Uncontrolled Copy. Table of Contents. 1. Principle of Assay Specimens Reagents and Materials... 3

Uncontrolled Copy. Table of Contents. 1. Introduction and Principle of Assay Specimens Reagents and Materials...

Uncontrolled Copy. Table of Contents. 1. Principle of Assay Specimens Reagents and Materials... 4

Uncontrolled Copy. Table of Contents. 1. Principle of Assay Specimens Reagents and Materials... 4

Uncontrolled Copy. Table of Contents. 1. Principle of Assay Specimens Reagents and Materials... 4

Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Modified Adams Assay for Phenolics in Wine

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Scope: Materials: Labeled 1.5 ml or 2 ml Eppendorf tubes Appropriate pipette tips (see Table 1) Labeled LC vials

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

Determination of Cannabinoids (THC) in Biological Samples

Prepared By: R. W. Waggoner, Jr. Effective Date: August 18, 2008

METHADONE and EDDP in urine by GC-MS Code GC Method of Confirmation by GC-MS

NORBUPRENORPHINE (Buprenorphine s Metabolite ) BUPRENORPHINE in urine by GC/MS Code GC Method of Confirmation by GC-MS

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

Catecholamines Urine LC-MS/MS Analysis Kit User Manual

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Toxicology Extraction Procedure for Base Drugs Using United Chemical Technologies Clean Screen Extraction Columns

Sulfotepp impurities in Chlorpyrifos EC formulations

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

Method - Determination of aromatic amines in mainstream cigarette smoke

Limit-test of NDMA and NDEA in Sartans. by GC-MS (Liquid-direct-injection)

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C

Uncontrolled Copy. SOP-060 Chromatography, Tune Evaluation and Troubleshooting. Table of Contents. 1. Principle... 3

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

Benzoylecgonine in Urine by SAMHSA GC/MS

637. Thiamethoxam. HPLC method

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

The analysis of organic acid content of additives, premix, feed, and water.

Method for estimation of iodine in urine

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

FORMIC ACID in URINE by GC/MS-headspace - Code GC74010

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

The Nitrofurantoin Capsules Revision Bulletin supersedes the currently official monograph.

Micro Volume QuEChERS kit

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD

Psychoactive Drugs Urine LC-MS/MS Analysis Kit User Manual

STANDARD OPERATING PROCEDURES

ANALYTICAL METHOD PROCEDURES

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF

INTERNATIONAL CHEMISTRY TESTING. 258 Main Street- Suite 311- Milford, MA Tel: Fax:

Opiates in Urine by SAMHSA GC/MS

STANDARD OPERATING PROCEDURE (SOP) FOR ESTIMATION OF DIETARY INTAKE OF SODIUM (DIS) IN 24 HOUR URINE USING ELECTROLYTE ANALYZER

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

[Unless otherwise stated all tolerances are ±5%] SAMPLES

Chemistry Gas Chromatography: Separation of Volatile Organics

Spectrophotometric Determination of Ferrocyanide in Effluents

METHOD 3665 SULFURIC ACID/PERMANGANATE CLEANUP

Standard Operating Procedure. Immunology

ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 504.1/8011

TECHNICAL BRODIFACOUM

ENVIRONMENTAL analysis

4023 Synthesis of cyclopentanone-2-carboxylic acid ethyl ester from adipic acid diethyl ester

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS

Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats

Date 10/22/97 Page 1 Revision 4.1 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8000/8100 (MODIFIED) DIESEL RANGE ORGANICS (DRO)

Functional Genomics Research Stream. Lecture: February 17, 2009 Masses, Volumes, Solutions & Dilutions

BUTYL ALCOHOL in urine by GC-FID Code GC05510

HIPPURIC ACID, o-m-p-methylhippuric ACIDS and CREATININE IN URINE by UV CODE Z07510

Mercury, total-in-sediment, atomic absorption spectrophotometry, nameless, direct

STANDARD OPERATING PROCEDURES SOP: 1824 PAGE: 1 of 22 REV: 0.0 DATE: 04/21/95 ANALYSIS OF METHYL PARATHION IN SOIL SAMPLES BY GC/MS

Colorado Department of Public Health and Environment. Laboratory Services Division Toxicology Laboratory

camp Direct Immunoassay Kit

Method of confirmation by GC-MS

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

[Unless otherwise stated all tolerances are ±5%] SAMPLES

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

PYRIPROXYFEN TECHNICAL

Technical Procedure for ELISA Drug Screen Version 3 Toxicology Unit Effective Date: 02/15/2013. ELISA Drug Screen

METHOD 3510B SEPARATORY FUNNEL LIQUID-LIQUID EXTRACTION

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

BRIEFING. Pharmacopeial Discussion Group Sign Off Document Attributes EP JP USP Definition Loss on drying Readily carbonizable substances

Standard Operating Procedure

FORMALDEHYDE IN URINE by UV Code Z65010

Quantitative determination of cannabinoids in Hemp and Hemp products by High Performance Liquid Chromatography using a UV Detector

STANDARD OPERATING PROCEDURES

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

Density of Aqueous Sodium Chloride Solutions

Detection limit: grain, feed 500 ppb; milk 50 ppb; cream, cheese 5 ppb

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit

ab83360 Ammonia Assay Kit

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

CHEMISTRY DEPARTMENT

Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

DTNB-Thiols Assay Kit

Determination of Orthophosphate Ion

Transcription:

Table of Contents 1. Principle of Assay... 3 2. Specimens... 3 3. Reagents and Materials (HPLC grade)... 3 4. Standards, Controls, and Solutions... 4 5. Equipment and Special Supplies... 5 6. Instrumentation and Parameters... 6 7. Target Ions... 7 8. Procedure... 7 9. Calculations... 8 10. Quality Control... 9 11. Validation of Method... 10 12. Reporting... 11 13. Load Assignment Packet Preparation... 12 14. References... 13 1 of 13

SOP Name: Ethylene Glycol Quantitation by North Carolina Office of the Chief Medical Examiner Toxicology Laboratory Revision: 10.1.2 Updated RT acceptance range 10.1.7 & 10.1.8 Add calibrator acceptance criteria 11.1 Added validation table 4 Added standards prep instructions 6 Updated Instrumentation and instrument parameters SOP #: 209 Revision Date/Initials: MSF - 5/11/2015 MSF 04/13/17 MSF 5/25/2017 MSF 06/05/2017 Approving Authority Name Approving Authority Signature Approval Date Ruth E. Winecker, Ph.D. 04/08/2015 Ruth E. Winecker, Ph.D. 08/25/2016 Ruth E. Winecker, Ph.D. 12/07/2017 2 of 13

1. Principle of Assay 1.1. Ethylene glycol is a common component of commercial anti-freeze preparations, automotive cooling systems, and hydraulic brake fluids. This, coupled with its sweet taste, results in its relatively common appearance in emergency departments and medical examiner laboratories. 1.2. From an aliquot of biological sample, ethylene glycol is derivatized with phenylboronic acid to form a boronate ester. Following a basic wash, the sample is made acidic with concentrated HCl and extracted with methylene chloride. The extract is injected onto a GC/MS for quantitation. 2. Specimens 2.1. Approximately 0.1mL of blood, serum, vitreous, urine, or 0.2g tissue homogenate. 3. Reagents and Materials (HPLC grade) 3.1. Ethylene Glycol 3.2. Ethylene Glycol-d6 3.3. Acetone 3.4. Deionized water 3.5. Methylene Chloride 3.6. Phenylboronic acid 3.7. Potassium hydroxide 3.8. Concentrated hydrochloric acid 3.9. Sodium Sulfate, anhydrous 3.10. 1% Phenylboronic acid solution 3.11. 0.1% Phenylboronic acid solution 3.12. 0.1N potassium hydroxide 3 of 13

4. Standards, Controls, and Solutions 4.1. Prepare Calibration, IS, and Control standards, as needed, according to SOP 10. 4.2. Ethylene Glycol-d6 Stock Solution (5000 mg/dl) 4.2.1. Place a 10mL volumetric flask containing 1-2mL of DiH20 onto a rough balance and tare. 4.2.2. Using a pasture pipette, transfer 0.5g Ethylene Glycol-d6, dropwise, into the flask. 4.2.3. Fill to the line with DiH20, insert stopper and invert three times to mix. 4.2.4. Transfer to properly labeled 16x100mm screw topped test tubes and cap. 4.3. Ethylene Glycol-d6 Internal Standard (100 mg/dl) 4.3.1. Into a 10mL volumetric flask, add 0.2mL of Ethylene Glycol-d6 Stock Solution (5000 mg/dl) with a micropipette. 4.3.2. Fill to the line with DiH20, insert stopper and invert three times to mix. Transfer to properly labeled 16x100mm screw topped test tubes and cap. 4.4. Ethylene Glycol Stock Calibration Standard (10000mg/dL) 4.4.1. Place a 10mL volumetric flask containing 1-2mL of DiH20 onto a rough balance and tare. 4.4.2. Using a pasture pipette, transfer 1g Ethylene Glycol, dropwise, into the flask. 4.4.3. Fill to the line with DiH20, insert stopper and invert three times to mix. 4.4.4. Transfer to properly labeled 16x100mm screw topped test tubes and cap. 4.5. Ethylene Glycol Calibration Standard (Cal 2-6 - 100mg/dL) 4.5.1. Into a 10mL volumetric flask, add 0.1 ml of Ethylene Glycol Stock Calibration Standard (10000mg/dL) with a class: A volumetric pipette. 4.5.2. Fill to the line with DiH20, insert stopper and invert three times to mix. Transfer to properly labeled 16x100mm screw topped test tubes and cap. 4.6. Ethylene Glycol Calibration Standard (Cal 1-10mg/dL) 4.6.1. Into a 10mL volumetric flask, add 1 ml of Ethylene Glycol Calibration Standard (Cal 2-6 - 100mg/dL) with a class: A volumetric pipette. 4.6.2. Fill to the line with DiH20, insert stopper and invert three times to mix. Transfer to properly labeled 16x100mm screw topped test tubes and cap. 4 of 13

4.7. Ethylene Glycol Stock Control Standard (10000mg/dL) 4.7.1. Place a 10mL volumetric flask containing 1-2mL of DiH20 onto a rough balance and tare. 4.7.2. Using a pasture pipette, transfer 1g Ethylene Glycol, dropwise, into the flask. 4.7.3. Fill to the line with DiH20, insert stopper and invert three times to mix. 4.7.4. Transfer to properly labeled 16x100mm screw topped test tubes and cap. 4.8. Ethylene Glycol Control Standard (100mg/dL) 4.8.1. Into a 10mL volumetric flask, add 0.1 ml of Ethylene Glycol Stock Control Standard (10000mg/dL) with a class: A volumetric pipette. 4.8.2. Fill to the line with DiH20, insert stopper and invert three times to mix. Transfer to properly labeled 16x100mm screw topped test tubes and cap. 4.9. 1% Phenylboronic acid solution 4.9.1. Weigh 1 g phenylboronic acid, transfer to 100 ml volumetric flask, and qs to volume with acetone. Store at 2-8ºC for up to 1 year. 4.10. 0.1% Phenylboronic acid solution 4.10.1. Pipette 10 ml of 1% Phenylboronic acid solution into a 100 ml volumetric flask and qs to volume with acetone. Make fresh daily. 4.11. 0.1N potassium hydroxide 4.11.1. Weigh 2.8g potassium hydroxide, transfer to 500 ml volumetric flask and qs to volume with deionized H2O. Store at room temperature for up to two years. 5. Equipment and Special Supplies 5.1. Test tubes, 16x100 borosilicate glass 5.2. Micropipette (10-100µL) 5.3. Pipette tips (10-100µL) 5.4. Vortex mixer 5.5. Centrifuge (2000-3000 rpm) 5 of 13

5.6. GC autosampler vials 5.7. Nitrogen Evaporator 5.8. Pasteur Pipettes 6. Instrumentation and Parameters 6.1. Agilent GC/MS, Chemstation software, compatible computer and printer. 6.2. GC/MS parameters (The instrument and conditions may be changed to permit improved performance). 6.3. Instrument method (GCMS13, GCMS15): 13_EG_QUANT, 15_EG_QUANT click here for instrument parameters. 6.4. Acquisition mode: SIM 6.5. Column: DB5-MS 6.6. MS transfer line temperature: 280º C 6.7. Oven ramp program: 6.7.1. Equilibration time: 30 sec. 6.7.2. Initial temp: 50º C 6.7.3. Initial temp hold: 1.00 min. 6.7.4. Ramp 1: 35º C/min. 6.7.5. Ramp 1 final temp: 310º C 6.7.6. Ramp 1 hold: 0.5 min. 6.7.7. Total Run Time: 8.93 min. 6.8. Inlet Parameters 6.8.1. Mode: SPLIT 6.8.2. Split ratio: 50:1 6 of 13

6.8.3. Split vent: 677mL/min 6.8.4. Temperature: 275º C 6.8.5. Injection Volume: 1µL 6.8.6. Purge time: Constant 6.8.7. Column Flow: 1.4mL/min (Constant Flow) 7. Target Ions 7.1. Major Ions for Ethylene glycol phenyl borate ester: 91, 148, 118 7.2. Major Ions for Ethylene Glycol-d6 phenyl borate ester: 93, 152 8. Procedure 8.1. Prepare 0.1% phenylboronic acid solution. (See Section 4. Standards, Controls, and Solutions) 8.2. Add 100µL Ethylene Glycol-d6 IS to the appropriate number of 16x125mm test tubes. 8.3. Prepare calibrators/controls according to Standard & Control Worksheet. 8.4. Add 0.1mL blank blood to all calibrators and controls. 8.5. Aliquot case samples in duplicate at 0.1mL and 0.05mL. 8.6. Add 0.5mL DiH2O to each tube and vortex. 8.7. Add 3.0mL 0.1% phenylboronic acid solution to each tube and vortex for 30 seconds. 8.8. Centrifuge at 2000 x g for 10 minutes. 8.9. Transfer supernatants to clean 16x125mm test tubes. Discard bottom layer (blood pellet). 8.10. Evaporate supernatants to approximately 0.5mL in Turbo-Vap at 60 o C for 6-8 min. (The purpose is to remove acetone and leave aqueous portion). 8.11. Remove tubes from Turbo-Vap and add 1mL of 0.1N KOH to each tube and vortex for 10 seconds. 8.12. Add 2mL methylene chloride to each tube and vortex for 30 seconds. 7 of 13

8.13. Centrifuge at 2000 x g for 10 minutes. 8.14. Transfer upper (aqueous) layer to labeled 16x125mm tubes. Discard bottom organic layer into halogenated waste bottle. 8.14.1. Note: It is OK to transfer a small amount of methylene chloride in this step. 8.15. Acidify samples with concentrated HCl (1 drop per tube) and vortex. 8.16. Add 2mL methylene chloride to each tube and vortex for 30 seconds. 8.17. Centrifuge at 2000 x g for 10 minutes. 8.18. Aspirate upper aqueous layer to waste. 8.19. Dry bottom organic layer by adding ~100mg (2 small spatulas) of anhydrous sodium sulfate to each sample and vortex for 15 seconds. 8.20. Centrifuge at 2000 x g for 5 minutes. 8.21. Transfer ~60µL of remaining organic layer to GC autosampler vials for analysis. 9. Calculations 9.1. Calibration 9.1.1. A linear regression resulting from the 6 standards is used to quantitate the analytes in the case. The area of the analyte divided by the area of the internal standard is used in the resulting formula of the calibration curve. 9.2. Dilution Factor 9.2.1. D = Total volume/sample volume 9.3. Multiplier for homogenates/dilutions and non-standard volumes 9.3.1. M = (Vcurve / Vsamp) x D 9.3.1.1. M = Multiplier 9.3.1.2. D = dilution factor 8 of 13

9.3.1.3. Vcurve = matrix volume of calibration curve 9.3.1.4. Vsamp = matrix volume of specimen 9.4. Concentration 9.4.1. C = (A / V) * M 9.4.1.1. C = Concentration (ng/ml) of the analyte in the unknown case. 9.4.1.2. A = Amount of drug in sample 9.4.1.3. V = Volume of sample 9.4.1.4. M = Multiplier 9.5. Max/Min 9.5.1. Percent Difference = ((Rh / Rl)-1) x 100 9.5.1.1. Rh = high result 9.5.1.2. Rl = low result 9.6. Average 9.6.1. Average = (R1 + R2) / 2 9.6.1.1. R1 = first result 9.6.1.2. R2 = second result 9.7. Qualifier Ion Ratios 9.7.1.1.1. Ratio 1 = QL 1 /QN 9.7.2. QL 1 = response of the quantifying product ion 9.7.3. QN = response of the qualifying product ion 10. Quality Control 10.1. For an analysis to be acceptable the following criteria must be met: 9 of 13

10.1.1. Chromatography must be acceptable with a symmetrical (Gaussian) shape. Each analyte of interest must have near baseline resolution from any other peaks in the chromatogram. 10.1.2. The retention time of each analyte should be within ±2% of the expected retention time based on the calibrators and the relative retention time to the internal standard. 10.1.3. The quality control samples shall have an analytical value as stated on the standard and control worksheet, and shall not deviate ±20% from the expected value. 10.1.4. The blank shall represent a specimen of none detected and should not contain analyte signal with appropriate ion ratios above 10% of the low standard. 10.1.5. The internal standard areas of the samples shall not deviate more than 50-200% from the average internal standard areas of the calibrators. 10.1.6. Calculated ion ratios shall not deviate more than 20% to that of the average ion ratios of the calibrators. 10.1.7. Analytical curves must have a coefficient of determination (R 2 ) of 0.992 or greater. 10.1.8. Each calibrator, when calculated against the calibration curve, must not deviate outside ± 20% of the target value (± 25% at LOQ). 10.1.9. Results must fall within the linear range of the assay. Results above or below the linear range can only be reported as less than the low control or greater than the highest control. 10.1.10. Results are reported with two significant figures. All results will be truncated to the nearest significant figure. 11. Validation of Method 11.1. Parameter Bias (all matrix - blood, urine, hd liver) Precision (all matrix - blood, urine, hd liver) Calibration model Result L: -10.86% H: -9.37% L: 2.21% H: 0.83% Linear: Equal weighting 10 of 13

Carryover Ion Ratios Ionization/Suppression: (Not needed if IS coelutes within 0.05 min.) LOD (Determined experimentally) LOQ (Set to lowest calibrator with acceptable Bias/Precision). 12. Reporting Dilution Integrity No carryover observed following high calibrator. Ion ratios were noted to be concentration dependent. Ion ratio range for unknowns shall be determined using the closest calibrator concentration ion ratio(s) as the target. N/A - IS coelutes with target analyte (0.1 min). 1.35 mg/dl 10 mg/dl Average % difference of diluted specimens is 3.51% - lowest: 0.44% Highest 6.63% (n=8) 12.1. The percent difference of duplicate analysis for an analyte must be less than or equal to 25% (see Max/Min in Calculations section) 12.2. Reporting of duplicate analysis should be done according to the table below: Reporting Duplicates Dilution factors of 1 and 2 (or other) Scenario Dil 1 2 REPORT A In curve In curve Average B In curve BQL In value C AQL In curve In value D In curve ND (should be in) Repeat E AQL/BQL AQL/BQL Less than/greater than F BQL ND ND G In curve ND (should be BQL) In value 12.2.1. 11 of 13

12.2.1.1. In Curve = Measured concentration (pre-multiplier) falls within the calibration rang 12.2.1.2. AQL = Measured concentration (pre-multiplier) falls Above Quantitation Limit 12.2.1.3. BQL = Measured concentration (pre-multiplier) falls Below Quantitation Limit 12.2.1.4. ND = None Detected 12.3. Averaging reportable values 12.3.1. Results for duplicate analysis (both falling within calibration curve) shall be truncated prior to averaging. 12.3.2. Enter calculated concentration for each specimen into toxlog. 12.4. Significant figures 12.4.1. Concentrations are truncated and reported with two significant figures in mg/dl. 13. Load Assignment Packet Preparation 13.1. After completing all data generation and reviewing for corrections, the analyst will assimilate the data in the following order: 13.1.1. Load assignment sheets, followed by any additional notes to file pertaining to load. 13.1.2. Load specimen sheet. 13.1.3. Data summary 13.1.4. Chain of Custody. 13.1.5. Standard Sheet. 13.1.6. GC/MS Method and Calibration Report(s). 13.1.7. Running Sequences. 13.1.8. Chromatograms and results for calibrators, controls and specimens 12 of 13

13.1.9. The Load Checklist should be initialed and dated to acknowledge completion of load. The finished data package will be placed in the data review box. 14. References 14.1. Baselt, Randall C. Disposition of Toxic Drugs and Chemicals in Man. Foster City, CA: Chemical Toxicology Institute, 2000. 406-08. Print. 13 of 13