Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Similar documents
Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Method to 0.50 mg/l NH 3 N Powder Pillows

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Nitrogen, Total Inorganic

Chloramine (Mono) and Nitrogen, Free Ammonia

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Method to 0.50 mg/l NH 3 N Powder Pillows

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

Chlorine, Free and Total, High Range

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

Oxygen Demand, Chemical

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Oxygen Demand, Chemical

Chlorine, Free and Total, High Range

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 )

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Chlorine, Free and Total, Low Range

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

DR/4000 PROCEDURE NITRATE. 2. The display will show:

Trihalomethane Formation Potential (THMFP)

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

DR/4000 PROCEDURE. IRON, Total

Quaternary Ammonium Compounds

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

NITROGEN, AMMONIA, High Range, Test N Tube

CHLORINE, TOTAL (0 to 4.00 mg/l)

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure

PAR Method Method to 2.0 mg/l Pb TNTplus 850

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Mercaptoacetic Acid Method*

DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample.

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

TPH (Total Petroleum Hydrocarbons)

DR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys.

DPD Test N Tube Method *

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

DR/4000 PROCEDURE. HARDNESS, Total

NITRITE, Low Range (0 to mg/l NO 2- -N)

NITROGEN, TOTAL, HR, Test N Tube

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced.

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

SILICA, Low Range (0 to 1.60 mg/l)

NITROGEN, TOTAL, Test N Tube * Method 10071

Polychlorinated Biphenyls (PCB) in Soil

RAPID KJELDAHL BENCHNOTES

p-dimethylaminobenzaldehyde Method*

Ammonia Gas Sensing Combination Electrode

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Manganese III Digestion Method * (with chloride removal)

SIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER COLORIMETRIC DETERMINATION OF CYANIDE

Ammonia Gas Sensing Combination Electrode

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

METHOD #: Approved for NPDES and SDWA (Ed. Rev. 1974, 1978) Fluoride, Total (Colorimetric, SPADNS with Bellack Distillation)

ALLOWAY METHOD OUTLINE

#18 Analysis of Water (Chlorine, Phosphorus, Fluoride, and Color)

Sample Preparation LCW 909. Total Kjeldahl Nitrogen, TKN NB: NEW! LCW 909 GB

PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION

Steam-Assisted Gravity Drainage Water Analysis Handbook

# 12 ph-titration of Strong Acids with Strong Bases

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

Steam-Assisted Gravity Drainage Water Analysis Handook

CH2MHILL GILROY/MORGAN HILL LABORATORY LABORATORY STANDARD OPERATING PROCEDURE NO TOTAL KJELDAHL NITROGEN - TKN SM 4500-N org.

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

Cat. No

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV)

MERCURY (0.1 to 2.5 µg/l) Method 10065

GETTING THE END POINT TO APPROXIMATE. Two hours

ASTM Designation: D Standard Test Method for Determination of Iodine Number of Activated Carbon

Experiment 5: Analysis of Nutrients in Natural Waters CH3600 / ESP 5090: Environmental Chemistry, Plymouth State University

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

Transcription:

Nitrogen, Ammonia DOC316.53.01078 USEPA 1 Nessler Method 2 Method 8038 0.02 to 2.50 mg/l NH 3 N Reagent Solution Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater. 1 USEPA accepted for wastewater analysis (distillation required), Method 350.2. 2 Adapted from Standard Methods for the Examination of Water and Wastewater, 4500-NH 3 B & C, 15th Edition. Test preparation Instrument-specific information Table 1 shows all of the instruments that have the program for this test. The table also shows sample cell and orientation requirements for specific instruments. To use the table, select an instrument, then read across to find the applicable information for this test. Table 1 Instrument-specific information Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR 3900 The fill line is to the right. 2495402 The fill line is toward the user. Before starting Hold the reagent droppers and dropper bottles vertically, not at an angle, when the reagent is added. The reagents that are used in this test contain mercury. Collect the reacted samples for safe disposal. If the Pour-Thru Cell is used, clean the cell periodically. To clean, add several crystals of sodium thiosulfate pentahydrate into the cell funnel. Add deionized water to dissolve the crystals. Rinse fully with deionized water. Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective equipment. Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory agencies for further disposal information. Items to collect Description Quantity Ammonia Nitrogen Reagent Set 1 Water, deionized 25 ml Pipet, serological, 1-mL 1 Mixing cylinder, graduated, 25 ml, glass stopper 2 Sample cells (For information about sample cells, adapters or light shields, refer to Instrumentspecific information on page 1.) 2 1

Refer to Consumables and replacement items on page 5 for order information. Sample collection and storage Test procedure Collect samples in clean glass or plastic bottles. If the sample contains chlorine, add one drop of 0.1 N sodium thiosulfate for each 0.3 mg/l chlorine in 1 liter of sample. To preserve samples for later analysis, adjust the sample ph to less than 2 with concentrated sulfuric acid (approximately 2 ml per liter). No acid addition is necessary if the sample is tested immediately. Keep the preserved samples at or below 6 C (43 F) for a maximum of 28 days. Let the sample temperature increase to room temperature before analysis. Before analysis, adjust the ph to ~7 with 5 N sodium hydroxide solution. Correct the test result for the dilution caused by the volume additions. Start 1. Start program 380 N, Ammonia, Ness. For information about sample cells, adapters or light shields, refer to Instrumentspecific information on page 1. 2. Prepare the sample: Fill a mixing cylinder to the 25- ml line with sample. 3. Prepare the blank: Fill a mixing cylinder to the 25 ml line with deionized water. 4. Add 3 drops of Mineral Stabilizer to each mixing cylinder. 5. Put the stopper on the mixing cylinders. Invert the mixing cylinders several times to mix. 6. Add 3 drops of Polyvinyl Alcohol Dispersing Agent to each mixing cylinder. 7. Put the stopper on the mixing cylinders. Invert the mixing cylinders several times to mix. 8. Use a pipet to add 1.0 ml of Nessler Reagent to each mixing cylinder. 2 Nitrogen-Ammonia, Nessler Method (2.50 mg/l)

9. Put the stopper on the mixing cylinders. Invert the mixing cylinders several times to mix. 10. Start the instrument timer. A 1-minute reaction time starts. 11. Pour 10 ml from the blank cylinder into a sample cell. 12. When the timer expires, clean the blank sample cell. Zero 13. Insert the blank into the cell holder. 14. Push ZERO. The display shows 0.00 mg/l NH 3 N. 15. Pour 10 ml from the sample cylinder into a second sample cell. 16. Clean the prepared sample cell. Read 17. Insert the prepared sample into the cell holder. 18. Push READ. Results show in mg/l NH 3 N. Interferences Interfering substance Chlorine Hardness Iron Interference level Table 2 Interfering substances Remove residual chlorine from a 250 ml sample by adding 1 drop of sodium thiosulfate for each mg/l chlorine (Cl 2 ). Sodium arsenite can be used instead of sodium thiosulfate. Refer to Sample collection and storage on page 2. A solution containing a mixture of 500 mg/l CaCO 3 and 500 mg/l Mg as CaCO 3 does not interfere. If the hardness concentration is more than these concentrations, add extra Mineral Stabilizer. Interferes at all levels by causing turbidity with Nessler Reagent. Nitrogen-Ammonia, Nessler Method (2.50 mg/l) 3

Table 2 Interfering substances (continued) Interfering substance Seawater Sulfide Glycine, various aliphatic and aromatic amines, organic chloramines, acetone, aldehydes and alcohols Interference level Add 1.0 ml (27 drops) of Mineral Stabilizer to the sample before analysis. This complexes the high magnesium concentrations found in sea water, but the sensitivity of the test is reduced by 30% due to the high chloride concentration. For best results, make a calibration with standards that contain the same chloride concentration as seawater, or distill the sample. Interferes at all levels by causing turbidity with Nessler Reagent. May cause greenish or other off colors or turbidity. Distill the sample if these compounds are present. Pollution prevention and waste management The Nessler reagent contains mercuric iodide. The reacted samples and blanks will contain mercury and must be disposed of as a hazardous waste. Dispose of reacted solutions according to local, state and federal regulations. Distillation Accuracy check To eliminate most interferences, distill the sample, then use the distilled sample in the test procedure. 1. Set up the distillation apparatus for general purpose distillation. Refer to the Distillation Apparatus manual for proper assembly. 2. Measure 250 ml of sample into a 250-mL graduated cylinder. 3. Pour the sample into a 400-mL beaker. If the sample contains chlorine, add 1 drop of 0.1 N sodium thiosulfate solution for each 1 mg/l Cl 2 to remove the chlorine. 4. Add 25 ml of borate buffer solution and mix. Adjust the ph to approximately 9.5 with 1 N sodium hydroxide solution. Use a ph meter to monitor the ph. 5. Pour the solution into the distillation flask. 6. Add a magnetic stir bar and 5 glass beads. 7. Use a graduated cylinder to measure 25 ml of deionized water into a 250-mL Erlenmeyer flask. Add the contents of one Boric Acid Powder Pillow. Mix thoroughly. 8. Set the flask under the distillation apparatus drip tube. Elevate the flask so that the end of the tube is immersed in the solution. 9. Set the stirrer power to on. Set the stir control to 5. 10. With the thermometer inserted, set the heat control to 10. The yellow pilot lamp is an indication that the heater is on. 11. Turn on the water and adjust to maintain a steady flow through the condenser. 12. When 150 ml of distillate has been collected, turn the heater off. Immediately remove the collection flask. Measure the distillate to make sure 150 ml was collected (total volume = 175 ml). 13. Adjust the ph to approximately 7 with 1 N sodium hydroxide solution. Use a ph meter to monitor the ph. 14. Quantitatively transfer the distillate into a 250-mL volumetric flask. Dilute to the mark with deionized water. Mix well. Use the diluted distillate in the test procedure. Standard additions method (sample spike) Use the standard additions method (for applicable instruments) to validate the test procedure, reagents and instrument and to find if there is an interference in the sample. Items to collect: 50-mg/L Nitrogen-Ammonia Standard Solution 4 Nitrogen-Ammonia, Nessler Method (2.50 mg/l)

Mixing cylinders, 25-mL (3x) TenSette Pipet and pipet tips 1. Use the test procedure to measure the concentration of the sample, then keep the (unspiked) sample in the instrument. 2. Go to the Standard Additions option in the instrument menu. 3. Select the values for standard concentration, sample volume and spike volumes. 4. Open the standard solution. 5. Prepare three spiked samples: use the TenSette pipet to add 0.1 ml, 0.2 ml and 0.3 ml of the standard solution, respectively, to three 25-mL portions of fresh sample. Mix well. 6. Use the test procedure to measure the concentration of each of the spiked samples. Start with the smallest sample spike. Measure each of the spiked samples in the instrument. 7. Select Graph to compare the expected results to the actual results. Note: If the actual results are significantly different from the expected results, make sure that the sample volumes and sample spikes are measured accurately. The sample volumes and sample spikes that are used should agree with the selections in the standard additions menu. If the results are not within acceptable limits, the sample may contain an interference. Standard solution method Use the standard solution method to validate the test procedure, the reagents and the instrument. Items to collect: 1-mg/L Nitrogen-Ammonia Standard Solution 1. Use the test procedure to measure the concentration of the standard solution. 2. Compare the expected result to the actual result. Note: The factory calibration can be adjusted slightly with the standard adjust option so that the instrument shows the expected value of the standard solution. The adjusted calibration is then used for all test results. This adjustment can increase the test accuracy when there are small variations in the reagents or instruments. Method performance The method performance data that follows was derived from laboratory tests that were measured on a spectrophotometer during ideal test conditions. Users can get different results under different test conditions. Program Standard Precision (95% confidence interval) Sensitivity Concentration change per 0.010 Abs change 380 1.00 mg/l NH 3 N 0.99 1.01 mg/l NH 3 N 0.02 mg/l NH 3 N Summary of Method The Mineral Stabilizer complexes hardness in the sample. The Polyvinyl Alcohol Dispersing Agent helps the color formation in the reaction of Nessler Reagent with ammonia and certain other amines. A yellow color is formed that is proportional to the ammonia concentration. The measurement wavelength is 425 nm. Consumables and replacement items Required reagents Description Quantity/test Unit Item no. Ammonia Nitrogen Reagent Set, includes: 2458200 Nessler Reagent 2 ml 500 ml 2119449 Mineral Stabilizer 6 drops 50 ml SCDB 2376626 Nitrogen-Ammonia, Nessler Method (2.50 mg/l) 5

Consumables and replacement items (continued) Description Quantity/test Unit Item no. Polyvinyl Alcohol Dispersing Agent 6 drops 50 ml SCDB 2376526 Water, deionized varies 4 L 27256 Required apparatus Description Quantity/test Unit Item no. Mixing cylinder, graduated, 25 ml with stopper 1 each 2088640 Pipet, serological, 1 ml, glass 1 50/pkg 2093135 Pipet filler, safety bulb 1 each 1465100 Recommended standards Description Unit Item no. Nitrogen Ammonia Standard Solution, 1.0-mg/L NH 3 N 500 ml 189149 Nitrogen Ammonia Standard Solution, 10-mL Voluette Ampule, 50-mg/L NH 3 N 16/pkg 1479110 Wastewater Effluent Standard Solution, Mixed Parameter, for NH 3 -N, NO 3 -N, PO 4 3, COD, SO 4 2, TOC 500 ml 2833249 Optional reagents and apparatus Description Unit Item no. Ampule Breaker, 10-mL Voluette Ampules each 2196800 Distillation apparatus set, general purpose each 2265300 Distillation heater and support for apparatus set, 115 VAC option each 2274400 Distillation heater and support for apparatus set, 230 VAC option each 2274402 Pipet, TenSette, 0.1 1.0 ml each 1970001 Pipet tips for TenSette Pipet, 0.1 1.0 ml 50/pkg 2185696 Sodium Hydroxide Standard Solution, 5.0 N 100 ml MDB 245032 Sodium Thiosulfate, 0.1 N 100 ml 32332 Sulfuric Acid, ACS 500 ml 97949 FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company/Hach Lange GmbH, 2007, 2010, 2012, 2014, 2017. All rights reserved. 04/2017, Edition 9