Gate simulation of Compton Ar-Xe gamma-camera for radionuclide imaging in nuclear medicine

Similar documents
Compton Camera. Compton Camera

A Concept of Cylindrical Compton Camera for SPECT

Study of the feasibility of a compact gamma camera for real-time cancer assessment

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

A novel design of the MeV gamma-ray imaging detector with Micro-TPC

Photon and primary electron arithmetics in photoconductors for digital mammography: Monte Carlo simulation studies

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Study well-shaped germanium detectors for lowbackground

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques

INTERACTIONS OF RADIATION WITH MATTER

1420 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 57, NO. 3, JUNE /$ IEEE

Compton Camera with PositionSensitive Silicon Detectors

Physics of Radiotherapy. Lecture II: Interaction of Ionizing Radiation With Matter

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

III. Energy Deposition in the Detector and Spectrum Formation

Lecture 5: Tomographic nuclear systems: SPECT

Detection and measurement of gamma-radiation by gammaspectroscopy

Modeling of microporous silicon betaelectric converter with 63 Ni plating in GEANT4 toolkit*

Simulation of 4π HPGe Compton-Suppression spectrometer

Radiation Detection and Measurement

Temporal and lateral distributions of EAS neutron component measured with PRISMA-32

QUIZ: Physics of Nuclear Medicine Atomic Structure, Radioactive Decay, Interaction of Ionizing Radiation with Matter

Compton suppression spectrometry

Bases of radioisotope diagnostic methods

on basis of the Geant4 simulations

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

Quantitative Assessment of Scattering Contributions in MeV-Industrial X-ray Computed Tomography

Mitigation of External Radiation Exposures

A. I, II, and III B. I C. I and II D. II and III E. I and III

A Study On Radioactive Source Imaging By Using A Pixelated CdTe Radiation Detector

Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions

Radionuclide Imaging MII Detection of Nuclear Emission

Basic hands-on gamma calibration for low activity environmental levels

Pressure effect in the X-ray intrinsic position resolution in noble gases and mixtures

X-ray Energy Spectroscopy (XES).

Basic physics Questions

X-ray ionization yields and energy spectra in liquid argon

Comparative Analysis of Nuclear Cross Sections in Monte Carlo Methods for Medical Physics Applications

Structure of Biological Materials

Influence of gas conditions on electron temperature inside a pinch column of plasma-focus discharge

Precision neutron flux measurement with a neutron beam monitor

Introduction to SPECT & PET TBMI02 - Medical Image Analysis 2017

X-ray ionization yields and energy spectra in liquid argon

A scintillation dual-detector system featuring active Compton suppression

laser with Q-switching for generation of terahertz radiation Multiline CO 2 Journal of Physics: Conference Series PAPER OPEN ACCESS

Theoretical approach to estimate radiation damage within FEL irradiated samples. Beata Ziaja

arxiv:physics/ v1 3 Aug 2006

An autonomous underwater telescope for measuring the scattering of light in the deep sea

Interactive Web Accessible Gamma-Spectrum Generator & EasyMonteCarlo Tools

Effect of the time characteristics of the Compton camera on its performance

Interaction of particles with matter - 2. Silvia Masciocchi, GSI and University of Heidelberg SS2017, Heidelberg May 3, 2017

Applied Radiation and Isotopes

Evaluation of the Nonlinear Response Function and Efficiency of a Scintillation Detector Using Monte Carlo and Analytical Methods

Nuclear Medicine Intro & Physics from Medical Imaging Signals and Systems, Chapter 7, by Prince and Links

Radionuclide Imaging MII Positron Emission Tomography (PET)

Physics of Radiography

PHYS 3650L - Modern Physics Laboratory

Sample Spectroscopy System Hardware

The Compton Effect. Martha Buckley MIT Department of Physics, Cambridge, MA (Dated: November 26, 2002)

Determination of parameters of cascade showers in the water calorimeter using 3D-distribution of Cherenkov light

DR KAZI SAZZAD MANIR

11/10/2014. Chapter 1: Introduction to Medical Imaging. Projection (Transmission) vs. Emission Imaging. Emission Imaging

Prompt gamma measurements for the verification of dose deposition in proton therapy. Contents. Two Proton Beam Facilities for Therapy and Research

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Particle Energy Loss in Matter

nerix PMT Calibration and Neutron Generator Simulation Haley Pawlow July 31, 2014 Columbia University REU, XENON

Basic physics of nuclear medicine

Units and Definition

Laboratory experiments on the formation and recoil jet transport of aerosol by laser ablation

Radioisotopes in action. Diagnostic application of radioisotopes. Steps of diagnostic procedure. Information from various medical imaging techniques

Interaction of Ionizing Radiation with Matter

Validation of the MCNP computational model for neutron flux distribution with the neutron activation analysis measurement

Nuclear Physics and Astrophysics

Interaction of charged particles and photons with matter

Effects of femtosecond laser radiation on the skin

ISOCS / LabSOCS. Calibration software for Gamma Spectroscopy

1 Geant4 to simulate Photoelectric, Compton, and Pair production Events

Drift plane. substrate (20ÉIm polyimide) 200ÉIm. Back strip (180ÉIm width) Base (Ceramic) Anode strip (10ÉIm width) Cathode strip (100ÉIm width)

MCSHAPE: A MONTE CARLO CODE FOR SIMULATION OF POLARIZED PHOTON TRANSPORT

CHARGED PARTICLE INTERACTIONS

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN

Radiation Physics PHYS /251. Prof. Gocha Khelashvili

Interactions of Radiation with Matter

Scintillators 1. YEAR, I. CYCLE. AUTHOR: BPhys. Manja Ščetinec ADVISOR/MENTOR: Prof. Boštjan Golob

Particle Energy Loss in Matter

Performance of the Gamma-ray Imaging Detector with Micro-TPC

An idea for the future proton detection of (p,2p) reactions with the R 3 B set-up at FAIR

Jovian electrons as an instrument of investigation of the interplanetary medium structure

CHAPTER 2 INTERACTION OF RADIATION WITH MATTER

TRADITIONALLY, Compton imaging has been performed

Absolute activity measurement

Correction for PMT temperature dependence of the LHCf calorimeters

The TDCR method in LSC. P. Cassette Laboratoire National Henri Becquerel CEA/LNE, France

Lucite Hodoscope for SANE

K 40 activity and Detector Efficiency

EEE4106Z Radiation Interactions & Detection

Sensitivity of sodium iodide cryogenic scintillation-phonon detectors to WIMP signals

Interaction theory Photons. Eirik Malinen

Experimental study of some shielding parameters for composite shields

Transcription:

Journal of Physics: Conference Series PAPER OPEN ACCESS Gate simulation of Compton Ar-Xe gamma-camera for radionuclide imaging in nuclear medicine To cite this article: L Yu Dubov et al 2017 J. Phys.: Conf. Ser. 784 012019 View the article online for updates and enhancements. Related content - Development of event reconstruction algorithm for full-body gamma-camera based on SiPMs D E Philippov, V N Belyaev, P Zh Buzhan et al. - The Organization of the Internal Irradiation Monitoring System in Conditions of Nonstandard Radionuclide Intakes A V Ovchinnikov, K M Izmestyev, D G Demyanyuk et al. - Assessment of radionuclides and heavy metals in marine sediments along the Upper Gulf of Thailand S Khuntong, C Phaophang and W Sudprasert This content was downloaded from IP address 46.3.207.217 on 29/11/2017 at 22:20

Gate simulation of Compton Ar-Xe gamma-camera for radionuclide imaging in nuclear medicine L Yu Dubov, V N Belyaev, A K Berdnikova, A I Bolozdynia, Yu A Akmalova, and Yu V Shtotsky National Research Nuclear University «MEPHI», Moscow, Russian Federation E-mail: LYDubov@mephi.ru Abstract. Computer simulations of cylindrical Compton Ar-Xe gamma camera are described in the current report. Detection efficiency of cylindrical Ar-Xe Compton camera with internal diameter of 40 cm is estimated as1-3%that is 10-100 times higher than collimated Anger s camera. It is shown that cylindrical Compton camera can image Tc-99m radiotracer distribution with uniform spatial resolution of 20 mm through the whole field of view. 1. Introduction Taking all into consideration we conclude that behavior of S-factor in different cases has a strong influence on effective therapeutic activity. Method has high flexibility and reproducibility and gives opportunity to physician for wide retrospective in-depth analysis, thus such approach will lead to improvement of RSO effectiveness. Radionuclide imaging is a powerful tool for medical diagnostics [1]. Single photon emission computer tomography (SPECT) based on Anger scintillation gamma cameras with typical spatial resolution of about 10 mm is commonly used for imaging distribution of radioisotopes emitting single gamma quanta. However, the low efficiency of mechanically collimated gamma cameras significantly affects their performance, particularly it is not possible to reduced exposition time or to lower the activity of administered radionuclide. The absence of a collimator allows to use much more of the available radiation. Detectors with an electronic collimation based on the Compton effect, having a high luminosity and an acceptable spatial resolution of ~ 10 mm, were considered in a number of previous studies [3,4]. In the current work parameters of a full size cylindrical gaseous Argon-Xenon Compton gamma camera are optimized through Monte Carlo simulations in Gate/Geant4 environment [2].Also, the requirements for event selection and reconstruction algorithms for 3D Compton images are discussed. 2. Compton gamma camera The determination of the source location via the Compton effect is based on the dependence between the scattering angle Θ and the energy of the recoil electron E 1 after Compton scattering of a gamma quantum with the energy E: 2 1 1 cos( ) 1 m e c Е1 E Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. Published under licence by IOP Publishing Ltd 1

Compton camera (see Figure 1) consists of a scatter detector (SD) and an absorption detector (AD). If the energy of the gamma rays is known it is necessary to determine the energy of the recoil electron E1 in the RD for calculation of the scattering angle Θ. Compton gamma camera detects the arrival direction of the gamma ray within the events acceptance cone, which is determined by the scattering angle Θ. Event cone axis is the extension of straight line connecting the point of interaction in the SD (x1, y1, z1) and the AD (x2, y2, z2).the apex of the cone is the point of interaction (Compton scattering) in the RD. Uncertainty in the determination of the events acceptance cone are defined by a finite energy resolution of the SD, an error in the coordinates of events and by the Doppler broadening of the scattered γ-ray energy [3]. 3. Optimal parameters of gaseous detectors Figure 1. Compton camera operation principle. 3.1.General requirements for the Compton camera detectors The scatter detector (SD) determines the coordinates of Compton interaction and the energy of Compton electron. SD must have high energy resolution and the prevailing probability of Compton scattering of the gamma rays η C. The probability of photoelectric effect and multiple Compton scattering should be as low as possible. Total absorption detector (AD) determines the coordinates and energy of photoelectric absorption. AD should have a high probability of the photoelectric effect η PE for gamma rays. Contributions of single and multiple Compton scattering (followed by photoelectric effect) in AD should be minimal. The detector for medical imaging should enclose the investigated object in order to achieve maximal efficiency. The optimal geometry of the Compton gamma camera for this task is a coaxial system of two cylinder detectors (SD and AD) of considerable size (about 1 m in diameter and in length). The only economically available detectors with such a size may be gaseous detectors. The aim of the computer simulation is to optimize the parameters of the gaseous scatter and absorber detectors of the Compton gamma camera to obtain the highest possible registration efficiency for gamma rays emitted by radioisotope Tc-99m (140 kev). 2

3.2. Selection of the optimal parameters of the gaseous detectors Selection and optimization of parameters of gaseous detectors are carried out through Monte Carlo simulation in Gate\Geant4 environment, and the simulation results were analyzed in the MathLab environment. The results of simulation have demonstrated that the most suitable gas for the SD is argon (η C >80%) and for the AD is xenon (η PE > 90%). Required detection efficiency can be achieved by increasing the gas pressure or the thickness of the detector. Increasing the pressure from 10 bar to 20 bar in the SD at a constant scattering efficiency causes the grow of multiple Compton scattering. The ratio of probabilities of single (η C ) and multiple (η MC ) Compton scattering (κ C = η C / η MC ) reduce from 30 to 18. Multiple scattering worsen the quality of the reconstructed image. Optimal parameters for SD gaseous chamber are thickness of about 35 cm and argon pressure of 15 bar. It provides the scattering efficiency ε C = 10 % and the single/multiple ratio κ C = 22. Similar calculations for the AD have shown that increasing the pressure at the same photoelectric absorption efficiency (η Ph ) causes growing the probability of photoelectric effect following Compton scattering (η C+Ph ), which also worsen image quality. The ratio of probabilities (κ Ph = η Ph / η C+Ph ) decreases from ~ 11 to ~ 9 by increasing the pressure from 10 bar to 20 bar. The obtained results demonstrate that the optimal parameters for AD gaseous chamber is a thickness of ~ 30 cm and xenon pressure of 15 bar. This ensure the effectiveness of the photoelectric absorption ε Ph = 72% with the accept able contribution of multiple events κ Ph = 9.5. 3.3. Accuracy of the interaction position determination Position of Compton scattering or photo absorption events determined by the detector corresponds to the center of ionization region created by the secondary electrons. To estimate the spatial resolution of SD and AD the deviations of event position from the real point of interaction were calculated. a b Figure 2. Distribution of deviation ΔR of determined position from the real point of interaction: а) position accuracy for Compton scattering in Ne and Ar (50 kev), for photo absorption in Хе (110 kev); b) position accuracy for photo absorption in Хе (photoelectron 110 kev, characteristic X- rays 30 kev and the total photoelectric absorption (140 kev). For Ar-SD typical energy of secondary electron is 50 kev, while for Xe-AD 110 kev. The simulation results are presented in the Figure 2a. Average deviation ΔR е for Ar-SD and Xe-AD is ~ 1 mm with a mean square deviation of 0.3-0.4 mm. However, the photoelectric effect in Xe-AD is followed by a characteristic radiation with an energy of ~ 30 kev. It leads to an additional error in determination of the coordinates of the photo absorption. The deviation of the characteristic radiation absorption from the point of X-ray emission are shown in the Figure 2b. The mean deviation of the coordinate for X-ray in Xe-AD is ΔR γ = 8 mm. The energy 3

weighted average deviation of the coordinate for "total photo absorption " is ΔR е+γ = 2.4 mm, and the FWHM of this distribution is about 3 mm. If the detector electronics able to distinguish between single and multiple interactions, coordinate accuracy in Xe-AD can reach 1 mm. 4. Events selection in the gaseous compton camera For identification of the initial γ-ray direction two coincident events have to be registered: Compton scattering in SD and photo absorption in AD. But there are some other combinations of physical events which cause coincidence in SD and AD. Computer simulations of the cylindrical Compton gamma camera were carried out using Gate/Geant4 framework for the development of algorithm for the proper coincidence selection. Mathematical model of the camera consists of two enclosed cylinders - Ar-SD and Xe-AD (Figure 3). b Point-like gamma source Absorption detector (Xe) Z S Metal septum (1 mm) X Figure 3. Model of the cylindrical Ar-Хе Compton gamma camera. The gas pressure in both detectors is 15 bar. The inner diameter of the SD is 400 mm, outer diameter is 1100 mm. The inner diameter of the AD is 1100 mm, outer 1700 mm. Between the detectors is a thin (1 mm) metal septum is placed. Detectors length is 500 mm. A point-like isotropic mono energetic gamma radioactive source with energy of 140 kev is placed in the center of the gamma camera. 10 6 initial γ-rays were simulated. All the coincidences observed in Ar-Xe gamma camera were divided into eight types (see Table 1). Coincidences of the "0" type without Compton scattering occur with the photoelectric effect of the primary gamma rays in the Xe-AD and the subsequent registration of the characteristic radiation (~ 30 kev) in the Ar-SD. These coincidences do not provide any useful information and are background. Coincidences of the "1" type - "useful" coincidences, caused by Compton scattering of the primary gamma quanta in Ar-SD and following photoelectric absorption of the scattered radiation in the Xe- AD. Coincidences of the "2-7" type, resulting from the multiple Compton scattering of the primary gamma rays in the SD and the AD distort useful information. Table 1 shows the classification of different types of coincidences and their relative contribution. The efficiency of coincidences registration is ~ 3% when the restriction only on the total deposited energy (E SD +E AD ) > 100 kev is used. The efficiency become 1% when selecting the energy as follows 1 < E SD < 29 kev, E AD > 70 kev and (E SD +E AD ) > 100 kev. Table 1 demonstrate that such selection lead to the maximal contribution of "useful" coincidences of the type "1" is (87.9%); the contribution of coincidences of the type "0" is reduced from 15.8% to 0.2%; the contribution of coincidences of the 4

type "4" (two Compton scatterings + Photo absorption) is practically unchanged (~ 12%).The contribution of other types does not exceed 0.5%. It is important to take into account that about 50% of the gamma quanta emitted from the source escape through the open ends of the gamma camera without interaction. Table 1. Classification and the contribution of the different types of coincidences («C» Compton scattering, «Ph» photoelectric effect) Type 0 Without an energy selection in each detector 15,8%, (background) Contribution With an energy selection in each detector 0,2%, (background) Origin of coincidences 1 st event 2 nd event PE photoelectron registration (110 kev) and a characteristic γ-quantum (30 kev) with different detectors (Ph) 1 67,2%, 87,9%, (useful) (useful) SD (one C) AD (Ph) 2 1,5% 0% AD (one C) SD (Ph) 3 1,1% 0,1% SD (one C) AD (one C) 4 12,2% 11,5% SD (two C) AD (Ph) SD (one C) AD (one C+Ph) 5 0,5% 0% AD (two C) SD (Ph) AD (one C) SD (one C+Ph) 6 0,5% 0% AD (one C) SD (one C) 7 1,1% 0,3% More than 2C 5. Spatial Resolution of the point-like gamma sources To estimate the spatial resolution of cylindrical Compton camera the image of two point-like gamma sources was simulated. For the image reconstruction algorithm of simple cone back projections in combination with iterative maximum likelihood estimation method (MLEM) was used [4]. Figure 4 shows a reconstructed image of two point-like gamma sources (140 kev) placed symmetrically with respect to Z axis at a distance of 80 mm from each other. Reconstruction with a simple cone back projections method provides a spatial resolution of ~ 40 mm. A subsequent usage of the MLEM allowed to improve the spatial resolution up to ~ 20 mm. Selection of the most effective image reconstruction algorithm concerned with the design of the Compton gamma camera and largely determine its spatial resolution. It should be noted that unlike the flat Compton camera spatial resolution in the cylindrical Compton chamber is uniform throughout the field of view. 5

Figure 4. Reconstructed OXY plane images of two point-like isotropic gamma sources (140 kev) placed at the distance ±40 mm from Z: а) simple cone backprojection method; б) iterative maximum likelihood estimation method (after8iterations). 6. Conclusion The following results and operating parameters of the detector were obtained via computer simulations: argon scatter detector with a thickness of 350 mm (at 15 bar) provides an efficiency of 140 kev gamma radiation scattering ~ 10 % and a spatial resolution ~ 1 mm; xenon absorption detector with a thickness of 300 mm (at 15 bar) provides an efficiency photoelectric absorption of scattered radiation~ 70 % and a spatial resolution ~ 3 mm; efficiency of registration of coincidences which are selected based on the energy deposit in both detectors is ~ 1 %; FWHM of the point-like gamma source spread function (spatial resolution) obtained with a simple backprojection reconstruction method is ~ 40 mm, and ~ 20 mm with a reconstruction based on the iterative MLEM algorithm. Acknowledgements This article was prepared with the support of NRNU MEPhI Academic Excellence Project. References [1] «Webb s physics of medical imaging» / M. A. Flower. CRC Press, 2012 [2] GAMOS 3.0.0 User s Guide, 2011, <http://fismed.ciemat.es/gamos/ [3] Bolozdynya A., Egorov V., Koutchenokov A.at. al.,nim, A385, p. 225-238, 1997 [4] T. Hebert, R. Leahy, M. Singh, Opt. Soc. Amer., v. 7, N 7, p. 1305-1313, 1990 6