High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Similar documents
Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

LC-MS/MS Analysis of Phytocannabinoids and their

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Detection of Mycotoxins in Corn Meal Extract Using Automated Online Sample Preparation with Liquid Chromatography-Tandem Mass Spectrometry

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Determination of urea in ultrapure water by IC-MS/MS

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

The key to your laboratory s success

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Screening Method for 30 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Quantitative analysis of mitragynine in human urine by high performance liquid chromatography-tandem mass spectrometry

isobaric Barbiturates in Serum using LDTD- MS/MS combined with differential mobility

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Determination of trace anions in concentrated hydrofluoric acid

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Automated and accurate component detection using reference mass spectra

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Improved Automated Sample Preparation for the Analysis of 25-OH-Vitamin D3 by LC/MS/MS

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Fast methods for the determination of ibuprofen in drug products

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Tecan AC Extraction Plate

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Supported Liquid Extraction (SLE) Guide and FAQ s. User Guide

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

A S ENSIT IV E M E T HO D FO R T H E D E T E RM INAT IO N O F ENDO C RIN E- D IS RU P T ING COM P OUNDS IN RIV E R WAT E R BY L C / MS/MS

High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research

Chemical derivatization

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

LC-MS/MS in the Clinical Laboratory. Jo Adaway

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Patient samples Use of patient samples was according to approved guidelines of the University of Washington Human Subjects Division.

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Definitive EtG/EtS LC-MS/MS Analysis:

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

Psychoactive Drugs Urine LC-MS/MS Analysis Kit User Manual

Direct Analysis using Paper-Spray Mass Spectrometry: Method Development for the Rapid Screening of Drugs of Abuse for Forensic Toxicology

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Expanded capabilities in ammonium and amine detection

Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

A New Polymer-based SPE Sorbent to Reduce Matrix Effects in Bioanalytical LC-MS/MS

Enhancement of Linearity and Response in Charged Aerosol Detection

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

An Evaluation of Various High-Resolution, Accurate-Mass Scan Modes for In Vitro Drug Discovery Screening

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Abstract. Experimental Sample Preparation

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Supporting Information. Detection and Occurrence of Chlorinated By-products of Bisphenol A, Nonylphenol and

Supporting Information. Sweetened Swimming Pools and Hot Tubs

Relative quantification using TMT11plex on a modified Q Exactive HF mass spectrometer

A High Sensitivity Dual Solid Phase Extraction LC-MS/MS Assay for the Determination of the Therapeutic Peptide Desmopressin in Human Plasma

Analysis of Plasticizer Contaminants in Beverages and Milk using an Automated System Based on Turbulent-flow Chromatography Coupled to LC-MS/MS

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC

Thermo Scientific Pesticide Explorer Collection. Start-to-finish. workflows for pesticide analysis

Appendix II- Bioanalytical Method Development and Validation

Transcription:

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes Joe DiBussolo, Marta Kozak Thermo Fisher Scientific, San Jose, CA Application Note 652 Key Words Steroid, estrogens, estrone, estradiol, E1, E2, Transcend, TSQ Endura Goal To develop and evaluate a reliable and robust HPLC-MS/MS method for quantifying estrone and estradiol in human blood plasma/serum for clinical research purposes. Application Benefits Simple, economical extraction/derivatization procedure Reliable quantitation from 5 to 500 pg/ml with ion-ratio confirmation Throughput capabilities of 10, 20, or 40 injections per hour from 1-, 2-, or 4-channel HPLC systems Introduction Estrone (also known as E1) and estradiol (17-β-estradiol or E2) are two of several estrogens that are steroid hormones involved in the development and function of female anatomical and physiological characteristics and processes such as the menstrual cycle. Researchers studying the effects of E1 and E2 on such things need to quantify them within an analytical range of 5 to 500 pg/ml (18.5 to 1,850 pmol/l) in blood plasma or serum. E1 and E2 form negative ions by deprotonation in both electrospray ionization (ESI) and atmospheric-pressure chemical ionization (APCI) sources of mass spectrometers with low efficiency. In order to robustly achieve the needed quantitation limits, most researchers use dansyl chloride to form positively charged derivatives of these and other estrogens. 1,2 s Preparation Aliquots of fresh blood serum specimens (200 µl), as well as calibrators prepared in 1 BSA and quality control specimens (QCs), were spiked with estradiol-d 5 internal standard (IS) before being subjected to liquid-liquid extraction with methyl t-butyl ether (MTBE). After drying the ether extracts with heated nitrogen, dansyl chloride reagent was added to the residue of each to form positiveion dansylated derivatives. The preparations were diluted with water and acetonitrile (1:1) and then analyzed by reversed-phase liquid chromatography coupled to tandem mass spectrometry with a heated electrospray ionization (HESI) probe. Liquid Chromatography Using one or more channels of a Thermo Scientific Transcend LX-4 multichannel HPLC, chromatographic separation of the dansylated steroid analytes from unwanted sample components was achieved by gradient elution through a Thermo Scientific Accucore RP-MS column (2.6 µm, 50 x 2.1 mm, P/N 17626-052130) heated to 40 C. The chromatographic conditions are described in Figure 1. This method was multichanneled with other methods that utilized the same MS ion source at the same temperatures and gas flows. Column: Accucore RP-MS, 2.6 µ, 50 x 2.1 mm at 40 C Solvent A: Water + 0.1 Formic Acid Solvent B: Methanol Step Start Sec Flow Gradient A B 1 0.00 10 0.5 Step 50 50 2 0.17 20 0.5 Ramp 25 75 3 0.50 30 0.5 Step 25 75 4 1.00 90 0.5 Ramp 20 80 5 2.50 30 0.5 Ramp - 100 6 3.00 60 0.5 Step - 100 7 4.00 90 0.5 Step 50 50 Start data: 2.0 min Duration: 2.1 min Total run time: 5.5 min Figure 1. Chromatographic parameters. Mass Spectrometry and Data Analysis A Thermo Scientific TSQ Endura triple quadrupole mass spectrometer was used with a HESI probe. Dansylated E1, E2, and deuterated E2, which served as the IS for both analytes, were measured by selected reaction monitoring (SRM), as described in Figure 2. Data were acquired and processed using Thermo Scientific TraceFinder software.

2 Ion Source: HESI, + 3500 V, vaporizer temp: 400 C SRM Transitions: Q1 & Q3 resolutions: 0.4 & 0.7, respectively Analyte Q1 Q2 CE RF Dansyl-E1 (Confirm) 504.30 156.20 50 160 Dansyl-E1 (Quan) 504.30 171.25 33 160 Dansyl-E2 (Confirm) 506.30 156.20 50 160 Dansyl-E2 (Quan) 506.30 171.25 33 160 Dansyl-IS (Confirm) 511.35 156.20 50 160 Dansyl-IS (Quan) 511.35 171.25 33 160 Figure 2. TSQ Endura parameters. Evaluation precision was assessed by calculating percent coefficient of variation (CV) of peak areas from 20 replicate injections of two pools of test specimens. Carryover was measured in blanks immediately following injections of the highest calibrator. Matrix effects were evaluated by comparing IS peak areas in specimen samples to IS peak areas in calibrator and control samples. A method comparison experiment was performed by analyzing 40 donor samples (no informed consent was needed) following the procedures described in this study and comparing results with those obtained by a reference lab. Results and Discussion As shown in Figures 3 and 4, the desired analytical range from 5 to 500 pg/ml (18.5 to 1,850 pmol/l) was achieved for both E1 and E2. The ranges were consistently linear (r 2 0.995 with 1/X weighting). Carryover never exceeded 0.5. Confirmation/quantitation ion ratios among calibrators, QCs, and specimens were within 20 of averages calculated from the calibrators. Ion ratios for dansylated E1, E2, and E2-D 5 averaged 35, 32, and 35, respectively. Typical characteristics of chromatographic peaks from quantitation- and confirmation-ion transitions for E1 and E2 from a donor specimen are shown in Figure 5. For both E1 and E2, the intra- and inter-batch precisions were better than 6 and 7, respectively. IS peak areas among calibrators and QCs averaged 143,000 cps with an RSD of 11. Specimen IS peak areas ranged from 46,000 to 72,000 cps with an average recovery of 37, relative to the averaged IS peak areas in calibrators and QCs. However, the IS in each sample successfully compensated for matrix effects as proved by method comparison results. In the method comparison experiments, values of analyzed samples ranged from 16 to 156 pg/ml for E1 (Table 1) and from 11 to 356 pg/ml for E2 (Table 2). The percent difference between the two analytical methods for 95 of analyzed samples was 20 or less. Calibrator 5 pg/ml -0.56 20 pg/ml -4.93 50 pg/ml 0.27 100 pg/ml 7.59 200 pg/ml -1.71 500 pg/ml -0.66 1/X weighting r 2 = 0.9991 Figure 3. Typical E1 quantitation results.

3 Calibrator 5 pg/ml 10.70 20 pg/ml 1.91 50 pg/ml -5.74 100 pg/ml -5.81 200 pg/ml -4.36 500 pg/ml 3.30 1/X weighting r 2 = 0.9980 Figure 4. Typical E2 quantitation results. Estrone (E1) Dansylated Estradiol (E2) Dansylated Figure 5. Analyte chromatograms from donor specimen containing 10 pg/ml E1 & E2.

4 Table 1. E1 results comparison with reference lab. 1 19 20-6.1 2 34 39-12.9 3 19 21-7.7 4 130 165-21.5 5 24 28-13.4 6 39 45-15.0 7 30 33-10.5 8 91 112-18.6 9 36 34 3.2 10 139 138 0.7 11 19 23-18.2 12 109 108 1.0 13 36 42-14.8 14 19 18 3.8 15 30 29 2.4 16 47 53-10.8 17 91 98-7.7 18 18 16 14.8 19 123 149-17.5 20 39 50-21.8 21 33 40-18.9 22 26 29-7.7 23 22 25-12.9 24 20 21-5.2 25 80 91-12.3 26 60 63-4.6 27 23 20 16.9 28 55 50 9.7 29 51 54-6.8 30 83 83 0.2 31 40 38 4.5 32 19 19 2.1 33 83 83-0.2 34 29 27 9.4 35 60 57 4.6 36 18 17 7.6 37 47 48-1.5 38 155 156-0.7 39 23 19 19.5 40 43 35 21.8 Table 2. E2 results comparison with reference lab. 1 32 33-2.7 2 40 38 5.3 3 12 11 6.4 4 297 283 4.9 5 16 14 16.1 6 104 90 15.3 7 100 83 20.1 8 176 151 16.6 9 164 138 18.6 10 301 251 19.9 11 17 19-10.2 12 220 190 15.6 13 20 25-19.4 14 24 24-0.4 15 71 76-7.7 16 79 71 10.8 17 112 117-4.6 18 14 17-13.9 19 36 40-11.9 20 72 66 10.0 21 23 18 24.3 22 16 20-19.0 23 25 32-22.1 24 12 11 9.5 25 206 243-15.4 26 71 81-12.4 27 21 18 19.4 28 117 102 14.9 29 180 157 14.9 30 157 138 13.7 31 31 26 17.1 32 18 19-2.1 33 198 185 6.8 34 108 102 6.0 35 76 72 5.4 36 41 40 1.8 37 107 113-5.2 38 325 356-8.7 39 20 17 17.3 40 58 54 7.5

Conclusion Sensitive, robust quantitation assays of estrone (E1) and estradiol (E2) in blood plasma/serum are described in this application note. Some of the key features of this research method are: Sensitive quantitation from 5 to 500 ng/dl (18.5 to 1,850 pmol/l) Good reliability with inter- and intra-batch precisions less than 8 and carryover less than 0.5 High-throughput capabilities and multichanneling with other HESI methods Acknowledgement We thank Dr. Hashim Othman and his staff of Bio- oratories (Elmwood Park, NJ) for providing the donor specimens and resources to evaluate this research method. Application Note 652 s 1. Nelson, R.E. et al. Liquid chromatography tandem mass spectrometry assay for simultaneous measurement of estradiol and estrone in human plasma. Clin Chem, 2004, 50, 373-384. 2. Kushnir, M.M. et al. High-sensitivity tandem mass spectrometry assay for serum estrone and estradiol. Am J Clin Pathol, 2008, 129, 530-539. For research use only. Not for use in diagnostic procedures. To find a local representative, visit: thermofisher.com 2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. AN64659-EN 0616S