A Comprehensive, Fast, and Sensitive Method for the Quantitation of Synthetic Cannabinoids using the Elute UHPLC Coupled to an EVOQ LC-TQ MS

Similar documents
Application Note LCMS-110 Development of a Targeted Quantitative LC-MS/MS Method for 431 Positive and Negative Ion Pesticides in a Single Analysis

TargetScreener. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

At-a-Glance. Verapamil C 27 H 38 N 2 O 4 M+H + =

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS

Application Note LCMS-112 A Fully Automated Two-Step Procedure for Quality Control of Synthetic Peptides

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

MRMS axelerate rapidly detected micropollutants and plant response metabolites in poplar leaves

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Highly sensitive quantitative estimation of genotoxic impurities from API and drug formulation using LC/MS/MS

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Proven robustness for large-scale metabolomics studies using the Phenomics Workhorse

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Highly sensitive quantitative analysis of Felodipine and Hydrochlorothiazide from plasma using LC/MS/MS

materials analysis Solutions for Your Analytical Business Markets and Applications Programs

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Application Note GCMS-03 Reliable high-throughput quantification of melamine and related analogs below regulatory limits in food using the EVOQ GC-TQ

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis

LC-MS/MS Analysis of Phytocannabinoids and their

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

Validation Report Determination of isoprothiolane residues in rice by GC-MS/MS (QuEChERS method)

Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS

Theory and Practical Consequences.

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Fast Screening and Quantitation of Perfluorinated Sources from Textiles using Chemical Ionization GC-MS

LC-MS/MS in the Clinical Laboratory. Jo Adaway

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Validation Report 18

Guide to Peptide Quantitation. Agilent clinical research

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Characterization of petroleum samples via thermal analysis coupled to APCI FTMS

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Simultaneous Analysis of 15 Synthetic Cannabinoids in Human Urine by Using Liquid-Chromatography Tandem-Mass Spectrometry

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis

Multi-residue analysis of pesticides in crude food extracts using a simple extraction technique and LC/MS/MS

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Benzodiazepines and Barbiturates

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Determination of 198 pesticide residues in eggplant using Gas Chromatography tandem Mass Spectrometry/Mass Spectrometry

Application Note FTMS-56 Reproducibility of Crude Oil Characterization by Flow Injection APPI-FT-ICR Mass Spectrometry

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Quantitative Analysis and Identification of Migrants in Food Packaging Using LC-MS/MS

Chemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography.

Appendix II- Bioanalytical Method Development and Validation

Agilent 6495 Triple Quadrupole LC/MS System EXPERIENCE A NEW LEVEL OF CONFIDENCE

Supporting Information. Sweetened Swimming Pools and Hot Tubs

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Tecan AC Extraction Plate

Uncontrolled Copy. Table of Contents. 1. Introduction and Principle of Assay Specimens Reagents and Materials...

Analyst Software. Automatic Optimization Tutorial

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Uncontrolled Copy. Table of Contents. 1. Principle of Assay Specimens Reagents and Materials... 3

Designed for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS

Quantitative analysis of mitragynine in human urine by high performance liquid chromatography-tandem mass spectrometry

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Agilent LC & LC-MS Cannabis Testing

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS

Transcription:

A Comprehensive, Fast, and Sensitive Method for the Quantitation of Synthetic Cannabinoids using the Elute UHPLC Coupled to an EVOQ LC-TQ MS Rapid and Sensitive Quantitation of Synthetic Cannabinoids in Serum by UHPLC-Triple Quadrupole Mass Spectrometry Abstract This study demonstrates a sensitive, rapid and reliable method for the simultaneous quantitation of 99 synthetic cannabinoids in serum using the Bruker Elute TM UHPLC coupled to the EVOQ Elite TM triple quadrupole MS. Sample preparation was performed with liquid-liquid extraction. The method was fully validated. Keywords: Synthetic cannabinoids, serum, quantitation Authors: Rafaela Martin 1, Jürgen Kempf 2, Laura M. Huppertz 2 1 Bruker Daltonik, Bremen 2 Institute of Forensic Medicine, Medical Center - University of Freiburg

Introduction Synthetic cannabinoids first appeared on the recreational drug scene in 2004. The term synthetic cannabinoid refers to an increasing number of manmade mind-altering chemicals that are either sprayed on dried, shredded plant material so they can be smoked, or sold as liquids to be vaporized and inhaled in e-cigarettes and other devices. Many synthetic cannabinoids are full agonists of the cannabinoid receptors CB 1 and CB 2, and therefore their psychoactive effects are similar to those of cannabis. They include a lot of different chemical structures such as e.g. naphthoyl-, phenylacetyl-, benzoyl- and cyclopropyl-indoles, cyclohexylphenoles, and various other indazole and indole derivatives (Figure 1). In an attempt to circumvent current legislation synthetic cannabinoids are misleadingly marketed via the internet or in headshops as bath salts, plant food or research chemicals and are often sold in colorful small bags with creative names, for example, Unicorn Magic Dust, Be Happy, or Aliens. They are then judged by consumers to be safe and legal, despite the fact that their effects can lead to high blood pressure, nausea, hallucination, psychosis, physical addiction and even life-threatening conditions. In order to keep ahead of evolving drug prohibition laws and modern detection methods, new or slightly modified variants of synthetic cannabinoids are regularly synthesized and rapidly marketed to the recreational drug user market. This means that analytical methods need to be regularly updated to also cover the newly emergent analytes. Unlike UHPLC-triple quadrupole mass spectrometry, traditional detection methods such as immunoassay and LC-UV do not have the required specificity, sensitivity or flexibility to cope with this ever-growing analytical challenge. This manuscript describes a UHPLC-triple quadrupole mass spectrometry method to rapidly and reliably quantify synthetic cannabinoids in human serum at the sub ng/ml level. R 3 O R 2 Fig. 1: Representative structures of synthetic cannabinoids; left: naphtoylindoles, right: indazole derivatives Liquid chromatography Instrument Column Mobile phase A Mobile phase B Gradient Flow rate Injection volume 10 µl Column oven 40 C Bruker Elute TM UHPLC H 2 O O H Experimental Kinetex C18 100A, 2.6 µm (100 x 2.1 mm) Sample Preparation 0.5 ml buffer (ph 10), 1.5 ml hexane:ethylacetate (99:1) and 10 µl internal standard mix were added to 1 ml serum. After mixing and centrifugation 1 % acetonitrile, 0.1 % formic acid with 2 mm ammonium formate 99 % acetonitrile, 0.1 % formic acid with 2 mm ammonium formate 0.0 1.0 min 20 % B 1.0 2.5 min to 60 % B 2.5 4.0 min to 65 % B 4.0 5.5 min 65 % B 5.5 8.0 min to 90 % B 8.0 10.0 min 90 % B 10.0 10.1 min to 20 % B 10.1 12.0 min 20 % B 500 µl/min

Mass Spectrometry Instrument EVOQ Elite TM triple quadrupole mass spectrometer Ion source VIP H-ESI positive, 4700 V Probe gas 50 units at 400 C Cone gas 25 units at 350 C and recovery, limits of detection and quantitation (LOD and LOQ), linearity, precision and accuracy. To determine selectivity, blank serum from 10 individual donors was analyzed without the addition of analytes or internal standards and two blank serum samples with the addition of internal standards. ebulizing gas 50 units Active exhaust on Collision gas Argon, 1.5 mtorr MRM transitions 2 per analyte, 1 per IS 1 ml supernatant was transferred to an HPLC vial. The residue was mixed with 1.5 ml hexane:ethylacetate (80:20) and 1 ml supernatant was transferred to the first HPLC vial after centrifugation. The combined supernatants were evaporated and the residue reconstituted in 100 µl eluent A:B (50:50). Instrumentation Method validation The method was validated according to the guidelines of the German Society of Toxicological and Forensic Chemistry (GTFCh). Validation parameters included selectivity, matrix effects To examine matrix effects and recovery, five different serum samples from five different individuals spiked before extraction, five additional serum samples spiked after extraction and five neat standards were measured at a low (0.1 ng/ml) and a high concentration (1 ng/ml) and compared. The matrix effects and the recovery were then calculated according to Matuszewski et al 1. LOD and LOQ were evaluated from calibration curves constructed with equidistant calibrators in the range of Fig. 2: MRM builder; top: define compound name, precursor and number of product ions desired; bottom: results of optimization with optional export to method and/or user library

the expected LOD (5 to 50 pg/ml) and calculated according to DI 32645 2. To demonstrate linearity, six calibration curves with seven calibrators each were constructed, ranging from 50 pg/ml up to 1.25 ng/ml. Precision and accuracy were determined by the analysis of two replicates of low (0.05 ng/ml), medium (0.25 ng/ml) and high (1 ng/ml) QC samples on eight consecutive days. Results and Discussion For the development of the MRM method, the MRM builder tool was used (Figure 2). Optimization was performed by infusing mixtures of standard solutions of the analytes. 7 6 WI 48098 MEPIRAPIM AM-2233 AM-1220 JWH-200 A-796,260 AB-005 AB-PIACA 5F M-25 ADBICA-5F 5 4 kcps 3 2 1 0 2.25 2.50 2.75 3.00 3.25 minutes Fig. 3: Chromatogram of the first ten eluting compounds at 5 pg/ml

Curve Fit: Linear, Ignore, 1/X2 Resp. Fact. RSD: 3.905% by the use of an appropriate internal standard. Recovery was >50% for 79 analytes. 2.5 2.0 y = +2.296994x -0.002876 JWH-018 Coeff. Det.(r2):0.998365 The LOD was <5 10 pg/ml for 89 analytes with the highest LOD being 50 pg/ml. Figure 3 shows a chromatogram of representative analytes at the concentration of 5 pg/ml. The calculated LOQ was 30 pg/ml for 94 of the 99 analytes and was finally defined as the lowest calibrator concentration of 50 pg/ml for all compounds. Peak Size / PS Std. 1.5 1.0 0.5 For calibration curves a weighting factor of 1/x² was used. All analytes followed a linear calibration model (Figure 4), with a bias (accuracy) within ±15% being obtained. Though for six analytes the average bias met the validation criteria, single values exceeded the ±15% range, so findings for these six analytes are given as semi-quantitative results only. For interday and intraday precision all analytes fulfilled the validation criteria with RSD s of <15%. 0.0 Conclusions 0.25 0.50 0.75 1.00 1.25 Amount Std. (ng/ml) Fig. 4: Calibration curve of JWH-018 The MRM builder identified the optimal transitions and collision energies as Compound Based Scanning). points per peak (generally referred to for each analyte which were then directly exported to the method and The chromatographic separation of the also added to a user library for future 99 analytes was performed within 8.5 reference. Similarly, 99 synthetic cannabinoids and 17 deuterated internal utes by the new Elute UHPLC system. minutes with a total runtime of 12 min- standards were easily added to the method. Method validation As there were no interfering signals at the ion transitions of the ana- Scan times for each analyte were automatically calculated after defining the lytes, the method was deemed to be retention times, and the retention time selective. Matrix effects were within windows, as well as the average peak ± 25% for 89 of the 99 analytes. Higher width and required number of data matrix effects can be compensated A method for the sensitive quantitation and/or the semiquantitative determination of 99 synthetic cannabinoids including recently emerged substances has been developed using the Bruker EVOQ LC Triple Quad system. The method is easily customizable using the MRM builder to add new compounds to the user library. The low limits of detection obtained provide the necessary evidence to show recent use of synthetic cannabinoids in forensic cases such as driving under the influence of drugs or in post-mortem cases.

Bruker Daltonics is continually improving its products and reserves the right to change specifications without notice. Bruker Daltonics 02-2017, LCMS-123, 1849846 Learn More You are looking for further Information? Check out the Link or scan the QR Code. www.bruker.com/literature-elute References (1) Matuszewski, M.L. Constanzer, C.M. Chavez-Eng, Anal. Chem., 2003, 75 (13), 3019-3030 (2) Deutsches Institut für ormung. DI 32 645 achweis-, Erfassungs- und Bestimmungsgrenze, Beuth Verlag: Berlin, 1994 For research use only. ot for use in diagnostic procedures. Bruker Daltonik GmbH Bremen Germany Phone +49 (0)421-2205-0 Fax +49 (0)421-2205-103 Bruker Daltonics Inc. Billerica, MA USA Phone +1 (978) 663-3660 Fax +1 (978) 667-5993 ms.sales.bdal@bruker.com www.bruker.com