Experimental studies of THGEM in different Ar/CO 2 mixtures

Similar documents
Experimental research on the THGEM-based thermal neutron detector

Performance study of the ceramic THGEM *

Simulation and performance study of ceramic THGEM *

p/π + response of single layer THGEM detector in Ar/3%iso *

& SUN ZhiJia 1;2. Abstract

Spatial resolution measurement of Triple-GEM detector and diffraction imaging test at synchrotron radiation

Advances in the Micro-Hole & Strip Plate gaseous detector

GEM at CERN. Leszek Ropelewski CERN PH-DT2 DT2-ST & TOTEM

Aging effect in the BESIII drift chamber

Chinese Physics C. Study of n-γ discrimination in low energy range (above 40. kevee) by charge comparison method with a BC501A liquid

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection

Performance of a triple-gem detector for high-rate particle triggering

Key words: boron-coated straw, n-gamma discrimination, longitudinal resolution, three-dimensional positioning, neutron scattering spectrometer

GEM: A new concept for electron amplification in gas detectors

Micro Pixel Chamber with resistive electrodes for spark reduction

arxiv:physics/ v2 27 Mar 2001

Electron transparency, ion transparency and ion feedback of a 3M GEM

R&D on Astroparticles Detectors. (Activity on CSN )

Nuclear Instruments and Methods in Physics Research A

X-ray ionization yields and energy spectra in liquid argon

Development of New MicroStrip Gas Chambers for X-ray Applications

A Triple-GEM Telescope for the TOTEM Experiment

Rate Dependency Study on Gas Electron Multiplier Gain

Devices for high precision x-ray beam intensity

Simulating the Charge Dispersion Phenomena in Micro Pattern Gas Detectors with a Resistive Anode

Study of GEM-like detectors with resistive electrodes for RICH applications

Breakdown limit studies in high-rate gaseous detectors

Micro Pixel Chamber Operation with Gas Electron Multiplier

arxiv: v1 [physics.ins-det] 19 Apr 2014

X-ray ionization yields and energy spectra in liquid argon

First experimental research of low energy proton radiography

ATLAS New Small Wheel Phase I Upgrade: Detector and Electronics Performance Analysis

PHOTOELECTRON COLLECTION EFFICIENCY AT HIGH PRESSURE FOR A GAMMA DETECTOR ENVISAGING MEDICAL IMAGING

CCD readout of GEM-based neutron detectors

General Overview of Gas Filled Detectors

GEM-based photon detector for RICH applications

A fast triple GEM detector for high-rate charged-particle triggering

Simulation of GEM-TPC Prototype for the Super-FRS Beam Diagnostics System at FAIR

Research Article Analytic Approximation of Energy Resolution in Cascaded Gaseous Detectors

Development of a new MeV gamma-ray camera

A new detector for neutron beam monitoring

High Spatial Resolution in thermal Neutron Detection: from CASCADE to BASTARD. Markus Köhli. DPG Frühjahrstagung Münster

Particle Energy Loss in Matter

A Complete Simulation of a Triple-GEM Detector

Flux and neutron spectrum measurements in fast neutron irradiation experiments

Recent advances in gaseous imaging photomultipliers

Status and Challenges of CEPC Time Projection Chamber Detector. Huirong On behalf of CEPC Tracking Subgroup

Hybrid Gaseous Detector Module for CEPC-TPC

Precision neutron flux measurement with a neutron beam monitor

Microscopic Simulation of GEM Signals

Progress in the Development of Photosensitive GEMs with Resistive Electrodes Manufactured by a Screen Printing Technology

Signals in Particle Detectors (1/2?)

PoS(EPS-HEP2015)232. Performance of a 1 m 2 Micromegas Detector Using Argon and Neon based Drift Gases

Submitted to Chinese Physics C CSNS/RCS

Precision Calibration of Large Area Micromegas Detectors Using Cosmic Muons

Development of a Time Projection Chamber with GEM technology in IMP. Herun yang Gas detector group

Design and performance of Hall probe measurement system in CSNS

Imaging principle and experiment results of an 11 MeV low-energy proton radiography system

Neutron Structure Functions and a Radial Time Projection Chamber

SUPPLEMENTARY INFORMATION. Observation of tunable electrical bandgap in large-area twisted bilayer graphene synthesized by chemical vapor deposition

The Compact Muon Solenoid Experiment. Conference Report. Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

A proposal to study gas gain fluctuations in Micromegas detectors

STUDIES OF A GAMMA-BLIND FAST-NEUTRON IMAGING DETECTOR FOR BNCT APPLICATIONS

arxiv:physics/ v2 [physics.ins-det] 31 Oct 2003

Siberian Branch of Russian Academy of Science. A. Buzulutskov, A. Bondar, L. Shekhtman, R. Snopkov, Yu. Tikhonov

R&D of a novel gas electron multiplier the THGEM

High pressure xenon gas detector with segmented electroluminescence readout for 0nbb search

Detectors in Nuclear and High Energy Physics. RHIG summer student meeting June 2014

TitleDevelopment of an MeV gamma-ray ima. works must be obtained from the IEE

SILVER NEUTRON ACTIVATION DETECTOR FOR MEASURING BURSTS OF 14 MEV NEUTRONS

A complete simulation of a triple-gem detector

Status report of CPHS and neutron activities at Tsinghua University

GEM upgrade of the ALICE TPC. Markus Ball Physics Department E18 Technische Universität München On Behalf of the ALICE TPC upgrade

Neutron collimator design of neutron radiography based on the BNCT facility arxiv: v1 [physics.ins-det] 3 May 2013

Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector

Recent advances in gaseous imaging photomultipliers

ACCURATE MEASUREMENTS OF SURFACE EMISSION RATE FOR LARGE-AREA ALPHA AND BETA REFERENCE SOURCES

China high-intensity accelerator technology developments for Neutron Sources & ADS

Absorption and Backscattering ofβrays

Track Resolution Measurements for a Time Projection Chamber with Gas Electron Multiplier Readout

PHYS 3446 Lecture #12

Investigation of GEM space point resolution for a TPC tracker

PERFORMANCE IMPROVEMENT OF CZT DETECTORS BY LINE ELECTRODE GEOMETRY

Spatial Resolution of a Micromegas-TPC Using the Charge Dispersion Signal

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use

TPC-like analysis for thermal neutron detection using a GEMdetector

Scintillation Detector

Flexible Asymmetrical Solid-state Supercapacitors Based on Laboratory Filter Paper

arxiv: v1 [astro-ph.im] 25 Apr 2014

arxiv: v1 [physics.ins-det] 26 Jun 2013

Absorption and Backscattering of β-rays

Charge readout and double phase

Boron-based semiconductor solids as thermal neutron detectors

DESIGN OF NEUTRON DOSE RATE METER FOR RADIATION PROTECTION IN THE EQUIVALENT DOSE

Concept of a novel fast neutron imaging detector based on THGEM for fan-beam tomography applications

6th International Conference on Information Engineering for Mechanics and Materials (ICIMM 2016)

Lecture 18. New gas detectors Solid state trackers

GEANT4 simulation of the 10 B-based Jalousie detector for neutron diffractometers

2008, hm 2. ( Commodity Bundle) [ 6], 25 4 Vol. 25 No JOURNAL OF NATURAL RESOURCES Apr., , 2, 3, 1, 2 3*,

Design, construction and test of Boron Array Neutron Detector - Gas Electron Multiplier (BAND-GEM)

Transcription:

Experimental studies of THGEM in different Ar/CO 2 mixtures He Zhan-ying( 何占营 ) 1;2) Zhou Jian-rong( 周健荣 ) 2;3*) Sun Zhi-jia( 孙志嘉 ) 2;3) Yang Gui-an( 杨桂安 ) 2;3) Xu Hong( 许虹 ) 2;3) Wang Yan-feng( 王艳凤 ) 2;3) Liu Qian( 刘倩 ) 3;4) LIU Hong-bang( 刘宏邦 ) 3;4) Chen Shi( 陈石 ) 4) Xie Yi-gang( 谢一冈 ) 2;4) Zheng Yang-heng( 郑阳恒 ) 3;4) Wang Xiao-dong( 王晓冬 ) 1 Zhang Xiao-dong( 张小东 ) 1 Hu Bi-tao( 胡碧涛 ) 1 Chen Yuan-Bo( 陈元柏 ) 2;3) 1 School of Nuclear Science and Technology, Lanzhou University, Lanzhou 730000, China; 2 Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, China; 3 State Key Laboratory of Particle Detection and Electronics,Beijing 100049,China; 4 University of Chinese Academy of Sciences, Beijing 100049,China Abstract: In this paper, the performances of a type of the domestic THGEM (THick Gaseous Electron Multiplier) working in the Ar/CO 2 mixtures are reported in details. This kind of single THGEM can provide the gain range from 100 to 1000, which is very suitable for the application in the neutron detection. In order to study its basic characteristics as the references for the development of THGEM based neutron detector, the counting rate plateau, the energy resolution and the gain of the THGEM have been measured in the different Ar/CO 2 mixtures with the change of the electrical fields. For the Ar/CO 2 (90%/10%) gas mixture, a wide counting rate plateau is got from 720V to 770V with the plateau slope of 2.4% / 100 V and the excellent energy resolution about 22% is obtained at the 5.9keV full energy peak of the 55 Fe X-ray source. Key words: THGEM, counting rate plateau, gas gain, energy resolution; PACS: 29.40.Gx, 29.40.Cs, 28.20.Cz 1. Introduction Neutrons are used to investigate the structure and dynamics of a material. Many efforts have recently been devoted to the development of the next generation of neutron facilities, which include SNS in USA, J- PARC in Japan, ISIS in UK, CSNS (China Spallation Neutron Source) in China and ESS in Europe [1]. The neutron detector is one of the key components of the neutron scattering instruments. With the international development of the new generation neutron source, the traditional neutron detector based on He-3 has not been able to satisfy very well the demand of the application of high flux especially. And also facing the global crisis of He-3 * Supported by National Natural Science Foundation of China (11127508,11175199) *) E-mail: zhoujr@ihep.ac.cn Phone:010-88236411

supply as well [2], the research on the new style of the neutron detector which can replace the He-3 based detection technology becomes extremely urgent. As a good candidate, a boron coated GEM became the focus of attention recently [3], which was firstly designed by Martin Klein using CERN standard GEM in 2006[4]. It has the outstanding and excellent characteristics, such as high counting rate capability (>10MHz), good spatial resolution and timing properties, radiation resistance, flexible detector shape and readout patterns [5]. In 2011, IHEP and UCAS firstly developed successfully a kind of THGEM, manufactured economically by standard printed-circuit drilling and etching technology in China. Compared with the CERN standard GEM, THGEM has higher gain, sub-millimeter spatial resolution and the possibility of industrial production capability of large-area robust detectors, which is very suitable and adequate for the application of neutron detection. In this paper, the THGEM is provided by Zheng Yang-heng group of UCAS. It s a thinner-thgem with a thickness of 200μm, hole diameter of 200μm, pitch of 500μm and a very small rim of 5 10μm. The thinner-thgem is made of FR4 glass epoxy substrate, with 20μm thick copper- cladding on both sides and has an active area of 50mm*50mm [6].In order to study its basic characteristics as the references for the development of this kind of domestic THGEM based neutron detector, the performances of the counting rate plateau, the energy resolution and the gain have been measured in the different Ar/CO 2 mixtures with the different high voltages. According to the experiments, the working conditions optimized of the THGEM have been obtained, which would be very helpful for the design of this kind of THGEM based neutron detector in future [7,8]. 2. Experimental setup Fig.1 shows a schematic view of the detector configuration consisting of cathode, anode and a single THGEM. The THGEM detector was operated in different Ar/CO 2 mixtures at a normal pressure and temperature. Measurements were carried out by using a 55 Fe X ray source (activity 10mCi) which was positioned in such a way that a collimated beam (ф1 hole) of X-rays perpendicularly entered the upper drift region. The ionization electrons generated by the interaction of 55 Fe 5.9keV X-rays with Ar atoms were amplified in avalanche mode in the THGEM holes [8], and then entered the induction region, where they were finally collected by the anode. All the three electrodes of HV were supplied by the WIENER MPOD mini-hv power and the signals were readout with an ORTEC 142IH preamplifier followed by an ORTEC 572A amplifier (shaping time t=2μs) and an ORTEC multi-channel analyzer (trump-usb-8k).

3. Results and discussions 3.1 Plateau Fig.1 Schematic view of a single THGEM detector with drift region Dd=5.5mm and induction region Di=3.5mm. In order to know the suitable working voltage of the detector in different Ar/CO 2 mixture, its counter plateau was measured in different Ed (drift field) and Ei (induction field) and the results are shown on Fig.2. During the experiment, the total flow of Ar and CO 2 gas is 50 SCCM to ensure the amount of effective working gas in the chamber. The counts were recorded in every one minute and the voltage of THGEM was increased by the increment of 5V until the spark discharge occurred. As the Fig.2 shows, it has a longer plateau in the Ar/CO 2 mixture ratio of 90%/10% and gets shorter with the increase of the proportion of CO 2. In certain extent, the counting rate was independent on the Ed and increasing with the Ei, so that the plateau was shifted left with the increasing of the Ei. For the gas mixture Ar/CO 2 (90%/10%), the induction field Ei of 2.0 kv/cm and the drift field Ed of 0.5kV/cm, the plateau range of this kind of THGEM was from 720V to770 V and its plateau slope was about 2.4% / 100 V, as shown in the top left of the Fig.2. This optimization would be helpful to know the working range of the THGEM and find the conditions with the lower HV. Fig.2 Counting rate plateaus of the THGEM detector in the three different kinds of gas mixture (90%Ar/10%CO 2 80%Ar/20%CO 2 and 70%Ar/30%CO 2 ) with Ed=0.5kV/cm and Ei=1.0kV/cm 1.5kV/cm 2.0kV/cm and 3.0kV/cm.

3.2 Gain By using the 5.9keV full energy peak of the 55 Fe X-ray source, the effective gain was measured with the voltage of THGEM increased by the increment of 5V in different Ar/CO 2 mixture and different fields Ed and Ei. In certain extent, the stronger the induction field Ei is and the greater the voltage of the THGEM is, the gain is larger for the same gas mixture. As Fig.3 shows, the gain increases exponentially with the voltage of the THGEM for each kind of gas mixture and the single THGEM can provide a wide gain range from 100 to1000. The gain decreases with the increase of the proportion of CO 2 in the gas mixture. Fig.3 The experimental gain in the three different kinds of gas mixture (90%Ar/10%CO 2 80%Ar/20%CO 2 and 70%Ar/30%CO 2 ) with Ed=0.5kV/cm and Ei=1.0kV/cm 1.5kV/cm 2.0kV/cm and 3.0kV/cm. As mentioned above, the influence of drift field Ed to the gain is very small in certain extent. Fig.4 (a) shows the gain as a function of Ed in different Ei in the Ar /CO 2 (90%/10%) gas mixture. In different Ed, the gain keeps nearly no change at the voltage 770V of THGEM for each Ei and increases with the Ei. As the same, Fig.4 (b) shows the gain change with Ed in different Ei in the Ar/CO 2 (80%/20%) gas mixture and it shows the similar results as those in the Ar/CO 2 (90%/10%) gas mixture. For the Ar/CO 2 (70%/30%) gas mixture, the same regularity exists. Due to no common appropriate voltage of THGEM, the similar kind of this figure has not been presented.

Fig.4 Gain vs Ed is in different Ei and in same voltage of the THGEM. (a) Gain changes with the Ed in an Ar/CO 2 (90%/10% ) gas mixture and the voltage of the THGEM VGem=770V.(b) Gain changes with the Ed in an Ar/CO 2 (80%/20%) gas mixture and the voltage of the THGEM VGem=830V. 3.3 Energy resolution Energy resolution is one of the most important parameters related to the detector performance. By using the 5.9 kev full energy peak of the 55 Fe X-ray source, the energy resolution was measured. The Fig.5 shows the results of three kinds of gas mixture with the voltage of THGEM increased by the increment of 5V in different field Ei. There are three regions with clear boundary related to the ratio of CO 2 in the gas mixture. As the ratio of CO 2 decreases, the working voltage of the THGEM will get lower and the energy resolution will get smaller as well. The much better energy resolution is obtained in the Ar/CO 2 (90%/10%) gas mixture. With the voltage of THGEM increasing, the energy resolution will also get better obviously. The induction field Ei has a bit affection on the energy resolution and the drift field has nearly no effect on the energy resolution from 0.5 to 3kV/cm (measured and not included in the figure). As a summary for the optimization, the best working points of this kind of THGEM are recommended as following: the Ar/CO 2 (90%/10%) gas mixture, the drift field 0.5kV/cm, the induction field 2kV/cm and the voltage range of THGEM from 720V-770V, which will give the better energy resolution smaller than 25%. Fig.5 The energy resolution in the three different kinds of gas mixture (90%Ar/10%CO 2 80%Ar/20%CO 2 and 70%Ar/30%CO 2 ) with Ed=0.5kV/cm and Ei=1.0 kv/cm 1.5 kv/cm 2.0 kv/cm and 3.0kV/cm. Fig.6 shows a pulse height spectrum obtained with a 55 Fe source in the Ar/CO 2 (90%/10%) gas mixture. For obtaining energy resolution, it s fitted with the Gaussian function. It indicates the energy resolution (FWHM) of the detector based on THGEM is about 22%. With such an energy resolution, the detector can entirely separate the 3keV of Ar escape peak from the 55 Fe main X-ray peak located at 5.9keV.

4. Conclusion Fig.6 A pulse height spectrum in the Ar/CO 2 (90%/10%) gas mixture Ed=0.5kV/cm Ei=2kV/cm and the voltage of the THGEM VGem=770V with 55 Fe source. In this paper, the experimental performance of a kind of domestic THGEM working in the Ar/CO 2 mixtures are presented in details. The effective gain of single THGEM can reach about 1000 in the Ar/CO 2 mixture. As the ratio of CO 2 decreases from 30% to 10%, the working voltage of the THGEM will get lower, the plateau will get longer and the energy resolution will get much better as well. With the induction field Ei increasing from 1 to 3kV/cm, the performance of THGEM will be getting better and better. The drift field Ed (0.5 to 1.5kV/cm) has nearly no influence on the performance of THGEM. As a summary for the optimization, the best working points of this kind of THGEM are recommended as following: the Ar/CO 2 (90%/10%) gas mixture, the drift field 0.5kV/cm, the induction field 2kV/cm and the voltage range of THGEM from 720V-770V, which will give the better energy resolution smaller than 25%. According to the experiments, it would be very helpful for the design of this kind of THGEM based neutron detector in future. 5. Acknowledgements We are grateful to the supports from National Natural Science Foundation of China (item: 11127508 and 11175199), China Spallation Neutron Source, the State Key Laboratory of Particle Detection and Electronics and the State Key Laboratory of neutron detection and fast electronics technology in Dongguan University of Technology. References [1]B.Gebauer, Towards detectors for next generation spallation neutron sources, Nucl.Instr. and Meth. A,2004, 535: 65-78 [2] Dana A. Shea et al. The Helium-3 Shortage: Supply, Demand, and Options for Congress, Congressional Research Service, 2010 [3]H. Ohshita et al. Development of a neutron detector with a GEM, Nucl.Instr. and Meth. A, 2010,623:126-128 [4] C.Schmidt, M.Klein, The CASCADE Neutron Detector: A System for 2D Position

Sensitive Neutron Detection at Highest Intensities, Neutron News, 2006, 17(1):12-15 [5] M.Klein, C.Schmidt, CASCADE neutron detectors for highest count rates in combination with ASIC/FPGA based readout electronics, Nucl.Instr. and Meth. A, 2011, 628:9-18 [6] H.B. Liu, Y.H.Zheng et al. The performance of thinner-thgem, Nucl.Instr. and Meth. A,2011,659:237 241 [7] Zhou JianRong et al. Neutron beam monitor based on a boron-coated GEM, Chinese Physics C,2011,35(7):668-674 [8] WANG Yan-Feng, SUN Zhi-Jia, ZHOU Jian-Rong et al. Simulation study on the boron-coated GEM neutron beam monitor, Chinese Physics, SCIENCE CHINA Physics,Mechanics & Astronomy,2013(received) [9] A.Breskin, R.Alon et al.a concise review on THGEM detectors, Nucl.Instr. and Meth. A,2009,598:107 111