Modulated DSC Paper #8 Use Of Quasi-isothermal Mode for Improved Understanding of Structure Change

Similar documents
Characterization of Water Adsorption and Absorption in Pharmaceuticals

Moisture Sorption Analysis of Pharmaceuticals

Apparent Melting: A New Approach to Detecting Drug-Excipient Incompatibility

Moisture Sorption Analysis of Pharmaceuticals

Synergy of the combined application of thermal analysis and rheology in monitoring and characterizing changing processes in materials

APPLICATIONS OF THERMAL ANALYSIS IN POLYMER AND COMPOSITES CHARACTERIZATION. Wei Xie TA Instruments

DSC AS PROBLEM-SOLVING TOOL: BETTER INTERPRETATION OF Tg USING CYCLIC DSC

DSC Methods to Quantify Physical Aging and Mobility in Amorphous Systems: Assessing Molecular Mobility

TOPEM the new advanced multi-frequency TMDSC technique

DSC AND TG/DTA AS PROBLEM-SOLVING TOOLS: CHARACTERIZATION OF PHARMACEUTICAL COMPOUNDS

High Pressure DSC Differential Scanning Calorimeter

Universal Standard Protocols for Temperature and Material Characterization Calibration with Pharmaceuticals by Thermal Analysis

Melting and solidi cation of Pb nanoparticles embedded in an Al matrix as studied by temperature-modulated di erential scanning calorimetry

Thermal Methods of Analysis Theory, General Techniques and Applications. Prof. Tarek A. Fayed

Characterization of Solid State Drugs by Calorimetry

Common Definition of Thermal Analysis

Calorimetry: differential scanning calorimetry (DSC), isothermal titration calorimetry (ITC)

DEVELOPMENT OF IMPROVED METHODS FOR CHARACTERISING THE CURE OF COMPOSITE MATERIALS

2015 TA 儀器 DSC 免費訓練講座 2015 年 6 月 8 日 ( 一 ) 國立台灣科技大學國際大樓三樓 302 會議室 機台的原理與應用. Differential Scanning Calorimetry (DSC) 許炎山

Nucleation Behavior of Polypropylene with a Nan-O-Sil Additive Dr. Previn Shah, Rheoplast Associates

Chapter 31. Thermal Methods

Differential Scanning Calorimetry

New Castle, DE USA Lindon, UT USA Elstree, United Kingdom Shanghai, China Beijing, China Taipei, Taiwan Tokyo, Japan Seoul, South Korea Bangalore,

DATA ANALYSIS WITHOUT FOURIER TRANSFORMATION FOR SAWTOOTH-TYPE TEMPERATURE-MODULATED DSC

How to Use Kinetic Modeling Analysis To Predict Profile Part Failure

DSC PT 10. Applications

Differential Scanning Calorimeter Analysis of Hydrino-Producing Solid Fuel

Differential Scanning Calorimetry

Modelling of viscoelastic properties of a curing adhesive

Modulated-Temperature Thermomechanical Measurements

Q=mcDt. Chemistry 30 notes review of specific heat capacity. Thermodynamics the movement of thermal energy

Chemistry Heat Review. Heat: Temperature: Enthalpy: Calorimetry: Activation energy:

The Kinetics of B-a and P-a Type Copolybenzoxazine via the Ring Opening Process

CH3511 EXPERIMENT: Determination of Thermal Properties Using a Differential Scanning Calorimeter

Thermochemistry. The study of energy changes that occur during chemical reactions and changes in state.

THERMAL CHARACTERIZATION OF GYPSUM COMPOSITES BY USING DIFFERENTIAL SCANNING CALORIMETRY

A NEW MEASUREMENT AND EVALUATION METHOD FOR DSC OF PCM SAMPLES

INTERNATIONAL STANDARD

Comprehensive Handbook of Calorimetry and Thermal Analysis

TA Instruments NEW AR-G2 INTRODUCING THE WORLD S FIRST MAGNETIC BEARING RHEOMETER! The World Leader in Thermal Analysis and Rheology Systems

Heat Transfer in Polymers

APPLICATION NOTE. Characterization and Classification of Recycled Polyamides by Means of Identify. Dr. Ekkehard Füglein

D Y N A M I C M E C H A N I C A L A N A L Y S I S A N D I T S A D V A N T A G E S O V E R D E F L E C T I O N T E M P E R A T U R E U N D E R L O A D

Uncertainty of Thermal Characterization of Phase Change Material by Differential Scanning Calorimetry Analysis

Nutshells of Thermal Analysis. Heat it up! Burn it! Thermal Analysis

Unit 15 Energy and Thermochemistry Notes

Capturing the mechanical aging kinetics by thermal analysis

Thermal and UV-curing Behavior of Inks, Adhesives, and Coatings by Photo-, In-situ DEA and DMA.

Unit 15 Energy and Thermochemistry Notes

DAMPING PROPERTIES OF SOME EPOXY RESIN - LIQUID CRYSTALLINE DYESTUFF COMPOSITES

TECHNICAL UPDATE. Ricon Resins Peroxide Curing Data and Use as a Reactive Plasticizer in Polyphenylene Ether Based CCL and PWB

The Influence of Processing-History on the Main Melting Temperature and Secondary Melting Temperature Using DSC. Stephen Sansoterra

News & Trends for Thermal Analysis

CHAPTER 17 Thermochemistry

Single-ion BAB triblock copolymers as highly efficient electrolytes for lithium-metal batteries

Temperature-Modulated DSC

Thermal analysis unlocks the secrets of elastomers

EXPERIMENTALLY DETERMINING THE VISCOELASTIC BEHAVIOR OF A CURING THERMOSET EPOXY R. Thorpe 1, A. Poursartip 1*

Complete Materials Deformulation Using TGA-IR

Measuring the Reaction Kinetics of Polyester/PCA-501(HAA) Powder Coatings With Dielectric Analysis (DEA)

Section 16.3 Phase Changes

DYNAMIC MECHANICAL ANALYZER DMA 2980

ISO INTERNATIONAL STANDARD. Plastics Differential scanning calorimetry (DSC) Part 4: Determination of specific heat capacity

Ch 100: Fundamentals for Chemistry

K. Ishikiriyamaa and B. Wunderlich*

Polymer Chemistry Research Experience. Support: NSF Polymer Program NSF (PI: Chang Ryu)/RPI Polymer Center

Thermal Analysis. Short Courses POLYCHAR 25 Kuala Lumpur. Copyright 2017 by Jean-Marc Saiter

CHEM*3440. Thermal Methods. Thermogravimetry. Instrumental Components. Chemical Instrumentation. Thermal Analysis. Topic 14

Name Chemistry / / SOL Questions Chapter 9 For each of the following, fill in the correct answer on the BLUE side of the scantron.

What Happens When Rheological Properties Change? Looking into Rheological Properties with Simultaneous Collection of Microscopic Images

Cationic Cure of Epoxy Resin by an Optimum Concentration of N-benzylpyrazinium Hexafluoroantimonate

Material Testing Overview (THERMOPLASTICS)

Thermogravimetric Analysis Advanced Techniques for Better Materials Characterisation

THEORY AND APPLICATIONS OF MODULATED TEMPERATURE PROGRAMMING TO THERMOMECHANICAL TECHNIQUES

Thermochimica Acta 531 (2012) Contents lists available at SciVerse ScienceDirect. Thermochimica Acta

S T A T I O N 1 E X O T H E R M I C / E N D O T H E R M I C P R O C E S S E S

ARTICLES. Kinetics of Induced Crystallization of the LC 1-x Sil x System. Dipti Sharma* and Germano Iannacchione

DEA. Dielectric Cure Monitoring

QUANTITATIVE EVALUATION OF CURING SHRINKAGE IN POLYMERIC MATRIX COMPOSITES

A Thermal Analysis Technique That Combines Differential Scanning Calorimetry and Light Microscopy

Protective coatings for aluminum alloy based on. hyperbranched 1,4-polytriazoles

Rheological And Dielectric Characterization of Thermosetting Polymers. Jeffrey Gotro, Ph.D.

TAWN tests for quantitatively measuring the resolution and sensitivity of DSCs (version 2.1)

The Gibbs Phase Rule F = 2 + C - P

Liquids & Solids: Section 12.3

Honors Unit 9: Liquids and Solids

ab Uricase Assay Kit (Fluorometric) 1

Thermal and mechanical properties of several phthalonitrile resin system

Case study: molecular dynamics of solvent diffusion in polymers

The THT micro reaction calorimeter RC

A FINITE ELEMENT COUPLING MODEL FOR INTERNAL STRESS PREDICTION DURING THE CURING OF THICK EPOXY COMPOSITES

Microcalorimetric techniques

First-order transitions in glasses and melts induced by solid superclusters nucleated and melted by homogeneous nucleation instead of surface melting

Topic 05 Energetics : Heat Change. IB Chemistry T05D01

PCM specific heat capacity c p (T) measurements

Cure Kinetics of Ring-Opening Metathesis Polymerization of Dicyclopentadiene *

Technical Data Sheet: Thermokinetics 3

(Refer Slide Time: 00:58)

The Nature of Energy Energy is the ability to do work or produce Heat, q or Q, is ; flows due to temperature differences (always to )

Material Testing Report SN 7131 ISOFIL HK 30 TTH0 WT0822

Transcription:

Modulated DSC Paper #8 Use Of Quasi-isothermal Mode for Improved Understanding of Structure Change Leonard C. Thomas TA Instruments, 109 Lukens Drive, New Castle, DE 19720, USA ABSTRACT MDSC provides the ability to measure heat capacity and changes in heat capacity during isothermal operation and during kinetic processes such as crystallization and chemical reaction. INTRODUCTION DSC has been used for isothermal experiments since its conception. The analyst selects the constant temperature and time period over which the sample will be maintained. The sample can be ramped at a heating / cooling rate to that temperature or the method can be set to jump to that temperature as fast as the DSC can heat or cool. The result is a plot of heat flow versus time. Since it is impossible to measure heat capacity at a constant temperature, it is only possible to study the rate and degree of reaction for kinetic processes by isothermal DSC. The explanation for this is easily seen in the basic equation used to describe heat flow from DSC or MDSC. dh dt = Cp dt dt + f (T, t) Where: dh/dt = heat flow (mw or W/g) rate as measured by the DSC Cp = sample heat capacity (J/ºC) dt/dt = heating rate (ºC/min) f(t,t) = heat flow (mw or W/g) that is a function of temperature and time If a zero heating rate is used, as in a DSC isothermal experiment, the heat capacity term becomes zero, and the only measured heat flow comes from the kinetic term f(t,t). As discussed in an earlier paper (1), MDSC has a unique advantage since it uses two simultaneous heating rates. If the average heating rate is set to zero, the average temperature remains constant. The modulated heating rate can be set large for better sensitivity or small for improved resolution. Since the modulated heating rate results from selection of a modulated temperature amplitude and a modulation period, the user selects the size (amplitude) of the temperature oscillation to be superimposed on the TP 013 1

constant temperature and the time period (seconds) for each cycle. The term Quasiisothermal describes this type of experiment in which a small temperature oscillation (typically <1 C) is applied to a constant average temperature. Figure 1 shows both the temperature and heating rate from an MDSC quasi-isothermal experiment. Figure 1 Even though the average temperature remains constant in MDSC, it is possible to measure heat capacity because of the modulated heating rate. EXPERIMENTAL CONDITIONS There are two different types of quasi-isothermal experiments and it is easy to select MDSC experimental conditions for each. The most common type is one in which a single isothermal temperature is used such as illustrated in Figure 1. A method for such an experiment might look like the following: 1. Equilibrate 38ºC 2. Modulate +/- 0.5ºC every 60 seconds 3. Isothermal for 160 minutes Figure 2 illustrates a typical MDSC quasi-isothermal data on a sample of epoxy resin. The plot shows the Total heat flow, Reversing Heat Capacity and temperature versus time. The exothermic cure peak is seen in the Total signal and the amount of reaction during that time calculated in software to be 256 J/g. The Reversing Cp signal shows how the heat capacity decreases towards the end of the curing exotherm. This is a result of cross-linking of the molecules, which results in lower mobility and therefore, lower heat capacity. This decrease in mobility is what causes the reaction rate to slow as TP 013 2

the reaction moves from chemical control to diffusion control. This can be verified by heating the sample above 100 ºC once the isothermal experiment is completed. This is seen at the end of Figure 2 and in the temperature plot displayed in Figure 3. Once the sample is heated above the start of Tg near 100 ºC, the reaction begins again and there is an additional 31 J/g of reaction. 1.5 2.6 Heat Flow (mw) 1.0 0.5 0.0-0.5 50.73min 256.4J/g Cure Exotherm @ 100 C 75.30min Decrease in Cp Due to Crosslinking (Kinetics become Diffusion Controlled) Sample: Epoxy Size: 10.85 mg Method: MDSC Iso at 100 C Residual Cure Figure 2 Heating @ 3 C/min [ ] Temperature ( C) 350 300 250 200 150 2.4 2.2 2.0 1.8 1.6 [ ] Rev Cp (J/g/ C) -1.0 Iso @ 100 C for 160 min Temperature 31.06J/g 100 1.4-1.5 1.2 0 50 100 150 200 250 Exo Up Time (min) Universal V3.8A TA Instrume 0.2 Heating Experiment at 3 C/min after 160min Isothermal Cure @ 100 C 0.4 Note Onset of Decomosition before Complete Cure Heat Flow (mw) -0.2-0.6-1.0 Figure 3 Nonreversing Total Reversing 110.75 C 117.14 C 31.08J/g Note Inability to Measure Tg [ ] Nonrev Heat Flow (mw) 0.4 0.0-0.4 0.0-0.4-0.8 [ ] Rev Heat Flow (mw) 119.12 C(H) 0.2810J/g/ C All Signals at Same Sensitivity -1.4 Exo Up -1.2 52 102 152 202 252 Temperature ( C) Universal V3.8A TA Instruments TP 013 3

Note that it is not possible to see or measure the glass transition of the partially cured epoxy in the Total signal, which is equivalent to what would be obtained by standard DSC. The reason for this is that the endothermic glass transition is hidden by the more energetic exothermic reaction. Important Note: Many new MDSC users are often confused by the fact that the Reversing and Nonreversing Heat Flow signals have a value of zero during quasi-isothermal experiments. This is due to the fact that all signals that use a calculation that involves subtraction, multiplication or division by zero are automatically set to zero by the software. For example: Reversing Heat Flow = Reversing Cp X Avg. Heating Rate o Since the average heating rate is zero during a quasi-isothermal experiment, the Reversing Heat Flow is also zero Nonreversing Heat Flow = Total - Reversing o Since Reversing Heat Flow is zero, Nonreversing Heat Flow is also set to zero and the only Heat Flow signal that can be obtained during quasiisothermal operation is the Total signal Total Heat Capacity = Total Heat Flow Avg. Heating Rate o Since the average heating rate is zero during quasi-isothermal operation, the Total Heat Capacity signal is also zero The second type of quasi-isothermal experiment uses a series of isothermal experiments that involve a change in the average temperature at fixed time increments. It can be used to study both thermodynamic events such as the glass transition(s) or kinetic processes such as thermoset cure or crystallization. Examples of a typical method and applications are provided below. Example of Quasi-isothermal Method With Increasing Temperature Steps 1.Data Storage OFF 2.Equilibrate 50C 3.Modulate +/- 0.5ºC every 100 seconds 4.Isothermal for 5 minutes 5.Data Storage ON 6.Isothermal for 5 minutes 7.Data Storage OFF 8.Increment 1ºC 9.Repeat Segment 4 for 100 times TP 013 4

An explanation for some of the segments in the above method will help the reader become more familiar with the technique. Segments 1 and 5, Data Storage; the quality of the data can be significantly improved if data is collected only after equilibration at the set-point temperature. In this example, equilibration periods of 5 minutes are used before data storage is turned on and the data collected for 5 additional minutes at that temperature. Segment 3, Modulate Segment; although smaller amplitudes can be used, good sensitivity and resolution are obtained over the range of 0.5 C to 1.0 C. Since there is no concern with obtaining a sufficient number of cycles over a transition, a longer period (100 seconds) is recommended to insure optimum heat flow between the sample and sensor. Segment 8, Increment; although larger or smaller steps can be made, steps of 1 C are a good starting point. If the mature of the sample requires decreasing temperature steps, simply use negative values (-1 C). Segment 9, Repeat; this segment defines the number of steps. If a value of one (1 C) degree is used for the increment segment then 100 such steps will result in quasi-isothermal measurements over a 100 C range. Although not specified in the method described above, a data collection rate of 1 point every 10 seconds is sufficient for this type of measurement. This compares to the recommended data collection rate of 1 point every 2 seconds for normal MDSC heating / cooling experiments. Time and temperature plots from a quasi-isothermal experiment on a sample of quench-cooled PET are shown in Figures 4 and 5. A few features from these figures deserve comment: Except for the first few minutes of the experiment, the Total heat flow signal only changes by about 30 microwatts over almost 20 hours and most of that is due to the exothermic crystallization. This is an example of extremely good baseline stability. The Reversing Cp signal is very stable in the middle of the glass transition. This is due to the fact that the glass transition is caused by a distribution of molecular motion. The Reversing Cp signal decreases with time during crystallization. This is due to conversion of amorphous material to a crystalline structure, which has lower molecular mobility and therefore lower heat capacity. The Total signal, like standard DSC, cannot measure heat capacity changes isothermally or during kinetic processes. TP 013 5

Figure 4 Figure 5 TP 013 6

DSC has been used to characterize food products for many years but its utility has been limited by the complexity of observed transitions. MDSC provides numerous advantages for analysis of foods and quasi-isothermal MDSC is particularly useful in understanding structure and how that structure changes as a function of time and temperature. An example of this is seen in Figure 6, a comparison of the effect of cooling rate on the structure of a frozen sucrose solution. The quench-cooled sample shows a higher Reversing Cp and there is a single step in Cp associated with the glass transition of the frozen solution (melts above 30 C). The sample cooled at the low rate of 1 C/min has a lower heat capacity and goes through two steps. A detailed description for the difference in structure caused by the cooling rate is beyond the scope of this paper but there is additional information below that provides insight into the cause. The data for the quench-cooled sample shows that the heat capacity changes (range of values) during the 30-minute isothermal over the temperature range from -44 to 35 C. It was suspected that this resulted from crystallization of water that did not have time to freeze due to the high cooling rate of the quench-cooled sample. The suspicion was verified when the state diagram shown in Figure 7 was published by Roos and co-workers in a 1996 article in Food Technology. It clearly shows that ice forms between about -47 to 35 C. The conversion of water to ice would cause a decrease in Cp as seen in the MDSC results in Figure 6. Sample: 40% w/w sucrose in water Pan: Hermetic Alum. Weight: approx 15mg. Amp. +/- 0.5 C Period 100 sec Step, + 1 C Data Store 30 min Reversing Heat Capacity J/g C Figure 6 TP 013 7

Figure 7 SUMMARY As with many of the applications shown in previous papers in this series, MDSC overcomes the many natural limitations of DSC. In this case, MDSC provides the ability to measure heat capacity and changes in heat capacity during isothermal operation and during kinetic processes such as crystallization and chemical reaction. This ability is the direct result of using two simultaneous heating rates instead of the single heating rate in traditional DSC. KEY WORDS modulated differential scanning calorimetry, mdsc, quasi-isothermal, glass transition, foods COPYRIGHT 2005 TA INSTRUMENTS TA INSTRUMENTS United States, 109 Lukens Drive, New Castle, DE 19720 Phone: 1-302-427-4000 Fax: 1-302-427-4001 E-mail: info@tainstruments.com Spain Phone: 34-93-600-9300 Fax: 34-93-325-9896 E-mail: spain@tainstruments.com United Kingdom Phone: 44-1-293-658900 Fax: 44-1-293-658901 E-mail: uk@tainstruments.com Belgium/Luxembourg Phone: 32-2-706-0080 Fax: 32-2-706-0081 E-mail: belgium@tainstruments.com TP 013 8

Netherlands Phone: 31-76-508-7270 Fax: 31-76-508-7280 E-mail: netherlands@tainstruments.com Germany Phone: 49-6023-9647-0 Fax: 49-6023-96477-7 E-mail: germany@tainstruments.com France Phone: 33-1-304-89460 Fax: 33-1-304-89451 E-mail: france@tainstruments.com Italy Phone: 39-02-27421-283 Fax: 39-02-2501-827 E-mail: italia@tainstruments.com Sweden/Norway Phone: 46-8-594-69-200 Fax: 46-8-594-69-209 E-mail: sweden@tainstruments.com Japan Phone: 813 5479 8418 Fax: 813 5479 7488 E-mail: nurayama@taij.po-jp.com Australia Phone: 613 9553 0813 Fax: 61 3 9553 0813 E-mail: steve_shamis@waters.com To contact your local TA Instruments representative visit our website at www.tainst.com TP 013 9