Sensitive multiplex spectroscopy in the molecular fingerprint 2.4 µm region with a Cr 2+ :ZnSe femtosecond laser

Similar documents
Supercontinua for high resolution absorption multiplex infrared spectroscopy

Graphene mode-locked Cr:ZnS chirped-pulse oscillator

Molecular fingerprinting with femtosecond lasers

Observation of spectral enhancement in a soliton fiber laser with fiber Bragg grating

Optical Frequency Comb Fourier Transform Spectroscopy with Resolution beyond the Path Difference Limit

DIODE- AND DIFFERENCE-FREQUENCY LASER STUDIES OF ATMOSPHERIC MOLECULES IN THE NEAR- AND MID-INFRARED: H2O, NH3, and NO2

Remote Measurement of Emissions by Scanning Imaging Infrared Spectrometry

Ho:YLF pumped HBr laser

Microjoule mode-locked oscillators: issues of stability and noise

Weak water absorption lines around and 1.66 lm by CW-CRDS

Nonlinear Optics (WiSe 2016/17) Lecture 9: December 16, 2016 Continue 9 Optical Parametric Amplifiers and Oscillators

Strongly enhanced negative dispersion from thermal lensing or other focusing effects in femtosecond laser cavities

Multipulse Operation and Limits of the Kerr-Lens Mode-Locking Stability

Frequency combs for ultrasensitive molecular detection

Richard Miles and Arthur Dogariu. Mechanical and Aerospace Engineering Princeton University, Princeton, NJ 08540, USA

MODELING OF NO AND OH INFRARED RADIATION ON A CONVERGENT-DIVERGENT NOZZLE

High sensitivity CW-cavity ring down spectroscopy of N 2 O near 1.5lm (I)

Line Intensities in the ν 6 Fundamental Band of CH 3 Br at 10 µm

MEFT / Quantum Optics and Lasers. Suggested problems Set 4 Gonçalo Figueira, spring 2015

Multi-cycle THz pulse generation in poled lithium niobate crystals

New Concept of DPSSL

Infrared quantitative spectroscopy and atmospheric satellite measurements

Group interactions of dissipative solitons in a laser cavity: the case of 2+1

Let us consider a typical Michelson interferometer, where a broadband source is used for illumination (Fig. 1a).

High Sensitivity Gas Sensor Based on IR Spectroscopy Technology and Application

Experimental studies of the coherence of microstructure-fiber supercontinuum

Dark Soliton Fiber Laser

3.5 Cavities Cavity modes and ABCD-matrix analysis 206 CHAPTER 3. ULTRASHORT SOURCES I - FUNDAMENTALS

Fourier Transform Infrared. Spectrometry

Laser Physics OXFORD UNIVERSITY PRESS SIMON HOOKER COLIN WEBB. and. Department of Physics, University of Oxford

Dissipative soliton resonance in an all-normaldispersion erbium-doped fiber laser

Fiber Gratings p. 1 Basic Concepts p. 1 Bragg Diffraction p. 2 Photosensitivity p. 3 Fabrication Techniques p. 4 Single-Beam Internal Technique p.

Time resolved optical spectroscopy methods for organic photovoltaics. Enrico Da Como. Department of Physics, University of Bath

No. 9 Experimental study on the chirped structure of the construct the early time spectra. [14;15] The prevailing account of the chirped struct

DEVELOPMENT OF A NEW RADIATION SCHEME FOR THE GLOBAL ATMOSPHERIC NWP MODEL

Alexander Gaeta Department of Applied Physics and Applied Mathematics Michal Lipson Department of Electrical Engineering

Linear pulse propagation

Eximer Lasers UV laser sources Consist two atom types which repel each other eg nobel gas and halide or oxide When excited/ionized atoms attract

Femtosecond Spectral Hole Burning Spectroscopy as a Probe of Exciton Dynamics in Quantum Dots

IMPLEMENTATION OF A LONG PATH MULTI-REFLECTION OPTICAL CELL WITH A MID-INFRARED FREQUENCY COMB LASER SOURCE FOR SENSITIVE MOLECULAR SPECTROSCOPY

Optical Spectroscopy of Advanced Materials

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626

IN RECENT YEARS, Cr -doped crystals have attracted a

Calibrating our Wavemeter for Spectroscopy

An assessment of the accuracy of the RTTOV fast radiative transfer model using IASI data

Spectral Fraunhofer regime: time-to-frequency conversion by the action of a single time lens on an optical pulse

Hiromitsu TOMIZAWA XFEL Division /SPring-8

Bound-soliton fiber laser

Vibrational Spectra of Chloroform, Freon-11 and Selected Isotopomers in the THz Frequency Region

Feedback-controlled Raman dissipative solitons in a fiber laser

Multi-Line Fit Model for the Detection of Methane at ν 2 + 2ν 3 Band using Hollow-Core Photonic Bandgap Fibres

ABSTRACT 1 INTRODUCTION

LASERS. Amplifiers: Broad-band communications (avoid down-conversion)

Self-Phase Modulation in Optical Fiber Communications: Good or Bad?

Mid-IR Photothermal Imaging with a Compact Ultrafast Fiber Probe Laser

Performance Limits of Delay Lines Based on "Slow" Light. Robert W. Boyd

Numerical Analysis of Soft-Aperture Kerr-Lens Mode Locking in Ti:Sapphire Laser Cavities by Using Nonlinear ABCD Matrices

Introduction to FT-IR Spectroscopy

Ultrafast Laser Physics

Direct measurement of spectral phase for ultrashort laser pulses

Vector dark domain wall solitons in a fiber ring laser

B 2 P 2, which implies that g B should be

XUV frequency comb development for precision spectroscopy and ultrafast science

RECOMMENDATION 1 (CI-2002): Revision of the practical realization of the definition of the metre

Dispersion and how to control it

Schemes to generate entangled photon pairs via spontaneous parametric down conversion

Ultrafast laser oscillators: perspectives from past to futures. New frontiers in all-solid-state lasers: High average power High pulse repetition rate

Detection of ozone for use as an extrasolar biosignature

Quantum Electronics Laser Physics PS Theory of the Laser Oscillation

APPLICATION NOTE. Supercontinuum Generation in SCG-800 Photonic Crystal Fiber. Technology and Applications Center Newport Corporation

Spectral phase optimization of femtosecond laser pulses for narrow-band, low-background nonlinear spectroscopy

Effect of cross-phase modulation on supercontinuum generated in microstructured fibers with sub-30 fs pulses

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

Survey on Laser Spectroscopic Techniques for Condensed Matter

Highly Efficient and Anomalous Charge Transfer in van der Waals Trilayer Semiconductors

Where are the Fringes? (in a real system) Div. of Amplitude - Wedged Plates. Fringe Localisation Double Slit. Fringe Localisation Grating

Astronomy 203 practice final examination

CHAPTER 7 SUMMARY OF THE PRESENT WORK AND SUGGESTIONS FOR FUTURE WORK

Slow, Fast, and Backwards Light: Fundamentals and Applications Robert W. Boyd

Supplementary Information for Mid-Infrared Optical Frequency Combs at 2.5 µm based on Crystalline Microresonators

Excimer Lasers Currently best UV laser sources Consist two atom types which repel each other eg nobel gas and halide or oxide which normally do not

All-Optical Delay with Large Dynamic Range Using Atomic Dispersion

Graphene Based Saturable Absorber Modelockers at 2µm

PRESSURE-DEPENDENT LINE PARAMETERS OF CO 2 BY MULTISPECTRUM LEAST-SQUARES FITTING

Terahertz absorption spectra of highly energetic chemicals

1 Mathematical description of ultrashort laser pulses

PAPER No. 12: ORGANIC SPECTROSCOPY MODULE No. 7: Instrumentation for IR spectroscopy

FEMTOSECOND MID-INFRARED SPECTROSCOPY OF HYDROGEN-BONDED LIQUIDS

Fourier transform infrared spectroscopy (FTIR) is a method used to obtain an infrared

Isotopic Ratio Measurements of Atmospheric Carbon Dioxide Using a 4.3 µm Pulsed QC Laser

Coherent control of light matter interaction

Hefei

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES

Development of Polarization Interferometer Based on Fourier Transform Spectroscopy for Thomson Scattering Diagnostics

Optical solitons and its applications

The near infrared ( µm) absorption spectrum of methane down to 77 K

Molecular spectroscopy, besides its own interest. Infrared Spectroscopy and the Atmosphere

LINC: an Interdepartmental Laboratory at ENEA for Femtosecond CARS Spectroscopy

Edward S. Rogers Sr. Department of Electrical and Computer Engineering. ECE318S Fundamentals of Optics. Final Exam. April 16, 2007.

SUB-NATURAL-WIDTH N-RESONANCES OBSERVED IN LARGE FREQUENCY INTERVAL

Transcription:

Sensitive multiplex spectroscopy in the molecular fingerprint 2.4 µm region with a Cr 2+ :ZnSe femtosecond laser E. Sorokin, I. T. Sorokina 2, J. Mandon 3, G. Guelachvili 3, and N. Picqué 3* Institut für Photonik, TU Wien, Gusshausstrasse 27/387, A-040 Vienna, Austria, 2 Department of Physics, NTNU, N-749 Trondheim, Norway, 3 Laboratoire de Photophysique Moléculaire, CNRS, Université Paris-Sud, Bâtiment 350, 9405 Orsay Cedex, France * Corresponding author: nathalie.picque@u-psud.fr; http://www.laser-fts.org e.sorokin@tuwien.ac.at; sorokina@ntnu.no Abstract: An ultrashort-pulse Cr 2+ :ZnSe laser is a novel broadband source for sensitive high resolution molecular spectroscopy. A 30-fs pulse allows covering of up to 380 cm - spectral domain around 2.4 µm which is analyzed simultaneously with a 0.2 cm - (3.6 GHz) resolution by a Fouriertransform spectrometer. Recorded in 3 s, from 70-cm length absorption around 450 cm -, acetylene and ammonia spectra exhibit a 3800 signal-tonoise ratio and a 2.4 0-7 cm - Hz -/2 noise equivalent absorption coefficient at one second averaging per spectral element, suggesting a 0.2 ppbv detection level for HF molecule. With the widely practiced classical tungsten lamp source instead of the laser, identical spectra would have taken more than one hour. 2007 Optical Society of America OCIS codes: (40.4050) Mode-locked lasers; (40.3070) Infrared and far-infrared lasers; (40.7090) Ultrafast lasers; (300.6300) Spectroscopy, Fourier transforms; (300.6390) Spectroscopy, molecular; (320.0320) Ultrafast optics. References and links. J. Mandon, G. Guelachvili, N. Picqué, F. Druon, P. Georges, Femtosecond laser Fourier transform absorption spectroscopy, Opt. Lett. 32, 677-679 (2007). 2. M. J. Thorpe, D. D. Hudson, K. D. Moll, J. Lasri, and J. Ye, Cavity-ringdown molecular spectroscopy based on an optical frequency comb at.45-.65 µm, Opt. Lett. 32, 307-309 (2007). 3. S. A. Diddams, L. Hollberg, and V. Mbele, Molecular fingerprinting with the resolved modes of a femtosecond laser frequency comb, Nature 445, 627-630 (2007). 4. M. J. Thorpe, K. D. Moll, R. J. Jones, B. Safdi, and J. Ye, Broadband cavity ringdown Spectroscopy for sensitive and rapid molecular detection, Science 3, 595-599 (2006). 5. A. Schliesser, M. Brehm, F. Keilmann, and D. van der Weide, Frequency-comb infrared spectrometer for rapid, remote chemical sensing, Opt. Express 3, 9029-9038 (2005). 6. J. Mandon, G. Guelachvili, and N. Picqué are preparing a manuscript to be called Doppler-limited multiplex frequency comb spectrometry. 7. J. Hult, R. S. Watt, and C. F. Kaminski, High bandwidth absorption spectroscopy with a dispersed supercontinuum source, Opt. Express 5, 385-395 (2007). 8. I. T. Sorokina, E. Sorokin, and T. Carrig, Femtosecond pulse generation from a SESAM mode-locked Cr:ZnSe Laser, CLEO/QELS, Technical Digest on CD, (Optical Society of America, 2006) paper CMQ2. 9. I. T. Sorokina and E. Sorokin, Chirped-mirror dispersion controlled Femtosecond Cr:ZnSe Laser, in Advanced Solid-State Photonics, OSA Technical Digest Series (CD) (Optical Society of America, 2007), paper WA7. 0. I. T. Sorokina, Broadband mid-infrared solid-state lasers, in Mid-Infrared Coherent Sources and Applications, M. Ebrahim-Zadeh and I. T. Sorokina, eds., (Springer-Verlag 2007), pp. 225 260.. N. Picqué, F. Gueye, G. Guelachvili, E. Sorokin, and I. T. Sorokina, Time-resolved Fourier transform intracavity spectroscopy with a Cr 2+ :ZnSe laser, Opt. Lett. 30, 340-342 (2005). 2. R. D'Cunha, Y. A. Sarma, G. Guelachvili, R. Farrenq, Q. Kou, V. M. Devi, D. C. Benner, and K. N. Rao, Analysis of the high-resolution spectrum of acetylene in the 2.4 µm region, J. Mol. Spectrosc. 48, 23-225 (99). 3. L. S. Rothman, D. Jacquemart, A. Barbe, D. Chris Benner, M. Birk, L. R. Brown, M. R. Carleer, C. Chackerian, Jr., K. Chance, L. H. Coudert, V. Dana, V. M. Devi, J.-M. Flaud, R. R. Gamache, A. Goldman, J.-M. Hartmann, K. W. Jucks, A. G. Maki, J.-Y. Mandin, S. T. Massie, J. Orphal, A. Perrin, C. P. Rinsland, (C) 2007 OSA 0 December 2007 / Vol. 5, No. 25 / OPTICS EXPRESS 6540

M. A. H. Smith, J. Tennyson, R. N. Tolchenov, R. A. Toth, J. Vander Auwera, P. Varanasi, G. Wagner, The HITRAN 2004 molecular spectroscopic database, J. Quantum Spectrosc. Radiat. Transf. 96, 39-204 (2005). 4. Š. Urban, N. Tu, K. Narahari Rao, and G. Guelachvili, Analysis of high-resolution Fourier transform spectra of 4 NH 3 at 2.3 µm, J. Mol. Spectrosc. 33, 32-330 (989). 5. J. Mandon, E. Sorokin, I. T. Sorokina, G. Guelachvili, and N. Picqué, Infrared frequency combs and supercontinua for multiplex high sensitivity spectroscopy, Annales de Physique, submitted (2007). 6. E. Sorokin, S. Naumov, and I. T. Sorokina, Ultrabroadband Infrared Solid-State Lasers, IEEE J. Sel. Top. Quantum Electron., 690-72 (2005).. Introduction Mode-locked lasers and femtosecond frequency combs actively enter [-6] the field of broadband spectrometry. In particular, their high brightness allows improving the signal to noise ratio (SNR) of the traditional absorption spectroscopy []. For spectroscopic applications, the mid-infrared spectral region is of special interest, as most of molecules have intense vibration-rotation transitions lying in this domain. Developing sensitive techniques in this region results therefore in improved detection levels. High resolution is also advantageous because the Doppler profiles of the infrared lines are narrow and their resolved observation increases the sensitivity. The accuracy of the absolute frequency reading is required for unambiguous identification of the lines and species. Additionally, the broadband simultaneous detection is often necessary, for instance, for multi-species trace gas sensing or for accurate renewed molecular theoretical approaches. Most of the traditional spectrometric approaches encounter severe limitations if these four features are to be met simultaneously. For example, the multichannel grating spectrometers have poor efficiency both from the point of view of spectral resolution and broadband coverage; and wavelength agile laser techniques [7] have limited accuracy. In addition, the mid-infrared region has been suffering from a lack of convenient broadband laser sources, while the spectral radiance of incoherent sources is insufficient for sensitive measurements. A broadband mid-infrared approach to high resolution sensitive spectroscopy is therefore highly desirable. A solution to this challenging problem may be found in the multiplex high resolution spectral analysis using the newly developed femtosecond mode-locked lasers and frequency comb sources. In this paper, we report the implementation of a new spectrometric method based on a femtosecond laser taking the full benefit of the direct access to the mid-infrared wavelength range. We apply the first femtosecond solid-state mode-locked laser source in the molecular fingerprint region above 2 µm to high resolution spectroscopy. The radiation of a Cr 2+ :ZnSe mode-locked laser is absorbed by the gas sample and analyzed by a commercial Fourier transform (FT) spectrometer. Examples of acetylene and ammonia spectra covering up to 380 cm - in the 450 cm - region are shown. 2. Experiment Atten. Fourier-transform spectrometer R=00 Cr 2+ :ZnSe R=75 f=40 YAG R=50 Er:fiber λ=.6 µm Absorption cell OC 2% SESAM on heatsink CM Fig.. Experimental set-up. CM: chirped mirror. OC: output coupler. Atten: attenuator. The first sub-ps mode-locked operation of Cr 2+ :ZnSe lasers has been reported recently [8], with pulse duration down to 80 fs at 80 mw of output power [9,0]. In the present experiment, (C) 2007 OSA 0 December 2007 / Vol. 5, No. 25 / OPTICS EXPRESS 654

we have set up a prismless femtosecond laser following the design of Ref. [9], optimized for a long-term hands-free stable operation. The laser is based on a 4 mm thick Brewster-cut Cr 2+ :ZnSe crystal (Fig. ). The crystal temperature is maintained at 20.5 C by a recirculating water chiller system. The astigmatically compensated X-fold cavity consists of 75-mm and 00-mm radius-of-curvature dichroic mirrors, a 50-mm radius of curvature chirped mirror focusing the light onto an InAs/GaSb semiconductor saturable absorber mirror (SESAM), and an output coupler with transmission of 2 % at 400 cm -. Optical pumping is achieved with a 607 nm Er-doped fiber laser, focused onto the crystal by an uncoated lens of 40 mm focal length. The self-starting mode-locking results from the use of the SESAM, which has several hundreds of picoseconds relaxation time. Dispersion compensation is provided by a combination of a 6.5 mm thick uncoated YAG plate and the spherical chirped mirror. Combined with the dispersion of the Cr:ZnSe active medium, this provides about -500 fs 2 total round-trip group delay dispersion (GDD), which is practically constant over 500 cm -. The cavity length is about 75 cm, corresponding to 200 MHz pulse repetition rate. In order to get rid of the strong absorption lines of water vapor, which are present in the vicinity of 2.4 µm, the oscillator is placed inside a sealed enclosure. The enclosure is equipped with windows made of BK7 on the Er:fiber laser side and of CaF 2 on the Cr 2+ :ZnSe laser output side. The enclosure is first evacuated under primary vacuum conditions and then filled with dry nitrogen. Under these conditions, with.9 W of pumping power, stable mode-locked operation is obtained during hours with ~50 mw average output power. The main stability limitation has been identified as arising from temperature increase of the SESAM heatsink, which was not attached to the cooling circuit, and resulted in slow decrease of the output power (and pulse broadening) during the laser operation. Fig. 2 provides a low resolution (30 GHz - cm - ) spectrum and an interferometric autocorrelation trace of the laser pulses. The full width at half maximum (FWHM) of the autocorrelation trace is 250 fs, which leads to a pulse width of the order of 30 fs, consistent with the 76 cm - measured FWHM of the corresponding spectrum, which has almost perfect sech 2 profile, as shown on Fig. 2. The pulse duration could be decreased by dispersion optimization [9], but we opted to improve the operation stability by providing some additional negative GDD. (a) (b) Measured sech 2 fit Spectral intensity (arb. units) Autocorrelation (arb. units) 4000 400 4200 4300 Wavenumber (cm - ) -400-200 0 200 400 Time delay (fs) Fig. 2. (a). Spectrum (FWHM: 76 cm - ) and (b) interferometric autocorrelation (FWHM: 250 fs) of a 30-fs pulse of the Cr 2+ :ZnSe mode-locked laser. Pump power:.9 W, output power: 50 mw, repetition rate 200 MHz. The laser emission is centered at 457 cm -. As compared to the open-air operation, the emission peak is not shifted towards longer wavelengths. This is different from our previous (C) 2007 OSA 0 December 2007 / Vol. 5, No. 25 / OPTICS EXPRESS 6542

observation of the non-stationary continuous-wave operation of the Cr 2+ :ZnSe laser under secondary vacuum [], where the decrease of the water absorption (growing towards smaller wavenumbers) after air removal resulted in a shift of the laser wavelength by ~75 cm - to the infrared. More surprisingly, the pulse spectral width also doesn t increase when the air humidity absorption disappears. Our explanation is that the high losses of the mode-locked setup (2% initial and ~6% saturated absorption in the SESAM alone [9]) significantly exceed the integrated absorption in the narrowband water vapor lines, thus making the setup almost insensitive to the air humidity. After the exit window, the laser pulses pass through a 70 cm-long single-pass absorption cell, filled with the gas of interest, and ~5 m of open-air propagation before finally reaching the spectrometer. The laser radiation is analyzed by a commercial FT spectrometer (max. resolution 0.2 cm - ) equipped with a fluorine beam-splitter and a thermoelectrically cooled InAs detector. The laser beam has to be attenuated in order not to saturate the detector. R ee (8) ν +ν 5 ν 3 (7,5,3)-(6,3,4) H 2 O ν 3 (8,5,3)-(7,3,4) H 2 O R ee (9) ν +ν 5 433 434 (cm - ) 435 4050 400 (cm - ) 450 Fig. 3. Part of a C 2 H 2 spectrum at 0.2 cm - resolution in the 2.4 µm region. The baseline modulation is not noise-related but due to a channel spectrum from an unwedged window. The strongest spectral features are the P, Q and R branches of the ν +ν 5 band [2] of 2 C 2 H 2, centered at 4090 cm -. The intensity alternation is obvious in the R-branch. The upper left inset is a zoom showing acetylene and residual water vapor lines coming from a 5 meters air path between the cell and the interferometer. The H 2 O lineshapes reveal an atmospheric-pressure broadening. The rotational assignment for the H 2 O lines [3] gives (J,K a,k c ). As a first demonstration of the capabilities of the source for broadband absorption spectroscopy, the rovibrational spectra of acetylene and ammonia have been recorded around 450 cm -. Figure 3 gives an illustration of the C 2 H 2 spectrum, in the region of the ν +ν 5 band [2]. The full width at half maximum of the spectrum is 75 cm -. Resolution is limited by the spectrometer to 3.6 GHz (0.2 cm - ). With the total recording time T=3 s we benefit from an uppermost SNR better than 3800. With spectral boundaries defined by a SNR relatively degraded by a factor 000, the spectral range extends over 380 cm -, from 3980 cm - to 4360 cm -. This corresponds to a number of spectral elements M equal to 367. The cell is filled with acetylene at 23 hpa (7 Torr) pressure. The absorption lines shown on Fig. 3 are due to the acetylene bands and the water vapor lines [3] from the open-air propagation. Figure 4 displays a portion of a NH 3 recorded absorption spectrum, with a NH 3 pressure equal to 26 (C) 2007 OSA 0 December 2007 / Vol. 5, No. 25 / OPTICS EXPRESS 6543

hpa (96 Torr). The spectral lines of ammonia belong to the ν + ν2 and ν2 + ν3 rovibrational combination bands [4]. As the spectrum is complex and crowded, high resolution is essential to discriminate against the various lines. The recording conditions are the same as for the acetylene spectrum. 3. Results and discussion In these spectra, the noise equivalent absorption coefficient (NEA) reaches 3.7 0-6 cm -. NEA at one second averaging per spectral element (L SNR) - (T/M) /2 = 2.4 0-7 cm -. Hz -/2, where L is the absorption path length. In the present spectral region, this level of sensitivity allows for the high resolution detection of 22 parts per billion by volume (ppbv) of C 2 H 2, and 60 ppbv of NH 3 at s of integration time per spectral element. These detection limit values are still relatively high because the combination bands of C 2 H 2 and NH 3 in this region are weak: their intensity, of the order of 0-2 cm molecule -, is not higher than in the.5 µm region. For hydrogen fluoride HF (-0 band, with intensity of the R(2) line at 4075 cm - equal to 2.3 0-8 cm molecule - [3]), the corresponding detection level is 200 parts per trillion by volume (pptv) in the 2.4 µm spectral region. 4000 400 (cm - ) 4200 4300 00% Transmission 50% 430 440 (cm - ) 450 460 Fig. 4. Spectrum of NH 3 in the 2.4 µm region illustrating the spectral bandwidth and SNR capabilities of the spectrometric technique. The ν + ν 2 and ν 2 + ν 3 combination bands are observed. The upper graph shows the whole spectral domain covered in a singe recording while the lower graph shows a portion expanded in wavenumber and intensity scales. For a fair comparison, we recorded a spectrum using a traditional tungsten lamp, which is the most widely used broadband source in the near- and mid-infrared absorption FTS. Under identical experimental conditions, the SNR was degraded by about 7. The recording time would have then been 300 times longer to get identical results. This is not surprising as this 50 mw femtosecond laser source has a spectral radiance which is about 2.8 0 5 times stronger than a 3000 K blackbody source. Actually, a signal-to-noise ratio enhancement of about 500 should be achieved. This could not be demonstrated in the present experiment: as the 50 mw (C) 2007 OSA 0 December 2007 / Vol. 5, No. 25 / OPTICS EXPRESS 6544

power of the laser saturates the detector, we had to attenuate the beam by more than two orders of magnitude before entering in the interferometer. To make a proper use of the abundant power, one should e.g. employ a classical multipass cell. In this case increase of the absorption path by more than two orders of magnitude would then improve the sensitivity by the same factor, without lowering the SNR. It is also interesting to compare our experimental results with a recent near-infrared combbased experiment [2]. By coupling a 550 nm Er-fiber based frequency comb to a cavity with a finesse higher than 300, Ref. [2] obtains a NEA at one second averaging equal to 2 0-8 cm -.Hz -/2, with an equivalent absorption path length higher than 3000 m. This absorption path length is about 4300 times higher than ours whereas the resulting sensitivity is only one order of magnitude better. Moreover, the resolution of the grating spectrometer is limited to 25 GHz. The acquisition of a broad spectral domain is a sequential, long and inaccurate process, whereas in the present experiment, 3 seconds are enough to measure simultaneously 367 individual spectral elements spanning 380 cm -, that are only limited by the laser emission, at 3.6 GHz resolution. Actually, during these 3 seconds, the FT spectrometer samples a 5800 cm - broad spectral domain by 3650 independent spectral elements. A more detailed discussion of the advantages of using a FT spectrometer for this kind of experiments may be found in []. Further improvements of the present experiment can be made in several directions. The resolution can easily be improved up to the Doppler width of the lines by using a higher resolution interferometer. It is also possible to significantly increase the measured spectral domain. The enormous gain bandwidth of the Cr 2+ :ZnSe laser (850 nm FWHM) allows indeed coverage of a much broader spectral region, either by tuning, or more interestingly by construction of a few-cycle laser source to observe hundreds of nm simultaneously. Judging from the demonstrated cw tuning range of 2000-300 nm [6], coverage of the entire 2-3 µm region is feasible in a Kerr-lens modelocked configuration with proper mirrors [0]. Alternatively, generation of a mid-infrared supercontinuum in highly nonlinear fibers looks promising, as we demonstrated it recently in the near infrared spectral region [5]. Using of a broadband source is especially attractive for FT spectroscopy, where measurement time is independent of the spectral domain width. As discussed above, sensitivity improvement can benefit from absorption path length enhancement either by use of a classical multipass cell or by injection of a fs frequency comb in a high finesse cavity. Use of a comb can bring additional sensitivity gain by performing high frequency synchronous detection, which also leads to the advantage of allowing simultaneous measurement of the absorption and dispersion associated with the spectral features, as experimentally established in [6]. 4. Conclusion Summarizing, we have demonstrated the first spectroscopic application of a mid-infrared femtosecond mode-locked laser used as a broadband infrared source with a high resolution FT spectrometer. This simple experimental setup already exhibits high sensitivity and resolution over a broad spectral domain. Sub-ppb detection levels should be easily obtained for a large panel of molecules thanks to absorption path length enhancement and high frequency detection. Further improvements in resolution and acquisition time are also in progress. This will include taking advantage of the comb structure of a stabilized mode-locked Cr 2+ :ZnSe laser by combination of the complementary methods developed in [4] and in [6]. Acknowledgments P. Jacquet (Laboratoire de Photophysique Moléculaire, Orsay) is warmly thanked for his participation to the experiments. I. T. Sorokina acknowledges Université Paris-Sud for a position of invited professor. This work is accomplished in the framework of the Programme Pluri-Formation de l Université Paris-Sud Détection de traces de gaz 2006-2009 and the Austrian Fonds zur Förderung der wissenschaftlichen Forschung project P7973. (C) 2007 OSA 0 December 2007 / Vol. 5, No. 25 / OPTICS EXPRESS 6545