The active background method in XPS data peak-fitting

Similar documents
Peak shape analysis of core level photoelectron spectra using UNIFIT for WINDOWS

Quantification of JEOL XPS Spectra from SpecSurf

X-Ray Photoelectron Spectroscopy (XPS)-2

Advanced Lab Course. X-Ray Photoelectron Spectroscopy 1 INTRODUCTION 1 2 BASICS 1 3 EXPERIMENT Qualitative analysis Chemical Shifts 7

Stability of Organic Cations in Solution-Processed CH3NH3PbI3 Perovskites: Formation of Modified Surface Layers

X-Ray Photoelectron Spectroscopy (XPS)-2

Error estimation in peak-shape analysis of XPS core-level spectra using UNIFIT 2003: how significant are the results of peak fits?

A Beginners Guide to XPS

X- ray Photoelectron Spectroscopy and its application in phase- switching device study

This article was published in

Name: (a) What core levels are responsible for the three photoelectron peaks in Fig. 1?

Figure 1: Symmetrical lineshapes generated form the LA definition in CasaXPS.

CasaXPS Manual Rev 1.2. Copyright 2009 Casa Software Ltd

Photoelectron Spectroscopy. Xiaozhe Zhang 10/03/2014

Lecture 7 Chemical/Electronic Structure of Glass

Practical Surface Analysis

An introduction to X- ray photoelectron spectroscopy

for XPS surface analysis

Methods of surface analysis

Supporting Information: Analysis of protein coatings on gold nanoparticles by XPS and liquid-based particle sizing techniques

AES Features in CasaXPS

Photoelectron Peak Intensities in Solids

An axis-specific rotational rainbow in the direct scatter of formaldehyde from Au(111) and its influence on trapping probability

0.8 b

FRAM V5.2. Plutonium and Uranium Isotopic Analysis Software

(a) (b) Supplementary Figure 1. (a) (b) (a) Supplementary Figure 2. (a) (b) (c) (d) (e)

Standard Practice for Calibration of the Electron Binding-Energy Scale of an X-Ray Photoelectron Spectrometer 1

XPS: Issues with Data Acquisition and Data Processing

Why Did We Spend 10 Years to Make the Digital SpecMaster XPS Spectral Data-Base System?

Estimation of IMFP except for a constant factor using only XPS background-optimized peak intensities and cross sections

Experimental Supporting Information (ESI) XPS measurements

Basic and Advanced Processing. Tim Carney

Supplementary Figure 1. Visible (λ = 633 nm) Raman spectra of a-co x layers. (a) Raman spectra of

Electron spectroscopy Lecture Kai M. Siegbahn ( ) Nobel Price 1981 High resolution Electron Spectroscopy

Asymmetry of Peaks in the XPS of Polymers

NIST Database for the Simulation of Electron Spectra For Surface Analysis (SESSA)*

Supplementary Information

Inelastic background analysis of HAXPES spectra: towards enhanced bulk sensitivity in photoemission Tougaard, Sven Mosbæk

X-Ray Photoelectron Spectroscopy (XPS)

Supplementary Figure 1 PtLuSb RHEED and sample structure before and after capping layer

X-ray Photoelectron Spectroscopy/ Electron spectroscopy for chemical analysis (ESCA), By Francis Chindeka

ELECTRON SPECTROSCOPY STUDIES OF VACUUM DEPOSITED CHROMIUM AND COBALT LAYERS. Jani Sainio

Low Energy Electrons and Surface Chemistry

Journal of Electron Spectroscopy and Related Phenomena

Introduction to X-ray Photoelectron Spectroscopy (XPS) XPS which makes use of the photoelectric effect, was developed in the mid-1960

Supporting Online Material for

Influence of Surface Roughness on Results of XPS Measurements

5-Sep-15 PHYS101-2 GRAPHING

The Benefit of Wide Energy Range Spectrum Acquisition During Sputter Depth Profile Measurements

A COMPARATIVE CHEMICAL STUDY OF THE TOP LAYERS OF TWO DIFFERENT QUALITY SILICON SOLAR CELLS

Universality (and its limitations) in Cosmic Ray shower development

Spin-resolved photoelectron spectroscopy

An Introduction to Auger Electron Spectroscopy

Lecture 17 Auger Electron Spectroscopy

Birck Nanotechnology Center XPS: X-ray Photoelectron Spectroscopy ESCA: Electron Spectrometer for Chemical Analysis

Nearly Free Electron Gas model - II

Supplementary Information

X-ray Photoelectron Spectroscopy (XPS)

XMM-Newton SOC Technical Note

Defense Technical Information Center Compilation Part Notice

X-Ray Photoelectron Spectroscopy (XPS)

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu

Auger Electron Spectroscopy

EDS User School. Principles of Electron Beam Microanalysis

An XPS study on the surface reduction of V 2 O 5 (001) induced by Ar + ion bombardment

Lecture 10. Transition probabilities and photoelectric cross sections

Spectroscopies for Unoccupied States = Electrons

Electron Spettroscopies

An Introduction to XAFS

Reviewers' Comments: Reviewer #1 (Remarks to the Author)

5) Surface photoelectron spectroscopy. For MChem, Spring, Dr. Qiao Chen (room 3R506) University of Sussex.

Probing Matter: Diffraction, Spectroscopy and Photoemission

Peak-Fitting of High Resolution ToF-SIMS Spectra: A Preliminary Study

Theta Probe: A tool for characterizing ultra thin films and self assembled monolayers using parallel angle resolved XPS (ARXPS)

Photoemission Spectroscopy

Auger Analyses Using Low Angle Incident Electrons

Lattice dynamics, phase transitions and spin relaxation in [Fe(C 5 H 5 ) 2 ]PF 6

Experiment 1 (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with 1,10- Phenanthroline

Keywords: electron spectroscopy, coincidence spectroscopy, Auger photoelectron, background elimination, Low Energy Tail (LET)

P627/Chem 542 Surface/Interface Science Spring 2013 Rutherford Backscattering Lab: Answers to Questions

Supporting Information. for. Advanced Materials, adma Wiley-VCH 2005

Spatial Coherence Properties of Organic Molecules Coupled to Plasmonic Surface Lattice Resonances in the Weak and Strong Coupling Regimes

Photoelectron Spectroscopy

Appearance Potential Spectroscopy

SUPPLEMENTARY INFORMATION

Inelastic soft x-ray scattering, fluorescence and elastic radiation

Photoelectric Effect Experiment

Supporting Information

PHOTOELECTRON SPECTROSCOPY PROBLEMS:

Lawrence Berkeley National Laboratory Recent Work

the Fermi surface of a metallic material can be represented in reciprocal space using angle-resolved photoemission experiments. In this image, the

Local Anodic Oxidation of GaAs: A Nanometer-Scale Spectroscopic Study with PEEM

Determining Chemical Composition. Of Sputtered Uranium Oxide Thin Films. through X-Ray Photoelectron Spectroscopy

Supporting Information

X-ray Spectroscopy. Interaction of X-rays with matter XANES and EXAFS XANES analysis Pre-edge analysis EXAFS analysis

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

Shu Hu 1,2, Matthias H. Richter 1,2, Michael F. Lichterman 1,2, Joseph Beardslee 2,4, Thomas Mayer 5, Bruce S. Brunschwig 1 and Nathan S.

Organic semiconductor heterointerfaces containing bathocuproine

Transcription:

CENTRO DE INVESTIGACIÓN Y DE ESTUDIOS AVANZADOS DEL IPN UNIDAD QUERÉTARO Internal Report (Created: 3/2011. Last Update: 9/2012) The active background method in XPS data peak-fitting Alberto Herrera-Gomez CINVESTAV- Queretaro Content Abstract...1 I. Traditional or static background method in XPS data peak-fitting...1 II. The active background method...3 III. Active versus static background...5 III.1. Underestimation of the peak area with the static method...6 III.2. Dependence of the Lorentzian and Gaussian widths with the fitting energy range...8 III.3. Convergence of the static iterative-shirley background...10 Acknowledgements...11 References...11 Abstract A practical implementation of the active background method is introduced, one which is employed for fitting photoelectric spectra, such as in x-ray (XPS) or ultraviolet (UPS) photoelectron spectroscopy. In contrast with the standard or static approach, in which the background is defined prior to peak-fitting, the active background intensity and shape are defined during the peak-fitting process. The active method provides better fits than the static method. I. Traditional or static background method in XPS data peak-fitting Quantifying the intensity of peaks in x-ray photoelectron spectroscopy (XPS) data requires the proper modeling of the background signal. There are various background types, namely Shirley, Tougaard, linear, exponential and polynomial, that are employed on a case-by-case basis. The traditional, or static, peak-fitting methodology has the following distinguishable characteristics: One of the background types is chosen depending on the shape of the spectrum. Some software allows for a certain predefined combination of two of them. Two points, one at each side of the main features of the spectra (i.e., at each side of the peak or peaks), are chosen. The background is forced to pass though these two points. The background is first subtracted from the original data and then the modified, backgroundless spectrum, is peak-fitted. In some commercially available software, the user

is allowed to manually modify the background by adding a polynomial contribution and forcing the background to pass through additional intermediate points. Sometimes it is necessary to make additional corrections to the data before peak-fitting. For example, if the background on the right (high kinetic-energy side) shows a non-negligible slope, it is necessary to fit a straight line in that region and subtract the extrapolated line from the whole spectrum. 1 The term static is employed for the traditional method in the sense that the background does not play an active role during peak-fitting since, as mentioned in the third point above, it is subtracted prior to peak-fitting. This contrasts with the active method, where the background shape and intensity is defined during peak-fitting. Figure 1 shows a Au 4p 3/2 spectrum from a clean sample. Under the static methodology, a line going through the points on the right of the peak is extrapolated and subtracted to the whole spectrum. Besides the Shirley-type step at the peak, the removal of the linear dependence leaves behind a flat background on both sides since, for this particular case, the slope of the background on the left is the same as that on the right. The remaining stepwise background shape is subtracted employing the iterative Shirley method. 2 The remaining feature is then fitted with a singlet peak. 2/11

80 70 a) Au 4p 3/2 Original spectrum 60 50 40 Points employed to calculate the slope Line subtracted 30 20 10 140 120 Static Shirley Background 860 880 900 920 940 960 980 1000 1020 1040 b) Background Limits Kinetic Energy (ev) Modified Spectrum Au 4d 100 80 60 Au 4p 3/2 40 20 860 910 960 1010 1060 1110 1160 1210 Kinetic Energy (ev) Figure 1. a) Au 4p 3/2 XPS spectrum from a clean gold surface. Some of the steps employed for the static iterative Shirley background removal are shown. b) This figure illustrates that the Au 4p 3/2 peak rides on the decreasing background of the 4d peak. II. The active background method An alternative for treating the background is the active method. The active approach is applied by first constructing a series of functions in the fitting energy range. For the example shown in Figure 1, the functions are a singlet peak, a constant term, a linear term, and an iterative-shirley background. The contribution from each term is found through an optimization process. Since the parameters of the peak change during the fitting process, the relative contributions from the constant, linear, and Shirley backgrounds also change. The background shape and intensity vary during the fitting process. If the position of the proposed peak is off, as illustrated in Figure 2a for the same data as shown in Figure 1, the background is modified to minimize the fitting error for that particular position of the proposed peak. As the peak-fitting procedure continues, the 3/11

shape of the background evolves to that shown in Figure 2b. For this particular example, the static and the active background methods provide essentially the same fit. As shown in the next section, this is not the case for most spectra. a) Au 4p 3/2 b) b) Kinetic Energy (ev) Figure 2. a) The same Au 4p 3/2 spectrum as in Figure 1. Pictured is the first iteration of the active background fit illustrating how the background varies in an active manner responding to the position of the peak. The background is not the same as in (b), but it evolves to be the same as the fit is optimized. b) Optimized fit employing the active method. The background employed is the sum of a baseline, a linear term, and Shirley (k Shirley = 0.032 ev -1 ). The slope of the background is the same on both sides of the peak. The fit is slightly better than in Figure 1a; the area 28.8 a.u., however, is slightly larger than in the static case (26.9 a.u.). The active background method has been implemented in the software AAnalyzer. 3 Under the active method, choosing various backgrounds to work simultaneously can be as easy as selecting the corresponding checkboxes. This is illustrated in Figure 3, which shows the parameters tab for the fit shown in Figure 2b. Note that the Baseline (constant term), 1st Order (linear term), and Shirley backgrounds are chosen within the Active Background group box. The number of iterations for the Shirley background was set to six. 4/11

Figure 3. The fitting parameters tab of AAnalyzer, showing the parameters employed for the fit shown in Figure 1. As indicated in the box next to the Shirley parameter box, six iterations were employed to generate the background. This number was chosen since, generally, further iterations have negligible effects on the background. The implementation of the active background, and its importance, had been covered in other reports. 4,5 It should be mentioned that, besides those in the Active Background group box (circled in Figure 3), there is another background type, the Peak-Shirley background (last column in Figure 3), that can also be used in an active way. This is a variant of the Shirley method that does not require iterations. Its form is very similar to that suggested in another report, 5 and a detailed description can be found elsewhere. 6 The Extrinsic-Intrinsic Simplified background is described in Reference 7, where it is shown a number of examples in which this is combined with the Shirley and 1st order backgrounds under the active method. III. Active versus static background The active method is always better, and often much better, than the traditional or static method for the treatment of the background in photoemission data analysis. The advantages of the active method are many: The area of the peaks is usually underestimated when the static method is employed. As mentioned above, with the static method it is necessary to choose two points, one at each side of the spectrum features or peaks. The operator has to guarantee or assume that there is no appreciable contribution to the signal at those two points from any of the peaks. The method 5/11

does not allow the consideration of peaks that extend beyond either of the chosen points. In contrast, the active treatment does not restrain this possibility since the strength of the background can be varied to account for contributions from peaks that extend beyond the chosen points. The active method should to be employed when the data contains a peak that extends beyond the measurement range. The extent of the errors incurred by using the static method in the quantification of the peak intensity is illustrated in Reference 8. The relative contributions of the Gaussian and Lorentzian widths to the FWHM of the peaks are dependent on the fitting energy range when the static method is employed. As shown below, this is not the case for the active approach. As is well known, the traditional (static) iterative-shirley background algorithm fails for some types of spectra. An example is provided below. Another issue is that it takes longer to converge, that is, more iterations are required with the static than with the active method. The active treatment allows a simultaneous combination of backgrounds. For example, the strength of a linear background employed simultaneously with a Shirley background can be decided by the optimization process. In contrast, the static method requires that the operator chooses one among the background types available (linear, Shirley, Tougaard, etc.), or a fixed combination of them. When the background is subtracted before peak-fitting, the Poisson-character of the data is no longer applicable, so the uncertainty of the peak parameters cannot be calculated employing the covariant matrix method. 9 The operator might forget about the background once it has been subtracted, which can lead to errors in the static method. In the active method, the changing role of the background is accounted for as the operator considers other options during peak-fitting. Last, but not least, the active method always provides better fits than the static method. III.1. Underestimation of the peak area with the static method Figure 4 shows an example in which the use of the static method seems harmless since the background at both sides of the peak is apparently flat. At first look, both fitting methods, static and active, provide approximately the same fit. However, the difference in the calculated area is ~10%. This is because, in the static method, the background is forced to pass through the points P 1 & P 2 (shown in the figure). In contrast, in the active approach the peak is allowed to have a (small) contribution at those points, so the background is slightly lower, resulting in a larger area of the peak. 6/11

a) Static Background b) Active Background C 1s C 1s Lorentzian width: 0.26 ev Gaussian width: 1.25 ev Total width: 1.27 ev Area: 245 a.u. Lorentzian width: 0.52 ev Gaussian width: 1.1 ev Total width: 1.21 ev Area: 269 a.u. P 1 P 2 P 1 P 2 Binding Energy (ev) Figure 4. Fitting of a C 1s peak with the background treated a) using the static method (first subtracted and then fitted) and b) using the active method (optimized during the fitting process). Although the fits look equally good, the difference in the peak area is ~10%. The background is the main source of the difference. Using the active method, the background is allowed to go below the signal at the chosen end points (P 1 at 288.8 and P 2 at 280.9 ev). This freedom allowed for larger values of the Lorentzian width during the optimization process. The problem might be more serious for asymmetric peaks for which the tail on the left could be considerably prolonged. This is exemplified in Figure 5 with an Fe 2p spectrum. (The asymmetry in Figure 5 was accounted for through the double-lorentzian line shape described in Reference 10). Although the signal is already flat for binding energies above 730 ev, the analysis performed with the active background method shows that the last peak to the left extends well beyond 732.5 ev, the left extreme of the energy range in which the data was acquired. The inset in Figure 5b shows that the fit done with the active background method is excellent, even when the 1:2 ratio for the intensity of the 2p 1/2 and 2p 3/2 branches is forced during the fitting. A good fit is not possible with the static method if the branching ratio restriction is applied (see inset in Figure 5a). Another important difference is that the ratio of the intensities of the oxidized and metallic peaks changes by 50%. This would directly affect the assessment of the thickness of the oxide layer by about that same percentage. 7/11

18000 16000 14000 Fe 2p a) Oxidized 12000 10000 8000 P 1 Metallic 6000 P 2 4000 2000 0 18000 16000 14000 730 Fe 2p 725 2p 1/2 720 Traditional (static) Shirley background 715 710 Binding Energy (ev) b) 2p 3/2 705 700 695 12000 10000 8000 6000 4000 Active Shirley background 2000 0 730 725 720 715 710 705 700 695 Binding Energy (ev) Figure 5. Fe 2p XPS spectrum from an iron substrate with a thin iron oxide layer. The Shirley background is shown for various iterations for the a) static method, and b) active method. The thicker lines correspond to the final convergence shape. The background takes longer to converge and arrives at a very different value with the static method. The reason is that, under this method, the background is obligated to go through the point P 1. The active method allows for the possibility that a peak could contribute to the signal at that position. This is indeed the case, as the data was acquired over an energy range that was not wide enough. The insets show that the fit is much better with the active than with the static method. III.2. Dependence of the Lorentzian and Gaussian widths with the fitting energy range The spectrum shown in Figure 6 for In 3d 5/2 was fitted employing the five energy range sets indicated by the pairs of numbered vertical lines. 8/11

In 3d 5/2 1 2 3 4 5 5 4 3 2 1 Binding Energy (ev) Figure 6. In 3d 5/2 spectrum (dots). The pairs of vertical lines indicate the energy range over which peakfitting was carried out. The results of the fits for the different energy ranges, and employing both the static and active background methods, are shown in Figure 7. In this figure, the lines correspond to the fit carried out employing the active background method and the energy range 1. Figure 7 shows the results of the fitting of the data shown in Figure 6 for the various energy ranges and employing both the static and active background methods. The peak height and the Lorentzian and Gaussian widths were allowed to vary during the analysis process to optimize the fit. Besides being underestimated, the area assessed with the static method has a large dependence on the fitting range. Likewise, the Lorentzian and Gaussian widths also show a large dispersion when the static method is employed. 9/11

Peak Area (a.u.) Peak Width (ev) 500 400 300 200 100 0 0.6 0.4 0.2 0.0 Static FWHM Active FWHM Static Gaussian Active Gaussian Static Lorentzian Active Lorentzian Static Active 1 2 3 4 5 Fitting Range Figure 7. Results of the fits performed employing the static and active background methods for the data and energy ranges shown in Figure 6. The results employing the active method show much lower dispersion than with the static background method. III.3. Convergence of the static iterative-shirley background For some spectra, when the energy range is large or it contains large, sharp peaks with low regions in between, the traditional iterative-shirley algorithm fails. This is illustrated in Figure 8 for an Fe 2p spectrum from a clean iron surface. The static Shirley background does not converge and has large oscillations between even and odd iterations, whereas the active iterative- Shirley algorithm converges quickly to produce an excellent fit. 10/11

Fit employing the active iterative-shirley background Static iterative-shirley background Active iterative-shirley background Binding Energy (ev) Figure 8. Fe 2p spectrum for a clean iron sample. It shows the lack of convergence of the traditional or static iterative-shirley background method. The asymmetry was modeled with the Double- Lorentzian line shape. 10 There are many advantages and no apparent drawbacks when the active method is employed instead of the traditional (static) background methodology. The active method provides better fits with much lower dependence on operator choices, such as the fitting energy range. Acknowledgements The technical support from Mariela Bravo-Sanchez and Milton Vazquez-Lepe is deeply appreciated. The editions and comments to the manuscript from Michael S. Boldrick, from Peace Corps, are also deeply appreciated. References 1 2 3 4 5 6 7 8 9 10 S. Tougaard. Surface Science 216 (1989) 343-360. P.M.A. Sherwood, in: D. Briggs, M.P. Seah (Eds.), Practical Surface Analysis by Auger and X-ray Photoelectron Spectroscopy, Wiley, Chichester, 1983 (Appendix 3). More information about the software can be found at www.rdataa.com/aanalyzer. A.M. Salvi and J.E. Castle. J. Elec. Spec. Rel. Phen. 95 (1988) 45. J. Vegh. J. Elec. Spec. Rel. Phen. 46 (1988) 411. A. Herrera-Gomez. The Peak-Shirley Background. Internal Report. CINVESTAV-Unidad Queretaro (2011). http://www.qro.cinvestav.mx/~aherrera/reportesinternos/peakshirley.pdf A. Herrera-Gomez. Simplified near-peak background for extrinsic and intrinsic electron scattering in XPS data peak-fitting. To be submitted to J. Elec. Spec. Rel. Phen. (2012). J. Muñoz-Flores, A. Herrera-Gomez. Underestimation of the peak areas for the static and active background methods in XPS peak-fitting. To be submitted to J. Elec. Spec. Rel. Phen. (2012). J. Muñoz-Flores, A. Herrera-Gomez. Resolving overlapping peaks in ARXPS data: the effect of noise and fitting method. J. Elec. Spec. Rel. Phen. 184 (2012) 533 541, and references therin. A. Herrera-Gomez. A double Lorentzian shape for asymmetric photoelectron peaks. Internal Report. Cinvestav-Queretaro.(2011). http://www.qro.cinvestav.mx/~aherrera/reportesinternos/doublelorentzian.pdf 11/11