Chapter 2. Interaction with Soft Tissue

Similar documents
Doppler echocardiography & Magnetic Resonance Imaging. Doppler echocardiography. History: - Langevin developed sonar.

EE 5345 Biomedical Instrumentation Lecture 6: slides

Bioengineering 280A Principles of Biomedical Imaging. Fall Quarter 2006 Ultrasound Lecture 1

A small object is placed a distance 2.0 cm from a thin convex lens. The focal length of the lens is 5.0 cm.

Sound wave bends as it hits an interface at an oblique angle. 4. Reflection. Sound wave bounces back to probe

Structure of Biological Materials

Basic System. Basic System. Sonosite 180. Acuson Sequoia. Echo occurs at t=2z/c where c is approximately 1500 m/s or 1.5 mm/µs

Unit 4 Parent Guide: Waves. What is a wave?

Physics and Knobology

D. BARD DIVISION OF ENGINEERING ACOUSTICS, LUND UNIVERSITY

Waves Review Checklist Pulses 5.1.1A Explain the relationship between the period of a pendulum and the factors involved in building one

Nicholas J. Giordano. Chapter 13 Sound

Signal Loss. A1 A L[Neper] = ln or L[dB] = 20log 1. Proportional loss of signal amplitude with increasing propagation distance: = α d

I have nothing to disclose

ULTRASOUND. Ultrasound physical phenomenon properties basics of medical applications, History. History

Chapter 16 Traveling Waves

Technical University of Denmark

Medical Imaging Physics Spring Quarter Week 3-2

School. Team Number. Optics

Quest Chapter 29. turned back think bounce. Reread the definition of the law of reflection on page 444.

SECTION A Waves and Sound

Light and Sound. Some questions. All waves can... Why is the ocean blue? How would you hide a submarine?

Polarization. If the original light is initially unpolarized, the transmitted intensity I is half the original intensity I 0 :

P5 Revision Questions

ISUOG Basic Training The Principles of Doppler Ultrasound. Juriy Wladimiroff

The physics US and MRI. Prof. Peter Bogner

Light.notebook May 03, 2016

Optics in a Fish Tank Demonstrations for the Classroom

Chapter 1. The Nature of Sound. The Nature of Sound

LECTURE 1: ELECTROMAGNETIC RADIATION

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS

During part of the journey the car is driven at a constant speed for five minutes.

Chapter 33: ELECTROMAGNETIC WAVES 559

Waves ~ Learning Guide

Name Date Class _. Please turn to the section titled The Nature of Light.

Physics Test Pack WALCH PUBLISHING

Supplement (videos)

CHAPTERS: 9.1, 10.1 AND 10.2 LIGHT WAVES PROPERTIES

Ch Guided Reading Sound and Light

Physics General Physics. Lecture 25 Waves. Fall 2016 Semester Prof. Matthew Jones

Light as a Transverse Wave.

The Physics of Doppler Ultrasound. HET408 Medical Imaging

Name: Section: Forms of Energy Practice Test

General Physics (PHY 2130)

AP Waves/Optics ~ Learning Guide

Today s menu. Last lecture. Ultrasonic measurement systems. What is Ultrasound (cont d...)? What is ultrasound?

Department of Engineering Science. Robin Cleveland

General Physics (PHY 2130)

Rad T 290 Worksheet 2

EL-GY 6813/BE-GY 6203 Medical Imaging, Fall 2016 Final Exam

(INCLUDING THIS FRONT PAGE)

31545 Medical Imaging systems

GOZO COLLEGE. Half Yearly Examinations for Secondary Schools FORM 4 PHYSICS TIME: 1h 30min

Mandatory Assignment 2013 INF-GEO4310

Outline of today Medical Imaging systems. Wave types. 1. Discussion assignment on B-mode imaging

UNIVERSITY OF SOUTHAMPTON

Donald School Journal of Ultrasound Physical Bases in Obstetrics of Medical and Ultrasound Gynecology, April-June 2009;3(2):1-9

PHYSICS. Chapter 16 Lecture FOR SCIENTISTS AND ENGINEERS A STRATEGIC APPROACH 4/E RANDALL D. KNIGHT Pearson Education, Inc.

Waves ~ Learning Guide

Chapter 20: Mechanical Waves

Physics 202 Quiz 1. Apr 8, 2013

Angular Spectrum Decomposition Analysis of Second Harmonic Ultrasound Propagation and its Relation to Tissue Harmonic Imaging

Plasma Physics Prof. V. K. Tripathi Department of Physics Indian Institute of Technology, Delhi

Topic 4: Waves 4.3 Wave characteristics

Sixth Grade Science BLOCK 3 ASSESSMENT Heat, Light, and Sound

Today s menu. Last lecture. Measurement of volume flow rate. Measurement of volume flow rate (cont d...) Differential pressure flow meters

Wave Motion and Sound

PHYSICS 220. Lecture 21. Textbook Sections Lecture 21 Purdue University, Physics 220 1

1. Waves and Particles 2. Interference of Waves 3. Wave Nature of Light

PHYSICS. Ray Optics. Mr Rishi Gopie

SECTION 3 & 4 LIGHT WAVES & INFORMATION TRANSFER

Downloaded from

3. Which of the following statements is (are) TRUE about detector crystals in Anger cameras?

Physics 142 Mechanical Waves Page 1. Mechanical Waves

Properties of Waves. Before You Read. What are the features of a wave?

Revision checklist SP4 5. SP4 Waves. SP4a Describing waves. SP4b Wave speeds. SP4c Refraction

Corso di Laurea in Fisica - UNITS ISTITUZIONI DI FISICA PER IL SISTEMA TERRA. Wave propagation FABIO ROMANELLI

Electrical Engineering 3BA3: Structure of Biological Materials

BNG/ECE 487 FINAL (W16)

Acoustic Shadowing Due to Refractive and Reflective Effects

Lab #13: Polarization

Chemistry Instrumental Analysis Lecture 2. Chem 4631

10. A Kelvin thermometer and a Fahrenheit thermometer both give the same reading for a certain sample. The corresponding Celsius temperature is: A)

Physics Mechanics. Lecture 34 Waves and sound II

3/9/2011. Outline Chapter 7 Waves Water Waves Water Waves. Water waves are really circular. They are an example of Mechanical waves.

Physics 30: Chapter 5 Exam Wave Nature of Light

CHAPTER 4 BASICS OF ULTRASONIC MEASUREMENT AND ANFIS MODELLING

Technical University of Denmark

Chapter 12 Sound in Medicine

0.4 s 0.8 s 1.5 s. 2.5 s. 2. A beam of light from a ray box spreads out as shown in the diagram and strikes a plane mirror.

The diagram below. to the by the. outlet into. calculation. Since TRANSMISSION VIA STRUCTURE. Vibration Via Supports Duct Breakout

General Physics II Summer Session 2013 Review Ch - 16, 17, 18

Revision checklist. Step Learning outcome Had a look Nearly there Nailed it! Step Learning outcome Had a look Nearly there Nailed it!

Physics Spring 2009 Midterm #1 Solution

Physics 210 Medical Physics Midterm Exam Winter 2015 February 13, 2015

Decomposing white light into its components:

Fineman CP Physics Final Study Guide

Some properties of waves: Huygens principle Superposition Coherence Interference Young s double-slit experiment Thin-film interference

We have seen how the Brems and Characteristic interactions work when electrons are accelerated by kilovolts and the electrons impact on the target

Transcription:

Chapter 2 Interaction with Soft Tissue Ultrasound interacts with human soft tissue in several predictable ways that allow engineers to design instruments that provide diagnostic information that forms the basis of medical sonography. These parameters include: 1. Attenuation 4. Diffraction 2. Reflection 5. Divergence 3. Refraction 6. Interference 1. Attenuation (decibels, db) Attenuation is the decrease in amplitude and intensity of a sound wave as it travels through a medium. A person standing at the front of a large room speaking in a normal tone (amplitude) of voice will typically be heard by people close to the speaker. People at the back of the room, however, will have a hard time hearing the speaker because sound energies are attenuated as they travel through the air. In ultrasonic imaging the energy carried by the wave is converted into minute amounts of heat. Approximately 80% of the energy that is lost is converted into heat by the process of absorption. The other 20% is lost through other attenuative processes such as reflection and scattering that will be discussed in this chapter. Reflection provides useful echo data for producing two-dimensional sonographic images. Scattering generally dissipates imaging energies. It does, however, provide information that can be utilized in Doppler displays and it contributes to the display of tissue textures in solid parenchymal organs. In general, however, any pathologic change that increases scattering in a structure will most likely result in a loss of information available for high-quality imaging. Clinical examples of significant energy loss due to increased scattering can be found in sonographic examination of a fatty liver or a fibroid uterus. UNITS OF MEASUREMENT: db (decibels) Note: Units for attenuation must be negative since the intensity is DECREASING. Attenuation Coefficient The attenuation coefficient is a number that refers to the amount of energy lost per unit of travel, or per centimeter. In measuring attenuation it is important to remember that it is measured as a round trip event; energies are lost both on the first pass into tissue and also on the return trip by the echoes. 23

UNITS OF MEASUREMENT: db/cm The total amount of attenuation is dependent upon both the frequency of the transducer (MHz) the distance traveled (cm) as well as the characteristics of the tissue through which the sound is traveling. As the frequency of the transducer increases, attenuation also increases. FREQUENCY ATTENUATION FREQUENCY ATTENUATION For example, relative to clinical practice, a 5.0 MHz transducer, while producing very high-resolution images, cannot penetrate deeply into the body. Because ultrasound energies are attenuated more rapidly, less is available for imaging deeper. Low frequency probes, on the other hand, penetrate deeply because there is less attenuation per unit of distance (cm) but the overall resolution of the image is poorer. Another example of the relationship between attenuation and frequency can be found in daily life. People living in apartment houses with noisy neighbors who enjoy playing their stereos loudly can hear low frequency sounds (bass) because these energies travel much more effectively than the high-frequency (treble) sounds. Low frequencies penetrate the walls and floors better than the high frequencies. Attenuation is a significant factor in determining which transducer should be selected for various sonographic examinations. This is because frequency of the sound beam places limits on the depths that can be effectively and adequately imaged. The other factor which directly influences the attenuation coefficient is the distance traveled (cm). As this distance increases, the attenuation coefficient also increases. : DEPTH ATTENUATION DEPTH ATTENUATION A good approximation of attenuation is described in the following formula: FORMULA: A = 0.5 db/cm/mhz (one way) = 1.0 db/cm/mhz (round trip) EXAMPLE: Calculate the overall attenuation of a 3.5MHz ultrasound beam traveling through 5 cm of soft tissue (eliminate all units for ease of calculation): A = 0.5dB x 5cm x 3.5MHz = 8.75 db (one way) = 8.75 x 2 (both ways) = 17.5 db (round trip) 24

The exact composition of human tissue type also affects the loss of ultrasound energies during the imaging process. This loss is the result of reflection, scattering, refraction and absorption. Absorption is the conversion of ultrasound into heat whereas the other processes refer to the redirection of the sound beam from its original trajectory. There is very little difference in the total attenuation of sound in what arerefered to as human soft tissues ( soft indicating that the tissue is not hard such as bone). In the chart below, it can be seen that there is only a 0.5dB/cm difference between fat and heart tissue, while there is a full 19dB/cm difference in attenuation between bone and fat. This difference is largely because dense bone absorbs nearly all of the acoustic energy impinging on it and converts most of it into heat. Attenuation Coefficients Tissue A.C (db/cm) Water 0.0022 Blood 0.18 Fat 0.6 Brain 0.6 Liver 0.7 Kidney 0.9 Muscle 1.0 Heart 1.1 Bone 20.0 Lung 41.0 Acoustic Impedance (Rayl) Acoustic impedance is the resistance to travel that a sound beam encounters as it passes through a medium, such as human soft tissue. Just as velocity is dependent on density, so is acoustic impedance. As the density of a material increases, so does acoustic impedance. Therefore, acoustic impedance is directly proportional to tissue density. DENSITY ACOUSTIC IMPEDANCE DENSITY ACOUSTIC IMPEDANCE 25

In human soft tissue, acoustic impedance is defined as the product of the density and velocity. Z = DV FORMULA: Where: Z = acoustic impedance D = density of medium V = velocity of sound UNITS OF MEASUREMENT: Rayl Defining rayl simply can be, at best, difficult. When sound passes through a medium, the pressure of the incident wave causes the particle within the medium to move (oscillate). The impedance, or resistance, to the passage of sound is the ration between the pressure and the particle velocity it produces. The impedance is 1 rayl if unit pressure produces unit velocity. Two ways of stating this relationship (one using the meter-kilogram-second method, the other using the centimeter-gramsecond method) are as follows: 1 rayl = 1 paschal-second/meter = Pa s/m = 1 newton-second/cubic centimeter = N s/cm 3 Slight differences in acoustic impedance between two tissues create an interface that will cause a portion of the ultrasound wave to be reflected. These acoustic mismatches are the basis of the echoes that are used in creating all twodimensional sonographic images. The larger the acoustic mismatch, the greater will be the intensity of the reflected portion of the beam. The acoustic impedances of some common tissues are summarized below: Typical Impedances Tissue Impedance (Rayl X 10-5 ) Air 0.0004 Fat 1.38 Water 1.48 Blood 1.51 Kidney 1.62 Soft tissue (average) 1.63 Liver 1.65 Muscle 1.70 Bone 7.8 26

This chart demonstrates that the difference in values for acoustic impedance is practically non-existent in most of the body's tissues and, typically, less than 1% of the incident sound beam will be reflected from any given interface. The exceptions are bone and air. The large difference in acoustic impedance is the reason that bony or air-filled structures cannot be imaged. For example, in scanning the lung, highly reflective surfaces can be seen on the screen, but no information will be displayed provided beyond that point. That is because approximately 98% of the ultrasound beam is reflected due to the high magnitude of difference in acoustic impedance between the air-containing lung and the intervening skin, musculature and soft tissue. 2. Reflection Ultrasound images depend upon a balance between reflection and transmission of the sound beam, or, more specifically the ultrasonic pulse. As the pulsed beam enters the human body, it encounters tissues of varying, or mismatched, acoustic impedances. When it encounters such an impedance mismatch, a portion of the incident beam is reflected or sent back toward the transducer. These reflected energies are called echoes and are detected, analyzed, and displayed by the ultrasound machine as part of the sonographic information obtained. A portion of the incident beam continues the journey, interacting with subsequent impedance mismatches. Several factors control these two phenomena. Reflection requires a large enough smooth surface so that the sound beam doesn't split and go around the structures; differences in acoustic impedance must also exist. If the mismatch is too large, however, most sound will be reflected and none will remain to penetrate deeper. The percentage of sound reflected depends on the angle of incidence and the difference in acoustic impedance between the two media. FORMULA: R = EXAMPLE: Assuming normal incidence, calculate the percent of the beam reflected when an ultrasound pulse travels from the liver into the kidney. Since the pulse is traveling from the liver (Z 1 ) into the kidney (Z 2 ) the formula, using the impedance chart above, can be written as: (Z 2 - Z 1 ) 2 (Z 1 + Z 2 ) 2 Where: R = % beam reflected Z 1 = acoustic impedance (medium 1) Z 2 = acoustic impedance (medium 2) 27

Angle of Insonation The angle at which the sound beam strikes the interface, which does not figure into the above equation, is the other important factor that affects the magnitude of reflection. When the sound beam is at 90, or perpendicular to the reflector, it is said to possess NORMAL INCIDENCE. Maximum reflection of wave energy occurs when the incident beam is normal to the interface. It's like shining a bright flashlight directly into a mirror. Also, according to Snell's Law (discussed more under Refraction) when the angle of the incident beam is normal (90 ), the angle of the transmitted sound is also normal (or straight ahead) and there will be no refraction (bending of the beam). According to a variation on Snel s Law, the angle of incidence equals angle of reflection. If a sound beam strikes a reflector significantly off 90 0, the reflected sound will be bounced away from the returning path to the transducer. This inhibits the ability to obtain good images. As always with ultrasound reflections, there must be a difference in acoustic impedance for an echo to occur. OBLIQUE INCIDENCE is the term applied to any angle of incidence that isn't normal (90 ). Generally, it is only of interest when considering refraction or in optimizing transducer position in imaging applications. 28

Reflection Coefficients Reflection can be measured in terms of the difference in amplitude that exists between the incident beam and the returning echoes. Assuming that the incident beam is the entire beginning entity (unity, or 1) then transmitted energies are those remaining after a reflection occurs at an interface. Mathematically, transmitted energies (T) are the difference between what went into the interface (1) and what was reflected back (R). This relationship is described by the equation: FORMULA: T = 1 - R where: 1 = incident sound intensity T = transmitted energies R = amount reflected EXAMPLE: Using the calculations above for reflection from a liver/kidney interface, what percentage of the ultrasound beam intensity is transmitted beyond this interface? T = 1 - R = 1-0.00008 = 0.99992 Virtually all of the original intensity remains after encountering the first interface in this hypothetical medium through which the wave is traveling. Again, for purpose of example, adequate energy remains to continue on to the next interface and produce yet another reflection. Obviously, transmission and reflection intensities, when added together, will always equal incident intensity at the interface. It is possible to describe reflection and transmission intensities as a percentage of incident intensity. The following two coefficients do just that: INTENSITY REFLECTION COEFFICIENT (IRC) is the percentage of the beam that is reflected when the sound beam passes from one medium to another. It is a direct function of the difference in acoustic impedance. Mathematically, the reflection coefficient is calculated by the formula given above for standard reflection (using acoustic impedance, or Z, values). However, it can be stated another way as a ratio, or percentage, of reflected intensity to incident intensity: FORMULA: IRC = Reflected intensity Incident intensity 29

INTENSITY TRANSMISSION COEFFICIENT (ITC) is the percentage of the beam that is transmitted through the interface and goes on to interact with subsequent impedance mismatches. It is inversely related to the reflection coefficient; as the magnitude of reflection increases, the magnitude of transmission decreases. Remember that, at normal incidence, when there is no change in the acoustic impedance, no reflection will occur. In this case, the transmitted energies are equal to the incident energies. This value is expressed mathematically as: FORMULA: ITC = Transmitted intensity Incident intensity Both of these coefficients are unitless and can range from 0-100%. When added together, however, they always equal 100%. Some important points about reflectors found in the human body: 1. Strongest reflections arise from the greatest changes in mechanical or elastic properties of adjacent tissue. It is mathematically related to the difference in acoustic impedance. 2. Approximately 1% of the incident beam can be reflected from smooth connective tissue such as that found in the renal capsule, surrounding major muscle bundles, and in the walls of small blood vessels. 3. At normal incidence (F o = 90 ) the amount of reflection is proportional to the difference in acoustic impedance between two tissues. MUSCLE - BLOOD INTERFACES REFLECT <1% FAT - MUSCLE INTERFACES REFLECT <1% TISSUE - BONE INTERFACES REFLECT 43% Specular Reflection Specular means mirror-like, and while specular reflection doesn't truly occur in diagnostic ultrasound, the word has come to connote bright, smooth reflectors. True specular reflection would send all energies back to the transducer and, as you know, some of the energy must be transmitted for deeper imaging. For a reflector to be considered specular, it must have a large, flat surface greater in size than the resolving capabilities of the scanner (usually 1mm). Secondly, an acoustic mismatch must exist. Finally, the reflecting surface must be nearly perpendicular to the transducer face in order for the reflection to be received by the transducer. 30

The smoother the interface, the more the magnitude of the reflection depends upon the incident angle. Maximum reflection occurs at 90. For most anatomic structures, the wavelength is considerably smaller than the interface. Therefore, the best reflection occurs when the transducer is at 90 to the interface. As the transducer is moved away from 90, the intensity of the reflection diminishes until, finally, it is not detectable. There is no specific angle at which this occurs. Rather, it is a function of impedance mismatch, sensitivity of the unit, transducer diameter, and depth of the reflector. This schematic demonstrates a specular reflector, which is a large interface that is close to perpendicular (90 0 ) to the beam. Non-Specular Reflection (Scattering) Scattering occurs when the sound beam interacts with interfaces that are rough, irregular, or smaller than the wavelength of the sound beam. For example, small reflectors that are about the same size as a wavelength fall into this category. This phenomenon is most widely used in Doppler ultrasound and in producing parenchymal tissue textures. Small, uneven reflectors produce low-level echoes. In the liver, breast, and other parenchymal structures, small lobules provide acoustic interfaces and generate different textural patterns. A major scatterer in the human body is the red blood cell (RBC), each of which has a dimension of about 8-10 m. Each RBC will absorb a certain amount of ultrasound energy and will re-radiate it in all directions. This means that a blood vessel can be insonated and the resulting scattering can be detected from any angle as long as the receiver is pointed at the area of insonation. The reduction of the incident energies over 360 is called Rayleigh scattering. It is analogous to the concentric formation and distribution of light when the headlights of an automobile interact with fog. No longer are the beams of the headlights focused directly on the road, but rather a glowing cloud is seen in front of the car. Raleigh scattering most frequently occurs in circulating red blood cells and forms the basis for Doppler and Color Doppler Imaging (CDI) techniques. 31

FACTORS INFLUENCING SCATTERED INTENSITY: 1. Dimension of the scatterer. As size increases, scattered intensity increases. 2. Number of scatterers present. As number increases, intensity increases. Studies have shown that scattering from blood is proportional to hematocrit (HCT) when measured at low levels. 3. Inhomogeneity of scatterer, or difference in acoustic impedance between scatterer and surrounding material. 4. Frequency of incident beam. As frequency increases, intensity increases. This schematic demonstrates scattering, or the re-direction of sound energies when the incident beam encounters an irregular surface. 3. Refraction Refraction is the change in direction of a wave as it passes obliquely from one medium into a second medium in which the propagation speed is different. Refraction occurs with light as well as sound. An example of light refraction is demonstrated in some science classes where the teacher places a pencil into a vase filled with water. To the eye of the observer, it appears that the pencil is bent. In reality, the difference in the direction of light waves passing through water and air bends, or refracts, the light. The same can happen with ultrasound waves. Simply stated, refraction is the change in direction of sound as it passes from one medium to another. 32

SNELL'S LAW OF OPTICS calculates this bending effect. It states that the ratio of the sine of the incident angle and the sine of the refracted angle are almost equal. When a sound beam obliquely encounters an interface, part of the beam is reflected and part of it passes through. Snel s Law of Refraction calculates the angle of the transmitted beam. The sound beam is always bent TOWARD the low velocity medium. FORMULA: sin t = C 2 sin i C 1 Where: sin t = transmitted angle sin i = incident angle C 2 = velocity through medium 2 C 1 = velocity through medium 1 DIFFERENCE IN VELOCITY REFRACTION DIFFERENCE IN VELOCITY REFRACTION EXAMPLE : Calculate the angle of refraction of a beam passing at an angle of 10 o from fat into soft tissue. 1. Obtain sin for incident angle (sin i ) from sin table. (sin 10 o =.173) 2. Obtain velocity for fat (C 1 ) and soft tissue (C 2 ) from velocity table. (C 1 = 1460 m/s; C 2 = 1540 m/s) sin t.173 = 1540 1460 sin t.173 = 1.06 sin t = 1.06 x.173 sin t = 0.183 From sin chart: sin.183 = 10 o t = 10 o 33

Refraction is a significant consideration in clinical applications of ultrasound technology. First of all, refraction can only occur if there is oblique incidence and a change in propagation speed between two media. When this happens, however, the bending of the sound beam can cause a variety of imaging artifacts and energy losses that may confuse the novice sonographer. With experience, techniques can be developed to compensate for this problem. (These are described further in the Chapter 14: Imaging Artifacts). One such phenomenon is called CRITICAL ANGLE. If the speed of sound in the material on the second side of an interface (Z 2 ) is significantly greater than that in the material on the first side of the interface (Z 1 ) and there is a severe or critical angle of incidence, the sound beam may be refracted at 90 to incident. If all the energy is sent off at 90, the transducer cannot detect echoes from the interface and "drop-off" or loss of imaging will occur in this area. In clinical practice, this is frequently seen near the dome of the full urinary bladder, at the edges of the BPD, or in any situation where a curved anatomical structure possessing different acoustic impedance from surrounding tissue exists. Sound waves refract or bend away at 90 o away from incident when the beam encounters a curved structure of different acoustic impedance. See p. 56 for Exercises 2: Interaction with Soft Tissue 34