3WHB10K 3-Wire Half Bridge Terminal Input Module

Similar documents
3WHB10K 3-Wire Half Bridge Terminal Input Module

INSTRUCTION MANUAL. CR10XTCR Thermocouple Reference. Revision: 1/03. Copyright (c) Campbell Scientific, Inc.

3WHB10K 3-WIRE HALF BRIDGE TERMINAL INPUT MODULE INSTRUCTION MANUAL

IRTS-P Precision Infrared Temperature Sensor

CVD20 20:1 Voltage Divider Revision: 5/12

INSTRUCTION MANUAL. Model 207 Temperature and Relative Humidity Probe. Revision: 2/94. Copyright (c) Campbell Scientific, Inc.

INSTRUCTION MANUAL. CSI Model TB4 Rain Gage. Revision: 8/04. Copyright (c) Campbell Scientific, Inc.

Model 108 Temperature Probe Revision: 4/07

INSTRUCTION MANUAL. CS500 Temperature and Relative Humidity Probe. Revision: 7/04. Copyright (c) Campbell Scientific, Inc.

Model ASPTC Aspirated Shield with Fine Wire Thermocouple

WXT510 Weather Transmitter Revision: 4/07

INSTRUCTION MANUAL. LI200X-LC Pyranometer for MetData1 4/97. Copyright (c) Campbell Scientific, Inc.

INSTRUCTION MANUAL. CS100 Barometric Pressure Sensor 3/05. Copyright (c) Campbell Scientific, Inc.

CR10XTCR THERMOCOUPLE REFERENCE

CR10TCR THERMOCOUPLE REFERENCE

INSTRUCTION MANUAL. CS547 Conductivity and Temperature Probe and A547 Interface. Revision: 9/00. Copyright (c) Campbell Scientific, Inc.

CS100 Barometric Pressure Sensor

MODEL 107 TEMPERATURE PROBE INSTRUCTION MANUAL

Alter-Type Wind Screen for Tipping Bucket Rain Gages Revision: 1/12

IRTS-P PRECISION INFRARED TEMPERATURE SENSOR INSTRUCTION MANUAL

4WFB120, 4WFB350, 4WFB1K 4 Wire Full Bridge Terminal Input Modules Revision: 5/07

Model HFP01 Soil Heat Flux Plate Revision: 7/12

CR10XTCR THERMOCOUPLE REFERENCE

PTA-427 BAROMETRIC PRESSURE TRANSDUCER OPERATOR S MANUAL

227 Delmhorst Cylindrical Soil Moisture Block Revision: 5/13

MODEL HMP45C TEMPERATURE AND RELATIVE HUMIDITY PROBE INSTRUCTION MANUAL

SI-111 Precision Infrared Radiometer Revision: 7/13

Using CSIM11 ph and ORP Probes with Campbell Scientific Dataloggers

SP-LITE SILICON PYRANOMETER INSTRUCTION MANUAL

INSTRUCTION MANUAL. Model HFP01SC Self-Calibrating Soil Heat Flux Plate Revision: 3/14. Copyright Campbell Scientific, Inc.

INSTRUCTION MANUAL. SP Lite2 Pyranometer. August Copyright 2017 Campbell Scientific (Canada) Corp.

TDR Probes CS605, CS610, CS630, CS635, CS640, CS645 Revision: 9/13

Using CSIM11 ph and ORP Probes with Campbell Scientific Dataloggers Revision: 4/10

MET ONE RAIN GAGE MODELS 380, 385, 380M, 385M

HMP60 Temperature and Relative Humidity Probe Revision: 7/12

WARRANTY AND ASSISTANCE

CM3 PYRANOMETER INSTRUCTION MANUAL

INSTRUCTION MANUAL. 229 Heat Dissipation Matric Water Potential Sensor Revision: 5/09. Copyright Campbell Scientific, Inc.

Eppley PIR Precision Infrared Radiometer

INSTRUCTION MANUAL. WXT520 Weather Transmitter Revision: 9/15. Copyright Campbell Scientific, Inc.

INSTRUCTION MANUAL. CS616 and CS625 Water Content Reflectometers Revision: 2/14. Copyright Campbell Scientific, Inc.

INSTRUCTION MANUAL. BlackGlobe Temperature Sensor for Heat Stress Revision: 1/15. Copyright Campbell Scientific, Inc.

TEMPERATURE COMPENSATION FOR MEAS PRESSURE SENSORS APPLICATION NOTE

NOVALYNX CORPORATION MODEL 110-WS-16BP BAROMETRIC PRESSURE SENSOR INSTRUCTION MANUAL

INSTRUCTION MANUAL. CS100 Barometric Pressure Sensor Revision: 7/16. Copyright Campbell Scientific, Inc.

INSTRUCTION MANUAL. TE525 Tipping Bucket Rain Gage Revision: 12/15. Copyright Campbell Scientific, Inc.

INSTRUCTION MANUAL. TE525 Tipping Bucket Rain Gage Revision: 11/17. Copyright Campbell Scientific, Inc.

Temperature Compensation for MEAS Pressure Sensors

INSTRUCTION MANUAL. EE181 Temperature and Relative Humidity Probe Revision: 3/17. Copyright Campbell Scientific, Inc.

ISOCON-6. 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER

89BSD CALCULATION METHOD APPLICATION NOTE

32 Input AC Block I/O Module

Program Notes Use 2:1 voltage divider on Differential Input channel 1 to get full output range from C7

LM135 LM235 LM335 LM135A LM235A LM335A Precision Temperature Sensors

16 Input AC Block I/O Module

Installation Manual. Models 4911A/4911. VW Rebar Strain Meters

BMXART0814 analog input module M340-8 inputs - temperature

BMXART0814 analog input module M340-8 inputs - temperature

Instruction Manual HI Refractometer for Invert Sugar Measurements

Small, Gauge Pressure Sensor

LM135/LM235/LM335, LM135A/LM235A/LM335A Precision Temperature Sensors

Exercise 1: Thermocouple Characteristics

HR-Z SERIES Mild Radiation Resistance LVDT

INSTRUCTION MANUAL. Laser Diode Sarcomere Length. Aurora Scientific Inc. 360 Industrial Parkway South, Unit 4 Aurora, Ontario, Canada L4G 3V7

Small Gage Pressure Sensor

Drive electronics pq12

SPECIFICATION FOR APPROVAL. Approved by

Designing a Thermostat Worksheet

INSTRUCTION MANUAL ANALOG COLORIMETER

ISOCON-3 MAINS POWERED ISOLATING SIGNAL CONVERTER

LUDLUM MODEL 42-5 NEUTRON BALL CART. August 2015 Serial Number PR and Succeeding Serial Numbers

Small Absolute Pressure Sensor

792A AC/DC Transfer Standard

792A AC/DC Transfer Standard

AN3002 Thermocouple measurement

LM /LM /LM Micropower Voltage Reference Diode

Instruction Manual HI Refractometer for Sucrose in Wine and Grape Products Measurements

IntraSense TM Series 1-French Wire-Connected Pressure Sensor SMI-1A AAUU

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

DS 300. Electronic Pressure Switch. with IO-Link interface. Stainless Steel Sensor. accuracy according to IEC 60770: 0.35 % FSO.

Route and retain the clinometer cable so that it does not interfere with moving parts where the unit is mounted.

Parallel Circuits. Chapter

User s Manual: MODEL 904-T CLINOMETER PAK

CERTIFICATE OF CALIBRATION

HIGH CURRENT BRIDGE DRIVER and 4-20mA Transmitter

WELDED STEEL PIPE NIPPLE PRICE SHEET

WELDED STEEL PIPE NIPPLE PRICE SHEET

IMPROVED PRODUCT Z, Z (Z-Foil)


Exercise 1: Thermistor Characteristics

LM135/LM235/LM335, LM135A/LM235A/LM335A Precision Temperature Sensors

LM /LM /LM Micropower Voltage Reference Diode

CS705 Snowfall Adapter Revision: 5/13

Presco AG. Standard Air Capacitor 10 nf / 2 kv. Type CL-0N10-K201. User's manual

APPLICATION NOTES FOR MULTILAYER CERAMIC CAPACITORS

THERMOCOUPLE CHARACTERISTICS TRAINER

Low Pressure Sensor Amplified Analog Output SM6295-BCM-S

Series FEATURES AND BENEFITS INTRODUCTION. Figure 1 Power Derating Curve. Table 1 Tolerance and TCR vs. Resistance

Bulk Metal Foil Technology Industrial Grade Miniature Voltage Divider with TCR Tracking of 1.5 ppm/ C and Ratio Stability of 0.

LD1117A SERIES LOW DROP FIXED AND ADJUSTABLE POSITIVE VOLTAGE REGULATORS

Transcription:

3WHB10K 3-Wire Half Bridge Terminal Input Module User Manual Issued 19.9.07 Copyright 1996-2007 Campbell Scientific Inc. Printed under licence by Campbell Scientific Ltd. CSL341

Guarantee This equipment is guaranteed against defects in materials and workmanship. This guarantee applies for twelve months from date of delivery. We will repair or replace products which prove to be defective during the guarantee period provided they are returned to us prepaid. The guarantee will not apply to: Equipment which has been modified or altered in any way without the written permission of Campbell Scientific Batteries Any product which has been subjected to misuse, neglect, acts of God or damage in transit. Campbell Scientific will return guaranteed equipment by surface carrier prepaid. Campbell Scientific will not reimburse the claimant for costs incurred in removing and/or reinstalling equipment. This guarantee and the Company s obligation thereunder is in lieu of all other guarantees, expressed or implied, including those of suitability and fitness for a particular purpose. Campbell Scientific is not liable for consequential damage. Please inform us before returning equipment and obtain a Repair Reference Number whether the repair is under guarantee or not. Please state the faults as clearly as possible, and if the product is out of the guarantee period it should be accompanied by a purchase order. Quotations for repairs can be given on request. When returning equipment, the Repair Reference Number must be clearly marked on the outside of the package. Note that goods sent air freight are subject to Customs clearance fees which Campbell Scientific will charge to customers. In many cases, these charges are greater than the cost of the repair. Campbell Scientific Ltd, Campbell Park, 80 Hathern Road, Shepshed, Loughborough, LE12 9GX, UK Tel: +44 (0) 1509 601141 Fax: +44 (0) 1509 601091 Email: support@campbellsci.co.uk www.campbellsci.co.uk

Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat bookmarks tab for links to specific sections. 1. Function... 1 2. Specifications... 1 3. Wiring... 2 4. Programming Examples... 2 4.1 CR9000X...3 4.2 CR1000...3 4.3 CR10(X)...4 4.4 21X...4 4.5 CR7...5 5. 100 Ohm PRT in 3-Wire Half Bridge... 5 5.1 Excitation Voltage...5 5.2 Calibrating a PRT...6 5.3 Compensation for Wire Resistance...6 Figures Tables 1-1. Terminal Input Module...1 2-1. Schematic...1 3-1. 3-Wire Half Bridge Used to Measure PRT...2 3-1. 3WHB10K Connections to Campbell Scientific Dataloggers...2 4-1. Excitation Voltage for 100 Ohm PRT in 3WHB10K Based on Maximum Temperature and Input Voltage Range...3 i

This is a blank page.

3WHB10K 3-Wire Half Bridge Terminal Input Module 1. Function Terminal input modules connect directly to the datalogger's input terminals to provide completion resistors for resistive bridge measurements, voltage dividers, and precision current shunts. H L G H L AG H L AG Figure 1-1. Terminal Input Module 2. Specifications 10 kohm Completion Resistor Tolerance @ 25 C ±0.01% Temperature coefficient 0-60 C ±4 ppm/ C -55-125 C ±8 ppm/ C Power rating @ 70 C 0.25 W Vx 10 kω H H L L or AG G Figure 2-1. Schematic 1

3WHB10K 3-Wire Half Bridge Terminal Input Module 3. Wiring Datalogger Vx 10 kω H H A L L or AG G B or G Shield Figure 3-1. 3-Wire Half Bridge Used to Measure PRT Function Table 3-1. 3WHB10K Connections to Campbell Scientific Dataloggers Label/Lead CR10X, CR510 CR23X, CR1000, CR800, CR850, CR3000 21X, CR7, CR9000X Excitation Black Wire E1 EX1 Excitation 1 V1 Reference H SE1 SE1 1H V2 Sense L SE2 SE2 1L Ground G AG 4. Programming Examples The following examples simply show the two instructions necessary to 1) make the measurement and 2) calculate the temperature. The result of the 3-wire half bridge measurement as shown is R s /R o, the input required for the PRT algorithm to calculate temperature. All the examples are for a 100 Ohm PRT in the 3WHB10K. The excitation voltages used were chosen with the assumption that the temperature would not exceed 50 C. Table 4-1 lists excitation voltage as a function of maximum temperature and the input voltage ranges used with the different dataloggers. Calculation of optimum excitation voltage is discussed in Section 5.1. The multiplier shown is for a 100 Ohm PRT. The multiplier for a 1000 Ohm PRT is 10. 2

User Manual Table 4-1. Excitation Voltage for 100 Ohm PRT in 3WHB10K Based on Maximum Temperature and Input Voltage Range Max. Temp C PRT Resistance Ohms Excitation Voltage, mv ±25 mv Input Range, CR10(X), CR800, CR850, CR1000, ±50 mv Range, 21X, CR7, CR3000, CR9000X 50 119.4 2119 4237 100 138.5 1830 3660 150 157.31 1614 3228 200 175.84 1447 2893 250 194.07 1313 2626 300 212.02 1204 2408 350 229.67 1113 2227 400 247.04 1037 2074 450 264.11 971 1943 500 280.9 915 1830 550 297.39 866 1731 600 313.59 822 1644 650 329.51 784 1567 700 345.13 749 1499 750 360.47 718 1437 800 375.51 691 1381 850 390.26 666 1331 4.1 CR9000X 'CR9000X Datalogger Public RS_Ro, Temp_F DataTable (Temp_F,1,-1) DataInterval (0,0,0.10) Sample (1,Temp_F,FP2) EndTable BeginProg Scan (1,mSec,0,0) BrHalf3W (Rs_Ro,1,V50,5,1,6,1,1,4200,True,30,40,100,0) PRT (Temp_F,1,Rs_Ro,1.8,32) CallTable Temp_F NextScan EndProg 4.2 CR1000 'CR1000 Series Datalogger Public Rs_R0, Temp_C DataTable (Hourly,True,-1) DataInterval (0,60,Min,0) 3

3WHB10K 3-Wire Half Bridge Terminal Input Module Average (1,Temp_C,IEEE4,0) EndTable BeginProg Scan (1,Sec,0,0) BrHalf3W (Rs_R0,1,mV25,1,Vx1,1,2100,True,0,250,100,0) PRT (Temp_C,1,Rs_R0,1.0,0) CallTable Hourly NextScan EndProg 4.3 CR10(X) 1: 3W Half Bridge (P7) 1: 1 Reps 2: 23 ± 25 mv 60 Hz Rejection Range 3: 1 SE Channel 4: 1 Excite all reps w/exchan 1 5: 2100 mv Excitation 6: 1 Loc [ Rs_R0 ] 7: 100 Mult 8: 0 Offset 2: Temperature RTD (P16) 1: 1 Reps 2: 1 R/RO Loc [ Rs_R0 ] 3: 2 Loc [ Temp_C ] 4: 1.0 Mult 5: 0.0 Offset 4.4 21X 1: 3W Half Bridge (P7) 1: 1 Reps 2: 3 ± 50 mv Slow Range 3: 1 SE Channel 4: 1 Excite all reps w/exchan 1 5: 4200 mv Excitation 6: 1 Loc [ Rs_R0 ] 7: 100 Mult 8: 0 Offset 2: Temperature RTD (P16) 1: 1 Reps 2: 1 R/RO Loc [ Rs_R0 ] 3: 2 Loc [ Temp_C ] 4: 1.0 Mult 5: 0 Offset 4

User Manual 4.5 CR7 1: 3-Wire Half Bridge (P7) 1: 1 Reps 2: 4 ± 50 mv Slow Range 3: 1 In Card 4: 1 SE Channel 5: 1 Ex Card 6: 1 Ex Channel 7: 1 Meas/Ex 8: 4200 mv Excitation 9: 1 Loc [ Rs_R0 ] 10: 100 Mult 11: 0 Offset 2: Temperature RTD (P16) 1: 1 Reps 2: 1 R/RO Loc [ Rs_R0 ] 3: 2 Loc [ Temp_C ] 4: 1 Mult 5: 0 Offset 5. 100 Ohm PRT in 3-Wire Half Bridge The advantages of the 3-wire half bridge over other measurements that correct for lead wire resistance such as a 4-wire half bridge, are that it only requires 3 lead wires going to the sensor and takes 2 single-ended input channels, whereas the 4-wire half bridge requires 4 wires and 2 differential channels. The result of the 3-wire half bridge instruction is equivalent to the ratio of the PRT resistance, R s to the resistance of the 10 k fixed resistor, R f. R R The RTD Instruction (16) computes the temperature ( C) for a DIN 43760 standard PRT from the ratio of the PRT resistance at the temperature being measured (R s ) to its resistance at 0 C (R 0 ). Thus, a multiplier of R f /R 0 is used with the 3-wire half bridge instruction to obtain the desired intermediate, R s /R 0 = (R s /R f x R f /R o ). When R f = 10,000 and R 0 = 100, the multiplier is 100; when R 0 is 1000 the multiplier is 10. The fixed resistor must be thermally stable. Over the -55 to 85 C extended temperature range for the datalogger, the ±4 ppm/ C temperature coefficient would result in a maximum error of ±0.04 C at 60 C. The ±8 ppm/ C temperature coefficient would result in a maximum error of ±0.13 C at -55 C. s f 5.1 Excitation Voltage The best resolution is obtained when the excitation voltage is large enough to cause the signal voltage to fill the measurement voltage range. The voltage drop across the PRT is equal to the current, I, multiplied by the resistance of the PRT, R s, and is greatest when R s is greatest. For example, if it is desired to measure a temperature in the range of -10 to 40 C, the maximum voltage drop will be at 40 C when R s = 115.54 Ohms. To find the maximum excitation voltage that can 5

3WHB10K 3-Wire Half Bridge Terminal Input Module be used when the measurement range is ±25 mv, we assume V2 equal to 25 mv and use Ohm's Law to solve for the resulting current, I. I = 25 mv/r s = 25 mv/115.54 Ohms = 0.216 ma V x is equal to I multiplied by the total resistance: V x = I(R s +R f ) = 2.18 V If the actual resistances were the nominal values, the 25 mv range would not be exceeded with V x = 2.18 V. To allow for the tolerances in the actual resistances and to leave a little room for higher temperatures, set V x equal to 2.1 volts. 5.2 Calibrating a PRT The greatest source of error in a PRT is likely to be that the resistance at 0 C deviates from the nominal value. Calibrating the PRT in an ice bath can correct this offset and any offset in the fixed resistor in the Terminal Input Module. With the PRT at 0 C, R s = R 0. Thus, the above result becomes R 0 /R f, the reciprocal of the multiplier required to calculate temperature, R f /R 0. By making a measurement with the PRT in an ice bath, errors in both R s and R 0 can be accounted for. To perform the calibration, connect the PRT to the datalogger and program the datalogger to measure the PRT with the 3-wire half bridge as shown in the example section. For a 100 Ohm PRT use a multiplier of 100; for a 1000 Ohm PRT use a multiplier of 10. Place the PRT in an ice bath (@ 0 C; R s = R 0 ). Read the result of the bridge measurement. The reading is R s /R f, which is equal to R 0 /R f since R s = R 0. The correct value of the multiplier, R f /R 0, is the multiplier used divided by this reading. For example, if, with a 100 Ohm PRT, the initial reading is 0.9890, the correct multiplier is: R f /R 0 = 100/0.9890 = 101.11. 5.3 Compensation for Wire Resistance The 3-wire half bridge compensates for lead wire resistance by assuming that the resistance of wire A is the same as the resistance of wire B (Figure 3-1). The maximum difference expected in wire resistance is 2%, but is more likely to be on the order of 1%. The resistance of R s calculated with Instruction 7, is actually R s plus the difference in resistance of wires A and B. For example, assume that a 100 Ohm PRT is separated from the datalogger by 500 feet of 22 awg wires. The average resistance of 22 AWG wire is 16.5 Ohms per 1000 feet, which would give each 500 foot lead wire a nominal resistance of 8.3 Ohms. Two percent of 8.3 Ohms is 0.17 Ohms. Assuming that the greater resistance is in wire B, the resistance measured for the PRT (R 0 = 100 Ohms) in the ice bath would be 100.17 Ohms, and the resistance at 40 C would be 115.71. The measured ratio R s /R 0 is 1.1551; the actual ratio is 115.54/100 = 1.1554. The temperature computed by Instruction 16 from the measured ratio would be about 0.1 C lower than the actual temperature of the PRT. This source of error does not exist in a 4-wire half bridge where a differential measurement is used to directly measure the voltage across the PRT. 6

CAMPBELL SCIENTIFIC COMPANIES Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com info@campbellsci.com Campbell Scientific Africa Pty. Ltd. (CSAf) PO Box 2450 Somerset West 7129 SOUTH AFRICA www.csafrica.co.za sales@csafrica.co.za Campbell Scientific Australia Pty. Ltd. (CSA) PO Box 444 Thuringowa Central QLD 4812 AUSTRALIA www.campbellsci.com.au info@campbellsci.com.au Campbell Scientific do Brazil Ltda. (CSB) Rua Luisa Crapsi Orsi, 15 Butantã CEP: 005543-000 São Paulo SP BRAZIL www.campbellsci.com.br suporte@campbellsci.com.br Campbell Scientific Canada Corp. (CSC) 11564-149th Street NW Edmonton, Alberta T5M 1W7 CANADA www.campbellsci.ca dataloggers@campbellsci.ca Campbell Scientific Ltd. (CSL) Campbell Park 80 Hathern Road Shepshed, Loughborough LE12 9GX UNITED KINGDOM www.campbellsci.co.uk sales@campbellsci.co.uk Campbell Scientific Ltd. (France) Miniparc du Verger - Bat. H 1, rue de Terre Neuve - Les Ulis 91967 COURTABOEUF CEDEX FRANCE www.campbellsci.fr info@campbellsci.fr Campbell Scientific Spain, S. L. Psg. Font 14, local 8 08013 Barcelona SPAIN www.campbellsci.es info@campbellsci.es Campbell Scientific Ltd. (Germany) Fahrenheitstrasse13, D-28359 Bremen GERMANY www.campbellsci.de info@campbellsci.de Please visit www.campbellsci.com to obtain contact information for your local US or International representative.