Rheological measurements using microcantilevers

Similar documents
Design and Analysis of Various Microcantilever Shapes for MEMS Based Sensing

Torsional frequency response of cantilever beams immersed in viscous fluids with applications to the atomic force microscope

Noninvasive determination of optical lever sensitivity in atomic force microscopy

Point mass approximation. Rigid beam mechanics. spring constant k N effective mass m e. Simple Harmonic Motion.. m e z = - k N z

Basic Laboratory. Materials Science and Engineering. Atomic Force Microscopy (AFM)

Noise, AFMs, and Nanomechanical Biosensors

CHAPTER 4 DESIGN AND ANALYSIS OF CANTILEVER BEAM ELECTROSTATIC ACTUATORS

Optimal Design and Evaluation of Cantilever Probe for Multifrequency Atomic Force Microscopy

Parametric Investigation of the Common Geometry Shapes for Added Mass Calculation

Digital processing of multi-mode nano-mechanical cantilever data

FEM-SIMULATIONS OF VIBRATIONS AND RESONANCES OF STIFF AFM CANTILEVERS

The Fluid-Coupled Motion of Micro and Nanoscale Cantilevers

3.052 Nanomechanics of Materials and Biomaterials Thursday 02/08/06 Prof. C. Ortiz, MIT-DMSE I LECTURE 2 : THE FORCE TRANSDUCER

Scanning Force Microscopy

SURFACE-DISPLACEMENT IMAGING USING OPTICAL BEAM DEFLECTION. S.E. McBride* and G.C. Wetsel, Jr.* Southern Methodist University Dallas, Texas 75275

Chapter 2 Correlation Force Spectroscopy

2d-Laser Cantilever Anemometer

Instrumentation and Operation

SUPPLEMENTAL MATERIAL Dynamic Similarity of Oscillatory Flows Induced by Nanomechanical Resonators

Plasmonic sensing of heat transport and phase change near solid-liquid interfaces

Lecture Note October 1, 2009 Nanostructure characterization techniques

Lecture 4 Scanning Probe Microscopy (SPM)

Lorentz Contact Resonance for viscoelastic measurements of polymer blends

Measuring the spring constant of atomic force microscope cantilevers: thermal fluctuations and other methods

AC Electrothermal Characterization of Doped-Si Heated Microcantilevers Using Frequency-Domain Finite Element Analysis

Effect of AFM Cantilever Geometry on the DPL Nanomachining process

Effect of Liquid Viscosity on Sloshing in A Rectangular Tank

Dynamic Analysis of AFM in Air and Liquid Environments Considering Linear and Nonlinear Interaction Forces by Timoshenko Beam Model

Mapping the mechanical stiffness of live cells with the scanning ion conductance microscope

Introduction to Scanning Tunneling Microscopy

IN 1986, Binnig, Quate, and Gerber invented the atomic force

Differential criterion of a bubble collapse in viscous liquids

Effects of Variation of the Flame Area and Natural Damping on Primary Acoustic Instability of Downward Propagating Flames in a Tube

Micro-Rheology Measurements with the NanoTracker

Methodology for sloshing induced slamming loads and response. Olav Rognebakke Det Norske Veritas AS

Lorentz Contact Resonance for viscoelastic measurements of polymer blends

The Reynolds experiment

Supporting Online Material for

EE 5344 Introduction to MEMS CHAPTER 6 Mechanical Sensors. 1. Position Displacement x, θ 2. Velocity, speed Kinematic

Module 26: Atomic Force Microscopy. Lecture 40: Atomic Force Microscopy 3: Additional Modes of AFM

Calibration of a Quartz Tuning Fork for Low Temperature Measurements

Experimental validation of incipient failure of yield stress materials under gravitational loading

Microstructure cantilever beam for current measurement

Comparative dynamics of magnetically, acoustically, and Brownian motion driven microcantilevers in liquids

Nonlinear Vibration of the Double-Beams Assembly Subjected to A.C. Electrostatic Force

Chemical and Biomolecular Engineering 150A Transport Processes Spring Semester 2017

Pulsed Photoacoustic Spectroscopy of I 2 and NO 2

developed piezoelectric self-excitation and selfdetection mechanism in PZT microcantilevers for dynamic scanning force microscopy in liquid

Modeling of Trace Gas Sensors

MEMS-based Measurement of Rheological Fluid Properties

Optical sensor based on hybrid LPG/FBG in D-fiber for simultaneous refractive index and temperature measurement

The Large Amplitude Oscillatory Strain Response of Aqueous Foam: Strain Localization and Full Stress Fourier Spectrum

Time Resolved Faraday Rotation Measurements of Spin Polarized Currents in Quantum Wells

and Joachim Peinke ForWind Center for Wind Energy Research Institute of Physics, Carl-von-Ossietzky University Oldenburg Oldenburg, Germany

Experimental and numerical investigation on particle-induced liquid metal flow using Lorentz force velocimetry

Studies on flow through and around a porous permeable sphere: II. Heat Transfer

APPENDIX 1: SPECTRAL METHODS FOR MEASURING THE NORMAL CANTILEVER SPRING CONSTANT k

REPULSIVE CASIMIR AND VAN DER WAALS FORCES: FROM MEASUREMENTS TO FUTURE TECHNOLOGIES

General concept and defining characteristics of AFM. Dina Kudasheva Advisor: Prof. Mary K. Cowman

Metamaterials. Peter Hertel. University of Osnabrück, Germany. Lecture presented at APS, Nankai University, China

Multi-Dimensional IR Spectroscopy of Acetic Acid Dimers and Liquid Water

BDS2016 Tutorials: Local Dielectric Spectroscopy by Scanning Probes

Thermoacoustic Instabilities Research

Scanning Tunneling Microscopy

HYPER-RAYLEIGH SCATTERING AND SURFACE-ENHANCED RAMAN SCATTERING STUDIES OF PLATINUM NANOPARTICLE SUSPENSIONS

arxiv:cond-mat/ v2 [cond-mat.other] 2 May 2007

The sound generated by a transverse impact of a ball on a circular

SUPPLEMENTARY INFORMATION

Transactions on the Built Environment vol 22, 1996 WIT Press, ISSN

Ultrasonic Method for Measuring Frictional Temperature Rise in Rubbing Surface

Particles, drops, and bubbles. Lecture 3

Measuring rheological properties using a slotted plate device

Dopant Concentration Measurements by Scanning Force Microscopy

Measurement of hardness, surface potential, and charge distribution with dynamic contact mode electrostatic force microscope

CFD DESIGN OF A GENERIC CONTROLLER FOR VORTEX-INDUCED RESONANCE

Scanning Force Microscopy II

BIRLA INSTITUTE OF TECHNOLOGY AND SCIENCE, Pilani Pilani Campus

Measurement Techniques for Engineers. Motion and Vibration Measurement

Università degli Studi di Bari "Aldo Moro"

Formation and Long Term Evolution of an Externally Driven Magnetic Island in Rotating Plasmas )

Mobility and Impedance Methods. Professor Mike Brennan

Added-Mass Effect in Modeling of Cilia-Based Devices for Microfluidic Systems

Identification of crack parameters in a cantilever beam using experimental and wavelet analysis

SUPPLEMENTARY INFORMATION

Scanning capacitance spectroscopy of an Al x Ga 1Àx NÕGaN heterostructure field-effect transistor structure: Analysis of probe tip effects

Amplitude Change of a Quartz Crystal Microbalance

Article for the 29 th Sensing Forum

Scanning Tunneling Microscopy

Development of PC-Based Leak Detection System Using Acoustic Emission Technique

TAILORING THE GEOMETRY OF MICRON SCALE RESONATORS TO OVERCOME VISCOUS DAMPING MARGARITA VILLA

Accurate determination of the thickness or mass per unit area of thin foils and single-crystal wafers for x-ray attenuation measurements

On modeling two-mode vibration sensors for combined density and viscosity measurements

Modeling the Response of Photoacoustic Gas Sensors

INTRODUCTION TO SCA\ \I\G TUNNELING MICROSCOPY

DYNAMIC RESPONSE OF MINI CANTILEVER BEAMS IN VISCOUS MEDIA. Anamika Sharmin Mishty

LESSON PLAN. B.Sc - FIRST YEAR ( REGULATION)

How Does a Microcantilever Work?

Chaos suppression of uncertain gyros in a given finite time

Miao Boya and An Yu Department of Physics, Tsinghua University, Beijing , People s Republic of China

Sound Radiation Of Cast Iron

Transcription:

Rheological measurements using microcantilevers S. Boskovic Department of Chemical Engineering and School of Chemistry, The University of Melbourne, Victoria, 3010 Australia J. W. M. Chon and P. Mulvaney School of Chemistry, The University of Melbourne, Victoria, 3010 Australia J. E. Sader a) Department of Mathematics and Statistics, The University of Melbourne, Victoria, 3010 Australia (Received 11 September 2001; final revision received 25 February 2002) Synopsis The use of microcantilevers in rheological measurements of gases and liquids is demonstrated. Densities and viscosities of both gases and liquids, which can range over several orders of magnitude, are measured simultaneously using a single microcantilever. The microcantilever technique probes only minute volumes of fluid 1nL, and enables in situ and rapid rheological measurements. This is in direct contrast to established methods, such as cone and plate and Couette rheometry, which are restricted to measurements of liquid viscosity, require large sample volumes, and are incapable of in situ measurements. The proposed technique also overcomes the restrictions of previous measurements that use microcantilevers, which are limited to liquid viscosity only, and require independent measurement of the liquid density. The technique presented here only requires knowledge of the cantilever geometry, its resonant frequency in vacuum, and its linear mass density. A simple yet robust calibration procedure is described to determine the latter two parameters, from a single measurement of the resonant frequency and quality factor of the cantilever in a reference fluid such as air, if these parameters are unknown. 2002 The Society of Rheology. DOI: 10.1122/1.1475978 I. INTRODUCTION Since the invention of the atomic force microscope AFM, there has been extensive exploration of the force measurement and sensor applications of the instrument. In the latter class, AFM microcantilevers have been used as sensors for gas detection, mass measurement, temperature measurement, spectroscopy, and calorimetry, to name just a few; the reader is referred to Berger et al. 1997 for a detailed review. One application that has received attention recently is the use of AFM microcantilevers in rheological measurements on fluids Oden et al. 1996 ; Bergaud and Nicu 2000 ; Ahmed et al. 2001. By measuring the frequency response of a microcantilever a Author to whom correspondence should be addressed; electronic mail: jsader@unimelb.edu.au 2002 by The Society of Rheology, Inc. J. Rheol. 46 4, 891-899 July/August 2002 0148-6055/2002/46 4 /891/9/$25.00 891

892 BOSKOVIC ET AL. immersed in a fluid, information regarding the rheological properties of the fluid, in principle, can be determined. However, this application is predicated on the frequency response of the microcantilever being sensitive to the rheological properties of the fluid under consideration. It is well known that the frequency responses of cantilevers of macroscopic size, say, 1 m in length, are insensitive to the viscosities of the fluids in which they are immersed Lindholm et al. 1965 ; Fu and Price 1987. In contrast, for AFM microcantilevers, which are approximately 100 m in length, fluid viscosity can significantly affect their frequency response. This property has permitted the use of microcantilevers as small-scale viscometers Oden et al. 1996 ; Bergaud and Nicu 2000 ; Ahmed et al. 2001. However, this previous rheological application of microcantilevers is restricted to the measurement of liquid viscosity only, with density and the properties of gases being outside the realm of its applicability. In addition, a priori and independent measurement or knowledge of the liquid density is required, which in turn further restricts the utility of the microcantilever technique. In this paper, we demonstrate for the first time that the frequency response of microcantilevers can be used to measure the density and viscosity of both gases and liquids. This is achieved by employing the theoretical model of Sader 1998 for the frequency response of AFM cantilevers immersed in viscous fluids. We emphasize that, unlike previous works that use microcantilevers, this application is not restricted to the viscosity of liquids, and does not require a priori knowledge of the density of the fluid. Using a single microcantilever, the density and viscosity of both gases and liquids are measured simultaneously. Importantly, this application also permits in situ measurements and only probes minute fluid volumes 1nL. In contrast, established rheological techniques such as cone and plate and Couette rheometry Bird et al. 1960 measure the viscosity of liquids only, require comparatively large sample volumes, and are incapable of in situ measurements. To illustrate the utility of microcantilever rheometry, we present detailed measurements on a range of different gases and liquids. The measured densities and viscosities are compared against published literature values, where good agreement is found. II. METHODS AND FUNDAMENTAL PRINCIPLES We begin by outlining the principles involved in using microcantilevers in rheological measurements, and discuss the theoretical model Sader 1998 used to determine the density and viscosity from the measured frequency response. A schematic depiction of the principles of the technique is given in Fig. 1. A. Theoretical model In this model, the microcantilever is assumed to have a uniform cross section over its entire length, and its length greatly exceeds its width. For a cantilever beam of rectangular cross section, it is also assumed that the width greatly exceeds the thickness of the cantilever. These assumptions are satisfied by many microcantilevers found in practice. It has been established that the frequency response of a cantilever beam is well approximated by that of a simple harmonic oscillator, when the quality factor Q greatly exceeds unity. That is, in the neighborhood of each resonant peak, the amplitude frequency response A( ) is given by A A 0 R 2 2 2 2 2 R 2 R Q 2, 1

RHEOLOGICAL MEASUREMENTS USING MICROCANTILEVERS 893 FIG. 1. Schematic illustrating the technique of using a microcantilever in rheological measurements of fluids. where A 0 is the zero-frequency amplitude of the response, is the radial frequency, R is the radial resonant frequency of the mode in question in the absence of dissipative effects, and Q is the quality factor. We emphasize that, in general, R does not correspond to the frequency at which the peak in the resonance occurs. The frequency R shall, henceforth, be referred to simply as the resonant frequency. The respective formulas for R and Q are vac R, 2a b 2 1 4 r R 4 b 2 r R Q, 2b i R where r and i are, respectively, the real and imaginary components of a hydrodynamic function ( ), vac is the resonant frequency of the cantilever in vacuum for the mode in question, b is the width of the cantilever, is the density of the fluid, and is the mass per unit length of the cantilever; for a rectangular cantilever c bh, where c is the density of the cantilever and h its thickness. The hydrodynamic function ( ) is detailed by Sader 1998 and depends on the geometry of the cross section of the cantilever and the viscosity and density of the fluid only. For a rectangular cross section, whose width b greatly exceeds its thickness h, an explicit analytical formula for the hydrodynamic function ( ) is given in Eq. 20 of Sader 1998, and numerical data in Sader et al. 1999. For this case, the hydrodynamic function depends only on the width of the cantilever b, and the density and viscosity of the fluid. Note that the accuracy of Eq. 2 decreases as the mode number increases, and the best accuracy is expected for the fundamental mode. In addition, Chon et al. 2000 demonstrated that Eqs. 1 and 2 describe the shape and position of the resonance peak accurately, provided the frequencies are restricted to the immediate neighborhood of the peak and the quality factor Q 1. Consequently, Eqs. 1 and 2 can be used to predict the frequency response of cantilevers in gases

894 BOSKOVIC ET AL. where Q 1, typically and liquids where Q 1, typically, and in turn can be employed to determine the density and viscosity of both gases and liquids. Even so, the complete theory from which Eqs. 1 and 2 are derived can be used to determine the density and viscosity, but this is significantly more complex and is, therefore, not examined here. We now provide details on determining the density and viscosity from the measured frequency response. B. Calibration of cantilever The density and viscosity are determined using Eq. 2, where it is evident that the linear mass density and vacuum frequency vac of the cantilever in question are required. Importantly, if these quantities are unknown, they can be calculated easily from measurements of the resonant frequency R and quality factor Q of the cantilever in a single reference fluid, whose density and viscosity are known, e.g., air. From Eqs. 2a and 2b we obtain the following explicit and independent formulas for the vacuum frequency and linear mass density: vac R 1 1/2 r R, 3a Q i R b2 4 Q i R r R. 3b Substituting the measured resonant frequency R and quality factor Q in the reference fluid into Eq. 3, then enables the explicit calculation of the vacuum frequency vac and linear mass density for the cantilever in question. C. Measurement of viscosity and density To determine the density and viscosity of an unknown fluid, we measure the frequency response of the cantilever in this fluid. Equation 1 is then fitted to the immediate neighborhood of the resonant peak to determine the parameters R and Q. This is best performed using a nonlinear least-squares-fitting procedure for example, using MATHEMATICA, Wolfram Research Inc., IL. The fitted values for R and Q are then substituted into Eq. 2 and the two simultaneous equations solved numerically, using a root finding procedure such as Newton s method Scheid 1968, to determine the density and viscosity. Given the simplicity of the equations and this procedure, the viscosity and density can be determined very rapidly. III. EXPERIMENTAL RESULTS We now examine the utility of the proposed microcantilever technique by presenting measurements on a range of different fluids with known densities and viscosities. A single microcantilever was used throughout to demonstrate the universal applicability of the technique to both gases and liquids. This cantilever was procured from Park Scientific Instruments, Sunnyvale, CA, is made of single-crystal silicon, and has a highly uniform rectangular geometry, with accurately controlled material properties and dimensions. Its length is 397 m, width 29 m, and thickness 2 m, which satisfy the geometric criterion of the theoretical model, described above. This cantilever has no imaging tip, and its vibration properties in fluids were examined in detail by Chon et al. 2000. This cantilever was chosen to eliminate any possible systematic error that would otherwise result

RHEOLOGICAL MEASUREMENTS USING MICROCANTILEVERS 895 FIG. 2. Measured thermal noise amplitude spectrum of microcantilever in air, showing fit solid line to Eq. 1. The measured fitting parameters, resonant frequency R, and quality factor Q, are also given. Spectrum is normalized so that the peak is at unity. from a nonideal geometry. We also emphasize that complete details of the microcantilever material composition are not needed, since all that is required is its linear mass density and vacuum frequency vac, which can be easily measured using a reference fluid, as was discussed above. Indeed, microcantilevers constructed from composite materials may also be used. Several different gases and liquids were used to assess the validity and accuracy of the technique: gases: air, CO 2, Ar, He, H 2 ; liquids: acetone, CCl 4, water, 1-butanol. Density values of these fluids range from 0.08 to 1590 kg m 3, whereas viscosity values range from 8.8 10 6 to 2.5 10 3 kg m 1 s 1 Weast 1985. Measurements of gas and liquid properties were made in the fluid cell of the AFM, which consists of a white silicone rubber O ring as the cell wall and a lucite cell roof and silicon wafer base. For gas measurements, gas of purity 99.9% was bubbled through the cell and then into a water solution, which prevented diffusion of air back into the cell. All measurements were performed at room temperature and pressure. The temperature was monitored using a small type-k thermocouple that was inserted into the fluid cell of the AFM, and was found to be 27 C. To measure the frequency response in fluids, the motion of the cantilever due to thermal fluctuations was monitored, i.e., the thermal noise spectrum was measured. This ensured that any external effects, such as modification of the cantilever frequency response due to acoustic modes in the fluid cell, were eliminated. Thermal noise spectra were measured using a National Instruments data acquisition card AT-MIO-16E-1 board, National Instruments, Austin, TX, carrying out a digital fast Fourier transform of the signal using LABVIEW software National Instruments, Austin, TX. Details of this procedure are given by Chon et al. 2000. To measure the density and viscosity of the fluid in question, only the fundamental resonance peak was considered, for reasons given above. Equation 1 was then fitted to the immediate neighborhood of this resonance peak using a nonlinear least-squares-fitting procedure MATHEMATICA, Wolfram Research Inc., IL. In line with Chon et al. 2000, a white-noise floor, i.e., noise with a constant spectral density, was also included to ensure accurate fits. An example of the resulting fit for a sample fluid air is given in Fig. 2. Note the excellent fit of Eq. 1 to the experimental data, in the neighborhood of the resonant peak. With R and Q determined, the viscosity and density were then deter-

896 BOSKOVIC ET AL. TABLE I. Viscosities and densities of five gases and measured resonant frequencies f R ( R /2 ) and quality factors Q, obtained by fitting Eq. 1 to fundamental resonant peaks. Results using microcantilever MCR, and published literature data exact Weast 1985. All values are at a temperature of 27 C. Air was used as the reference fluid to determine the vacuum frequency and mass density of the cantilever, namely, 17.48 khz and 2325 kg m 3, respectively. Gas f R khz Q Viscosity ( 10 5 kg m 1 s 1 ) Density kg m 3 Exact MCR Exact MCR Air 17.31 55.5 1.9 1.9 1.2 1.2 CO 2 17.27 54.7 1.5 1.5 1.8 1.7 Ar 17.27 43.5 2.3 2.4 1.6 1.5 He 17.42 90.4 2.0 2.2 0.16 0.15 H 2 17.45 185 0.9 1.0 0.08 0.08 mined using the above procedure. The vacuum frequency and density of the cantilever were also measured using the method described above, using air as the reference fluid, and were found to be 17.48 khz, and 2325 kg m 3, respectively. This measured density of the cantilever compares favorably with the literature value of 2328 kg m 3 for singlecrystal silicon Weast 1985. In Table I, the measured densities and viscosities of the five different gases are presented, together with a comparison with published literature values and the measured resonant frequencies R and quality factors Q. Results for air are included only for completeness, since air was used as a reference fluid to determine the vacuum frequency and cantilever density. In all other gases, note the good agreement for both viscosity and density. Even for the lightest gases, helium and hydrogen, where the resonant frequency of the cantilever is very close to the calculated vacuum frequency see Table I, good agreement with published literature values for the density and viscosity is found. We note that any error in the vacuum frequency, which is determined from the resonant frequency in air as detailed above, will manifest itself most significantly for these gases. The good agreement observed in these cases illustrates the accuracy of the calibration procedure described above for determining the vacuum frequency and mass density of the cantilever, and the robustness of the procedure. It is important to note that the viscosity and density of fluids with very similar properties, such as CO 2 and Ar, can be easily distinguished with this technique; measured spectra for these gases are presented in Fig. 3, and the corresponding measured rheological properties are given in Table I. The reason for this arises from the property that the frequency response, and hence the resonant frequency R and quality factor Q, are determined by combination of the effects of density and viscosity. Hence, only fluids with identical density and viscosity will give the same resonant frequencies R and quality factors Q; a difference in only one of these parameters will affect both the density and viscosity. As an example, consider CO 2 and Ar, which have similar yet significantly different rheological properties. Interestingly, we find that the measured spectra in these two fluids have nearly identical resonant frequencies, yet their quality factors are different, see Table I and Fig. 3. Given the excellent fit of Eq. 1 to the experimental data, as illustrated for air in Fig. 2, the difference in the quality factors between these two fluids is easily resolved. This difference in the quality factors greatly exceeds the accuracy 1% with which it can be measured, and this enables the rheological properties to be distinguished easily, see Table I.

RHEOLOGICAL MEASUREMENTS USING MICROCANTILEVERS 897 FIG. 3. Measured thermal noise amplitude spectra of microcantilever in Ar and CO 2. The resonant frequencies and quality factors are given in Table I and are measured to within accuracies of 0.1% and 1%, respectively. Spectra are normalized so that the peaks are at unity. In Table II, corresponding measurements for four liquids are presented. Again, note the good agreement between the microcantilever results and the published literature values, with errors of less than 14% in all cases. These errors are consistent with the errors in the theoretical model for the frequency response, which were found to be less than approximately 10% Chon et al. 2000. Taken together with the above results for gases, Table I, these results demonstrate that a single microcantilever can be used to measure the density and viscosity of fluids whose values span several orders of magnitude. Furthermore, only single measurements of the frequency response of the cantilever are required, since these are extremely reproducible. Therefore, multiple measurements of the frequency response on a single sample do not influence the accuracy of the density and viscosity. We note that similar results and accuracy to those presented in Tables I and II were obtained using a rectangular microcantilever with dimensions 197 29 2 m Chon et al. 2000. However, the accuracy of the theoretical model, and hence the rheological technique, decreases as the ratio of the cantilever length to its width decreases. Thus, cantilevers of high length-to-width aspect ratio should always be used. TABLE II. Viscosities and densities of four liquids and measured resonant frequencies f R ( R /2 ) and quality factors Q, obtained by fitting Eq. 1 to fundamental resonant peaks. Results using microcantilever MCR, and published literature data exact Weast 1985. All values are at a temperature of 27 C. Liquid f R khz Q Viscosity ( 10 4 kg m 1 s 1 ) Density ( 10 3 kg m 3 ) Exact MCR Exact MCR Acetone 6.55 2.9 3.1 3.2 0.79 0.75 CCl 4 4.51 2.0 8.8 9.9 1.6 1.5 Water 5.39 2.0 8.6 8.6 1.0 0.97 1-butanol 4.86 1.2 25 28 0.81 0.80

898 BOSKOVIC ET AL. IV. DISCUSSION AND CONCLUSIONS All measurements presented above were performed using the AFM, and the deflections of the microcantilevers were measured using the optical deflection technique. However, there is nothing precluding the use of microcantilevers in isolation from the AFM. All that is required is a detection system to monitor the frequency response of the microcantilevers in fluid. Depending on the detection system used, this would also allow the use of microcantilevers in rheological measurements of opalescent or turbid media. In addition, the microcantilever could be actively excited to eliminate any signal-to-noise problems, as would be the case in monitoring the thermal noise spectrum. The above practically realizable modifications would result in a technique that is applicable to a wide range of media, permitting in situ and rapid monitoring of the rheological properties of fluids. It should be noted that the model used to extract the fluid density and viscosity from the measured frequency response requires the quality factor Q 1. If this is not the case, then the analogy with the response of a simple harmonic oscillator is not valid. This immediately indicates that Eqs. 1 and 2 cannot be used in such cases. We have found previously Chon et al. 2000 that for fluid/cantilever combinations where heavy damping is present (Q 1), it is impossible to fit the measured spectra to the response of a simple harmonic oscillator. In such cases, the density and viscosity of the fluid cannot be measured using the equations given here. However, this does not place a fundamental restriction on the method, since by using a different cantilever in the same fluid a different quality factor is obtained. Provided this quality factor is larger than unity, the methodology presented here is applicable. Consequently, to fully utilize the capabilities of the microcantilever technique, it would be desirable to use a range of different cantilevers, preferably in a cantilever array configuration. This would enable fluids with a wider range of properties to be measured than those accessible using a single microcantilever, as demonstrated here. Finally, we note that microcantilevers only probe minute volumes of fluid 1nL. This is because the hydrodynamic length scale of a cantilever vibrating in fluid is dictated by its width and length. Consequently, any fluid at a distance greater than several cantilever widths away from the cantilever does not influence the measurement. This favorable property then allows local and in situ rheological measurements of both gases and liquids, which probe less than 1 nl of fluid, to be performed on large volumes of fluid. Spatially resolved rheological measurements are, therefore, possible using microcantilevers. These favorable properties are in stark contrast to those of conventional rheological techniques Bird et al. 1960, which probe comparatively large volumes of fluid, are not applicable to both gases and liquids, and are incapable of in situ measurements. Microcantilever rheometry, therefore, provides the exciting possibility for novel rheological measurements on small fluid samples of wide ranging properties. ACKNOWLEDGMENTS The authors would like to thank Professor David Boger for many interesting discussions. This research was supported in part by the Particulate Fluids Processing Center of the Australian Research Council and by the Australian Research Council Grants Scheme.

RHEOLOGICAL MEASUREMENTS USING MICROCANTILEVERS 899 References Ahmed, N., D. F. Nino, and V. T. Moy, Measurement of solution viscosity by atomic force microscopy, Rev. Sci. Instrum. 72, 2731 2734 2001. Bergaud, C. and L. Nicu, Viscosity measurements based on experimental investigations of composite cantilever beam eigenfrequencies in viscous media, Rev. Sci. Instrum. 71, 2487 2491 2000. Berger, R., Ch. Gerber, H. P. Lang, and J. K. Gimzewski, Micromechanics: A toolbox for femtoscale science: Towards a laboratory on a tip, Microelectron. Eng. 35, 373 379 1997. Bird, R. B., W. E. Stewart, and E. N. Lightfoot, Transport Phenomena Wiley, New York, 1960. Chon, J. W. M., P. Mulvaney, and J. E. Sader, Experimental validation of theoretical models for the frequency response of atomic force microscope cantilever beams immersed in fluids, J. Appl. Phys. 87, 3978 3988 2000. Fu, Y. and W. G. Price, Interactions between a partially or totally immersed vibrating cantilever plate and the surrounding fluid, J. Sound Vib. 118, 495 513 1987. Lindholm, U. S., D. D. Kana, W.-H. Chu, and H. N. Abramson, Elastic vibration characteristics of cantilever plates in water, J. Ship Res. 9, 11 22 1965. Oden, P. I., G. Y. Chen, R. A. Steele, R. J. Warmack, and T. Thundat, Viscous drag measurements utilizing microfabricated cantilevers, Appl. Phys. Lett. 68, 3814 3816 1996. Sader, J. E., Frequency response of cantilever beams immersed in viscous fluids with applications to the atomic force microscope, J. Appl. Phys. 84, 64 76 1998. Sader, J. E., J. W. M. Chon, and P. Mulvaney, Calibration of rectangular atomic force microscope cantilevers, Rev. Sci. Instrum. 70, 3967 3969 1999. Scheid, F. J., Theory and Problems of Numerical Analysis McGraw-Hill, New York, 1968. Weast, R. C., CRC Handbook of Chemistry and Physics CRC, Boca Raton, FL, 1985.