Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

Similar documents
Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Headspace Technology for GC and GC/MS: Features, benefits & applications

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Quantification of Rice Aroma, 2-Acetyl-1-Pyrroline (2-AP), Using TurboMatrix Headspace Trap Coupled with GC/NPD and GC/MS

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Benzoylecgonine in Urine by SAMHSA GC/MS

Application. Gas Chromatography February Introduction

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

HYPHENATED TECHNOLOGY GUIDE

HYPHENATED TECHNOLOGY GUIDE

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Study of Residual Solvents in Various Matrices by Static Headspace

Separation and Characterization of Indian and Australian Sandalwood Oils

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Analysis of Polyphenols in Saffron Petal Extracts with UHPLC/UV/MS

The determination of residual solvents in pharmaceuticals using the Agilent G1888 headspace/6890n GC/5975 inert MSD system Application Note

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline

Opiates in Urine by SAMHSA GC/MS

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

[ a ppl ic at ion no t e ]

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products.

Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP <467> method

Methanol Extraction of high level soil samples by USEPA Method 8260C

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

AUTOMATED HEADSPACE. Headspace Principles. Graham Broadway. Headspace Gas Chromatography Background to the HS Technique

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Whole Tablet Measurements Using the Frontier Tablet Autosampler System

A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS

A Capillary Gas Chromatographic Procedure for the Analysis of Nine Common Residual Solvents in Water-Insoluble Bulk Pharmaceuticals

High Purity Chromasolv Solvents

Chemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography.

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS

Application. Gas Chromatography March 1998

U.S. EPA Method 8270 for multicomponent analyte determination

Spring 2010 updated March 24 Determination of Aromatics in Gasoline by Gas Chromatography- Mass Spectrometry: Comparison of Grades and Brands

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis

ISSN: ; CODEN ECJHAO E-Journal of Chemistry , 7(2),

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Definitive EtG/EtS LC-MS/MS Analysis:

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput

Volatile Organic Compounds in Every Day Food

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.

High-Speed Gas and Headspace Analysis for the Process-Line and Laboratory: SIFT- MS IFPAC 2017

Title Experiment 7: Gas Chromatography and Mass Spectrometry: Fuel Analysis

Verification of Pharmaceutical Raw Materials Using the Spectrum Two N FT-NIR Spectrometer

GC Methods for Cannabis Safety and Potency Tes6ng

Product Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

METROLOGY IN CHEMISTRY MEETING. AZIZ MUKOTA Chemistry Laboratory Uganda National Bureau of Standards

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

STANDARD OPERATING PROCEDURES

Thermal Desorption Technical Support

GAS CHROMATOGRAPHY (GC)

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Determination of Carbonyl Compounds In Water by Dinitrophenylhydrazine Derivatization and HPLC/UV*

Pittcon P

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25)

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Analytical Trap Comparison for USEPA Method 8260C

Analysis of BTEX in Natural Water with SPME

Fast Gas Chromatography using Heated Headspace Gas Auto Sampling Techniques: Ethylene Oxide and 1,4-Dioxane in Fatty Acids

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Chemistry Gas Chromatography: Separation of Volatile Organics

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

We don t have anything against mass spectrometry. We just think it s time for a worthy alternative.

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

CHEM 429 / 529 Chemical Separation Techniques

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Transcription:

APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Lee Marotta Tom Kwoka David Scott Miles Snow Toby Astill PerkinElmer, Inc. Shelton, CT Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates Introduction Compared to the more traditional cannabis flower, cannabis concentrate products, such as extracts, tinctures, edibles, waxes, and oils are becoming the most commonly used cannabis products that are legally manufactured for both medicinal and recreational purposes. Most concentrates are extracted using a solvent such as supercritical CO 2, butane, propane, other hydrocarbons, water, or alcohol. These solvents are used to extract out the cannabinoids and terpenes from the plant material. In some cases, the solvent and impurities from the solvent remain in the extracted material. These are called residual solvents and are the byproducts of the extraction process. In some cases, these impurities can be toxic, which is why residual solvent analysis is a critical element of cannabis testing. The method of choice for measuring residual solvents is headspace (HS) gas chromatography (GC) coupled with mass spectrometry (MS) detection, so false positives are not reported.

The major benefit of this approach is that headspace is a fast, simple, accurate and precise technique that allows the components of interest (e.g. residual solvents and terpenes) to be introduced into the analytical system. The non-volatile matrix components remain in the sample vial and do not enter the GC, which results in a mostly maintenance free system, and faster analysis time. In addition, the technique is mature and already has been accepted for quantitation in several regulatory industries including pharmaceutical forensics, environmental and food, where its results have routinely stood up to scrutiny in a court of law. 1-4 It s also important to emphasize that because there are currently no federal regulations in the U.S., the allowable concentration limits for each residual solvent are defined by the individual state or country where the cannabis is grown. For example, Table 1 shows the list of proposed residual solvents and action levels for cannabis products in the State of California. 5 This study will focus on the analysis of residual solvents using pressure-balanced headspace (HS) sample introduction coupled with Gas Chromatography/Mass Spectrometry (GC/MS). In addition, it will discuss the objective of unambiguous separation of all compounds while maximizing sample throughput. Table 1. List of proposed residual solvents and action limits in cannabis products for the State of California. Compound CA Action Levels (ppm) Propane 1000 Butane 1000 Methanol 600 Ethylene Oxide 1 Pentane 1000 Ethanol 1000 Ethyl Ether 1000 Acetone 1000 Isopropyl Alcohol 1000 Acetonitrile 80 Methylene Chloride 1 Hexane 60 Ethyl Acetate 1000 Chloroform 1 Benzene 1 1,2-Dichloroethane 1 Heptane 1000 Trichloroethylene 1 Toluene 180 Xylenes total 430 Instrumentation The TurboMatrix HS sampler and a Clarus SQ 8 GC/MS (PerkinElmer Inc., Shelton, CT) was the system used for this analysis. The benefits of headspace sampling are well-recognized in the public domain, but it is essentially a separation technique in which volatile material such as residual solvents and terpenes, is extracted from a heavier sample matrix and injected directly into a GC for analysis. 6 The major reason for using MS detection is that many of the organic compounds associated with residual solvents elute at the same time (co-elute), so the unique mass spectrum of each compound means they can be optimally separated and detected without using additional detectors. Identifying and quantifying all the residual solvents using this solution, results in a faster analysis, enhanced productivity, quicker release of product, and maximized return on investment. Experimental Sample Preparation After Extraction Many states require taking multiple sampling points from non-homogenous samples (such as waxes and edibles) to ensure a representative sample for analysis. If this is the requirement, five sampling points from one sample are recommended. For example, if 500 mg is the regulatory requirement for testing, then five - 100 mg portions should be placed in a vial and brought to a final volume of 10 ml with Dimethylacetamide (DMA). Twenty µl of the diluent is then inserted into the HS vial, which is capped and placed onto the HS autosampler for analysis. However, if an average sampling is not required, a 40 mg aliquot of the extract can be directly weighed into the HS vial; capped; and placed onto the autosampler tray which is the preferred and easiest approach. A Turnkey Solution A fast, accurate, robust GC/MS-HS solution, and SOP, was developed to separate the required analyte compounds in each of the concentrates being tested using the mass and/or time domains for identifying the target compound and then quantifying the specific compounds using the following commercially-available standards (Emerald Scientific, San Luis Obispo, CA). California Residual Solvent Mix #1 (Inhalation) reference number STRS01102 California Residual Solvent Mix #2 (Inhalation) reference number STRS01103 2

Figure 1 demonstrates the chromatographic separation of all the residual solvents previously listed. To show the benefits of mass spectrometric detection, the chromatographic peaks of two pairs of compounds that co-elute (Pentane/Ethanol at about 3.0 minutes and Benzene/1,2 -Dichloroethane at about 4.8 minutes) exemplifies how they have been further separated by mass. Note, all compounds have been eluted and identified in about seven and a half minutes. A multi-level concentration suite of standards was prepared which represented the required ranges for quantitation of the sample, and met the required action levels. Repeatability was performed preparing eight vials with 20 µl of the same concentration standard. These data are shown in Table 2, which demonstrates the linearity and precision achieved using this method, together with the method reporting limits and California action levels. Figure 1. Chromatographic separation of all the compounds listed in Table 1. Table 2. Linearity and precision achieved using this method, together with the reporting and California action limits for cannabis products. Note: Reporting limits are based on a 1 to 20 sample dilution; therefore if no dilution is carried out, reporting limits are 20x lower. Compound Corelation Coefficient Precision (n=8) CA Action Levels (ppm) PerkinElmer Reporting Levels (ppm)* Propane 0.9996 2.30 1000 3.20 Butane 0.9991 1.08 1000 57.60 Methanol 0.9996 1.29 600 19.20 Ethylene Oxide 0.9994 2.29 1 1.00 Pentane 0.9997 1.95 1000 14.40 Ethanol 0.9997 1.41 1000 19.20 Ethyl Ether 0.9998 0.59 1000 9.60 Acetone 1.0000 0.94 1000 14.40 Isopropyl Alcohol 0.9996 1.33 1000 9.60 Acetonitrile 0.9998 0.45 80 1.16 Methylene Chloride 0.9999 1.08 1 1.00 Hexane 0.9996 1.08 60 0.48 Ethyl Acetate 0.9999 1.02 1000 6.80 Chloroform 0.9996 1.68 1 0.60 Benzene 1.0000 1.02 1 0.96 1,2-Dichloroethane 0.9993 2.15 1 0.96 Heptane 0.9997 1.08 1000 9.60 Trichloroethylene 0.9998 2.12 1 0.48 Toluene 0.9998 1.46 180 2.88 Xylenes total 0.9999 0.86 430 2.88 3

Discussion of Results As seen in Figure 1, the chromatographic peaks are well separated with a runtime of about seven and a half minutes and a sample to sample cycle time of less than 11 minutes. Using mass spectrometry, it allows for the identification of components without concern for false positives, while still maintaining extremely fast run times. The two pairs that co-elute in time, ethanol/pentane and benzene/1,2- dichloroethane have very unique spectra and quantitation ions as seen in greater detail in Figure 3. Since quantitation is performed on the mass chromatogram (also referred to as the quantitation ion) of the unique mass, it offers the advantage of interference free integration and quantitation, which would not be possible if the analysis was carried out by flame ionization detection. Figure 2 displays an example of the calibration curve for the target compound benzene. A table is inserted in this graphic documenting the quantitation of each point using this curve. The % deviation calculated for each point clearly shows excellent correlation with the calibration standards. Std Conc Calculated Conc % Dev 0.048 0.049 7 0.12 0.11-6 0.24 0.24-1 0.72 0.73 1 1.8 1.8-1 3.0 3.0 0 Figure 2. Calibration curve for benzene, showing the % deviation for each standard. 4 Figure 3. The mass chromatogram of two pairs of compounds that co-elute in time, ethanol/pentane and benzene/1,2- dichloroethane, show they have very unique spectra and quantitation ions.

As demonstrated in Table 2, linearity across the compound range is excellent using a multi-level calibration with all targets having a correlation coefficient value greater than 0.9993. The turnkey solution is also precise with the relative standard deviation of less than 2.3 % for all compounds. It s also important to emphasize that the reporting limits are all based upon a 1:20 sample dilution factor; therefore the reporting limit could be improved even further if smaller dilutions or no dilutions were used. Conclusion This study has clearly shown that headspace technology coupled with GC/MS is the perfect solution for identifying and measuring residual solvents in cannabis concentrates and its many products. The benefit of headspace sample introduction over other approaches is that it requires minimum sample preparation, with very little interaction required by the operator. When headspace sampling is combined with GC/MS, it allows for rapid, unambiguous and interference free integrations, with very little likelihood of false positives. In addition, MS provides the ability to identify unknown components that may be present in the sample that are not target compounds. This capability offers significant benefits over a single detector such as flame ionization detection (FID) where a non-targeted compound eluting at the same time as a targeted compound would produce a result which was over the action limit, resulting in a failed batch and a cannabis product not viable for market. It s also important to emphasize that this technique is fast and capable of quantifying residual solvents in all concentrate samples and other required matrices. Combined with essentially maintenance-free operation, a GC/MS-HS method will enhance productivity and strengthen the lab s business operations. References 1. Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology, David Scott, PerkinElmer Inc. Application Note, https://www.perkinelmer.com/lab-solutions/resources/docs/ APP_Residual_Solvents_In_Pharmaceuticals_by_ USP_467_013617_01.pdf. 2. Increasing Accuracy of Blood-Alcohol Analysis Using Automated Headspace-Gas Chromatography, John Musselman, Anil Solanky, William Arnold, PerkinElmer Inc. 3. Measuring Environmental Volatile Organic Compounds Using US EPA Method 8260B with headspace trap GC/MS, Heidi Griffith, PerkinElmer Inc. Application Note, http://www.perkinelmer. com/lab-solutions/resources/docs/app_ GasChromaUSEPA8260B.pdf. 4. Monitoring Volatile Organic Compounds in Beer Production Using the Clarus SQ 8 GC/MS and TurboMatrix Headspace Trap Systems, Lee Marotta, Andrew Tippler, PerkinElmer Application Note, http://www.perkinelmer.com/pdfs/ Downloads/App_FoodBeerVolatileCompounds.pdf. 5. California Bureau of Cannabis Control Proposed Text of Regulations, July, 2018; https://cannabis.ca.gov/wp-content/ uploads/sites/13/2018/07/bureau-of-cannabis-control- Proposed-Text-of-Regulations.pdf. 6. An Introduction to Headspace Sampling Gas Chromatography Fundamentals and Theory: Andrew Tippler, PerkinElmer Inc. Application Note, http://www.perkinelmer.com/lab-solutions/ resources/docs/gde_intro_to_headspace.pdf. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/contactus Copyright 2018, PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 014321_01 PKI