Vector blood velocity estimation in medical ultrasound

Similar documents
Recent advances in blood flow vector velocity imaging

Spectral Velocity Estimation in the Transverse Direction

High Frame Rate Vector Velocity Estimation using Plane Waves and Transverse Oscillation

Multi-dimensional spectrum analysis for 2-D vector velocity estimation

A new estimator for vector velocity estimation [medical ultrasonics]

Non-invasive Measurement of Pressure Gradients in Pulsatile Flow using Ultrasound

Estimation of velocity vector angles using the directional cross-correlation method

Two-Dimensional Blood Flow Velocity Estimation Using Ultrasound Speckle Pattern Dependence on Scan Direction and A-Line Acquisition Velocity

Estimation of blood velocity vectors using transverse ultrasound beam focusing and cross-correlation

ieee transactions on ultrasonics, ferroelectrics, and frequency control, vol. 53, no. 5, may Investigation of Transverse Oscillation Method

31545 Medical Imaging systems

Plane Wave Medical Ultrasound Imaging Using Adaptive Beamforming

1194 IEEE TRANSACTIONS ON ULTRASONICS, FERROELECTRICS, AND FREQUENCY CONTROL, VOL. 64, NO. 8, AUGUST 2017

Lateral Blood Flow Velocity Estimation Based on Ultrasound Speckle Size Change With Scan Velocity

Investigation of Transverse Oscillation Method

Pirnia, Elham; Jakobsson, Andreas; Gudmundson, Erik; Jensen, Jörgen

Feasibility of non-linear simulation for Field II using an angular spectrum approach

Paper published in IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control: A New Method for Estimation of Velocity Vectors

746 ieee transactions on ultrasonics, ferroelectrics, and frequency control, vol. 54, no. 4, april 2007

A Transverse Oscillation Approach for Estimation of Three-Dimensional Velocity Vectors, Part I: Concept and Simulation Study

Ultrasound Vector Flow Imaging: Part I: Sequential Systems

Ultrasound Vector Flow Imaging: Part II: Parallel Systems

THIS two-part paper gives a review of vector flow

31545 Medical Imaging systems

Computational fluid dynamics using in vivo ultrasound blood flow measurements

Output intensity measurement on a diagnostic ultrasound machine using a calibrated thermoacoustic sensor

Pixel-based Beamforming for Ultrasound Imaging

SCATTERING OF ULTRASONIC WAVE ON A MODEL OF THE ARTERY J. WÓJCIK, T. POWAŁOWSKI, R. TYMKIEWICZ A. LAMERS, Z. TRAWIŃSKI

(12) Patent Application Publication (10) Pub. No.: US 2015/ A1

The Physics of Doppler Ultrasound. HET408 Medical Imaging

31545 Medical Imaging systems

AN ITERATIVE ADAPTIVE APPROACH FOR BLOOD VELOCITY ESTIMATION USING ULTRASOUND

Safety Assessment of Advanced Imaging Sequences II: Simulations

Today s menu. Last lecture. Ultrasonic measurement systems. What is Ultrasound (cont d...)? What is ultrasound?

Outline of today Medical Imaging systems. Wave types. 1. Discussion assignment on B-mode imaging

Curve Sketching. The process of curve sketching can be performed in the following steps:

Non-invasive measurement of pressure gradients using ultrasound

Ultrasound (US) Doppler techniques are widely

New method for two-point nonuniformity correction of microbolometer detectors

B-MODE IMAGES. Jrgen Arendt Jensen, Jan Mathorne, Torben Gravesen and. Bjarne Stage. Electronics Institute, bldg Technical University of Denmark

Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar.

Simulation of Contrast Agent Enhanced Ultrasound Imaging based on Field II

Calculation of Pressure Fields from Arbitrarily Shaped, Apodized, and Excited Ultrasound Transducers

Fast simulation of nonlinear radio frequency ultrasound images in inhomogeneous nonlinear media: CREANUIS

In vivo investigation of filter order influence in eigen-based clutter filtering for color flow imaging

Today s menu. Last lecture. Measurement of volume flow rate. Measurement of volume flow rate (cont d...) Differential pressure flow meters

CHAPTER 8 ANALYSIS OF AVERAGE SQUARED DIFFERENCE SURFACES

IEEE Ultrasonic symposium 2002

An Alternative Poincaré Section for Steady-State Responses and Bifurcations of a Duffing-Van der Pol Oscillator

Correlation analysis of three-dimensional strain imaging using ultrasound two-dimensional array transducers

Ultrasonic Measurement of Minute Displacement of Object Cyclically Actuated by Acoustic Radiation Force

MEDT8007 Simulation methods in ultrasound imaging

HYDROGEN SPECTRUM = 2

ECMUS The Safety Committee of EFSUMB : Tutorial

Structure of Biological Materials

between electron energy levels. Using the spectrum of electron energy (1) and the law of energy-momentum conservation for photon absorption:

DOPPLER-BASED ULTRASONIC BLOOD VELOCITY ESTIMATION

Hussain, B; Wong, EY; Yiu, BYS; Yu, ACH; Poepping, TL. IEEE Ultrasonics Symposium. Proceedings. Copyright IEEE.

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam

A Systematic Approach to Frequency Compensation of the Voltage Loop in Boost PFC Pre- regulators.

Numerical calculation of the electron mobility in ZnS and ZnSe semiconductors using the iterative method

Elec Eng 3BA3: Structure of Biological Materials

Products and Convolutions of Gaussian Probability Density Functions

Electrical Engineering 3BA3: Structure of Biological Materials

X-ray Diffraction. Interaction of Waves Reciprocal Lattice and Diffraction X-ray Scattering by Atoms The Integrated Intensity

Change in Ultrasonic Backscattered Energy for Temperature Imaging: Factors Affecting Temperature Accuracy and Spatial Resolution in 3D

UNCERTAINTY EVALUATION OF SINUSOIDAL FORCE MEASUREMENT

( x) f = where P and Q are polynomials.

Analysis of Friction-Induced Vibration Leading to Eek Noise in a Dry Friction Clutch. Abstract

Technical University of Denmark

2/11/2006 Doppler ( F.Robilliard) 1

Sakura Pascarelli European Synchrotron Radiation Facility, Grenoble, France

Figure 18: Top row: example of a purely continuous spectrum (left) and one realization

Sound Intensity. Sound Level. Doppler Effect. Ultrasound. Applications

T man body has been greatly eased by the introduction of

Ultra Fast Calculation of Temperature Profiles of VLSI ICs in Thermal Packages Considering Parameter Variations

2 Frequency-Domain Analysis

RESOLUTION MSC.362(92) (Adopted on 14 June 2013) REVISED RECOMMENDATION ON A STANDARD METHOD FOR EVALUATING CROSS-FLOODING ARRANGEMENTS

774. Tribological adhesion of particles in acoustic field

Echocardiography is a widely used modality for. Effect of Subaperture Beamforming on Phase Coherence Imaging. Hideyuki Hasegawa and Hiroshi Kanai

References Ideal Nyquist Channel and Raised Cosine Spectrum Chapter 4.5, 4.11, S. Haykin, Communication Systems, Wiley.

Technical University of Denmark

An introduction to X-ray Absorption Fine Structure Spectroscopy

Acoustic forcing of flexural waves and acoustic fields for a thin plate in a fluid

Physics 5153 Classical Mechanics. Solution by Quadrature-1

University of Lübeck, Medical Laser Center Lübeck GmbH Optical Coherence Tomography

CROSS-TALK RESPONSE ANALYSIS ON A PIEZOELECTRIC ULTRASONIC MATRIX ARRAY

sensors ISSN

Feasibility of a Multi-Pass Thomson Scattering System with Confocal Spherical Mirrors

Chapter 4 Imaging. Lecture 21. d (110) Chem 793, Fall 2011, L. Ma

Mathematical Notation Math Calculus & Analytic Geometry III

APPLICATION OF A SCANNING VIBROMETER FOR THE PERIODIC CALIBRATION OF FORCE TRANSDUCERS

Non-newtonian Rabinowitsch Fluid Effects on the Lubrication Performances of Sine Film Thrust Bearings

AH 2700A. Attenuator Pair Ratio for C vs Frequency. Option-E 50 Hz-20 khz Ultra-precision Capacitance/Loss Bridge

Modified Phased Tracking Method for Measurement of Change in Thickness of Arterial Wall

The achievable limits of operational modal analysis. * Siu-Kui Au 1)

BNG/ECE 487 FINAL (W16)

9.1 The Square Root Function

Finger Search in the Implicit Model

Phase-Sensitive Lateral Motion Estimator for Measurement of Artery-Wall Displacement Phantom Study

Transcription:

Vector blood velocity estimation in medical ultrasound Jørgen Arendt Jensen, Fredrik Gran Ørsted DTU, Building 348, Technical University o Denmark, DK-2800 Kgs. Lyngby, Denmark Jesper Udesen, Michael Bachmann Nielsen, Kristina Rue Nielsen Department o Radiology, Section o Ultrasound, Rigshospitalet, Blegdamsvej 9, DK-2100 Copenhagen Ø, Denmark Two methods or making vector velocity estimation in medical ultrasound are presented. All o the techniques can ind both the axial and transverse velocity in the image and can be used or displaying both the correct velocity magnitude and direction. The irst method uses a transverse oscillation in the ultrasound ield to ind the transverse velocity. In-vivo examples rom the carotid artery are shown, where complex turbulent low is ound in certain parts o the cardiac cycle. The second approach uses directional beam orming along the low direction to estimate the velocity magnitude. Using a correlation search can also yield the direction, and the ull velocity vector is thereby ound. An examples rom a low rig is shown. Keywords: Medical ultrasound, vector velocity estimation, directional beam orming. 1 INTRODUCTION TO VECTOR VELOCITY ESTIMATION Imaging o the blood velocity has been made or the last 20 years. The original paper by Kasai et al. [1] demonstrated that the velocity could be estimated in the direction o the ultrasound beam. This oten gives problems, since the vessel run along the skin surace and the ultrasound beam is, thus, perpendicular to the low direction. This can be corrected by tilting the beam, but oten the geometry o the vessels is complicated and a single correction is not suicient. There is thereore a real need in medical ultrasound or vector low techniques. This paper will present two techniques or vector low imaging developed in our laboratory. Many authors have developed techniques or vector velocity estimation. Fox [1,2] developed the crossing beams technique, where the velocity is ound along two beam directions. The approach is, however, diicult to use or imaging, as it is diicult to get crossing beams simultaneously over the ull image. Another approach is speckle tracking [3], where two images acquired at dierent times are correlated to ind the vector motion. A number o other techniques have also been introduced, but none have so ar made it into commercial products. 2 TRADITIONAL VELOCITY ESTIMATION Traditional velocity estimation is perormed by emitting a pulsed ield a number o times in the same direction as shown in Fig. 1. Here the let graph shows the received signal rom a single scatterer moving away rom the transducer. The consecutive signals are shown with an increasing y- axis oset. The signal is sampled at the depth o interest resulting in the signal on the right. The requency o this signal is proportional to the velocity as [4] 2 v cosθ 2vz = 0 (1) c c p 0 = where Θ is the low angle between the ultrasound beam and the velocity vector, v is the velocity magnitude, v z is the axial blood velocity, c is the speed o sound, and 0 is the emitted ultrasound center requency. A larger velocity will, thus, give a larger shit between signals and a higher requency. The generating mechanism is the oscillation in the ultrasound signal and the shit in position between pulse emissions. Figure 1: Received RF signal or a single scatterer moving away rom the transducer. The let graph shows the individual traces received ater each pulse emission and the right graph shows the sampled signal obtained at the dashed line in the let graph at 3.2 µs (rom [4]). Figure 2 shows the major problem with this approach. The velocity has here been determined at a number o places in the image rom the requency given by (1). A blue color shows velocity towards the transducer and a red color away rom the transducer. The jugular vein and carotid artery in the neck has been scanned. A change in color is seen 3

around the green line. This is due to the cos Θ actor in (1) as the low angle changes along the green line. This is easily detected in this image, but very diicult to compensate or in more complicated low patterns. A major problem is also that the projected velocity is shown. This will always make the velocity lower than the true velocity, and this makes quantitative velocity imaging diicult. oscillation period in the signal. Such a transverse oscillation can be introduced by splitting the transducer aperture in two and emitting waves rom each part as shown in Fig. 5. The two ields will then interere and generate a transverse oscillation. Figure 2: Velocity image o the jugular vein (top) and carotid artery (bottom). The blue color indicates velocity towards the transducer and red away rom it. The change in color along the blue center line is due to the change in angle between the ultrasound beam and the velocity direction (rom [16]). 3 TRANSVERSE OSCILLATION APPROACH The important eature or estimating the velocity in traditional systems is the sinusoidal pulse in the axial directions. This generates the received requency and makes it possible to determine the velocity. The transverse velocity cannot be estimated, since there is no corresponding transverse oscillation, as shown in Fig. 3. Here the point spread unction (ps) is depicted with a black value corresponding to a negative pressure and a white value is positive pressure. A scatterer going through the ps in the axial direction will give an oscillation as described in the previous Section with a requency given by (1). A corresponding lateral movement will give the signal shown in the right graph in Fig. 3. Here no oscillation is seen and the mean requency o this signal will be zero. The basic idea in the transverse oscillation approach [5, 13] is to introduce a lateral oscillation in the ield as shown in Fig. 4. A lateral motion will then give the signal shown in the right graph in Fig. 4, with an oscillation requency proportional to the lateral velocity as v x l = (2) λx where v x is the lateral oscillation and λx is the lateral Figure 3: Point spread unction (ps) or ordinary ultrasound velocity estimation. The top let graph shows the ps and the lower graph shows the response rom a scatterer transversing the ield in the axial direction. The right graph shows the signal or the scatterer transversing the ield in the lateral direction (rom [5]). Figure 4: Point spread unction (ps) or transvers oscillation method. The top let graph shows the ps and the lower graph shows the response rom a scatterer transversing the ield in the axial direction. The right graph shows the signal or the scatterer transversing the ield in the lateral direction (rom [5]). The received signal can then be processed to yield both the axial and lateral velocities independently as described in [6]. An example o such an image is shown in Fig. 6, which has been measured on the carotid artery o a human volunteer. A 128 element, 7 MHz linear array 4

probe was used connected to the RASMUS experimental ultrasound scanner [14]. The data was stored in the scanner and then processed in a Linux cluster to estimate both the axial and lateral velocities. The image shows the velocity magnitude as a red color with the magnitude corresponding to color intensity. The arrows on the image indicate both direction and magnitude. The low angle is nearly 90 degrees and an estimate o the velocity is obtained. For this simple case the ultrasound beam could just be titled to give an estimate with the conventional method, but this is not possible or more complex patterns. Figure 5: Transducer aperture used or transmission and the corresponding ield as simulated by the Field II program [12]. Figure 7: Vector velocity image o the carotid artery biurcation at peak systole (top) and 0.1 s ater peak systole. (rom [7]). This is demonstrated in Fig. 7, where the low patterns in the carotid biurcation are shown around the maximum contraction o the heart. The low at peak systole is airly laminar with a lower velocity at the lowest part o the vessel, whereas there is a very disturbed, rotating low right ater peak systole. The velocity arrows here rotate ull circle within 100-200 ms demonstrating the importance o showing the ull velocity vector to capture and display the very complex low patterns in the human body. Figure 6: Vector velocity image o the carotid artery. The color indicates velocity magnitude and arrows indicate both velocity direction and magnitude (rom [7]). 4 DIRECTIONAL BEAMFORMING APPROACH There are also other methods or inding the velocity vector. The directional beam orming approach uses ocusing along the low direction to ind the velocity 5

vector [8,9]. Here a normal ocused ultrasound ield is emitted and the scattered signal is sampled rom all the receiving elements. The direction o the low is determined rom the B-mode image and signals are then ocused along this direction as shown in Fig. 8. The process is repeated and a number o signals are ound and the signals are crosscorrelated to yield the shit in position between the signals. The spatial shit will be equal to the velocity magnitude times the time between the signals. Finding the shit and dividing by the time will then yield the velocity. This is the principle in the directional beam orming method. determining the correct velocity and that vector velocity techniques are needed in order to display the ull complexity o low in the human body. An example o such an image is shown in Fig. 9. It was made using a circulating low rig with a mass low meter as reerence. A 128 element, 7 MHz linear array probe was used connected to the RASMUS experimental ultrasound scanner and 20 pulse echo lines were acquired or each o the 65 imaging lines. The velocity was then ound by ocusing lines perpendicular to the ultrasound beam, cross-correlating them, and then inding the velocity. Figure 9: Velocity image at 90 degrees in a low rig (rom [9]). ACKNOWLEDGEMENT This work was supported by grant 9700883, 9700563 and 26-04-0024 rom the Danish Science Foundation, the Academy o Technical Sciences, the Ministry o Science, Technology and Innovation, and by B-K Medical A/S. Figure 8: Directional ocusing along the low direction (rom [10]). The standard deviation across the parabolic proiles is 4.3% relative to the peak velocity in the vessel, which is comparable to the accuracy obtained in traditional scanners, that here would show no velocity at 90 degrees. The angle or the low can also be determined or this approach. This is done by beam orming in a number o directions and then ind the velocity and normalized peak cross-correlation or the dierent angles. The highest correlation coeicient then denotes the correct angle, since the signals will decorrelate or others angles as the velocity is dierent in dierent parts o the directional signal. Such an approach has been investigated in [15]. 5 SUMMARY Two dierent methods or vector velocity estimation have been presented. It has been shown that conventional techniques can lead to cross errors in REFERENCES [1] M. D. Fox. Multiple crossed-beam ultrasound Doppler velocimetry. IEEE Trans. Son. Ultrason., SU-25:281-286, 1978. [2] B. Dunmire, K. W. Beach, K-H Labs., M. Plett, and D. E. Strandness. Cross-beam vector Doppler ultrasound or angle independent velocity measurements. Ultrasound Med. Biol., 26:1213-1235, 2000 [3] G. E. Trahey, J. W. Allison, and O. T. von Ramm. Angle independent ultrasonic detection o blood low. IEEE Trans. Biomed. Eng., BME-34:965-967, 1987. [4] J. A. Jensen. Estimation o Blood Velocities Using Ultrasound: A Signal Processing Approach. Cambridge University Press, New York, 1996 [5] J. A. Jensen and P. Munk. A new method or estimation o velocity vectors. IEEE Trans. Ultrason., Ferroelec., Freq. Contr., 45:837-851, 1998. [6] J. A. Jensen. A new estimator or vector velocity estimation. IEEE Trans. Ultrason., Ferroelec., Freq. Contr., 48(4):886-894, 2001. [7] J. Udesen, M. B. Nielsen, K. R. Nielsen, J. A. Jensen, Blood vector velocity estimation using an autocorrelation approach: In vivo Investigation., IEEE Ultrasonics Symposium 2005, pp 162-165, 2005 [8] J. A. Jensen. Directional velocity estimation using ocusing along the low direction: I: Theory and simulation. IEEE Trans. Ultrason., Ferroelec., Freq. Contr., pages 857-872, 2003. 6

[9] J. A. Jensen and R. Bjerngaard. Directional velocity estimation using ocusing along the low direction: II: Experimental investigation. IEEE Trans. Ultrason., Ferroelec., Freq. Contr., pages 873-880, 2003. [10] J. A. Jensen and S. I. Nikolov. Directional synthetic aperture low imaging. IEEE Trans. Ultrason., Ferroelec., Freq. Contr., pages 1107-1118, 2004. [11] J. A. Jensen and N. B. Svendsen. Calculation o pressure ields rom arbitrarily shaped, apodized, and excited ultrasound transducers. IEEE Trans. Ultrason., Ferroelec., Freq. Contr., 39:262-267, 1992. [12] J. A. Jensen. Field: A program or simulating ultrasound systems. Med. Biol. Eng. Comp., 10th Nordic- Baltic Conerence on Biomedical Imaging, Vol. 4, Supplement 1, Part 1:351-353, 1996b. [13] M. E. Anderson. Multi-dimensional velocity estimation with ultrasound using spatial quadrature. IEEE Trans. Ultrason., Ferroelec., Freq. Contr., 45:852-861, 1998. [14] J. A. Jensen, O. Holm, L. J. Jensen, H. Bendsen, S. I. Nikolov, B. G. Tomov, P. Munk, M. Hansen, K. Salomonsen, J. Hansen, K. Gormsen, H. M. Pedersen, and K. L. Gammelmark. Ultrasound research scanner or real-time synthetic aperture image acquisition. IEEE Trans. Ultrason., Ferroelec., Freq. Contr., 52 (5), May 2005. [15] J. Kortbek and J. A. Jensen. Determination o velocity vector angles using the directional cross-correlation method. In Proc. IEEE Ultrason. Symp., pages 1339-1343, 2005. [16] P. Munk, J. A. Jensen, Perormance o a vector velocity estimator, IEEE Ultrasonics Symposium Proceedings, pp 1489-1493, IEEE, 1998 7