FUTURE CONFIRMATORY CRITERIA

Similar documents
WADA Technical Document TD2003IDCR

WADA Technical Document TD2015IDCR

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS

Proposal: criteria for positive identification in LC-MS analysis

Multi-residue analysis of pesticides by GC-HRMS

Introduction to GC/MS

The Emergence of Benchtop Tandem Quadrupole Mass Spectrometers as a Routine Tool in Quantitative GC/MS/MS Applications

Analytical Technologies and Compound Identification. Daniel L. Norwood, MSPH, PhD SCĪO Analytical Consulting, LLC.

MassHunter TOF/QTOF Users Meeting

Analyst Software. Peptide and Protein Quantitation Tutorial

LC-MS Based Metabolomics

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

Theory and Practical Consequences.

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Making Sense of Differences in LCMS Data: Integrated Tools

TargetScreener. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS

Agilent TOF Screening & Impurity Profiling Julie Cichelli, PhD LC/MS Small Molecule Workshop Dec 6, 2012

Quantitative Screening of 46 Illicit Drugs in Urine using Exactive Ultrahigh Resolution and Accurate Mass system

4-Methyldiethcathinone (4-MDEC)

Introduction. Chapter 1. Learning Objectives

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

5Cl-AKB48 and 5Br-AKB48

2-[[2-(4-ethyl-2,5-dimethoxy-phenyl)ethylamino]methyl]phenol

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

TANDEM MASS SPECTROSCOPY

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Profiling of Diferulates (Plant Cell Wall Cross- Linkers) Using Ultrahigh-performance Liquid. Chromatography-Tandem Mass Spectrometry

MassHunter Software Overview

LC-HRMS: Challenges for Routine Implementation

Choosing the metabolomics platform

Analyzing Compounds of Environmental Interest Using an LC/Q-TOF Part 1: Dyes and Pigments. Application. Introduction. Authors. Abstract.

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

N-(1-carbamoyl-2-methyl-propyl)-1-(5-chloropentyl)indazole-3- carboxamide

- Supporting Information -

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Methyl 1-[(4-fluorophenyl)methyl]indazole-3-carboxylate

Confirmation of In Vitro Nefazodone Metabolites using the Superior Fragmentation of the QTRAP 5500 LC/MS/MS System

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

THE IMPLEMENTATION OF A SCREENING WORKFLOW FOR ION MOBILITY QUADRUOPOLE TIME-OF-FLIGHT MASS SPECTROMETRIC ANALYSIS OF PFOS ISOMERS

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions

Atomic masses. Atomic masses of elements. Atomic masses of isotopes. Nominal and exact atomic masses. Example: CO, N 2 ja C 2 H 4

Agilent MassHunter Profinder: Solving the Challenge of Isotopologue Extraction for Qualitative Flux Analysis

Choosing Fit-For-Purpose Food Safety Methods

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

Highly Sensitive and Rugged GC/MS/MS Tool

Application Note. Edgar Naegele. Abstract

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

De Novo Metabolite Chemical Structure Determination. Paul R. West Ph.D. Stemina Biomarker Discovery, Inc.

Chemistry Instrumental Analysis Lecture 37. Chem 4631

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. MAGLAB Summer School Ryan P. Rodgers

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

1-(1,3-benzodioxol-5-yl)-2-(ethylamino)hexan-1-one

Application Note LCMS-116 What are we eating? MetaboScape Software; Enabling the De-replication and Identification of Unknowns in Food Metabolomics

[ instrument specifications ]

INTRODUCTION OVERVIEW

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

4-methyl-1-phenyl-2-pyrrolidin-1-yl-pentan-1-one

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

The Use of Tandem Quadrupole GC/MS/MS for the Determination of Polycyclic Aromatics Hydrocarbons (PAHs) in Food Products

Features or compounds? A data reduction strategy for untargeted metabolomics to generate meaningful data

Identifying Disinfection Byproducts in Treated Water

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Methods for proteome analysis of obesity (Adipose tissue)

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Simplified Workflow and Improved Confidence in Screening Methods on Time of Flight Mass Spectrometry

Cerno Application Note Extending the Limits of Mass Spectrometry

Designed for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS

Implementing GC-HRAM MS for More Efficient and Effective Routine Pesticide Residues Analysis

An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application

materials analysis Solutions for Your Analytical Business Markets and Applications Programs

Signal, Noise, and Detection Limits in Mass Spectrometry

Types of Analyzers: Quadrupole: mass filter -part1

Methyl 2-[[1-(5-fluoropentyl)indole-3-carbonyl]amino]-3,3- dimethyl-butanoate

MASS SPECTROMETRY. Topics

UNIFI - Targeted and Untargeted Screening MS Software

Fereshteh Zandkarimi, Samanthi Wickramasekara, Jeff Morre, Jan F. Stevens and Claudia S. Maier

Automated and accurate component detection using reference mass spectra

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Agilent ESI and APCI sources: for polar to non-polar compounds

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Hplc Lc Ms And Gc Method Development And Validation Guideline For Academic And Industrial Scientists Involved In Method Development And Validation

Hplc Lc Ms And Gc Method Development And Validation Guideline For Academic And Industrial Scientists Involved In Method Development And Validation

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS

Transcription:

FUTURE CONFIRMATORY CRITERIA Revision CD 2002/657 Marco Blokland, Bjorn Berendsen, Hans Mol, Leen van Ginkel, Saskia Sterk

REVISION 2

Revision 2002/657 EU Commission has explicitly asked for a revision of 2002/657 Status of new document: not yet decided on EURLs in the area are tasked with making a proposal State of the art criteria for confirmatory analysis Should stand for a long time 3

EXAMPLE: Ambient Ionization-MS Ambient ionization methods permit direct ionization of samples outside the mass spectrometer in the open atmosphere There are >70 different types of AIMS described Minimal sample preparation No chromatography 4

What would this mean for new criteria? Since no chromatography is involved it will break the retention time criteria Without the use of retention time the number of false positives will increase?? Solution 1: The most selective MS measurement should be used in confirmatory analysis using AIMS Solution 2: Inclusion of another selective orthogonal technique 5

Ion Ratio 6

LC-MS/MS relative ion intensity In 2002/657 MS/MS criteria are based on expert opinions, at that time no scientific data available Experimental data is now available Relative intensities Allowed Dev. [rel] >50 % ± 20 % >20 % to 50 % ± 25 % >10 % to 20 % ± 30 % <10 % ± 50% Relative Intensity Berendsen et al: Drug Testing and Analysis, 2016, 8,477-90 7

LC-MS/MS relative ion intensity The data shows deviations depend on the intensity of the less abundant ion From: Berendsen et al: Drug Testing and Analysis 2016, 8, 477-490 8

LC-MS/MS relative ion intensity The S/N of the lowest product ion is very important in obtaining stable ion ratio s Maximum allowed deviation a compromise between the false positive and false negative rate New guideline proposed ± 40 % (relative) for all relative ion intensity In case the precursor selection has a mass selection window of more than one Dalton (e.g. in case of Data Independent Acquisition) the technique is considered as full-scan confirmatory analysis. 9

Retention Time 10

Revision: Retention times Almost all current guidelines have a relative requirement (e.g. 2.5%) for the maximum deviation of the Rt between sample and standard Using gradient elution, empirical studies (Mol et al. Berendsen et al.) show deviation is absolute over the whole retention time range 11

Revision: Retention times From: Mol et al: Analytica Chimica Acta, 2015, 873, 1-13 From: Berendsen et al: Drug Testing and Analysis 2016, 8, 477-490 12

Fast chromatography: one-minute gradiënt 1 minute Oven: 60 C Typical run times 30-60 seconds on 2-5 mm ID column Flow 1-1.5 ml/min 13

Revision: Retention times All mass spectrometric analyses are combined with a sufficiently orthogonal separation (e.g. chromatography) or any separation technique that shows enough separation power for the specific application Possible to use of an absolute or relative criterion ± 0.1 minute for LC and GC In case retention time is below 1 minute then 5%, allowing the use of fast chromatography At least twice the retention time corresponding to the void volume of the column If an internal standard is used the relative retention time of the analyte, shall correspond to that of the calibration standard at a tolerance of ± 0.5 % for GC and ± 1 % for LC 14

Separation in the first dimension 1 n c 15

Coupling of second separation dimension 1 n c 16

Fully coupling of two separation techniques When truly orthogonal n c,2d = 1 n c x 2 n c 1 n c 17

Example added value of LC as orthogonal technique Kaabia, Z., Laparre, J., Cesbron, N., Le Bizec, B., & Dervilly-Pinel, G. (2018). Comprehensive steroid profiling by liquid chromatography coupled to high resolution mass spectrometry. The Journal of Steroid Biochemistry and Molecular Biology, 183, 106-115. 18

Ion Mobility 19

Ion Mobility Mass spectrometry Ions separated by mass-to-charge ratio Ion mobility Ions separated by size, shape, and charge Size and charge: ions driven through gas filled chamber using electric field Shape: compact ions travel faster, less hindered by drift gas 20

Ion Mobility 2+ 1+ 2+ 1+ Drift times related to collisional cross sections (CCS) 21 Drift time

Collisional Cross Sections (CCS) CCS = Effective area of interaction between individual ion and neutral gas molecule Related to chemical structure and dimensions Can be obtained from experimental ion mobility data, and by molecular modelling Not all types of ion mobility yield CCS data Different protonation sites can yield different CCS values Boschmans et al., Analyst, 2016, 141, 4044-4054 22

Drift tube ion mobility (DT-IMS) Types of ion mobility High field asymmetric IMS (FAIMS) Differential Mobility Spectrometry (DIMS) Traveling wave ion mobility (TWIM) Trapped ion mobility (TIMS) López et al., New J. Chem., 2013, 37, 1283-1289 Silveira et al., Analytical Chemistry, 2014, 86, 5624-5627 23

Drift tube ion mobility (DT-IMS) Types of ion mobility Agilent 6560 High field asymmetric IMS (FAIMS) Differential Mobility Spectrometry (DIMS) AB Sciex SelexIon Traveling wave ion mobility (TWIM) Waters Synapt Trapped ion mobility (TIMS) Bruker timstof López et al., New J. Chem., 2013, 37, 1283-1289 Silveira et al., Analytical Chemistry, 2014, 86, 5624-5627 24

Separation of isomers: Diethylstilbestrol cis-des DTIMS trans-des Observed CCS difference ~ 5% TWIMS 25

Experimental determination of CCS value CCS can be calculated from drift time using Mason-Schamp equation: Ω / z = charge state of the analyteion e= charge of an electron N= number density of the drift gas µ= reduced mass of the ion-neutral pair k B = Boltzmann constant T = gas temperature K 0 = reduced mobility; mobility measured at standard temperature and pressure Only valid for drift tube instruments TWIMS and TIMS instruments need to be calibrated CCS calculation is not possible from DIMS data 26

CCS repeatability using TWIMS CCS of eight compounds determined on 7 different days using TWIMS instrument Compound Detected ion m/z Average Experimental CCS (Å 2, n=7) Cimaterol [M+H] + 220.1450 154.7 0.4% Clenbuterol [M+H] + 277.0874 162.8 0.6% Isoxsuprine [M+H] + 302.1756 174.6 0.6% THC [M+H] + 315.2324 187.7 0.4% RR-p-chloramphenicol [M-H] - 321.0045 160.7 1.1% THC-COOH [M-H] - 343.1909 195.4 0.6% Salmeterol [M+H] + 416.2801 210.0 0.3% Okadaicacid [M-H] - 803.4582 310.8 0.9% Repeatability is < 1% for all compounds, with the exception of RR-p-CAP (1.1%) RSD No significant effect of matrix observed Literature: <2%, inter-laboratory <5% Article in preparation 27

CCS as extra identification point? + Highly reproducible, no effect of matrix + Possibility to compare with literature or database + Reduction of matrix - Large deviation between DTIMS and TWIMS data found for some compounds -High correlation between CCS and m/z, which reduces uniqueness of CCS value Drift time (bins) 100 90 β-agonists 80 70 60 R² = 0.9678 50 40 30 20 200 250 300 350 400 450 m/z 28

Orthogonality of ion mobility Hernández-Mesa, M., Le Bizec, B., Monteau, F., García- Campaña, A. M., & Dervilly- Pinel, G. (2018). Collision Cross Section (CCS) database: An additional measure to characterize steroids. Analytical chemistry, 90(7), 4616-4625. 2% error bars Only [M+H] + 29

Evaluation future Ion Mobility 1 cycle DTX-2 Okadaic acid 2 cycle 3 cycle 4 cycle 5 cycles 30

Serpentine UltralongPath with Extended Routing (SUPER) High Resolution Traveling Wave Ion Mobility-MS using Structures for Lossless Ion Manipulations (SLIM) 1 pass 13.5 m R=47 300 ms peak capacity >200 40 pass R=340 31

Examples of SLIM 32

Ion-mobility MS Cross-section additional confidence in confirmatory analysis Future alternative for chromatography? Criteria proposed: Cross-section is within 2 percent of a reference standard 0.5 identification point Not (yet) in compliance with orthogonality criteria 33

High Resolution Mass Spectrometry 34

Definition full scan hrms Current Definition of hrms: 657/2002 Resolution > 10,000 10% valley FDA FVM guidance Resolution > 10,000 FWHM SANTE/11945 not defined 35

full scan hrms Define resolving power, mass accuracy or both? Resolving power: Sante/11945: use enough resolving power to achieve a mass accuracy <5 ppm Mass accuracy: 657/2002: no mass accuracy requirement SANTE/11945: 5 ppm (< m/z 200 1 mda) FDA FVM: MS: 5 ppm MS/MS: 10 ppm 36

Proposal: full scan hrms 2002/657 revision In high-resolution mass spectrometry (HRMS), the resolution shall typically be greater than 10,000 for the entire mass range at 10 % valley or 20,000 at FHWM. Only diagnostic ions with a relative intensity of more than 10 % in the reference spectrum of the calibration standard or MMS are suitable For confirmation of identity the mass deviation of all diagnostic ions should be below 5 ppm (or in case M/z <200 below 1mDa). Adducts and isotopes of selected diagnostic ions are excluded. In full scan mode, the isotopic pattern should match the molecular formula. 37

Suggestions for identification points Technique IPs earned Chromatography (GC or LC) 1 Ion mobility 0.5 LR-MS ion 1 Precursor ion selection at <±0.5 Da mass range 1 (indirect) LR-MS n product ion 1.5 HR-MS ion 1.5 HR-MS n product ion 2.5 Non MS techniques 1 38

Examples Technique(s) Separation Number of ions IP GC-MS (EI or CI) GC n GC-MS (EI and CI) GC 2 (EI) + 2 (CI) GC-MS (EI or CI) 2 derivates GC 2 (Derivate A) + 2 (Derivate B) LC-MS LC n (MS) GC-or LC-MS/MS GC or LC 1 precursor + 2 products GC-or LC-MS/MS GC or LC 2 precursor + 2 products GC-or LC-MS 3 GC or LC 1 precursor + 1 MS 2 product + 1 MS 3 product GC-or LC-HRMS GC or LC n IM-HRMS IM 3 GC-or LC-HRMS/MS GC or LC 1 precursor (<±0.5 Da mass range) + 2 products GC-or LC-HRMS and HRMS/MS GC or LC 1 full scan ion + 1 HRMS product ion a GC-and LC-MS GC and LC 2 ions (GCMS) + 1 ion (LCMS) 39

Examples Technique(s) Separation Number of ions IP GC-MS (EI or CI) GC n 1+n GC-MS (EI and CI) GC 2 (EI) + 2 (CI) 1+4 GC-MS (EI or CI) 2 derivates GC 2 (Derivate A) + 2 (Derivate B) 1+4 LC-MS LC n (MS) 1+n GC-or LC-MS/MS GC or LC 1 precursor + 2 products 1 + 1 + 2*1.5 = 5 GC-or LC-MS/MS GC or LC 2 precursor + 2 products 1 + 2 + 2*1.5 = 6 GC-or LC-MS 3 GC or LC 1 precursor + 1 MS 2 product + 1 MS 3 product 1 + 1 + 1.5 + 1.5 = 5 GC-or LC-HRMS GC or LC n 1 + n*1.5 IM-HRMS IM 3 0.5 + 3*1.5 = 5 GC-or LC-HRMS/MS GC or LC 1 precursor (<±0.5 Da mass range) + 2 products 1 + 1 + 2*2.5 = 7 GC-or LC-HRMS and HRMS/MS GC or LC 1 full scan ion + 1 HRMS product ion a 1 + 1.5 + 2.5 = 5 GC-and LC-MS GC and LC 2 ions (GCMS) + 1 ion (LCMS) 1 + 1 + 2 + 1 = 5 40