Determination of Benzo(a)pyrene in Sausage and Preserved Ham

Similar documents
Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Simultaneous Determination of Aromatic Amines and Pyridines in Soil

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

Determination of Carbonyl Compounds in Workplace Air

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Determination of Aniline and Nitroanilines in Environmental and Drinking Waters by On-Line SPE

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Determination of trace anions in concentrated hydrofluoric acid

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Determination of Methylamine in Drug Products

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column

Fast methods for the determination of ibuprofen in drug products

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

APPLICATION NOTE Goal. Introduction. Keywords

Monitoring Protein PEGylation with Ion Exchange Chromatography

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

[ Care and Use Manual ]

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

HILIC Method Development in a Few Simple Steps

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Determination of silicate in high-purity water using ion chromatography and online sample preparation

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

of mass spectrometry

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Comprehensive Neurochemical Profiling of Brain Tissue Samples

Screening Method for 30 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Thermo Scientific. Anion-Exchange Column

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Determination of urea in ultrapure water by IC-MS/MS

Advances in Sample Preparation for Accelerated Solvent Extraction

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Determination of Additives in Carbonated Beverages

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Determination of Monochloroacetic Acid in Carbocisteine

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Inline sample preparation for the determination of anions in sodium hydroxide

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep

Supported Liquid Extraction (SLE) Guide and FAQ s. User Guide

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Dionex IonPac AS28-Fast-4µm column

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

HPLC Praktikum Skript

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Enhancement of Linearity and Response in Charged Aerosol Detection

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Determination of 24 PAHs in Drinking Water

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Transcription:

Determination of Benzo(a)pyrene in Sausage and Preserved Ham Ran Liangji, Jin Yan, Xu Qun, Liang Lina, and Jeffrey Rohrer 2 Thermo Fisher Scientific, Shanghai, People s Republic of China; 2 Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 093 Key Words PAHs, On-Line SPE, HPLC, Fluorescence Detection, Food Safety Analysis Goal To develop an efficient high-performance liquid chromatography (HPLC) method for sensitive and rapid determination of benzo(a)pyrene in meat products such as sausage and preserved ham using on-line solid-phase extraction (SPE) for sample preparation instead of the commonly used off-line SPE method Introduction Benzo(a)pyrene (structure shown in Figure ) is a polycyclic aromatic hydrocarbon (PAH) compound formed from the incomplete combustion of organic matter. Due to its potential carcinogenic and mutagenic properties, most countries have regulations limiting its concentration in drinking water, food additives, cosmetics, workplaces, and factory emissions. Some samples such as sausage and preserved ham have too complex a matrix that interferes with the separation of benzo(a)pyrene, thus preventing use of a routine HPLC method and making benzo(a)pyrene quantification difficult. Therefore, sample preparation steps including extraction and extract cleanup are very important for the determination of benzo(a)pyrene by HPLC. Liquid-liquid extraction and off-line SPE are among the commonly used sample preparation techniques for meat products and have been applied to many regulatory methods (Table ). -3 Figure. Structure of benzo(a)pyrene. Table. Regulations for benzo(a)pyrene analysis required by the China Institute of Standardization (CNIS) in the People s Republic of China. Regulation Sample Pretreatment Determination Method NY/T 666-2008 SC/T 304-2008 2 GB/T 5009.27-2003 3 *Method detection limit Meat (Barbecue, Smoked) Aquatic Products Food (Meat) Cyclohexane Extraction and Liquid-Liquid Extraction Saponification, N-Hexane Extraction, and Off-Line SPE Purification Hexane Extraction and Off-Line SPE Purification HPLC with Fluorescence Detection HPLC with Fluorescence Detection MDL * (μg/kg) 0.5 Fluorospectrophotometry.0

2 Thermo Scientific Application Notes (ANs) 96, 23, and 085 provide on-line SPE HPLC methods to quantify low concentrations of PAHs in oil and water samples. 4-6 The work shown here provides a sensitive and rapid HPLC method with on-line SPE and fluorescence detection for the determination of benzo(a)pyrene in sausage and preserved ham. This method reduces costs associated with the SPE cartridge, labor, time, and reagents; and results are more consistent because the cleanup step is performed by on-line SPE rather than manual off-line SPE. Equipment Thermo Scientific Dionex UltiMate 3000 Dual Rapid Separation LC (RSLC) system, including: DGP-3600RS Dual Gradient Rapid Separation Pump (P/N 5040.0066) SRD-3600 Integrated Solvent and Degasser Rack (P/N 5035.9230) WPS-3000TRS Rapid Separation Wellplate Sampler, Thermostatted (P/N 5840.0020), equipped with a 00 μl sample loop and a 25 μl syringe TCC-3000RS Rapid Separation Thermostatted Column Compartment (P/N 5730.000), equipped with one 2 6p valve FLD-3400RS Rapid Separation Fluorescence Detector (P/N 5078.0025) Thermo Scientific Dionex Chromeleon Chromatography Data System (CDS) software, version 7. or above Thermo Scientific Sorvall ST6 Centrifuge (P/N 75004240) RV 0 Basic Rotary Evaporator with Dry Ice Condenser (IKA P/N 80300) T 8 Digital ULTRA-TURRAX Disperser (IKA P/N 3720000) Consumables Thermo Scientific Target2 Polypropylene Syringe Filters (0.45 μm, 30 mm, P/N F2502-9) Reagents and Standards Deionized (DI) water, 8.2 MΩ-cm resistivity, generated by the Thermo Scientific Barnstead GenPure Pro UV-TOC (total organic carbon) Water Purification System (P/N 503948) Methanol, 99.8%, HPLC Grade (Fisher Scientific P/N AC60090040) Acetonitrile, HPLC Grade (Fisher Scientific P/N AC6000040) Ethyl Acetate, HPLC Grade (Fisher Scientific P/N E95-4) Sodium Sulfate Anhydrous, Powder (Fisher Scientific P/N S429-500) Benzo(a)pyrene, 99.6% (Dr. Ehrenstorfer GmBH P/N 20635000) Conditions For On-Line SPE Column: Thermo Scientific Acclaim 20, C8, 5 µm Analytical, 4.6 50 mm (P/N 05938) Mobile Phase: A. Acetonitrile B. Water C. Ethyl acetate Gradient: Table 2 Flow Rate:.0 ml/min Injection Volume: 0 μl on the on-line SPE column Table 2. Gradient program for loading pump. Time (min) For Separation Column: % B (Water) Mobile Phase: Flow Rate: Temperature: 35 C Detection: Fluorescence Excitation: 290 nm Emission: 40 nm Response Time: 2.0 s Data Collection Rate: 5.0 Hz Sensitivity: % C (Ethyl Acetate) Valve Position 0 20 0 _2 0.3 20 0 6_.8 20 0 _2 2.0 20 0 _2 2.5 0 75 _2.5 0 75 _2 2 20 0 _2 5 20 0 _2 Acclaim Phenyl-, 3 μm Analytical, 4.6 50 mm (P/N 07969)* Acetonitrile:Water (8:2, v/v).0 ml/min * The Thermo Scientific Hypersil Green PAH Column, 3 μm, 3.0 50 mm (P/N 303-53030) can be used as a substitute. Preparation of Standard Solutions Stock Standard Solution Dissolve ~2.5 mg of benzo(a)pyrene in 50 ml of methanol in a 250 ml volumetric flask and dilute to the mark with methanol. The final concentration of benzo(a)pyrene will be 50 mg/l. Stock Standard Solution 2 Add ml of Stock Standard Solution to a 50 ml volumetric flask and dilute to the mark with methanol. The final concentration of benzo(a)pyrene will be mg/l.

Table 3. Preparation of calibration standards. 3 Stock Standard Solution of Benzo(a)pyrene Volume of Stock Standard Solution (µl) Volume of Methanol (ml) Final Volume of Calibration Standard (ml) Final Concn of Calibration Standard (µg/l) Stock Standard Solution 2 ( mg/l) Stock Standard Solution 3 (20 µg/l) 500 9.50 50 00 9.90 0 50 9.95 05 0.0 500 9.50 0 250 9.75 0.5 50 9.95 0. Stock Standard Solution 3 Add ml of Stock Standard Solution 2 to a 50 ml volumetric flask and dilute to the mark with methanol. The final concentration of benzo(a)pyrene will be 20 µg/l. Working Standard Solutions for Calibration For calibration, prepare six working standard solutions with different concentrations by diluting the proper amounts of Stock Standards 2 and 3 with methanol. The volumes of each solution needed to make the calibration standards are shown in Table 3. Sample Preparation Two kinds of meat product samples, sausage and preserved ham, were purchased from a supermarket in Shanghai, China. Add 20 ml of ethyl acetate to 2 g of anhydrous sodium sulfate and 2 g of mashed meat product sample in a 50 ml centrifuge tube. Mix for min to extract, centrifuge the extract for 2 min at 8000 rpm, remove the supernatant, then add 20 ml of ethyl acetate to the residue and extract a second time in the same manner. Combine the two supernatants (total volume ~40 ml) and use rotary evaporation at 55 C to condense the volume to 2 ml. Transfer the condensed solution to a 0 ml volumetric flask and dilute to the mark with ethyl acetate. Prior to injection, filter the solution through a 0.45 μm filter. Results and Discussion Optimization of Sample Preparation Four solvents with different polarities acetonitrile, isopropanol, ethyl acetate, and n-hexane were evaluated as extraction solvents for extracting benzo(a)pyrene from a spiked sausage sample. The method recoveries of benzo(a)pyrene obtained using acetonitrile and isopropanol were less than 50% and, therefore, were not further pursued. When using n-hexane a solvent with low polarity more fat can be extracted from the sample, which can reduce the efficiency of on-line SPE as well as the precision of the method. Using ethyl acetate as the extraction solvent, the average recovery of benzo(a)pyrene was good (98%), obtained by spiking five preserved ham samples with 5 μg/kg benzo(a)pyrene, using one injection for each spiked sample. The peak area RSD was 2.40 and there was a time savings of approximately 5 min, compared to using n-hexane in the SPE column regeneration. Therefore, ethyl acetate was chosen as the extraction solvent. Note: To dehydrate the sample during sample preparation, anhydrous sodium sulfate was added, which can reduce the extraction of high-polarity compounds. Evaluation of On-Line SPE Figure 2 shows a typical flow schematic of the on-line SPE column directly coupled to the HPLC column using one 6-port (2 6p) valve. The filtered sample is directly injected onto the system and delivered to the SPE column for enrichment (_2 position) using one pump of the dualpump module (labeled For On-Line SPE); the analytical column is simultaneously equilibrated using the other pump of the dual-pump module (labeled For Separation). After analytes are bound to the SPE column and impurities washed out, the SPE column is switched into the analytical flow path (6_ position) to elute bound analytes separated on the analytical column and detected by the fluorescence detector. This method can be easily performed using an UltiMate 3000 Dual RSLC system. Figure 2. Flow schematic of on-line SPE.

4 Selection of On-Line SPE and Analytical Columns As discussed in previous work (ANs 96, 23, and 085), the Acclaim PolarAdvantage II column is a good choice for on-line SPE preconcentration of large volumes of 00% aqueous samples. 4-6 However, the Acclaim 20 C8 column was chosen as the on-line SPE column for this study because the meat samples have high fat content and fat can be easily eluted from this column during its regeneration. Baseline separation of 20 PAHs (including the 6 PAHs specified in U.S. Environmental Protection Agency [EPA] Methods 550, 550., and 60) in a water sample can be accomplished using the Hypersil Green PAH column that features a specially tailored alkyl-bonded silica with high carbon loading. 7-9 However, the Acclaim Phenyl- column, based on covalent modification of silica particles with silane ligand-bearing proprietary alkyl aromatic functionality, can yield good chromatography for a single PAH (e.g., benzo(a)pyrene). As shown in Figure 3, using a combination of the Acclaim 20 C8 column for on-line SPE and the Acclaim Phenyl- column for separation, on-line SPE followed by HPLC of benzo(a)pyrene can be accomplished in 5 min. Therefore, the Acclaim Phenyl- column was used for the work shown here. 50,000 Peak:. Benzo(a)pyrene µg/l Calibration linearity for fluorescence detection of benzo(a)pyrene was investigated by making five consecutive 0 µl injections of a standard prepared at six different concentrations (i.e., 30 total injections). Linearity was observed from 0. to 50 μg/l when plotting concentration versus peak area. The linear regression equation was A = 8254c + 208.3, where A represents peak area and c represents concentration of analyte. The coefficient of determination was.000. This calibration curve was used to quantify benzo(a)pyrene in sausage and preserved ham samples. Five replicate injections of a benzo(a)pyrene standard with a concentration of 0. μg/l were used for estimating the MDL using a signal-to-noise ratio = 3. The MDL of benzo(a)pyrene was 0.02 μg/l. Analysis of Meat Product Samples Table 4 summarizes the results of analyzing sausage and preserved ham samples. Benzo(a)pyrene was found in four of the samples. However, no benzo(a)pyrene was found in any of the meat samples when using the fluorescence spectrophotometric detection method described in GB/T 5009.27-2003. 3 Figure 4 shows a chromatogram of preserved ham samples. To judge method accuracy, five injections were made of Preserved Ham Sample 3 spiked with 5 µg/kg of benzo(a)pyrene standard. The recovery range was from 95 to 05%. Table 4. Sample analysis results. Counts 0-5000 0 2 4 6 8 0 2 4 5 Minutes Figure 3. On-line SPE-HPLC determination of benzo(a)pyrene using a combination of the Acclaim 20 C8 and Phenyl- columns. Reproducibility, Linearity, and Detection Limits Method reproducibility was estimated by making five consecutive injections of a calibration standard with a concentration of μg/l benzo(a)pyrene. The retention time reproducibility RSD is 0.04 and peak area reproducibility RSD is 0.96, demonstrating good shortterm precision for this on-line SPE HPLC method. Sample preparation reproducibility was estimated by spiking five preserved ham samples with 5 μg/kg benzo(a)pyrene, using one injection for each spiked sample. The retention time reproducibility RSD was 0.07 and peak area reproducibility RSD was 2.40, demonstrating good sample preparation reproducibility. Sample Detected (μg/kg) Method in this Work GB/T 5009.27-2003 Sausage 0.49 ND * Sausage 2.36 ND Preserved Ham 0.98 ND Preserved Ham 2 0.53 ND Preserved Ham 3 ND ND *ND = Not Detected 392,000 Counts 0 C B A Samples: A. Preserved Ham Sample 2 B. Preserved Ham Sample 3 C. Preserved Ham Sample 3 spiked with 5 µg/kg benzo(a)pyrene Peak:. Benzo(a)pyrene -208,000 0 2 4 6 8 0 2 4 Minutes Figure 4. Preserved ham samples.

Conclusion This work describes an on-line SPE HPLC method with fluorescence detection for a rapid and sensitive determination of benzo(a)pyrene in meat products. The determination was performed using an UltiMate 3000 Dual RSLC system controlled by Chromeleon CDS software. References. NY/T 666-2008: Determination of Benzo(a)pyrene in Meat Products, High Performance Liquid Chromatography Method; CNIS, Ministry of Agriculture of the People s Republic of China: Beijing, 2008. 2. SC/T 304-2008: Determination of Benzo(a)pyrene in Aquatic Products, High Performance Liquid Chromatography; CNIS, Ministry of Agriculture of the People s Republic of China: Beijing, 2008. 3. GB/T 5009.27-2003: Determination of Benzo(a) pyrene in Foods; CNIS, Standardization Administration of China (SAC), National Health and Family Planning Commission of the People s Republic of China: Beijing, 2003. 4. Dionex (now part of Thermo Scientific) Application Note 96: Determination of Polycyclic Aromatic Hydrocarbons (PAHs) in Edible Oils by Donor- Acceptor Complex Chromatography (DACC)-HPLC with Fluorescence Detection. Sunnyvale, CA, 2008. [Online] www.thermoscientific.com/content/dam/tfs/ ATG/CMD/CMD%20Documents/AN-96- Determination-PAH-Edible-Oil-LPN998.pdf (accessed April 28, 204). 5. Dionex (now part of Thermo Scientific) Application Note 23: Determination of Polycyclic Aromatic Hydrocarbons (PAHs) in Tap Water Using On-Line Solid-Phase Extraction Followed by HPLC with UV and Fluorescence Detections. Sunnyvale, CA, 2009. [Online] www.thermoscientific.com/content/dam/tfs/ ATG/CMD/CMD%20Documents/AN-23- Determination-PAH-Tap-Water-LPN228.pdf (accessed April 29, 203). 6. Thermo Scientific Application Note 085: Sensitive and Rapid Determination of Polycyclic Aromatic Hydrocarbons in Tap and River Waters. Sunnyvale, CA, 204. [Online] www.thermoscientific.com/ content/dam/tfs/atg/cmd/cmd%20documents/ Application%20&%20Technical%20Notes/ Chromatography/Liquid%20Chromatography/ AN-085-HPLC-PAHs-Tap-Water-AN70923-EN.pdf 7. Method 550: Determination of Polycyclic Aromatic Hydrocarbons in Drinking Water by Liquid-Liquid Extraction and HPLC with Coupled Ultraviolet and Fluorescence Detection; U.S. EPA, Environmental Monitoring and Support Laboratory: Cincinnati, OH, 990. 8. Method 550.: Determination of Polycyclic Aromatic Hydrocarbons in Drinking Water by Liquid-Solid Extraction and HPLC with Coupled Ultraviolet and Fluorescence Detection; U.S. EPA, Environmental Monitoring and Support Laboratory: Cincinnati, OH, 990. 9. Method 60: Polynuclear Aromatic Hydrocarbons; U.S. EPA, Environmental Monitoring and Support Laboratory: Cincinnati, OH, 990. Application Note 093 www.thermofisher.com/chromatography 2046 Thermo Fisher Scientific Inc. All rights reserved. ULTRA-TURRAX is a registered trademark of IKA-Werke GmBH & Co. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa +43 333 50 34 0 Australia +6 3 9757 4300 Austria +43 80 282 206 Belgium +32 53 73 42 4 Brazil +55 373 540 Canada + 800 530 8447 China 800 80 58 (free call domestic) 400 650 58 AN7069-EN 076S Denmark +45 70 23 62 60 Europe-Other +43 333 50 34 0 Finland +358 9 329 0200 France +33 60 92 48 00 Germany +49 603 408 04 India +9 22 6742 9494 Italy +39 02 950 59 Japan +8 6 6885 23 Korea +82 2 3420 8600 Latin America + 56 688 8700 Middle East +43 333 50 34 0 Netherlands +3 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00 Russia/CIS +43 333 50 34 0 Singapore +65 6289 90 Sweden +46 8 556 468 00 Switzerland +4 6 76 77 00 Taiwan +886 2 875 6655 UK/Ireland +44 442 233555 USA + 800 532 4752