CALIBRATION REPORT FOR THERMOHYDROMETER

Similar documents
2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

CALIBRATION CERTIFICATE # 503

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

Certificate of Accreditation to ISO/IEC 17025:2005

Technical Procedure for General Laboratory Equipment

CERTIFICATE OF CALIBRATION

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

The Fundamentals of Moisture Calibration

Uncertainty Propagation for Force Calibration Systems

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05

Calibration Traceability Guidelines

CERTIFICATE OF CALIBRATION

Document No: TR 12 Issue No: 1

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium,

Calibration of Temperature Block Calibrators

Metallic materials Brinell hardness test. Part 3: Calibration of reference blocks

ANALYSIS CERTIFICATE OF

New Thermometer Guidance Document from AASHTO Accreditation Program. Maria Knake Asphalt Binder ETG Meeting May 10, 2018

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

Process Accuracy Ratio vs Process Uncertainty Ratio (Risk Mitigation: Calibration and the Customer s Process)

Certificate of Accreditation

Accuracy Calibration Certificate

g/cm³ DAkkS Calibration Laboratory for Temperature and Density High-precision calibration for temperature and density

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

A Unified Approach to Uncertainty for Quality Improvement

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Certificate of Accreditation

Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018

MEASUREMENT TRACEABILITY AND ERRORS RELATED TO FORCE MEASUREMENT

Uncertainty of Measurement A Concept

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

Withdrawn at December 31, Guideline DKD-R 5-7. Calibration of Climatic Chambers DEUTSCHER KALIBRIERDIENST

R SANAS Page 1 of 7

Traceable UV-Vis Reference Materials

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SAMPLE. Accuracy Calibration Certificate. Mettler Toledo 1900 Polaris Parkway Contact: Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC WI v.1.

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities

Metallic materials Brinell hardness test. Part 2: Verification and calibration of testing machines

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

IMPORTANT FEATURES OF ESTIMATION & INTERPRETATION OF MEASUREMENT UNCERTAINTY. B. Bhattacharya, K. Ramani, P.H. Bhave

Part 5: Total stations

Technical Procedure for Measurement Assurance

INTER-COMPARISONON TEMPERATURE MEASUREMENT USING FLUKE THERMOMETER MODEL 1502A

Determination of particle size distribution Single particle light interaction methods. Part 4:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

ANALYSIS CERTIFICATE OF 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar

STATUS OF THE WIGOS DEMONSTRATION PROJECTS

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Certificate of Accreditation

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. ALLOMETRICS, INC Bayport Blvd. Seabrook, TX Terry Baldwin Phone: CALIBRATION

Infrared Thermometer Calibration A Complete Guide

OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

An Introduction to the Differences Between the Two Most Recognized Force Standards

Certificate of Accreditation

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

INDIANA DEPARTMENT OF TRANSPORTATION OFFICE OF MATERIALS MANAGEMENT. VERIFYING THERMOMETERS ITM No T

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Test Report EN (2008)

CERTIFICATE OF. Single Analyte Custom Grade Solution g/ml (measured at 20 ± 4 ºC) Graviametric NIST SRM Lot Number: See Sec. 4.

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3163 Lot Number:

MOY/SCMI/36 SPECIFICATION OF ACCURACY FOR A PRECISION CLINOMETER

Calibration and verification: Two procedures having comparable objectives and results

REVISIONS SYMBOL DESCRIPTION DATE APPROVAL

This document is a preview generated by EVS

CERTIFICATE OF CALIBRATION

Certificate of Accreditation

Certificate of Accreditation

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3165 Lot Number: EDTA NIST SRM 928 Lot Number: 928

PRODUCT GUIDE PRECISION INSTRUMENTS FOR Temperature, Density, Time, Weather and Calibration

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

SCOPE OF ACCREDITATION

Certificate of Accreditation

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT#

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3103a Lot Number:

Spa & Wellness for Pipettes

1551A Ex/1552A Ex Intrinsically Safe Stik Thermometer. The new gold standard of industrial temperature calibration

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

CERTIFICATE OF. Single Analyte Mass Spec Solution ± 0.1 µg/ml g/ml (measured at 20 ± 4 ºC)

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

CERTIFICATE OF ACCREDITATION

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

A case study on how to maintain confidence of thermal properties test: Thermal conductivity of building insulation materials

Transcription:

CALIBRATION REPORT FOR THERMOHYDROMETER SAMPLE CUSTOMER S/N: XXXXX Report Number: Y205188 Page 1 of 5 Report No. Y205188

ICL CALIBRATION LABORATORIES, INC. ISO/IEC 17025 and ANSI/NCSL Z540-1 accredited The specialists in ASTM and laboratory thermometers & hydrometers Members: A2LA ASTM API NCSLI ASQ NCWM Setting new standards in calibration excellence! Cert 526.01 Calibration 1501 SE Decker Avenue, Suite 118, Stuart, FL 34994 USA Tel: 772-286-7710 1-800-713-6647 Fax: 772-286-8737 email: sales@iclcalibration.com Internet: www.icllabs.com CALIBRATION REPORT FOR THERMOHYDROMETER The instrument or device identified below was examined and calibrated in ICL's metrology laboratory, using NIST traceable standards, following the calibration procedure referenced below. This calibration fulfills the requirements of ISO/IEC 17025-2005, 'General Requirements for the Competence of Testing and Calibration Laboratories' and ANSI/NCSL Z540-1-1994, 'Calibration Laboratories and Measuring and Test Equipment - 'General Requirements'. CLIENT SAMPLE CUSTOMER STREET ADDRESS CITY, STATE ZIP CODE Purchase order number: NOT AVAILABLE Submitted by: SAMPLE COMPANY ICL internal reference (SO): 444444 DATES Date report issued: 03-13-2015 Next due (specified by client): March 13, 2016 UUT (Unit Under Test) INFORMATION Model number: 40054HL-C THERMOHYDROMETER ASTM 54HL Manufacturer, brand or mark: LSW Serial No: XXXXX HYDROMETER PORTION OF INSTRUMENT Engineering units: API Gravity for petroleum products Range: 29/41 API Scale divisions: 0.10 API Accuracy tolerance (maximum scale error per ASTM E100 or E2995, as appropriate): 0.10 API THERMOMETER PORTION OF INSTRUMENT Engineering units: F Scale range: 0 to 150 F Scale divisions: 2.0 F Immersion: TOTAL Accuracy tolerance (maximum scale error per ASTM E100 or E2995, as appropriate): +/- 1.0 RESULTS OF PHYSICAL EXAMINATION This thermohydrometer was examined under a polariscope, and strains in the glass, if any, were judged to be minimal and of no detriment to the function of the instrument. The instrument was then examined under magnification, looking for cracks, fissures, scratches, irregularities or other defects in the glass which might adversely affect its function. None of concern were observed, unless indicated below. The capillary of the integral thermometer was examined under magnification. No foreign material, moisture, oxidation, or other evidence of contamination could be seen. No discernable capillary irregularities were noted. There was no loose ballast or other material observed within the hydrometer. A scale slippage indicator of a type required by ASTM E100 and ASTM E2995 was present and correctly positioned, unless noted otherwise below. It was determined that this instrument is in apparent good working order and is therefore suitable for calibration. CALIBRATION PROCEDURES Hydrometer portion: ICL Procedure 02, which is based on ASTM E126. Thermometer portion: ICL Procedure 01, which is based on ASTM E77 LABORATORY ENVIRONMENTAL CONDITIONS Temperature: 23 C +/- 5 C, Relative humidity: between 30% and 80% Page 2 of 5 Report No. Y205188

RESULTS OF CALIBRATION FOR HYDROMETER NOTE: The indications of this instrument cannot be adjusted or modified by ordinary means; accordingly, the readings given in the table below should be considered, in effect, to be both 'As Found' and 'As Left' readings. Nominal UUT reading Correction Tolerance Accept Limit* P/F/Ind Uncertainty 30.00 API 29.97 API 0.03 API ± 0.10 API ± 0.093 API Pass ± 0.037 API 35.00 API 35.00 API 0.00 API ± 0.10 API ± 0.093 API Pass ± 0.038 API 40.00 API 39.94 API 0.06 API ± 0.10 API ± 0.092 API Pass ± 0.039 API The test points listed in the above table satisfy the requirements of ASTM E126 (current revision). The scale reading is determined by the intersection of the horizontal plane, tangent to the bottom of the meniscus, on the stem scale. See ASTM D1298 for drawings, discussion, and instruction on making accurate hydrometer readings. The readings of this hydrometer were made under magnification, estimated and resolved to one-tenth of one scale division. The readings (and corrections) presented above represent the average of (a minimum of) three independent observations of the UUT and at least six observations of the reference hydrometer at each test point. GUARD BANDING ISO/IEC 17025:2005(E) requires, in Section 5.10.4.2., that, When statements of compliance are made, the uncertainty of measurement shall be taken into account.'' One valid way of complying with this requirement is applying a 'guard band' to the device's tolerance. The guard band is calculated as a function of the test uncertainty ratio (TUR), the ratio of the tolerance of the UUT to the measurement uncertainty. Basically, the smaller the uncertainty is relative to the tolerance, the smaller the guard band. A TUR of 5:1 or even 4:1 results in a guard band of zero, or very close to zero. A 3:1 TUR results in a modest guard band. As TUR declines, the guard band becomes larger. The use of the guard band in the decision process is designed to reduce the probability of a false acceptance (PFA), or a false failure, to 2% or less. The method and equations we use for calculation of the guard band comply with the requirements of ANSI/NCSL Z540.3 The *Accept Limit(s) are calculated by subtracting the guard band from the tolerance. The Accept Limit is essentially a new tolerance, for this calibration only, which we use to make a declaration of Pass, Fail, or Indeterminate, as explained below: Pass The measured value falls within the interval described by the test point plus or minus the Accept Limit. Fail The measured value falls outside the interval described by the test point plus or minus (the tolerance + the guard band). Ind (Indeterminate) The measured value is indeterminate, falling in that grey area too close to permit a credible determination; it is statistically and metrologically imprudent to declare that the instrument is definitively either 'in-tolerance' or 'out-of-tolerance'. LIMITATIONS OF USE This is a full range calibration. No limitations of use are imposed on this instrument. MEASUREMENT UNCERTAINTY The measurement uncertainty reported is the expanded uncertainty at 2 sigma (k=2), to provide a confidence level of approximately 95%. The uncertainty is calculated considering both Type A and Type B contributors. Type A contributors include the precision and repeatability of the comparison process obtained from process statistics, and the standard deviation of repeated measurements made during the performance of the calibration. Type B contributors include the uncertainty of the calibration of the NIST standard, the uncertainty of the calibration of the transfer standard, the resolution of the standard, and the resolution of the UUT, among other contributors. The Type A and B contributors are combined using the rootsum-square method to obtain the standard uncertainty at 1 sigma. The standard uncertainty is then multiplied by 2 to obtain the expanded uncertainty at 2 sigma (k=2). This uncertainty calculation is consistent with the requirements of the ISO Guide to the Expression of Uncertainty in Measurement (the 'GUM') and NIST Technical Note 1297. The expanded uncertainties (k=2) reported here do not contain estimates for (1) any effects that may be introduced by transportation of the instrument between ICL and the user's facility, (2) drift of the instrument, (3) hysteresis of the instrument, or (4) any measurement uncertainties introduced by the user. TRACEABILITY INFORMATION This calibration is traceable to NIST through an unbroken chain of comparisons. The NIST reference standards are used only for calibration of the transfer standard hydrometers, which in turn are used to calibrate the client's hydrometer(s). Measurement uncertainty has been calculated at each step in the chain and fully documented. The NIST primary reference hydrometer(s) and transfer standard hydrometers(s) utilized in the performance of this calibration are listed below. Transfer standard hydrometers are calibrated annually against the NIST reference hydrometers. Test point NIST reference Transfer standard MTE# Manufacturer Next Due 30.00 API NIST 13506008 & 117269 ASTM 4H 12486737 364 LSW 09/16/15 35.00 API NIST 13506008 & 117269 ASTM 4H 12486737 364 LSW 09/16/15 40.00 API NIST 13506008 & 117269 ASTM 4H 12486737 364 LSW 09/16/15 (End of hydrometer section. Thermometer information appears on the following page.) Page 3 of 5 Report No. Y205188

RESULTS OF CALIBRATION FOR THERMOMETER NOTE: The indications of this instrument cannot be adjusted or modified by ordinary means; accordingly, the readings given in the table below should be considered, in effect, to be both 'As Found' and 'As Left' readings. Nominal UUT reading Correction Tolerance Accept Limit* P/F/Ind Uncertainty 32.00 F 32.0 F 0.0 F ± 1.0 F ± 0.88 F Pass ± 0.46 F 100.00 F 100.0 F 0.0 F ± 1.0 F ± 0.88 F Pass ± 0.46 F The above readings were made under magnification and resolved to one-tenth of one scale division. Unless otherwise stated, the thermometer was permitted to stabilize for a minimum of 5 minutes at each test temperature prior to reading. GUARD BANDING ISO/IEC 17025:2005(E) requires, in Section 5.10.4.2., that, When statements of compliance are made, the uncertainty of measurement shall be taken into account.'' One valid way of complying with this requirement is applying a 'guard band' to the device's tolerance. The guard band is calculated as a function of the test uncertainty ratio (TUR), the ratio of the tolerance of the UUT to the measurement uncertainty. Basically, the smaller the uncertainty is relative to the tolerance, the smaller the guard band. A TUR of 5:1 or even 4:1 results in a guard band of zero, or very close to zero. A 3:1 TUR results in a modest guard band. As TUR declines, the guard band becomes larger. The use of the guard band in the decision process is designed to reduce the probability of a false acceptance (PFA), or a false failure, to 2% or less. The method and equations we use for calculation of the guard band comply with the requirements of ANSI/NCSL Z540.3 The *Accept Limit(s) are calculated by subtracting the guard band from the tolerance. The Accept Limit is essentially a new tolerance, for this calibration only, which we use to make a declaration of Pass, Fail, or Indeterminate, as explained below: Pass The measured value falls within the interval described by the test point plus or minus the Accept Limit. Fail The measured value falls outside the interval described by the test point plus or minus (the tolerance + the guard band). Ind (Indeterminate) The measured value is indeterminate, falling in that grey area too close to permit a credible determination; it is statistically and metrologically imprudent to declare that the instrument is definitively either 'in-tolerance' or 'out-of-tolerance'. LIMITATIONS OF USE This is a limited, or partial-range calibration, and accordingly, this instrument may be used with confidence only within the range bracketed by the test points. MEASUREMENT UNCERTAINTY The measurement uncertainty reported is the expanded uncertainty at 2 sigma (k=2), to provide a confidence level of approximately 95%. The uncertainty is calculated considering both Type A and Type B contributors. Type A contributors include the standard deviation of the measurement process from check standard control charts, the standard deviation of monthly Triple Point of Water calibrations of the standard, and UUT variability observed during the calibration. Type B contributors include comparator uniformity, uncertainty of the calibration of the reference standard, stem conduction and other immersion effects, the sensitivity and accuracy of the reference standard thermometer's readout, resolution of the reference standard and resolution of the UUT. The Type A and B contributors are combined using the root-sum-square method to obtain the standard uncertainty at 1 sigma. The standard uncertainty is then multiplied by 2 to obtain the expanded uncertainty at 2 sigma (k=2). This uncertainty calculation is consistent with the requirements of the ISO Guide to the Expression of Uncertainty in Measurement (the 'GUM') and NIST Technical Note 1297. The expanded uncertainties (k=2) reported here do not contain estimates for (1) any effects that may be introduced by transportation of the instrument between ICL and the user's facility, (2) drift of the instrument, (3) hysteresis of the instrument, or (4) any measurement uncertainties introduced by the user. TRACEABILITY INFORMATION This calibration is traceable to NIST through an unbroken chain of comparisons. Our primary reference standard, a NIST calibrated SPRT, is used only to calibrate our transfer standards, which in turn are used to calibrate our clients' instruments. Measurement uncertainty has been calculated at each step in the chain and is fully documented. ICL maintains three Rosemount model 162CE 25.5 Ohm SPRTs, for redundancy and to permit sequential rotation to NIST for calibration. As of this date, traceability is conveyed through S/N 5058, our MTE-262, calibrated by NIST on August 17, 2012. NIST GMP-11 recommends a 36 month calibration cycle for SPRTs. PRT transfer standards are calibrated annually against this SPRT, per NIST GMP-11 recommendations, and are monitored continually using measurement assurance strategies including check standards, control charts and monthly triple point of water checks. The comparators and transfer standards used in the performance of this calibration are indicated below, organized by test point. Test point Comparator MTE# Manufacturer Transfer standard MTE# Manufacturer Next Due 32.00 F 9510 Glycol bath 002 PolyScience 5614 PRT 597010 135 Hart Scientific 09/06/15 100.00 F 7310 Water bath 006 PolyScience 5614 PRT 576776 130 Hart Scientific 09/06/15 (End of thermometer section) Page 4 of 5 Report No. Y205188

NOTES AND SUPPLEMENTAL INFORMATION All temperatures given in this report are those defined by the International Temperature Scale of 1990 (ITS-90). For additional information on hydrometers and usage see ASTM E100, ASTM E126, ASTM D1298 and ASTM D6822. TECHNICIANS Hydrometer portion: ROB FOX Thermometer portion: DEBORAH M. WEBER ICL CALIBRATION LABORATORIES, INC. An ISO/IEC 17025 & ANSI/NCSL Z-540-1 accredited laboratory - American Association for Laboratory Accreditation Certificate #526.01 Approved by: Reviewed by: J. Jeff Kelly, Technical Director Deborah M. Weber, Quality Deputy This report document was prepared by Lori J. Parr Date report issued: 03-13-2015 Recalibration date specified by client: March 13, 2016 Caution: Users should be aware that both density and temperature indications of thermohydrometers may change with rough handling, shock, exposure to aggressive liquids, and thermal cycling, among other factors. Consequently, test results and performance obtained at time of calibration may not necessarily apply throughout an extended period of use. Periodic recalibration of this instrument, in accordance with procedures set forth in ASTM E126 (for the hydrometer), and ASTM E77 (for the thermometer), is recommended. This calibration report may not be reproduced except in full without the express written permission of ICL Calibration Laboratories, Inc. This report applies only to the item calibrated. This calibration report shall not be used to claim product endorsement by the A2LA. End of report. Page 5 of 5 Report No. Y205188