Supporting Information

Similar documents
Supporting Information for:

An insulin-sensing sugar-based fluorescent hydrogel

SUPPORTING INFORMATION

g, 2.5 mol%) were placed in a sealed tube and then N-methylpyrrolidone (NMP) (1.00 ml)

Electronic Supplementary Information. Iridium(III) phosphors with bis(diphenylphorothioyl)amide ligand for

Supporting Information for. Metallonaphthalocyanines as Triplet Sensitizers for Near-Infrared. Photon Upconversion beyond 850 nm

DATA THAT YOU MAY USE UNITS Conventional Volume ml or cm 3 = cm 3 or 10-3 dm 3 Liter (L) = dm 3 Pressure atm = 760 torr = Pa CONSTANTS

Chapter 17: Fundamentals of Spectrophotometry

CD Basis Set of Spectra that is used is that derived from comparing the spectra of globular proteins whose secondary structures are known from X-ray

What the Einstein Relations Tell Us

Adiabatic Expansion/Compression

3) In CE separation is based on what two properties of the solutes? (3 pts)

Chapter 17: Fundamentals of Spectrophotometry

Effects of Temperature and Concentration on the Rate of Photo-bleaching of Erythrosine in Water

FIRST PUBLIC EXAMINATION SUBJECT 3: PHYSICAL CHEMISTRY

Exponential Growth (Doubling Time)

LABORATORY OF ELEMENTARY BIOPHYSICS

WEEK 6. Multiphase systems

The Second Law of Thermodynamics (Chapter 4)

Introduction to Chemical Thermodynamics. (10 Lectures) Michaelmas Term

Supporting Information

Introduction to Chemical Thermodynamics. D. E. Manolopoulos First Year (13 Lectures) Michaelmas Term

Tutorial 1 (not important for 2015)

Luminescence basics. Slide # 1

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Electronic Supplementary Information

Supporting Information: Self-assembled benzophenone bis-urea macrocycles facilitate selective oxidations by singlet oxygen.

Singlet. Fluorescence Spectroscopy * LUMO

Ziessel a* Supporting Information (75 pages) Table of Contents. 1) General Methods S2

Room temperature phosphorescence vs thermally activated delayed fluorescence in carbazole pyrimidine cored compounds

Chemistry 102 Spring 2019 Discussion #4 Chapters 11 and 12 Student name TA name Section

Chem Homework Set Answers

Chapter 18. Fundamentals of Spectrophotometry. Properties of Light

CHEMISTRY Topic #2: Thermochemistry and Electrochemistry What Makes Reactions Go? Fall 2018 Dr. Susan Findlay See Exercises in Topic 8

Search for a monochromatic component of solar axions using Fe-57. Toshio Namba ICEPP, University of Tokyo

Questions on Instrumental Methods of Analysis

SUPPLEMENTARY INFORMATION

Chemistry 122 Wrap-Up Review Kundell

Ideal Gas Law. Deduced from Combination of Gas Relationships: PV = nrt. where R = universal gas constant

[ A] 2. [ A] 2 = 2k dt. [ A] o

Last Updated: April 22, 2012 at 7:49pm

Supplementary Figure 1. Potential energy, volume, and molecular distribution of the

Ch. 19 Entropy and Free Energy: Spontaneous Change

Chemical Thermodynamics. Chapter 18

Spiro-Configured Bifluorenes: Highly Efficient Emitter for UV Organic Light-Emitting Device and Host Material for Red Electrophosphorescence

Supporting Information

CHEMISTRY - BROWN 13E CH.7 - PERIODIC PROPERTIES OF THE ELEMENTS

INTRODUCTION Radiation differs from conduction and convection in that it does not require the presence of a material medium to take place.

Exam 4, Enthalpy and Gases

CHEM 305 Solutions for assignment #4

A 3 Vapor pressure of volatile liquids

Chemistry Chapter 16. Reaction Energy

1. The most important aspects of the quantum theory.

Isatin phenylhydrazones: Anion enhanced photochromic behaviour

Highly Luminescent -Conjugated Dithienometalloles: Photophysical Properties and Application to Organic Light-Emitting Diodes

Electronic Supporting Information for

Chem 105 Final Exam. Here is the summary of the total 225 points plus 10 bonus points. Carefully read the questions. Good luck!

Fundamentals of Mathematics (MATH 1510)

Version 001 Practice Final Exam 1

Intensity / a.u. 2 theta / deg. MAPbI 3. 1:1 MaPbI 3-x. Cl x 3:1. Supplementary figures

AP Chemistry. Free-Response Questions

Mechanisms of Visible Photoluminescence from Size-Controlled Silicon Nanoparticles

Chapter Eighteen. Thermodynamics

Energy Transfer Upconversion Processes

Supplementary Materials

Supporting Information for. Near infrared-to-blue photon upconversion by exploiting direct. S-T absorption of a molecular sensitizer

A. 24 B. 27 C. 30 D. 32 E. 33. A. It is impossible to tell from the information given. B. 294 mm C. 122 mm D. 10 mm E. 60 mm A. 1 H B. C. D. 19 F " E.

There are eight problems on the exam. Do all of the problems. Show your work

Electronic supplementary information

Lecture 7 Enthalpy. NC State University

A fluorescent ph probe for acidic organelle in living cells

Name: First three letters of last name

A long-lived iridium(iii) chemosensor for the real-time

UNIVERSITY OF SOUTHAMPTON

Molecular Luminescence. Absorption Instrumentation. UV absorption spectrum. lg ε. A = εbc. monochromator. light source. Rotating mirror (beam chopper)

Blending conjugated polymers without phase separation for fluorescent colour tuning of polymeric materials through FRET

Chem 325 NMR Intro. The Electromagnetic Spectrum. Physical properties, chemical properties, formulas Shedding real light on molecular structure:

Express the transition state equilibrium constant in terms of the partition functions of the transition state and the

Lecture 2. Review of Basic Concepts

Experiment 3: Radiative Lifetime Determination by Laser-Induced Fluorescence

Excited States in Organic Light-Emitting Diodes

Physics 213. Practice Final Exam Spring The next two questions pertain to the following situation:

Name: Class: Date: SHORT ANSWER Answer the following questions in the space provided.

Chemistry Instrumental Analysis Lecture 3. Chem 4631

Supporting Information for A potential-controlled switch on/off mechanism for selective excitation in mixed electrochemiluminescent systems

Neeraj K. Giri, Abhinandan Banerjee, Robert W. J. Scott, Matthew F. Paige,* and Ronald P. Steer* Supporting Information

Lecture 3: Light absorbance

Chapter 10 Liquids, Solids and Phase Changes

Enthalpies of Reaction

Saturation Absorption Spectroscopy of Rubidium Atom

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

'GEOi-C GIST Exam-^0> A-GSE-P-DIB Serial No. CHEMISTRY Paper II Time Allowed': Three Hours Maximum Marks : 200

Spectroscopy. Page 1 of 8 L.Pillay (2012)

Chapter 15 Molecular Luminescence Spectrometry

CHE 105 Fall 2016 Exam 3

Supplementary Figure S1 Definition of the wave vector components: Parallel and perpendicular wave vector of the exciton and of the emitted photons.

Excitation-Wavelength Dependent and Time-Resolved Photoluminescence Studies of Europium Doped GaN Grown by Interrupted Growth Epitaxy (IGE)

Thermodynamic condition for equilibrium between two phases a and b is G a = G b, so that during an equilibrium phase change, G ab = G a G b = 0.

where R = universal gas constant R = PV/nT R = atm L mol R = atm dm 3 mol 1 K 1 R = J mol 1 K 1 (SI unit)

Generation of light Light sources

Transcription:

Supporting Information Photophysical Properties of the Series fac- and mer-(-phenylisoquinolinato-n^c ) x (- phenylpyridinato-n^c ) 3-x Iridium(III) (x = to 3) Joseph C. Deaton *,, Ralph H. Young, Jerome R. Lenhard, Manju Rajeswaran, and Shouquan Huo *, Eastman Kodak Company, Rochester, NY, 4650; Department of Chemistry, East Carolina University, Greeneville, NC 7858 Work carried out at Eastman Kodak Company, Rochester, New York 4650, was done prior to December, 009 * To whom correspondence should be addressed. Email: jdeaton@rochester.rr.com (J.C.D); huos@ecu.edu (S. H.).

Synthesis Additional procedures for thermal isomerization of mer-ir(-piq) (ppy) in solid state. Method C: A sample of mer-ir(-piq) (ppy) (0 mg, mer/fac ratio >99:) was placed in a small ampoule. The ampoule was sealed under high vacuum and placed in an oven. The temperature of the oven was set to 300 C. After 3 h, the ampoule was taken out and cooled to room temperature, the mer/fac ratio of the sample was determined by HPLC to be 4:96 (area). Method D: A sample of mer-ir(-piq) (ppy) (50 mg, mer/fac ratio >99:, HPLC area) was placed in a opened small vial. The vial was placed into a heating apparatus with an inlet connected with a dry argon line and with an outlet connected to a bubbler. The heating apparatus was purged for 30 min before heating then maintained with a slow argon flow during heating. The material was heated at 90 C for 48 h. After cooling to room temperature, the vial was taken out and the mer/fac ratio of the sample was determined by HPLC to be 7:93 (HPLC area). Sublimation. The vapor pressures of fac-(piq) 3 and fac-ir(piq) (ppy) at various temperatures were measured by the Knudsen effusion technique (Knudsen, M. Ann. Phys. 909, 8, 999 06) and are shown in Figure S. The enthalpy and entropy of sublimation for the median temperature of measurement, T m, were calculated from a linear fit to the experimental data using the integrated form of the Clausius-Clayperon equation, eq SI- (P 0 = atm, R = gas constant). The values are listed in Table S, along with those of fac-ir(ppy) 3 for reference.

3 ln( P / P0 ) = S / R H RT (SI-) sub m sub m / Table SI-. Enthalpy and Entropy of Sublimation for fac-ir(ppy) 3, fac-ir(piq) 3 and fac- Ir(piq) ( ppy) at median temperatures of measurement. Compound sub H m (kj mol ) sub S m (J K mol ) T m (K) fac-ir(ppy) 3 75.8 ±.8 9. ± 3. 478 fac-ir(piq) 3 03.9 ± 0.6 38.6 ±. 568 fac-ir(piq) (ppy) 0.4 ±.3 58.3 ±.3 543 Interestingly, the data in Figure SI- and Table SI- reveal that the greater volatility of fac- Ir(piq) (ppy) relative to that of fac-ir(piq) 3 arises from an increased entropy of sublimation, not from any significant difference in enthalpy of sublimation. The vapor pressure of fac- Ir(piq)(ppy) was not measured, but it was found in vapor deposition of OLEDs (vide infra) that the temperature required for deposition at the same rate followed in the order fac-(piq) 3 > fac-ir(piq) (ppy) > fac-ir(piq)(ppy). Figure SI-. Vapor pressures determined by Knudsen effusion technique for fac-ir(piq) 3 and fac-ir(piq) (ppy).

4 Photophysical Properties Figure SI-. Absorption spectra of the series fac-ir(piq) x (ppy) 3-x (x = to 3) in -MeTHF at 95 K. Figure SI-3. Phosphorescent decays of fac-ir(piq)(ppy) in PMMA (0.05 wt %) at selected temperatures. Emission detected at 60 nm after pulsed excitation at 485 nm.

5 Fig SI-4. Absorption of the series mer-ir(piq) x (ppy) 3-x in -MeTHF at 95 K. Fig SI-5. Absorption comparison on an expanded scale of mer- and fac-ir(piq)(ppy) in - MeTHF at 95 K.

6 Figure SI-6. Phosphorescent decays of mer-ir(piq)(ppy) in PMMA ( wt %) at selected temperatures. Emission detected at 580 nm after pulsed excitation at 485 nm. Correction of the spectra and quantum yields for fac-ir(piq)(ppy). The spectra and quantum yields reported in the text for fac-ir(piq)(ppy) represent a sample containing a 5% impurity of fac-ir(piq) (ppy). As shown below, the effects of the impurity are quite small. Quantities relating to the pure compounds, fac-ir(piq)(ppy) and fac-ir(piq) (ppy) are indicated by a subscript or ; quantities relating to the impure fac-ir(piq)(ppy) are indicated by a subscript i. The concentrations were in the ratio c /c = 5/95. A straightforward calculation leads to eq SI-. Thus, the correction to the absorption spectrum of fac-ir(piq)(ppy)

7 ε ε 055 ( ε ) i + 0. i ε (SI-) amounts to 5.5% of the difference between the spectra of the impure fac-ir(piq)(ppy) and the pure fac-ir(piq) (ppy). At room temperature and λ ex = 355 nm, the extinction coefficient of fac-ir(piq) (ppy) is almost equal to that of fac-ir(piq)(ppy) (Figures and SI-). A straightforward calculation starting from the quantum yields in Table 4 shows that the correction to the quantum yield of fac-ir(piq) (ppy) is less than 0.00 in -MeTHF. Since the quantum yields of Ir(piq) (ppy) and the impure fac-ir(piq)(ppy) are equal in PMMA (Table 5), the quantum yield of pure Ir(piq) (ppy) is the same. The emission spectra of fac-ir(piq)(ppy) and fac-ir(piq) (ppy) shown in Fig 3 are quantitatively so similar that an appropriately weighted subtraction of one from the other cannot have an appreciable effect. The emission spectra at low temperature, Figs. 4 and 5, are more different, especially in the blue edges, so it is perhaps worthwhile to consider the correction for the impurity. Let s (λ), s (λ), and s i (λ) be the emission spectra as functions of wavelength, normalized to as in Figs 4 and 5. The spectrum of pure fac-ir(piq)(ppy) can be evaluated using eq SI-3a, where the coefficient X is given by eq SI-3b. s ( s ) = s + X (SI-3a) i i s X c = c ε ε Φ Φ A A (SI-3b) The quantities A and A are the areas under the curves in the normalized spectra of Fig. 5. Their ratio is approximately at each temperature. The ratio of quantum yields is also

8 approximately at each temperature. The ratio of extinction coefficients pertains to the wavelength of excitation, 458 nm, where ε /ε =.3. The concentration ratio is 5/95. Therefore, X 0.07, and the normalized emission spectrum of pure fac-ir(piq)(ppy) is given by eq SI-3c. ( s ) s si + 0. 07 i s (SI-3c) The correction amounts to about 7% of the difference between the spectra of the impure fac- Ir(piq)(ppy) and the pure fac-ir(piq) (ppy). The main effect is to increase the intensity of the blue edge by several percent. Correction of the zero-field parameters for fac-ir(piq)(ppy). The zero-field parameters reported for fac-ir(piq)(ppy) in Table 6 were evaluated without taking the 5% fac- Ir(piq) (ppy) impurity into account. The decay constants k obs were evaluated from the initial slopes of the photoluminescence decay curves on semilogarithmic plots. The corrected values cited in the text were obtained by fitting the data to a formula for k obs that includes the contributions of the impurity to the observed decays. The derivation of this formula is given in the following. For the Ir(piq)(ppy) and Ir(piq) (ppy) components of the impure sample, let k and k be the respective decay constants, c and c be the molar concentrations, ε and ε be the extinction coefficients, Φ and Φ be the quantum yields, and f and f be the fractions of emitted photons that are counted by the detector. The initial concentrations of excited states for the two components of the impure sample are Aε c and Aε c, respectively, where A is a proportionality constant. The subsequent concentrations are represented by decaying

9 exponentials, Aε c exp(-k t) and Aε c exp(-k t). The radiative rate constants are k Φ and k Φ, and the numbers of photons emitted per second are k Φ Aε c exp(-k t) and k Φ Aε c exp(-k t). The numbers of photons detected per second are f k Φ Aε c exp(-k t) and f k Φ Aε c exp(-k t). The total counting rate, I(t), is the sum of the two individual rates, eq SI-4. ( t) = A[ f k Φ c exp( k t) + f k Φ ε c ( k t) ] I exp ε (SI-4) Evaluating d ln I( t) dt at t = 0 gives an expression for the observed decay rate of the impure mixture as an initial slope, eq SI-5a. The temperature-dependent values of k, the decay constant for the impurity, are known from experiment. The coefficient B is given in terms of other experimental quantities by eq SI-5b. k + B k k obs = k (SI-5a) k + B k c B= c ε ε f f Φ Φ (SI-5b) A Boltzmann formula, eq SI-5c, expresses k in terms of the temperature and the zero-field splittings and decay rates of the triplet sublevels for pure fac-ir(piq)(ppy). k I + k II exp( EII / kt ) + k III exp( EIII / kt ) k = (SI-5c) + exp( E / kt ) + exp( E / kt ) II III

0 Combining eqs SI-5a to SI-5c provides a theoretical formula for the initial slope of the photoluminescence decay of the impure sample of fac-ir(piq)(ppy). The ratio c /c was 5/95 from the HPLC analysis. At 485 nm and room temperature, the ratio ε /ε is.40 (see Figs. and SI-). Because the absorption spectra are not highly structured, the ratio ε /ε was assumed to be independent of temperature. Because the emission spectra are similar, the ratio f /f is approximately at each temperature (see text and Figs. 3 and 5). At room temperature, the pure fac-ir(piq) (ppy) and the impure fac- Ir(piq)(ppy) have equal quantum yields in PMMA, so the pure fac-ir(piq)(ppy) must have the same value, and Φ /Φ =. The temperature dependence of B is determined by the temperature dependence of Φ /Φ. Individually, Φ and Φ are proportional to the integrated intensities of the spectra, and the temperature dependences of the observed intensities were used to determine the ratio Φ /Φ as a function of temperature. The zero-field splittings and decay rates of the triplet sublevels for pure fac- Ir(piq)(ppy) were evaluated from the experimental data by fitting the theoretical expression for the initial slopes, eq SI-5a to SI-5c, to the observed values for the sample of fac- Ir(piq)(ppy) containing 5% of fac-ir(piq) (ppy). A least-squares procedure was used, assuming equal relative uncertainties for the data points. The best-fit curve was similar to that shown in Fig. 6, with a slightly better fit (smaller r.m.s. deviation from the data). Although eq SI-4 is formally biexponential, the decays I(t) evaluated with the corrected parameters appeared approximately single-exponential because the second term in eq SI-4 is relatively small. The corrected parameters for pure fac-ir(piq)(ppy) are shown in

comparison to the uncorrected values in Table SI- below, and are also reported in the text. The corrections are not significantly larger than the standard deviations of the parameters. Table SI-. Zero-field splittings and decay rate constants for triplet sublevels of fac- Ir(piq)(ppy) without and with correction for the effect of the 5% fac-ir(piq) (ppy) impurity. E II (σ) E III (σ) k I (σ) k II (σ) k III (σ) (cm ) (cm ) (0 4 s ) (0 5 s ) (0 6 s ) uncorrected 4.5(.9) 00.6(5.9).58(0.08).4(0.6).6(0.7) corrected 5.6(3.) 06.6(7.4).56(0.07) 0.95(0.).78(0.0)

Figure SI-7 H NMR spectrum of fac-ir(piq) 3 (in CH Cl ).

Figure SI-8 H NMR spectrum of fac-ir(piq) (ppy) (in CH Cl ). 3

Figure SI-9. H NMR spectrum of fac-ir(piq)(ppy) (in CH Cl ). 4