Anion Analysis of Ophthalmic Irrigation Solutions with the Shimadzu Prominence IC System Craig Young, PhD, Sr. Tech Support Specialist

Similar documents
anion-exchange column

ION CHROMATOGRAPHY SYSTEM S 150

Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

Pelagia Research Library

Rapid Post-Blast Inorganic Explosive Analysis by Suppressed Ion Chromatography

Determination of Chloride, Fluoride, Bromide, Nitrate, Sulphate and Phosphate in Water Sample...

Theory of Ion Chromatography

groups I and II cations, amines, divalent transition metals

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Determination of trace anions in concentrated hydrofluoric acid

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

C190-E105 Printed in Japan ATD

IC Columns Introduction

PHARMACY CALCULATIONS

Buffered and Isotonic Solutions

columns IonPac AS17-C Anion Exchange Column

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

TP Analytical Chemistry II 7 th, December 2012

Suppression in anion chromatography. More sensitive analysis of anions and organic acids

Isotonic and Buffer Solutions

Dionex IonPac AS28-Fast-4µm column

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

4. Ion chromatography (IC) 4.1. Educational aims and objectives

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Waters. Ion Chromatography

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Benzodiazepines and Barbiturates

Analysis of various clays with Metrohm Combustion IC

columns IonPac AS22 Anion-Exchange Column

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Inline sample preparation for the determination of anions in sodium hydroxide

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

SKKU Physical Pharmacy Laboratory 성균관대학교물리약학연구실

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

881 Compact IC pro 882 Compact IC plus. Compact ion chromatography for intelligent, reliable and precise analysis

Cyanide and sulfide analysis using amperometric detection and Metrosep A Supp /4.0

Concentrator and Trap Columns. Column: IonPac AG15, AS15, 2 mm. Eluent Source: EG40 Concentrator. Temperature: 30 C Column: IonPac AC15 (2 50 mm)

"Analysis of Inorganic Anions in Potomac Water, Sediment and Floc by Ion Chromatography" FINAL REPORT. by James E. Girard, Ph.D.

columns IonPac AS18 Anion-Exchange Columns

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

Chromatographic Methods of Analysis Section - 4 : Ion Exchange Chrom. Prof. Tarek A. Fayed

Ion Chromatography (IC)

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

IC5000 Ion Chromatography System

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

881 Compact IC pro / 882 Compact IC plus for ion chromatography. Intelligent. Reliable. Precise.

Ion Chromatography. Anion Exchange. Chromatography Ion Exchange Theory. Dr. Shulamit Levin

Determination of Trace Cations in Power Plant Waters Containing Morpholine

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

ProPac WCX-10 Columns

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Prominence for Ion Analysis

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE)

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

Ion Analysis of Hydraulic Fluids in Fracking Process

Automated Sample Preparation in Quality Control of Eye-Drop Formulation

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Expanded capabilities in ammonium and amine detection

Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography

A novel approach to the analysis of multivitamin by online supercritical fluid extraction/supercritical fluid chromatography

Determination of urea in ultrapure water by IC-MS/MS

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers.

Ion chromatography. Ion chromatography

Subject: Ionic Analysis of PrintedCircuit Boards, Printed board Assemblies and Materials, Ion Chromatography Method

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Thermo Scientific. Anion-Exchange Column

Determination of Adsorbable Organic Halogen in Wastewater

Air Monitoring. Semi-continuous determination of ambient air quality

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Supporting Information. Labeled Ligand Displacement: Extending NMR-based Screening of Protein Targets

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

High Performance Liquid Chromatography

suppressors SRS -ULTRA Self-Regenerating Suppressor

Overcoming Challenging Matrices in Ion Chromatography

International Journal of Pharmacy and Pharmaceutical Sciences Vol 2, Issue 1, 2010

Ion Analysis of Hydraulic Fluids in Fracking Process

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 25: CHROMATOGRAPHIC METHODS AND CAPILLARY ELECTROPHORESIS

LC III: HPLC. Originally referred to as High-Pressure Liquid Chromatography. Now more commonly called High Performance Liquid Chromatography

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Transcription:

High Performance Liquid Chromatography SSI HPLC 5 Anion Analysis of Ophthalmic Irrigation Solutions with the Shimadzu Prominence IC System Craig Young, PhD, Sr. Tech Support Specialist Introduction Ophthalmic solutions suitable for eye irrigation, eye strain relief, contact lens storage and tear replacement abound in the marketplace. While some of these serve as convenient agents for administering medicines in minute quantity, many are formulated to act as cleaning agents and may contain only dissociated ions like chloride or sulfate and organic preservatives such as benzalkonium chloride. The physiologic ph of blood and tears is approximately 7.4, but a ph range of 6.5 to 8.5 will prevent damage to the cornea. This is accomplished using buffers. To prevent eye tissue damage and maximize comfort, solutions are made to be iso-osmotic (equal osmotic pressure) with tears. The common term for this is isotonic. A.9% NaCl solution in pure water is said to be isotonic with tears. In this application note we will aid in the quality control of such ophthalmic solutions for their content of common anions. Conductivity detection has long been used to determine aqueous ions separated by ion exchange. Although the combination of ion exchange separation and conductivity detection is a form of HPLC, it is known as ion chromatography (IC). In its simplest form, an IC is an HPLC system equipped with an appropriate separation column and conductivity detector. Because of the inherently high conductivity of IC mobile phases, a suppressor is commonly placed between the analytical column and the detector to reduce mobile phase background conductivity and increase the analyte signal to noise ratio. Suppressed IC results in 1x greater sensitivity for anionic species when compared to non-suppressed IC. In suppressed IC, the suppressor is a post-column reactor, frequently a strong cation exchange resin which chemically alters the column effluent and lowers the conductivity of the mobile phase. At the same time, a heightening of the conductivity of anions like chloride and sulfate occurs. With such a system, detection limits for the common anions are at the ppb level, making it suitable for the determination of anionic species in EPA-regulated drinking water or ground water supplies. The Shimadzu Prominence HPLC system, when equipped with the SeQuant IC package, yields such anion analysis results reliably and accurately. For purposes of this application note, EPA Method 3.1 parameters will serve as a guideline for the quantitation of the common anions.

SeQuant SAMS /CARS system is comprised of a suppressor column (strong cation exchange resin initially in the Resin-H+ form) and chemical regeneration system. Ion exchange reactions of the suppressor with mobile phase and analyte are as follows: Resin-H + + NaHCO 3 Resin-Na + + H 2 CO 3 Resin-H + + NaCl Resin-Na + + HCl This chemical exchange gradually depletes the resin as a protic source and it must be regenerated. This is done automatically by the CARS regenerant (a high-molecular weight sulfonic acid solution) which flows continually while IC analysis is being performed. Sample Preparation The standard test mixture of seven anions was purchased from Alltech Associates, Inc. and used without modification. Over-the-counter ophthalmic irrigation solutions were purchased from local retailers and samples were diluted 5:1 without further modification. All injection volumes were 2μL. Analytical Conditions (Suppressed IC) Mobile Phase: 3.2mM sodium carbonate / 1.mM sodium bicarbonate Column(s): Shodex IC SI-5 4E, 4. x 25mm, 5μm (polyvinylalcohol gel column) Injection Volume: 2μL Flow Rate:.7mL/min. Column Temp: 4 C Detector Settings: Gain =.1 (1μS/V); Cell Temp = 45 C; Response = 1 Results Under the mobile phase conditions described above, baseline drift was less than 2nS/min. (Method 3.1 specifies drift 5nS/min.). Standards were prepared in triplicate over 5 calibration levels by dilution of stock solution (5ppm chloride): Anion Level 1 Level 2 Level 3 Level 4 Level 5(ppm) R 2 25. 2.5.5.25.5.9994 5. 5. 1..5.1.9994 Nitrite 5. 5. 1..5.1.9999 Bromide 5. 5. 1..5.1.9998 5. 5. 1..5.1.9997 5. 5. 1..5.1.9998 5. 5. 1..5.1.9998 Figure 1: LCsolution calibration screen for bromide standard curve

Figure 2: Method 3.1 Anions Standard Mixture 9 8 7 6 5 4 3 2 1 Nitrite Bromide. 5. 1. 15. 2. 25. 3. min 2μL injection 25ppm; 5ppm; Nitrite 5ppm; Bromide 5ppm; 5ppm; 5ppm; 5ppm 1. 7.5 5. 2.5 Nitrite Bromide. -2.5-5.. 5. 1. 15. 2. 25. 3. min 2μL injection 25ppb; 5ppb; Nitrite 5ppb; Bromide 5ppb; 5ppb; 5ppb; 5ppb

Figure 3 Eye Wash 1 (5:1 dilution) Label Claim: hydroxylalkyl phosphonate, polyoxypropylene, EDTA, boric acid, sodium chloride, polyaminopropyl biguanide 11 1 9 8 7 6 5 4 3 2 1 15. 12.5 1. 7.5 5. 2.5. Analyte RT Estimated Conc. (ppm) Corrected Conc. (ppm) --.. 8.176 57.47 2874 Nitrite --.. Bromide --.. 13.973.124 6.2 21.976.63 3.15 26.644.17 5.35

Eye Wash 2 (5:1 dilution) Label Claim: Sodium chloride, sodium borate, boric acid 1 9 8 7 6 5 4 3 2 1 Bromide 27.5 25. 22.5 2. 17.5 15. 12.5 1. 7.5 5. 2.5. Bromide -2.5 Analyte RT Estimated Conc. (ppm) Corrected Conc. (ppm) 5.67.44 2.2 8.152 5.51 2526 Nitrite --.. Bromide 12.185 2.43 12.2 13.943.316 15.8 --.. 26.583.176 8.8

Eye Wash 3 (5:1 dilution) Label Claim: Buffering soln., polyhexamethylene biguanide, sodium phosphate 35 3 25 2 15 1 5 Analyte RT Estimated Conc. (ppm) Corrected Conc. (ppm) --.. 8.234 9.99 455 Nitrite --.. Bromide --.. 13.974.26 1.3 21.983 8.99 449.5 26.637.139 9.5

Eye Wash 4 (5:1 dilution) Label Claim: Polyoxamine, sodium hydranate, polyaminopropyl biguanide, boric acid, EDTA, sodium borate, sodium chloride 11 1 9 8 7 6 5 4 3 2 1 35 3 25 2 15 1 5 Analyte RT Estimated Conc. (ppm) Corrected Conc. (ppm) 5.611.6 3. 8.175 59.3 2965 Nitrite --.. Bromide --.. --.. 21.974.623 31.15 26.655.45 2.25

System Configuration Part # Description 22-91398-2 CBM-2A System Controller 228-4519-32 DGU-2A5 On-Line Degasser 228-45-32 LC-2AD Pump 22-91397-2 SIL-2A Autosampler 228-451-32 CTO-2AC Column Oven 228-4554-32 CDD-1Avp Conductivity Detector 22-9725-44 Windows XP Pro Data Station 22-91449-5 LCsolution 1.23 software 221-99999-G1 SeQuant (SAMS /CARS ) Anion Suppressor & Regeneration Station References 1. General guide to pharmaceutical formulations, Univ. of North Carolina School of Pharmacy, Chapel Hill, NC 26. 2. Raaidah Saari-Nordhaus and James M. Anderson Jr., Recent Advances in Ion Chromatography Suppressor Improve Anion Separation and Detection, J. of Chromatogr. A, 956 (22) 15. 3. Derick R. Douglas, Raaidah Saari-Nordhaus, Philippe Despres, and James M. Anderson, Jr., New Suppressor Technology Improves Trace Level Anion Analysis with Carbonate/Bicarbonate Mobile Phases, J. of Chromatogr. A, 956 (22) 47. 4. Rakesh Bose, Raaidah Saari-Nordhaus, and James M. Anderson Jr., Solving Complex Anion Separation with carbonate/bicarbonate Gradient, J. of Chromatogr. A, 956 (22) 71. Shimadzu Scientific Instruments Shimadzu Scientific Instruments 712 Riverwood Drive, Columbia, MD 2146 Southern California Regional Office Phone: 8-477-1227, Fax: 41-381-1222 1817 Aston Ave., Carlsbad, CA 928 www.ssi.shimadzu.com Phone: 866-862-1677, Fax: 76-931-9854 webmaster@shimadzu.com