Evaluation of Capillary Columns for General Performance Parameters

Similar documents
Column Dimensions. GC Columns and Consumables. Mark Sinnott Application Engineer. March 12, 2010

Secrets of GC Column Dimensions

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Trajan SGE GC Columns

Ultra-Inert chemistry for Trace Level Analysis

New ZB-5HT Inferno The World s Highest Temperature Non-Metal GC Column

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

Author. Abstract. Introduction

Water Injections in GC - Does Water Cause Bleed?

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Product Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas

Selection of a Capillary

Activity in the FID Detection Port: A Big Problem if Underestimated

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

Selection of a Capillary GC Column

The Analysis of Organophosphate Pesticides by LC/MS Application

Chemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography.

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

Understanding Gas Chromatography

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Theory and Instrumentation of GC. Chromatographic Parameters

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

Introduction to Capillary GC

Volatile organic compounds (VOCs):

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Speakers. Moderator. John V Hinshaw GC Dept. Dean CHROMacademy. Tony Taylor Technical Director CHROMacademy. Dave Walsh Editor In Chief LCGC Magazine

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications

Virtually Particle-Free Rt -Silica BOND Columns

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

GC Analysis of Polybrominated Flame Retardants Application

Introduction to Capillary GC

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

How To Select the Correct GC Column. Simon Jones Application Engineer

Application Note. Abstract. Authors. Introduction

Understanding the Capillary GC Column: How to Choose the Correct Type and Dimension

Continuous Improvement in Petroleum/Petrochemical Analysis HP s Family of Innovative PLOT Columns

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Fast and Accurate Analysis of PBDEs in a Single Run, Including BDE-209

Hypersil BDS Columns TG 01-05

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline

Fast and Accurate Analysis of PBDEs in a Single Run, Including

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

The Suite for Environmental GC Analysis

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products.

Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System

So Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Improved Peak Shape Performance and Consistency for Analysis of Challenging Polar Compounds

CHAPTER 6 GAS CHROMATOGRAPHY

GAS CHROMATOGRAPHY. Mobile phase is a gas! Stationary phase could be anything but a gas

Turkish raki is the most popular traditional distilled

Headspace Technology for GC and GC/MS: Features, benefits & applications

GAS CHROMATOGRAPHY (GC)

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Industrial Applications with a New Polyethylene Glycol-Based GC Column

environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Application. Gas Chromatography February Introduction

Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Application. Introduction. Authors. Abstract. Hydrocarbon Processing

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Characterization and Classification of Heroin from Illicit Drug Seizures Using the Agilent 7200 GC/Q-TOF

Application Note. Abstract. Authors. Environmental

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Size Exclusion Chromatography: Method Development

A New Web-Based Application for Modeling Gas Chromatographic Separations. Dr. Hansjoerg Majer Restek Corporation Market Development Manager Europe

AN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION

Transcription:

Evaluation of Capillary s for General Performance Parameters Application Authors Mitch Hastings, Eberhardt R. Kuhn and Allen K. Vickers Agilent Technologies, Inc 91 Blue Ravine Road Folsom, CA 95630 USA Abstract In recent years, new GC capillary columns have entered the marketplace accompanied by claims of performance equivalent to those from established manufacturers. This application note examines a sampling of these columns, comparing their performance using standard and practical test procedures. Tests were for column bleed, retention index, film thickness, and trace level acid and base performance. Both 5% phenyl- and wax columns were included in this study. Introduction The purpose of this application note is to compare, in a rigorous and quantitative way, the performance of Agilent DB- and HP- brand capillary GC columns against a variety of competitive columns claiming performance equivalence. See Figure 1. Figure 1, derived from a competitor's presentation, purports to make a valid comparison. It concludes that the only difference between the columns presented is their price. Is this a valid conclusion? Closer examination shows that it is not. In this example, the sample is a mixture of phenols, containing 10 ng per component. A lower concentration, yielding an on-column amount between 1 and 5 ng per component, would present a more discriminating challenge. Given the small graphical size of the chromatograms depicted and the fact that no numerical data is presented, it is difficult to discern important differences between the columns that may exist. Figure 1. Competitor s column comparison. Additionally, analysis of phenols tells only part of the story. In the real world, general utility columns such as the 5% phenyl type columns (HP-5 and DB-5 for instance) see a broad range of analytes. While there are two difficult phenols in this mixture (,4-dinitrophenol and pentachlorophenol), no other class of compounds, such as bases, are examined. Also, there are claims of low column bleed, but no data is presented to support this claim. The reason you do not see them being compared will be shown.

The brands compared for this study were DB-, HP-, and two column brands new to the marketplace that are being aggressively marketed based on price. The dimensions of all columns tested were 30 m 0.5 mm id, 0.5 µm film thickness. General Comparison Methodology First, all columns were compared using Agilent Technologies quality control (QC) conditions and parameters for the DB-5 column. The parameters tested were column bleed (offset from 135 C to 35 C after hours of conditioning at 35 C), Retention Indices, and film thickness (measured through k). Then, the effects of thermal-cycling stress on column inertness were monitored with replicate injections of chromatographically active analytes at trace levels (5 ng on-column) in the form of an active acid (,4-dinitrophenol), an active base (noctylamine), a very difficult to chromatograph ethanolamine (N,N'-diethylethanolamine), and a chromatographically inert internal standard (naphthalene) at 10 ng on-column. On-column injection was used to eliminate the inlet as a source of activity. A flame ionization detector (FID) was used to ensure long term stability. Test order was randomized and replicate injections (n=6) were performed for inertness testing. Oncolumn probe amounts were the same, verified by naphthalene to within a range of ±3%, and determined to be statistically the same using comparison testing at the 95% confidence level. Flow rate was adjusted in order to elute naphthalene at 9.18 ±.10 min for all columns so that valid peak height comparisons could be made. Six replicates of the standard were injected at each of the following points: Initial After 5 cycles to 350 C After 50 cycles to 350 C After 5 cycles to 370 C After 50 cycles to 370 C Finally, after the temperature cycling, the columns were compared again using the DB-5 QC test criteria as described above. Initial Bleed Out/Condition Times pa 50 40 30 0 10 0 Figure. : 5% Phenol 30m X 0.5mm I.D., 0.5µm Oven: 75 C for 0.5 min 75-350 C max ramp rate 350 C hold for min Carrier: Hydrogen, ml/min, constant pressure Injector: Cool on-column, air blank Detector: FID, 350 C Manufacturer Z 4 6 8 10 Time (min) Entire chromatogram of initial bleed profile. General QC results-bleed FID Signal (pa) 0 18 16 14 1 10 8 6 4 0 3.9.61 1.54 1.05 4.35 4.89 HP-5 DB-5 Brand Z 18.05 16.5 Manufacturer V Agilent HP-5 Agilent DB-5 Some competitors claim that their columns bleed equally or less than DB- and HP- columns. The bleed specifications for DB-5 and DB-5ms are 6 and 4 pa respectively, using an FID as described above. See Figure 3. Figure 3. One-hour bleed Two-hour bleed bleed.41.31 bleed at one and two hours of conditioning. While one competitor s claim is valid, two others did not meet their low-bleed claims. However, the one that did meet the DB-5ms bleed specification,, had very serious inertness problems. See Figure 5. Additional QC Results The columns did perform similarly to DB-5 and HP-5 columns in the parameters of Retention Index and film thickness (as measured by k), all falling within the allowable range for the J&W brand column.

Inertness activity (lack of inertness) is solute adsorption or worse, reaction of the solute by or with the column solute adsorption is chromatographicaly evident as peak shape anomolies (e.g., tailing) or in extreme cases, complete loss of solute response. It is particularly problematic at trace levels, especially on the competitors' columns. See Figure 4. Why is this important? If you are analyzing a broad range of compounds, for which the 5% phenyl columns are especially useful, you may be getting incomplete data. If you are trying to determine trace amounts of impurities, you may miss them because the column retains them or changes them chemically. NOTE: The diethylethanolamine is likely a reaction/decomposition product of the N,N'-diethylethanolamine according to MS analysis. This was seen on all competitors' columns, and to a small degree on the DB-5, especially after thermal cycling to 370 C. The competitors columns perform well only for acidic compounds, but performed poorly for basic compounds and compounds with active, mixed functional groups. GC run conditions for Figure 4 were as follows: Injector: s: Cool-on-column, 0.5 µl, MeCl solvent 30 m 0.5 mm, 0.5 µm film Temperature program: 40 C, 0.5 min 40 175 C at 10 /min 175 300 C at 30 /min 300 35 C at 0 /min 35 C for min Carrier: Detector: Hydrogen, (99.995 plus big universal trap), t r of Naphthalene = 9.18 min ±0.10 min FID, 35 C, N makeup, column compensation flow pa 3.5 HP-5 30.0 7.5 1 3 4 5.0.5 0.0 17.5 15.0 1.5 10.0 4 6 8 10 1 14 min pa 3.5 30.0 7.5 5.0.5 Manufacturer B 3 4 Peak Identification 1. Diethylethanolamine. Octylamine 3. Naphthalene 4.,4-Dinitrophenol 0.0 17.5 15.0 1.5 1? 10.0 4 6 8 10 1 14 min Figure 4. Chromatograms showing typical performance for the trace level compounds on the DB-5 and HP-5 (top) columns, and all of the other competitors' columns (bottom). 3

inertness toward trace active analytes can be measured and compared in two ways: Peak area less on one column means permanent loss to the column (irreversible adsorption). Peak height less on one column means more activity, but doesn t necessarily mean irreversible loss to the column. The average relative response factors (ARRF) to naphthalene are used to minimize for injection variation from column to column. a. d. Area RRF of N,N'-Diethylethanolamine (5 ng on column) Figure 5 is a collection of charts showing the initial height, area, and ARRFs for all active probes on all columns tested after conditioning but prior to thermal cycling. Note the extremely poor base and active mixed functional group performance on all competitors' columns compared to DB-5 and HP-5 columns, but essentially equivalent acid (,4-dinitrophenol) performance. Practical interpretation of these results is that you can chromatograph trace acid, base, and mixed functional group compounds on HP-5 and DB-5 columns at trace levels, but reliably, you cannot on the competitors columns. Height RRF of Octylamine (5 ng on column) 0.160000 0.50000 0.140000 RRF N,N'-Diethylethanolamine 0.10000 0.080000 0.060000 Initial height HP-5 DB-5 Brand Z 0.040000 0.00000 HP-5 DB-5 Brand Z b. e. 0.160000 0.140000 0.10000 0.080000 0.060000 0.040000 0.00000 Height RRF of N,N'-Diethylethanolamine (5 ng on column) Height RRF N,N'-Diethylethanolamine (5 ng on column) 0.350000 0.50000 Area RRF of,4-dinitrophenol (5 ng on column) Initial area HP-5 DB-5 Brand Z HP-5 DB-5 Brand Z type c. f. Area RRF of Octylamine (5 ng on column) 0.500000 0.450000 0.400000 0.350000 Initial area 0.50000 HP-5 DB-5 Brand Z 0.50000 Height RRF of,4-dinitrophenol (5 ng on column) Initial height HP-5 DB-5 Brand Z Figures 5a-f. Initial AARF for all columns for active probes. n=6 for all samples, n = 1 for each column type. 4

Effects of Thermal Cycling Lifetime Factually and definitively there is not a good way of predicting any given column's lifetime. Relatively speaking, there are many factors that influence useful column lifetime. Most columns fail during use due to contamination from sample matrices, degradation from injection of inorganic acids, bases or salts and oxygen in the carrier gas at high temperatures. The durability of siloxane polymers against abuse is debatibly equal for all columns. Take that away, and the only valid means of comparing column lifetime is by comparing how well columns hold up to thermal stresses in terms of inertness and column bleed under carefully controlled conditions. Bleed is nothing more than degraded stationary phase eluting from the column and creating a signal on the detector. This is more a function of the column manufacturers proprietary deactivation procedures and the control they have over their manufacturing processes. A way of academically measuring column life is to measure and use the stationary phase bleed rate at a given column temperature, correlate it with the response for a known column bleed component, and make relative comparisons of columns assuming only thermal stress. Using the -hour conditioning bleed values shown in Figure 3, we can compare the relative lifetimes of the columns. See Table 1. This is an academic exercise and not meant to imply lifetime for reasons already stated. The fact that some competitors columns bleed at rates so much higher suggests that when the variables of sample matrices and oxygen are removed, DB-5 and HP-5 columns have the longest life, does not perform too poorly, but Brands Z and V make a poor showing. Table 1. Summary of Bleed Characteristics HP-5 DB-5 Brand Z Bleed at 35 C in.61 1.05 4.89 16.5.31 pa ( h value) Total Phase in 4.71 4.71 4.71 4.71 4.71 column (mg) Bleed rate (ng/s) 0.37 0.095 0.445 1.477 0.10 Thermal column 30 5711 3 36.9 60 Life at 35 C Effects of Thermal Cycling-Inertness A column rarely gets better, and in particular, more inert with use once it is properly conditioned. What follows are the ARRF for height and area plotted throughout the thermal cycling test for each one of the probes on each of the columns. Use the following key when viewing the Figures 6a-c: 1 = Initial = 5 cycles 350 C 3 = 50 cycles 350 C 4 = 5 cycles 370 C 5 = 50 cycles 370 C Unfortunately, two of Brand Z s columns rated to 370 C broke between the first and second rounds of replicate testing, so only the first round is truly validly comparable. However, the data is still shown. Unless special high-temperature polyimide tubing is used in capillary columns, a thermal limit of 360 C for the tubing will give a reasonable lifetime. The fact that two columns broke and also appeared very brittle, and that nearly all tubing used in GC columns comes from one manufacturer, indicates that the robustness of these columns may be compromised by what could be their higher than appropriate thermal limit (370 C). A few generalizations can be made about Figure 6. HP-5 and DB-5 are superior columns for analyzing bases and mixed function compounds and appear to offer stable inertness when staying within their normal operating range (up to 350 C). appears to improve slightly with continued operation, which suggests that the column may not be fully stabilized by the manufacturer prior to shipping. All columns are fairly close in response for the,4-dinitrophenol, with the HP-5 showing a dip after cycling to 370 C and a continued drop in response, but it was still close to that of DB-5 and other brand columns. Bleed One competitor exhibited much higher bleed. Only one competitor met the J&W DB-5ms bleed specification. Comparable bleed comes at a price with competitors columns. Assume value of 11 pas/nanogram for bleed components 5

Area RRF N,N'-Diethylethanolamine vs. replicate set 0.350000 0.50000 6a. HP-5 DB-5 Brand Z 0 1 3 4 5 6 Replicate set Area RRF of Octylamine vs. replicate set 0.600000 0.500000 0.400000 6b. HP-5 DB-5 Brand Z 1 = Initial = 5 cycles 350 C 3 = 50 cycles 350 C 4 = 5 cycles 370 C 5 = 50 cycles 370 C 0 1 3 4 5 6 Replicate set Area RRF,4-Dinitrophenol vs. replicate set 0.350000 0.50000 6c. HP-5 DB-5 Brand Z Figure 6. 0 1 3 4 5 6 ARRF for all columns as a function of thermal cycling. 6

Inertness The price of lower bleed is a significant response loss for bases and di-functional compounds. Selectivity All appear equivalent within normal variation, before and after thermal cycling (data not shown). Mechanical Stability Higher temperature limit, competitor s columns seem more prone to breakage. Problem may be a side affect of overstressing the tubing s temperature limits to achieve low bleed status. Activity at trace levels HP-5 is most universal ( analyte ph neutral ) and therefore is most robust for acids, bases, and mixed function compounds. DB-5 is less robust, but it exhibits some low-level universal performance. All other columns are good for acids only. Base and mixed function compound performance is very poor. Performance for bases diminishes for all columns with extended thermal cycling to 370 C. Additional Comparsions Design of the column comparison methods above is fairly exhaustive work, so the scope of this project was limited to comparing only 5% phenyl capillary columns. However, in early applications, comparison of other competitors wax columns with our DB- Wax column makes it worth noting that the implied equivalence in their advertising statements is unsupported. For such claims to have any use, most analysts would expect to be able to substitute one of these replacement columns for a DB- or HP- column and essentially obtain the same results. Resolution, relative elution order, and column selectivity should be nearly equal under the same conditions. However, as the following data show, replacement does not mean equivalent, and not recognizing this semantic distinction could jeopardize a lab's time, money, and potentially its reputation. A comparison of the industry standard, DB-WAX column and another wax column from a manufacturer, who claimed that their columns replace DB-WAX columns, was made. A typical application, adulterants and impurities commonly found in distilled alcohol from biological sources, using a sample containing only 18 compounds was chosen. The separation was relatively simple and was readily accomplished on a DB-WAX column. See Table. Table. Comparison of Retention Time and Retention Index (RI) for Test Compounds on DB-WAX and "Replacement" s DB-WAX Replacement DB-WAX Replacement RI Compound time (min) time (min) RI RI difference 1 Methane 1.184 1.195 Pentane 1.30 1.36 500.0000 500.0000 n/a 3 Acetaldehyde 1.430 1.436 704.1436 715.015 11.0579 4 Methyl formate 1.587 1.890 764.40 775.36 10.9906 5 Propionaldehyde 1.700 1.694 794.3359 803.693 9.934 6 Acetone 1.818 1.806 819.410 88.317 _8.815 7 Methyl acetate 1.889 1.87 83.334 840.6943 8.3601 8 Butyraldehyde.13.181 878.3749 886.3760 8.001 9 Ethyl acetate.317.71 890.0975 896.9889 6.8913 10 Acetal.366.71 895.65 896.9889 1.7364 11 Nonane.413.98 900.0000 900.0000 n/a 1 Methanol.446.39 13 Methyl propionate.505.450 14 Isopropanol.85.744 15 Ethanol 3.01.90 16 1-Propanol 5.150 4.879 17 Isobutanol 7.653 7.117 18 1-Butanol 10.64 9.917 19 Active amyl alcohol 16.41 15.303 0 Isoamyl alcohol 16.44 15.303 7

The comparison was done under identical conditions of sample, temperature, column dimensions, and carrier gas linear velocity. The separations were all performed isothermally, thus minimizing any minor differences in column dimensions that may affect this comparison. In addition to these components, we incorporated a few hydrocarbons to check retention indices (RI), a measure of selectivity. If these measurements are the same (within normal variation, generally ± RI units for this phase type), then the replacement column really could be considered an equivalent, and the manufacturer s claim could be considered practical. However, as the data in Table show, the selectivity of these columns is very different. The chromatograms in Figures 7 and 8 show that these same columns are far from direct replacements. If you replaced a DB-Wax with one of these other wax columns, you would: Lose resolution of ethyl acetate and acetal Lose all resolution existed between active amyl and isoamyl alcohol Further, you can optimize separation between ethanol and isopropanol without compromising other separations simply by increasing the linear velocity on a DB-WAX column. 6 Resolved 3 4 5 7 9 1 13 DB-WAX 10 1 8 11 C1096a 1. 1.4 1.6 1.8.0..4 Time (min) 6 9,10 3 4 5 7 1 Not resolved 13 Replacement wax 1 8 11 C1096b 1. 1.4 1.6 1.8.0..4 Time (min) Figure 7. Comparison of DB-WAX and a Replacement column Chromatograms; Earliest eluants. 8

1-13 shown in Figure 7 15 16 17 18 14 Partially resolved 19 0 DB-WAX C1097a 4 6 8 10 1 14 16 18 Time (min) 1-13 shown in Figure 7 15 16 17 18 14 19,0 Replacement wax Not resolved C1097b 4 6 8 10 1 14 16 18 Time (min) Figure 8. Comparison of DB-WAX and a Replacement column Chromatograms; Later eluants. Ethyl acetate and acetal and active amyl and isoamyl alcohol could not be resolved on the replacement column under the range of conditions used for this comparison. Often, with these replacement columns, the manufacturer claims that the only difference is the price. For even simple applications like that shown in Figure 7, when you factor in the time it takes to reverify elution order and the potential risk that either the column will not work for your application or an elution order change might be missed, minor savings in the price of a column are erased and may rapidly become major liabilities. Most analysts would agree that consistent column performance from one column to the next is the most prized column performance characteristic. For More Information For more information on our products and services, visit our web site at www.agilent.com/chem. 9

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 003 Printed in the USA March 0, 003 5988-899EN