INE-DDB-303A Portable Conductivity Meter Instruction Manual

Similar documents
INE-PHS-3E ph Meter Operating Manual

PC100 Handheld ph/cond./tds/sal. Meter. Instruction Manual

PHS-25CW/3BW Benchtop ph Meter Instruction Manual

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

OPERATING INSTRUCTIONS

P S H - S 3G m t e e t r e (C ( E mar ar e k d e ) P B H - B 4 - Por or ab ab e l PH M t e e t r

Rapid RH Portable ph Meter. User's Manual

symphony Reference Electrodes

This sensor is now also available with a sturdier flat electrode which enables testing even smaller samples than the conventional electrode.

Bante210 Benchtop ph Meter Instruction Manual

Operation Manual. SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

[Agilent Technologies] [Electrochemistry Products] Quick Reference

STANDARD OPERATING PROCEDURE

Chemistry MultiMeasure Sensor PS-2170

DIGITAL ABBE REFRACTOMETER INE-WYA-2S OPERATING INSTRUCTION PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

рн PRO Controller The device is used to control and regulate ph level in hydroponic systems and water preparing units automatically USER MANUAL

ph electrode Instruction Manual For use with the RAH-210 & RPH-250

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Bright Advance Corporation

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Operating Instructions for Digital Manometer. Model: MAN-SD

PH700 Benchtop ph Meter Instruction Manual

INSTRUCTION MANUAL Milwaukee Refractometer MA881 Refractometer for Invert Sugar Measurements

symphony Multisensor Probes

Acceleration/Velocity/Displacement VIBRATION METER

Electromagnetic Flowmeter

Appendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL. Milwaukee Refractometer

10 samples per second

QAM-I-111 Operation and Calibration of the Conductivity Meter

Operating Manual. Turb 355 IR/T. Portable turbidimeter. ba75454e03 09/

OPERATION MANUAL. 3 in 1, IR, Type K/J/R/E/T, Pt 100 ohm THERMOMETER. Model : TM-2000

GENERAL INSTRUCTIONS GENERAL PREPARATION

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

GENERAL INSTRUCTIONS GENERAL PREPARATION

VIBRATION METER. Instruction Manual.

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

Acceleration, Velocity, Separate probe VIBRATION METER Model : VB-8202

ph / Temperature Meter for Cheese with Application Specific Probe

473 Dew Point Hygrometer

One-hand ph/temperature measuring instrument

INSTRUCTION MANUAL. Type: ORP500 C, ORP500 CD, ORP222 CD, ORP223 CD EN Introduction Helpful Operating Tips. 2

473 Dew Point Hygrometer

DOC /2013, Edition 1. User Manual

Metso Concentration Measurement 3300

ph/redox Measurement CPM 280

Richter Optica. Instructions for Model: S850 Stereo Zoom Microscope

Compact ph Measuring Instruments With Pro Engineering

473-SHX Dew Point Mirror

Guide to ph Analysis.

NO 500 DIN, NO 500 BNC NO 800 DIN, NO 800 BNC

Instruction Manual HI Refractometer for Sodium Chloride Measurements

HI Refractometer for Sodium Chloride Measurements INSTRUCTION MANUAL

Measure - Conductivity STANDARD Last std: Sep 10:17 am. Touch meas to measure sample. Touch std to access standardize mode ATC. 25.

Conductivity sensor for hygienic applications

Instruction Manual HI Refractometer for Invert Sugar Measurements

MULTIFUNCTION COMPUTER METER CX-742 USERS MANUAL

BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER. Technical Description Version 2 BGRIMM AUTOMATION

Understanding ph Troubleshooting and Diagnostic information

Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A

Digital Dissolved Oxygen, Conductivity, TDS and ph Meter

Portable ph/orp/ion Meter D-73 Instruction Manual CODE:GZ A

DIGITAL THERMOMETER OPERATOR S MANUAL

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

ph Electrode basics ph Electrodes Model 19 B /1 =1 m 24 B /1 =1 m

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

Operating instructions Magnetic-inductive flow sensor SM6004 SM7004 SM / / 2014

Instruction Manual HI Refractometer for Sucrose in Wine and Grape Products Measurements

CTC 155 / 350 / 660 Dry-block calibrator

Advanced Sensor Technologies, Inc. TEL: TOLL FREE: WOW-ASTI ( ) FAX:

Latest generation testo air velocity & IAQ probes with cable handle. Instruction manual

INSTRUCTION MANUAL Milwaukee Refractometer MA886 Refractometer for Sodium Chloride Measurements

Basic Analytical Techniques, Calorimeter, and Conductivity Meter

Unauthorized reprinting or copying of this operation manual

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL

Dewpoint Meter Data Sheet / Instructions

ph/o.r.p. probe Type 8619 multicell transmitter/controller General data Measuring range Bürkert ph probe Bürkert O.R.P. probe Medium temperature

1) A ph/mv meter or an ion meter, either line operated or portable.

SECTION 1 - WHAT IS A BTU METER? BTU's = Flow x ΔT Any ISTEC BTU Meter System consists of the following main components:

Digital Multi-Day Soil Temperatures Protocol

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

Data Sheet / Instructions. Dewpoint MeterO

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL. Milwaukee Refractometer

Elcometer 116C. Sling Hygrometer. Operating Instructions

Instruction Manual. HI Refractometer for Sucrose in Wine and Grape Products Measurements

Instruction Manual Portamess 911 ph

Digital Multi-Day Max/ Min/Current Air and Soil Temperatures Protocol

National Optical & Scientific Instrument Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210)

Operating Instructions for Ammonium ISE Specifications

ea980r Digital Anemometer User Manual

VIBRATION METER Acceleration & Velocity Model : VB-8201HA

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

WATER ANALYSIS The ph Value. The Redox Potential. ALMEMO ph and Redox Measurement. The Electrical Conductivity

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049

Instruction Manual. Waterproof Portable Colorimeter C401 / 301 / 201 / 101 / 102 / 103 / 104 / 105. Part of Thermo Fisher Scientific

PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION

INSTRUCTION MANUAL Milwaukee Refractometer MA872 Refractometer for Fructose Measurements

INSTRUCTION MANUAL Milwaukee Refractometer MA888 Refractometer for Ethylene Glycol Measurements

Transcription:

INE-DDB-303A Portable Conductivity Meter Instruction Manual Operation ManualPLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com,3 MRC.9.16

Welcome to choose the INE-DDB-303A Portable Conductivity meter, read this instruction in detail before operating the meter for the first time or after a long time from the last operation, it will help you to operate the meter better. The meter must be sent to metrological service or other departments that are qualified to check for checking after being used for one year, and only those qualified ones can be used again. For short-term not in use, immerse the Conductive probe platinum in DI water. If the operating interval more than 6 hours or for long-term storage, wash the probe and store it in a empty protective bottle. After stored or used for some time, re-calibrate the conductive probe because the constant of probe may change. Abbreviation list Abbreviation in meter STD COND Const Coef Temp Meas Cali Explanation Standard Conductivity Cell Constant Temperature Coefficient Temperature Measurement Calibration 1

CONTENT 1Installation 1.1 Tools and Components needed for Installation 1.2 Installation of the Conductivity Meter 1.2.1Installation of the Electrode holder 1.2.2 Installing the conductive probe 2Operation 2.1 Introduction 2.1.1 Definition of Terms Cell constant Temperature coefficient 2.12 Features 2.13 Specifications 2.1.4 Physical overview The Display The Keypad 2.1.5 Uses for the Meter 2.2 Instrument Control Turn on the Meter Setup Calibrating Measuring Mode 2.3 Notice 2.4 How to use back-light key 2.5 To turn off the meter 3 Maintenance 3.1 Maintenance 3.2 Probe cleaning & Storage 3.2.1 Probe cleaning 3.2.2 Probe storage 4 Consumables and Ordering Information 5 2

1 Installation 1.1 Tools and Components needed for Installation To provides the tools needed for installation. The following accessories are included in the shipping case. one pc of Electrode Holder One pc of Conductive Probe, for example, DJS-1C Conductive Probe Battery: two pcs of 5 AA battery 1.2 Installation of the Conductivity Meter 1---- Electrode socket 2----Back- light 3--- Power key 4--- Mode selector 5--- Enter key 6---Digit plus key 7---Digit minus key 8---Electrode holder 1.2.1Installation of the probe holder To put the electrode holder(8) into the electrode holder socket in the left of the meter 1.2.2 Installing the conductive probe To put the conductive probe into the electrode holder(8) To put the conductive probe into the electrode socket in the top of the meter 3

2Operation 2.1 Introduction 2.1.1 Definition of Terms Cell Constant : The cell constant of a pair of platinum sheets is the ratio of distance between the two sheets to the area of each sheet. The ratio is measured in cm-1. Temperature coefficient: The temperature coefficient is the % change of conductivity temperature in %/ C. 2.1.2 Features Model INE-DDB-303A Conductivity Meter is used for measuring conductivity of water solution in laboratories. It is applied widely in the fields of petroleum chemical industry, biological pharmacy, sewage treatment, environmental monitoring, metallurgy, universities and scientific research institutes. If it is fitted with the conductive electrode with suitable constant, it can also be used to measure conductivity of pure water or ultra pure water in electronic semiconductor plant, nuclear energy industry and power stations. Main characters: Four kinds of range, automatic range changing, automatic carryover display, convenient operation, high precise measurement; Power supply: two pieces of No.5 batteries, portable for field operation, with low power consumption; 1 3 bit large screen LCD display, with displaying conductivity value and 2 temperature value simultaneously, smart appearance, clear display. Three kinds of constants electrode (0.01, 1.0, 10) can be selected to match the meter, the meter can not only measure the conductivity of ordinary water solution, but also measure the conductivity of ultra pure water. 4

2.1.3 Specifications Measurement range: (0~10 5 μs cm -1 ). With four kinds of ranges which can be switched automatically. They are: 0.00~19.99μs/cm; 20.0~199.9μs/cm; 200~1999μs/cm; 2000~10000μs/cm; 10.0~100.0mS/cm(to be matched with K=10 conductive probe) Note: To measure high conductivity, a large constant conductive electrode is commonly used. If conductivity is 10000μs cm -1, the electrode with constant 10 is used. Accuracy of electronic unit: ±1.0% (FS) ±1 bit. Accuracy of the meter: 1.5% (FS) ±1 bit. Accuracy of temperature coefficient of electronic unit: ±0.5% / Manual temp compensation range:(15~35), basic standard temp 25. Stability accuracy of electronic unit: ±0.66% (FS) ±1 bit/3h Accuracy of repeatablity of electronic unit: ±0.33% (FS) ±1 bit. Outside dimension: l*b*h(mm) 170*80*35. Weight: 0.3kg Normal operating conditions: a) Ambient temperature: (5~35) ; b) Relative humidity: 85%; c) Power supply DCAA 1.5V No. 5 batteries 2 pcs.; d) No obvious oscillation; e) No strong electromagnetic interference except the geomagnetic field. 5

2.1.4 Physical overview The Conductivity Meter is composed of a meter and a probe system. See Figure as following: The Display : The Keypad (picture) The keypad has five keys. The Meter is usually controlled by the keypad. Functionality automatically changes based on context. Key functions are listed in the Table 1: Table 1 Keypad functions key Function 1 On/Off Meter power switch. Move the selection cursor upward when selecting. Move the selection cursor down when selecting. Mode Switch the display window or parameter when measuring. Enter Confirm the last operation 6

2.1.5 Uses for the Meter Conductivity is a measure of a material s ability to conduct an electric current. The meter measures the conductivity of solutions with high accuracy. When equipped with a probe with the proper cell constant, the conductivity meter can measure the conductivity of ultra pure water. Users can select different probes each with a different cell constant to match the expected conductivity of samples to be measured. Refer to the final packing list for the probes that are Shipped. This section describes the operation of the Meter. There are four major steps to measure conductivity with the meter: Meter setup ( Setup on page ) Prepare the conductive probe ( Prepare Probes on page ) Set or calibrate the cell constant ( Calibrate on page ) Measure conductivity ( Perform a Measurement on page ) 2.2 Instrument Control Turn on the Meter Insert as No.5 batteries; Press ON/OFF key and the meter enters into the state of measurement; Pre-warm for 15 minutes. Picture Setup Setup Temperature In the state of measurement ("MEAS" is displayed on the right bottom), press " " or" " key to enter solution temperature adjusting state (" " indication light is flashing ), press " " or" " key again to adjust temp. Press "Enter" key to exit from temperature adjusting state and enter measuring state. Picture 7

Prepare conductive probe 1. Wash the electrode with DI water fully; then rinse the electrode as 4.3 Method of cleaning and reserve. If it is unstable or mistake continuously, please refer to 4 Maintenance and service. 2. Check samples: If the electrode working properly in standard buffer solution but not normal in samples, please check if there is interfering substance in samples or if the conductivity cell has been mechanical damaged. 3. Platinum black electrode: There is some loose platinum black layer on the surface of platinum slice. It can only be wash with DI water, otherwise the platinum black layer might be damaged and cause inaccuracy. 4. If the performance of the electrode reduces, rinse the platinum slice with absolute alcohol and DI water in proper order. 5. Since the platinum black layer may absorb samples when measuring, must wash the electrode after using. 6. For long-term in use or not in use, the constant cell may change and need to recalibrate. Set or calibrate the cell constant The cell constant for each probe was calibrated and printed on the probe label before shipping. However, transportation, storage, and continued use may change the cell constant to some degree. Recalibrate the conductivity probe before measurement. 1 Select the conductive probe with different cell constants and with different measuring method according to the value of measured medium resistance or the value of conductivity. Note1: When measuring conductivity, to select conductive probe constant correctly, in order to get high precise measurement. Three kinds of conductive probes with constant 0.01, 1.0, 10 can be matched to the meter. To select conductive probe with relative constant according to reference table1 of measuring range. Note2: The conductive probe with constant 0.1or10 are classified into "bright" and "platinum black", platinum probe is usually called platinum black conductive probe,the measuring range of bright conductive probe is usually0~300μs/cm. 8

Table 2 No Cond of solution (μs/cm) Corresponding resistance rate (Ω) Matched electrode Constant Actual conductivity of measured solution 1 0~0.2 ~5000000 Titanium 0.01 Displayed digit 0.01 2 alloy 0~2 ~500000 0.01 Displayed digit 0.01 electrode 3 0~20 ~50000 0.01 Displayed digit 0.01 4 0~200 ~5000 0.01 Displayed digit 0.01 5 0~20 ~50000 DJS-1C 1 Displayed digit 1 6 0~200 ~5000 bright electrode 1 Displayed digit 1 7 0~2000 ~500 DJS-1C 1 Displayed digit 1 8 0~10000 ~100 platinum black electrode 1 Displayed digit 1 9 0~200 ~5000 DJS-10C 10 Displayed digit 10 10 platinum 0~2000 ~500 10 Displayed digit 10 black 11 0~20000 ~50 electrode 10 Displayed digit 10 12 0~100000 ~10 10 Displayed digit 10 2 The constant 1, 10, and 0.01 have been calibrated, strictly when ex-factory, the conductive probe of calibrated cell constant is a constant value that is marked on the conductive probe, (constant labels are stuck on the electrode when ex-factory). Press "Mode" key to enter cell constant calibrating state, display C= ; Press " " or" " key again to adjust cell constant value as shown as following; picture Press Enter key to exit from constant adjusting state and enter measuring state. 9

The adjusting method: If the conductive probe with constant 0.95, adjust the digit to be 950; If it is the conductive probe with constant 11, adjust the digit to be 1.100; If it is titanium alloy conductive probe with constant 0.012, adjust the digit to be 1.200. 2.2.3 Calibration In order to calibrate, you need to make a standard solution. Read the printed cell constant from the label. Use Table 3 to find the suitable KCl concentration of the standard solution. Table 3: Concentration of Standard solution based on the cell constant Electrode constant cm -1 0.01 0.1 1 10 KCL lose to concentration (mol/l) 0.001 0.01 0.0 0.1 or 1 1 Connect conductive probe into the meter; 2 Set temperature value by manual to 25 ; 3 Control temperature of solution to (25±0.1) ; 4 Wash the conductive probe with DI water; 5 Soak the conductive probe in the standard solution. 6 Get the conductivity value is Kmeasurement 7 Calculate the cell constantj: J=K/Kmeasurement IN which: K is the standard conductivity value in Table 5. Calibrating with standard electrode 1 Use a standard conductive probe with known cell constant as Jstandard 2 Select standard solution( see Table 3) according to the cell constant and the Composition of Standard Solution (see in Table 4), 3 Soak the conductive probe with unknown cell constant as J1 and the standard conductive probe with known cell constant(all have been clean) both into solution at same depth; 4 Connect them to conductivity meter in turn, set the measured conductivity to be K1, and K standard respectively; 5 calculate cell constant: J1=J*K standard/k1 In which: K1= it is the conductivity value that the conductive probe with unknown cell constant measured Kstandard = constant measured it is the conductivity value that the conductive probe with known cell 10

Table 4 Composition of Standard Solution Concentration (mol/l) Concentration of KCl in g/l (20 C) 1mol/L: KCl 74.2650g/L at 20 0.1mol/L: KCl 7.4365g/L at 20 0.01mol/L: KCl 0.7440g/L at 20 0.001mol/L: Dilute 0.01mol/L solution from100mol to 1L Table 5 STD Value (μs/cm) Temperature 15.0 C 18.0 C 20.0 C 25.0 C 30.0 C 1 mol/l 92120 97800 101700 11310 131100 0.1 mol/l 10455 11163 11644 12852 15353 0.01 mol/l 1141.1 1220.0 1273.7 1408.3 1687.6 0.001 mol/l 118.5 126.7 132.2 146.5 176.5 Note: 1 Only choose one method of above two. 2 Manually entering a cell constant will overwrite a previous cell constant calibration. 2.2.4 Measurement conductivity During measurement, make sure the tip of the probe is completely submersed in the solution. Place the probe in a location where the solution can flow freely around the measuring tip. To measure ultra pure water, install the conductivity probe in a sealed container with ultra pure water flowing through. The flow rate should be not too high. Choose the correct probe Before measuring, choose the proper conductivity probe. Refer to Table 6 when choosing a conductivity probe. There are two kinds of conductivity probes with cell constant K=1.0: Platinum black - the platinum sheet has a platinum black coating. Bright - the platinum sheet does not have a platinum black coating. The Bright probe is suitable for wider measurement range (2 to 3000 μs/cm). If conductivity is 3000 μs/cm, the measurement may not be as accurate. To measure high conductivity, use a conductivity probe with a high cell constant. 11

When conductivity is 200.00 ms/cm, use a conductivity probe with cell constant = 5 or 10. When conductivity is 500.00 ms/cm, use a conductivity probe with cell constant = 10. Table 7 Conductivity Range and Corresponding Cell Constant Conductivity range Resistivity range Recommended cell (Ω-cm) constant (cm-1) 0.000 19.99 μs/cm 20.00 M 50.0k 0.01 0.20 00 μs/cm 5.00 M 5.00k 0.1 2.000 0 ms/cm 500 5 1.0 20.00 00 ms/cm 50 5 10 After calibrating the conductive probes with different constants, press "Enter" key to exit from calibrating state and then enters measuring state,; Soak the conductive probe into solution, the displayed value is the conductivity value of measured solution at this moment. When using the conductive probe with cell constant 1 to measure, the displayed value on the screen is the actual conductivity of measured solution. When using the conductive probe with constant 10 to measure, the displayed value multiply by 10 is the actual conductivity of measured solution. When the meter is used to measure the conductivity of pure water or ultra pure water in industrial flow, select the titanium alloy conductive probe with constant 0.01, together with the sealed corrosion-proof flow through(see diag.2). The displayed value should be multiplied by 0.01. The concluded value is the actual conductivity. Operating procedure: A) Clean the conductive probe with ultra water; B) Revolve the conductive probe into the measuring cell, connect water in and water out pipes according to diag.2; C) Make sure that there is no air coming into the flowing water, there is no bubble in thepipeeither; D) Flow speed should become faster gradually until conductivity value is stable and the value doesn't change under faster flow speed. When using titanium alloy electrode with constant 0.01, the displayed value should be multiplied by 0.01. The concluded value is the actual conductivity. 12

Measuring electrode Water out Measuring trough Water in Diag2. Measuring cell operating diagram 2.3 ATTENTION To avoid pollution when measuring ultra pure water. It is better to measure with sealing flowing method. Since fixed 2% temperature coefficient compensation is adopted in temperature compensation, make measurement without compensation(adjusting temperature to be 25 ) when measuring ultra pure water, after that, consult the table. In order to guarantee precise measurement, clean conductive probe with distilled water of 0.5μs/cm (or de-ion water) twice, then wash it with measured water three times before measurement. Conductive probe socket must be prevented from moisture to avoid any unnecessary error. Conductive probe needs constant calibration at regular intervals. Take out batteries if the meter won't be used for a long time; If "LOBAT" is displayed on the top left corner, or the displayed value is not stable, it's necessary to replace batteries. 2.4 How to use Back-light When measuring at night, turn on back-light key at above left of the meter. It is not recommended to use in the daytime as it will reduce the battery life. 2.5 Turn off the meter After using the meter, Press [On/Off] to turn off the meter.when not in use, all probes should be soaked in distilled water. 13

If the meter is out of use for a long time: Keep the socket of the meter clean and dry. Avoid contact of the socket with acid, alkaline and salt solution. Keep the input terminal of the meter (connector of probe) clean and dry. If the probe has been exposed to high humidity or water, wipe the cable interface of the probe with dry gauze or Kimwipe. When the conductivity probe will be out of use for a short time, soak it in DI water. If the probe will be out of use for 6 hours or longer, rinse it with DI water and store it in a protective bottle. 3 Maintenance The meter adopts DC batteries power supply and liquid crystal display, with good reliability. Inspect as follows if the meter doesn't work regularly. The meter can't work regularly when measuring A) Inspect if the conductive probe has inserted into the socket, the plug positioning groove should be aligned with the socket positioning pin; B) The cable line of Conductive probe is not correct or the conductive probe is damaged; C) Batteries' voltage is lower than 2.5V; D) There is some trouble inside the meter. When the meter is switched on, there is no display on screen. A) Socket of batteries is loose, contact is not good; B) There is some trouble inside the meter. C) No life of battery. 3.2 Cleaning and Storage for conductive probe This section describes how to maintain and store the probe. Probe cleaning General electrode cleaning includes inorganic cleaning, organic cleaning, grease cleaning, and protein precipitation cleaning. The type of cleaning needed depends on the contaminants and the probe. After one or more cleaning procedures, rinse the outside of the probe with distilled water. Probe storage Soak the measuring tip with distilled water or deionized water for temporary storage.when the electrode is not used for a long time, put the electrode back in the box and store it a dry place at ambient temperature. 14

4 Consumables and Ordering Information Model Name Description INE-DDB-303A Conductivity meter Standard Configuration INE-DDB-303A Electrode holder Standard Configuration DJS-1C Conductive Probe(Platinum-black) Standard Configuration (2-20000uS/cm) DJS-1C Conductive Probe(Bright) Standard Configuration (2-3000uS/cm) DJS-0.01C Conductive Probe(Titanium-alloy) Optional Configuration (0-2uS/cm) DJS-0.1C Conductive Probe(Platinum) Optional Configuration (0.05-200uS/cm) DJS-10C Conductive Probe Optional Configuration (2000-200000uS/cm) DJS-10 Conductive Probe Optional Configuration (2000-200000uS/cm) 5 APPENDIX Appendix 1, accord with the determination regulation of conductivity meter and international standard of conductivity meter, the symbol and unit of conductivity meter, conductance and electrode constant are as follows: Name Symbol Unit Remarks Conductivity meter K μs/cm 1S=10 3 ms=10 6 μs Conductance G μs Electrode constant J 1/cm 15

Packing List Model INE-DDB-303A Conductivity Meter No. Descriptions and Specifications Quantity 1 Model INE-DDB-303A Portable Conductivity Meter 1 set 2 DJS-1C Platinum-black Conductive Probe 1 3 Electrode H older (8.042.178) 1 4 AA 1.5V alkali batteries (No.5) 2 5 Operating Instruction 1 16