Determination of Elemental Sulfur in Explosives and. Explosive Residues by Gas Chromatography-Mass Spectrometry. Kimberly S.

Similar documents
JWH-018 and JWH-022 as Combustion Products of AM2201

GC-CI-MS analysis of TMS derivatives

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

NCFS Sample Preparation and Instrumental Parameters

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

5-METHOXY-N,N-DIISOPROPYLTRYPTAMINE Latest Revision: June 20, 2005

Sulfotepp impurities in Chlorpyrifos EC formulations

SECOBARBITAL Latest Revision: February 15, 1999

Author. Abstract. Introduction

5-METHOXY- -METHYLTRYPTAMINE Latest revision: June 20, 2005

Determination of Volatile Substances Proof of Food Adulteration

Analysis of Pressure Sensitive Adhesives by GC/MS and GC/AED with Temperature Programmable Pyrolyzer

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Selective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of 2,2 Dinitrobiphenyl with TiO 2 and UV light irradiation

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

The Focus Robotic Sample Processor as a Tool for the Multiple Analysis of Samples using Complementary Techniques

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Multi-residue analysis of pesticides by GC-HRMS

Chapter 27: Gas Chromatography

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Report AFK0242/18 TABLE OF CONTENTS

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

DEVELOPMENT AND VALIDATION OF GC-FID METHOD FOR THE DETERMINATION OF ETHANOL RESIDUE IN MARJORAM OINTMENT

ANALYTICAL REPORT. 25B-NBOH (C17H20BrNO3) 2-({[2-(4-bromo-2,5-dimethoxyphenyl)ethyl]amino}methyl)phenol RSUNJYKZRKIBNB-UHFFFAOYSA-N

Chromatography. Gas Chromatography

BRIEFING. Pharmacopeial Discussion Group Sign Off Document Attributes EP JP USP Definition Loss on drying Readily carbonizable substances

Volatile organic compounds (VOCs):

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Spectra Library Electron Capture Negative Ion (ECNI) Mass Spectra of Selected Polybrominated Diphenyl Ethers (PBDEs)

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

U.S. EPA Method 8270 for multicomponent analyte determination

Chemistry Gas Chromatography: Separation of Volatile Organics

Analysis of Sulfur-Containing Flavor Compounds By GC/MS With A PFPD

STANDARD OPERATING PROCEDURES

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Rapid Post-Blast Inorganic Explosive Analysis by Suppressed Ion Chromatography

Analyst

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Xinyu LIU, Peter JOZA, Andrew MASTERS, Bill RICKERT

SPE AND GC MS INVESTIGATION OF ORGANIC CONTAMINANTS IN ATMOSPHERIC PRECIPITATION

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

1-(2-METHOXYPHENYL)PIPERAZINE Latest revision: June 27, 2005

BUFOTENINE Latest Revision: August 16, 2005

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Supplementary Materials: Model-based peak alignment of metabolomic profiling from comprehensive two-dimensional gas chromatography mass spectrometry

Agilent G3212 GC-APCI Source

GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL

GAS CHROMATOGRAPHY (GC)

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Deposition and Detection System. Trace Level Analysis of Chemical Warfare Agents (CWAs) Related Chemicals by GC-IR Solid Phase System

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010

Methyl 1-[(4-fluorophenyl)methyl]indazole-3-carboxylate

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

4-methyl-1-phenyl-2-pyrrolidin-1-yl-pentan-1-one

Expectations for GC-MS Lab

Selection of a Capillary GC Column

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Determination of Butyltin Compounds in Environmental Samples by Isotope Dilution GC-MS. Application. Authors. Abstract. Introduction.

TERTIARY BUTYLHYDROQUINONE

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

ADVANTAME (TENTATIVE)

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

2,5-DIMETHOXY-4-IODOPHENETHYLAMINE Latest Revision: August 22, 2005

Analyzing Semi-Volatiles by GC-MS

Selection of a Capillary

Benzoylecgonine in Urine by SAMHSA GC/MS

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

This method describes the identification of the following prohibited colorants in cosmetic products:

ISSN: ; CODEN ECJHAO E-Journal of Chemistry , 7(2),

Shodex TM ODP2 HP series columns

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

2-[[2-(4-ethyl-2,5-dimethoxy-phenyl)ethylamino]methyl]phenol

4-Methyldiethcathinone (4-MDEC)

N-(1-carbamoyl-2-methyl-propyl)-1-(5-chloropentyl)indazole-3- carboxamide

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

BENZYLPIPERAZINE Latest Revision: June 1, 2005

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

Introduction to Capillary GC

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Transcription:

Determination of Elemental Sulfur in Explosives and Explosive Residues by Gas Chromatography-Mass Spectrometry Kimberly S. Bradley Illinois State Police, Springfield Forensic Science Laboratory 24 Hill Meadows Drive, Springfield, IL 6272 Abstract A method for the positive identification of elemental sulfur in explosives and explosive residues is presented. Following a carbon disulfide wash of explosives or explosive residues, a sample of the extracted material is injected onto a gas chromatography (GC) column, then analyzed via mass-selective (MS) detection. A positive identification of elemental sulfur is based on both retention time and fragmentation pattern. The GC-MS method is demonstrated to be useful in detecting and positively identifying elemental sulfur from both burned and unburned explosive mixtures. With a detection limit of 2.5 ng of elemental sulfur on the column, this method is 4 times more sensitive than the presumptive chemical color test that has previously been the method employed for the detection of small amounts of sulfur.(j. Forensic Sci.,25:5,96-13.) Introduction Elemental sulfur is commonly used as a fuel component in inorganic explosive mixtures such as black powder, flash powder, and a variety of black powder substitutes. Consequently, residues from such explosive mixtures also contain elemental sulfur. When analyzing these residues, positive identification of elemental sulfur, in conjunction with other characteristic reaction products, is essential in determining the composition of the originating explosive mixture. In many instances, however, only a very small amount of elemental sulfur is recovered, thereby limiting the options of analysis technique for positive identification. Currently, the most 1

commonly used techniques are X-ray diffractometry (XRD), energy dispersive X-ray analysis (EDX), and a chemical color test. Although both XRD and EDX require relatively small amounts of solid sample for analysis, it is often much greater than what can be isolated from an explosive residue. Thus, the chemical color test which requires far less sample than XRD or EDX is frequently utilized. In 21, the Technical Working Group for Fire and Explosions (TWGFEX) released an action document (1) suggesting that the chemical color test for sulfur is insufficient for the positive identification of elemental sulfur in explosives and explosive residues. This means that the chemical color test cannot be the sole analytical technique used to determine the presence of sulfur. Consequently, an alternative method for the positive identification of elemental sulfur was sought. In this study, which has been published previously (2), GC-MS is shown to be capable of positive identification of elemental sulfur in explosives and explosive residues, and to be 4 times more sensitive for sulfur than the chemical color test. Methods GC-MS analyses were performed on an Agilent 689 GC coupled with an Agilent 5973 MS detector. A 25 m Agilent HP-1 capillary column (cross-linked methyl siloxane,.2 mm I.D.,.5 :m film thickness) was used with helium as the carrier gas, with a flow rate of 1. ml/min. Several combinations of GC-MS settings were tested in order to determine an optimized method for sulfur analysis, resulting in the following: injector port, 25 C; MS interface, 28 C; MS source, 2 C; initial column T, 23 C; initial hold time, 6 min; T ramp, 1 C/min to 25 C; final hold time, 2 min; split ratio, 4:1; MS scan mode, electron ionization; ionization energy, 7eV; solvent delay, 2 min; MS scan range, 3-33m/z; total run time, 1 2

min; retention time of elemental sulfur peak, 5.3 min. Samples were analyzed for "-S 8, the most thermodynamically stable allotrope of elemental sulfur. (3) Test samples were prepared by performing a carbon disulfide (CS 2 ) wash of the explosive or explosive residue. The CS 2 was then allowed to evaporate completely, in order to confirm insufficient material to perform other instrumental tests for sulfur. For GC-MS analysis, the CS 2 extract was dissolved in ~.5 ml CS 2, and then transferred to a shell vial. This solution was then evaporated to a small volume (~5 µl), and a 1 µl sample was injected into the GC-MS. Positive identification of sulfur was based on retention time and identification of the fragmentation pattern for "-S 8. Fragmentation patterns were compared to an in-house reference for "-S 8 and to references in the Wiley database (4), using a probability-based matching search routine. The search result statistics include a value called match quality, which in general refers to the probability that the sample spectrum has been identified correctly. Match quality values greater than 9 are considered very good matches, while values less than 5 suggest substantial differences between the sample and the reference spectra (5). For this study, a positive identification of sulfur required a match quality value of 9 or greater. For chemical color test analysis, the CS 2 extract was dissolved in ~.5 ml pyridine, swirled gently by hand to mix, and then warmed on a hot plate at ~15 C for ~1 min. Then one drop of 2 N NaOH was added, and any color change was noted. A positive identification of sulfur was based on a color change to any hue ranging from very pale blue to deep blue-greenish brown. (6) Results and Discussion Identification of Elemental Sulfur from Explosives and Explosive Residues In order the investigate the potential use of GC-MS as a method of detecting and positively identifying elemental sulfur as part of explosives analysis, samples of unburned and 3

burned explosive mixtures were examined. The two explosive mixtures chosen for this study were a flash powder (composed of potassium chlorate, sulfur, and aluminum) and Pyrodex (a black powder substitute composed of potassium perchlorate, potassium nitrate, sulfur, carbon, sodium benzoate, and cyanoguanidine). Samples of unburned flash powder (2 mg) and Pyrodex (5 mg), and samples of burned flash powder (32 mg before burning) and Pyrodex (3 mg before burning), were extracted with CS 2 and processed as described previously. In no instance was there sufficient extracted material to perform XRD or EDX analysis. Figure 1 presents data for Pyrodex only. Figures 1(A) and 1(C) are the total ion current chromatograms (TICs) for unburned and burned Pyrodex. For unburned Pyrodex, there is one large peak with a retention time of ~5.3 min, and its fragmentation pattern is shown in Figure 1(B). The peak is positively identified as "-S 8, with a match quality value of 96. For burned Pyrodex, there is a small peak at ~5.3 min that is superimposed upon a broad, noisy background envelope of peaks. The fragmentation pattern for this peak is shown in Figure 1(D). Although this fragmentation pattern also contains many contributions from the background envelope of peaks, fragments indicative of sulfur are clearly visible, and "-S 8 is positively identified with a match quality value greater than 9. Following analysis by GC-MS, the remaining extracted material from each sample was tested for elemental sulfur using the chemical color test. The chemical color test confirmed the presence of sulfur in every sample. It should be noted that the fronting of the peak at ~5.3 min in the TIC for unburned Pyrodex is commonly seen when relatively large amounts of sulfur are loaded onto this particular column. However, the fronting does not interfere with the ability to positively identify "-S 8. Diluting the sample would result in minimization of fronting. 4

Detection Limits Detection limits for both the chemical color test and the GC-MS method were determined experimentally. In each case, the same standard sulfur stock solution was used. This stock solution was created by adding 5. mg precipitated sulfur powder to 25. ml CS 2. The resulting solution has a concentration of 7.8 1-4 M "-S 8 (2 ng "-S 8,/µL solution). The detection limit for the chemical color test for sulfur was determined using the sample preparation procedure as indicated previously. The blank control was a clean shell vial that was processed in the same manner as all of the trials, and the sulfur control was a known, but not quantified, sample of sulfur that was also processed in the same manner as all of the trials. All experimental observations were witnessed and confirmed by another analyst, in addition to the author. Table 1 summarizes the results, including the amounts of sulfur used and any observed color change. As indicated in the table, the smallest amount of sulfur that gave a positive result was 1 µg. Trials with.8 µg and.6 µg were inconclusive, meaning that either one or both of the analysts observing the experiment were unable to detect a definitive color change. Thus, the detection limit for the chemical color test for sulfur was determined to be 1 µg. The detection limit for the GC-MS method for sulfur was determined by injecting various amounts of the standard sulfur stock solution in order to deliver 5 ng, 2.5 ng, and 1 ng of "-S 8 onto the column. The resulting TICs are shown in Figure 2. Wiley database searches were performed for the peak at ~5.3 min in each TIC. The peaks in both the 5 ng and 2.5 ng trials were positively identified as "-S 8, while the peak in the 1 ng sample was too weak to be positively identified. Thus, the detection limit for the GC-MS method for sulfur was determined to be 2.5 ng, making it 4 times more sensitive for elemental sulfur than the chemical color test. 5

The GC-MS detection limit for sulfur could be decreased even further by changing the instrument parameters and extending the total runtime. However, for the purpose of explosives analysis, a compromise between total runtime and detection limit was made. Reproducibility The reproducibility of the analytical method was tested by injecting 1. µl of the standard sulfur stock solution five times. For the five trials, the peak retention time ranged from 5.327 5.331 min, the abundance of the peak ranged from 33, 43, counts, and the minimum match quality was 94. These data demonstrate that the sulfur peak consistently appears at a retention time of ~5.33 min, and that it is consistently positively identifiable by the Wiley database search criteria. Although explosives analysis typically only requires identification of elemental sulfur and not quantification, it is interesting to note statistics regarding the abundance of the sulfur peak in each case. For the five trials, the mean for the abundance was 38,2 counts, and the standard deviation of the population was 3,22 counts. The relative standard deviation was found to be 8%. Advantages and Limitations The primary advantages of using GC-MS to identify elemental sulfur from explosives and explosive residues include the following: 1. It is an instrumental technique that does not require the human eye for detection, as is the case for the chemical color test. 2. It is a very sensitive technique. The GC-MS method as reported here is capable of detection and positive identification of as little as 2.5 ng of sulfur on the column, making it 4 6

times more sensitive than the chemical color test, which has a detection limit of 1 µg of sulfur. 3. It is a conclusive technique which allows the collection and archival of hard copy data to support or refute the positive identification of sulfur. 4. GC allows the separation of other compounds that may be present in the sample, and that can interfere with the chemical color test (e.g. organic dyes). Therefore, inconclusive results due to interference are minimized. The only limitation of using GC-MS to identify sulfur from explosives and explosive residues that was uncovered during this study involved analysis time. The total time required for casework identification of sulfur (i.e. runs for sensitivity check, blank, and sample) is approximately two to three times longer than that for the chemical color test in instances where the chemical reagents necessary for the color test are already prepared. However, it can be nearly comparable in instances where fresh chemical reagents for the color test must be prepared. Moreover, in reference to the total explosives analysis procedure, which typically involves both a water and a CS 2 extraction of the residue, the GC-MS analysis for sulfur can easily be performed while waiting for the water extract to evaporate. Consequently, the use of GC-MS will not increase the total time necessary for a complete explosives case analysis. Conclusion This study has demonstrated GC-MS to be a very promising tool to replace the chemical color test in the detection and positive identification of elemental sulfur from explosives and explosive residues. Its use is recommended for cases in which there is insufficient material to perform XRD or EDX analysis. The GC-MS method is efficient, conclusive, and 4 times more sensitive for sulfur than the chemical color test. 7

Acknowledgements The author expresses appreciation to Mr. George Dabdoub, Ms. Julie Glasner, Mr. Blair Schultz, and the Illinois State Police Forensic Sciences Command for their assistance and support of this work. References 1. Recommended Guidelines for Forensic Identification of Intact Explosives, August 21 Revision, http://www.twgfex.org. 2. Bradley, K.S. Determination of Elemental Sulfur in Explosives and Explosive Residues by Gas Chromatography-Mass Spectrometry. J. Forensic Sci. 25;5:96-13. 3. Miessler, G.L.; Tarr, D.A. Inorganic Chemistry. Englewood Cliffs (NJ): Prentice Hall, 1991. 4. McLafferty, F. Wiley Registry of Mass Spectral Data, 7 th Edition. Software version. Hoboken (NJ): John Wiley & Sons, May 2. 5. Productivity Tools [help screen, computer program] in Enhanced ChemStation software. Version C... Palo Alto (CA): Agilent, 1999. 6. Parker, R.G.; Stephenson, M.O.; McOwen, J.M.; Cherolis, J.A. Analysis of Explosives and Explosive Residues. Part 1: Chemical Tests. J. Forensic Sci. 1975;2:133-14. Table 1. Data from determination of detection limit of chemical color test for elemental suflur. Amount of Sulfur Present Observations µg Negative; no color 1 µg Positive; blue 2 µg Positive; light blue 1 µg Positive; very pale blue.8 µg Inconclusive.6 µg Inconclusive.4 µg Negative; no color 8

25 A 6 B 64 2 15 1 5 4 2 96 128 16 192 256 2 3 4 5 6 7 8 9 1 Retention Time (min) 12 C 8 4 2 3 4 5 6 7 8 9 1 Retention Time (min) 5 1 15 2 25 3 m/z 4 35 64 D 256 3 128 16 25 2 192 15 96 1 5 5 1 15 2 25 3 m/z Figure 1. (A) TIC for sample solution from CS 2 extraction of unburned Pyrodex. (B) Fragmentation pattern for peak at retention time of 5.3 min in (A). (C) TIC for sample solution from CS 2 extraction of burned Pyrodex. (D) Fragmentation pattern for peak at retention time of 5.3 min in (C). 9

12 1 A 8 6 4 2 2 3 4 5 6 7 8 9 1 Retention Time (min) 12 1 B 8 6 4 2 2 3 4 5 6 7 8 9 1 Retention Time (min) 12 1 C 8 6 4 2 2 3 4 5 6 7 8 9 1 Retention Time (min) Figure 2. TICs for (A) 5 ng, (B) 2.5 ng, and (C) 1 ng of sulfur on the column. 1