Potentiometry: Ion-Selective Electrodes

Similar documents
Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

Titration of a strong acid with a strong base with Cobra4

Chem 321 Lecture 17 - Potentiometry 10/24/13

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

Operating Instructions for Ammonium ISE Specifications

Related concepts Electrolyte, electrical conductance, specific conductance, ion mobility, ion conductivity, conductometry, volumetry.

Electrochemistry LEC Potentiometric ph titration (phosphoric acid in a soft drink) What you need: What you can learn about

EXPERIMENT 6. Properties of Buffers INTRODUCTION

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

1225 Lab # 11 Fluoride Speciation and Analysis by Flow Injection using Ion-Selective Electrode: Measurement of Total Fluoride in Water

Buffer Preparation. Learning Objectives:

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Determination of the K a of a Weak Acid and the K b of a Weak Base from ph Measurements

GENERAL INSTRUCTIONS GENERAL PREPARATION

# 12 ph-titration of Strong Acids with Strong Bases

Experiment C-10 Titration of a Strong Acid and a Strong Base

Conductometric Titration & Gravimetric Determination of a Precipitate

Electrochemistry LEC 06. The Nernst equation. What you need: What you can learn about. Principle and tasks

Determining the K sp of Calcium Hydroxide

READING A. INTRODUCTION CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE. Skoog, Holler and Crouch: Chapter 23 and Appendix 3.

Electrochemistry. Conductivity of strong and weak electrolytes LEC 06. What you need: What you can learn about. Principle and tasks

#13 ph-titration of Weak Acids with Strong Bases

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

6 Acid Base Titration

PRACTICAL 3 ph AND BUFFERS

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base

Standardizing a Solution of Sodium Hydroxide. Evaluation copy

POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA

Acid-Base Titration. Sample

ph Probe ReallyEasyData com

(b) Write the chemical equation for the dissolution of NaOH in water.

Thermodynamics and the Solubility of Sodium Tetraborate Decahydrate

Electrochemical Techniques: Cyclic Voltammetry

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

Analysis of cations and anions by Ion- Selective Electrodes (ISEs)

Determining the Conductivity of Standard Solutions

Mixtures of Acids and Bases

Chapter 12 Tex-617, Determining Chloride in Concrete

EXPERIMENT 8 POTENTIOMETRY: DIRECT-MEASUREMENT OPTION

ALLOWAY. Method Outline. Standard Laboratory Method: Total Fluoride via Distillation. Distillation then Specific Ion. Date Issued: 01/04/82

Dr. White Chem 1B Saddleback College 1. Experiment 15 Thermodynamics of the Solution Process

Freezing point depression (Item No.: P )

Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE

Chemistry MultiMeasure Sensor PS-2170

Acid-Base Titration. Evaluation copy

1) A ph/mv meter or an ion meter, either line operated or portable.

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

The Synthesis and Analysis of Aspirin

Shown below is a sample titration curve for a diprotic acid. Note the two equivalence points.

Experiment 7 Buffer Capacity & Buffer Preparation

POTENTIOMETRIC TITRATIONS & SOLUBILITY EQUILIBRIA. Background

Chemistry 213. Electrochemistry

PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION

Electricity. Electrolysis. Current and the transport of charge DETERMINATION OF THE FARADAY CONSTANT BASIC PRINCIPLES

GETTING THE END POINT TO APPROXIMATE. Two hours

GENERAL INSTRUCTIONS GENERAL PREPARATION

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

Electrochemical Techniques: Anodic Stripping Voltammetry

CHEMISTRY DEPARTMENT

Title: Gravimetric verification of chloride concentration from a precipitate of silver nitrate.

iworx Sample Lab Experiment GB-2: Membrane Permeability

#30 Thermochemistry: Heat of Solution

Conductimetric Titration and Gravimetric Determination of a Precipitate

Titration of Acids and Bases

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15

Conductimetric Titration and Gravimetric Determination of a Precipitate

Chemistry with Mr. Faucher. Acid-Base Titration

METHOD 13B - DETERMINATION OF TOTAL FLUORIDE EMISSIONS FROM STATIONARY SOURCES (SPECIFIC ION ELECTRODE METHOD)

Density of Aqueous Sodium Chloride Solutions

Experiment DE: Part II Fisher Esterification and Identification of an Unknown Alcohol

The Determination of an Equilibrium Constant

National standard of People s Republic of China

Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic.

Acid-Base Titration. Computer OBJECTIVES

Sodium Chloride - Analytical Standard

Introduction. Concepts Kinetics Order of reaction Reaction rate Colorimetry. Background

User Manual. Thermo Scientific Orion Fluoride Ion Selective Electrode

Synthesis of Benzoic Acid

To understand concept of limiting reagents. To learn how to do a vacuum filtration. To understand the concept of recrystallization.

12.01 Determination of the isoelectric point of an amino acid (glycine)

Sodium Chloride - Analytical Standard

7-A. Inquiry INVESTIGATION. 322 MHR Unit 3 Quantities in Chemical Reactions. Skill Check. Safety Precautions

Advanced Unit 7: Chemistry Practical Examination (SET A) Candidates must have: Scientific calculator Ruler

To see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide:

Modified Adams Assay for Phenolics in Wine

O H 3 O 1 1 A. O 1 1 OH (K w

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

perfection Guidebook perfection Combination Fluoride Electrode Successful Ion Measurement

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

Ka Acid Dissociation Constant Minneapolis Community and Technical College Principles of Chemistry II, C1152 v.1.16

Determination of the K a Value and Molar Mass of an Unknown Weak Acid

aa + bb cc + dd Equation 1

Ion Selective Electrode Probe

maect PRODUCTION SOP M/1 JUNE 2007 version page: 1 Preparation of 2 liters of 10x concentrated PBS stock solution (phosphate buffered saline)

SPECIFICATION & TEST PROCEDURE SODIUM SALICYLATE Technical. Molecular weight : Reference : In-house

Measuring Enthalpy Changes

Transcription:

Potentiometry: Ion-Selective Electrodes Determination of the sodium monofluorophosphate content of a toothpaste sample using a fluoride ion-selective electrode. 1. Aims The aims of the experiment are to: 1.1 Calibrate a fluoride ion-selective electrode (FISE) and determine over what range of fluoride ion concentrations the electrode has a Nernstian response. 1.2 Use the calibrated FISE to determine the amount of sodium monofluorophosphate present in a toothpaste sample, and see if the levels are consistent with the amount reported by the manufacturer. 1.3 Characterise inteferences to the FISE, including interferences from chloride ions, iron (III) ions and ph variations. 2. Background Ion-selective electrodes are potentiometric electrodes with a high degree of specificity for a single ion or class of ions. They are used in conjunction with a reference electrode and a high impedance voltmeter such as the EA165 ph/mv Amp, to complete the measuring circuit in Figure 2.1. Figure 2.1 Potentiometric measuring circuit Ion-Selective Electrodes Page 1 of 13 EXP006_0305

The measured potential (E) of the ion-selective electrode is given by the Nernst equation E = E 0 + 2.3 RT nf log a i where E 0 (standard potential) is a constant dependant upon the type of measuring and reference electrodes being used, R is the universal gas constant, T is the temperature in Kelvin, F is the Faraday constant, n is the valence of the ion and a i is the ionic activity of the ion being measured. The ionic activity, a i, of a solution is related to the analytical concentration, [i], by the following relationship a i = i i where i is the ion activity coefficient. In dilute solutions, i is frequently close to unity, thus activity and concentration can often be used interchangeably. However in most analytical situations, i varies significantly from unity, and is principally a function of the ionic strength of the solution. Ionic strength adjustment buffers are often used in conjunction with an ion-selective electrode to fix the ionic strength of the solution high enough to make the activity coefficients of unknowns and standards virtually identical. These buffers are carefully selected to have a ph that favours the free form of the ion, and often contain complexing agents that preferentially complex with interfering ions, releasing the analyte ion. Adjusting the ph and activity coefficients to optimum and constant values allows the concentration of an ion to be interpreted from measurements of ion activity. Fluoride ion selective electrodes (Figure 2.2) consist of a tube with a lanthanum fluoride (LaF 3 ) crystal membrane forming an interface between the external solution and the internal electrolyte. The internal electrolyte consists of 0.1 M NaCl, 0.1 M NaF and saturated AgCl, all in contact with the internal Ag/AgCl reference electrode and LaF 3 crystal membrane. Figure 2.2 Fluoride ion-selective electrode showing lanthanum fluoride membrane and internal electrode construction. The cell notation for the electrode is: Ag / AgCl (s),cl (aq) (0.1M),F (aq) (0.1M) /LaF /F (" x "M) / ERE 3 (aq) Internal Solution Test Solution Ion-Selective Electrodes Page 2 of 13 EXP006_0305

The overall cell potential is the result of several contributions, but only activity of the fluoride in the test solution is variable. A change in the fluoride ion concentration in the test solution results in a change of potential over the LaF3 crystal membrane. This membrane potential is the result of a dynamic exchange of fluoride ions between both the test solution and internal filling solution, and the crystal membrane. This change in potential is measured in the experiment and calibrated against fluoride ion concentration. 3. Hardware Required Per Group: a computer system an e-corder unit an EA165 ph/mv Amp a fluoride ion-selective electrode with BNC connector a Ag/AgCl reference electrode with 4 mm banana plug connector 17 x 150 ml polypropylene beakers* 19 x 100 ml volumetric flasks* 1 x 25 ml pipette 1 x 10 ml pipette 1 x 5 ml pipette 2 x stainless steel spatulas a small glass funnel a pipette bulb a magnetic stirrer with 20 mm stirrer bar a retort stand with two boss heads and clamps attached a permanent marker pen a wash bottle filled with distilled water a thermometer * Less may be used, however they will need to be rinsed thoroughly with deionised water between uses. Per Class: a microwave oven oven mitts an electronic balance (0.0001 g readability) a printer Ion-Selective Electrodes Page 3 of 13 EXP006_0305

4. Reagents Required a tube of toothpaste with sodium monofluorophosphate as the active ingredient sodium monofluorophosphate (Na 2 FPO 3 ) 1 M sodium fluoride (NaF) solution sodium chloride (NaCl) Total Ionic Strength Adjustment Buffer (TISAB) 0.2 M potassium nitrate solution (KNO 3 ) distilled water 0.01 M iron chloride FeCl 3 solution 1.0 M HCl solution 1.0 M NaOH solution 5. Safety Notes & Instrument Care Wear gloves and protective eyewear throughout the experiment Take care removing contents from the microwave oven, use the oven mitts supplied Avoid spilling liquids on the hardware by mixing reagents away from the units Avoid touching the lanthanum fluoride membrane of the FISE 6. Hardware Setup 6.1 Setup the e-corder and EA165 ph/mv Amp as per Instructions in the edaq hardware manuals found on the edaq installer CD. Connect the Amp to Channel 1 of the e-corder. Connect the FISE to the BNC input of the ph/mv Amp and the reference electrode into the 4 mm socket. The configuration should be warmed up for at least 10 minutes prior to the laboratory session. 6.2 Secure the electrode(s) to the retort stand with the clamps. Place magnetic stirrer unit below the electrodes and set to a medium rate of stirring. The polypropylene cups are placed on top of the unit and secured with another clamp connected to the retort stand. 7. Procedure 7.1. Preliminary Steps 7.1.1 Record the ambient temperature of the laboratory. For best results, all reagents should be equilibrated to the laboratory temperature. 7.1.2 Accurately weigh out around 400 mg of toothpaste in a 150 ml polypropylene beaker (remembering to tare out the mass of the beaker). Record the mass and Ion-Selective Electrodes Page 4 of 13 EXP006_0305

make a note of the reported sodium monofluorophosphate content of the brand of toothpaste. 7.1.3 Mix the toothpaste with 50 ml of TISAB and heat in a microwave oven at maximum power for two minutes. Wait for a minute and carefully remove the beaker from the oven. WARNING: THE CONTENTS WILL BE HOT! Set aside to cool to room temperature whilst carrying out the calibration of the FISE. 7.1.4 To determine the proportion of fluoride that is liberated from the sodium monofluorophosphate matrix, accurately weigh around 400 mg of pure sodium monofluorophosphate (record the exact weight), transfer powder to a 100 ml volumetric flask, and dilute to the mark with distilled water. Then serially dilute 5 ml of this solution in another 100 ml volumetric flask. 7.1.5 In another 150 ml polypropylene beaker, repeat the steps in 7.1.3 substituting the toothpaste with 5 ml of the sodium monofluorophosphate solution prepared in 7.1.4. 7.1.6 Calculate and make a note of the theoretical fluoride ion concentration of the solution prepared in 7.1.4, if 100% of the fluoride bound in the sodium monofluorophosphate matrix is converted to fluoride during the digestion step. 7.1.7 Open Chart software and hide all channels except Channel 1. Use the following settings throughout the rest of the experiment. Range 200 mv Sampling Speed 2 /s Filtering Select a 1 Hz digital filter in the input amplifier dialog box for the ph/mv Amp Trigger Set Chart to record a 5 min block of data when prompted by the user. This should be enough time for the response of the FISE to stabilise. Adjust the recording time if it takes longer than 5 minutes for the signal to stabilise. 7.2. FISE Calibration 7.2.1 From the 1 M NaF top standard, prepare fluoride standards by serial dilution with concentrations of 10-1, 10-2, 10-3, 10-4, 10-5, 10-6 and 10-7 M. This can be conveniently achieved by diluting 10 ml of the previous higher concentration standard to 100 ml in a volumetric flask. 7.2.2 The fluoride standards should be analysed from least to most concentrated. Pipette 25 ml of the 10-7 M standard and 25 ml of TISAB buffer into a 150 ml polypropylene beaker. 7.2.3 Put the stirrer bar into the beaker and place the beaker onto the magnetic stirrer set to a medium rate of stirring. Lower the electrodes into the solution. Ion-Selective Electrodes Page 5 of 13 EXP006_0305

7.2.4 Click on the Start button to begin recording a block of data. When the block of data has been recorded, lift the electrodes and the stirrer bar from the solution. Rinse the electrodes and stirrer bar with distilled water and pad dry with lint free tissue. Use the comment function of Chart to make a note of the fluoride concentration with the corresponding block of data. 7.2.5 Repeat steps 7.2.2 to 7.2.4 with the next standard, until all standards have been analysed. 7.2.6 Adjust the view buttons so that all the blocks of data collected for the standards can be seen on the screen. With the mouse, highlight an area that encompasses the voltage response for all the standards. With the area highlighted, select MPCalibration from the Channel Function pop-up menu to display the Multipoint Calibration dialog box. Figure 7.1 Typical electrode response for NaF standards and the Multipoint Calibration dialog box. 7.2.7 For each block of data, highlight a region where the signal has stabilised, i.e. towards the end of the block. With the region highlighted, click on one of the arrows on the far LHS of the Multipoint Calibration dialog box. This enters the average millivolt reading for the selected data into the left hand column. Enter the corresponding pf for that solution in the right hand column. If Chart has not been previously used to measure pf from a FISE, you will need to define the unit pf. pf = -log F - Ion-Selective Electrodes Page 6 of 13 EXP006_0305

7.2.8 Click on the fit button in the Multipoint Calibration dialog box and select a linear fit. Inspect the resulting fitted curve and determine over which range of pf the electrode has a linear response (i.e., Nernstian response). A full fluoride standard range sample is given in Figure 7.2. Print the fitted curve over the full standard range for inclusion in your report. Figure 7.2 Linear fit over the full standard range. 7.2.9 In the Multipoint Calibration dialog box, delete the data points with pf values above which there is no longer a Nernstian response (probably the lowest 2 or 3 fluoride standards) In the case shown in Figure 5, the Nernstian response appeared to break down above a pf of 4 (i.e. [F - ] below 10-4 M). Click on fit again and inspect the new fitted linear curve for the reduced number of data points. Print the fitted curve over the Nernstian response range for inclusion in your report. Ion-Selective Electrodes Page 7 of 13 EXP006_0305

Figure 7.3 Linear fit over the Nernstian range. 7.2.10 Use the Nernst equation to calculate the theoretical FISE Nernstian response and use this value to calculate the percent Nernstian response for the curves obtained in 7.2.8 and 7.2.9. Note that the slope of the equation in the Multipoint Calibration Fit window is in pf.mv -1. Convert the slope to mv.pf -1 to calculate the percent Nernstian response. Hint: Theoretical Nernstian Slope = 2.3RT nf Upon returning to the Chart window, you will notice that the millivolt traces have been converted to pf traces. Chart has now been calibrated to record pf directly from the FISE. 7.3. Sample Dilution and Analysis The toothpaste and sodium monofluorophosphate samples prepared in 7.1, need to be diluted with TISAB prior to analysis. 7.3.1 Quantitatively transfer the cooled toothpaste sample to a 100 ml volumetric flask and dilute to the mark with distilled water. 7.3.2 Repeat steps 7.2.2 to 7.2.4, substituting the standard with 25 ml of the diluted toothpaste sample. Note the pf value for the toothpaste sample when the electrode response has stabilised. Convert this value into a fluoride ion concentration. 7.3.3 With the sodium monofluorophosphate solution prepared in 7.1.4. Repeat step 7.3.2 above, substituting the diluted toothpaste sample with 25 ml of sodium monofluorophosphate solution. Ion-Selective Electrodes Page 8 of 13 EXP006_0305

7.3.4 Compare the measured fluoride concentration in the sodium monofluorophosphate solution with the theoretical concentration calculated in 7.1.6. Determine the percentage of fluoride liberated from the sodium monofluorophosphate matrix. 7.3.5 Assume that the percentage of fluoride liberated from the pure sodium monofluorophosphate sample is the same as the percentage liberated from the sodium monofluorophosphate in the toothpaste sample. Using this assumption, the measured fluoride concentration in 7.3.2 and the various dilution factors used throughout the experiment, calculate the sodium monofluorophosphate content of the toothpaste sample. 7.4. Interferences 7.4.1 Pipette 10 ml of the 10-1 M NaF standard into a 100 ml volumetric flask. Add 10 g of NaCl, swirl to dissolve and dilute to the mark with distilled water. Repeat steps 7.2.2 to 7.2.4 substituting the standard with 25 ml of the above solution. Report the mv value for the response, and compare with the value obtained for the 10-2 M standard in 7.2.4. Turn off Multipoint Calibration to return to the raw mv recording. 7.4.2 Pipette 10 ml of the 10-1 M NaF standard into a 100 ml volumetric flask. To this flask add 5 ml of the Fe 3+ solution and dilute to the mark with distilled water. Repeat steps 7.2.2 to 7.2.4 substituting the standard with 25 ml of the above solution. Report the mv value for the response, and compare with the value obtained for the 10-2 M standard in 7.2.4. 7.4.3 Pipette 10 ml of the 10-1 M NaF standard into a 100 ml volumetric flask. To this flask add 10 ml of the 1 M HCl solution and dilute to the mark with distilled water. Repeat steps 7.2.2 to 7.2.4 substituting the standard with 25 ml of the above solution. Report the mv value for the response, and compare with the value obtained for the 10-2 M standard in 7.2.4. 7.4.4 Pipette 10 ml of the 10-1 M NaF standard into a 100 ml volumetric flask. To this flask add 10 ml of the 1 M NaOH solution and dilute to the mark with distilled water. Repeat steps 7.2.2 to 7.2.4 substituting the standard with 25 ml of the above solution. Report the mv value for the response, and compare with the value obtained for the 10-2 M standard in 7.2.4. 7.4.5 Repeat steps 7.4.1 to 7.4.4 substituting the TISAB solution with the 0.2 M KNO 3 solution as the supporting electrolyte. Report the mv value for the response, and compare with the value obtained for the 10-2 M standard in 7.2.4. Ion-Selective Electrodes Page 9 of 13 EXP006_0305

8. Results Brand of toothpaste: Mass of toothpaste: mg, Mass of Na 2 FPO 3 : mg, Reported Na2FPO3 content: 100% dissolution concentration: M Range of FISE Nernstian response: pf Theoretical Nernstian slope:.pf 1 Percent Nernstian response of FISE: % Toothpaste sample solution fluoride concentration: M Na 2 FPO 3 sample solution fluoride concentration: M % Na 2 FPO 3 dissolution: % Measured Na 2 FPO 3 content of toothpaste sample: % Difference between measured and reported fluoride content of toothpaste sample: % Response 10-2 M standard with TISAB Response 10-2 M standard with TISAB and NaCl Response 10-2 M standard with TISAB and 1 M HCl Response 10-2 M standard with TISAB and 1 M NaOH Response 10-2 M standard with 0.2 M KNO 3 Response 10-2 M standard with 0.2 M KNO 3 and NaCl Response 10-2 M standard with 0.2 M KNO 3 and 1 M HCl Response 10-2 M standard with 0.2 M KNO 3 and 1 M NaOH Ion-Selective Electrodes Page 10 of 13 EXP006_0305

9. Questions 1. What is the Nerstian slope of an ideal ph electrode at 25 C? Over what range of ph is the response of a ph electrode typically Nernstian? How does this range compare with the Nernstian response range of the FISE? As the hydronium ion charge is +1, the Nernstian slope is opposite to the slope for a FISE, i.e. 59.160 mv.ph -1. ph electrodes typically have a Nernstian response between ph 0 and 14, which is several orders of magnitude wider than the response of a FISE. 2. What is the purpose of using an ionic strength adjustment buffer when measuring fluoride ion concentrations in solution with a FISE? In 7.4.5 what influence does substituting the TISAB with 0.2 M KNO 3 have on the electrode response? The purpose of the TISAB solution is threefold: to adjust ph to a constant value, to provide a background of constant ionic strength, and to release fluoride from its complexes with a range of metal ions, principally Al 3+ and Fe 3+. Substituting TISAB with 0.2 M KNO 3 increases interferences due to ph fluctuations, and the presence of Fe 3+. 3. To what degree do chloride ions interfere with a FISE signal? Propose a mechanism for the interference. The presense of an excess of chloride ions causes a slight positive bias in the determination of total fluoride. Fluoride ions preferentially diffuse across the LaF 3 membrane, but other species such as chloride and other halogens do also to a lesser extent. 4. To what degree do Fe 3+ ions interfere with a FISE signal? Propose a mechanism for the interference. The addition of Fe 3+ ions cause a negative bias in the determination of total fluoride. This is because FISE only respond to free fluoride, and Fe 3+ tends to form complexes with fluoride. This effect is more pronounced in the sample without TISAB, because this reagent includes a much stronger metal complexing agent. 4. How does ph influence the response of the FISE? Propose a mechanism for the interference. Low ph causes a negative bias, as the free fluoride concentration is reduced by the formation of HF and HF 2 -. High ph causes a positive bias as the hydroxide ion interferes with the LaF 3 membrane. This effect is more pronounced when TISAB is not used. Ion-Selective Electrodes Page 11 of 13 EXP006_0305

10. Instructors Notes 10.1 Computer System and Software A PC (running Windows XP or later) or Mac (running OS X or later) computer with USB support is required. Chart must be installed along with the Multipoint Calibration extension, which is available for download from the edaq website: http://www.edaq.com/chart_extns_list.html 10.2 Availability of Fluoride Ion-Selective Electrodes There are many companies that manufacture fluoride ion-selective electrodes. See the ph and ion-selective electrodes page on the edaq website for a list of some manufacturers: http://www.edaq.com/research_ph.html The majority of FISE manufactured require an additional external reference electrode. A Ag/AgCl reference electrode is the most practical type to use in this experiment. Some companies manufacture combination FISE, but these tend to be more expensive than the standard type. A common feature of both types of FISE is a single crystal lanthanum fluoride, LaF 3, membrane, which permits the transport of fluoride ions only. The crystal is doped with europous fluoride, EuF 2, in order to reduce its electrical resistance. 10.3 Potentiometers This practical has been written assuming a EA165 ph/mv Amp is to be used with a standard FISE and reference electrode, however the EP303 ph Pod can be used with a combination FISEs. Most other standard ph meters with an analog output can also be used in this experiment. To use the edaq ph/mv Amp, the FISE must be fitted with a standard BNC connector, and the reference electrode must be fitted with a 4 mm banana plug. 10.4 Reagents The following solutions should be prepared prior to the laboratory session in quantities suitable for the size of the class. All reagents need to be prepared in deionised water. Total Ionic Strength Adjustment Buffer (TISAB)* 1 M NaF solution 0.01 M FeCl3 solution 1.0 M HCl solution 1.0 M NaOH solution * To Prepare 1 L of TISAB buffer, boil 500 ml of distilled water in a 1 L beaker, add 57 ml of glacial acetic acid, 58 g of NaCl and 2 g of 1,2 cyclohexlenedinitrotetraacetatic acid (CDTA). Stir Ion-Selective Electrodes Page 12 of 13 EXP006_0305

to dissolve solids and cool the solution to room temperature. Adjust the ph to between 5 and 5.5 by adding 5 M NaOH. Quantitatively transfer to a 1 L volumetric flask and fill to the mark. Solutions containing fluoride should be stored in polypropylene bottles. NaCl and sodium monofluorophosphate powder also needs to be available to the class. 11. Bibliography D. Midgley, and K.Torrance, Potentiometric Water Analysis, 2nd Edition. Ion-Selective Electrodes Page 13 of 13 EXP006_0305