Analysis by laser-induced breakdown spectroscopy of complex solids, liquids, and powders with an echelle spectrometer

Similar documents
Comparative study of two new commercial echelle spectrometers equipped with intensified CCD for analysis of laser-induced breakdown spectroscopy

Comparisons between LIBS and ICP/OES

Single pulse laser-induced breakdown spectroscopy of bulk aqueous solutions at oceanic pressures: interrelationship of gate delay and pulse energy

Laser-Induced Breakdown Spectroscopy: Application to Nuclear Waste Management Seong Yong Oh, Fang Yu Yueh, Jagdish P.

Approach to Detection in Laser-Induced Breakdown Spectroscopy

Oceanic Applications of Laser Induced Breakdown Spectroscopy: Laboratory Validation

Laser-Induced Breakdown Spectroscopy for space exploration. applications : Influence of the ambient pressure on the calibration

Quantitative Analysis of Carbon Content in Bituminous Coal by Laser-Induced Breakdown Spectroscopy Using UV Laser Radiation

Calibration free laser-induced breakdown spectroscopy (LIBS) identification of seawater salinity

LASER-INDUCED BREAKDOWN SPECTROSCOPY FOR REAL TIME AND ONLINE ELEMENTAL ANALYSIS

Laser Ablation for Chemical Analysis: 50 Years. Rick Russo Laser Damage Boulder, CA September 25, 2012

Outline. LIBS Background. LIBS Developments. LIBS Overview. Atomic Emission Spectroscopy

Accuracy improvement of quantitative analysis by spatial confinement in laser-induced breakdown spectroscopy

LIBSlab ANALYZERS ANALYZERS

Enhancement of optical emission from laserinduced plasmas by combined spatial and magnetic confinement

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

JAAS Accepted Manuscript

Optimally enhanced optical emission in laserinduced breakdown spectroscopy by combining spatial confinement and dual-pulse irradiation

Generation of high-temperature and low-density plasmas for improved spectral resolutions in laser-induced breakdown spectroscopy

Measurements of plasma temperature and electron density in laser-induced copper plasma by time-resolved spectroscopy of neutral atom and ion emissions

Evaluation of laser-induced breakdown spectroscopy (LIBS) as a new in situ chemical sensing technique for the deep ocean

Chemical warfare detection by LIBS

Influence of laser energy on the electron temperature of a laser induced Mg plasma

Correction of Matrix Effects in Quantitative Elemental Analysis With Laser Ablation Optical Emission Spectrometry

Study of the Matrix Effect on the Plasma Characterization of Heavy Elements in Soil Sediments using LIBS with a Portable Echelle Spectrometer

Absolute characterization of laser-induced breakdown spectroscopy detection systems

Time-Resolved Emission Spectroscopic Study of Laser-Induced Steel Plasmas

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Effect of applying static electric field on the physical parameters and dynamics of laser-induced plasma

OES - Optical Emission Spectrometer 2000

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Lead and Arsenic concentration in the Marseille Calanques measured by Laser Induced Breakdown Spectroscopy. by T. Sarnet and J.

Use of a High-Resolution Overview Spectrometer for the Visible Range in the TEXTOR Boundary Plasma

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

ICP-OES Application Note Number 35

Analysis Repeatability of Trace and Major Elements in a Water Sample

Development of advanced optical techniques for verification measurements Igor Jovanovic University of Michigan

Remote laser-induced plasma spectrometry for elemental analysis of samples of environmental interest

The importance of longer wavelength reheating in dual-pulse laser-induced breakdown spectroscopy

Visualization of Xe and Sn Atoms Generated from Laser-Produced Plasma for EUV Light Source

Remote Raman & Fluorescence Capabilities for Chemical Detection at University of Hawaii

Title duration on laser ablation in liqui. Rightc 2009 Elsevier B.V. All rights res

Temporally resolved laser induced plasma diagnostics of single crystal silicon effects of ambient pressure

Measurements of egg shell plasma parameters using laser-induced breakdown spectroscopy

Underwater Raman Sensor for Detecting High Explosives and Homemade Explosives (HMEs)

Investigation of fundamental mechanisms related to ambient gas heating and hydrodynamics of laser-induced plasmas

NEW DEVELOPMENTS OF AUTORADIOGRAPHY TECHNIQUE TO IMPROVE ALPHA AND BETA MEASUREMENTS FOR DECOMMISSIONING FACILITIES

Electron Microprobe Analysis 1 Nilanjan Chatterjee, Ph.D. Principal Research Scientist

Electron Microprobe Analysis 1 Nilanjan Chatterjee, Ph.D. Principal Research Scientist

Laser-induced breakdown spectroscopy in analysis of Al 3+ liquid droplets: On-line preconcentration by use of flow-injection manifold

Overview of X-Ray Fluorescence Analysis

TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER

L E A. Laser Elemental Analyzer

DIAGNOSTIC OF A LASER-INDUCED OPTICAL BREAKDOWN BASED ON HALF-WIDTH AT HALF AREA OF H LINES , H , AND H

Design, construction and assessment of a field-deployable laser-induced breakdown spectrometer for remote elemental sensing

ICP-Mass Spectrometer

This article appeared in a journal published by Elsevier. The attached copy is furnished to the author for internal non-commercial research and

LASER INDUCED BREAKDOWN SPECTROSCOPY FOR SEMI-QUANTITATIVE ELEMENTAL ANALYSIS IN SOILS AND MARINE SEDIMENTS

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES

In Situ Analysis of Geochemistry and Mineralogy on the Venus Surface

Chapter 4 LIBS system: Performance evaluation and optimization of conditions for quantitative analysis

The Evolution of Inductively Coupled Plasma - Optical Emission Spectroscopy (ICP-OES)

Ultrafast Dynamics and Single Particle Spectroscopy of Au-CdSe Nanorods

Surface analysis techniques

In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer*

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Assessment of the Upper Particle Size Limit for Quantitative Analysis of Aerosols Using Laser-Induced Breakdown Spectroscopy

Environmental impact. Improve resource efficiency. Quality secondary raw materials. Pollution. CO 2 emission. Biodiversity. Policy and regulations

IMPLEMENTATION OF THE MONTE CARLO-LIBRARY LEAST- SQUARES APPROACH TO ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS

Portable type TXRF analyzer: Ourstex 200TX

Determination of the activity of radionuclides

Resonance Raman measurements utilizing a deep UV source

Laser and pinching discharge plasmas spectral characteristics in water window region

This work was performed under the auspices of the U.S. Department of Energy by Lawrence Livermore National Laboratory under contract

Hristina Nikolova, Miguel Angel Aguirre, Montserrat Hidalgo and Antonio Canals. 5-6 June 2012, Plovdiv, Bulgaria. 5-6 June 2012, Plovdiv, Bulgaria

CARBON NANOTUBE ANALYSIS USING THE TSI LIBS DESKTOP ANALYZER

Florida Institute of Phosphate Research 1855 West Main Street Bartow, Florida (863) Fax: (863)

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Supplemental material for Bound electron nonlinearity beyond the ionization threshold

Fundamental investigation on CO 2 laser-produced Sn plasma for an EUVL source

Elemental analysis by microwave-assisted laser-induced breakdown spectroscopy: Evaluation on ceramics

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

Differential Spectral Imaging of the CN Violet Band in Laser-Induced Plasmas on TNT Simulant Molecules

Introduction to LIBS COMMUNITY USER WORKSHOP ON PLANETARY LIBS (CHEMCAM) DATA. Sam Clegg and the ChemCam team

MICHAEL J. WITTE 1 AND CHRISTIAN G. PARIGGER 1,a

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Elemental analysis by X-ray f luorescence. Sequential benchtop WDXRF spectrometer

Measurements of Deuterium Retention and Surface Elemental Composition with Double Pulse Laser Induced Breakdown Spectroscopy

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

UNIVERSITI SAINS MALAYSIA. Second Semester Examination Academic Session 2004/2005. March KAA 502 Atomic Spectroscopy.

High-Resolution Survey of the Visible Spectrum of NiF by Fourier Transform Spectroscopy

PFC/JA Precision Measurements of the Wavelengths of Emission Lines of Mg-like and Na-like Kv in Alcator C Plasmas

Lecture 16 Instrumentation for ICP AES-VIII-Instruments

EUV Reflectivity measurements on Acktar Sample Magic Black

Chapter 9. Atomic emission and Atomic Fluorescence Spectrometry Emission spectrophotometric Techniques

Set-up for ultrafast time-resolved x-ray diffraction using a femtosecond laser-plasma kev x-ray-source

Transcription:

Analysis by laser-induced breakdown spectroscopy of complex solids, liquids, and powders with an echelle spectrometer Pascal Fichet, Denis Menut, René Brennetot, Evelyne Vors, and Annie Rivoallan One of the most promising approaches to laser-induced breakdown spectroscopy LIBS experiments involves the use of an echelle spectrometer coupled with an intensified CCD. Even if drawbacks remain with its use, the echelle spectrometer facilitates a multielemental analysis that is more rapid than can be obtained with the more-conventional spectrometer and, moreover, does not sacrifice reliability. Quantitative results obtained with such apparatus for solids, liquids, powders, and gases are described and when possible compared with results from spectrometers. Liquid analysis by LIBS with echelle spectrometers has allowed a spectral database to be compiled. Once the qualitative spectra of pure elements in aqueous solutions, are obtained, they can be used for qualitative analysis of unknown samples. 2003 Optical Society of America OCIS codes: 140.3440, 300.6210 1. Introduction In recent years laser-induced breakdown spectroscopy LIBS has been studied because of the numerous advantages of its use in remote analysis in various environments, including hostile ones. As an example, a recent paper mentions the use of LIBS to characterize solids submerged in water. 1 Moreover, in this paper applications of LIBS to control the cooling systems of nuclear plants are proposed. Experiments seem to indicate that LIBS will have important applications in nuclear waste control. 2,3 But all these applications require not only a qualitative but also a quantitative rapid multielemental analysis, which implies a large spectral coverage with high resolution. 2 Of course, high spectral resolution can be obtained with conventional spectrometers but only in a restricted portion of the spectrum. It takes many laser shots and therefore much time to make the sequential measurements that are needed for the analysis of many elements. The authors are with the Commissariat à l Energie Atomique Saclay, Fuel Cycle Division, Laboratoire d Analyse par Laser et d Etude de Surface, Bâtiment 391, 91191 Gif Sur Yvette, France. P. Fichet s e-mail address is fichet@carnac.cea.fr. Received 2 December 2002; revised manuscript received 7 May 2003. 0003-6935 03 306029-07$15.00 0 2003 Optical Society of America In contrast, in recent years an echelle spectrometer combined with an intensified CCD camera has offered a large spectral domain for investigation. The objective of the research described in this paper is to highlight the great potential of echelle spectrographs not only for qualitative analysis but also for quantitative analysis. To our knowledge, there are three commercial apparatuses relevant to LIBS analysis. 4 6 Our aim in this paper is not to compare all these systems but to give examples of ways in which an echelle spectrometer can be an interesting system. Early in the recent development of echelle spectrometers for LIBS analysis, Bauer et al. 7 tried to connect an echelle spectrometer and a Princeton Instruments intensified CCD camera. Al and Fe spectra were produced, but to obtain large spectral coverage with this first apparatus, they used a mobile mirror inside the device. The same problem was encountered by Haisch et al., 8 who developed an echelle spectrometer for analysis of colloid materials. These early publications reveal that a large intensified CCD camera is necessary for important spectral coverage without any moving components. At the same time, Lindblom 9 presented a glass spectrum obtained with his new echelle spectrometer the Mechelle spectrometer. Large spectral coverage 190 nm, 1100 nm with no gaps illustrates the wide applicability of echelle spectrometers. Recently a versatile spectrometer was proposed by Florek et al. 10 20 October 2003 Vol. 42, No. 30 APPLIED OPTICS 6029

In parallel with the development of the apparatus, with a multichannel Mechelle 7500 commercial instrument 11 Goode et al. showed an interesting ability to classify quickly and qualitatively various alloys, taking into account the entire emission spectrum and using a principal-component analysis. Almost the same time, another application was described by Fink et al. 12 They obtained good results from a LIBS experiment that included an echelle spectrometer Model ESA 3000, LLA GmbH, Berlin to investigate recycled thermoplasts. The LIBS technique is known as a destructive analytical method. But Bicchieri et al. 13 applied the technique in research with a medieval manuscript because of its utility in terms of depth resolution and local analysis. Therefore the Mechelle 7500 spectrometer was useful for simultaneous multielement analysis and production of significant amounts of information. Recently, applications of LIBS to liquids, 14,15 wood, 16 aluminum alloys, 17 glass, 18 and geological samples 19,20 have been described, proving the existence of a growing interest in echelle spectrometers. At the same time, some developments of large broadband optical range spectrometers have also been described. 21 25 These new applications which use Ocean Optics Dunedin, Fla. miniature spectrometers, highlight the necessity to collect as much information as possible after plasma formation to develop powerful analytical applications of LIBS. The goal of the present research was to determine whether LIBS experiments that use echelle spectrometers are appropriate for a variety of applications: investigations of liquids, powders, solids, and gases. We built homemade software with which to identify automatically various emission lines among the huge quantity of data produced by the system. As a consequence, a library of LIBS spectra was obtained for 46 elements and used to investigate unknown samples. 2. Experimental Setup For each application described below a standard LIBS experiment was built. A Nd:YAG laser Quantel Brilliant B operating at 532 nm was used as the ablation source, with an energy of 60 mj pulse. Three reflective mirrors guided the laser to a focusing lens. The closest mirror to the lens, made from quartz that is reflective at 45, permitted plasma recording in the same direction. The laser beam was focused by a plano convex quartz lens with a different focal length for each application for liquids, f 30 cm; for powders, f 50 cm; for solids and gases, f 10 cm. For liquids and powders longer focal lengths were chosen to prevent splashing from damaging the optics during the experiment, and the laser frequency was fixed at 1 instead of 10 Hz for solid and gas applications. Only for liquid applications was the laser s optical pathway tilted at an angle of 15 for the same reason, which is described in detail in Ref. 14. For gas applications an airtight chamber was designed with one quartz window to guide the laser Fig. 1. Experimental setup for analysis of liquids. inside and to collect the plasma emission in the same direction. Once the plasma was formed, its emission went through the laser focusing lens and the quartz dichroic mirror and was then focused onto a 1-m-long fused-silica optical fiber numerical aperture, 0.22; core, 1 mm placed behind the quartz mirror closest to the lens. As an example, the experimental setup used for investigations of liquids is presented schematically in Fig. 1. This setup is described in full detail in Ref. 14. For some experiments described below, the plasma emission was imaged at the entrance slit of a Czerny Turner spectrometer Jobin Yvon Model THR 1000, Longjumeau, France; not shown in Fig. 1 or at the entrance of an echelle spectrometer. The ESA 3000EV spectrometer LLA GmbH, Berlin, Germany was used for liquids, solids, and powders its spectral range is 200 780 nm and its resolution is 10,000. For gases a Catalina SE 200 high order version, Catalina Scientific Corporation, Tucson, Arizona echelle spectrometer provided spectra its spectral range is 190 1100 nm and resolution is 3,000. 3. Results and Discussion A. Echelle for Investigations of Liquids Even if LIBS analytical applications to liquids were studied less frequently than those to solids, standard sample preparations are obviously much easier. The solutions used for this study were purchased from the Spex Certiprep Company Metuchen, New Jersey. With several monoelemental solutions chosen at concentrations of 10,000 g ml 1, the corresponding spectra were obtained with the echelle spectrometer detector. The echelle spectrometer produced a huge quantity of data, an outcome that first was considered a drawback; therefore 6030 APPLIED OPTICS Vol. 42, No. 30 20 October 2003

Table 1. Part of the LIBS Spectral Database Obtained for Co in This Work Compared to NIST Data Co LIBS Spectral Database NIST Spectral Data b Wavelength nm Intensity a.u. Wavelength nm Intensity a.u. 345.344 41,128 345.35 21,000 340.505 28,341 340.512 11,000 356.931 26,178 356.938 8,800 352.981 23,716 352.981 7,300 350.626 21,636 350.632 7,000 358.717 19,932 358.719 6,700 350.222 19,069 350.228 9,600 a Ref. 26. b The source in all cases was Co I. homemade software VISUAL BASIC 6.0, Microsoft was written to extract automatically the various emission lines of interest and to accumulate the spectral database obtained by LIBS. So far, LIBS spectra for 46 elements have been gathered in our spectral database. As an example, some of the results obtained for Co are compared in Table 1 with those that we extracted from Ref. 26. The echelle spectrometer is reliable in terms of wavelength calibration because the difference between the National Institute of Technology NIST database and the experimental data was found for Co and also for the other elements to be 15 pm. Because there are no moving parts in the echelle spectrometer, automatic wavelength calibration is a great advantage for qualitative and obviously for quantitative application. We have used this spectral database built for LIBS experiments to interpret unknown spectra. Examples are provided for investigation of powders. After several tests for wavelength calibration, quantitative studies of liquids were performed by dilution of known quantities of standard solutions in distilled water. The ability of an echelle spectrometer see Fig. 2 for-an-example showing a calibration curve for Na in water; the detection limit was 0.55 g ml 1 to give quantitative results was assessed. Each datum of the calibration curve represents the mean value of six duplicate experiments. Each measurement corresponds to the area of the emission line of interest obtained with an accumulation on the detector of 300 laser shots. The reproducibility of each point in the calibration curve was 5 10%. The accuracy of the analysis was evaluated to be 10%, assuming that the unknown concentration is far from the detection limit and of course within the limits of the calibration curve. The sensitivity of such system has also to be evaluated by direct comparison of the detection limits obtained in the same conditions laser energy and experimental setup with a classic spectrometer see particularly Ref. 14, which described the same experimental setup. A direct comparison of all the essential parameters gate delay, gate width, gain of the camera remains difficult. Therefore for the limits of detection LOD obtained with the echelle and the systems we must admit that optimum parameters were chosen. Table 2 shows the different limits of detection obtained for Na and Mg in the research reported here and that reported previously in the literature. For Mg the result that we obtained has the lowest LOD, only Ref. 27 presents a better LOD for Na. With these results we can conclude that detection limits obtained with an echelle spectrometer are of the same order of magnitude as those obtained with a spectrometer. The possibility of using systematically an internal standard that can have an emission line that is not close to the element of interest is also an essential feature of the echelle spectrometer. This opportunity, very often used for inductively coupled plasma optical emission spectroscopy, can be important to be used to improve the stability of the system with time reproducibility. In Fig. 3, two calibration curves are shown as examples of calibration curves of Ba in Fig. 2. Calibration curve of Na in an aqueous solution obtained with an echelle spectrometer equation of the line, y 1899x 3831; R 2 0.9983. 20 October 2003 Vol. 42, No. 30 APPLIED OPTICS 6031

Table 2. Limits of Detection g ml 1 of Na and Mg in Water Element Present Work Echelle LOD Echelle a b c a Ref. 15. b Ref. 14. c Ref. 27. Mg 0.4 1 1 3 Na 0.55 2 0.5 0.08 aqueous solutions with and without Sc as an internal standard. The experimental parameters used for Na were kept constant for the Ba study. We are not aware of any previous studies of the time evolution of the calibration curve. With the results shown in Fig. 3 our conclusion is that no improvement was achieved by the use of an internal standard. The reliability of the LIBS technique for the analysis of liquids is under study, and thus the usefulness of an internal standard will have to be proved. B. Echelle for Analysis of Solids Having established the utility of an echelle spectrometer for the analysis of liquids, we also employed the detector for analysis of Al alloys. This apparatus was an interesting choice for recording and choosing the most sensitive lines for various trace elements to be analyzed but also for the less-sensitive lines of the matrix. The matrix lines can be used as internal standards in the calibration curves of trace elements if no saturation or self-absorption occurs. With different Al alloys, 12 calibration curves were obtained by LIBS. Each datum obtained for the calibration curve corresponds to five duplicate experiments at different positions on the sample surface to prevent potential sample in homogeneity. Two hundred laser shots were accumulated on the detector. The same reproducibility and accuracy were found for applications to solids and to liquids. An example of a calibration curve for Ni is shown in Fig. 4. The detection limits are close to those that can be found in the literature, but obviously all trace elements can be investigated at the same time and thus quickly. Ta- Fig. 3. Calibration curves, with detection limit, 0.3 g ml; equation of the line, y 0.12x 0.15, R 2 1 and without detection limit, 0.36 g ml; equation of the line, y 6093x 8455, R 2 1 Sc as an internal standard, for Ba in an aqueous solution obtained with an echelle spectrometer. Fig. 4. Calibration curve obtained for Ni in Al alloys equation of the line, y 19.3x 0.3, R 2 0.99. 6032 APPLIED OPTICS Vol. 42, No. 30 20 October 2003

Table 3. Detection Limits Obtained for Trace Elements in Al Alloys a Element This work g g 1 Echelle Echelle Ref. 17 ppm Ref. 28 ppm Ref. 29 ppm Ref. 30 ppm and Femtosecond Laser Be 4 1.6 0.3 Cr 2 3 Cu 2 33 10 56 10 7.0 Fe 9 30 22.3 Mg 0.134 9 0.5 7 0.5 2.1 Mn 2 65 7 90 2 6.7 Ni 7 Pb 15 Si 10 90 14 14 80 Ti 3 Zn 15 38 a A dash means no data available; ppm means parts in 10 6. ble 3 lists not only our results but also those reported in the literature. The type of spectrometer is specified for each reference. The values obtained for Cu, Mg, Si, Fe, and Zn are the lowest. The LOD reported in Ref. 17 for Be with a spectrometer is better than that obtained with the twoechelle apparatus. This result can be due to the sensitivity of the various detectors to a specific wavelength. For Cr and Mn, detection limits obtained with the two systems are similar. For Ni, Pb, and Ti no other values have been published to our knowledge. C. Echelle for Analysis of Powders Because LIBS instruments can be portable, applications to the screening of various soils can be interesting. A goal of our experiment was to show the capability of the echelle spectrometer for qualitative analysis of unknown samples. A spectrum was obtained with our standard LIBS experimental setup. Homemade software was built to extract the different emission lines of the experimental spectrum. Once they were obtained, all the wavelengths of the emission lines were automatically compared to our LIBS spectral database obtained with liquids and described above. The intensities of the emission lines were also considered in our interpretation of the results. Observation of the three main lines in the experimental spectrum, as reported in the spectral database, is the criterion chosen for our calculation. Obviously the three main lines in the LIBS spectra of solids and liquids may be different, but generally lines that are sensitive for liquids are sensitive for solids. With this criterion, and only for a 1-s calculation duration, the elements Si, Fe, Mg, Ti, Ca, Na, Mn, Sr, Ba, O, and Cr were qualitatively observed in the unknown sample. For example, Fig. 5 shows a qualitative comparison of the lines observed with 10,000 g ml of Ti in water reference spectrum and those recorded for the soil sample. A quantitative interpretation of the results is more difficult, but use of an echelle spectrometer detector for a LIBS experiment is a powerful apparatus just for qualitative analysis. D. Echelle for Control of Gases In the present study an echelle spectrometer the Catalina SE 200 was tested for control of gases. A mixture of 5% Kr in N 2 was detected, a portion of the spectrum is depicted in Fig. 6. Other gas lines such as Xe 529.222 nm and O 777.194 nm were also observed with our experimental setup. We have performed no quantitative analysis for this gas study. Fig. 5. Comparison of a reference spectrum of Ti in an aqueous solution and lines observed in the unknown powder spectrum. 20 October 2003 Vol. 42, No. 30 APPLIED OPTICS 6033

Fig. 6. Spectrum of a portion of a N 2 Kr mixture. 4. Conclusions For several analytical applications of laser-induced breakdown spectroscopy to liquids, solids, powders, and gases, the use of a spectrometer that possesses a wide spectral range and particularly of an echelle spectrometer is an attractive choice. Because a large portion of the spectrum is investigated simultaneously, many elements can be qualitatively and quantitatively studied at the same time. No moving parts could be found in the echelle spectrometers tested, which ensures stable wavelength calibration. Moreover, the difference between our database and the NIST spectral database for emission spectroscopy was 15 pm. The quantitative results described for liquids and solids are quite similar to those that have been reported in the literature for systems that use a standard apparatus. The main difference lies in the fact that fewer experiments are necessary for multielemental analysis to produce similar results. All the trace elements were investigated at the same time. Because it is easy to find commercial pure monoelemental solutions of many elements, a spectral database dedicated to the LIBS technique was easily produced with the echelle spectrometer for 46 different elements. With criteria such as the presence or lack of sensitive lines, the spectral database can be used to analyze qualitatively unknown samples. References 1. D. C. S. Beddows, O. Samek, M. Liska, and H. H. Telle, Singlepulse laser-induced breakdown spectroscopy of samples submerged in water using a single-fibre light delivery system, Spectrochim. Acta Part B 57, 1461 1471 2002. 2. P. Fichet, P. Mauchien, and C. Moulin, Determination of impurities in uranium and plutonium dioxides by laser-induced breakdown spectroscopy, Appl. Spectrosc. 53, 1111 1117 1999. 3. A. I. Whitehouse, J. Young, I. M. Botheroyd, S. Lawson, C. P. Evans, and J. Wright, Remote material analysis of nuclear power station steam generator tubes by laser-induced breakdown spectroscopy, Spectrochim. Acta Part B 56, 821 830 2001. 4. LLA Instruments GmbH, Scwarzschildstrasse 10, D-12489 Berlin-Adlershof, Germany, Model ESA 3000; www.lla.de. 5. Catalina Scientific Corporation, 2930 West Saint Tropez Avenue, Tucson, Ariz. 85713, Model SE 200 echelle spectrograph; www.catalinasci.com. 6. Multichannel Instruments AB, Pilotgatan 2, S-12832 Skarpnäck Stockolm Sweden, Models Mechelle 900 and Mechelle 7500; www.multichannel.se. 7. H. E. Bauer, F. Leis, and K. Niemax, Laser induced breakdown spectrometry with an echelle spectrometer and intensified charge coupled device detection, Spectrochim. Acta 53, 1815 1825 1998. 8. C. Haisch, U. Panne, and R. Niessner, Combination of an intensified charge coupled device with an echelle spectrograph for analysis of colloidal material by laser-induced plasma spectroscopy, Spectrochim. Acta Part B 53, 1657 1667 1998. 9. P. Lindblom, New compact echelle spectrographs with multichannel time-resolved recording capabilities, Anal. Chim. Acta 380, 353 361 1998. 10. S. Florek, C. Haisch, M. Okruss, and H. Becker-Ross, A new, versatile echelle spectrometer relevant to laser induced plasma applications, Spectrochim. Acta Part B 56, 1027 1034 2001. 11. S. R. Goode, S. L. Morgan, R. Hoskins, and A. Oxsher, Identifying alloys by laser-induced breakdown spectroscopy with a time-resolved high resolution echelle spectrometer, J. Anal. At. Spectrom. 15, 1133 1138 2000. 12. H. Fink, U. Panne, and R. Niessner, Analysis of recycled thermoplasts from consumer electronics by laser-induced plasma spectroscopy, Anal. Chim. Acta 440, 17 25 2001. 13. M. Bicchieri, N. Nardone, P. A. Russo, A. Sodo, M. Corsi, G. Cristoforetti, V. Palleschi, A. Salvetti, and E. Tognoni, Characterization of azurite and lazurite bazed pigments by laser induced breakdown spectroscopy, Spectrochim. Acta Part B 56, 915 922 2001. 14. P. Fichet, P. Mauchien, J. F. Wagner, and C. Moulin, Quantitative elemental determination in water and oil by laser induced breakdown spectroscopy, Anal. Chim. Acta 429, 269 278 2001. 15. B. Charfi and M. A. Harith, Panoramic laser-induced breakdown spectrometry of water, Spectrochim. Acta Part B 57, 1141 1153 2002. 16. A. Uhl, K. Loebe, and L. Kreuchwig, Fast analysis of wood preservers using laser induced breakdown spectroscopy, Spectrochim. Acta Part B 56, 795 806 2001. 17. V. Detalle, R. Héon, M. Sabsabi, and L. St Onge, An evaluation of a commercial echelle spectrometer with intensified charge-coupled device detector for materials analysis by laserinduced plasma spectroscopy, Spectrochim. Acta Part B 56, 1011 1025 2001. 18. J. I. Yun, R. Klenze, and J.-I. Kim, Laser-induced breakdown spectroscopy for on-line multielement analysis of highly radioactive glass melt. I. Characterization and evaluation of the method, Appl. Spectrosc. 56, 437 448 2002. 19. R. Brennetot, J. L. Lacour, E. Vors, A. Rivoallan, D. Vailhen, and S. Maurice, MALIS Mars analysis by laser-induced breakdown spectroscopy : influence of Mars atmosphere on plasma emission. Study of factors influencing plasma emission with the use of Doehlert designs, Appl. Spectrosc. to be published. 6034 APPLIED OPTICS Vol. 42, No. 30 20 October 2003

20. D. A. Cremers, R. C. Wiens, M. J. Ferris, J. D. Blasic, R. Brennetot, and S. Maurice, Development of laser-induced breakdown spectroscopy LIBS for analysis of geological samples on planetary missions, presented at the Lunar and Planetary Science XXXIII meeting, Houston, Tex., 11 15 March 2002. 21. B. C. Castle, A. K. Knight, K. Visser, B. W. Smith, and J. D. Winefordner, Battery powered laser induced plasma spectrometer for elemental determinations, J. Anal. At. Spectrom. 13, 589 595 1998. 22. I. B. Gornushkin, B. W. Smith, H. Nasajpour, and J. D. Winefordner, Identification of solid materials by correlation analysis using a microscopic laser-induced plasma spectrometer, Anal. Chem. 71, 5157 5164 1999. 23. D. Kossakovski and J. L. Beauchamp, Topographical and chemical microanalysis of surfaces with a scanning probe microscope and laser-induced breakdown spectroscopy, Anal. Chem. 72, 4731 4737 2000. 24. G. Galbacs, I. B. Gornushkin, B. W. Smith, and J. D. Winefordner, Semi-quantitative analysis of binary alloys using laser-induced breakdown spectroscopy and a new calibration approach based on linear correlation, Spectrochim. Acta Part B 56, 1159 1173 2001. 25. S. I. Gornushkin, J. M. Anzano, B. W. Smith, and J. D. Winefordner, Effective normalization technique for correction of matrix effects in laser-induced breakdown spectroscopy detection of magnesium in powdered samples, Appl. Spectrosc. 56, 433 436 2002. 26. NIST Atomic Spectra Database; www.physics.nist.gov cgibin atdata main_asd. 27. O. Samek, D. C. S. Beddows, J. Kaiser, S. V. Kukhlevsky, M. Liska, H. H. Telle, and J. Young, Application of laser-induced breakdown spectroscopy to in-situ analysis of liquid samples, Opt. Eng. 39, 2248 2262 2000. 28. A. K. Rai, H. Zhang, F. Y. Yueh, J. P. Singh, and A. Weisburg, Parametric study of a fiber-optic laser-induced breakdown spectroscopy probe for analysis of aluminum alloys, Spectrochim. Acta Part B 56, 2371 2383 2001. 29. M. Sabsabi and P. Cielo, Quantitative analysis of aluminum alloys by laser-induced breakdown spectroscopy and plasma characterization, Appl. Spectrosc. 49, 499 507 1995. 30. B. Le Drogoff, J. Margot, M. Chaker, M. Sabsabi, O. Barthelemy, T. W. Johnston, S. Laville, F. Vidal, and Y. Von Kaenel, Temporal characterization of femtosecond laser pulses induced plasma for spectrochemical analysis of aluminium alloys, Spectrochim. Acta Part B 56, 987 1002 2001. 20 October 2003 Vol. 42, No. 30 APPLIED OPTICS 6035