arxiv: v1 [nucl-ex] 16 Feb 2012

Similar documents
Radioactivity measurements for the ERMES project at the STELLA facility

Dark matter search with the SABRE experiment

Direct measurement of the 2H(α,γ)6Li cross section at energies of astrophysical interest

Cosmogenic background for the GERDA experiment. Luciano Pandola INFN, Laboratori del Gran Sasso, Italy

arxiv: v1 [nucl-ex] 20 Dec 2008

Determination of the activity of radionuclides

Sodium-iodide with Active Background REjection. Irene Bolognino

Underground nuclear astrophysics and the Sun

Hands on LUNA: Detector Simulations with Geant4

Status of the CUORE experiment at Gran Sasso

The 46g BGO bolometer

THE CRYOGENIC UNDERGROUND OBSERVATORY FOR RARE EVENTS: STATUS AND PROSPECTS

High Purity Germanium Detector Calibration at ISOLDE

Esperimenti bolometrici al Gran Sasso: CUORE e CRESST

Evidence of Solar Influences on Nuclear Decay Rates

Direct Dark Matter and Axion Detection with CUORE

HEROICA: a test facility for the characterization of BEGe detectors for the GERDA experiment

LOW ENERGY SOLAR NEUTRINOS WITH BOREXINO. Lea Di Noto on behalf of the Borexino collaboration

arxiv: v1 [physics.ins-det] 22 Dec 2016

Testing Dynamical Reduction Models at the Gran Sasso underground laboratory

Periods Detected During Analysis of Radioactivity Measurements Data

Two Neutrino Double Beta (2νββ) Decays into Excited States

Testing Dynamical Reduction Models at the underground Gran Sasso Laboratory

Direct Dark Matter and Axion Detection with CUORE

arxiv: v1 [physics.ins-det] 4 Nov 2017

Status of the CUORE and CUORE-0 experiments at Gran Sasso

Background Free Search for 0 Decay of 76Ge with GERDA

The Gerda Experiment for the Search of Neutrinoless Double Beta Decay

Studies of the XENON100 Electromagnetic Background

arxiv: v1 [physics.ins-det] 3 Feb 2011

Nuclear astrophysics at Gran Sasso Laboratory: LUNA experiment

BUREAU INTERNATIONAL DES POIDS ET MESURES

College Station, TX 77843, USA *Corresponding author. Tel ; fax address:

Double Beta Present Activities in Europe

1/1. Concerning the variability of beta-decay measurements. P.A. Sturrock a,*, E. Fischbach b, A. Parkhomov c, J.D. Scargle d, G.

DarkSide-50: performance and results from the first atmospheric argon run

PoS(ICHEP2016)474. SoLid: Search for Oscillations with a Lithium-6 Detector at the SCK CEN BR2 reactor

Disproof of solar influence on the decay rates of 90 Sr/ 90 Y

Status of CUORE and Results from CUORICINO. SERGIO DI DOMIZIO UNIVERSITÀ & INFN GENOVA On behalf of the CUORE Collaboration

CRaTER Science Requirements

arxiv: v1 [physics.ins-det] 26 May 2015

Recent results and status of the

Characterization and Monte Carlo simulations for a CLYC detector

Search for Possible Solar Influences in Ra-226 Decays

Low Background NAA: A powerful tool for Physics of Rare Experiments and Environmental Sciences

Light Dark Matter and XENON100. For the XENON100 Collaboration Rafael F. Lang Columbia University

COUNT RATE EFFECTS ON PEAK SHAPE AND THROUGHPUT IN DIGITAL AND ANALOG GAMMA RAY SPECTROMETRY

Cosmogenic and primordial radioisotopes in copper bricks shortly exposed to cosmic rays

Results on geoneutrinos at Borexino experiment. Heavy Quarks and Leptons Yamagata Davide Basilico

E.Fiorini, Neutrino 2004 Paris, June 17, For searches on neutrinoless ββ decay, WIMPs and axions interactions and on rare nuclear events

L esperimento GERDA. GERDA - gruppo INFN. L Aquila, Congresso SIF, 30/09/2011. OUTLINE: GERDA Motivations GERDA Status GERDA Future plans

Comparison of the Photo-peak Efficiencies between the Experimental Data of 137 Cs Radioactive Source with Monte Carlo (MC) Simulation Data

Results from 730 kg days of the CRESST-II Dark Matter Search

Technical Specifications and Requirements on Direct detection for Dark Matter Searches

Sensitivity of sodium iodide cryogenic scintillation-phonon detectors to WIMP signals

Hands on DarkSide-50: Low Energy Calibration

Status of Cuore experiment and last results from Cuoricino

Laboratori Nazionali del GranSasso Alfredo G. Cocco Istituto Nazionale di Fisica Nucleare Sezione di Napoli

anti-compton BGO detector

Study well-shaped germanium detectors for lowbackground

Nuclear Astrophysics Underground Status & Future

Status of the CUORE and CUORE-0 experiments at Gran Sasso

arxiv:nucl-ex/ v2 21 Jul 2005

EFFICIENCY AND RESOLUTION OF GERMANIUM DETECTORS AS A FUNCTION OF ENERGY AND INCIDENT GEOMETRY

First results from DAMA/LIBRA and the combined results with DAMA/NaI

Introduction to Environmental Measurement Techniques Radioactivity. Dana Pittauer 1of 48

Results from Borexino 26th Rencontres de Blois

Background optimization for a new spherical gas detector for very light WIMP detection

ACCURATE MEASUREMENTS OF SURFACE EMISSION RATE FOR LARGE-AREA ALPHA AND BETA REFERENCE SOURCES

Design, Construction, Operation, and Simulation of a Radioactivity Assay Chamber

Publications of Francesco Arneodo: journal articles

1220 QUANTULUS The Ultra Low Level Liquid Scintillation Spectrometer

Gamma background measurements in the Gran Sasso National Laboratory

The GERmanium Detector Array

Latest results from LUNA

NUCL 3000/5030 Laboratory 2 Fall 2013

Educational cosmic ray experiments with Geiger counters

Determination of the shielding power of different materials against gamma radiation

E15. Background suppression in GERDA Phase II. and its study in the LARGE low background set-up

Hands on CUORE: investigation on the vibrations and temperature control of the cryostat

Contents. General Introduction Low background activities Pixel activities Summary and Outlook. Kai Zuber

Investigation of pulse shapes and time constants for NaI scintillation pulses produced by low energy electrons from beta decay

Status of the GERDA experiment

arxiv: v1 [nucl-ex] 4 Feb 2009

Direct measurement of the 17 O(p,α) 14 N reaction at energies of astrophysical interest at LUNA

A Trial of Neutrino Detection from Joyo Fast Research Reactor

AGATA preamplifier performance on large signals from a 241 Am+Be source. F. Zocca, A. Pullia, D. Bazzacco, G. Pascovici

GERDA experiment A search for neutrinoless double beta decay. Roberto Santorelli (Physik-Institut der Universität Zürich)

Radon-Thoron mixed atmosphere: realization, characterization, monitoring and use for detector calibration.

F. Cappella INFN - LNGS

Background simulations and shielding calculations

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

On the HalfiLife of LABORATORY ERNEST ORLANDO LAWRENCE BERKELEYNATIONAL. and Particle Astrophysics Nuclear Science and Physics Divisions

Nuclear Recoil Scintillation and Ionization Yields in Liquid Xenon

DETECTORS. I. Charged Particle Detectors

Measurement of nuclear recoil responses of NaI(Tl) crystal for dark matter search

Experimental study of the 14 N(p,γ) 15 O reaction

Detection and measurement of gamma-radiation by gammaspectroscopy

New results of CUORICINO on the way to CUORE

arxiv: v1 [nucl-ex] 7 Jun 2011

Transcription:

Search for time dependence of the 137 Cs decay constant E. Bellotti a, C. Broggini b, G. Di Carlo c, M. Laubenstein c, R. Menegazzo b a Università degli Studi di Milano Bicocca and Istituto Nazionale di Fisica Nucleare, Sezione di Milano, Milano, Italy b Istituto Nazionale di Fisica Nucleare, Sezione di Padova, Padova, Italy c Istituto Nazionale di Fisica Nucleare, Laboratori Nazionali del Gran Sasso, Assergi (AQ), Italy arxiv:1202.3662v1 [nucl-ex] 16 Feb 2012 Abstract Starting from June 2011, the activity of a 137 Cs source has been measured by means of a HPGe detector installed deep underground in the Gran Sasso Laboratory. In total about 5100 energy spectra, one hour measuring time each, have been collected. These data allowed the search for time variations of the decay constant with periods from a few hours to 1 year. No signal with amplitude larger than 9.6 10 5 at 95% C.L. has been detected. These limits are more than one order of magnitude lower than the values on the oscillation amplitude reported in literature. In particular, for 1 year period an oscillation amplitude larger than 8.5 10 5 has been excluded at 95% C.L., independently of the phase. The same data give a value of 29.96±0.08 years for the 137 Cs half life, in good agreement with the world mean value of 30.05±0.08 years. Keywords: Radioactivity, Beta Decay, Gran Sasso 1. Introduction The possible time dependence of the radioactive nuclei decay constant has been investigated since the infancy of the science of radioactivity [1]. Recently, the interest in the time dependence has strongly increased [2] since various experiments have reported evidence of such an effect. In particular, in [3] the constancy of the activity of a 226 Ra source has been measured with an ionization chamber. An annual modulation of amplitude 0.15%, having the maximum in February and the minimum in August, has been observed. The activity of a 152 Eu source has also been measured by means of a Ge(Li) detector and an even larger annual modulation (0.5%) has been detected in the counting rate at the full energy peak of the 1408 kev line. Alburger et al. [4] measured the half-life of 32 Si, which is interesting for many applications in Earth Science. Data, collected in four years, show an evident annual modulation, with an amplitude of about 0.1%. In [2, 5] the existence of new and unknown particle interaction has been advocated to explain the yearly variation in the activity of radioactive sources [3, 4]. The Authors correlate these variations, whose amplitude is about 10 3, the maximum being around February and the minimum in August, to the Sun-Earth distance. The paper by Jenkins and Collaborators [2] triggered a strong interest in the subject. Old and recent data have been analyzed, or reanalyzed, to search for periodic and sporadic variations with time. Some of the measurements and analysis confirm the existence of oscillations [6, 7] whereas others contradict this hypothesis [8, 9, 10]. For instance, [6] presents the time dependence of the counting rate for 60 Co, 90 Sr and 90 Y sources, measured with Geiger Müller detectors and for a 239 Pu source, measured with silicon detectors. While beta sources show an annual (and monthly) variation with amplitude of about 0.3%, the count rate from the Pu source is fairly constant. More recently, an anti-correlation between decay rate and intensity of solar flares has been suggested. As a matter of fact, in [11] the decay rate of a 54 Mn source has been measured and a significant dip in the count rate in coincidence with solar flares has been observed. According to the Authors, this suggests that nuclear decay rates may be affected by solar activity. However, in [12] no anti-correlation is found between decay rate and intensity of solar flares. As also suggested by many authors, we believe that more dedicated experiments are needed to clarify if the observed effects are physical or are due to systematic Preprint submitted to Elsevier February 17, 2012

effects not taken into account. Therefore, we performed a dedicated experiment using a 137 Cs source. Time dependence of the order of 0.2% has already been reported [13] in the decay constant of 137 Cs measured with a germanium detector. The special feature of our experiment is the set-up installed deep underground in the INFN Gran Sasso Underground Laboratory. The laboratory conditions are very favorable: the cosmic ray flux is reduced by a factor 10 6 and the neutron flux by a factor 10 3 with respect to above ground. As a consequence, we do not have to take into account possible time variations of these fluxes since their contribution to the counting rate is completely negligible. Moreover, the laboratory temperature is naturally constant, maximum variation being of a few degrees Celsius during the year. Ventilated NIM Crate HV & LV Input Amplifier PUR Busy Output PT100 137Cs source HPGe Detector PT100 EasyMCA Pressure Gauge USB Computer USB Agilent 34970A Isolation Transformer 2. The measurement The aim of the present experiment is the measurement of a 137 Cs source activity as a function of time by the detection and the counting of the emitted gamma rays. 2.1. The set-up The set-up (Fig. 1) is installed in the low background facility STELLA (SubTErranean Low Level Assay) located in the underground laboratories of LNGS. The source is a 137 Cs standard source with Cs embedded in a plastic disk of 1 diameter and 1/8 thick. The source nominal activity was 4.14 kbq (July 16 th, 1998). At the beginning of the measurement (June 6 th, 2011) its activity has been estimated to be 3.0 kbq. The detector is a p-type High Purity Germanium manufactured by Ortec and powered at the nominal bias of 3500 V. The crystal has 78.4 mm diameter and is 84.6 mm high with a relative efficiency of 96%. The source is firmly fixed to the copper end-cap of the germanium detector in order to minimize variations in the sourcedetector relative positions. As a matter of fact, from Monte Carlo simulations we have that a variation of 1 micron in the source-detector distance would cause a variation of 5 10 5 in the counting efficiency. The Germanium detector is surrounded by at least 5 cm of copper followed by 25 cm of lead to suppress the laboratory gamma ray background. Finally, shielding and detector are housed in a polymethylmetacrilate box flushed with nitrogen at slight overpressure and which is working as an anti-radon shield. The signal from the detector pre-amplifier goes first to an Ortec amplifier (mod.120-5f) where it is shaped 2 Anti Radon Box N2 flushed Figure 1: Layout of the setup. The whole system is powered through an isolation transformer. with 6 µs shaping time, and then to a Multi Channel Analyser (Easy-MCA 8k Ortec, 0.21 kev/channel). Also the busy and pile-up signals from the amplifier are sent to the MCA. The pile-up rejection circuit of the amplifier is able to recognize two signals if they are separated in time by at least 0.5 µs. In order to minimize the noise the whole set-up, detector and electronics modules, is powered through a 3 kva AC-AC isolation transformer. Finally, temperature and atmospheric pressure are the two important parameters we have to monitor since they could change the detector efficiency by varying the relative distance Cs source-detector. In particular, the temperature is continuously measured with a Pt100 sensor located in the shielding very close to the Cs source, whereas the atmospheric pressure is measured in a nearby hall of the underground laboratory. The temperature at the position of the electronics modules is also continuously monitored since it can affect the electronics performance. 2.2. Data taking Spectra from the MCA are collected every hour (real time provided by the MCA), except during the liquid nitrogen refilling of the detector. These interruptions occur two times a week and last less than 2 hours. The quality of the measurement is monitored by checking

Rate [counts/(h ch)] 100000 10000 1000 100 10 1 0.1 0.01 137 Cs 137 60 Cs(Σ) Co 60 Co 1200 1300 1400 152 Eu 40 K 0.001 0 200 400 600 800 1000 1200 1400 1600 E γ [kev] Figure 2: Measured γ-ray spectrum of the 137 Cs source after 5124 hours. A weak activity from 60 Co, 152 Eu and 40 K contaminants is observed at high energy (see inset). The 40 K line is also evident in the background spectrum. The peak at 1.323 MeV results from the accidental coincidence of two 661.6 kev γ-lines when two 137 Cs decays occur within the pulse pair resolution of the spectroscopy amplifier. the energy resolution at the Cs line (661.6 kev). Its value, 1.79±0.02 kev, is stable, this way proving that the electronic noise does not change with time. This is also confirmed by the counting rate in the lowest MCA channels, below 7 kev, where the noise is dominant and whose rate remains constant with time. Rate [counts/(h kev)] 5 4 3 2 1 0-1 -2 260 270 280 290 300 E γ [kev] Figure 3: Weak branching in the 137 Cs spectrum from the decay of the (1/2) + level in 137 Ba. The spectrum is obtained after a background subtraction. The background component has been evaluated from a polynomial fit of regions on both sides of the peak. Statistical uncertainties are indicated. 3 The position of the Cs peak is also constant, within 0.21 kev, proving the stability of the electronics. Finally, we monitored the dead time, provided by the MCA. It smoothly changed from the initial value of 5.10% to 4.98% as a consequence of the decreased activity of the Cs source. 2.3. Energy spectrum The spectrum measured with the present set-up is shown in Fig. 2. Modest signals from 40 K, 60 Co and 152 Eu are visible in the energy spectrum above 1 MeV. The 40 K line is also present in the background spectrum, while the 60 Co and 152 Eu activities are related to the source. Their contribution to the total count rate, estimated by Monte Carlo simulation, is 1.7 10 2 Hz for 60 Co and 2.7 10 3 for 152 Eu and 40 K. The total count rate above the threshold of 7 kev is of about 700 Hz. The intrinsic background, i.e shielded detector without Cs source, has been measured during a period of 70 days. Thanks to the underground environment and to the detector shielding, it is very low, down to about 40 counts/hour above the 7 kev threshold (0.01 Hz). The signature of the 137 Cs source is given by the outstanding peak at 661.6 kev, by the peak at 1.323 MeV due to the sum of two 661.6 kev gamma rays too close in time to be recognized by the pile-up circuit and, finally, by the modest peak at 283 kev (Fig. 3). This last peak is due to the (1/2) + level of 137 Ba populated by the beta decay of 137 Cs with a branching ratio of 5.3(14) 10 6 [14]. 3. Data analysis Data discussed in this letter have been continuously collected during 217 days, starting from June 6th 2011

2.44 Rate [counts/(96h 10 8 )] 2.43 2.42 2.41 2.40 Residual 10 5 0-10 5 0 50 100 150 200 Time [days] Figure 4: Detected activity of the 137 Cs source. Dead-time corrected data are summed over 96 hours. The first two points correspond to the beginning of data taking, when the set-up was stabilizing, and they are not considered in the analysis. Dotted lines represent a 0.1% deviation from the exponential trend. Residuals (lower panel) of the measured activity to the exponential fit. Error bars include statistical uncertainties and fluctuations in the measured dead time. to January 9th 2012. The activity of the source can be calculated for any interval of time multiple of the hour. In particular, Fig. 4 shows the activity per 4 days as a function of time, i.e. the integral from 7 kev to 1.7 MeV of the MCA spectra collected during 4 days and corrected for the dead time. A total error equal to the linear sum of the statistical uncertainty (6.4 10 5 relative) and of the fluctuations of the dead time (2.8 10 5 relative) has been assigned to each data point. The activity decrease is due to the source decay. The continuous line in Fig.4 is the exponential fit obtained with the mean life which minimizes the Chi Squared: 43.22 years (we take the year of 365.25 days of 86400 seconds each). The reduced Chi Squared per degree of freedom is 1.02. During the 7 month period of data taking the temperature at the Cs source smoothly increased by 0.4 K with a few larger variations up to 0.7 K during one week. There is no effects on the data correlated with such temperature variations. In the same period the pressure inside the 4 laboratory varies in the range ±10 mbar. Also for these pressure changes we can exclude any significant effect on the rate. 3.1. 137 Cs half-life The 137 Cs half-life has been determined by many authors during more than fifty years (for a critical review of available data see [15]). Data collected in this measurement allow for a new and precise estimate of the 137 Cs half-life. The data are fitted with an exponential function leaving two parameters free: initial decay rate and half-life. The resulting half-life is 10942±30 days, to be compared with the recommended value of 10976±30 days. 3.2. Time modulations We searched for time variations with periods from 6 hours to 400 days. For short periods, up to 40 days, the

use of the discrete Fourier transform of the hourly data, after the subtraction of the exponential trend, is appropriate. For longer period we performed a Chi Squared analysis of the daily count rate, fitting with a superposition of the exponential decay function (with our mean life time estimate of 43.22 years) and a sine function of fixed period. The initial activity, the amplitude of the oscillation and its phase are left free. We scanned all the periods from 40 to 400 days, studying in particular the 1 solar year period. No significant improvement of the Chi Squared per degree of freedom has been observed in this range. Our analysis excludes any oscillation with amplitude larger than 9.6 10 5 at 95% C.L.. In particular, for an oscillation period of 1 year the amplitude is 3.1(2.7) 10 5, well compatible with zero; and a limit of 8.5 10 5 at 95% C.L. on the maximum allowed amplitude is set independently of the phase. As a consequence, the claimed existence of an oscillation with amplitude of about 10 3 due to the variation of the Sun to Earth distance is rejected by the present data. [4] D.E. Alburger et al., Earth an Planetary Science Lett. 78 (1986) 168. [5] E. Fischbach et al., arxiv:1106.1470 (2011), Rencontre de Moriond 2011 (3-2011) La Thuile. [6] A.G. Parkhomov, arxiv:1004.1761 (2010). [7] D. Javorsek II et al., Astropart. Phys. 34 (2010) 173. [8] J.C. Hardy et al. arxiv:1108.5326 (2011). [9] P.S. Cooper, Astropart. Phys. 31 (2009) 267. [10] E. B. Norman et al, Astroparticle Physics 31 (2009) 135. [11] J.H. Jenkins et al., Astroparticle Physics 31 (2009) 407. [12] A.G. Parkhomov, arxiv:1006.2295 (2010). [13] Y.A. Baurov, Mod. Phys. Lett. A 16 (2001) 2089. [14] I. Bikit et al., Phys. Rev. C 54 (1996) 3270. [15] R.G. Helmer and V.P. Chechev, Bureau International des Poids et Mesures Monographies-5, vol. 1-6 (2011) 4. Conclusion The half-life of a 137 Cs source has been estimated to be 10942±30 days, in agreement with its recommended value. Moreover, from our measurement we can exclude the presence of time variations in the source activity, superimposed to the expected exponential decay, larger than 9.6 10 5 at 95% C.L. for oscillation periods in the range 6 hours-1 year. In particular, we exclude an oscillation amplitude larger than 8.5 10 5 at 95% C.L. correlated to the variation of the Sun-Earth distance, in clear contradiction with previous experimental results and their interpretation as indication of a novel field (or particle) from the Sun. The data taking with the improved experimental set-up is going to continue for at least another 6 months. 5. Acknowledgments Thanks are due to our colleagues C. Cattadori, L. Pandola and C. Rossi Alvarez for many useful discussions; to M. Dedeo who provided us data on atmospheric pressure, to L. Votano, director of the LNGS, for the constructive hospitality. References [1] M. Curie, Doctoral Dissertation, Sorbonne University, Paris (1903). [2] J.H. Jenkins et al., Astropart. Phys. 32 (2009) 42. [3] H. Siegert et al., Appl. Radiat. Isot. 49 (1998) 1397. 5