EUROMET PROJECT FINAL REPORT

Similar documents
Derivation of Calibration Factors and Dial Settings Michaela Baker

BIPM.RI(II)-K1.Am-241, CCRI(II)-K2.Am-241, and COOMET.RI(II)-K2.Am.241

BIPM.RI(II)-K1.F-18, BIPM.RI(II)-K4.F18, CCRI(II)-K3.F-18 and APMP.RI(II)-K3.F-18

PROGRESS IN ORGANIZING NATIONAL AND INTERNATIONAL COMPARISONS FOR NUCLEAR MEDICINE MEASUREMENTS

Progress in organizing national and international comparisons for nuclear medicine measurements

The metrology of radioactivity. C. Michotte, BIPM

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs

Intercomparison of 131 I and 99m Tc Activity Measurements in Brazilian Nuclear Medicine Services

G. Ratel 1, C. Michotte 1, M. Sahagia 2, A. Yunoki 3. Abstract

Activity measurements of the radionuclide 153 Sm for the ANSTO, Australia in the ongoing comparison BIPM.RI(II)-K1.Sm-153

RADIOLOGICAL CHARACTERIZATION Laboratory Procedures

C. Michotte*, G. Ratel* and S. Courte*, T. Dziel #, A. Listkowska #, I. Csete, K. Rózsa, L. Szücs, A. Zsinka, C. Fréchou**, C. Bobin**, S.

THE SUPPORT OFFERED BY THE ROMANIAN PRIMARY ACTIVITY STANDARD LABORATORY TO THE NUCLEAR MEDICINE FIELD *

Activity measurements of the radionuclide 111 In for the LNE-LNHB, France in the ongoing comparison BIPM.RI(II)-K1.In-111

NATIONAL AND INTERNATIONAL COMPARISONS ON RADIOPHARMACEUTICALS ACTIVITY MEASUREMENT *

Bureau International des Poids et Mesures (BIPM), Sèvres. National Institute of Standards and Technology (NIST), USA. Becquerel (LNE-LNHB), France

Metrology for Molecular Radiotherapy MetroMRT V. Smyth (coordinator) Lena Johansson, Andrew Fenwick, Kelley Ferreira et al.

Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kv to 300 kv

Final report on CCQM-K36.1 with erratum

European Project Metrology for Radioactive Waste Management

Ionizing Radiation Group K.B. Lee

Bureau International des Poids et Mesures (BIPM),

Half-life data a critical review of TECDOC-619 update

Euramet project 1187 Comparison of Instrument Current Transformers up to 10 ka. Technical protocol (March 2012)

G. Ratel*, C. Michotte*, Y. Hino** *BIPM and **NMIJ, Japan

Activity measurements of the radionuclide 99m Tc for the NIST, USA in the ongoing comparison BIPM.RI(II)-K4.Tc-99m. C. Michotte 1, R.


ɣ-radiochromatography

BUREAU INTERNATIONAL DES POIDS ET MESURES

Radioactivity standardization in South Africa

EMRP JRP ENV54 METROLOGY FOR DECOMMISSIONING NUCLEAR FACILITIES (WP2, WP3)

G. Ratel 1, C. Michotte 1, D. Kryeziu 1, M. Moune 2, A. Iwahara 3. Abstract

Gamma background measurements in the Boulby Underground Laboratory

Bonding Practice Problems

abstractnumber.doc The Support Offered by the Romanian Primary Activity Standard Laboratory to the Nuclear Medicine Field

The first model was based only on the CANBERRA information which we. drawing. Canberra refused any other information.

RADIOACTIVITY. 95 Alpha Particle Solution Standards. 96 Beta Particle and Electron Capture Solution Standards. 97 Gamma Point Source

THE ENVIRONMENTAL RADIATION ANALYSIS ON THE RESIDENTIAL AREA AROUND NUCLEAR POWER PLANTS

Name: Chemistry 400: General Chemistry Miller Exam I September 20, 2017 Approximately 150 points

Which statement best describes the energy change as bonds are formed or broken in this reaction?

Measurement of Tritium in Helium

ADVANCED PLACEMENT CHEMISTRY CHAPTERS 7, 8 AND 9 REVIEW QUESTIONS

Distillation purification and radon assay of liquid xenon

Certified Reference Material - Certificate of Analysis

EURAMET Project no Inter-comparison of a 1000 L proving tank

Work Programme. of the International Bureau of Weights and Measures. for the four years Comité international des poids et mesures

EUROMET Project 702 EUROMET.M.D-K4

Analysis of domestic sludge using the Agilent 4200 MP-AES

Select Assay Capabilities at PNNL

Nuclear Medicine Physics Course Laboratory Number 3: Dose Calibrator Quality Control

NPL REPORT ENG 13. Report on EUROMET key comparison of multiples and submultiples of the kilogram (EUROMET.M.M-K2) M Perkin NOT RESTRICTED

Comparison of the standards of air kerma of the GUM and the BIPM for 60Co 'Y rays

2011, Robert Ayton. All rights reserved.

THE ACTIVITY CALIBRATOR

Radionuclide Imaging MII Detection of Nuclear Emission

CHEM 171 EXAMINATION 1. October 9, Dr. Kimberly M. Broekemeier. NAME: Key

Chemistry 31A Autumn 2004 Professors Chidsey & Zare Exam 4 Name: 9:00am 10:00am 2:15pm 3:15pm 7:00pm 8:00pm S02 OC103 Charles

M [scale units/s] of the system

C. Michotte 1, G. Ratel 1, S. Courte 1, Y. Nedjadi 2, C. Bailat 2, L. Johansson 3, Y. Hino 4. Abstract

Summary of the ICRM Life Sciences Working Group Meeting

Bureau International des Poids et Mesures

CHEMISTRY 102B Hour Exam II. Dr. D. DeCoste T.A.

Bringing metrology to Swiss hospitals: CIR II / TCIR project

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW

g. Looking at the equation, one can conclude that H 2 O has accepted a proton from HONH 3 HONH 3

Faculty of Natural and Agricultural Sciences Chemistry Department. Semester Test 1 MEMO. Analytical Chemistry CMY 283

AS CHEMISTRY 7404/1. Paper 1: Inorganic and Physical Chemistry

6.02 x x baseballs would cover the Earth to a height of several hundred miles.

Development of Radioactivity Standards for Quantitative Positron Emission Tomography

Chem Hughbanks Exam 2, March 10, 2016

INTERLABORATORY COMPARISON OF INSTRUMENT CURRENT TRANSFORMER STANDARDS IN THE RANGE UP TO 800 A AT 50 HZ

Measurements and Dispersion Calculations by the Deutscher Wetterdienst Regarding the Release of Radionuclides at Fukushima Daiichi Nuclear Power Plant

Project 887 Force Uncertainty Guide. Andy Knott, NPL EUROMET TC-M Lensbury, Teddington, United Kingdom 1 March 2007

FINAL REPORT. European co-operation for Accreditation (EA) Interlaboratory Comparison IR3: Calibration of a Radiation Protection Dosimeter

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the LNE-LNHB, France and the BIPM in 60 Co gamma radiation

CHEMISTRY 107 Section 501 Final Exam Version A December 12, 2016 Dr. Larry Brown

GENERAL CHEMISTRY I CHEM SYSTEM FINAL EXAM VERSION A Fall 2016

Chemistry 141 Samuel A. Abrash Chemical Reactions Lab Lecture 9/5/2011

E5 Lewis Acids and Bases: lab 2. Session two lab Parts 2B, 3, and 4. Session one lab Parts 1and 2A. Aquo Complex Ions

Results of the 2015 national indoor radon intercomparison measurements in Serbia

E5 Lewis Acids and Bases: lab 2. Session two lab Parts 2B, 3, and 4. Session one lab Parts 1and 2A. Aquo Complex Ions. Aquo Complex Ion Reactions

INTERNATIONAL KEY COMPARISON OF MEASUREMENTS OF NEUTRON SOURCE EMISSION RATE ( ) CCRI(III)-K9.AmBe

SOLUTIONS. Solutions - page

Radiometric Measurements of Environmental Radioactivity

Activity measurements of the radionuclide 99m Tc for the CNEA, Argentina and the LNMRI/IRD, Brazil in the ongoing comparison BIPM.RI(II)-K4.

DOE S RADIOLOGICAL TRACEABILITY PROGRAM FOR RADIOANALYTICAL MEASUREMENTS: WORK IN PROGRESS

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Last 4 Digits of USC ID:

ISOCS / LabSOCS. Calibration software for Gamma Spectroscopy

19:00 21:30 Registration and reception at Hotel ETAP Altinel. Welcome

Chem 1A Dr. White Fall 2015 Exam 3 Practice Problems

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

Chem Exam 1. September 26, Dr. Susan E. Bates. Name 9:00 OR 10:00

RADON EQUILIBRIUM MEASUREMENT IN THE AIR *

The following examination contains 38 questions valued at 3 points/question. Name

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

Name of research institute or organization: Bundesamt für Gesundheit, Sektion Umweltradioaktivität, Bern

LSC-BASED APPROACH FOR RADON IN SOIL GAS MEASUREMENTS. The Marzeev Institute of Hygiene and Medical Ecology 50, Popudrenko str., Kiev 02094, Ukraine.

Uncertainty of Radionuclide Calibrator Measurements of 99

Transcription:

1. Ref. No: 634 (please leave blank) EUROMET PROJECT FINAL REPORT 2. Subject Field: Ionizing radiation 3. Type of collaboration: comparison 4A: Partners: (institutions) ČMI, BEV, SMÚ 5. Participating countries: Czech Republic, Austria, Slovak Republic 6. Title: Common calibration of ionizing chambers 7. Progress: All devices (4 ionization chambers) were brought to the Czech Metrology Institute and calibrated in a week 18 to 21 September 2001. At the same week there were provided and prepared standards with radionuclides 18 F, 67 Ga, 99m Tc, 201 Tl, 123 I, 131 I in the form of 5 or 4 ml water solution in the usually two P6-like ampoules. The standards of 133 Xe and 85 Kr were prepared in the gas ampoules. The 137 Cs was in the form of gel. Specific activity of the solutions and gases were measured either by calibrated spectrometer or calibrated ionization chamber. The 131 I was standardized by absolute method and sent to S.I.R Troubles were found in the calibration of 123 I and 201 Tl. The more details are described in the technical final report which is an attachment of this final report. 8. Coordinator s name: Address: Pavel Dryak, ČMI - IIZ Praha, Radiová 1, Czech Republic Telephone +420 2 66020497 Fax: +420 2 66020466 e-mail: pdryak@cmi.cz 9. Completion Date: 29 March 2002 10. Coordinator s signature:

Technical Final Report of the EUROMET Project E634 The aim of the project was a calibration of the secondary chambers especially by short living radionuclides. If the secondary chambers were calibrated by the producer, the resulting activities from producer s calibration were compared to the CMI standards. Table 1: Participants, persons and devices laboratory Responsible persons Calibrated devices Czech Metrology Institute Inspectorate for Ionizing Radiation Czech Republic Pavel Dryák, Veronika Olšovcová 1. NPL-CRC No. 400001 2. BQM 4 No. 33 Slovak Metrology Institute Slovak Republic Bundesamt fuer Eich- und Vermessungswessen, Austria Anton Švec Petra Jachs, Michael Kreuziger VacuTec 70 129 No.98012 with steel liner 2,2 mm connected to Electrometer Keithley 6517A No.0695463 ISOCAL IV, model 286/SA/102 All devices were brought to the Czech Metrology Institute and calibrated by responsible persons, given in Table 1, in the week from 18 th to 21 st September 2001. During the same week there were provided and prepared standards with radionuclides 18 F, 67 Ga, 99m Tc, 201 Tl, 123 I and 131 I in the form of 5 or 4 ml water solution in the usual two P6-like ampoules. While the standards of 133 Xe and 85 Kr were prepared in the gas ampoules, the 137 Cs was in the form of gel. Specific activities of solutions and gases were measured either by calibrated spectrometer or calibrated 4π-γ ionization chamber. The 131 I was standardized by absolute method and sent to S.I.R. Details of the standards used are in the following Table 2. 2

Table 2: Standards prepared for E634 Radionuclide Vial number Activity [MBq] (σ) Reference time Day, hour, min Volum e [ml] Weight [g] Chemical composition Impurities at reference time 137 Cs 591-15 11,682(105) 18.9.2001 12:00 5 ml 5,0720 gel 1) 137 Cs 591-16 11,903(107) 18.9.2001 12:00 5 ml 5,0394 gel 134 Cs 137 Cs 591-17 11,566(104) 18.9.2001 12:00 5 ml 5,0250 gel 0,09 ± 10 137 Cs 591-18 12,070(109) 18.9.2001 12:00 5 ml 5,0338 gel 137 Cs 591-19 11,828(106) 18.9.2001 12:00 5 ml 5,0029 gel 131 I 799-01 19,870(199) 18.9.2001 12:00 5 ml 5,0783 NaI in H 2 O 126 I 2) 131 I 799-02 19,448(194) 18.9.2001 12:00 5 ml 5,0545 NaI in H 2 O < 0,05 18 F 800-002 52,494(787) 18.9.2001 13:00 4 ml 4,0727 KF + Na 2 S 2 O 3 < 0,1 3) 18 F 800-003 52,091(781) 18.9.2001 13:00 4 ml 4,0473 +Na 2 CO 3 123 I 801-002 91,00(109) 18.9.2001 15:00 5 ml 4,9991 124 I 0,004 123 I 801-003 91,45(110) 18.9.2001 15:00 5 ml 5,0136 KI + Na 2 S 2 O 3 126 I 0,01 3) 131 I 0,007 67 Ga 802-001 103,32(155) 19.9.2001 10:00 5 ml 5,0261 GaCl 3 + HCl 66 Ga 3) 67 Ga 802-002 101,34(152) 19.9.2001 10:00 5 ml 5,0214 < 0.015 201 Tl 803-001 74,571(820) 19.9.2001 12:00 5 ml 5,0083 KCl + Tl 2 SO 4 200 Tl 0,113 3) 201 Tl 803-002 73,993(814) 19.9.2001 12:00 5 ml 4,9949 +HNO 3 202 Tl 0,251 201 Tl 803-003 75,289(828) 19.9.2001 12:00 4 ml 4,0173 99m Tc 804-002 96,510(965) 19.9.2001 13:00 5 ml 5,0031 NaCl + 99 Mo 9.10-5 3) 99m Tc 804-003 97,637(976) 19.9.2001 13:00 5 ml 5,0160 NH 4 OH 131 I < 3.10-6 133 Xe 188,6(70) 19.9.2001 12:00 33 ml nitrogen 131m Xe < 0,66 4) 85 Kr 2437(120) 20.9.2001 12:00 33 ml nitrogen None gamma 5) Remarks Remark 1) The solution was standardized by absolute method. The composition of the gel is similar to water, with density 1 g.cm -3. 2) The solution was standardized by absolute method. It was sent to S.I.R.. The difference from the last four records (NIST, OMH, NPL, LNMRI) was 0,5. 3) The solutions were standardized by calibrated CMI ionization chamber which was calibrated by CMI standard solutions. 4) The activity of gas in the 33 ml glass ampoule (length 55 mm, int. diam. 27 mm, glass wall 1,5 mm) was standardized by calibrated HpGe spectrometer. The calibration factor is valid for 33 ml glass ampoule. 5) The activity of gas in the 7 ml glass ampoule (length 5 cm, int. diam. 1,3 cm, glass wall 1 mm) was standardized by calibrated HpGe spectrometer. The calibration factor is valid for 7 ml glass ampoule. The calibration factors given by producers and used in the comparison are listed in the following Table 3. The comparison of the CMI values and response of the devices is given in the Table 4. Table 3: The producers calibration factors NPL-CRC BQM * VacuTec ISOCAL IV * (pa/bq) 137 Cs 5,751 5,102 6,812 5,751 131 I 3,999 3,984 7,347 3,999 18 F 10,35 9,524 13,947 10,35 123 I 1,685 4,000 10,83 1,685 67 Ga 1,547 2,849 10,52 1,547 201 Tl 0,8557 4,762 10,89 0,8557 99m Tc 1,227 2,146 12,58 1,227 133 Xe Not given 3,610 4,878 Not given 85 Kr 0,0320 0,613 0,376 0,0320 * recalculated from the original calibration factors which were given as kbq/pa recalculated from the original calibration factors which were given as MBq/pA equivalent to NPL-CRC calibration 3

Table 4: CMI activity values and responses of the NPL-CRC, BQM, VacuTec and ISOCAL IV secondary chambers. In case of NPL-CRC secondary chamber the uncertainty of the calibration factors were not given, type A uncertainty is less than 0,5. Radionuclide Vial number CMI activity, A MBq NPL A r (A r A)/A BQM A 1 (A 1 A)/A VacuTec A 2 (A 2 A)/A ISOCAL IV A 3 (A 3 A)/A 137Cs 591-15 11,682(105) 11,509 * -1,48 11,47(31) -1,82 11,82(12) +1,2 11,51(25) * -1,44 137Cs 591-16 11,903(107) 11,729 * -1,46 11,79(31) -0,96 12,06(12) +1,3 11,73(25) * -1,44 137Cs 591-17 11,566(104) 11,389 * -1,53 11,35(31) -1,88 11,69(12) +1,1 11,39(25) * -1,50 137Cs 591-18 12,070(109) 11,889 * -1,50 11,89(31) -1,50 12,21(12) +1,1 11,89(25) * -1,48 137Cs 591-19 11,828(106) 11,669 * -1,34 11,63(31) -1,68 11,98(12) +1,2 11,65(25) * -1,50 131I 799-01 19,870(199) 20,025 +0,78 19,30(58) -2,86 20,52(30) +3,3 19,99(15) +0,62 131I 799-02 19,448(194) 19,587 +0,71 18,91(58) -2,78 20,11(30) +3,4 19,56(15) +0,58 18F 800-02 52,494(787) 53,14 +1,23 50,0(20) -4,78 54,96(82) +4,7 52,93(107) +0,82 18F 800-03 52,091(781) 52,857 +1,47 49,6(20) -4,88 54,28(82) +4,2 52,52(107) +0,82 123I 801-02 91,00(109) 96,554 +6,10 96,74 +6,31 97,5(1,5) +7,1 95,8(21) +5,27 123I 801-03 91,45(110) 97,025 +6,10 97,16 +6,25 98,0(1,5) +7,1 96,3(21) +5,31 67Ga 802-01 103,32(155) 103,827 +0,49 94,3(47) -8,70 102,1(1,5) -1,2 103,0(28) -0,32 67Ga 802-02 101,34(152) 101,665 +0,32 92,4(47) -8,80 100,2(1,5) -1,1 100,9(27) -0,42 803-01 74,571(820) 83,78 * +12,3 72,82 * -2,4 80,0(1,2) * +7,3 84,33(89) * +13,1 803-02 73,993(814) 83,76 * +13,2 72,16 * -2,5 79,3(1,2) * +7,2 83,77(89) * +13,2 803-03 75,289(828) 84,14 * +11,7 72,82 * -3,4 84,23(89) * +10,6 99mTc 804-02 96,510(965) 98,171 +1,72 97,7(29) +1,25 98,9(1,5) +2,4 97,34(165) +0,86 99mTc 804-03 97,637(976) 99,288 +1,69 98,8(29) +1,16 100,0(1,5) +2,4 98,49(167) +0,88 133Xe 188,6(70) 189(9) +0,4 202,1 +7,1 85Kr 2437(120) 2569 +5,4 * * corrected to impurities not measured gel compound; provided calibration factor for liquid solution only 4

From the data presented in Table 4 it is clear that there is some problem in the calibration of the CMI chamber for 123 I and 201 Tl. When the activity measured with NPL-CRC chamber is taken as the reference activity of 123 I and 201 Tl, the resulting activities and deviations from the NPL-CRC reference value for BQM, VacuTec and ISOCAL are shown in the following Table 5. Table 5: Activities of 123 I and 201 Tl measured by BQM, VacuTec, ISOCAL when compared with NPL-CRC chamber results. Radionuc lide Vial number 123I 801-02 123I 801-03 803-01 803-02 803-03 NPL-CRC activity, A r MBq BQM A 1 (A 1 -A r )/A r VacuTec A 2 (A 2 -A r )/A r ISOCAL IV A 3 (A 3 -A r )/A r 96,554 96,74 +0,19 97,5 +0,98 95,801-0,78 97,025 97,16 +0,14 98,0 +1,0 96,301-0,75 83,78 72,82-13 80,0-4,5 84,333 +0,66 83,76 72,16-14 79,3-5,3 83,768 +0,01 84,14 72,82-13 84,234 +0,11 Comment The calibration of 137 Cs, 131 I, 18 F, 67 Ga and 99m Tc is acceptable. The individual deviation can be used for the new calibration of chambers. The calibration of 133 Xe and 85 Kr was possible only for BQM and VacuTec, due to the dimensions of well of chambers. However, results are not comparable due to different calibration conditions. It is generally acceptable for both chambers. The real problem is in the calibration of CMI chamber for 123 I and 201 Tl. The calibration was done 5 years ago by commercially available material. This knowledge leads to necessity of recalibration of CMI chamber for the above-mentioned radionuclides. Then a new comparison of activity measurements of these radionuclides would be advisable. 5