Radiation shielding for undulator beamline in Indus-2 synchrotron radiation source

Similar documents
Shielding calculations for the design of new Beamlines at ALBA Synchrotron

Radiation Protection Dosimetry (2006), Vol. 118, No. 3, pp Advance Access publication 6 October 2005

Shielding Design for the Imaging and Medical Beamline at the Australian Synchrotron

Radiation Safety Assessment of the CLS Beamlines Using FLUKA Monte-Carlo Code

Radiation Protection Considerations for the Cryogenic In-Vacuum Undulator of the EMIL Project at BESSY

Radiation Protection At Synchrotron Radiation Facilities

Occupational Radiation Protection at Accelerator Facilities: Challenges

Vacuum System of Synchrotron radiation sources

SLAC-PUB Submitted to Radiation Protection and Dosimetry. Work supported by Department of Energy contract DE-AC02-76SF00515

FLUKA Calculations for the Shielding Design of the SPPS Project at SLAC*

Comparison of FLUKA and STAC8 for shielding calculations of the hard X-ray line of the LCLS

In vacuum ID beam line shielding commissioning and direct gasbremsstrahlung measurements at Synchrotron SOLEIL

Radiological safety studies for the TeraFERMI beamline at

SHIELDING CALCULATIONS FOR THE HARD X-RAY GENERATED BY LCLS MEC LASER SYSTEM R. QIU, J. C. LIU, S. H. ROKNI AND A. A. PRINZ

Measurements of Radiation Doses Induced by High Intensity Laser between and W/cm 2 onto Solid Targets at LCLS MEC Instrument

Radiological Implications of Top-up Operation at Canadian Light Source: Dose Computations and Measurements at the Vulnerable Points

RADIATION PROTECTION AT SYNCHROTRON RADIATION FACILITIES* James C. Liu 1 and Vaclav Vylet 2. Abstract

Gas Bremsstrahlung at the Diamond Light Source. P. Bonner, R. Ryder

Bulk shielding design for the MAX IV facility

PHOTON-NEUTRON SHIELDING CALCULATIONS FOR ELECTRON STORAGE RINGS. J. C. Liu, W. R. Nelson and K. R. Kase ~~~_

Radiation Safety Considerations for the TPS Accelerators

FLUKA calculations for the beam dump system of the LHC : Energy deposition in the dump core and particle spectra in the beam loss monitors

Dosimetric Quantities and Neutron Spectra Outside the Shielding of Electron Accelerators

Radiation Safety at LCLS: The Photon Beam s Maximum Capability and Material Damage Potential

New irradiation zones at the CERN-PS

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

European Organisation for Nuclear Research European Laboratory for Particle Physics

The photoneutron yield predictions by PICA and comparison with the measurements

Radiation Protection Considerations *

Radiation measurements around ESRF beamlines

Research with Synchrotron Radiation. Part I

How Does It All Work? A Summary of the IDEAS Beamline at the Canadian Light Source

Secondary Particles Produced by Hadron Therapy

Estimation of Radioactivity and Residual Gamma-ray Dose around a Collimator at 3-GeV Proton Synchrotron Ring of J-PARC Facility

Y. P. Feng, B. Lai, I. McNulty, R. J. Dejus. E(.J. Randall,

Synchrotron Facilities And Free Electron Lasers. Introduction

ATLAS NOTE. July 15, TGC Background Hit Rate: Comparison Between 2012 Data and Simulation. Abstract

Machine Protection. Lars Fröhlich DESY. CERN Accelerator School on FELs and ERLs June 9, 2016

Since the beam from the JNC linac is a very high current, low energy beam, energy loss induced in the material irradiated by the beam becomes very lar

Advanced Storage Photon Ring Source Upgrade Project:

RADIATION SOURCES AT SIBERIA-2 STORAGE RING

8 th International Workshop on Radiation Safety at Synchrotron Radiation Sources

Radiation Protection Dosimetry Advance Access published October 28, Radiation Protection Dosimetry (2010), pp. 1 8

Journal of Radiation Protection and Research

RADIOLOGICAL IMPACT OF THE TRIGAACCELERATOR-DRIVEN EXPERIMENT (TRADE)

The FLUKA study of the secondary particles fluence in the AD-Antiproton Decelerator target area.

Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry

Secondary Radiation and Shielding Design for Particle Therapy Facilities

Synchrotron Methods in Nanomaterials Research

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

Radiation background simulation and verification at the LHC: Examples from the ATLAS experiment and its upgrades

Developments for the FEL user facility

Estimate of Undulator Magnet Damage Due to Beam Finder Wire Measurements

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Insertion Devices Lecture 2 Wigglers and Undulators. Jim Clarke ASTeC Daresbury Laboratory

Beam diagnostics: Alignment of the beam to prevent for activation. Accelerator physics: using these sensitive particle detectors.

Radiological Issues at JLab

Radiation protection considerations along a radioactive ion beam transport line

Accelerator Design and Construction Progress of TPS Project

Comparison with simulations to experimental data for photoneutron reactions using SPring-8 Injector

Shielding Design Considerations for Proton Therapy Facilities

1. RADIOACTIVITY AND RADIATION PROTECTION

3. Synchrotrons. Synchrotron Basics

Calibration of the GNU and HSREM neutron survey instruments

Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials*

Synchrotron radiation: A charged particle constrained to move in curved path experiences a centripetal acceleration. Due to it, the particle radiates

Colliders and the Machine Detector Interface

Part E. Radiation monitors. Radiation Safety - JUAS 2014, X. Queralt. Part E. Radiation monitors

COST MP0601 Short Wavelength Laboratory Sources

Simulation of the radiation levels and shielding studies at the BDI positions in IR4

Theoretical Assessment of Aircrew Exposure to Galactic Cosmic Radiation Using the FLUKA Monte Carlo Code

Radiation Shielding of Extraction Absorbers for a Fermilab Photoinjector

Introduction to Accelerator Physics Part 1

Secondary Neutron Dose Measurement for Proton Line Scanning Therapy

Collimator Designed for the APS Front End Operating in the Top-up Mode. D. Shu, T. Sanchez, and T. Kuzay

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Delta undulator magnet: concept and project status

First propositions of a lattice for the future upgrade of SOLEIL. A. Nadji On behalf of the Accelerators and Engineering Division

A2. Light Source. ( i ) Units of light intensity

2.6 Electron transport lines

Introduction to Accelerator Physics

Physics sources of noise in ring imaging Cherenkov detectors

Abstract. 1. Introduction

Measurements with the new PHE Neutron Survey Instrument

Study on collimation and shielding of the back-streaming neutrons at the CSNS target

PoS(KAON09)023. Beam Hole Photon Veto For J-PARC K O TO experiment. Yosuke Maeda Kyoto University

Calculation of the Dose Equivalent Rate from Induced Radioactivity Around the CNGS Target and Magnetic Horn

Extension of CASCADE.04 to estimate neutron fluence and dose rates and its validation

Light Source I. Takashi TANAKA (RIKEN SPring-8 Center) Cheiron 2012: Light Source I

Optimization studies of photo-neutron production in high-z metallic targets using high energy electron beam for ADS and transmutation

Code inter-comparison and benchmark for muon fluence and absorbed dose induced by an 18 GeV electron beam after massive iron shielding

SURVEY AND ALIGNMENT FOR THE SYNCHROTRON LIGHT SOURCE ELETTRA

MAGNET INSTALLATION AND ALIGNMENT FOR THE FUJI TEST BEAM LINE AT KEKB

Synchrotron radiation: A charged particle constrained to move in curved path experiences a centripetal acceleration. Due to this acceleration, the

CHARACTERISTICS OF DEGRADED ELECTRON BEAMS PRODUCED BY NOVAC7 IORT ACCELERATOR

Accelerator Upgrade Problem

Jan. 5, 2006 Development of a Helical Undulator for ILC Positron Source

ICTP-IAEA Joint Workshop on Nuclear Data for Science and Technology: Medical Applications. 30 September - 4 October, 2013

Radiation Quantities and Units

Transcription:

Radiation shielding for undulator beamline in Indus-2 synchrotron radiation source P. K. Sahani 1,5, A. K. Das 2, Haridas G. 3, A. K. Sinha 4,5, B. N. Rajasekhar 2,5, T. A. Puntambekar 1 and N K Sahoo 2,5 Abstract 1 Beam Diagnostics and Coolant System Division, Raja Ramanna Centre for Advanced Technology, Indore-452013, India 2 Atomic and Molecular Physics Division, Bhabha Atomic Research Centre, Mumbai-400085, India 3 Health Physics Division, Bhabha Atomic Research Centre, Mumbai-400085, India 4 Synchrotrons Utilization Section, Raja Ramanna Centre for Advanced Technology, Indore-452013, India 5 Homi Bhabha National Institute, Training School Complex, Anushakti Nagar, Mumbai 400094, India Indus-2 synchrotron radiation source (SRS) is operational at Raja Ramanna Centre for Advanced Technology (RRCAT), India with electron beam current up to 200mA at 2.5 GeV and is being used for conducting research in basic and applied branches of science and engineering. Two planar undulators and one APPLE-II type undulator have been installed and commissioned in the straight sections of Indus-2 SRS to increase brilliance required for enhancing utilization. One of the insertion devices, a planar permanent magnet (PPM) undulator (U 1 ) will be used as a source of ultraviolet to soft X-ray photons for a beamline dedicated for atomic, molecular and optical science (AMOS) research. Appropriate radiation shielding enclosure (hutch) is mandatory requirement to operate this beamline for conducting experiments. The shielding requirement of hutch for this beamline in Indus-2 experimental hall was evaluated using FLUKA Monte Carlo code. The radiation dose due to gas bremsstrahlung radiation, photo-neutron and synchrotron radiation were simulated and the shielding parameters were optimized to achieve regulatory dose limit. This paper describes the design and optimization studies of a gas bremsstrahlung stop and shielding hutch using FLUKA Monte Carlo code. 1. Introduction Indus-2, 2.5 GeV electron storage ring is operational at Raja Ramanna Centre for Advanced Technology (RRCAT), Indore, India and is a dedicated synchrotron radiation source (SRS) for scientific and technological research applications. This facility comprises of 20 MeV microtron as electron injector, a booster synchrotron for delivering 550 MeV electrons and a 2.5 GeV electron storage ring. 550 MeV electrons from booster synchrotron are injected, stored and ramped to 2.5 GeV in Indus-2 SRS with an operating beam current of 200 ma. The facility has provision for accommodating 21 bending magnet and 5 insertion device beamlines. At present, thirteen bending magnet based beamlines (10 hard X- ray beamlines & 3 soft X-ray beamlines) are regularly being used for conducting research in medicine, material science, geology, biology etc. As a part of the program for enhancing utilization of Indus-2 SRS to perform advanced experiments requiring higher brilliance over broad energy range (ultraviolet to hard X ray through soft X- ray), many insertion devices have been planned to be installed. Two planar and One APPLE-II type undulators are installed in 10.4 m long straight sections of Indus-2 storage ring. One such device, a planar permanent magnet (PPM) undulator (U1) has been installed in the LS-2 long straight section. This undulator is going to be used as the photon source for a beamline to conduct Atomic, Molecular and Optical Science (AMOS) research [1]. Operation of this beamline requires commissioning of an appropriate radiation shielding enclosure (hutch) to keep the radiation dose in and around this beamline within permissible limits. Gas bremsstrahlung, produced due to interaction of high energy electron beam with residual gas molecules present inside the storage ring chambers travels in to the beamline (along with the synchrotron radiation) is a potential radiation hazard to the working personnel in beamline [2, 3]. Gas bremsstrahlung is significantly higher in beamlines based on insertion devises compared to bending magnet beamlines as the intensity of gas bremsstrahlung depends on beam energy and length of the straight sections through which the electrons move. Gas bremsstrahlung radiation is highly focused in the forward direction, but on interacting with beamline components, the radiation gets scattered in all directions. Radiation protection at synchrotron radiation beamlines, especially undulator/wiggler based beamlines therefore is challenging due to the presence of high energy gas bremsstrahlung. In addition to the scattered bremsstrahlung, photo-neutrons produced by gas bremsstrahlung complicate radiation issues further. To contain the radiation dose within allowed limits, beamline components are contained in a shielded enclosure made of materials like steel and

lead along with additional local shielding for radiation safety. In case of AMOS beamline, the first plane mirror acts as the first scattering element. In the present work, FLUKA Monte Carlo code [4] was used to study the gas bremsstrahlung characteristics like spectrum, angular distribution and dose profile in the beamline. Photo-neutron component is also evaluated and accordingly shielding requirements are finalised. 2. Radiation shielding evaluation The layout of the AMOS beamline along with the shielding hutch and beamline components are shown in Fig-1. Mirror was kept inside the optics hutch of the beamline and is at a distance of 18.5 m from the centre of U1 undulator. This acts as the first scattering element for gas bremsstrahlung radiation which channels along with synchrotron radiation. Radiation shielding evaluation is carried out in such a manner that radiation is confined within the optics hutch. Monochromator hutch and experimental hutch are made normally accessible. The dose limit for normally accessible areas is kept less than 1 µsv/h as per the regulatory requirement [5]. 2.1 Gas bremsstrahlung Fig-1: Layout of AMOS beamline with shielding hutch 2.1.1 Unshielded gas bremsstrahlung dose FLUKA simulation was performed to estimate the unshielded gas bremsstrahlung dose due to 2.5 GeV electrons passing through 10.4 m air target. A pencil beam of electrons of energy 2.5 GeV was allowed to incident on air target at NTP and the absorbed energy was scored in an ICRU tissue phantom (radius =15 cm and length =30 cm) placed at a distance of 19.5 m from the centre of straight section using USRBIN scoring estimator. The ICRU tissue [6] comprises of four elements: Hydrogen (10.12%), Carbon (11.1%), Nitrogen (2.6%) and Oxygen (76.18%). The schematic diagram of the geometry is shown in figure-2. Electron is killed at the end of straight section 2.5 GeV electron Air (10.4 m) Vacuum Tissue phantom Straight section 19.5 m Fig.2 Schematic of unshielded gas bremsstrahlung dose estimation in water phantom.

The average pressure in Indus-2 storage ring is maintained at 1 ntorr. The number of air molecules present at this pressure is significantly less compared to air at atmospheric pressure. This will lead to widening of emitted gas bremsstrahlung cone due to multiple scattering effects. Therefore multiple scattering effects are suppressed for all the simulation studies. The threshold for Moller scattering of electrons was set at 10 MeV to minimize angular divergence due to production of δ-rays [7, 8]. Results of simulations performed at atmospheric pressure (for better statistics), were then normalized to actual pressure of 1 ntorr. In actual scenario, an electron gets bent by a bending magnet at the end of straight section. In simulation, electrons are killed intentionally at the end of straight section using high electron transport cut off (2.5 GeV) in order to avoid the electron contribution in gas bremsstrahlung dose. The transport cut off for photon, electron and positron in air and tissue media were set at 1, 10 and 10 kev respectively. Tissue phantoms are binned longitudinally into 30 bins (1cm per bin) for estimation of dose equivalent. Dose equivalent rate was scored through USRBIN scoring card using fluence to dose equivalent conversion coefficients through AUXSCORE card. The fluence spectrum was obtained using USRTRACK scoring card and output data were scaled to actual pressure of 1 ntorr. The statistical uncertainties in the results are found to be within ± 3.0 %. The estimated gas bremsstrahlung dose rate is found to be 102.73 msv/h-ntorr at a distance of 19.5 m from the straight section for 200 ma stored current in Indus-2. Gas bremsstrahlung energy spectrum and corresponding depth dose inside tissue phantom are shown in figure-3. Fig.3- Gas bremsstrahlung energy spectrum (L) and the depth dose profile (R) inside tissue phantom due to 2.5 GeV electrons passing through 10.4 m straight section 2.1.2 Scattered gas bremsstrahlung dose from the mirror assembly In order to evaluate scattered gas bremsstrahlung contribution from mirror assembly, direct spectrum is allowed to incident on the silicon mirror which is inclined at 10 degrees with respect to synchrotron beam axis. This mirror is located at a distance of 18.5 m from the centre of the straight section of Indus-2 storage ring. A schematic diagram of the geometry used for the simulation of scattered gas bremsstrahlung dose rate and spectrum is shown in figure-4. Two detectors (tissue phantoms of radius = 15 cm and length= 30 cm) placed at 1 m along forward direction (0 0 ) and lateral direction (90 0 ) from the silicon mirror are used to score gas bremsstrahlung dose and spectrum. The gas bremsstrahlung energy spectrum along 0 0 and 90 0 are shown in figure-5 and the corresponding estimated dose equivalent rate in the detectors placed along (0 0 ) and perpendicular (90 0 ) to beam axis for 200 ma stored current are shown in figure-6. Figure-6 shows that the dose in forward direction (0 0 ) is increasing with respect to depth in tissue phantom whereas a decreasing trend is observed in lateral direction (90 0 ). This is in consistent with the energy spectrum shown in figure-5. The increase in dose is due to the high energy photons in the forward direction and relatively low energy in lateral direction as shown in figure-5. The maximum dose equivalent rate at 1 m along (0 0 ) and perpendicular (90 0 ) to beam axis is found to be 64.3 msv/h-ntorr and 0.31 μsv/h-ntorr respectively.

Electron Air @ 1 atm Electron is killed at the end of straight section Vacuum 1 m Mirror Chamber Silicon Mirror Tissue Phantom (90- degree) 1 m Tissue Phantom Straight Section (10.4 m) 18.5 m Fig.4- Geometry used for bremsstrahlung dose evaluation in 0 and 90 degree from the mirror Fig.5 - Scattered gas bremsstrahlung spectra from pre-mirror in 0 and 90-degree direction Fig.6- Depth dose profile inside 30 cm tissue phantom in 0 0 (L) and 90 0 (R) 2.1.3 Scattered gas bremsstrahlung dose outside beamline hutch For optimization of gas bremsstrahlung stop, first beamline component with plane mirror (made of silicon), mirror chamber (made of stainless steel) and complete hutch wall (made of lead) are simulated. Initially the calculation for bremsstrahlung stop is performed with a lead shielding of 10 and 15 cm thickness after the mirror. The direct gas bremsstrahlung spectrum is allowed to fall on the mirror and dose equivalent rate outside the hutch wall is estimated. Preliminary estimates with 10 to 15 cm of lead gave a forward (0 0 ) dose rate of 254 and 15.4 μsv/h-ntorr for 200 ma stored current at 2.5 GeV. This exercise helped in optimizing the final shielding required for bremsstrahlung stop. An optimized thickness of 20 cm is found out as the bremsstrahlung radiation shield (BR stop) thickness along the central axis. Size of BR stop is optimized in lateral direction taking the advantage of angular distribution of gas bremsstrahlung. In the lateral direction, a

thickness of 10 cm is found to be adequate to maintain radiation levels outside the hutch within acceptable limit. In order to evaluate scattered dose around optics hutch, four detectors (Tissue phantoms- D1, D2, D3 and D4) are placed and dose is scored. Schematic layout of the beamline with the proposed lead shielding for optics hutch is shown in figure-7. Fig.7- Geometry used for bremsstrahlung dose evaluation outside hutch area. 2.2 Photo-neutron dose outside the beamline hutch Gas bremsstrahlung on hitting the mirror/mirror chamber will generate photo-neutrons. Photo-neutron dose is scored outside the beamline hutch in the four detectors D1, D2, D3 and D4. For photo-neutron simulation, LAMBIAS card is used for reducing the photon interaction length for nuclear inelastic interactions by a factor of 50, since the photo nuclear cross-section is smaller as compared to cross-section for electromagnetic interactions. From the study, it has been observed that though 20 cm thick lead (BR stop) is sufficient enough to reduce gas bremsstrahlung dose within regulatory limit (1 µsv/h) but it enhances the photo-neutron dose rate. Therefore 15 cm polyethylene (density 0.93 g/cc) is added to the BR stop for neutron shielding. The dose equivalent rate due to photons and photo-neutrons with optimized shielding from four detectors for 200 ma at 2.5 GeV are shown in Table-1. Location & Detector Dose Equivalent rate (μsv/h-ntorr) for 200 ma stored current in Indus-2 Unshielded With Mirror assembly With BR stop and hutch walls With BR stop, Neutron shielding and hutch walls Total Gas BR Gas BR Gas BR Gas BR Photoneutron Photoneutron Total Forward (0-degree) D1 102730 64300 0.58 6.63 7.21 0.22 0.07 0.29 Beam pipe hole D2 - - 1.12 1.49 2.61 0.94 0.33 1.27 Corner of the hutch D3 - - 0.24 0.15 0.39 0.24 0.26 0.50 Lateral (90-degree) D4-0.31 0.06 0.11 0.17 0.06 0.12 0.18 Table-1: Scattered gas bremsstrahlung and photoneutron dose rate around the hutch

2.3 Synchrotron radiation dose evaluation outside the beamline hutch Synchrotron radiation in the energy range 1 to 100 kev from undulator U1 is generated using SPECTRA [9] code. To evaluate the dose due to direct synchrotron spectrum, this complete spectrum is programmed as incident beam using a subroutine (source.f) in FLUKA input. To evaluate the unshielded synchrotron dose, the spectrum is allowed to incident directly on the 30 cm thick tissue phantom (placed at D1 location). Input synchrotron spectrum and the corresponding depth dose profile inside the water phantom for 200mA stored current in Indus-2 are shown in figure-8. Maximum dose equivalent dose equivalent rate of 7.24 x 10 4 Sv/h for 200 ma stored current in Indus-2 was observed at the surface of the tissue phantom. Fig-8: Synchrotron radiation spectrum (L) and the depth dose profile (R) inside tissue phantom for 200mA stored current in Indus-2 To evaluate the scattered dose outside the beamline, the synchrotron photon spectrum (Fig-8) is allowed to transport through the beamline components inside the lead hutch wall (Fig-7). The dose equivalent rate was scored using the tissue phantoms D1, D2, D3 and D4 placed along the hutch wall (5 mm front and 2 mm sidewall). The scattered synchrotron dose rate in all the detectors placed outside the hutch wall are found to be less than 0.1 µsv/h for 200 ma stored current in Indus-2. 3. Conclusion The radiation shielding requirements for insertion device based AMOS beamline in Indus-2 is evaluated using Monte Carlo code, FLUKA. The dose rate due to direct gas bremsstrahlung, synchrotron radiation and scattered radiation are evaluated. Photo-neutron dose rate is also simulated. Based on the study, the recommendations for shielding enclosure of the beamline are as follows. Front and lateral wall of the hutch should have 5 and 2 mm thick lead shielding respectively. The optimised gas bremsstrahlung shielding (BR stop) behind the mirror assembly should be 20 cm thick lead along the central axis. Lateral thickness above the central axis (5 cm above) can be reduced to 10 cm. Additional 15 cm Polyethylene should be added to BR stop to reduce photoneutron dose rate. The beam pipe hole showed dose equivalent rate 1.45 µsv/h. Annular space between the beam pipe and the hutch wall near the beam hole needs to be plugged with lead wool to reduce the dose rate to acceptable levels and stop leakage of radiation. The scattered synchrotron dose outside hutch was found to be within the permissible limit. Beyond optics hutch, there is no shielding requirement for monochromator and experimental hutch of the beam line and the areas will be accessible in beam ON condition. Acknowledgement The authors are thankful to Dr. P. A. Naik, Director, RRCAT for his encouragement in the work. We are also grateful to Dr. R.M.Tripathi, Head, Health Physics Division, BARC and Dr. Tapas Bandyopadhyay, Head, Accelerator Radiation Safety Section, BARC for fruitful technical discussions.

References [1] Asim Kumar Das, B.N. Rajasekhar and N.K. Sahoo, An undulator based high flux and high resolution beamline for Atomic, Molecular and Optical Science (AMOS) research at INDUS-2 synchrotron radiation source (SRS), BARC Report BARC/2014/E/008. [2] W. P. Swanson, Radiological safety aspects of the operation of electron linear accelerators, IAEA technical Report Series No.188 (1979). [3] James C. Liu and Vaclav Vylet, Radiation protection at synchrotron radiation facilities, SLAC-PUB- 9006 (2001). [4] A. Fasso, A. Ferrari, J. Ranft and P. R. Sala, FLUKA A multi-particle transport code, CERN- 2005-10, INFN/TC_05/11, SLAC-R-773 (2005). [5] AERB Safety Directive, No.2 (1991). [6] http://physics.nist.gov/cgi-bin/star/compos.pl?matno=262. [7] Y. Asano, Simulation of angular distribution of gas bremsstrahlung depending on the residual gas pressure of storage ring, Proceedings of the second International Workshop on EGS, Tsukuba, Japan, KEK Proceedings 200-20, 286-292 (2000). [8] A. Ferrari, M. Pelliccioni and P.R. Sala, Estimation of fluence rate and absorbed dose rate due to gas bremsstrahlung from electron storage rings, Nuclear Instruments and Methods in Physics Research B83, 518-524 (1993). [9] Takashi Tanaka and Hideo Kitamura, SPECTRA: a synchrotron radiation calculation code, Journal of Synchrotron Radiation, Vol. 8, 1221 1228 (2001).