQUEST search engine was used for all searches of the yeast database. Raw files generated by the Q Exactive instrument were searched directly using a 10-ppm precursor mass tolerance and a 20-mmu fragment mass tolerance. Raw files from the TripleTOF 5600 were processed into.mgf files using ProteinPilot 4.0 software and searched with MASCOT against the yeast database using a 10-ppm precursor mass tolerance and a 50-mmu fragment mass tolerance. Ion injection times for each of the full-ms and MS/MS scans of an entire Q Exactive LC-MS/MS run were generated using RawMeat software (Vast Scientific) version 2.0. Results and Discussion Protein abundance levels were previously experimentally determined for the entire yeast proteome of more than 4000 proteins through genome-wide tagging experiments 8. Since then, yeast tryptic digests have been used routinely as a standard complex mixture for research purposes as well as for benchmarking instrument performance by several leading proteomics research groups 9,10,11. Traditionally, for in-depth analysis of whole-cell lysates, protein and/or peptide fractionation is employed prior to LC-MS/MS. In the past few years, the resolution, mass accuracy and MS/MS scan rate of mass spectrometers have improved significantly, often allowing for in-depth proteome analysis without extensive sample fractionation 12,13. Here we used a single-dimension LC-MS/MS experiment to investigate the performance of the Q Exactive instrument for profiling the yeast proteome. Evaluation of Q Exactive Performance for Protein Identification As shown in Figure 2, nearly 2000 protein groups and over 12,000 unique peptides were identified at 1% FDR from 1 µg of yeast digest using an online one-dimensional separation LC-MS/MS run with 110-minute on-column chromatographic separation time. Combining results of two replicate experiments produced 2180 protein groups and over 15,000 unique peptides (Table 2). When the amount of sample loaded on the column was reduced to 10 ng of yeast digest, more than 1000 protein groups and 6000 unique peptides were still identified (Figure 2). The MS/MS success rate (percentage of MS/MS scans that resulted in confident identifications) for the Q Exactive instrument was approximately 50% for the 1-µg sample load (Table 2). The high success rate is attributed to the high quality of the MS/MS scans. Protein groups Unique peptides Figure 2. Number of proteins and peptides identified from varying amounts of yeast whole-cell lysate digests on the Q Exactive MS

Table 2. Number of identifications and MS/MS success rate from 1000 ng of yeast digest using two search engines 3 File 1000 ng yeast digest_01 1000 ng yeast digest_02 Combined (All 4 searches) Search engine # MS/MS spectra PSMs* Protein groups Unique peptides Success rate, % SEQUEST 33022 17597 1898 12340 53.3 Mascot 33022 16805 1846 11878 51.0 Combined 66044 34402 1945 12612 52.1 SEQUEST 33257 17236 1906 12204 51.8 Mascot 33257 16435 1847 11726 49.4 Combined 66514 33671 1930 12466 50.6 Combined 132558 68073 2180 15270 51.3 *PSM (peptide spectrum match) is the number of spectra that resulted in 1% FDR identification. MS/MS sensitivity The higher sensitivity of the MS/MS scans on the Q Exactive instrument is achieved via the incorporation of an S-lens, which provides higher transmission of all ions into the mass spectrometer. The increase in ion flux both 1) improves the duty cycle/scan rate by lowering the parent ion injection time for high intensity ions and 2) increases the sensitivity for low-intensity ions by increasing the number of precursor ions in the Orbitrap mass analyzer for a given maximum ion injection time. Analysis of the raw data files indicated that the average ion injection times in full MS and MS/MS were only 0.5 msec and 36 msec, respectively, when 1 µg of yeast digest was analyzed. With all Thermo Scientific ion trap and Orbitrap instruments, ion injection time is automatically adjusted depending on the ion flux using the patented automatic gain control (AGC) technique. This dynamic control of ion accumulation, unlike the fixed beam time on Q-TOF instruments, allows the ion accumulation time to be inversely proportional to parent ion abundance. This means that the quality of the MS/MS spectra is the same for low-abundance ions as high-abundance ions and that the spectra usually look similar at the bottom of the chromatographic peak as they do at the top of the chromatographic peak. Dynamic range The AGC function allows the Q Exactive instrument to effectively identify peptides from complex samples with a large dynamic range of abundances. As a result, a dynamic range of >100,000 was achieved for precursor intensity of identified peptides from yeast digests when 1 µg of sample was analyzed (Figure 3). With 10 ng of yeast digest, the Q Exactive instrument reached a dynamic range of over 4 orders of magnitude for peptide precursor intensity. Mass accuracy Characteristic of Orbitrap mass analyzers, the data generated by the Q Exactive mass spectrometer has very high mass accuracy. As shown in Figure 4, the calculated root mean square (RMS) of delta mass between the theoretical and experimental masses for all of the peptides identified at a 1% FDR in 10 ng of yeast digest was 1.63 ppm. This high mass accuracy ensured confident identification of peptides. Figure 3. The Q Exactive mass spectrometer achieved 4 orders and 5.5 orders of inter-scan dynamic range at 10-ng and 1000-ng sample loads, respectively. Data in this figure spreads out in the Y-axis direction over normalized MASCOT ion score. Ion Score 140 120 100 80 60 40 20 RMS = 1.63 PPM 0-15 -10-5 0 5 10 15 Delta Mass [ppm] Figure 4. Distribution of mass error of peptides identified at 1% FDR in 10 ng of yeast digest by the Q Exactive MS. The graph was generated by Proteome Discoverer software version 1.3. Calculated RMS is 1.63 ppm.

Data acquisition rate Another significant feature of the Q Exactive instrument is its advanced signal processing which leads to 1.8-fold improvement in resolution for a given acquisition speed. This allows for scan rates of up to 12 Hz at 17.5K resolution. In addition, Orbitrap detection and ion injection into the C-trap for the next detection cycle are performed concurrently. Thus, the practical duty cycle is only limited by the duration of the Orbitrap detection. Figure 5 shows the actual duty cycle of the data-dependent top-10 HCD method that was used for this study. A cycle including 1 full MS scan at 70K resolution and 10 datadependent HCD MS/MS scans at 17.5K resolution was completed in 1.1 sec. The 10 HCD MS/MS spectra were acquired in only 0.83 seconds of this 1.1 sec total cycle time. Most importantly, the fast HCD MS/MS generated extremely high quality spectra that resulted in confident identification of low-abundance peptides. The improved scan speed and sensitivity combined with the other benefits of the Orbitrap mass analyzer provide the basis of the Q Exactive instrument s outstanding performance for protein identification. Performance comparison of Q Exactive and TripleTOF 5600 instruments To further benchmark the Q Exactive MS for protein identification, especially in identifying low-abundance proteins, a series of carefully controlled comparison experiments was conducted using the Q Exactive instrument in our lab and a TripleTOF 5600 run by a skilled operator in a core lab. The LC separation time in this study was 75 minutes for both instruments (described in the Experimental section). To ensure a fair and unbiased comparison of the instrument performance, exactly the same sample, LC column, and LC conditions were used for the analyses on the two platforms. Both the Q Exactive instrument and the TripleTOF 5600 were operated under the conditions that were optimum for this type of study based on manufacturer recommendations. The results of this comparison are shown in Figures 6, 7 and 8. 4 100 90 Relative Abundance 80 70 60 50 40 30 20 10 0 10 20 30 40 50 60 70 80 90 100 Time (min) 1.5 33.8222 One full scan 10 HCD MS/MS in 1.1 sec 33.8409 Figure 6. Numbers of protein and peptide identifications from various amounts of yeast whole cell lysate digests using the Q Exactive MS and TripleTOF 5600 MS Relative Abundance 1.0 0.5 33.8308 33.8283 33.8270 10 HCD in 0.83 sec 33.8377 33.8457 33.8320 33.8350 33.8394 33.8365 33.8474 0.0 33.820 33.825 33.830 33.835 33.840 33.845 Time (min) Figure 5. Practical duty cycle of Q Exactive mass spectrometer for peptide identification from whole cell lysate. Top: base peak LC-MS/MS chromatograph of 1000 ng of yeast digest. Bottom: stick view of scan events showing time window of a complete cycle of one full scan and 10 HCD scans. Figure 7. Comparison of inter-scan dynamic range of identified precursor abundances between the Q Exactive instrument and the TripleTOF 5600. Data in this figure spreads out in the Y-axis direction over normalized MASCOT ion score.

Dynamic range comparison As shown in Figure 7, the dynamic range of precursor intensity for peptides identified by the Q Exactive mass spectrometer is at least one order of magnitude larger than that generated by the TripleTOF 5600. This suggests that, compared to TripleTOF 5600, the Q Exactive instrument not only has lower detection limits in full-scan MS thus giving access to more low-abundance precursors, but also has higher sensitivity MS/MS, generating higher quality fragmentation spectra from those precursors. 5 Intensity [counts] (10^3) 50 40 30 20 10 0 y 1 175.11826 y 2 322.18576 y 3 435.26990 L G N D D E V I L F R b 5 - HN3 y 4 548.35284 y 5 647.42157 y 6 - H 2 O 498.17874 758.45410 776.46594 891.48932 MASCOT IonScore: 90 Exp Value: 3.8 E-9 1006.51752 1120.55273 1177.57886 200 400 600 800 1000 1200 m/z Figure 8. Comparison of sequencing depth between the Q Exactive mass spectrometer and the TripleTOF 5600 using 10 ng of yeast digest. Top: Distribution of identified proteins over cellular copy number. Bottom: Spectrum of a peptide of yeast protein YOR020C, with estimated 149 copies per cell, identified from 10-ng yeast digest using the Q Exactive instrument. Peptide and protein ID comparison As shown in Figure 6, the Q Exactive mass spectrometer consistently identified significantly more proteins and peptides in 10-ng, 100-ng and 1000-ng loads of yeast digest. The difference in the number of identifications was largest for the lowest sample load. This is due to the ability of the Q Exactive instrument to accumulate lowabundance ions and will be most profoundly evident when the amounts of peptides are at relatively low levels. From 10 ng of yeast digest, the Q Exactive instrument identified 691 proteins and 2540 unique peptides, nearly twice as many as were identified by the TripleTOF 5600. The Q Exactive instrument required less than 100 ng to identify the same number of proteins identified by the TripleTOF 5600 in 1000 ng. The Q Exactive mass spectrometer also demonstrated a much higher MS/MS success rate than the TripleTOF 5600. At 1 µg of sample load, 50% of the MS/MS spectra generated by the Q Exactive instrument resulted in a 1% FDR identification while only 24% of the TripleTOF 5600 MS/MS spectra resulted in confident identifications. The TripleTOF 5600 achieved a 7-ppm RMS mass accuracy for the identified peptides in both MS and MS/MS scans, suggesting that mass accuracy is not the limiting factor in the performance of this instrument. y 6 y 7 y 8 y 9 y 10 Identification of low-abundance proteins comparison The largest performance difference between the Q Exactive mass spectrometer and the TripleTOF 5600 in protein identification experiments was on the lower end of the range of cellular protein abundance (Figure 8). While both instruments identified about the same number of highabundance proteins, the Q Exactive instrument identified four times more proteins that were below 10,000 copies per cell (Figure 8, top). The bottom panel of Figure 8 shows a Q Exactive MS/MS spectrum of a peptide that belongs to the YOR020C protein; a protein with an experimentally determined cellular copy number of 149. 8 This high quality spectrum, acquired from 10 ng of the yeast digest, contains every ion in the y ion series and three additional b ions. This demonstrates the ability of the Q Exactive MS to confidently identify low-abundance proteins from minute sample loads of complex mixtures. Conclusions The performance of the Q Exactive hybrid mass spectrometer was evaluated for bottom-up protein identification in complex mixtures. In discovery experiments, the Q Exactive instrument demonstrated a practical acquisition speed of up to 12 high-quality MS/MS scans per second at 17,500 resolution. With 110-minute, single-column LC-MS/MS experiments, the Q Exactive instrument identified more than 1,000 protein groups and 6,000 unique peptides from 10 ng of yeast digest, and 2,000 protein groups and more than 12,000 unique peptides from 1 µg of yeast digest. Data generated by the Q Exactive MS achieved an average mass accuracy of <2 ppm RMS. The Q Exactive instrument also provided more than 5 orders of inter-scan dynamic range of identified peptides, which was at least one order of magnitude better than the dynamic range provided by the TripleTOF 5600 mass spectrometer used for comparison. Results from our carefully conducted comparison not only indicated that the Q Exactive instrument significantly out-performed the TripleTOF 5600 for identification of proteins in complex samples but, also revealed that the performance difference between the two instruments lies in the Q Exactive instrument s far superior ability to identify low-abundance proteins. The advantages of the Orbitrap mass analyzer such as high resolution, accurate mass and large dynamic range, combined with the increased sensitivity and acquisition rate, enable the Q Exactive MS to deliver more information to researchers looking to understand important processes in complex biological mixtures.

References 1. Hilger, M.; Bonaldi, T.; Gnad, F.; Mann, M. Systems-wide analysis of a phosphatase knock-down by quantitative proteomics and phosphoproteomics. Mol. Cell. Proteomics 2009, 8,1908-1920. 2. Dephoure, N.; Zhou, C.; Villén, J.; Beausoleilm S.A.; Bakalarski, C.E.; Elledge, S.J.; Gygi, S.P. A quantitative atlas of mitotic phosphorylation. Proc. Natl. Acad. Sci. USA 2008, 105, 10762 10767. 3. de Godoy, L.M.; Olsen, J.V.; Cox, J.; Nielsen, M.L.; Hubner, N.C.; Fröhlich, F.; Walther, T.C.; Mann, M. Comprehensive mass-spectrometry-based proteome quantification of haploid versus diploid yeast. Nature 2008, 455, 1251 1254. 4. Merrihew, G.E.; Davis, C.; Ewing, B.; Williams, G.; Käll, L.; Frewen, B.E.; Noble, W.S.; Green, P.; Thomas, J.H.; MacCoss, M.J. Use of shotgun proteomics for the identification, confirmation, and correction of C. elegans gene annotations. Genome Res 2008, 18,1660 1669. 5. Makarov, A. Electrostatic axially harmonic orbital trapping: A high performance technique of mass analysis. Anal. Chem. 2000, 72, 1156-1162. 6. Mann, M. and Kelleher, N.L. Precision proteomics: The case for high resolution and high mass accuracy. Proc. Natl. Acad. Sci. USA. 2008, 105, 18132-18138. 7. Zhang, Y.; Hao, Z.; Kellmann, M. and Huhmer, A. HR/AM targeted peptide quantification on Q Exactive: A unique combination of high selectivity, high sensitivity and high throughput. Thermo Fisher Scientific Application Note to be published. 8. Ghaemmaghami, S.; Huh, W.K.; Bower, K.; Howson, R.W.; Belle, A.; Dephoure, N.; O Shea, E.K.; Weissman, J.S. Global analysis of protein expression in yeast. Nature 2003, 16, 737 7341. 9. de Godoy, L. M.; Olsen, J. V.; Cox, J.; Nielsen, M. L.; Hubner, N. C.; Frohlich, F.; Walther, T. C.; and Mann, M. Comprehensive mass-spectrometry-based proteome quantification of haploid versus diploid yeast. Nature 2008,455, 1251-1254. 10. Michalski, A.; Damoc, E.; Hauschild, J.P.; Lange, O.; Wieghaus, A.; Makarov, A.; Nagaraj, N.; Cox, J.; Mann, M.; Horning, S. Mass spectrometry-based proteomics using Q Exactive, a high-performance benchtop quadrupole Orbitrap mass spectrometer. Mol Cell Proteomics. 2011, 10, M111.011015. 11. Washburn, M.P.; Wolters, D.; Yates, J.R. 3rd. Largescale analysis of the yeast proteome by multidimensional protein identification technology. Nat Biotechnol, 2001, 19, 242-247. 12. Domon, B. and Aebersold, R. Mass spectrometry and protein analysis. Science, 2006, 312, 212-217. 13. Lange, O.; Hauschild, J.P.; Makarov, A.; Fröhlich, U.; Crone, C.; Xuan, Y.; Kellmann, M.; Wieghaus, A. ASMS Conference 2011, Poster 437. Application Note 552 www.thermofisher.com 2016 Thermo Fisher Scientific Inc. All rights reserved. TripleTOF is a registered trademark of AB Sciex PTE. LTD. Mascot is a trademark of Matrix Science Ltd. SEQUEST is a registered trademark of the University of Washington. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa-Other 27 11 570 1840 Australia 61 2 8844 9500 Austria 43 1 333 50 34 0 Belgium 32 53 73 42 41 Canada 1 800 530 8447 China 86 10 8419 3588 Denmark 45 70 23 62 60 Europe-Other 43 1 333 50 34 0 AN63515_E 08/16S Finland /Norway/Sweden 46 8 556 468 00 France 33 1 60 92 48 00 Germany 49 6103 408 1014 India 91 22 6742 9434 Italy 39 02 950 591 Japan 81 45 453 9100 Latin America 1 608 276 5659 Middle East 43 1 333 50 34 0 Netherlands 31 76 579 55 55 South Africa 27 11 570 1840 Spain 34 914 845 965 Switzerland 41 61 716 77 00 UK 44 1442 233555 USA 1 800 532 4752