HI Propylene Glycol Refractometer INSTRUCTION MANUAL

Similar documents
HI Refractometer for Sodium Chloride Measurements INSTRUCTION MANUAL

Instruction Manual HI Refractometer for Sodium Chloride Measurements

HI Propylene Glycol Refractometer

Instruction Manual HI Refractometer for Sucrose in Wine and Grape Products Measurements

Instruction Manual HI Refractometer for Invert Sugar Measurements

INSTRUCTION MANUAL Milwaukee Refractometer MA888 Refractometer for Ethylene Glycol Measurements

Instruction Manual. HI Refractometer for Sucrose in Wine and Grape Products Measurements

INSTRUCTION MANUAL Milwaukee Refractometer MA881 Refractometer for Invert Sugar Measurements

HI Seawater Refractometer

INSTRUCTION MANUAL Milwaukee Refractometer MA886 Refractometer for Sodium Chloride Measurements

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL. Milwaukee Refractometer

INSTRUCTION MANUAL Milwaukee Refractometer MA872 Refractometer for Fructose Measurements

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL. Milwaukee Refractometer

HI 96811, HI HI 96813, HI 96814

MA882, MA883, MA884, MA885

Instruction Manual HI 96759C Maple Syrup Transmittance Analyzer

HI REDUCING SUGARS ISM for wine analysis

Instruction Manual HI 96738C Chlorine dioxide ISM

Instruction Manual HI 96728C Nitrate ISM

HI Hardness ISM

Instruction Manual HI Potassium LR and MR ISM

Instruction Manual HI 96720C Calcium Hardness ISM

Instruction Manual HI 96710C ph & Chlorine ISM

SUGAR/BRIX REFRACTOMETER INSTRUCTION MANUAL

IR/Thermistor Pen. Instruction Manual

HI HI 8633 HI HI 8734 HI

Today s Weather 24 Hour Forecaster

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

OPERATING INSTRUCTIONS

HI Free & Total Chlorine ISM

PHOTOMETER FOR BOAR SEMEN ANALYSIS

Rapid RH Portable ph Meter. User's Manual

Non - Contact Forehead Infrared Thermometer. Care Q7 USER MANUAL - 1 -

Bante210 Benchtop ph Meter Instruction Manual

INSTRUCTION MANUAL. Laser Diode Sarcomere Length. Aurora Scientific Inc. 360 Industrial Parkway South, Unit 4 Aurora, Ontario, Canada L4G 3V7

Eclipse Refractometer General Models

DHS1900 DHS3000 High Temperature Hand Held Infrareds User Manual

PHS-25CW/3BW Benchtop ph Meter Instruction Manual

ph electrode Instruction Manual For use with the RAH-210 & RPH-250

WS-7047TWC Wireless 433 MHz Weather Station With Rainfall and Temperature. Instruction Manual

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

ph / Temperature Meter for Cheese with Application Specific Probe

One-hand ph/temperature measuring instrument

IR F ( C) ENGLISH INSTRUCTION MANUAL. Thermometer

PC100 Handheld ph/cond./tds/sal. Meter. Instruction Manual

InfraRed Thermometer with Laser Pointer

Operating Manual. Turb 355 IR/T. Portable turbidimeter. ba75454e03 09/

Fischer Instruments Chrome and Black Wood Base Weather Station with Barometer, Hygrometer, Thermometer and Quartz Clock User Manual

Model ST-616CT. Instruction Manual. Infrared Thermo-Hygrometer. reedinstruments. www. com

рн PRO Controller The device is used to control and regulate ph level in hydroponic systems and water preparing units automatically USER MANUAL

473-SHX Dew Point Mirror

Owner's Manual. for AUTOMATIC LEVEL. NBL Series. FOR CUSTOMER SERVICE, PARTS & REPAIR, CALL Toll Free:

ph 5 Meter Manual The LaMotte 5 Series

HL-800K Infrared Laser Thermometer. -50 C to +800 C (13:1 ratio) User Manual

VIBRATION METER. Instruction Manual.

Portable ph/orp/ion Meter D-73 Instruction Manual CODE:GZ A

Model: WT-3181PL Instruction Manual DC: Indoor/Outdoor Atomic Wall Clock. Get Started

Mini Environmental Quality Meter

WS-7211U Wireless 433 MHz Weather Station. Instruction Manual

Fischer 1508BTH-45 5" Brass Barometer with Temperature & Humidity User Manual

Chemistry MultiMeasure Sensor PS-2170

SKELESCOPE REFLECTOR TELESCOPE WITH TABLE TOP TRIPOD SK-100 INSTRUCTION MANUAL

KD2. User s Manual. Thermal Properties Analyzer. version 6. Decagon Devices, Inc.

Owner's Manual NSL100B BUILDERS LEVEL NSL500B TRANSIT LEVEL

Guide to ph Analysis.

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base

473 Dew Point Hygrometer

1450 First Ave. Page 1 of 10 Chippewa Falls, WI (800) or (715) Fax: (715)

ICE MANTLE MAKER MODEL 452 User Maintenance Manual/Handbook

473 Dew Point Hygrometer

WeatherHub2 Quick Start Guide

Advanced Sensor Technologies, Inc. TEL: TOLL FREE: WOW-ASTI ( ) FAX:

Fischer Instruments Stainless Steel Barometer User Manual

DIGITAL ABBE REFRACTOMETER INE-WYA-2S OPERATING INSTRUCTION PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

Chematest 25. Operator s Manual A Swan Analytical Instruments AG CH-8340 Hinwil/Switzerland ANALYTICAL INSTRUMENTS

HS515G-M. PARALOW Circle Dot Sight. User's Manual. Holosun Technologies Inc. Phone: (909) Fax: (909)

INSTRUCTION MANUAL DTH Fax: (503)

Anemometer Thermometer Hygrometer

Field Pro Operation and Installation Manual

POTASSIUM ION SELECTIVE ELECTRODE

Instruction Manual. Qubit Fluorometer. probes.invitrogen.com/qubit

INE-PHS-3E ph Meter Operating Manual

STANDARD OPERATING PROCEDURE

This sensor is now also available with a sturdier flat electrode which enables testing even smaller samples than the conventional electrode.

Ambient Weather WS-152B 6" Contemporary Barometer User Manual

Fischer Banjo Weather Station with Thermometer, Hygrometer, Barometer User Manual

GENERAL INSTRUCTIONS GENERAL PREPARATION

Instruction Manual C401/301/201/102/101. Portable Colorimeter. 99 Washington Street Melrose, MA Phone Toll Free

LEAPERS, INC. RED/GREEN DOTS

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

product manual H-3220A Benkelman Beam

MODEL 107 TEMPERATURE PROBE INSTRUCTION MANUAL

N) manual. Biomaster Operating manual

Density of an Unknown

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049

GENERAL INSTRUCTIONS GENERAL PREPARATION

Ambient Weather WS-228TBH 9" Brushed Aluminum Traditional Barometer with Temperature and Humidity, Radiant Blue User Manual

New! 2010 Model. Model No

Coating Thickness Gauge TW-8. art. nr Thoma Tel./Fax: Operation Manual

Transcription:

HI96832 Propylene Glycol Refractometer INSTRUCTION MANUAL

Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using this instrument. This manual will provide you with the necessary information for correct use of this instrument, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e mail us at tech@hannainst.com or view our worldwide contact list at www.hannainst.com. All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.

Preliminary Examination... 4 General Description... 5 Specifications... 6 Principle of Operation... 7 Measurement Units... 8 Functional Description... 9 Display Elements... 10 Measurement Guidelines... 11 Calibration Procedure... 11 Measurement Procedure... 12 To Change Measurement Unit... 14 To Change Temperature Unit... 14 Making a Standard Solution... 15 Battery Replacement... 16 Error Messages... 17 Accessories... 18 TABLE OF CONTENTS 3

PRELIMINARY EXAMINATION Remove the instrument from the packing materials and examine carefully to ensure no damage has occurred during shipping. If any damage has occurred, contact your local Hanna Instruments Office. Each HI96832 instrument is supplied with: 9 V battery Instruction manual Plastic pipette Quality Certificate Note: Save all packing material until you are sure that the instrument functions correctly. A defective instrument must be returned in its original packing. 4

The HI96832 Digital Refractometer is a rugged portable, water resistant device that utilizes the measurement of refractive index to determine the % volume and freezing point of propylene glycol based solutions. Propylene glycol is used in many applications including: pharmaceutical solvent humectant food additive emulsification agent moisturizer in cosmetics, toothpaste, tabacco products, etc. hand sanitizer and lotion antifreeze The HI96832 device benefits from Hanna Instruments years of experience as a manufacturer of analytical instruments. The digital refractometer eliminates the uncertainty associated with mechanical refractometers and is easily portable for use in the field. The HI96832 refractometer is an optical device that is simple and quick to use. Samples are measured after a simple user calibration with distilled or deionized water. Within seconds, the refractive index and temperature are measured and converted into one of two measurement units: % Volume or Freezing Point. The instrument utilizes internationally recognized references for unit conversion and temperature compensation for propylene glycol solutions (e.g. CRC Handbook of Chemistry and Physics, 87 th Edition). The temperature (in C or F) is also displayed on the large dual level display along with helpful message codes. Key features include: IP65 waterproof protection Automatic Temperature Compensation (ATC) Battery operation with Low Power indicator (BEPS) Automatically turns off after 3 minutes of non-use GENERAL DESCRIPTION 5

SPECIFICATIONS Range Resolution Accuracy Temperature Compensation Measurement Time Minimum Sample Volume Light Source Sample Cell Case Material Enclosure Rating Battery Type / Life Auto-Off % Volume: 0.0 to 100.0 %V/V Freezing Point: 0.0 to -51.0 C (32.0 to -59.8 F) C ( F): 0.0 to 80.0 C (32.0 to 176.0 F) % Volume: 0.1 % Freezing Point: 0.1 C (0.1 F) C ( F): 0.1 C (0.1 F) % Volume: ±0.3 % Freezing Point: ±0.5 C (±1.0 F) C ( F): ±0.3 C (±0.5 F) Automatic between 0.0 and 40.0 C (32.0-104.0 F) Approximately 1.5 seconds 100 µl (cover prism totally) Yellow LED Stainless Steel ring and flint glass prism ABS IP65 9V / 5000 readings After 3 minutes of non-use Dimensions 192 x 102 x 69 mm (7.6 x 4.1 x 2.7 ) Mass 350 g (12.3 oz.) 6

Propylene Glycol determinations are made by measuring the refractive index of solution. Refractive Index is an optical characteristic of a substance and the number of dissolved particles in it. Refractive Index is defined as the ratio of the speed of light in empty space to the speed of light in the substance. A result of this property is that light will bend, or change direction, when it travels through a substance of different refractive index. This is called refraction. When passing from a material with a higher to lower refractive index, there is a critical angle at which an incoming beam of light can no longer refract, but will instead be reflected off the interface. PRINCIPLE OF OPERATION The critical angle can be used to easily calculate the refractive index according to the equation: sin (θ critical ) = n 2 / n 1 Where n 2 is the refractive index of the lower-density medium; n 1 is the refractive index of the higher-density medium. In the HI96832 refractometer, light from an LED passes through a prism in contact with the sample. An image sensor determines the critical angle at which the light is no longer refracted through the sample. Specialized algorithms then apply temperature compensation to the measurement and convert the refractive index to: % Volume or Freezing Point. 7

MEASUREMENT UNITS Freezing point is displayed as a temperature from 0.0 to -50.0 C corresponding to 0.0-60.0 % by volume. When the display shows LO, the freezing point is at a minimum (below -51 C). 8

TOP VIEW FUNCTIONAL DESCRIPTION 1. Liquid Crystal Display (LCD) 2. Secondary Display 3. Primary Display 4. READ Key (User Measurement) 5. ZERO Key (User Calibration) 6. RANGE Key (User Measurement Unit) 7. ON/OFF 8. Stainless Steel Sample Well and Prism BOTTOM 9. Battery Cover 10. Battery Compartment 9

DISPLAY ELEMENTS 1. Battery (blinks when low battery condition detected) 2. Primary Display (displays measurement and error messages) 3. Measurement in Progress Tag 4. SETUP: Factory Calibration Tag 5. Calibration Tag 6. Measurement Unit 7. Automatic Temperature Compensation (blinks when temperature exceeds 0.0-40.0 C / 32.0-104.0 F range) 8. Temperature Units 9. Secondary Display (displays temperature measurements; when blinking, temperature has exceeded operation range: 0.0-80.0 C / 32.0-176.0 F) 10

Handle instrument carefully. Do not drop. Do not immerse instrument under water. Do not spray water to any part of instrument except the sample well located over the prism. The instrument is intended to measure propylene glycol solutions. Do not expose instrument or prism to solvents that will damage it. This includes most organic solvents and extremely hot or cold solutions. Particulate matter in a sample may scratch the prism. Absorb sample with a soft tissue and rinse sample well with deionized or distilled water between samples. Use plastic pipettes to transfer all solutions. Do not use metallic tools such as needles, spoons or tweezers as these will scratch the prism. Cover sample well with hand if measuring in direct sun. To reduce the effects of evaporation or absorption of water when taking readings over a period of time, the prism and sample well can be covered with plastic wrap. MEASUREMENT GUIDELINES Calibration should be performed daily, before measurements are made, when the battery has been replaced, between a long series of measurements, or if environmental changes have occurred since the last calibration. 1. Press the ON/OFF key, then release. Two instrument test screens will be displayed briefly; all LCD segments followed by the percentage of remaining battery life. The meter will then briefly display an indication of the measurement units set. When LCD displays dashes, the instrument is ready. CALIBRATION PROCEDURE 2. Using a plastic pipette, fill the sample well with distilled or deionized water. Make sure the prism is completely covered. 11

CALIBRATION PROCEDURE Note: If the ZERO sample is subject to intense light such as sunlight or another strong source, cover the sample well with your hand or other shade during the calibration. 3. Press the ZERO key. If no error messages appear, your unit is calibrated. (For a description of ERROR MESSAGES see page 17). Note: The 0.0 screen will remain until a sample is measured or the power is turned off. 4. Gently absorb the ZERO water standard with a soft tissue. Use care not to scratch the prism surface. Dry the surface completely. The instrument is ready for sample measurement. Note: If the instrument is turned off, the calibration will not be lost. 12

Verify the instrument has been calibrated before taking measurements. 1. Wipe off prism surface located at the bottom of the sample well. Make sure the prism and sample well are completely dry. 2. Using a plastic pipette, drip sample onto the prism surface. Fill the well completely. MEASUREMENT PROCEDURES Note: If the temperature of the sample differs significantly from the temperature of the instrument, wait approximately 1 minute to allow thermal equilibration. 3. Press the READ key. The results are displayed as % Volume or Freezing Point. Note: The last measurement value will be displayed until the next sample is measured or the instrument is turned off. Temperature will be continuously updated. The ATC tag blinks and automatic temperature compensation is disabled if the temperature exceeds the 0.0-40.0 C / 32.0-104.0 F range. 4. Remove sample from the sample well by absorbing with a soft tissue. 5. Using a plastic pipette, rinse prism and sample well with distilled or deionized water. Wipe dry. The instrument is ready for the next sample. 13

TO CHANGE MEASUREMENT UNIT Press the RANGE key to select the measurement units. The instrument toggles between the two measurement scales each time the key is pressed and the primary display indicates VoL % Volume or FrPt for Freezing Point. When the instrument displays the screen with 4 dashes, the instrument is ready for measurement. TO CHANGE TEMPERATURE UNIT To change the temperature measurement unit from Celsius to Fahrenheit (or vice versa), follow this procedure. 1. Press and hold the ON/OFF key continuously for approximately 8 seconds. The LCD will display the all segment screen followed by a screen with the model number on the primary display and the version number on the secondary display. Continue pressing the ON/OFF key. 8seconds 2. While continuing to hold the ON/OFF key, press the ZERO key. The temperature unit will change from C to F or vice versa. C or F Note: The selected unit will be used in Freezing Point mode. 14

To make a Propylene Glycol Solution, follow the procedure below: To make an X % propylene glycol solution volumetrically add X ml of high purity propylene glycol (CAS #: 57-55-6; MW 76.09) to a 100 ml Grade A volumetric flask. Use distilled or deionized water to bring the flask close to the total volume, mix, allow the solution to return to room temperature. Once the solution has returned to room temperature use distilled or deionized water to bring the total volume to 100 ml. Mix solution well before use. Propylene Glycol Total Volume Expected Freezing Point Value 10 % V 10.00 ml 100.00 ml -3.4 C (25.9 F) 40 % V 40.00 ml 100.00 ml -21.3 C (-6.3 F) MAKING A STANDARD SOLUTION 15

BATTERY REPLACEMENT To replace the instrument s battery, follow these steps: Make sure the instrument is off. Turn instrument upside down and remove the battery cover by turning it counterclockwise. Extract the battery from its location. Replace with a new 9V battery making sure to observe polarity. Insert the back battery cover and fasten it by turning clockwise to engage. 16

Error Code Err LO primary display HI primary display Description General failure. Cycle power to instrument. If error persists, contact your local Hanna Instruments Office. Sample exceeds minimum measurement range. Sample exceeds maximum measurement range. ERROR MESSAGES LO primary display Cal segment ON HI primary display Cal segment ON t LO primary display Cal segment ON t HI primary display Cal segment ON Air Wrong solution used to zero instrument. Use deionized or distilled water. Press ZERO. Wrong solution used to zero instrument. Use deionized or distilled water. Press ZERO. Temperature exceeds ATC low limit (10.0 C) during calibration. Temperature exceeds ATC high limit (40.0 C) during calibration. Prism surface insufficiently covered. ELt Too much external light for measurement. Cover sample well with hand. nlt Battery segment blinking LED light is not detected. Contact your local Hanna Instruments Office. <5% of battery life is remaining. Temperature values are blinking 0.0 C or 80.0 C ATC segment blinking SETUP segment blinking Outside temperature compensation range (10.0 to 40.0 C). Temperature measurement out of range (0.0 to 80.0 C). Factory calibration lost. Contact your local Hanna Instruments Office. 17

ACCESSORIES Code HI740157P HI740029P Description Plastic refilling pipette (20 pcs.) 9V battery (10 pcs.) 18 MAN96832 04/17

Warranty HI96832 is warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. This warranty is limited to repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service is required, contact your local Hanna Instruments Office. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packed for complete protection. To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase. Hanna Instruments reserves the right to modify the design, construction or appearance of its products without advance notice.

World Headquarters Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA www.hannainst.com Local Office Hanna Instruments USA 270 George Washington Highway Smithfield, RI 02917 Phone: 800.426.6287 Fax: 401.765.7575 e-mail: tech@hannainst.com MAN96832 Printed in ROMANIA