HPLC analysis of VFAs, Furfural and HMF

Similar documents
for Acclaim Mixed-Mode HILIC-1 Column

Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Appendix II- Bioanalytical Method Development and Validation

Determination of trace anions in concentrated hydrofluoric acid

Scope: Materials: Labeled 1.5 ml or 2 ml Eppendorf tubes Appropriate pipette tips (see Table 1) Labeled LC vials

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

LC and LC/MS Column Selection Flow Chart

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

The analysis of organic acid content of additives, premix, feed, and water.

Available online Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article

Agilent LC & LC-MS Cannabis Testing

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Dilution(*) Chromatography

Definitive EtG/EtS LC-MS/MS Analysis:

FAQ's. 1 - Which column would be the most appropriate for my application?

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Experiment Protocol 201. Liquid Chromatography) TAMAGAWA SEIKI CO., LTD

Packings for HPLC. Packings for HPLC

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

[ CARE AND USE MANUAL ]

The Theory of HPLC. Quantitative and Qualitative HPLC

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

What actually are dietary supplements? They are consumed in large quantities, but they

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

INTERNATIONAL CHEMISTRY TESTING. 258 Main Street- Suite 311- Milford, MA Tel: Fax:

C18 Column. Care & Use Sheet

Unexpected Peaks in Chromatograms - Are They Related Compounds, System Peaks or Contaminations? From the Diary of an HPLC Detective

Quantification of Furocoumarins

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Work plan & Methodology: HPLC Method Development

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

GPC/SEC Practical Tips and Tricks. Thomas Dent Applications Scientist Agilent Technologies. October, 2011 Gulf Coast Conference

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

ACQUITY UPLC Peptide BEH C 18, 130Å and 300Å Columns

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

ProPac WCX-10 Columns

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

Shodex TM ODP2 HP series columns

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Reversed Phase Solvents

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

LUMEFANTRINUM LUMEFANTRINE

Solid Phase Extraction Method Development Tips and Tricks

penta-hilic UHPLC COLUMNS

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

CIPAC. CIPAC Free relevant impurities methods:

High Performance Liquid Chromatography

Acclaim Mixed-Mode WAX-1 Columns

11,12-EET/DHET ELISA kit

LC Technical Information

DEVELOPMENT AND VALIDATION OF NEW ANALYTICAL METHOD FOR BIOACTIVE COMPOUNDS

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

Supporting Information

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

LC-MS compatible Separation of the Fungicide Spiroxamine

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

DISCLAIMER: This method:

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Chromatography. Gas Chromatography

Too Polar for Reversed Phase What Do You Do?

Chromatographic Analysis

[ CARE AND USE MANUAL ] XSELECT HSS XP 2.5 µm COLUMNS I. INTRODUCTION CONTENTS I. INTRODUCTION II. GETTING STARTED III. COLUMN USE

Supporting Information

PYRIPROXYFEN TECHNICAL

High Performance Liquid Chromatography

High Performance Liquid Chromatography

High Performance Liquid Chromatography

HPLC Winter Webinars Part 2: Sample Preparation for HPLC

How proteins separate on reverse-phase HPLC

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION

2.1 2,3 Dichloro Benzoyl Cyanide (2,3 DCBC) and survey of. manufactured commonly for the bulk drug industry, few references were

H 3 CO H 3 CO S CH 3

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

LC800. Smart HPLC. Until now UHPLC From now Smart HPLC

columns IonPac AS17-C Anion Exchange Column

The Theory of HPLC. Ion Pair Chromatography

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

ACETONE IN URINE BY UV CODE Z42010

IDENTIFICATION OF STEROIDS IN COSMETIC PRODUCTS BY TLC AND HPLC 1 02/12/2005 ACM 007 A. THIN LAYER CHROMATOGRAPHY (TLC)

ARTEMETHER AND LUMEFANTRINE ORAL SUSPENSION:Final text for addition to The International Pharmacopoeia (November 2008)

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

ARTEMETHER AND LUMEFANTRINE TABLETS: Final text for addition to The International Pharmacopoeia (July 2008)

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

LC Column Troubleshooting. Do you have an equivalent column for my LC column?

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

Comprehensive 2D-LC Analysis of Chinese Herbal Medicine

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Value of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs

penta-hilic UHPLC COLUMNS

Development And Validation Of Rp-Hplc Method For Determination Of Velpatasvir In Bulk

Transcription:

HPLC analysis of VFAs, Furfural and HMF Introduction The following chromatographic analysis is developed in order to analyse volatile fatty acids and carbohydrate degradation products; furfural, hydroxymethylfurfural (HMF) and levulinic acid. The purpose of this method is to perform routine analysis of biogas and biomass pretreatment samples. Other purposes may also be applicable. The principle of separation of these analytes is that they are organic molecules with different hydrophobicity, and the separation principle is therefore based on reversed phase (RP) separation. It is important for efficient separation that ph in the mobile phase is below the pka value of the analyte with the lowest pka value as this will cause the analytes to be predominantly protonated. Due to a complex sample matrix and due to high water solubility for the shortest VFAs, the gradient profile is designed with relatively weak eluting conditions. Volatile fatty acids can be analyzed up to at least C 7. The sensitivity of the analysis is dependent on the injection volume and on the analyte. For VFAs at least 1 mm of each analyte is quantifiable at 1 µl injection. The sensitivity for HMF/furfural is approx. 1 times better than for VFAs and therefore down to.1 mm is quantifiable in 1 µl injection. If lower concentrations of analytes are expected in the samples, larger injection volume may be applied (up to 1µL). HPLC system The analysis is performed on the Dionex Ultimate 3 chromatographic system with UV detection. See relevant instrumental SOP for operation of this system. Analytical column The analytical column applied is Zorbax Eclipse Plus C18 from Agilent Technologies. For this SOP a 15x2.1 mm column is used (3.5 µm particles), equipped with guard column of the same brand, 12.5x2.1 mm, (5 µm particles). The column is based on usual reversed phase C18 technology (dimethyl-noctadecyl double endcapped silica). The columns ph range is ph 2-9 and is fully compatible with usual RP-eluent systems like acetonitrile and methanol. Column precautions The normal operating back pressure of the column incl. guard column is in the range of 1-14 bars at.3 ml/min, depending on the concentration of organic solvent. Maximum operating pressure for the Page 1 of 7

column is 4 bars. Between runs, the column should be stored in 5/5 water/acetonitrile or methanol for shorter time storage, and 1% organic solvent for longer time storage. Operating conditions Column temperature: 4 C Flow conditions:.3 ml/min Eluents: 1% Methanol (A) 2.5 mm H 2 SO 4 (B) Injection volume: 1 µl Instrumental method: ZORBAX_BIOGAS_VFAs (Table 1). Note that there are two versions of the method, dependent on which pump you will use (method name ends by either RightPump or LeftPump). Table 1: Gradient profile for Instrument method ZORBAX_BIOGAS_VFAs. Time (min) A (%) B (%) : 1 2:5 1 2:51 15 85 25: 15 85 25:1 1 35: 1 The analytes of interest in this method have very different hydrophobic properties and due to the nature of the samples (sample matrix), the elution profile is starting with weak elution strength (isocratic 1% 2.5mM H 2 SO 4 ) to achieve satisfactory retention/separation between the early eluting species (e.g. formic and acetic acid). After a short period of time, 2.5 min, the elution strength is increased slightly in order to elute the more hydrophobic analytes. Since the elution occurs at relatively mild conditions, re-equilibration to initial conditions is crucial and set to 1 minutes of the run time. Page 2 of 7

The analytes have different wavelength optima and therefore the instrumental method includes detection at two different wavelengths, namely 21 (for the organic acids) and 278 nm (HMF and furfural). Chromatographic run 69,2 65, mau 1 6, 55, 5, 45, 5mM H2SO4: 1, % 2 6 7 8 1, 4, 35, 3, 3 4 9 25, 2, 15, 5 1, 5, 85,, -5, Flow:,3 ml/min -1,8, 2, 4, 6, 8, 1, 12, 14, 16, 18, 2, 22, 24, 26, 28, 3, 32, 34, 35, min Figure 1: Chromatographic profile. 1 Formic acid, 2 Acetic acid, 3 Propionic acid, 4 Levulinic acid, 5 Impurity from levulinic acid std., 6 HMF, 7 Furfural, 8 Butyric acid, 9 Valeric acid. 1 µl injection, 21 nm Sample preparation Samples must be acidified prior to analysis (ph <2.5). If the samples are not acidic by nature, H 2 SO 4 is an appropriate acid to use. Apply as little volume as possible to avoid unnecessary sample dilution. NB! make sure signals are within the calibration range. Most often biogas samples do not need dilution, on the other hand you may need to increase injection volume for samples with low VFA-concentrations. Samples must be free from any particulate/insoluble matter and should preferably be colorless. If samples contain insoluble matter this MUST be removed either by centrifugation or by.22/.45 µm filtration. Particularly for biogas samples an appropriate sample preparation is: 1. Centrifuge the samples as hard as possible. 2. Take an aliquot of the supernatant into an eppendorf tube and centrifuge a second time at max. speed (15. rpm for table centrifuges). 3. Take an aliquot of the supernatant and add a little amount of strong sulfuric acid (72%) (2 µl to approx. 2 µl sample) and leave the sample on the table for a few minutes. The added volume Page 3 of 7

does not need to be exactly precise as long as the volume of acid is much smaller (< 1/1 times) than volume of sample. Carbon dioxide formation will occur and result in a pressure buildup (fizz) and pressure may need to be released from the eppendorf tubes. 4. The acidification will also precipitate appreciable amounts of colorants etc. and therefore the samples should be centrifuged hard or filtered one last time before dispensed into HPLC vials (note that air easily is entrapped in the bottom of the HPLC vials, air is removed by flicking tubes repeatedly). Calibration The calibration range is different for VFAs and HMF/furfural. VFAs should be analysed in a range from 1 mm to 1 mm. The calibration curve is linear to at least 5 mm at 1 µl injection, but poor peak shape is observed for butyric and valeric acid at concentrations above 25 mm. For low concentration samples, larger injection volume may be applied. Up to 1 µl injection is applicable, but this could affect the peak shape negatively. Figure 2 shows an example of a calibration curve for VFAs. Limit of detection (LOD) and limit of quantification (LOQ) in absolute amounts for VFAs are listed in table 2. Table 2: LOD/LOQ values for VFAs measured at 21 nm VFA LOD (nmol) LOQ (nmol) Formic acid.2.6 Acetic acid.3 1. Propionic acid.6 1.8 Butyric acid.4 1.4 Valeric acid 1.2 3.9 Page 4 of 7

8 7 6 FA: y =,1357x R² =,9999 Ace: y =,162x R² = 1 FA Ace Area (mau*min) 5 4 3 2 1 2 4 6 Prop: y =,1263x R² = 1 But: y =,1488x R² = 1 Val: y =,1312x R² =,9995 Prop But Val Linear (FA) Linear (Ace) Linear (Prop) Linear (But) Linear (Val) Concentration (mm) Figure 2: Calibration curve for VFAs at 1 µl injection and detection at 21 nm The sensitivity for HMF and furfural are significantly better then for VFAs and at 278 nm a suitable calibration range for these two analytes is from.1 mm to 5 mm. At high concentrations base line separation may not be achieved, and calibration based on peak height may be more applicable than calibration on peak area. Both methods provide good fit, since the peak shapes for HMF and furfural are fairly good. Figure 3 and 4 show examples of calibration for HMF and furfural both for area and peak height calibration. HMF and furfural have similar limit of detection (LOD) and limit of quantification (LOQ) and these are.3 nmol and 1 nmol, respectively. Always provide a new set of standards for the analytes of interest in every sequence and perform a recalibration in the processing method. The calibration curves provided here are meant as inspiration. An existing processing method may be found on the instrument and is named VFA-Furdehydes-biogas. Page 5 of 7

Area (mau*min) 25 2 15 1 5 2 4 6 Conc (mm) 2 18 16 14 12 1 8 6 4 2 Height (mau) HMF-Area HMF-Height Linear (HMF-Area) Linear (HMF-Height) Area: y = 43,742x R² =,9999 Height: y = 371,78x R² =,9996 Figure 3: Calibration curve for HMF at 1 µl injection and detection at 278 nm Area (mau*min)/height (mau) 2 18 16 14 12 1 8 6 4 2 2 4 6 Conc (mm) Furfural-Area Furfural - height Linear (Furfural-Area) Linear (Furfural - height) Heigth: y = 352,24x R² =,9998 Area: y = 43,76x R² =,9999 Figure 4: Calibration curve for furfural at 1 µl injection and detection at 278 nm Trouble shooting Identify if any immediate precautions should be taken and then call instrument super users (currently Jane W. A. or Bjørge W., IKBM-PEP). Page 6 of 7

Column regeneration Biogas samples are complex matrices that will contain many different components other than those of interest for the analysis. These can be observed either as background noise (unstable baseline) or they will simply stay on the column because the eluting conditions are too weak to remove them all during a normal analytical run. It is therefore necessary to end EVERY sequence with a regeneration/cleaning program so that residual compounds sticking to the column will be removed. If this is not done, the quality of the analysis will decrease very fast because column capacity will be exceeded. The cleaning program contains a longer elution profile with high elution strength and should be included as the last program in a sequence running a blank sample. The name of the cleaning program is: ZORBAX_BIOGAS_CLEAN_RightPump/LeftPump Column storage conditions Column storage times longer than 5 days, should be done in 1% ACN/MeOH. For shorter storage times lower organic solvent concentration is applicable, but please replace any buffers with water. This is important for the column, but also for the instruments lifetime! Editorial log First edition. 212.1.27 Page 7 of 7